Understanding the colloidal stability of protein therapeutics using dynamic light scattering

Similar documents
Use of DLS/Raman to study the thermal unfolding process of lysozyme

Understanding the conformational stability of protein therapeutics using Raman spectroscopy

DEVELOPING A BIOFORMULATION STABILITY PROFILE

Introduction to the calculators in the Zetasizer software

The power of multi-detector SEC in the analysis of antibodies

Executive Summary. Introduction WHITEPAPER

Component specific particle characterisation of the active components in a pharmaceutical topical formulations

Characterization of Single Wall Carbon Nanotubes (SWNTs) by combining Dynamic Light Scattering and Raman Spectroscopy

GPC/SEC Analysis of Polymer Degradation

Discriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS)

Optimization of PLA/PLGA molecular weight and structure to suit final application requirements

Characterization of protein aggregates in suspension and on a filter membrane by Morphologically-Directed Raman Spectroscopy

10 REASONS THAT THE ZETASIZER NANO IS CHOSEN FOR SCIENTIFIC SUCCESS

Differentiation of polymer branching and composition using the Mark Houwink plot

Overcoming and quantifying Wall Slip in measurements made on a rotational rheometer

Nanobubble Applications and Characterization by Nanoparticle Tracking Analysis

GPC/SEC Analysis of Polymer Solutions Used in Inkjet Printing

How to use GPC/SEC for compositional analysis

Studying the Effect of Steric Layer Thickness on Emulsion Stability

Determining cure profile and post-cure shrinkage of photopolymers using UV accessory on a rotational rheometer

How switching mobile phases can improve your OMNISEC results

Using rheology to study the hardness and spreadability of butter

Using the Zetasizer Nano to optimize formulation stability

Evaluating the rheological properties of hyaluronic acid hydrogels for dermal filler applications

MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

Measurements of protein electrophoretic mobility using the Zetasizer Nano ZSP

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20

The ultimate in desktop particle characterization

Measuring the rheology of thermoplastic polymer melts

Sample measurements to demonstrate Zetasizer specifications (Zetasizer Nano, Zetasizer APS, Zetasizer μv)

Zetasizer Nano ZSP: A Perfect Tool For Life Science Applications

The ultimate in desktop particle characterization

Zeta potential - An introduction in 30 minutes

Protein-Protein Interaction Measurement: Balancing Complete Characterization with High- Throughput Capability in Formulation Development

ZS90. The ultimate in desktop particle characterization. Particle size. Zeta potential. Molecular weight

Meet Our Uncle: 12 Stability Applications on One Platform

Laser diffraction and automated imaging: complementary techniques for nasal spray development

Biophysical characterization of SMALPs and nanodiscs

The Diffusion Barrier Technique, Practical Aspects and Data interpretation

Meet Stunner: The one-shot protein concentration and sizing combo

Electrophoretic Light Scattering Overview

Isothermal experiments characterize time-dependent aggregation and unfolding

Measuring the rheology of polymer solutions

ZETASIZER NANO Series PERFORMANCE, SIMPLICITY, VERSATILITY

Gold Nanoparticle Applications and Characterization by Nanoparticle Tracking Analysis

Malvern Macromolecular Solutions. Innovative Solutions in Material Characterization

A basic guide to particle characterization

KINEXUS DSR SERIES REDEFINING RHEOMETER CAPABILITIES FOR ASPHALT TESTING

next generation Welcome to the Dispersion options Smarter particle sizing Your sample, expertly processed and prepared Particle size

Generating Aggregation Rates to Determine Stability of Monoclonal Antibodies

Inform is a series of white papers designed to provide advice on material characterization issues. Mie theory The rst 100 years

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR

A Basic Guide to INFORM WHITE PAPER. Abstract:

Litesizer Sheet 1. [Home]

Investigating the Relationship Between the Rheological Properties of Hyaluronic Acid and its Molecular Weight and Structure using Multidetector

Bohlin. Rheological instruments backed with rheological experience. Rheological properties

Particle Analysis at the Touch of a Button. Litesizer series

Highly automated protein formulation development: a case study

Introduction to Dynamic Light Scattering with Applications. Onofrio Annunziata Department of Chemistry Texas Christian University Fort Worth, TX, USA

REVERSE ENGINEERING PHARMACEUTICAL FORMULATION USING THE MORPHOLOGI G3-ID.

