Thin Layer Coulometry with Ionophore Based Ion-Selective Membranes

Size: px
Start display at page:

Download "Thin Layer Coulometry with Ionophore Based Ion-Selective Membranes"

Transcription

1 Anal. Chem. 2010, 82, Thin Layer Coulometry with Ionophore Based Ion-Selective Membranes Ewa Grygolowicz-Pawlak and Eric Bakker* Department of Chemistry, Nanochemistry Research Institute, Curtin University of Technology, Perth, WA 6845, Australia Downloaded via on March 25, 2019 at 20:33:38 (UTC). See for options on how to legitimately share published articles. We are demonstrating here for the first time a thin layer coulometric detection mode for ionophore based liquid ion-selective membranes. Coulometry promises to achieve the design of robust, calibration free sensors that are especially attractive for applications where recalibration in situ is difficult or undesirable. This readout principle is here achieved with porous polypropylene tubing doped with the membrane material and which contains a chlorinated silver wire in the inner compartment, together with the fluidically delivered sample solution. The membrane material consists of the lipophilic plasticizer dodecyl 2-nitrophenyl ether, the lipophilic electrolyte ETH 500, and the calcium ionophore ETH Importantly and in contrast to earlier work on voltammetric liquid membrane electrodes, the membrane also contains a cationexchanger salt, KTFPB. This renders the membrane permselective and allows one to observe open circuit potentiometric responses for the device, which is confirmed to follow the expected Nernstian equation. Moreover, as the same cationic species is now potential determining at both interfaces of the membrane, it is possible to use rapidly diffusing and/or thin membrane systems where transport processes at the inner and outer interface of the membrane do not perturb each other or the overall composition of the membrane. The tubing is immersed in an electrolyte solution where the counter and working electrode are placed, and the potentials are applied relative to the measured open circuit potentials. Exhaustive current decays are observed in the range of 10 to 100 µm calcium chloride. The observed charge, calculated as integrated currents, is linearly dependent on concentration and forms the basis for the coulometric readout of ion-selective membrane electrodes. Modern directions in chemical sensing demand the realization of self-contained, robust sensing principles that do not require frequent supervision or recalibration in the field. Environmental remote chemical sensing applications with a large network of inexpensive chemical sensors, for instance, would have tremendous promise in monitoring programs. Current technologies are often restricted to passive, nonselective optical systems. 1 While state of the art in electrochemical sensing technology may offer adequate * To whom correspondence should be addressed. (1) Miller, R. L.; Del Castillo, C. E.; McKee, B. A., Eds. Remote Sensing of Coastal Aquatic Environments. Technologies, Techniques and Applications; Springer: Secaucus, NJ, selectivity and sensitivity in many cases, 2,3 it still utilizes traditional transduction principles, including potentiometry and amperometry. These have never been intended for recalibration free and robust sensing approaches. In the clinical laboratory setting, for instance, where sensors based on such principles are employed in vast numbers, a one point recalibration step is typically performed with each sample reading to compensate for any drift in the sensor output signal. 4 Clearly, the transduction principles of electrochemical sensors must be reconsidered from fundamental principles to achieve the ambitious goals mentioned above. 5 Absolute measurement principles in the analytical sciences are rare, but they do include gravimetry and coulometry. In controlled potential coulometry, the current associated with a specific redox reaction is monitored until it decays to near-zero values. 6 The observed current integrated over the time period gives the overall charge, which ideally relates to the number of moles of material converted in the sample using Faraday s law. The use of thin layer samples, as in thin layer coulometry, improves the speed of analysis relative to the use of voluminous samples. 7 The key challenge in employing this readout principle for chemical sensors has been that redox reactions at metal electrodes tend not to be sufficiently selective. Recently, however, a coulometric readout was successfully used in enzymatic glucose bioassays, for example, allowing one to exploit the confined volume for the exhaustive electrolysis of potential interferences before the actual glucose assay. 8 The electrochemical detection of ionic species is preferably accomplished with ion-selective electrodes. Polymeric or liquid membrane systems are now particularly well developed, and their sensing selectivities are a function of membrane extraction and complexation processes. 9,10 These membranes are traditionally read out by zero current potentiometry, often giving extremely low detection limits. 11 Moreover, the use of voltammetric methods on membrane electrodes has increased in recent years, including (2) Lee, H. J.; Yoon, I. J.; Yoo, C. L.; Pyun, H.-J.; Cha, G. S.; Nam, H. Anal. Chem. 2000, 72, (3) Wang, J.; Tian, B. Anal. Chem. 1992, 64, (4) Jacobs, E.; Ancy, J. J.; Smith, M. Point Care: J. Near-Patient Test. Technol. 2002, 1, (5) Bakker, E.; Bhakthavatsalam, V.; Gemene, K. L. Talanta 2008, 75, (6) Bard, A. J. Anal. Chem. 1966, 38, 88R 98R. (7) Oglesby, D. M.; Anderson, L. B.; McDuffie, B.; Reilley, C. N. Anal. Chem. 1965, 37, (8) Tsujimura, S.; Kojima, S.; Ikeda, T.; Kano, K. Anal. Bioanal. Chem. 2006, 386, (9) Buhlmann, P.; Pretsch, E.; Bakker, E. Chem. Rev. 1998, 98, (10) Bakker, E.; Buhlmann, P.; Pretsch, E. Chem. Rev. 1997, 97, (11) Bakker, E.; Pretsch, E. Angew. Chem., Int. Ed. 2007, 46, /ac100524z 2010 American Chemical Society Analytical Chemistry, Vol. 82, No. 11, June 1, Published on Web 04/29/2010

