Determination of the boron content in polyethylene samples using the reactor Orphée

Similar documents
Neutrons production on the IPHI accelerator for the validation of the design of the compact neutron source SONATE.

This paper should be understood as an extended version of a talk given at the

Quality control of neutron-absorber materials for the nuclear fuel cycle, Principle of the JEN-3 neutron Backscattering gauge

Control of the fission chain reaction

Measurements of liquid xenon s response to low-energy particle interactions

GAMMA RAY SPECTROSCOPY

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center

Determination of the activity of radionuclides

Neutron capture and fission reactions on. sections, -ratios and prompt -ray emission from fission. 1 Introduction and Motivation

Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor

A new detector for neutron beam monitoring

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Reactor radiation skyshine calculations with TRIPOLI-4 code for Baikal-1 experiments

Preparatory experiments for cold-neutron induced fission studies at IKI

Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor. Abstract

EEE4106Z Radiation Interactions & Detection

Neutron Interactions with Matter

TRITIUM PRODUCTION CROSS SECTION FOR NEUTRON INTERACTION WITH 10 B NUCLEI

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

M.Cagnazzo Atominstitut, Vienna University of Technology Stadionallee 2, 1020 Wien, Austria

Measurement of prompt fission γ-ray spectra in fast neutroninduced

Neutron capture cross section of 25 Mg and its astrophysical implications

Fission Fragment characterization with FALSTAFF at NFS

EXPERIMENTAL STUDY OF NEUTRON FIELDS PRODUCED IN PROTON REACTIONS WITH HEAVY TARGETS. Nuclear Physics Institute AS CR, Rez Czech Republic

Precision neutron flux measurement with a neutron beam monitor

6 Neutrons and Neutron Interactions

An introduction to Neutron Resonance Densitometry (Short Summary)

Detection and measurement of gamma-radiation by gammaspectroscopy

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices

Bi β + Po Bismuth-214 is radioactive. It has a half-life of 20 minutes. (a) The nuclide notation for bismuth-214 is Bi.

Calibration of the GNU and HSREM neutron survey instruments

in2p , version 1-28 Nov 2008

PHYSICS A2 UNIT 2 SECTION 1: RADIOACTIVITY & NUCLEAR ENERGY

Chapter V: Interactions of neutrons with matter

CONCEPT MAP ATOMS. Atoms. 1.Thomson model 2.Rutherford model 3.Bohr model. 6. Hydrogen spectrum

Activation Analysis. Characteristic decay mechanisms, α, β, γ Activity A reveals the abundance N:

POSITION SENSITIVE DETECTION OF CONCEALED SUBSTANCES EMPLOYING PULSED SLOW NEUTRONS

Fast neutron inelastic scattering Roland Beyer, Forschungszentrum Dresden-Rossendorf

THE COMPTON EFFECT Last Revised: January 5, 2007

Application of prompt gamma activation analysis with neutron beams for the detection and analysis of nuclear materials in containers

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

Neutrons For Science (NFS) at SPIRAL-2 (Part II: pulsed neutron beam)

Neutron capture cross sections of 69,71 Ga at n TOF EAR1

Neutron Based Techniques for the Detection of Illicit Materials and Explosives

Nuclear cross-section measurements at the Manuel Lujan Jr. Neutron Scattering Center. Michal Mocko

A new scintillator detector for nuclear physics experiments: the CLYC scintillator

Determination of research reactor fuel burnup

Detekce a spektrometrie neutronů. neutron detection and spectroscopy

Characterization and Monte Carlo simulations for a CLYC detector

Introduction to Nuclear Engineering

Radiation and Radioactivity. PHYS 0219 Radiation and Radioactivity

DETERMINATION OF THE HYDROGEN CONTENT OF OIL SAMPLES FROM NIGERIA USING AN Am-Be NEUTRON SOURCE.

