Industrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application

Similar documents
FEATURE ARTICLE. The CS-100 Series High-precision Chemical Solution Monitor in the Semiconductor Cleaning Process. Takaaki Yada.

Kentaro INOUE. Introduction. Measurement principle (membrane polarographic method)

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Answers to spectroscopy questions. 1. Consider the spectrum below. Questions a f refer to this spectrum.

GENERAL PHARMACOPOEIA MONOGRAPH

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer

Solar irradiance measurement up to 2500nm with the Arcoptix FT-NIR

MOLEBIO LAB #4: Using a Spectrophotometer

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Effective testing for wafer reject minimization by terahertz analysis and sub-surface imaging

Spectroscopy Primer. for ultraviolet and visible absorbance spectroscopy. by Stephanie Myers Summer 2015

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Application of IR Raman Spectroscopy

Investigating Transition Metal Complexes

UV-VIS Spectroscopy and Fluorescence Spectroscopy (Part 2 of 2)

Basic Digestion Principles

Development and Applications of Ruggedized VIS/NIR Spectrometer System for Oilfield Wellbores

RCA cleaning: SC1, SC2, DHF, BOE, SPM

C101-E111. Talk Letter. Vol.2 February 2009

UV Spectroscopy Determination of Aqueous Lead and Copper Ions in Water

Behavior and Energy States of Photogenerated Charge Carriers

UV3000. Accurate, precise, and portable ambient gas point analyzer

Near infrared reflectance spectroscopy

EXPERIMENTAL DETERMINATION OF SPECTRAL AND ANGULAR DEPENDENT OPTICAL PROPERTIES OF INSULATING GLASSES

The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer s Law.

Chapter 18. Fundamentals of Spectrophotometry. Properties of Light

Infra Red Spectroscopy

MULTICHANNEL NADIR SPECTROMETER FOR THEMATICALLY ORIENTED REMOTE SENSING INVESTIGATIONS

APPLIED SPECTROSCOPY. A Compact Reference for Practitioners. Jerry Workman, Jr. Art W. Springsteen LABSPHERE, INC. NORTH SUTTON, NH.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology

Lecture 0. NC State University

Study of static electricity in wafer cleaning process M. Wada 1a, T. Sueto 1b, H. Takahashi 1c, N. Hayashi 1d, and A. Eitoku 1e

INTRODUCTION The fundamental law of spectrophotometry is known as the Beer-Lambert Law or Beer s Law. It may be stated as: log(po/p) = A

PREPARATION OF LUMINESCENT SILICON NANOPARTICLES BY PHOTOTHERMAL AEROSOL SYNTHESIS FOLLOWED BY ACID ETCHING

Reprinted from February Hydrocarbon

Physical Change - alters the form or appearance of a substance but does not change it into a new, different substance

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Basics of UV-Visible Spectroscopy *

C101-E145 TALK LETTER. Vol. 17

Chapter 3 Stoichiometry

SOME ASPECTS OF PYROMETRY OF "WHITE" OBJECTS V.E.

Lab #12: Determination of a Chemical Equilibrium Constant

Photolithography II ( Part 1 )

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur

CHEM 254 EXPERIMENT 9. Chemical Equilibrium-Colorimetric determination of equilibrium constant of a weak acid

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Spectroscopy tools for PAT applications in the Pharmaceutical Industry

Spectrophotometry. Introduction

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample

Ultraviolet-Visible Spectroscopy

The importance of the on-line oil layer measurement

Danyal Education (Contact: ) A commitment to teach and nurture

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

Taking Fingerprints of Stars, Galaxies, and Other Stuff. The Bohr Atom. The Bohr Atom Model of Hydrogen atom. Bohr Atom. Bohr Atom

Q1. As the world population increases there is a greater demand for fertilisers.

NIRS Getting started. US Dairy Forage Research Center 2004 Consortium Annual Meeting

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

S2. INTRODUCTION TO ULTRA-VIOLET / VISIBLE SPECTROSCOPY

High-Performance Semiconducting Polythiophenes for Organic Thin Film. Transistors by Beng S. Ong,* Yiliang Wu, Ping Liu and Sandra Gardner

Outline of Recombinant DNA technology. Application of UV spectroscopy in recombinant DNA technology

Spectroscopy in Transmission

C101-E100. Talk Letter. Vol.1 October 2008

LIBSlab ANALYZERS ANALYZERS

Materials of Construction and Chemical Compatibility for Sensing and Control Products

Chemical formula - tells you how many atoms of each element are in a compound example: CO 2 (carbon dioxide) has one carbon atom and two oxygen atoms

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

The electrolysis of sodium chloride solution produces useful substances. covalent ionic non-metallic

Fresnel Equations cont.

