Characterization of the Lanthanum Chloride Scintillation Detector

Similar documents
New Scintillators for the Border Monitoring Equipment

Design of a Lanthanum Bromide Detector for TOF PET

Lawrence Berkeley National Laboratory Lawrence Berkeley National Laboratory

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

Gamma Spectroscopy. References: Objectives:

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Fast inorganic scintillators - status and outlook -

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Inorganic Scintillators

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Improved scintillation proportionality and energy resolution of LaBr 3 :Ce at 80K

Scintillation Products Technical Note

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

Chapter 4 Scintillation Detectors

Neutron Response of LaBr 3 :Ce and CsI:Na Phoswich Crystal Scintillators for PARIS

Timing and Energy Response of Six Prototype Scintillators

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Non-proportional response and energy resolution of pure SrI 2 and SrI 2 :5%Eu scintillators

Crystals for the HHCAL Detector Concept Rihua Mao, Member, IEEE, Liyuan Zhang, Member, IEEE, and Ren-Yuan Zhu, Senior Member, IEEE

OPPORTUNITY TO JOIN IEEE AND NPSS

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 55, NO. 4, AUGUST /$ IEEE

Scintillation Detector

Pulse height non-linearity in LaBr 3 :Ce crystal for gamma ray spectrometry and imaging

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

The Scintillation properties of Pr 3+ doped and Pr 3+, Ce 3+

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Timing resolution measurement of a 3 Lanthanum Bromide detector

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

INORGANIC crystal scintillators are widely used in high

Scintillation Detectors

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums

Precision Crystal Calorimeters in High Energy Physics: Past, Present and Future

2.5 Physics of the Universe, Astrophysics, Nuclear Planetology Dark Matter and Double Beta Decay Study Planetary Nuclear

Preparatory experiments for cold-neutron induced fission studies at IKI

Precision Crystal Calorimeters in High Energy Physics: Past, Present and Future

Journal of Asian Scientific Research, 1(3),PP

Radiation Dose, Biology & Risk

A new timing model for calculating the intrinsic timing resolution of a scintillator detector

arxiv: v2 [physics.ins-det] 8 Feb 2013

Precision Crystal Calorimetry in High Energy Physics

Dual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT

DETECTORS. I. Charged Particle Detectors

Recent advances and future perspectives of gamma imagers for scintimammography

Platinum resistance. also wirewound versions. eg

Quality Assurance. Purity control. Polycrystalline Ingots

Measurements of CsI(Tl) Crystals with PMT and APD. ipno.in2p3.fr Jean Peyré Milano - October 2006

Radiation Detection and Measurement

Light yield non-proportionality and energy resolution of Lu 1.95 Y 0.05 SiO 5 :Ce and Lu 2 SiO 5 :Ce scintillation crystals

IN the last decade, cerium doped silicate based heavy

Characterization of the new scintillator Cs 2 LiYCl 6 :Ce 3+ *

Detection and measurement of gamma-radiation by gammaspectroscopy

GAMMA RAY SPECTROSCOPY

Europium-doped barium halide scintillators for x-ray and -ray detections

Sample Spectroscopy System Hardware

The fast light of CsI(Na) crystals

Chapter 6: Basic radiation detectors

Investigation of Ce-doped Gd2Si2O7 as a scintillator. Author(s) Kurashige, Kazuhisa; Ishibashi, Hiroyuki; Furusaka, Instructions for use

Chemical Engineering 412

Radioactivity. Lecture 6 Detectors and Instrumentation

NUCL 3000/5030 Laboratory 2 Fall 2013

Scintillators General Characteristics

Survey of the performance of scintillation materials at low temperatures

arxiv: v1 [physics.ins-det] 29 Jun 2011

REFERENCE SOURCES FOR THE CALIBRATION OF THE AUTOCORRELATION SINGLE-CRYSTAL SCINTILLATION TIME SPECTROMETER

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

3. Perturbed Angular Correlation Spectroscopy

LANTHANUM BROMIDE SCINTILLATION DETECTOR FOR GAMMA SPECTROMETRY APPLIED IN INTERNAL RADIOACTIVE CONTAMINATION MEASUREMENTS

Gamma ray coincidence and angular correlation

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Week 6: Ch. 8 Scintillation Counters

LSO/LYSO Crystal Development Ren-yuan Zhu

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET

Characterization and Monte Carlo simulations for a CLYC detector

New Photonis XP20D0 photomultiplier for fast timing in nuclear medicine

Light yield non-proportionality and intrinsic energy resolution of doped CsI scintillators