Morphologically directed chemical identification Coupling particle characterization by morphological imaging with Raman spectroscopy

One-stop stability. T m & T agg. T m with SYPRO (DSF) Isothermal stability. Thermal recovery. Sizing & polydispersity. Sizing with thermal ramp B 22

KINEXUS SERIES REDEFINING RHEOMETER CAPABILITIES

KINEXUS SERIES REDEFINING RHEOMETER CAPABILITIES

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Page 1 of 5. Is it alright to estimate dñ/dc in SLS measurements?

Introduction to Dynamic Light Scattering for Particle Size Determination

YOUR PARTNER IN PARTICLE CHARACTERIZATION

Measuring Lysozyme Monomer at 0.1 mg/ml Concentration. Equipment used : Sample Preparation and Measurement :

Suspension Stability; Why Particle Size, Zeta Potential and Rheology are Important

Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics

Novel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison

South pacific Journal of Technology and Science

ORTHOGONAL PARTICLE CHARACTERIZATION TECHNIQUES FOR BIO-APPLICATIONS: AN INTRODUCTION TO DLS (DYNAMIC LIGHT SCATTERING), NTA (NANOPARTICLE

Analytical methods to characterize aggregates and subvisible particles

LS Spectrometer. The DLS and SLS system for sophisticated research

Particle Analysis at the Touch of a Button. Litesizer series

The Illumination of Structure using Light Scattering. Michael Caves Product Technical Specialist for Biophysical Characterisation

Setting the Standard for Advanced GPC/SEC Systems

Holographic Characterization of Protein Aggregates

Positive and Nondestructive Identification of Acrylic-Based Coatings

Using runt Disposable Cuvettes

Analytical Ultracentrifugation Macromolecule Characterization in Solution

Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis

Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacement To Gold Nanoparticles.

Analysis of demicellization data from isothermal titration calorimetry

Monitoring the Aggregation of Particles through Raman Spectroscopy

Surface Zeta Potential Cell (ZEN1020)

Dr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1

Particle Characterization Laboratories, Inc.

Use of SEC-MALS. (Size Exclusion Chromatography - Multi Angle. Light Scattering) for protein quality and characterization

Amended Specifications as per the Pre Bid meeting held on in Chapter 4

Clemens Stilling Characterisation, Analytical Sciences for Biologicals, UCB Pharma S.A. Alun, living with Parkinson s disease

Holographic Characterization of Agglomerates in CMP Slurries

Electronic Supplementary Material (ESI) for Nanoscale This journal is The Royal Society of Chemistry Characterization of CeO2 NP suspensions

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System

Gel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC)

Static and dynamic light scattering. Cy Jeffries EMBL Hamburg

Accessory Publication

Impurities In Biomolecules Institute for International Research. Monitoring & Predicting Biomolecular Aggregation Using Light Scattering

Transcription:

Understanding the colloidal stability of protein therapeutics using dynamic light scattering A Malvern Instruments' Bioscience Development Initiative Executive summary The combination of dynamic light scattering (DLS) with Raman spectroscopy provides the ability to extract a wealth of chemical, structural, and physical information about biotherapeutic proteins under formulation conditions. Dynamic light scattering (DLS) using noninvasive backscatter (NIBS) detection technology is capable of measuring the hydrodynamic radius of proteins over a large concentration range (0.1 mg/ml to 100 mg/ml). As a result, it is possible to determine, in actual biopharmaceutical formulations, the bulk viscosity/restricted diffusion interaction parameter (k D ), particle interaction parameter (B 22 ), melting temperature, onset temperature of protein aggregation and transition enthalpies to assess colloidal stability. The unique coupling of DLS with Raman spectroscopy provides the ability to simultaneously correlate colloidal parameters to protein structure under a variety of stress conditions, e.g., thermal, formulation, chemical degradation, extrinsic particulates, to enhance our understanding of protein therapeutic formulations. Introduction Dynamic light scattering (DLS) is a workhorse technique for determining the hydrodynamic diameter of biotherapeutic proteins in solution, sample polydispersity, and sample interactions. Raman spectroscopy derives information about protein secondary and tertiary structure by monitoring molecular vibrations. By combining these two analytical approaches into a single system, a wealth of chemical, structural and physical parameters can be determined. Protein secondary and tertiary structure, melting temperature, onset temperature of aggregation and transition enthalpy values can all be derived, as well as aggregation propensity, protein solubility, and the potential for high viscosities at formulated concentrations. The complementary nature of the results obtained from DLS and Raman spectroscopy on the same sample may provide unique insights into the mechanisms of aggregation and unfolding. Here we Malvern Instruments Worldwide Sales and service centres in over 65 countries www.malvern.com/contact 2014 Malvern Instruments Limited

