CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital.

Similar documents
Radionuclide Imaging MII Positron Emission Tomography (PET)

6: Positron Emission Tomography

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links

Acknowledgements. PET Fundamentals: Ideal Case. Why Are We Excited About PET/CT? PET Fundamentals: Real Case

PET. Technical aspects

Chapter 2 PET Imaging Basics

Solid State LightBurst New PET Technology GE PET/CT and PET/MR


Tomography is imaging by sections. 1

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

PET scan simulation. Meysam Dadgar. UMSU, Iran. IFMP, Elbasan, Fig 1: PET camera simulation in gate by cylindrical phantom

Lecture 5: Tomographic nuclear systems: SPECT

A Brief Introduction to Medical Imaging. Outline

Radioisotopes and PET

A. I, II, and III B. I C. I and II D. II and III E. I and III

Mitigation of External Radiation Exposures

Quantitative Imaging with Y-90 Bremsstrahlung SPECT (Single Positron Emission Computed Tomography)

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

The Physics of PET/CT scanners

Positron Emission Tomography

Positron Emission Tomography

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or

Welcome back to PHY 3305

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Technical University of Denmark

Mayneord-Phillips Summer School St Edmund Hall, University of Oxford July Proton decays to n, e +, ν

Dana-Farber Cancer Institute, 44 Binney Street, Boston, MA 02115, USA ramsey

Simulation of triple coincidences in PET

Development of Radioactivity Standards for Quantitative Positron Emission Tomography

Gamma ray coincidence and angular correlation

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging

Technical University of Denmark

Introduction to Medical Imaging. Medical Imaging

Positron Annihilation in Material Research

Nuclear Medicine RADIOPHARMACEUTICAL CHEMISTRY

Structure of Biological Materials

Procesamiento de Imágenes y Bioseñales

Bases of radioisotope diagnostic methods

Positron Emission Tomography (PET)

APPLIED RADIATION PHYSICS

CLINICALLY USEFUL RADIONUCLIDES:

Application of Nuclear Physics

Radiochemistry and Radiopharmacy III

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras?

III. Proton-therapytherapy. Rome SB - 2/5 1

Number of protons. 2. What is the nuclear symbol for a radioactive isotope of copper with a mass number of 60? A) Cu

Ba (Z = 56) W (Z = 74) preferred target Mo (Z = 42) Pb (Z = 82) Pd (Z = 64)

Radiation Detection and Measurement

Physics in Nuclear Medicine

Radiation Quantities and Units

Properties of the nucleus. 9.1 Nuclear Physics. Isotopes. Stable Nuclei. Size of the nucleus. Size of the nucleus

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx

Properties of the nucleus. 8.2 Nuclear Physics. Isotopes. Stable Nuclei. Size of the nucleus. Size of the nucleus

Two-Material Decomposition From a Single CT Scan Using Statistical Image Reconstruction

Nuclear Physics and Astrophysics

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Interaction of charged particles and photons with matter

Low Energy Medical Isotope Production. Naomi Ratcliffe IIAA, University of Huddersfield UK

CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration

Year 12 Notes Radioactivity 1/5

Nuclear Medicine: Physics and Imaging Methods (SPECT and PET)

β and γ decays, Radiation Therapies and Diagnostic, Fusion and Fission Final Exam Surveys New material Example of β-decay Beta decay Y + e # Y'+e +

(INCLUDING THIS FRONT PAGE)

Nuclear Chemistry. Background Radiation. Three-fourths of all exposure to radiation comes from background radiation.

12/1/17 OUTLINE KEY POINTS ELEMENTS WITH UNSTABLE NUCLEI Radioisotopes and Nuclear Reactions 16.2 Biological Effects of Nuclear Radiation

Sample Spectroscopy System Hardware

Part III Minor Option in Medical Physics 2018 Examples Sheet

Initial Studies in Proton Computed Tomography

Rad T 290 Worksheet 2

THE ACTIVITY CALIBRATOR

Dosimetry of patients injected with tracers Ga-68, Zr-89 and Lu-177. Bruno Vanderlinden

Wednesday 23 January 2013 Afternoon

Neural Network Approach for Photon-counting Detection The First Step: PPE Correction

Nuclear Reactions A Z. Radioactivity, Spontaneous Decay: Nuclear Reaction, Induced Process: x + X Y + y + Q Q > 0. Exothermic Endothermic

Nuclear Medicine: Physics and Imaging Methods (SPECT and PET)

Final exam questions ED

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

ELG7173 Topics in signal Processing II Computational Techniques in Medical Imaging

22.56J Noninvasive Imaging in Biology and Medicine Instructor: Prof. Alan Jasanoff Fall 2005, TTh 1-2:30

1. Which of the following statements is true about Bremsstrahlung and Characteristic Radiation?

EXPERIMENTAL DETERMINATION OF SHIELDING REQUIREMENTS FOR PET MEDICAL FACILITIES BRADLEY S. BRINKLEY

Nuclear Radiation. Natural Radioactivity. A person working with radioisotopes wears protective clothing and gloves and stands behind a shield.