COLORIMETER AND LAMBERT S-BEER S LAW. Shingala vaishali Sandha prafulla Tiwari Kuldeep

Portable type TXRF analyzer: Ourstex 200TX

Taking fingerprints of stars, galaxies, and interstellar gas clouds

Supplementary Information

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Chapter 9: Stoichiometry The Arithmetic ti Of Equations

Permeable Silica Shell through Surface-Protected Etching

Carbon Quantum Dots/NiFe Layered Double Hydroxide. Composite as High Efficient Electrocatalyst for Water

Ultraviolet and Visible Spectroscopy. interaction of materials with light at different electronic levels and the extent, to which such

An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13)

R O Y G B V. Spin States. Outer Shell Electrons. Molecular Rotations. Inner Shell Electrons. Molecular Vibrations. Nuclear Transitions

GENERAL PHARMACOPOEIA MONOGRAPH

A level Chemistry Preparation Work

Trapping Lithium into Hollow Silica Microspheres. with a Carbon Nanotube Core for Dendrite-Free

Questions on Instrumental Methods of Analysis

Chemistry Instrumental Analysis Lecture 15. Chem 4631

Chem 321 Lecture 18 - Spectrophotometry 10/31/13

TEST NAME: Chemistry TEST ID: GRADE:08 SUBJECT:Life and Physical Sciences TEST CATEGORY: My Classroom

The Claviature of Gas Analysis

3 - Atomic Absorption Spectroscopy

Supporting Information s for

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

PAPER No. 12: ORGANIC SPECTROSCOPY MODULE No. 4: Basic principles and Instrumentation for IR spectroscopy

ISA QATAR : 90 th Seminar

Visible-light Driven Plasmonic Photocatalyst Helical Chiral TiO 2 Nanofibers

Smart Sensing Embedded Spectroscopy Platform Botlek studiegroep 06-april-2017

Chem 310 rd. 3 Homework Set Answers

-:Vijay Singh(09CEB023)

Transcription:

FFeature Article Article Industrial In-line and Multi Component Monitor Using Absorption Spectroscopy and Its Application Yoko NAKAI HORIBA s CS-Series chemical concentration monitors that use ultraviolet (UV) and near-infrared (NIR) absorption spectroscopy are widely used in semiconductor wet processes, and make it possible to measure multiple components in line and in real time. The monitor can measure various types of samples, from general ammonia and hydrogen peroxide mixtures in semiconductor cleaning processes to 5-component chemistry including by-products from etching and metal ions in solutions, and things like CMP slurries including solid particles. This paper will introduce specific examples of applications. Introduction The HORIBA Group applies various fundamental technologies such as electrochemical, optical, and X-ray technology to provide analysis and measurement solutions. Out of those, an example that makes use of optical technology is the CS-series chemical concentration monitors that use ultraviolet (UV) and near-infrared (NIR) absorption spectroscopy, which have been widely used in semiconductor wet processes for many years and enables to take in-line, real-time measurements. This paper will introduce specific examples of applications. Measurement principle and structure of CS-Series chemical concentration monitors Absorption spectroscopy is a technique that is suited to measure chemical concentrations in a non-destructive, non-contact manner in real time. Incident light passes through sample cell and it is absorbed by the sample. The absorbed wavelength and absorption level differ based on the substance and concentration. The light absorbance A(λ) at a certain wavelength is expressed in Equation 1. A(λ) = log 10 [I 0 (λ)/i(λ)] (1) I 0 (λ): Incident light intensity I(λ): Transmitted light intensity λ: Wavelength The light absorbance and sample concentration is also expressed as follows, according to the Beer-Lambert law. A(λ) = α(λ)lc (2) α(λ): Absorption coefficient L: Sample cell length C: Sample concentration The Beer-Lambert law is the basic principle in the CS Series. The light absorbance is calculated by measuring the light intensity of each wavelength, and the chemical concentration is calculated based on the absorption coefficient and cell length that were set in advance. In actual measurement systems, it is difficult to directly measure the incident light, so we actually measure the reference light intensity I r instead of the incident light I 0. With regard to wavelengths for CS-series, the measurements are limited to the UV range and NIR range. The NIR range contains the absorption range for O-H bonding and N-H bonding, etc. The UV range contains the absorption range for H 2 O 2 and HNO 3 which are generally used in semiconductor cleaning processes. To make use of these characteristics, the monitor equips a Czerny-Turner spectrometer that can simultaneously conduct spectroscopy and detection in the UV and NIR ranges and use the multi-channel detector to simultaneously measure the amount of light at multiple wavelength points. In this way, we have achieved a fast response with a measurement interval of approximately 0.5-3 seconds, even when signal integration and moving average processing are included. There is also no moving part in the spectroscopy area, which means that measurements have good reproducibility. This design is suitable for industrial applications that require high reliability and reproducibility. [1] In addition, the sample cell is high-purity quartz or 2 English Edition No.47 June 2017