2 cyclic voltammetry and amperometry, 12 stripping voltammetry, 13 normal pulse voltammetry, 14 and pulsed chronopotentiometry. 15,16 When exploring these as sensor readout principles, the membrane was normally chosen not to exhibit ion-exchanger properties in order to achieve a sample-membrane interface that can be effectively polarized by the electrochemical perturbation step. The groups of Sanchez-Pedreno 17 and Osakai 18 were reportedly the first to use ion transfer voltammetry principles in a flow system to demonstrate the coulometric detection of very lipophilic ions. These two early works did not yet utilize chemical receptors (ionophores), and the coulometric efficiency was not yet optimal. Kihara and co-workers offered the first description of small ion detection with ionophores with this approach, using a porous Teflon tubing that was wholly immersed into an organic solvent that contained a lipophilic electrolyte and a chemical receptor. 19 Kihara and co-workers indeed showed that the absolute detection of ionic species is possible with this approach. Despite its elegance, this approach does not yet form the basis for a convincing practical device because of the required large volume of volatile organic solvent and high quantities of specialty chemicals. We believe that a membrane based sensing approach with modern ion-selective electrode materials is required to achieve this goal. We show here that a liquid membrane based coulometric ion detector can be achieved by utilizing membrane components typically used in polymeric ion-selective electrodes and with membranes that exhibit ion-exchanger properties. A common ion that is potential determining at both interfaces greatly simplifies the theoretical understanding of these systems relative to earlier cell designs, where the cation extraction at one interface was typically electrochemically coupled to an anion extraction step at the other side. 20 With thin membranes or membranes that have high ion mobilities, as desired in coulometry, this would be problematic because these two ions may eventually meet and result in significant water uptake with associated swelling and opacity increase of the membrane. 21 As an important other benefit, the approach introduced here also allows one to monitor the open circuit potential of the cell and to use this information to apply the correct potential for the coulometric readout step. EXPERIMENTAL SECTION Reagents, Materials, and Equipment. Dodecyl 2-nitrophenyl ether (DDNPE), tetradodecylammonium tetrakis(4-chlorophenyl)borate (ETH 500), potassium tetrakis[3,4-bis(trifluoromethyl)phenyl]-borate (KTFPB), calcium ionophore IV (ETH 5234), high molecular weight poly(vinyl chloride) (PVC), tetrahydrofuran (THF), calcium chloride dihydrate, and potassium chloride were purchased in the highest quality available from Fluka. Aqueous solutions were prepared by dissolving the appropriate salts in Milli- Q-purified distilled water. (12) Samec, Z.; Samcova, E.; Girault, H. H. Talanta 2004, 63, (13) Kim, Y.; Amemiya, S. Anal. Chem. 2008, 80, (14) Jadhav, S.; Meir, A. J.; Bakker, E. Electroanalysis 2000, 12, (15) Shvarev, A.; Bakker, E. Anal. Chem. 2003, 75, (16) Gemene, K.; Bakker, E. Anal. Chem. 2008, 80, (17) Sanchez-Pedreno, C.; Ortuno, J. A.; Hernandez, J. Anal. Chim. Acta 2002, 459, (18) Sawada, S.; Taguma, M.; Kimoto, T.; Hotta, H.; Osakai, T. Anal. Chem. 2002, 74, (19) Yoshizumi, A.; Uehara, A.; Kasuno, M.; Kitatsuhi, Y.; Yoshida, Z.; Kihara, S. J. Electroanal. Chem. 2005, 581, (20) Jadhav, S.; Bakker, E. Anal. Chem. 2001, 73, (21) Jadhav, S.; Bakker, E. Anal. Chem. 1999, 71, The polypropylene hollow fiber Accurel PP Q3/2 was supplied by Membrana GmbH (Wuppertal, Germany), and Sterling Silver Round wires of 250 and 500 µm i.d. were supplied by AEMETAL (Sydney, Australia). Unless otherwise specificied, all measurements were performed at room temperature vs a double-junction Ag/AgCl/sat. KCl/1 M LiOAc reference electrode (Mettler-Toledo AG, Schwerzenbach, Switzerland) placed in the outer aqueous solution. The ion-selective hollow fiber cell served as the working electrode, and a high surface area Pt electrode (MT Electrodes, Perth, Australia) served as the counter electrode in coulometric measurements. The potentiometric calibration was performed using a 16-channel EMF monitor (Lawson Laboratories, Inc., Malvern, PA) connected to a personal computer. The coulometric measurements were performed with an Autolab PGSTAT128N (Metrohm Autolab, Utrecht, The Netherlands) attached to a personal computer. Calibrations were performed by feeding the cell with appropriate solutions using an ISMATEC peristaltic pump (Glattbrugg, Switzerland) with a flow rate of 25 µl min -1. Preparation of the Coulometric System. Silver wires of 15 cm in length and 250 (for the twisted configuation) or 500 µm diameter were cleaned with acetone and water before chlorination. A 9 cm section of the cleaned wire was dipped in a1mhcl solution. A potential of 0.5 V against double junction reference electrode was applied between the silver wire and the high surface area platinum counter electrode. The process was carried out for 2 min. The silver/silver chloride wire was rinsed thoroughly with distilled water after chlorination. The polypropylene porous tube of 600 µm inner diameter, 200 µm wall thickness, and 7 cm length was washed inside and out with distilled water and THF. The chlorinated part of the wire was introduced into the polypropylene tube to its entire length. Two polypropylene pipet tips were used as fluidic connections for the flow-through system. The bare, nonchlorinated part of the silver wire was directed through the outlet tip to allow for the electrical connection. After assembly, the porous fiber was doped with 200 µl of the membrane cocktail composed of mg of DDNPE, 30 mg (10 wt %) of ETH 500, 2.4 mg (10 mmol kg -1 )of Ca-ionophore IV, 0.9 mg of KTFPB (30 mol % relative to the ionophore), and 1.5 ml of THF. The obtained ion-selective membrane was left to dry overnight, and any excess of the liquid membrane was wiped off with tissue paper. A PVC slurry was used to seal the inlet and outlet of the cell. The resulting inner active surface area of the membrane was about 113 mm 2, as estimated from the nominal dimensions. A silicone tube was fixed to the inlet of the cell and connected to the peristaltic pump. The inner side of the cell was washed for a day by continuously flowing 1 mm CaCl 2 solution at the flow rate mentioned above. Procedures. A potentiometric calibration was performed using a silver/silver chloride wire immersed in the outer solution of a fixed composition (10-3 M CaCl M KCl) connected to the potentiometer as indicator electrode while the silver/silver chloride inside the hollow fiber membrane was connected to the reference port. Sample solutions of a fixed 10-5,5 10-4, 10-4,5 10-3, and 10-3 M CaCl 2 concentration and with 10-2 M KCl as background electrolyte were continuously passed through the polypropylene tube, and the associated potential 4538 Analytical Chemistry, Vol. 82, No. 11, June 1, 2010

3 changes in the cell were recorded. The activities of calcium ion and the liquid junction potential were corrected with the Debye-Hückel approximation and the Henderson equation. All other experiments were performed in a 3-electrode system with the silver/silver chloride wire inside the hollow fiber, which now acted as the working electrode. The same outer solution as above was used (10-3 M CaCl M KCl). The peristaltic pump was always off during the following measurements. For the current interruption test, four alternative +1 µa and -1 µa 3 s long pulses were applied with 30 s regeneration pulses at 0 A current following and preceding each pulse. Normal pulse voltammetry was performed within a range of 250 mv starting from the open circuit potential (OCP) and going toward more positive values. Potential pulses (1 s) in 10 mv potential increment were followed by 5 s OCP regeneration pulses (baseline potential). For the coulometric analysis procedure, three pulses were applied. An initial 20 s pulse at the OCP confirmed the stability of the system, followed by a 30 s OCP + n 25 mv calcium extracting pulse (n ranging from 1 to 10). A third 60 s pulse at OCP was used to regenerate the system before a new portion of sample was delivered to the active electrode area. RESULTS AND DISCUSSION In order to maximize the ion mobility of the membrane, porous polypropylene tubing was doped with a membrane containing plasticizer without the use of traditional poly(vinyl chloride) (PVC). Similar hollow fiber membrane materials were introduced by the group of Buffle as permeation membrane sampling systems. 22 Preliminary experiments on similarly formulated 20 µm thick Celgard membranes showed that membranes without any PVC did not exhibit a long operational lifetime with the traditional plasticizer bis(2-ethyl hexyl sebacate) (DOS). This might be due to nonoptimal casting conditions, using the solvent THF in the membrane cocktail slurry. Earlier work on doping 20 µm thin Celgard membranes (based on porous polypropylene as well) with DOS yielded functional membranes for a period of 1 to 2 days, 23 which is expected to improve with the much thicker tubular membrane used here (200 µm). To eliminate potential issues with limited lifetime, the plasticizer dodecyl 2-nitrophenyl ether, a more lipophilic analog of o-npoe, was used. This solvent has been used in ion-selective electrode applications for over 30 years and exhibits an estimated octanol-water partition coefficient of log P ) 7.4, 2 orders of magnitude larger than that of o-npoe. 24 It was found earlier in optical bead based sensors to exhibit desirable lipophilicity characteristics as well, without apparent deteriorating influence on ion selectivity. 25 Note that polar (yet lipophilic) plasticizers are known to result in membranes with improved selectivity characteristics for divalent ions such as calcium 26 and, hence, was preferred here. The membranes also contained the lipophilic electrolyte ETH 500 to keep the resistance of the membranes as low as possible, (22) Ueberfeld, J.; Parthasarathy, N.; Zbinden, H.; Gisin, N.; Buffle, J. Anal. Chem. 2002, 74, (23) Malon, A.; Bakker, E.; Pretsch, E. Anal. Chem. 2007, 79, (24) Oesch, U.; Simon, W. Anal. Chem. 1980, 52, (25) Telting-Diaz, M.; Bakker, E. Anal. Chem. 2002, 74, (26) Anker, P.; Ammann, D.; Asper, R.; Dohner, R. E.; Simon, W.; Wieland, E. Anal. Chem. 1981, 53, Scheme 1. Schematic of the Thin Layer Ion-Selective Coulometric System Presented Here, Consisting of a Hollow Fiber Based Calcium-Selective Working Electrode, a High Surface Area Pt Counter Electrode, and a Double Junction Ag/AgCl Reference Electrode Immersed in 1 mm CaCl mm KCl a a The working electrode is a 500 µm thick Ag/AgCl wire within the 600 µm i.d. polypropylene hollow fiber doped with 10 wt % ETH 500, 10 mmol kg -1 Ca-ionophore IV, and 30 mol % (relative to the ionophore) of ion-exchanger KTFPB in DDNPE as a calcium-selective membrane. Sample solutions pass through the tubular electrode to form a thin sample layer (b) between the wire (a) and the membrane (c), while the wire is connected to the potentiostat at (d). in analogy to earlier work. 15,16,20 The resistance was estimated with the current interrupt method (±1 µa) as 20.3 ± 0.5 kohm. The use of the calcium ionophore ETH 5234 was explored in this initial study because of its high selectivity 27,28 (higher also compared to the well-known ETH 1001) and large complex formation constant with calcium, 29 which affords a wide potential window for electrochemically mediated calcium extraction. The membrane was planned to exhibit cation-exchanger properties, which was achieved with the well known tetraphenylborate derivative KTFPB. On the basis of established ion-selective electrode theory, 10 the countercation of the ion-exchanger will exchange with calcium during a conditioning step, which renders the calcium activity in the membrane largely independent of the calcium activity in the contacting aqueous phase. Under these circumstances, the open circuit potential across the membrane follows the established Nernst equation, which is used here to monitor the concentration changes in the cell and to apply a correct excitation potential. A chloridized silver wire of 500 µm diameter was inserted into the polyprolyene tubing of nominally 600 µm inner diameter, leaving sufficient room for the fluidic delivery of the thin layer sample solution. As shown in Scheme 1, the cell was completed by inserting the tubing into an aqueous solution of 1 mm CaCl mm KCl, in which the counter and reference electrodes were placed. Note that asymmetric bathing conditions on either side of the liquid membrane may result in undesired ion countertransport processes, and a potentiometric monitoring of the transmembrane potential allows one to conveniently evaluate concentration changes in the thin layer sample with time. Figure 1 demonstrates the potentiometric response as a function of the thin layer sample composition, ranging from 10-5 to 10-2 M CaCl 2 ina10mm KCl background electrolyte. Indeed, the near-nernstian response slope suggests that the thin layer sample composition is sufficiently unperturbed by the outside bathing solution. The slight deviation from Nernstian behavior at high concentration can be quantitatively explained by the increasing chloride (27) Gehrig, P.; Rusterholz, B.; Simon, W. Chimia 1989, 43, (28) Bhakthavatsalam, V.; Shvarev, A.; Bakker, E. Analyst 2006, 131, (29) Qin, Y.; Mi, Y.; Bakker, E. Anal. Chim. Acta 2000, 421, Analytical Chemistry, Vol. 82, No. 11, June 1,