Metal Artifact Reduction and Dose Efficiency Improvement on Photon Counting Detector CT using an Additional Tin Filter

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI)

Medical Physics. Nuclear Medicine Principles and Applications

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Differentiating Chemical Reactions from Nuclear Reactions

Novel detector systems for the Positron Emission Tomography

PHYS 3650L - Modern Physics Laboratory

Investigation of the relationship between linear attenuation coefficients and CT Hounsfield units using radionuclides for SPECT

1-D Fourier Transform Pairs

There are three mechanisms by which gamma rays interact with absorber atoms from which two are important for nuclear medicine.

Development of a High Precision Axial 3-D PET for Brain Imaging

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Overview of Nuclear Medical Imaging Instrumentation and Techniques*

INTERACTIONS OF RADIATION WITH MATTER

Oslo Cyclotron Laboratory

PET/MRI Principle, History, and Perspective. Main Imaging Techniques. X-ray Tube. History of X-ray & CT. How to Look inside the Human Body

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter

Transcription:

CT-PET calibration : physical principles and operating procedures Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital Topics Introduction to PET physics F-18 production β + decay and annichilation process γ γ coincidence detection true, random and scatter coincidences the NECR, image quality VS patient dose PET calibration and QC procedures the sinogram normalization and ECF calibration daily QC uniformity, sensitivity

PET POSITRON EMISSION TOMOGRAPHY Positron The positron (e + ) is emitted in the β + decay process: p n + e + + ν e The β + decay is forbidden for a free proton. The process must occur in a nucleus: (A,Z) (A,Z-1) + e + + ν e For example: 18 F 18 O + e + + ν e

Emission Emission (tomography) : the source of radiation is inside the patient: γ γ Transmission (tomography) : the source of radiation is outside the patient: x-ray Tomography Tómos = from Greek = (cross-) section Material from IAEA Educational Course Radiation Protection in Nuclear Medicine

How the F-18 is produced? H - H - acceleration inside the cyclotron (spiral trajectory) stripping e- e- p+ e- e- p+ e- e- H + p+ proton beam Carbon foil p + 18 O n + 18 F FDG synthesis FDG Main applications: Oncology (cancer detection and staging) Cardiology Neurology

FDG administration uptake time β + decays and annnihilation process 18 F 18 O + e + + ν e γ (511 kev) e + e - γ (511 kev)

e + range in the patient tissues e + Radioisotope e + Range (mm) 15 O 8.4 13 N 5.4 11 C 4.2 18 F 2.6 Radioisotope e + E max (kev) e + E mean (kev) T 1/2 (min) e + Range (mm) Decay 15 O 1720 740 2.07 8.4 100% β + 13 N 1190 490 9.96 5.4 100% β + 11 C 970 390 20.4 4.2 99% β + 18 F 635 250 110 2.6 97% β +, 3% E.C. Fermi s Golden Rule i f p i f 2π = M h if 2 dn f de Transition matrix (dynamics) density of final states (phase space)

e + e - annnihilation process and γ γ coincidence detection TRUE COINCIDENCE useful for the image reconstruction γ annihilation LOR γ constant fraction discriminator constant fraction discriminator coincidence window (6 ns) SCALER True, random and scatter coincidences TRUE COINCIDENCE useful for the image reconstruction RANDOM COINCIDENCE generated from two uncorrelated events of annihilation γ annihilation LOR γ wrong LOR constant fraction discriminator coincidence window (6 ns) constant fraction discriminator Compton scattering annihilation SCATTERING COINCIDENCE one (or both) of the two 512 kev photon is scattered by the patient wrong LOR SCALER