CHAPTER 6. Faculty of Applied Sciences, Delft University of Technology, Delft, Netherlands Radiation detectors complexity and relevance

Scintillating Crystals and their Applications in Particle and Nuclear Physics November 17-18, 2003

Application of Nuclear Physics

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY

DELAYED COINCIDENCE METHOD FOR PICOSECOND LIFETIME MEASUREMENTS

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

4.1. The Positron Sources

EEE4106Z Radiation Interactions & Detection

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2

Gamma and X-Ray Detection

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

Determination of the shielding power of different materials against gamma radiation

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

PARTICLES REVELATION THROUGH SCINTILLATION COUNTER

Radiation (Particle) Detection and Measurement

A new scintillator detector for nuclear physics experiments: the CLYC scintillator

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations

Stability and Characteristics of Large CZT Coplanar Electrode Detectors

Transcription:

Characterization of the Lanthanum Chloride Scintillation Detector François Kazadi Kabuya 1*, Zslot Podolyak 2 1 Commissariat Général à l Energie Atomique, PO BOX 868 Kinshasa XI, DR Congo 2 Department of Physics, University of Surrey, Guildford, GU2 7XH, United Kingdom Abstract. This paper reports about the investigations on the performance of the new earth-rare halide Scintillator, the LaCl 3 :10%Ce, which has been discovered with attractive scintillation properties: fast, efficient, and high-energy resolution. The combination of good brightness and linear response has naturally resulted in a very good energy resolution, whereas the fast decay time promoted counting applications at very high rate and very fast timing resolution. Energy resolutions of (2.99±0.02) %, (8.66±0.02) %, and (2.05±0.01) % have been achieved by exciting the detector with 662 kev 137 Cs, 122 kev 152 Eu, and 1332.5 kev 60 Co sources respectively, at room temperature. The variations of the resolution as a function of both shaping time and high voltage have been studied and found to be relatively linear. The timing resolution was also studied. The value of 1.07 ns FWHM has been recorded with LaCl 3 :10%Ce and BaF 2 detectors operating in coincidence mode, using 1333.5 kev γ-ray 60 Co peak. Compared to similar studies, this timing resolution was worse than expected even if the exciting radioactive sources used in the two experiments were different. The efficiency values of (0.24±0.01)%, (0.01±0.01)% and (0.09±0.01)% were obtained with laboratory 137 Cs, 152 Eu, 60 Co isotopes respectively. These detector efficiency values were very low, due to the smaller detector sizes used in these earlier crystals and to the large detector-source-distance used in this experiment. The Lanthanum Chloride Scintillator detector is hygroscopic and relatively expensive compared to established scintillators. In addition, it carries internal contamination in 138 La and 227 Ac for larger thicknesses. Nevertheless, the energy resolution figures achieved, twice as good as that of NaI(Tl), are in good agreement with those reported elsewhere. They credit the scintillation detectors as good challengers of semiconductor-based detectors, allowing their application fields to be extended. KEYWORDS: Scintillator, Energy Resolution, LaCl 3 :Ce, Gamma ray spectrometry 1. Introduction It was in November 1895 that W.C. Roentgen discovered the X-ray [1]. The use of this new and penetrating radiation found more applications with the development of the luminescent materials, which efficiently absorb X-rays (Phosphors). The discovery of the scintillator material was undertaken in three successive main phases [2]. The first and the second phases included experiences with the earliest scintillators CaWO 4 used by Becquerel, Crookes and Rutherford; discovery of scintillation in naphthalene and development, by Hofstadter, of the thallium-activated NaI. The third phase, corresponding to the two past decades, witnessed the renaissance in research and development of scintillator materials. This renewed interest to scintillator materials was motivated by two major applications in nuclear physics and medical imaging: the construction of the electromagnetic calorimeter of the CMS at Large Hadron Collider (LHC) and the increasing demand in Positron Emission Tomography (PET). Scintillation detectors are also widely used in the detection and spectroscopy of ionizing radiations. Common applications of scintillation spectrometers include non-destructive testing, nuclear nonproliferation, environmental monitoring, and X-ray diffraction. In these different applications, detector requirements vary considerably accordingly. In γ-ray counting and spectroscopy for example, the light yield, scintillation decay, and energy resolution are considered to be critical properties [3]. Today, Research and Development to introduce new scintillator materials has focused on better understanding of and improving the properties of two particular families of scintillator materials: (1) lanthanum halides doped-cerium [4] (LnX 3 :Ce 3+, with Ln = La, Gd, Lu and X = F, Cl, Br, I) and (2) PbWO4 [5] or Lu 2 SiO 5 (LSO) doped with Cerium. In the lanthanum halides based on the application of cerium Ce 3+ as a luminescence centre, the allowed 5d-4f transition makes possible the response * Presenting author, E-mail : kabuyafr@yahoo.ca