Neutronic Design on a Small Accelerator based 7 Li (p, n) Neutron Source for Neutron Scattering Experiments

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

Measurements of Neutron Total and Capture Cross Sections at the TOF spectrometers of the Moscow Meson Factory

Neutrons For Science

Neutron transmission and capture measurements for 241 Am at GELINA

Experimental study of the flux trap effect in a sub-critical assembly

The Attenuation of Neutrons in Barite Concrete

Digital simulation of neutron and gamma measurement devices

Introduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

Unit 3 Atomics Repair Problems

School on Pulsed Neutrons: Characterization of Materials October Neurton Sources & Scattering Techniques (1-2)

LECTURE 6: INTERACTION OF RADIATION WITH MATTER

Strong interplay between stripe spin fluctuations, nematicity and superconductivity in FeSe

Pulsed Neutron Interrogation Test Assembly - PUNITA

Neutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Nuclear Reactions A Z. Radioactivity, Spontaneous Decay: Nuclear Reaction, Induced Process: x + X Y + y + Q Q > 0. Exothermic Endothermic

CHAPTER 4: NUCLEAR INSTRUMENTATION MODULE 2: LOG RANGE DETECTORS

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

Chapter 2 Methods Based on the Absorption of Gamma-Ray Beams by Matter

Background optimization for a new spherical gas detector for very light WIMP detection

Neutron Activation of 74 Ge and 76 Ge

The investigation of the neutron total cross section & measurement of the nuclear radius with comparison of Ramsauer model

Small Angle Neutron Scattering in Different Fields of Research. Henrich Frielinghaus

Neutron cross sections in stellar nucleosynthesis: study of the key isotope 25 Mg

Measurement of the neutron capture cross section of gadolinium even isotopes relevant to Nuclear Astrophysics

Chem 481 Lecture Material 4/22/09

Flux and neutron spectrum measurements in fast neutron irradiation experiments

Neutron Physics. The De Broglie relation for matter is: h mv (1) λ =

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Gamma Spectroscopy. References: Objectives:

A new neutron monitor for pulsed fields at high-energy accelerators

Photoneutron reactions studies at ELI-NP using a direct neutron multiplicity sorting method Dan Filipescu

"Neutron Flux Distribution"

Detection of Neutron Sources in Cargo Containers

Method of active correlations in the experiment 249 Cf+ 48 Ca n

Considerations for Measurements in Support of Thermal Scattering Data Evaluations. Ayman I. Hawari

Rb, which had been compressed to a density of 1013

Compton suppression spectrometry

Measurement of the n_tof beam profile in the second experimental area (EAR2) using a silicon detector

K 40 activity and Detector Efficiency

Physics Dept. PHY-503-SEMESTER 112 PROJECT. Measurement of Carbon Concentration in Bulk Hydrocarbon Samples

Elastic Recoil Detection Method using DT Neutrons for Hydrogen Isotope Analysis in Fusion Materials. Abstract

Data Acquisition. What choices need to be made?

Today, I will present the first of two lectures on neutron interactions.

Scintillation Detector

Transcription:

Determination of the boron content in polyethylene samples using the reactor Orphée F. Gunsing, A. Menelle CEA Saclay, F-91191 Gif-sur-Yvette, France O. Aberle European Organization for Nuclear Research (CERN), Switzerland January 31, 2017 Abstract CERN-OPEN-2017-010 31/01/2017 The boron content of two unknown types of polyethylene has been determined relative to a known reference type. Samples of polyethylene, including a known boron-less one, were irradiated with thermal neutrons at the reactor Orphée at Saclay in France. Prompt gamma rays were measured with a CeBr 3 detector and the intensity of the 478 kev line from the 10 B(n,α 1 γ) 7 Li* reaction was extracted for each sample. Introduction Borated polyethylene is widely used as neutron shielding material. It is commercially available, usually with quantities of natural boron up to several percent. The isotopic composition of natural boron is 19.8% of 10 B and 80.2% of 11 B. Neutrons are moderated by elastic scattering from the hydrogen nuclei present in polyethylene and are subsequently absorbed by 10 B through the 10 B(n,α) reaction. With a probability of 0.937, the 7 Li nucleus is left in its first excited state by the 10 B(n,α 1 γ) 7 Li* reaction, with a cross section of about 3600 b at 25.3 mev. The excited 7 Li* and decays to its ground state by emitting a 478 kev gamma ray. The other possibility, with a probability of 0.063, is that the 7 Li nucleus is left in its ground state via the 10 B(n,α 0 ) reaction, with a cross section of about 240 b at 25.3 mev). No gamma radiation is emitted in this case. The neutron time-of-flight facility n TOF at CERN uses neutron shielding at various places along the two neutron beam lines. Sizeable quantities of two types of Technical Report: IRFU-17-05 / CERN-OPEN-2017-010 1