True, random and scatter coincidences LSO CT-PET Biograph Pico3D Cylindrical phantom water + F-18 filled 2.00E+0 1.80E+0 1.60E+0 1.40E+0 1.20E+0 cps 1.00E+0 8.00E+0 6.00E+0 4.00E+0 2.00E+0 0.00E+0 0.00 5.00 10.00 15.00 20.00 25.00 activity (mci) TOTAL TRUE SCATTER RANDOMS NECR Random coincidences : RANDOM COINCIDENCE they don t contribute to image they increase the noise 20 50 % of true coincidences How to reduce them? smaller coincidence window less activity C R = 2 τ C S 2 wrong LOR How to take into account? delayed window method

Scatter coincidences : don t contribute to image in the human thorax, 50% or more of the measured data may be scattered. reduce the contrast Compton scattering annihilation SCATTERING COINCIDENCE wrong LOR How to reduce them? 2D instead of 3D less activity How to take into account? scatter modelling and compensation algorithms Conventional Nuclear Medicine (planar, SPECT) N SNR = N N Positron Emission Tomography TRUE EVENTS contribute to SIGNAL and NOISE RANDOM AND SCATTERING EVENTS contribute only to NOISE NECR NECR = T T + S + 2R good LOR from TRUE wrong LOR from scatter 2 T NECR = T + S + 2R wrong LOR from random SNR = T T + S + 2R

IMAGE QUALITY NECR PET phantom for IQ 30 25 20 NEC [kcps] 15 10 10mCi inj. 20mCi inj. 5. 0 0 5 10 15 20 25 30 35 40 (courtesy of CPS Innovation /Siemens) Activity [kbq/ml] patients with: 2.50E+05 [52-62] kg weight and [155-170 cm] height TRUE + SCATTER RANDOM * PNECR 1 2 2.00E+05 3 1.50E+05 PNECR-MAX 4 1.00E+05 5 5.00E+04 0.00E+00 0.00E+00 5.00E+06 1.00E+07 1.50E+07 2.00E+07 2.50E+07 singles rate (cps) 6 7 COUNT-RATE ANALYSIS FROM CLINICAL SCANS IN PET WITH LSO DETECTORS F. Bonutti et al, Radiation Protection Dosimetry, (2008) 200 80 70 MISSING ACTIVITY = 70% (for central beds) 150 100 50 0 1 2 3 4 5 6 7 bed # bed # 1 2 3 4 5 6 7 but: 60 50 40 30 20 10 PNECR (SNR) IMPROVEMENT : 15-20% 0 0.0 0 20 40 60 80 100 % PNECR i mprovement

Attenuation correction Many true coincidences are lost because of scatter (Compton or Rayleigh interactions) absorption (photoelectric interaction) wrong LOR from scatter Scattered but detected scatter contribution to data attenuation of the true LOR intensity Scattered and not detected attenuation of the true LOR intensity How much is the attenuation? @ 511 kev every 7.2 cm of water half radiation intensity is lost: I I 2 I 4 7.2 cm 7.2 cm no AC AC Ge68 phantom Ge68 phantom

CT-PET Attenuation correction - the emitter distribution is distorted. - quantification is not possible (SUV) - lung lesions may not be visible in uncorrected images. Attenuation correction Attenuation factors range from < 5 in the brain to > 150 in the body. In PET the attenuation doesn t depend on the position of the source point along that LOR: 0 1 3 0 2 3 γ γ detection probability ~ γ γ detection probability ~ = 1 3 3 ( l) dl µ ( l) dl 0 1 = µ ( l) dl 0 e µ 2 3 e e = 3 µ l) dl µ ( l) dl µ ( l 0 2 0 e e e ( ) dl

CT-PET Attenuation correction Rescaling the attenuation coefficients µ with energy µ/ρ 0.3 Air Muscle Bone (cm 2 /g) 0.2 0.1 CT PET 0 0 100 200 300 400 500 Energy (kev) PET calibration and QC procedures The sinogram source sinogram 400 350 300 250 LOR rappresentation phi 200 150 100 50 0-1 -0.5 0 0.5 1 d

PET calibration and QC procedures The normalization procedure Ge-68 phantom, 1 2 mci BUCKET BLOCK Example : Siemens Biograph LSO Pico3D number of elementary detectors : (12 buckets) X (12 blocks) X (64 crystals) = 9216 (courtesy of CPS Innovation /Siemens) PET calibration and QC procedures uniform phantom sinogram for a un-normalized system uniform phantom sinogram for a normalized system