describe the utility of DLS to aid in the understanding the colloidal stability of protein therapeutics. The approach Malvern Instruments Zetasizer Helix (ZS Helix) integrates a fiber-coupled Raman spectrometer with a Zetasizer Nano ZSP to provide DLS (colloidal stability) and Raman (conformational stability) data sequentially on a single sample. The Zetasizer Nano system integrates proprietary non-invasive backscatter (NIBS) detector technology with dynamic (DLS), static (SLS) and electrophoretic (ELS) light scattering to measure the hydrodynamic radius of proteins from 0.15 nm - 5 µm, from 0.1 mg/ ml to 100 mg/ml. Raman spectra are collected using 785 nm excitation (~280 mw) from 150-1925 cm -1 at 4 cm -1 resolution. Sample aliquots (~120 µl) are placed into a 3 mm quartz cuvette and positioned in a temperature-controlled compartment that provides temperature control from 0 C - 90 C ± 0.1 C. Thermal ramp studies are conducted by collecting Raman and DLS data over a series of pre-defined 0.1 C - 5 C step-wise increments. Isothermal incubation studies are conducted by collecting a series of Raman and DLS data over a predefined time interval at a desired temperature set-point. Results and Discussion Measuring protein size at high concentration The formulation of monoclonal antibodies at high concentration has drawn great attention within the biopharmaceutical industry in recent years. However, for DLS, size measurement at formulated concentrations was limited by multiple experimental factors. As shown in Figure 1, increasing sample concentration could affect the observed scattering intensity and consequently the size determination. Factors affecting the measurement could include: multiple scattering, viscosity, positive/ negative virial effects, and self-association. Deconvolution of the above effects allows for interpretation of high-concentration sample results. Figure 1: The observed effects of increased sample concentration on the measured light scattering intensity. 2 Understanding the colloidal stability of protein therapeutics using dynamic light scattering

To better understand the effects of the aforementioned factors on the DLS data, it is convenient to consider an individual effect in isolation. Table I (below) summarizes the expected changes with increasing sample concentration to the reported size, polydispersity, and correlation function, as well as established correction methods, for each of the unwanted factors. Table 1: Summary of the effects of high concentration sample on DLS parameters Factors Apparent Size Apparent Polydispersity Correlation Intersection Correction Method Multiple Scattering Decrease Increase Decrease Backscattering Restricted Diffusion Increase No Change No Change Viscosity Correction Negative B 22 Increase Increase No Change No Positive B 22 Decrease No Change No Change No Figure 2 (below) demonstrates the effect of viscosity correction on the reported size of a sample. Using the viscosity of water as an estimate of the sample viscosity results in an increase in apparent size of the suspended particles (blue squares). However, when the true bulk solution viscosity is used to correctly fit the DLS results, no apparent size change is observed (green circles). Particle interaction terms including attractive/ repulsive, B 22 (second virial coefficient) or k d (diffusion interaction parameter) could be indicated and will be covered in the following section. It should be noted that for virial effects, there is currently no agreed approach to correct DLS data. Figure 2: The effects of bulk viscosity correction on the size measurement results. Detecting protein aggregation onset Static light scattering (SLS) or turbidity coupled with thermal ramping have been routinely used as rapid formulation screening tools for protein stability. Essentially, these methods measure the aggregation onset temperature T onset. DLS could also be utilized in this fashion. Figure 3 shows the aggregation onset for BSA is approximately 62 C by Z-averaged size. Additionally, as long as the correlogram remains valid, DLS offers the size distribution (Fig. 3, insert) and polydispersity (PDI), where SLS offers only the intensity. The size distribution trend shown here clearly reveals the 3 Understanding the colloidal stability of protein therapeutics using dynamic light scattering