times in the range of 10 ns to 100 ns. In addition to the ability of incorporating the Ce 3+ ions, the scintillators with high density and Z and a relatively small gap between the valence and the conduction bands, with potential to realize > 50 000 photons per MeV [6] were considered. The Cerium-doped lanthanum chloride (LaCl 3 :Ce) scintillator was discovered by Delft and Bern Universities [7]. Since 2003, it is commercially made available by Saint-Gobain Crystals with a trade name of BrilLance 350 (B350) crystals. The new scintillation detector has been extensively studied and found to show excellent luminescence and scintillation properties, superior to that for NaI(Tl) [8]. This paper investigates this new scintillation detector and addresses the essential question of how the energy resolution of an inorganic scintillator can be improved and, consequently, the scintillation applications can be extended. In view of its attractive properties, study and characterization of the LaCl 3 :(10%)Ce detector have been carried out for its applicability to gamma-ray spectroscopy. 2. Experimental and method 2.1. Experimental set-up The experimental set-up is shown figure 1.In table 1 properties of the investigated LaCl 3 :Ce detector are compared to other scintillators. Figure 1: BaF 2 scintillation detector (head to head) with LaCl 3 :Ce detector and spectroscopy chains. Table 1: Principal properties of the LaCl 3 :Ce against common Scintillators [9] Scintillators Light Yield (Photons/keV) Decay Time (ns) FOM * (t/ly) 1/2 Wavelength of maximum emission λm (nm) Refractive index at λm Density (gcm -3 ) Thickness (cm) for 50% attenuation (662 kev) NaI(Tl) 38 250 2.6 415 1.85 3.67 2.5 B350 49 28 0.8 350 ~ 1.9 3.79 2.4 B380 63 16 0.5 380 ~ 1.9 5.29 1.8 BaF 2 1.8 6.7 0.6 ~ 210 1.54 4.88 1.9 *Figure Of Merit LY = Light Yield 2.2. Method and Equipment 2.2.1. Energy resolution measurements Gamma-ray energy resolution of the LaCl 3 :Ce scintillation detector was measured using the experimental set-up on fig. 1. The detector was supplied with the Hamamatsu R6231 Mod PMT. The detector sample was irradiated with 661.54 kev γ-ray ( 137 Cs source at a detector-source-distance of approx. 10 cm) and the resulting signal was processed with the ORTEC 113 Preamplifier, shaped with 2

the spectroscopic amplifier ORTEC 572 and the pulse displayed through the MCA. Measurements were performed at room temperature, with the high voltage set at 750V and the shaping time at 6μs. The influence of both the high voltage and the shaping time on the resolution (FWHM) was studied through a series of other experiments. 2.2.2. Coincidence Timing resolution measurements Coincidence timing resolution for LaCl 3 :Ce detector was measured using the experimental set-up in figure 2. Figure 2: Experimental set-up for Coincidence Timing Measurements Two BaF 2 detectors, one in the START channel and the other in the STOP channel, were operated at -2400V in the first configuration set-up. Upon irradiation of each crystal with 1333.52 kev γ-rays 60 Co source, the signal from the Anode of each PMT was processed by an ORTEC-935 Quad Constant Fraction Discriminator (CFD). The delay in the STOP signal was set at 32 ns by using the ORTEC- 424A Delay Output. The resulting timing signals start and stop a Time to Amplitude Converter (ORTEC-TAC/SCA Model 567) whose the output was digitized and processed by the MCA through the Analog Digital Converter (ADC) card to produce a timing spectrum. The BaF 2 detector in the STOP channel was then replaced with the LaCl 3 :Ce detector in the second configuration set-up. 3. Results 3.1. Energy resolution Figure 3 shows the spectra recorded with BaF 2 and LaCl 3 :Ce detectors in set-up configuration of figure 1. Figure 3: showing the two spectra recorded with the two detectors irradiated with 152 Eu, 60 Co and 137 Cs sources during calibration procedure 3