PET calibration and QC procedures Calibration Factor (CF) make possible quantifications (SUV) Ge-68 phantom of known activity concentration CF number of counts inside voxel PET calibration and QC procedures Calibration Factor (CF) make possible quantifications (SUV) ROI ROI activity = ROI counts 0.969 x ECF β + p n + e + + ν e 18 F 18 O + ν e (branching ratio = 96.9 %) E.C. p + e - n + ν e (branching ratio = 3.1 %)

PET calibration and QC procedures Calibration Factor (CF) make possible quantifications (SUV) ROI SUV = ROI activity concentration mean (body) activity concentration PET calibration and QC procedures ECF behavior (Emission Calibration Factor) ECF behavior 3.00E+06 2.50E+06 2.00E+06 ECF 1.50E+06 1.00E+06 5.00E+05 0.00E+00 01/01/2004 31/12/2004 31/12/2005 31/12/2006 01/01/2008 31/12/2008 31/12/2009 time (Siemens Biograph LSO Pico3D)

PET calibration and QC procedures After normalization, the software gives the χ 2 If you normalize the system every 1-2 months, typical χ 2 are 0.5 1.5 If you do it less frequently, the χ 2 goes up χ 2 5 4.5 4 3.5 3 2.5 2 1.5 1 0.5 0 χ 2 behavior 0 20 40 60 80 100 days Optimal frequency may depend on the system In our Departement we do normalization at least every 2 months PET calibration and QC procedures The need for normalization come out from the daily QC test For this test, usually a Ge-68 cylindric phantom is used (1-2 mci)

PET calibration and QC procedures After the daily QC test, give a look to the sinograms : PET calibration and QC procedures SUV tests : with a single Ge-68 phantom (we do it monthly) CT PET Ge-68 phantom

PET calibration and QC procedures SUV tests : with two single Ge-68 phantom (we do it every 6 months) CT PET new Ge-68 phantom old Ge-68 phantom 1 2 mci 0.3 0.5 mci PET calibration and QC procedures SUV tests : with two F-18 bottles (we do it yearly) Check the SUV results on the workstationsd using the different available tools new F-18 phantom old F-18 phantom 1 2 mci 0.3 0.5 mci CT PET LEO2 min max averg 1.17 1.26 1.21 0.60 0.68 0.64 1.17 1.26 1.21 0.61 0.68 0.64 1.13 1.26 1.20 0.59 0.69 0.64 1.16 1.28 1.22 0.59 0.69 0.64 LEO3 min max averg 1.13 1.26 1.21 0.59 0.68 0.64 1.15 1.26 1.21 0.59 0.67 0.64 1.17 1.29 1.22 0.59 0.69 0.64 1.16 1.29 1.22 0.59 0.69 0.64 ESOFT min max averg 1.13 1.26 1.20 0.59 0.67 0.64 1.20 1.26 1.22 0.59 0.66 0.65 1.17 1.30 1.22 0.59 0.68 0.64 1.18 1.28 1.22 0.60 0.69 0.65 tool "elissoide" tool "3D isocontour" tool "multiframe polygon" tool "multiframe contour" tool "elissoide" tool "3D isocontour" tool "multiframe polygon" tool "multiframe contour" tool "elissoide" tool "3D isocontour" tool "multiframe polygon" tool "multiframe contour"

PET calibration and QC procedures Uniformity test : with Ge-68 phantom (we do it every 3 months) Uniformity Specification (from CPS/Siemens standards): The measured volume RMS variation will be < 10% NEMA : NU vol < 15% PET calibration and QC procedures Sensitvity test : with Ge-68 phantom (we do it every 3 months) Sensitivity Specification (from CPS/Siemens standards): The measured system sensitivity will be > 725000 counts/sec/µci/ml NEMA : S meas > 95% S ref

Daily : CT calibration and QC procedures X-ray tube check-up and calibration Every 3 months : Noise, constancy and uniformity of CT numbers CT numbers linearity, contrast scale High and low contrast spatial resolution Slice thickness Table accuracy movement Laser alignement Every 6 months : Dose (in head and body CT phantoms) CT-PET calibration and QC procedures Yearly : attenuation correction check Ge-68, no AC Ge-68, AC S.D. average avedev 5.00 167.47 2.99

CT-PET calibration and QC procedures Yearly : CT-PET co-registration check PET/CT UDINE SCAN PROTOCOL 62 mas; 130 kv p CT PET Survey CT X-Ray FUSION scatter correction attenuation correction WB PET: 20 min 370 MBq CT PET γγ γγ e+ e- Annihilation PET Fused PET/CT