Technical Reports Sample reference switching Halogen lamp Sample measurement Grating UV multi-channel detector Reference measurement NIR multi-channel detector Figure 1 Structure of CS-100 series high-purity sapphire, and the other wetting materials are PFA (perfluoroalkoxy polymer) and PTFE (polytetrafluoroethylene), that are suitable for use in semiconductor processes that require chemical resistance and high level of cleanliness. CS-Series chemical concentration monitor line-up The CS Series is categorized into 2 line-ups: (1) the builtin measurement cell type and (2) the separate measurement cell type. (1) Built-in measurement cell type This type includes the CS-100 Series (basic models) and the CS-700 Series (high-end models). The CS-700 Series makes use of the experience HORIBA gained from the CS-100 Series, and has a variety of added improvements in terms of hardware and algorism, etc. The most important point is the optical system, the CS-700 Series equips a detector with a larger number of channels than conventional systems for both UV and NIR, which has resulted in higher wavelength resolution, and improved spectrum separation performance. (2) Separate measurement cell type (optical fiber type) In this type, the measurement cell area and main unit area where equips electrical and optical system are separated, and both are connected using optical fibers. CS-100F1 is mainly used to measure corrosive chemicals, high-temperature and large flow rate chemicals. In the CS-600F Series, both the sample light and reference light can be measured by using an air-driven cylinder to move a mirror inside the measurement cell unit. In this way, we were able to significantly reduce the frequency of background corrections, which were necessary on the conventional CS-100F1 Series, from every 6 hours to more than one month. Preparing the calibration curve In the CS Series, for specific chemical types and concentration ranges of each component, there are individual conversion coefficients for converting concentrations from light absorbance spectra, and these are called calibration curves. These calibration curves are installed in individual monitors based on the specifications, various adjustments and inspections are conducted for the actual Table 1 CS series line up (CS-100, CS-700, CS-100F1, CS-600F) Build-in measurement cell type Separate measurement cell type CS-100 Series CS-700 Series CS-100F1 Series CS-600F Series Appearance Note Basic model Wider wavelength, higher resolution Basic model High stability model Temperature 20-30 C (Cooler unit can be attached) 20-30 C 20-30 C 20-80 C English Edition No.47 June 2017 3

Feature Article Example: SC-1 monitor Target concentration: NH 3 :6-11wt%, H 2O 2 8-13wt% 1 3 NH 3 H 2O 2 H 2O Calibration curve (Coefficient converting from absorbance to concentration) - Prepare concentration matrix to cover measurement range - Prepare chemical mixture based on the matrix Multivariate analysis Install H2O2 Concentration (wt%) 14 13 12 11 10 9 8 7 5 6 7 8 9 10 11 12 NH 3 Concentration (wt%) n n 3 Absorbance data 4 3 2 1 Absorbance 2 1 NH3 : x%, H2O2 : y%, H2O : z% T C Wavelength 2 Take absorbance spectrum of each samples Figure 2 Schematic flow of making calibration curve chemical, and then the monitors are shipped. To make a calibration curve, samples with known concentration are adjusted in matrix form within the concentration range for the applicable chemical, and these spectrums are obtained. Multivariate analysis is used to make the calibration curve based on this concentration information and the light absorbance spectrum. Measurement examples As a typical example of measurements with the CS-100 Series, we will give an example of measuring a chemical used in semiconductor cleaning processes that is mainly made from 2-3 components, such as NH 3 /H 2 O 2 /H 2 O or Figure 3 Absorbance spectra of metal ions HCl/H 2 O 2 /H 2 O. In recent years substances are not limited to this number of components, and chemicals are measured that are used in manufacturing processes in various fields, such as cases when by-products are generated, cases when organic substances are included, cases when metal ions are also present, and fields outside semiconductor, such as plating and lithium-ion batteries. Some of these are introduced below. Measuring metal ion concentrations in chemicals Here is an example of measuring the iron ion concentration. When iron dissolves into hydrochloric acid, the majority of it is bivalent, but some is trivalent, because it is oxidized in air. [2] We tried to measure the separation of Fe 2+ and Fe 3+ in hydrochloric acid. We set up the following as the measurement concentration range. HCl: 5-10 mass % Fe 2+ : 5-10 mass % Fe 3+ : 0.05-0.2 mass % To measure these components, when we made the calibration curve, we incorporated the iron chloride ions as one component in advance. That is, for Fe 2+, we used FeCl 2 and 4H 2 O powder, and for Fe 3+, we used an FeCl 3 aqueous solution, and mixed these with HCl and water, and made adjustments until the ion concentration became the prescribed concentration. When the iron ions dissolved, we saw that with Fe 2+, the solution turned a yellowish green color, and with Fe 3+ it 4 English Edition No.47 June 2017