Energy resolutions and uncertainties for both lanthanum chloride and Barium detectors are summarized in table 2. Table 2: Energy resolution for 137 Cs, 60 Co and 152 Eu sources Source Energy (kev) FWHM (kev) Resolution (%) Log E Log R 137 Cs 661.54 19.74 2.99±0.02 2.82 0.47 60 Co 1173.2 13.39 2.32±0.01 3.06 0.37 1332.5 13.57 2.05±0.01 3.12 0.31 152 Eu 121.8 10.62 8.66±0.02 2.08 0.93 344.3 18.19 5.23±0.01 2.53 0.73 The response of the lanthanum chloride detector to variations of both the high voltage and the shaping time are displayed figures 4 and 5, respectively. Figures 4 and 5: Response of the lanthanum chloride detector against variations of HV and shaping times with 137 Cs source 4

3.2. Coincidence Timing Resolution In figure 6 are shown spectra and results achieved following the set-up configuration of figure 2. Figure 6: Coincidence timing resolution for (a) BaF 2 - BaF 2 detectors and (b) LaCl 3 :Ce in STOP channel and BaF 2 in START channel 4. Discussions and Conclusions Experimental results and calculations provided energy resolution values achieved with LaCl 3 :Ce detector and shown in table 2. Apart from the 152 Eu spectrum, which showed relatively high energy resolution, LaCl 3 :Ce detector displayed excellent energy resolution figures. The figure of (2.99±0.02)% achieved in these experiments is three times better than that obtained with the BaF 2 detector. It is in good agreement with different values of energy resolution reported elsewhere for 662 kev 137 Cs source [10,11,7]. Resolution results with LaCl 3 :Ce are also good using 60 Co spectrum for 1173,2 kev peak. By achieving these results, inorganic Scintillators realize their best performance ever reported in the literature [7,10,11] and seriously are challenging semiconductor detectors such as CdTe and CdZnTe [12,10]. 5

Furthermore, using a high voltage of 735 V, the resolution value of 3.6 %( 662 kev) has been achieved by the manufacturer upon delivery of the detector. In order to study the relationship between the FWHM and HV, this figure (735V) was compared with the results achieved in our experiments. Detector performance was also studied through a set of measurements, including the relationship between FWHM and shaping times as illustrated in figures 4 and 5. Finally, what really explain the best performance (in terms of energy resolution) achieved with the LaCl 3 :Ce detector? It results from the graph plotted in figures 4 and 5 that resolution (FWHM) can be affected inversely by changing both high voltage and shaping time values. Using a value of 750V, the optimum operating point for this detector was reached even if the best resolution figure is achieved with higher HV values (850 V in these experiments). This figure is in agreement with the one (735V) achieved by the manufacturer during the calibration procedure. On the other hand, the best value of energy resolution (between 2.5% and 3%) is achieved with an optimum shaping time of 3μs. This dependence between these two quantities (HV and shaping times) and the energy resolution suggests the advantage of selecting optimum operating parameters of the detector during experiments. With regard to the energy resolution performance, for a typical scintillation detector, there are many factors that can contribute to achieving a good energy resolution. From the development viewpoint, the light output and the number of electron-holes (N phe ) pairs produced, and the detector size, are the most important factors to be considered. In particular for the LaCl 3 :10%Ce, resolution is also dependent on the dopant concentration [6, 11]. In halide scintillators, in particular with chloride compounds, a higher doping concentration (10% Ce) allowed to achieve a high light yield (~ 49000 photons/mev). In addition, the fundamental limit on the light yield is governed by the band gap and the value of β [13] (conversion efficiency for creating e-h pair). The light yield being proportional to Nphe, a good energy resolution requires a high light output. Those difficulties have been overcome by the incorporation of the Ce 3+ ion as a luminescence centre in the lattice material and also by the quality of the host material. Those two technological achievements allow a relatively small band gap between the valence and the conduction bands. However, the energy resolution is maximal when the value of the combined effect of intrinsic and transfer resolutions or the contribution of the non-proportionality effect is small [14]. Figure 6 provides a comparison for the coincidence timing resolution measurements performed with 1333.2 kev 60 Co peak for both Barium Fluoride and lanthanum Chloride detectors. Timing resolution value of 435 ps (FWHM) was achieved in BaF 2 - BaF 2 detectors configuration. In BaF 2 -LaCl 3 :Ce detectors set-up, the value of 1.07 ns FWHM was measured. In similar study [15], where detectors were irradiated with a 662 kev source, values of 273 ps HWHM and 264 ps FWHM were found for both BaF 2 and LaCl 3 :Ce detectors respectively. These values are up to 4 times better than those achieved in the present experiment for the same detectors and configurations. Compared to BaF 2 detector, which considered to be the benchmark for fast timing measurements (with ~ 0.6 ns), LaCl 3 :Ce detector is hence 4 times less faster. However, it should be noticed that both experiments were carried out using different radioisotope sources with different energies and different acquisition times. The efficiency results obtained are as close as possible to those displayed by conventional scintillation detectors. Absolute efficiency values are low. This is due the small detector thickness (2.4 cm). On the other hand, the intrinsic efficiency values obtained in these experiments are good and in agreement with those of most established scintillators. Nevertheless, scintillation detectors are known to be more efficient in detecting γ-rays than semiconductor-based detectors. The performance expected from the LaCl 3 :Ce detector being achieved, what now the future of inorganic scintillators looks like? With 70000 photons per MeV and a resolution of 2.8% and a decay time of 25 ns, the Cerium-activated Lanthanum Bromide detector (LaBr 3 :Ce 3+ ), is the best scintillator for the moment. However, the demand for better ones is always growing. The improvements to be addressed include the density of the host material, the light output and the related energy resolution. A record of high phonon yield approaching 95000 Photons/MeV has been already achieved with the LuI 3 :Ce 3+ scintillator [6]. The decay time of 24 ns was reached and the density value of 5.6 gcm -3 6