Figure 1: The layout of the neutron beam lines of the reactor Orphée [1,4]. Experimental area G4-3 was used for the boron determination experiment. polyethylene were localized in a storage area, but with unknown characteristics, in particular the possible boron content. In order to determine the boron content of the two unknown types of polyethylene relative to a known reference sample of borated polyethylene, an experiment was carried at the reactor Orphée [3] of the Laboratoire Léon Brillouin in Saclay, France. The goal was to detect the intensity of the 478 kev gamma ray emitted after neutron absorption by 10 B present in the polyethylene, relative to a known standard. Samples of polyethylene, including a reference sample with the same dimensions, were placed in the same position in a subthermal neutron beam. The emitted prompt gamma rays during irradiation were measured with a portable CeBr 3 detector. The measured intensity of the 478 kev gamma ray was then used to determine the boron content relative to the reference sample, similar as for example in ref. [2]. 1 The measurement setup The reactor Orphée [1, 3] is operated by the Laboratoire Léon Brillouin, a research institute jointly governed by the Commissariat à l Energie Atomique et aux Energies Alternatives (CEA) and the Centre National de la Recherche Scientifique (CNRS). The reactor is a 14 MW light-water moderated reactor with a compact core of highly enriched 235 U providing a flux of up to 3 10 14 ther- 2 Technical Report: IRFU-17-05 / CERN-OPEN-2017-010

Figure 2: One of the tested setups for the measurement. A white polyethylene sample is positioned in the neutron beam, shape by the final borated aluminum collimator. For the final measurement, the distance of the standalone Iris CeBr3 detector system was with respect to the sample was increased. mal neutrons cm 2 s 1, in the surrounding heavy water reflector tank. Neutron beams from 9 horizontal neutron beam lines are pointing tangentially towards the core, and from 6 curved neutron guides, are routinely used to study a large variety of solid state physics cases by neutron scattering for wavelengths ranging from 0.7 to 15 A. The layout of the beam lines [1, 3] is shown in figure 1. Even if neutron-induced gamma-ray spectroscopy is not part of the usual activities at Orphe e, it was possible to use one of the scattering setups which could be used with minor modifications. We have used experimental area G4-3 with the neutron beam deviated with a monochromator from neutron guide G4, resulting in a beam of monochromatic neutrons with a wavelength of 0.236 nm (14.7 mev) and a neutron flux of about 1.5 106 n cm 2 s 1 over an area of approximately 3 3 cm2. The neutron beam was shaped with the help of collimators made of boral, aluminium with a high concentration of boron carbide. The polyethylene samples were placed in the neutron beam using an aluminum sample holder, allowing to reproduce the geometrical position of each sample. Different configurations were tried changing the positions of the CeBr3 detector in order to minimize the activation of the crystal due to sample-scattered neutrons and to optimize the count-rate from the sample-emitted gamma rays in the detector. In figure 2 one of the tested setups is shown. In the setup used for the final measurements, the detector was placed further away from the sample. Technical Report: IRFU-17-05 / CERN-OPEN-2017-010 3

2 Polyethylene samples Several disc-shaped samples with a diameter of 60 mm and a thickness of 5 mm were produced for each of the 4 types of polyethylene: two types containing an unknown quantity of boron, labelled A and B, a reference sample R containing a known quantity of boron, and a sample of standard polyethylene with zero boron content labelled Z. The aspect of the reference sample R is white, softer than the other samples, and with visible, small dark particles in the material. The reference polyethylene sample was made from bulk material purchased from John Caunt Scientific Ltd., containing 5% of natural boron. The unknown type A and zero boron type Z polyethylene both have a clear white color and feel very smooth. The type B has a black color and is smooth as well. For each sample type the mass was measured and the density calculated as reported in table 2. Table 1: Properties of the different polyethylene samples, all of diameter 60 mm and thickness 5 mm. The uncertainty on the density is mainly due to the uncertainty of the thickness and is estimated as less than 2%. label color mass density (g) (g/cm 3 ) R white 14.55 1.03 A white 13.79 0.98 B black 13.42 0.95 Z white 13.33 0.94 3 Gamma-ray detector and calibration For the gamma-ray spectroscopy we have used a 2x2 inch CeBr 3 detector system Iris from Itech Instruments [5], which combines an on-board multi-channel analyzer (MCA), screen and high voltage supply into a portable device. The device was kindly put at our disposal for the measurements by the radioprotection unit of Orphée. Given the expected gamma-ray spectra, the resolution of the detector was sufficient for our purposes. The MCA registered gamma-rays up to about 2.4 MeV into 1024 channels. In order to characterize the detector, it was brought in the vicinity of a strong 192 Ir source. This source has several gamma ray rays, including the triplet at 296.0, 308.5 and 316.5 kev, a peak at 468.1 kev, and the triplet at 588.6, 604.4 and 612.5 kev. These energies cover the full energy range of interest for the present measurement. In figure 3 the measured spectrum of this source is shown. We fitted the three regions with closely lying peaks separately, while for each region fixing the known intensities and assuming a constant efficiency over the multiplet. An 4 Technical Report: IRFU-17-05 / CERN-OPEN-2017-010