4 Figure 1. Potentiometric calibration curve for the hollow fiber based calcium-selective electrode with an inner thin layer sample solution ranging from 10-5 to 10-2 M CaCl 2 in a 10 mm KCl background electrolyte. The outer solution is constant at 1 mm CaCl mm KCl. The straight line is the least-squares fit between logarithmic concentrations of -5 and -3, with a slope of 29.6 mv per 10-fold concentration change. activity, which has an influence on the potential at the contacting silver/silver chloride electrode, as evidenced by the corrected data on the same plot. On the basis of the concentration excess of potassium ion on either side of the membrane, the Nernstian response also demonstrates a high selectivity pot to calcium ions, as expected for this ionophore (logk Ca,K ) -9.3). 28 To our knowledge, this marks the first time that a hollow fiber membrane is reported as an ion-selective membrane in potentiometry. The ion-selective calcium membrane was subsequently explored for operation in a voltammetric detection mode. Earlier work with plasticized PVC membranes containing an ionexchanger salt exhibited essentially featureless ohmic response characteristics at elevated sample concentrations, according to the Nernst equation and ohmic drop. 21 This result was one key reason why further work on voltammetric sensors avoided ion-exchanger based membranes. 20 However, it is expected that a featureless ohmic response can only be observed when concentration polarizations in the contacting aqueous phases are negligible. 21 With thin layer samples or with low sample concentrations, one will likely observe a mass transport limited current response. This is schematically shown in Figure 2, where concentration dependent featureless ohmic responses at high concentrations for a primary ion I + and a background ion J + give way to a current-potential curve that assumes a well-defined limiting current within the potential window given by the two mentioned ohmic response curves. This behavior was first explored with normal pulse voltammetry, with a 200 µm twisted wire in the inner tubing compartment. Here, the inner sample volume is expected to be substantially Figure 2. Schematic presentation of the expected current-potential behavior of ion-exchanger based ion-selective membrane electrodes. The straight lines for the primary ion I + and the background ion J + are expected for high sample concentrations where sample depletion is negligible. The separation between the two lines is given by the membrane selectivity and the activity of the two ions. When mass transport of the primary ion in the aqueous phase becomes rate limiting, as with low concentrations or volumes, a limiting current that is linearly dependent on concentration is expected between the two straight lines that define the potential window. large, resulting in acceptably small sample perturbation between measurement pulses compared to a larger silver wire diameter. In normal pulse voltammetry, a correct baseline potential value needs to be applied between pulses that allows the imposed membrane concentration polarizations to essentially return back to its original, unperturbed state. 14 This baseline potential, of course, must here be the open circuit potential, 30 which is experimentally accessible. The normal pulse voltammetry protocol, therefore, consists of observing the open circuit potential, confirming the absence of potential drift, applying this potential for a fixed amount of time, confirming a near-zero current response, and applying 1 s potential excitation pulses at given increments relative to this baseline potential. Figure 3 demonstrates the resulting current responses for the mass transport of calcium from the inner compartment of the tubing into the membrane and outside solution. Clearly, the currents are limited by calcium mass transport, as evidenced by the linear relationship between current plateau and calcium concentration. Note that, at potentials near the baseline potential, an ohmic response to calcium can be discerned, while at large potential increments beyond the current plateau, the ohmic behavior toward the background ion potassium is observed. The potential window for the plateau is a function of the ion selectivity of the membrane and the composition of the sample solution. This behavior compares well to that schematically shown in Figure 2. To explore a thin layer coulometric operation of the tubular membrane, a thicker chloridized silver wire was inserted, as for the potentiometric data presented in Figure 1. The protocol for the coulometric detection was similar as for normal pulse voltammetry. The open circuit potential was monitored to confirm the absence of potential drifts, and subsequently, this potential was applied, confirming very low current readings. A potential incre- (30) Shvarev, A.; Bakker, E. Anal. Chem. 2005, 77, Analytical Chemistry, Vol. 82, No. 11, June 1, 2010