realized is also higher than that of LaBr 3 :Ce 3+. As mentioned above, with smaller band gap scintillator materials, high light output can be achieved. Therefore, the area of ultimate interest could be found among compounds presenting forbidden band gap energy Eg of about 3 ev. In this regards, potential candidates can be found among iodides, selenides and tellurides. With such materials, an ultimate energy resolution below 2% for 662 kev γ-rays can be predicted with Ce 3+ scintillators. That, of course inaugurates future fields of research. Acknowledgements This work was carried out thanks to the IAEA fellowship grant (ZAI/06001). The authors wish to thank the International Atomic Energy Agency (IAEA) and the government of the Democratic Republic of the Congo, through the authorities of the Congo Atomic Energy Commission (CGEA). 5. References [1] HOFSTADTER R., Alkali halide scintillation counters, Phys. Rev., 74 (1948) 100 [2] WEBER M.J., Inorganic Scintillators: Today and Tomorrow, J. Lumin., 100 (2002) 35-40 [3] VAN LOEF E.V.D. et al, Non-proportionality and energy resolution of a LaCl 3 :10%Ce 3+ scintillation crystal, IEEE Trans. on Nucl. Science, vol. 50, N 1, Feb. 2003. [4] Van LOEF E.V.D, Halide scintillators, Thesis, University of Technology, Delft (2003) [5] AMNENKOV A.A., KORZHIK M.V. and LECOQ P., Lead Tungstate scintillator material, Nucl. Inst. Meth. A, 490 (2002) 30. [6] KRAMER K.W., et al., Development and characterisation of highly efficient new cerium doped rare earth scintillator materials, J. Mater. Chemistry 16 (2006) 2773-2780 [7] VAN LOEF E.V.D. et al, High energy resolution scintillator: Ce 3+ -activated LaCl 3, Appl. Phys. Letter, 77, (2000) 1573 [8] ILTIS A. et al, Lanthanum halide scintillators: Properties and Applications, Nucl. Instr. and Meth. A, 563 (2006) 359 363. [9] ROWSA C.M et al, Performance Summary: BrilLance Scintillators LaCl 3 : Ce and LaBr 3 :Ce, Scintillator Products Whitepaper, saint-gobain Crystals, 2006 [10] SHAH, K.S. et al., High Energy resolution scintillation spectrometers, IEEE Trans. Nucl. Sc., 51, 5, October 2004 [11] VAN LOEF E.D.V., et al., Scintillation Properties of lacl 3 :Ce 3+ crystals: Fast, Efficient, and High- Energy resolution scintillators, IEEE Trans. on Nucl. Science, 48 3, June 2001 [12] ARYAEINEJAD R., HARTWELL and WADE W.S., High-resolution Compton-Suppressed CZT and LaCl 3 :Ce detectors for fission products identification, Ieeexplore.org/iel5/ [13] DORENBOS P., Light output and energy resolution of Ce 3+ -doped scintillators, Nucl. Nucl. and Meth. A 486 (2002) 208-213. [14] VAN EIJK C.W.E., New inorganic scintillators aspects of energy resolution, Nucl. Instr and Meth. A 471 (2001) 244 248. [15] SHAH K.S. et al., LaCl 3 :Ce scintillator for γ-ray detection, Nucl. Instr. Meth. Phys. Res. A 505 (2003) 76-81 7