example of the fit of the first region and the resulting energy calibration is also shown in figure 3. This energy calibration was found to be in agreement with a linear calibration using the single 662 kev gamma ray of a 137 Cs source, which was used for the final measurements. 4 Measurement results We measured the spectra of each of the 4 types of polyethylene, A, B, R and Z, as well as a spectrum with an empty sample position with beam and the ambient background without beam. Each spectrum was taken with an irradiation time of 180 seconds. The reactor power and hence the neutron flux was considered constant over the measurements. The resulting measured spectra are shown in figure 4. We see that the 478 kev gamma ray is also present in the measurement with the sample without boron (Z), due to the use of boral in the beam collimation. The 478 kev is has the same magnitude for the samples B and Z, from which we can already conclude that sample B does not contain boron. From this figure can also be observed that the intensity of the 559 kev peak and several other gamma rays, reduces when the the boron content in the sample is higher. This may be explained by neutron-induced reactions elsewhere in the experimental area by neutrons from the beam that are not absorbed by the boron, but scattered from the sample due to the high hydrogen content in polyethylene. In the same way, the intensity of the 511 kev annihilation peak is lower when the boron contents is higher. As an order of magnitude, the transmission for the 5% boron-loaded sample is about 0.3 due to boron only, where the main interaction occurs through the (n,α) reaction, and another transmission factor of 0.3 due to hydrogen, reacting mainly by (n,n) scattering, but also with a tiny part (1%) through (n,γ) creating gamma rays of 2223 MeV, visible in figure 4. Figure 3: The measured spectrum of the 192 Ir source (left panel), a zoom on the first triplet together with the fitted Gaussians with a fixed ratio of intensities (middle panel), and the resulting channel to energy calibration (right panel). Technical Report: IRFU-17-05 / CERN-OPEN-2017-010 5

Figure 4: The measured spectra of all samples, corresponding to an irradiation time of 180 seconds, shown as the full energy region (upper panel) and a zoom in the region containing the 478 kev peak from boron (lower panel). 6 Technical Report: IRFU-17-05 / CERN-OPEN-2017-010

Figure 5: A zoom of the measured spectra in the region of the 478 kev peak together with a fit of three gamma-rays from which the peak areas are determined, for each of the four samples A, B, Z, and R. From sample A the boron contents has to be determined relative to the reference sample R from the intensity of the 478 kev peak. This can be done either by fitting the subtracted spectra A-Z and R-Z, or by fitting the intensities I A, I Z, and I R of the 478 kev peak separately to obtain the ratio f as: f = I A I Z I R I Z (1) where the ratio f represents the fraction of boron relative to the known reference sample. In figure 5 the data and the fit with three Gaussian peak profiles are shown. From the peak areas the ratio f A for sample A can be calculated resulting in f A = 0.305 ± 0.010 (2) The quoted uncertainty is only due to the results of the fit, taking into account the correlation of the uncertainty on I Z in f. We already concluded that f B = 0. Technical Report: IRFU-17-05 / CERN-OPEN-2017-010 7

Conclusion From the present measurements we have determined the ratio f of boron content from two unknown samples relative to a known reference sample. Table 2: The ratio f of boron in the unknown samples A and B, relative to the reference sample R. Assuming a content of 5% in the reference sample, the nominal boron content is also given. sample ratio f nominal boron content (%) A 0.305 ± 0.010 1.5 B 0.0 0 The present measurement was performed with a portable CeBr 3 detector. Assuming a boron content of 5% in the reference sample, we determined a boron concentration of 1.5% in sample A and 0% in sample B. References [1] http://www-llb.cea.fr [2] T. Matsumoto and O. Aizawa, Appl. Radiat. Isot. 41 (1990) 897-903 [3] Ch. Alba-Simionesco, A. Menelle and J.-P. Visticot, Neutron News 22 (2011) 10-13 [4] P. Monceau and M. Alba, Neutron News 14 (2003) 10-12 [5] http://itech-instruments.com 8 Technical Report: IRFU-17-05 / CERN-OPEN-2017-010