5 Figure 3. Normal pulse voltammograms obtained for a hollow fiber based system with an increased volume of inner sample solution (smaller diameter silver wire) containing varying concentrations of CaCl 2 in a constant background of 10 mm KCl, while the outer solution consists of a constant 1 mm CaCl mm KCl. Note the concentration dependent limiting currents for this ion-exchanger based membrane system that compare to Figure 2. ment was then applied relative to the open circuit potential, now for a longer period to allow for the exhaustive depletion of the thin layer sample. Subsequently, the open circuit potential was again applied to return the membrane to its unperturbed state and to regenerate the silver chloride wire. 14,31,32 The thin layer sample solution was replaced with a fresh sample before applying the next measurement. Figure 4 shows the observed current decays for the indicated potential increments (shown as E, the applied potential minus the open circuit potential) for three different sample concentrations: (A) M, (B) M, and (C) M. Note that the initial current responses follow Ohm s law, similar to that shown in Figure 2, since thin layer sample depletion is negligible at short measurement times. Subsequently, the currents decay within the 30 s time period. For the 25 µm calcium sample at a E of 150 mv, for example, the initial current of 27.4 µa reduces to 1.2 µa. The data also show that, with increasing potential, the current decay curves start to essentially overlap, which is consistent with an exhaustive mass transport limited process. Note, however, the increasing current amplitudes for 10 µm calcium with larger applied potentials (above 175 mv relative to the open circuit potential) in Figure 4, which do not decay to zero at long times. This is clearly due to concurrent extraction of potassium at the higher applied potential, resulting in interference. It is anticipated that extended measurement protocols that incorporate an in situ background subtraction step may distinguish to some extent between electrolyte that exhaustively depletes and more concentrated background electrolyte that does not. Such studies will be forthcoming in future work. Figure 5A illustrates the calculated charge for the individual current responses as a function of applied potential (vs open circuit potential) for different calcium concentrations. Clearly, a potential window exists where the charge is strongly concentration dependent, which confirms that thin layer coulometric readout is indeed dependent on the calcium level in the thin layer sample. The width of this potential window is dependent on the selectivity (31) Bakker, E.; Meir, A. J. SIAM Rev. 2003, 45, (32) Zook, J. M.; Lindner, E. Anal. Chem. 2009, 81, Figure 4. Chronoamperograms obtained for a thin layer (500 µm thick Ag/AgCl wire) coulometric system for samples of (A) M, (B) M, and (C) M CaCl 2 all in a 10 mm KCl background during various 30 s potential pulses vs open circuit potential, as indicated on the graph. Other conditions as in Figure 3. of the membrane, with higher levels of interference diminishing the potential window. In a practical protocol, the measured open circuit potential and the available potential window may likely allow one to apply appropriate potentials that can be applied to samples of variable composition. Note that there is some increase of the observed charge with applied potential within the aforementioned potential window, which is mainly explained by non-faradaic charging currents, which have here not yet been corrected for. For comparison, the data are replotted in Figure 5B as applied potential vs Ag/AgCl, in which case the response toward the KCl background electrolyte can be better visualized. Further work will explore in situ background correction protocols to minimize this contribution. Analytical Chemistry, Vol. 82, No. 11, June 1,

6 Figure 6. Chronoamperograms of the thin layer system for the sample concentrations indicated on the graph (other conditions as in Figure 4). Here, the potential applied for individual concentrations was always 150 mv vs open circuit potential. The inset presents the calculated charge as a function of sample concentration from the chronoamperograms shown, confirming linear behavior. Repeated injections of three identical calcium samples and of samples with varying calcium concentration in a background of 10 mm KCl were performed and measured at an applied potential of 150 mv more positive than the open circuit potential. The observed current decays for the different concentrations are shown in Figure 6, while the calculated charge as a function of concentration is illustrated in the inset. No background correction was performed to best illustrate the raw performance of the system. The expected linear relationship between charge and calcium concentration is indeed observed. The nonspecific processes discussed above give a contribution of 25 µc to the background. The repeatability is very good for the initial system presented here, with relative standard deviations ranging from 0.11% for 10 µm calcium to 0.90% for 100 µm calcium (N ) 3). While the focus of this study was to introduce the underlying response principles, it is noted that higher calcium concentrations started to deviate from the calibration curve in Figure 6. For these concentrations, the electrolysis time of 30 s was not sufficiently long, as evidenced by current decays that were not exhaustive. Longer detection times, thinner membrane tubing, larger pore size of the membrane support, increased mobility within the membrane, and a thinner sample layer are all expected to increase the upper limit of detection. Higher detection limits were observed with the thin layer coulometry work of Kihara, for example, who used bulk organic solvents as the sensing phase. 19 The slope of the leastsquares fit of the calibration curve (0.882 µc/m) was found 10 days later as µc/m. As a typical example, the relative deviations in this time period of the determined charge were on the order of 2% for 50 and 75 µm concentrations of calcium. Figure 5. Calculated charge as a function of applied potential (A) vs open circuit potential and (B) vs Ag/AgCl wire for the calcium concentration indicated on the graphs. Other conditions as in Figure 4. CONCLUSIONS The study presented in this article demonstrates potential controlled thin layer coulometry with ionophore based ionselective membranes as a new technique for aqueous sample analysis. It was shown that the newly designed flow-through system consisting of a hollow fiber based ion-selective membrane and a silver/silver chloride inner electrode exhibits a potentiometric response behavior typical of a potentiometric ion-selective electrode. More importantly, it was found that an applied potential pulse forced the selective transport of the primary ion from the thin layer sample solution through the membrane and into the outer solution. The process occurs until the sample is exhausted which results in a monitored current decay. The charge calculated from the generated signal is indeed linearly related to the sample concentration. The relative standard deviation observed for several repeated sample measurements was found to be below 1% and within a time period of 10 days on the order of 2%, making this work a promising initial step toward calibration-free analysis. An effective, in situ evaluation and correction of non-faradaic charging currents and a more thorough theoretical study are underway to further develop this promising methodology. ACKNOWLEDGMENT The authors are grateful to the Australian Research Council (DP ), the CSIRO through the Flagship Cluster Sensors Systems for Analysis of Aquatic Environments, and the National Institutes of Health (R01-EB002189) for financial support of this research. Received for review February 26, Accepted April 20, AC100524Z 4542 Analytical Chemistry, Vol. 82, No. 11, June 1, 2010

Direct Ion Speciation Analysis with Ion-selective Membranes Operated in a Sequential Potentiometric/Time Resolved Chronopotentiometric Sensing Mode

Direct Ion Speciation Analysis with Ion-selective Membranes Operated in a Sequential Potentiometric/Time Resolved Chronopotentiometric Sensing Mode Supporting information for: Direct Ion Speciation Analysis with Ion-selective Membranes Operated in a Sequential Potentiometric/Time Resolved Chronopotentiometric Sensing Mode Majid Ghahraman Afshar a,b,

More information

Knowledge of total acidity and alkalinity is important for

Knowledge of total acidity and alkalinity is important for This is an open access article published under an ACS AuthorChoice License, which permits copying and redistribution of the article or any adaptations for non-commercial purposes. Cite This: Anal. Chem.

More information

READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.

READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3. CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE READING Skoog, Holler and Crouch: Chapter 23 and Appendix 3. A. INTRODUCTION Potentiometry is a static electroanalytical method in which the potential

More information

Electro Analytical Methods

Electro Analytical Methods CH 2252 Instrumental Methods of Analysis Unit II Electro Analytical Methods Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering Kalavakkam

More information

ANALYTICAL SCIENCES JANUARY 1999, VOL The Japan Society for Analytical Chemistry

ANALYTICAL SCIENCES JANUARY 1999, VOL The Japan Society for Analytical Chemistry 1999 The Japan Society for Analytical Chemistry 47 Additive-Salt Effect on Low Detection Limit and Slope Sensitivity in Response of Potassium- and Sodium-Selective Neutral Carrier Based Electrodes and

More information

SUPPLEMENTARY INFORMATION

SUPPLEMENTARY INFORMATION DOI: 10.1038/NCHEM.1858 Photocurrent Generation Based on a Light-Driven Proton Pump in an Artificial Liquid Membrane Xiaojiang Xie, Gastón A. Crespo, Günter Mistlberger and Eric Bakker* [ ] Department

More information

Fundamental molecular electrochemistry - potential sweep voltammetry

Fundamental molecular electrochemistry - potential sweep voltammetry Fundamental molecular electrochemistry - potential sweep voltammetry Potential (aka voltammetric) sweep methods are the most common electrochemical methods in use by chemists today They provide an efficient

More information

General terms relevant to ion-selective electrodes

General terms relevant to ion-selective electrodes 8.3.2.1 General terms relevant to ion-selective electrodes Activity, Activity coefficient and Concentration See sections 1.3.8, 1.3.10 and 3.3. Calibration plot This is a plot of the cell emf, E (i.e.