Technical Reports Figure 4 Measurement results of HCl, Fe 2+ and Fe 3+, using monitor which equips spectrometer for VIS region became a yellowish brown, and based on this we found that there was absorption in the visible range near blue, which is in a complementary color relationship. Figure 3 shows the Fe 2+ and Fe 3+ absorption spectrum with air as a reference. To measure the Fe 2+ and Fe 3+ absorption range at the same time, we adjusted the optical system so that we could measure the visible range, measured the above sample, and made a calibration curve. Figure 4 shows that calibration data. In this sample, all the components (HCl, Fe 2+, and Fe 3+ ) were adjusted so that they have variation, but the respective components separated and we were able to measure them. Measuring multiple components using the CS-700 Series Here we will discuss a case of a mixed acid with 3 or Figure 5 CS-700 measurement results of 5 components acid English Edition No.47 June 2017 5

Feature Article more components, where by-products were also generated. In this example, we monitored a case where silicon dissolved during etching into chemical mixture composed of 3 types of acid (HF, HNO 3, and acid), which is generally used in etching silicon wafers. Generally, the acid concentration decreases as etching progresses, and byproducts are generated from reactions with silicon. If the by-product concentration is high, it affects the absorption spectrum, and can cause errors in the concentration measurement results of each acid. As such, when we made the calibration curve, in addition to the water and 3 types of acid, we added the etching by-product H 2 SiF 6 as one component, prepared the sample as having 5 components, and made the curve based on this information. Figure 5 shows the measurement results using the multi-wavelength, high-resolution CS-700. None of the components were affected by the H 2 SiF 6, and the measurement results were good, and we were also able to measure the concentration of H 2 SiF 6. Example of measuring semiconductor CMP slurry In semiconductor manufacturing, a process called CMP (chemical mechanical polishing) is used, in which wafer surfaces are polished using chemical and mechanical reaction. In processes where metal layers such as copper are polished, hydrogen peroxide is added to the slurry, which increases the polishing efficiency due to the oxidation of the metal. At this time, controlling the concentration of hydrogen peroxide is an important point. In general CMP slurries, the particle diameter of the abrasive is often at the hundred or several-hundred nm level, but in the UV range, where the wavelength is short, the light is scattered by the particles, and this is not suitable for measurement when absorption spectroscopy is used. On the other hand, in the NIR range, the wavelength is long compared to the particle size, and light passes through without being scattered, so this is relatively suitable for measurement. We made use of the characteristic that H 2 O 2 also has absorption in the NIR range, although not as much as in the UV range, and tried to measure the H 2 O 2 concentration in the CMP slurry using only the NIR range. In addition to abrasive particles and water as a solvent, CMP slurries include various components such as ph regulators and inhibitors. We made a calibration curve to measure the H 2 O 2 concentration and set it up for 3 components, considering the overall slurry as one component, along with the H 2 O 2 and water [3]. As shown in Figure 6, there is good linearity, and we found that even in chemical mixture that include particles, if things like the particle sizes and concentrations are suitable, it is possible to take measurements. Summary Figure 6 Measurement result of H 2O 2 in CMP slurry This paper introduced concentration monitors that apply absorption spectroscopy that is used to measure cleaning chemical and etching chemical that are mainly used in semiconductor manufacturing processes. A wide variety of things can be measured, such as acid and metal ion concentrations dissolved in acid, concentrations of mixed acids that include by-products, hydrogen peroxide concentrations in CMP slurry that include particles, and also other substances that have not been listed here. In addition, there are many variations in the line-up for things like wavelength range and sampling system types, along with sample conditions and integration with customers manufacturing equipment. We would like to use our technology, starting with absorption analysis in various industrial fields, not just semiconductors, to contribute to solving problems in manufacturing processes and improving quality and production efficiency through in-line monitoring. References [ 1 ] Issei Yokoyama, Concentration monitor for semiconductor wet process in Instrumentation and Automation, vol. 36, No.13, JAPAN INDUSTRIAL PUBLISHING CO., LTD., Dec. 2008, pp.22-23 [ 2 ] Bruce H. Mahan, University Chemistry 2 nd edition, Hirokawa Publishing Co., Oct. 1998, pp.746 [ 3 ] Yoko Nakai, Hydrogen Peroxide Concentration Monitoring for CMP slurry, International Conference on Planarization/CMP Technology (2014), pp.227-228 Yoko NAKAI Application Development Dept. Research & Development Division HORIBA Advanced Techno, Co., Ltd. 6 English Edition No.47 June 2017