More information

Unit 2 Electrochemical methods of Analysis

Unit 2 Electrochemical methods of Analysis Unit 2 Electrochemical methods of Analysis Recall from Freshman Chemistry: Oxidation: Loss of electrons or increase in the oxidation number Fe 2 e - Fe 3 Reduction: Gain of electrons or decreases in the

More information

Chem 321 Lecture 17 - Potentiometry 10/24/13

Chem 321 Lecture 17 - Potentiometry 10/24/13 Student Learning Objectives Chem 321 Lecture 17 - Potentiometry 10/24/13 Electrodes The cell described in the potentiometric chloride titration (see 10/22/13 posting) consists of a Ag/AgCl reference electrode

More information

Chapter 3 Electrochemical methods of Analysis-Potentiometry

Chapter 3 Electrochemical methods of Analysis-Potentiometry Chapter 3 Electrochemical methods of Analysis-Potentiometry Electroanalytical chemistry Contents Introduction Galvanic and electrolytic cells Salt bridge Electrode potential and cell potential Indicator

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2014 Journal Name Chemicals Electronic Supplementary Information Two kinds of commercial anion exchange

More information

Ion-selective electrodes: challenges and opportunities for on-site/on-line measurements

Ion-selective electrodes: challenges and opportunities for on-site/on-line measurements EURACEM Workshop on Quality Assurance of Measurements from Field to Laboratory, MIKES, Espoo Finland, 20-21 May 2013 Ion-selective electrodes: challenges and opportunities for on-site/on-line measurements

More information

Instrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013

Instrumental Chemical Analysis. Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 0510212 Instrumental Chemical Analysis Dr. Abdul Muttaleb Jaber Professor Faculty of Pharmacy Philadelphia University Fall 2012/2013 Chapter 1 Electroanalytical Methods Electroanalytical Chemistry Electroanalytical

More information

LANTHANUM (III) SELECTIVE ELECTRODE BASE 1,10-DIAZA-4,7,13,16- TETRAOXACYCLOOCTADECANE-N,N -DIACETIC ACID AS IONOPHORE.

LANTHANUM (III) SELECTIVE ELECTRODE BASE 1,10-DIAZA-4,7,13,16- TETRAOXACYCLOOCTADECANE-N,N -DIACETIC ACID AS IONOPHORE. 1 LANTHANUM (III) SELECTIVE ELECTRODE BASE 1,10-DIAZA-4,7,13,16- TETRAOXACYCLOOCTADECANE-N,N -DIACETIC ACID AS IONOPHORE Suyanta a, *, a) Departement of Chemistry, Yogyakarta State University, Yogyakarta

More information

Supporting Information for: Potentiometric Sensing Array for Monitoring Aquatic Systems

Supporting Information for: Potentiometric Sensing Array for Monitoring Aquatic Systems Electronic Supplementary Material (ESI) for Environmental Science: Processes & Impacts. This journal is The Royal Society of Chemistry 2015 Supporting Information for: Potentiometric Sensing Array for

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required

More information

ELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY

ELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY ELECTROCHEMICAL TECHNIQUES, OSMOMETRY AND THE PRINCIPLES OF RADIOACTIVITY ELECTROCHEMISTY ELECTROCHEMISTRY IS THE STUDY OF CHEMICAL REACTIONS THAT RESULT IN THE FLOW OF ELECTRONS (CURRENT) OR THE DEVELOPMENT

More information

Chapter 24. Electrogravimetry and Coulometry

Chapter 24. Electrogravimetry and Coulometry Chapter 24 Electrogravimetry and Coulometry Dynamic Electrochemical Methods of analysis Electrolysis Electrogravimetric and Coulometric Methods For a cell to do any useful work or for an electrolysis to

More information

Fig. Electrochemical Cell/ Potentiometric Titration

Fig. Electrochemical Cell/ Potentiometric Titration Fig. Electrochemical Cell/ Potentiometric Titration The accurate, precise and effective potentiometric measurements can be made with the help of the following two types of electrodes namely : REFERENCE

More information

Chromoionophores are lipophilic ph indicators that are

Chromoionophores are lipophilic ph indicators that are This is an open access article published under an ACS AuthorChoice License, which permits copying and redistribution of the article or any adaptations for non-commercial purposes. Cite This: Anal. Chem.

More information

Cyclic Voltammetry. Fundamentals of cyclic voltammetry

Cyclic Voltammetry. Fundamentals of cyclic voltammetry Cyclic Voltammetry Cyclic voltammetry is often the first experiment performed in an electrochemical study of a compound, biological material, or an electrode surface. The effectiveness of cv results from

More information

Single Catalyst Electrocatalytic Reduction of CO 2 in Water to H 2 :CO Syngas Mixtures with Water Oxidation to O 2

Single Catalyst Electrocatalytic Reduction of CO 2 in Water to H 2 :CO Syngas Mixtures with Water Oxidation to O 2 Electronic Supplementary Material (ESI) for Energy & Environmental Science. This journal is The Royal Society of Chemistry 2014 Supporting Information Single Catalyst Electrocatalytic Reduction of CO 2

More information

Supporting Information

Supporting Information Supporting Information Electrochemical Synthesis of Ammonia from N 2 and H 2 O under Ambient Conditions Using Pore-Size Controlled Hollow Gold Nanocatalysts with Tunable Plasmonic Properties Mohammadreza

More information

Physico-Chemical Studies of the Pvc K + - Selective Membrane

Physico-Chemical Studies of the Pvc K + - Selective Membrane Leonardo Journal of Sciences ISSN 1583-0233 Issue 1, July-December 2002 p. 1-12 Physico-Chemical Studies of the Pvc K + - Selective Membrane COROIAN Ana 1, PICĂ Elena Maria 2*, GRECU Rodica 1, COSMA Viorica

More information

An effective nanostructured assembly for ion-selective electrodes. Ionophore covalently linked to carbon nanotubes for Pb 2+ determination

An effective nanostructured assembly for ion-selective electrodes. Ionophore covalently linked to carbon nanotubes for Pb 2+ determination Supplementary Material (ESI) For Chemical Communications An effective nanostructured assembly for ion-selective electrodes. Ionophore covalently linked to carbon nanotubes for Pb 2+ determination Enrique

More information

Chapter 11. Electrochemical Methods. In Chapter 10 we examined several spectroscopic techniques that take advantage of the.

Chapter 11. Electrochemical Methods. In Chapter 10 we examined several spectroscopic techniques that take advantage of the. Chapter 11 Electrochemical Methods Chapter Overview Section 11A Overview of Electrochemistry Section 11B Potentiometric Methods Section 11C Coulometric Methods Section 11D Voltammetric and Amperometric

More information

Ion-selective electrode for measuring low Ca 2+ concentrations in the presence of high K +,Na + and Mg 2+ background

Ion-selective electrode for measuring low Ca 2+ concentrations in the presence of high K +,Na + and Mg 2+ background Anal Bioanal Chem (2006) 385: 1477 1482 DOI 10.1007/s00216-006-0607-y ORIGINAL PAPER Iwona Bedlechowicz-Śliwakowska. Peter Lingenfelter. Tomasz Sokalski. Andrzej Lewenstam. Magdalena Maj-Żurawska Ion-selective

More information

Comparative Studies on Various Room-Temperature Vulcanizing-Type Silicone Rubbers as Ion-Selective Electrode Matrices

Comparative Studies on Various Room-Temperature Vulcanizing-Type Silicone Rubbers as Ion-Selective Electrode Matrices ANALYTICAL SCIENCES DECEMBER 1997, VOL. 13 969 Comparative Studies on Various Room-Temperature Vulcanizing-Type Silicone Rubbers as Ion-Selective Electrode Matrices Kyung Lae RHO*, Sang Hyun HAN*, Hakhyun

More information

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

The design and operational theory of ph electrodes is a very complex subject, explored only briefly here. What is important to understand is that thes

The design and operational theory of ph electrodes is a very complex subject, explored only briefly here. What is important to understand is that thes ph measurement A very important measurement in many liquid chemical processes (industrial, pharmaceutical, manufacturing, food production, etc.) is that of ph: the measurement of hydrogen ion concentration

More information

Current response of ion-selective solvent polymeric membranes at controlled potential

Current response of ion-selective solvent polymeric membranes at controlled potential Published in Journal of Electroanalytical Chemistry 571, issue 1, 27-35, 2004 which should be used for any reference to this work 1 Current response of ion-selective solvent polymeric membranes at controlled

More information

Analysis of cations and anions by Ion- Selective Electrodes (ISEs)

Analysis of cations and anions by Ion- Selective Electrodes (ISEs) Analysis of cations and anions by Ion- Selective Electrodes (ISEs) Purpose: The purpose of this assignment is to introduce potentiometric measurements of ionic species by ion selective electrodes (ISEs)

More information

Ion Selective Sensors and Electrodes Technologies for Industrially, Environmentally and Biologically Significant Ion Measurements

Ion Selective Sensors and Electrodes Technologies for Industrially, Environmentally and Biologically Significant Ion Measurements Ion Selective Sensors and Electrodes Technologies for Industrially, Environmentally and Biologically Significant Ion Measurements Introduction Advanced Sensor Technologies, Inc. 603 North Poplar Street

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 011, 3(4):551-555 Fabrication of Eu 3+ Polymeric Membrane Ion-Selective

More information

In-situ potentiometry and ellipsometry: a promising tool to study biofouling of potentiometric sensors

In-situ potentiometry and ellipsometry: a promising tool to study biofouling of potentiometric sensors In-situ potentiometry and ellipsometry: a promising tool to study biofouling of potentiometric sensors Grzegorz Lisak 1,2,3, *, Thomas Arnebrant 2,3, Andrzej Lewenstam 1,4, Johan Bobacka 1, Tautgirdas

More information

Detecting and manipulating ions

Detecting and manipulating ions Spotlight on: Bioanalytics Ion Sensors Detecting and manipulating ions From Potentiometry to the Nanoscale Xiaojiang Xie and Prof. Dr Eric Bakker, Department of inorganic, analytical and applied chemistry,

More information

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Ion Selective Electrodes for the Laboratory Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015 Outline Review ISE measurement technology How to properly calibrate

More information

Unit 2 B Voltammetry and Polarography

Unit 2 B Voltammetry and Polarography Unit 2 B Voltammetry and Polarography Voltammetric methods of Analysis What is Voltammetry? A time-dependent potential is applied to an electrochemical cell, and the current flowing through the cell is

More information

Preparation of Prussian blue-modified screen-printed electrodes via a chemical deposition for mass production of stable hydrogen peroxide sensors

Preparation of Prussian blue-modified screen-printed electrodes via a chemical deposition for mass production of stable hydrogen peroxide sensors Procedure 7 Preparation of Prussian blue-modified screen-printed electrodes via a chemical deposition for mass production of stable hydrogen peroxide sensors Francesco Ricci, Danila Moscone and Giuseppe

More information

ProPac WCX-10 Columns

ProPac WCX-10 Columns ProPac WCX-10 Columns Guidance for column use Tips to maximize column lifetime ProPac WCX-10 Column Tips and Tricks This guide provides essential information and invaluable guidelines for mobile phases,

More information

GENERAL INSTRUCTIONS GENERAL PREPARATION

GENERAL INSTRUCTIONS GENERAL PREPARATION GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions.

More information

CHAPTER 14: ELECTRODES AND POTENTIOMETRY

CHAPTER 14: ELECTRODES AND POTENTIOMETRY CHAPTER 14: ELECTRODES AND POTENTIOMETRY Chapter 14 Electrodes and Potentiometry Potentiometry : The use of electrodes to measure voltages that provide chemical information. (The cell voltage tells us

More information

Specific Determination of Hydrogen Peroxide With A Catalase Biosensor Based on Mercury Thin Film Electrodes

Specific Determination of Hydrogen Peroxide With A Catalase Biosensor Based on Mercury Thin Film Electrodes Turk J Chem 24 (2000), 95 99 c TÜBİTAK Specific Determination of Hydrogen Peroxide With A Catalase Biosensor Based on Mercury Thin Film Electrodes Nil ERTAŞ Ege University, Faculty of Science, Department

More information

Chapter 13 POTENTIOMETRIC ELECTRODES AND POTENTIOMETRY

Chapter 13 POTENTIOMETRIC ELECTRODES AND POTENTIOMETRY Chapter 13 POTENTIOMETRIC ELECTRODES AND POTENTIOMETRY POTENTIOMETRIC ELECTRODES AND POTENTIOMETRY Ch.13 In this chapter Understand the concept of the various types of electrodes that can be used for measuring

More information

A glance into the bulk of solvent polymeric ph membranes*

A glance into the bulk of solvent polymeric ph membranes* Pure Appl. Chem., Vol. 73, No. 1, pp. 17 22, 2001. 2001 IUPAC A glance into the bulk of solvent polymeric ph membranes* Erno" Lindner 1,, Robert E. Gyurcsányi 1,2, and Bradford D. Pendley 3 1 Joint Graduate

More information

METHODS FOR DETERMINATIONS OF ELECTROLYTES AND BLOOD GASES

METHODS FOR DETERMINATIONS OF ELECTROLYTES AND BLOOD GASES METHODS FOR DETERMINATIONS OF ELECTROLYTES AND BLOOD GASES ELECTROCHEMISTRY Basic principle Electrodes are used to selectively measure particular ions Instruments utilizing electrodes measure the potential

More information

Supplementary Information. Carolyn Richmonds, Megan Witzke, Brandon Bartling, Seung Whan Lee, Jesse Wainright,

Supplementary Information. Carolyn Richmonds, Megan Witzke, Brandon Bartling, Seung Whan Lee, Jesse Wainright, Supplementary Information Electron transfer reactions at the plasma-liquid interface Carolyn Richmonds, Megan Witzke, Brandon Bartling, Seung Whan Lee, Jesse Wainright, Chung-Chiun Liu, and R. Mohan Sankaran*,

More information

Highly efficient hydrogen evolution of platinum via tuning the interfacial dissolved-gas concentration

Highly efficient hydrogen evolution of platinum via tuning the interfacial dissolved-gas concentration Electronic Supplementary Material (ESI) for Chemical Communications. This journal is The Royal Society of Chemistry 2018 Supporting Information for Highly efficient hydrogen evolution of platinum via tuning

More information

Electrochemical Ion Transfer With Thin Films of Poly(3- octylthiophene)

Electrochemical Ion Transfer With Thin Films of Poly(3- octylthiophene) Supporting Information for: Electrochemical Ion Transfer With Thin Films of Poly(3- octylthiophene) Maria Cuartero #, Robert G. Acres &, Roland De Marco,,, Eric Bakker # and Gastón A. Crespo #,* # Department

More information

EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION

EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION I. INTRODUCTION This experiment introduces the direct-measurement approach to potentiometry. Principal purposes of the study are a) to understand quantitative

More information

Supporting Information for

Supporting Information for Supporting Information for Electrodeposition of Isolated Platinum Atoms and Clusters on Bismuth Characterization and Electrocatalysis Min Zhou, Jeffrey E. Dick, and Allen J. Bard Center for Electrochemistry,

More information

CHAPTER-5 CYCLIC VOLTAMETRIC STUDIES OF NOVEL INDOLE ANALOGUES PREPARED IN THE PRESENT STUDY

CHAPTER-5 CYCLIC VOLTAMETRIC STUDIES OF NOVEL INDOLE ANALOGUES PREPARED IN THE PRESENT STUDY CHAPTER-5 CYCLIC VOLTAMETRIC STUDIES OF NOVEL INDOLE ANALOGUES PREPARED IN THE PRESENT STUDY Page No. 175-187 5.1 Introduction 5.2 Theoretical 5.3 Experimental 5.4 References 5. 1 Introduction Electrochemical

More information

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE) (dr hab. inż. Andrzej Wasik, Gdańsk 2016) The aim of this laboratory exercise is to familiarise students with

More information

Investigation of the diffusion of ferricyanide through porous membranes using the SECM150

Investigation of the diffusion of ferricyanide through porous membranes using the SECM150 Investigation of the diffusion of ferricyanide through porous membranes using the SECM150 I INTRODUCTION Scanning ElectroChemical Microscopy (SECM) has the ability to provide spatially resolved, rather

More information

Supporting Online Material for

Supporting Online Material for www.sciencemag.org/cgi/content/full/332/6025/81/dc1 Supporting Online Material for Electrochemically Mediated Atom Transfer Radical Polymerization Andrew J. D. Magenau, Nicholas C. Strandwitz, Armando

More information

Combination ph, ORP, T Electrode with Transmitter

Combination ph, ORP, T Electrode with Transmitter Combination ph, ORP, T Electrode with Transmitter Technical Information and Manual 2014 Osorno Enterprises Inc. The content of this manual is protected by copyright. All rights reserved. OCS/1406 Combination

More information

Chapter 14: Electrodes and Potentiometry

Chapter 14: Electrodes and Potentiometry Yonsei University Chapter 14: Electrodes and Potentiometry The use of electrodes to measure voltages that provide chemical information is called potentiometry (ion-selective electrode, ion-sensing field

More information

Current based methods

Current based methods Current based methods Amperometric and voltammetric sensors More significant influence on analytical parameters (sensitivity, selectivity, interferences elimination) kind of method, potential range, electrode

More information

A simple on-plastic/paper inkjet-printed solid-state Ag/AgCl pseudoreference

A simple on-plastic/paper inkjet-printed solid-state Ag/AgCl pseudoreference Supporting Information A simple on-plastic/paper inkjet-printed solid-state Ag/AgCl pseudoreference electrode Everson Thiago Santos Gerôncio da Silva,, Sandrine Miserere, Lauro Tatsuo Kubota and Arben

More information

Supplementary Material

Supplementary Material Supplementary Material Digital Electrogenerated Chemiluminescence Biosensor for the Determination of Multiple Proteins Based on Boolean Logic Gate Honglan Qi*, Xiaoying Qiu, Chen Wang, Qiang Gao, Chengxiao

More information

Electrochemistry. Michael Faraday s law of electromagnetic induction says that whenever a conductor is

Electrochemistry. Michael Faraday s law of electromagnetic induction says that whenever a conductor is Surname 1 Name Course Instructor Date Electrochemistry 1. Faraday s Law Michael Faraday s law of electromagnetic induction says that whenever a conductor is positioned in a changeable magnetic field emf

More information

Flow-Injection Coulometric Detection Based on Ion Transfer and Its Application to the Determination of Chlorpromazine

Flow-Injection Coulometric Detection Based on Ion Transfer and Its Application to the Determination of Chlorpromazine Sensors 2008, 8, 3678-3688; DOI: 10.3390/s8063678 Article OPEN ACCESS sensors ISSN 1424-8220 www.mdpi.org/sensors Flow-Injection Coulometric Detection Based on Ion Transfer and Its Application to the Determination

More information

The Water Uptake of Plasticized Poly(vinyl chloride) Solid- Contact Calcium-Selective Electrodes

The Water Uptake of Plasticized Poly(vinyl chloride) Solid- Contact Calcium-Selective Electrodes Full Paper The Water Uptake of Plasticized Poly(vinyl chloride) Solid- Contact Calcium-Selective Electrodes Tom Lindfors,* a, b Fredrik Sundfors, a Lajos Hçfler, c, d Róbert E. Gyurcsµnyi c, e a Laboratory

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA Background In this experiment, students will familiarize themselves with potentiometric titration, practice using the first derivative to find the equivalence

More information

Influence of the Circulating Current on the Propagation of the Change in Membrane Potential

Influence of the Circulating Current on the Propagation of the Change in Membrane Potential ANALYTICAL SCIENCES JULY 2015, VOL. 31 677 2015 The Japan Society for Analytical Chemistry Influence of the Circulating Current on the Propagation of the Change in Membrane Potential Yuki KUSHIDA, Osamu

More information

Supporting Information for A potential-controlled switch on/off mechanism for selective excitation in mixed electrochemiluminescent systems

Supporting Information for A potential-controlled switch on/off mechanism for selective excitation in mixed electrochemiluminescent systems Supporting Information for A potential-controlled switch on/off mechanism for selective excitation in mixed electrochemiluminescent systems Egan H. Doeven, Elizabeth M. Zammit, Gregory J. Barbante, Paul

More information

Heterogeneous electron transfer at Au/SAM junctions in a room-temperature ionic liquid under pressure

Heterogeneous electron transfer at Au/SAM junctions in a room-temperature ionic liquid under pressure Supporting Information Heterogeneous electron transfer at Au/SAM junctions in a room-temperature ionic liquid under pressure Tina D. Dolidze, Dimitri E. Khoshtariya,* Peter Illner and Rudi van Eldik* a)

More information

CHAPTER 1 INTRODUCTION. precision agriculture (PA) (Oliver et al., 2013). PA is one of the top ten revolutions in

CHAPTER 1 INTRODUCTION. precision agriculture (PA) (Oliver et al., 2013). PA is one of the top ten revolutions in CHAPTER 1 INTRODUCTION 1.1 Overview on precision agriculture The world population is growing and lands available for agriculture are likely to decrease despite technological improvement that might extend

More information

Ion-Selective Microelectrodes

Ion-Selective Microelectrodes Daniel Ammann Ion-Selective Microelectrodes Principles, Design and Application With 153 Figures and 42 Tables Springer-Verlag Berlin Heidelberg New York Tokyo Table of Contents 1 Introduction 1 2 Classification

More information

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT METHOD 9040B ph ELECTROMETRIC MEASUREMENT 1.0 SCOPE AND APPLICATION 1.1 Method 9040 is used to measure the ph of aqueous wastes and those multiphase wastes where the aqueous phase constitutes at least

More information

Electronic Supplementary Material (ESI) for Dalton Transactions This journal is The Royal Society of Chemistry Supplementary Information

Electronic Supplementary Material (ESI) for Dalton Transactions This journal is The Royal Society of Chemistry Supplementary Information Supplementary Information 1 2-(2-Hydroxyphenyl)-1H-benzimidazole-manganese oxide hybrid as a promising structural model for Tyrosine 161/Histidine 190-manganese cluster in Photosystem II Mohammad Mahdi

More information

TEPZZ 6_Z6_ A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.:

TEPZZ 6_Z6_ A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.: (19) TEPZZ 6_Z6_ A_T (11) EP 2 6 612 A1 (12) EUROPEAN PATENT APPLICATION (43) Date of publication: 03.07.2013 Bulletin 2013/27 (51) Int Cl.: G01N 27/333 (2006.01) G01N 27/416 (2006.01) (21) Application

More information

Quantitative Electrophysiology

Quantitative Electrophysiology ECE 795: Quantitative Electrophysiology Notes for Lecture #1 Tuesday, September 18, 2012 1. INTRODUCTION TO EXCITABLE CELLS Historical perspective: Bioelectricity first discovered by Luigi Galvani in 1780s

More information

Reference electrode. Calomel electrode Hg in contact with Hg(I) chloride Ag/AgCl 15-2

Reference electrode. Calomel electrode Hg in contact with Hg(I) chloride Ag/AgCl 15-2 Potentiometry Potential measurements of electrochemical cells Ion selective methods Reference electrode Indicator electrode Potential measuring device Reference electrode Indicator electrodes Ion specific

More information

lect 26:Electrolytic Cells

lect 26:Electrolytic Cells lect 26:Electrolytic Cells Voltaic cells are driven by a spontaneous chemical reaction that produces an electric current through an outside circuit. These cells are important because they are the basis

More information

#13 Electrochemical Cells

#13 Electrochemical Cells #13 Electrochemical Cells If a copper strip is placed in a solution of copper ions, one of the following reactions may occur: Cu 2+ + 2e - Cu Cu Cu 2+ + 2e - The electrical potential that would be developed

More information

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL Cole-Parmer Instrument Company (800)323-4340 Fax:(847)247-2929 625 East Bunker Court, Vernon Hills, Illinois 60061 http://www.coleparmer.com

More information

Principles and Applications of Electrochemistry

Principles and Applications of Electrochemistry Principles and Applications of Electrochemistry Fourth edition D. R. CROW Professor of Electrochemistry and Dean of Research University of Wolverhampton BLACKIE ACADEMIC & PROFESSIONAL An Imprint of Chapman

More information

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste Experiment ISE: Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste 67 You have been hired by the government to check the fluoride concentration labelling on some major

More information

K + -Selective Optical Microsensors Based On Surface Modified Polystyrene Microspheres

K + -Selective Optical Microsensors Based On Surface Modified Polystyrene Microspheres Electronic Supplementary Material (ESI) for ChemComm. This journal is The Royal Society of Chemistry 214 Supporting information for K + -Selective Optical Microsensors Based On Surface Modified Polystyrene

More information

Thallium(I)-Selective Membrane Potentiometric Sensor Based on Dibenzyldiaza-18-Crown-6

Thallium(I)-Selective Membrane Potentiometric Sensor Based on Dibenzyldiaza-18-Crown-6 Thallium(I)-Selective Membrane Potentiometric Sensor Bull. Korean Chem. Soc. 2003, Vol. 24, No. 4 421 Thallium(I)-Selective Membrane Potentiometric Sensor Based on Dibenzyldiaza-18-Crown-6 Gholamreza Khayatian,

More information

Cyclic Voltammetry. Objective: To learn the basics of cyclic voltammetry with a well-behaved echem system

Cyclic Voltammetry. Objective: To learn the basics of cyclic voltammetry with a well-behaved echem system Cyclic Voltammetry Objective: To learn the basics of cyclic voltammetry with a well-behaved echem system Introduction Cyclic voltammetry (CV) is a popular electroanalytical technique for its relative simplicity

More information

Activated Carbon as a Pseudo-reference Electrode for Potentiometric Sensing Inside Concrete

Activated Carbon as a Pseudo-reference Electrode for Potentiometric Sensing Inside Concrete Delft University of Technology Activated Carbon as a Pseudo-reference Electrode for Potentiometric Sensing Inside Concrete Abbas, Y.; Pargar, Farhad; Olthuis, W; van den Berg, A. DOI 10.1016/j.proeng.2014.11.717

More information

Lecture 12: Electroanalytical Chemistry (I)

Lecture 12: Electroanalytical Chemistry (I) Lecture 12: Electroanalytical Chemistry (I) 1 Electrochemistry Electrochemical processes are oxidation-reduction reactions in which: Chemical energy of a spontaneous reaction is converted to electricity

More information

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL The Cole-Parmer Industrial Ammonium Ion Electrode is used to quickly, simply, accurately, and economically

More information

Electronic Supplementary Material (ESI) for Chemical Communications This journal is The Royal Society of Chemistry 2011

Electronic Supplementary Material (ESI) for Chemical Communications This journal is The Royal Society of Chemistry 2011 Supplementary Information for Selective adsorption toward toxic metal ions results in selective response: electrochemical studies on polypyrrole/reduced graphene oxide nanocomposite Experimental Section

More information

Quantitative Electrophysiology

Quantitative Electrophysiology ECE 795: Quantitative Electrophysiology Notes for Lecture #1 Wednesday, September 13, 2006 1. INTRODUCTION TO EXCITABLE CELLS Historical perspective: Bioelectricity first discovered by Luigi Galvani in

More information

NOTICE: this is the author s version of a work that was accepted for publication in Electrochemistry Communications. Changes resulting from the

NOTICE: this is the author s version of a work that was accepted for publication in Electrochemistry Communications. Changes resulting from the NOTICE: this is the author s version of a work that was accepted for publication in Electrochemistry Communications. Changes resulting from the publishing process, such as peer review, editing, corrections,

More information

BIOSENOSRS BIO 580. Electrochemical Biosensors - theory part 1 WEEK 1 Fall Semester

BIOSENOSRS BIO 580. Electrochemical Biosensors - theory part 1 WEEK 1 Fall Semester BIOSENOSRS BIO 580 Electrochemical Biosensors - theory part 1 WEEK 1 Fall Semester Faculty: Dr. Javed H. Niazi KM Faculty of Engineering & Natural Sciences Sabanci University Topics that will be covered

More information

Supplementary Materials

Supplementary Materials Atomic layer-deposited tunnel oxide stabilizes silicon photoanodes for water oxidation Yi Wei Chen 1, Jonathan D. Prange 2, Simon Dühnen 2, Yohan Park 1, Marika Gunji 1, Christopher E. D. Chidsey 2, and

More information

Chapter 22. Bulk Electrolysis: Electrogravimetry and Coulometry. Definition. Features of Bulk Electrolysis Cells

Chapter 22. Bulk Electrolysis: Electrogravimetry and Coulometry. Definition. Features of Bulk Electrolysis Cells Chapter 22 Bulk Electrolysis: Electrogravimetry and Coulometry Definition Bulk Electrolysis deals with methods that involve electrolysis producing a quantitative change in oxidation state Example: In a

More information

ANALYSIS OF LEAD IN SEAWATER

ANALYSIS OF LEAD IN SEAWATER ANALYSIS OF LEAD IN SEAWATER BY DIFFERENTIAL PULSE POLAROGRAPHY Introduction Electrochemical methods of analysis can be used for the quantitative analysis of any electroactive species any species that

More information

Supporting Information Reagents. Physical methods. Synthesis of ligands and nickel complexes.

Supporting Information Reagents. Physical methods. Synthesis of ligands and nickel complexes. Supporting Information for Catalytic Water Oxidation by A Bio-inspired Nickel Complex with Redox Active Ligand Dong Wang* and Charlie O. Bruner Department of Chemistry and Biochemistry and Center for Biomolecular

More information

Homogeneous Electrochemical Assay for Protein Kinase Activity

Homogeneous Electrochemical Assay for Protein Kinase Activity Homogeneous Electrochemical Assay for Protein Kinase Activity Ik-Soo Shin,,, Rohit Chand, Sang Wook Lee, Hyun-Woo Rhee, Yong-Sang Kim, * and Jong-In Hong* Corresponding Author *Prof. Dr. J.-I. Hong, Department

More information

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction The Cole-Parmer Calcium Ion Electrodes are used to quickly, simply, accurately, and economically

More information

Electroanalytical methods

Electroanalytical methods Electroanalytical methods G. Galbács Electrochemical methods In electrochemical methods of instrumental analysis, one measures voltage (potential) and/or current signals. A variety of electrochemical methods

More information

Secondary Ag/AgCl Pseudo-Reference Electrode on Silicon Substrate

Secondary Ag/AgCl Pseudo-Reference Electrode on Silicon Substrate Secondary Ag/AgCl Pseudo-Reference Electrode on Silicon Substrate F. L. Almeida a, M. B. A. Fontes b, C. Jimenez c and I. Burdallo c a Laboratório de Sistema Integráveis LSI, University of São Paulo, São

More information