appearance of a second peak at 90 nm near 60 C. The onset temperature from PID could be as early as 55 C, indicating aggregation may begin even earlier. Figure 3: DLS thermal ramping of BSA (50 mg/ml ph 7.4). Figure 4: Illustration of the intensity profile of two particles with 5 and 50 nm diameter. The intensity ratio will be 1:1000000. The ability to detect trace amounts of aggregates stems from the inherent sensitivity of DLS. Figure 4 demonstrates that for a simple mixture of two particles with an order of magnitude difference in diameter, the larger 50 nm particle will show a scattering intensity a million times stronger than the smaller 5 nm particle. This infers that even a single large particle mixed with a million small particles results in an equal detected intensity! Thus, DLS could be useful as the earliest detector for the appearance of aggregates. However, it should be noted that the resolution of DLS is inherently limited to a factor of three in size. Predicting colloidal stability 4 Understanding the colloidal stability of protein therapeutics using dynamic light scattering

Though T onset might be sufficient for initial formulation screening, no understanding of protein interaction can be reached from such experiments, let alone the prediction of protein stability. Recently, more attention has been focused on investigating intermolecular parameters as predictors of protein aggregation kinetics, which have been successfully correlated with B 22 and k D [1]. A large positive value of B 22 indicates that solutes prefer association with the solvent over self-association, either due to strong electrostatic repulsion between solute molecules or positive interaction between solute and solvent. In contrast, large negative values of B 22 indicate strongly selfattractive systems, where the particles prefer self-association or aggregation over complete solvation. In the case of DLS, k D can be measured directly and is related to B 22 by the following equations [2]: Here, C is the sample concentration and D refers to the diffusion coefficient measured by DLS. D 0 is the self-diffusion coefficient (the value of D at zero concentration). M w is the sample molecular weight, k f is the first order concentration coefficient of the friction and ν is the partial specific volume of the solute. The following experiments demonstrate how to apply k D to predict the protein stability for an IgG in four formulations. In Figure 5 only Soup 2 (red circles) has a positive k D, indicating a strong repulsive interaction between the protein samples. The other three samples show slightly negative k D. It should be mentioned that large positive k D usually indicates strong repulsive interactions, or positive B 22. However, when k D is close to zero it becomes difficult to directly link to B 22. Figure 5: k D plot measured for the same IgG solution in four formulations. 5 Understanding the colloidal stability of protein therapeutics using dynamic light scattering

Figure 6: DLS thermal ramping of one IgG sample in four different formulations. SLS can be measured at a fixed angle for B 22, and zeta potential measured to provide relative charge, as summarized in Table 2. Putting all of the information together, a comprehensive profile of protein properties can be obtained to predict protein stability. For example, we would predict that formulation 2 is the most stable formulation for this IgG, while formulation 1 is the least stable. To support this prediction, thermal ramping experiments were performed on the same set of samples, and are shown in Figure 6. Interestingly, sample 1 has the lowest T onset, close to 66 C, while formulation 2 is the most stable, showing no evidence of large aggregates up to 90 C. Table 2: Summary of predictive parameters for IgG stability IgG Formulation B 22 (10-5 ml mol/g 2 ) k D (ml/g) Charge Soup 1-1.5-5.2 3.3 Soup 2 127.5 31.9 9.7 Soup 3 10.4-9.7 5.1 Soup 4 2.3-4.7-2.8 Summary The combination and integration of DLS and Raman spectroscopy into a single instrumental platform provides unique analytical capabilities to determine protein secondary and tertiary structure, melting temperature, and size and onset of aggregation for the same sample, without altering the testing conditions. Here we have described the unique ability of DLS, when combined with noninvasive backscatter detector (NIBS) technology, to derive the bulk viscosity/restricted diffusion interaction parameter (k D ), particle interaction parameter (B 22 ), melting temperature, onset temperature of aggregation and transition enthalpies on high concentration formulations. A subsequent study describes the unique spectra-structure properties that may be correlated using Raman Spectroscopy (Understanding the Conformational Stability of Protein Therapeutics with Raman Spectroscopy). Future studies will report on the unique correlations that may be derived from the combination of DLS and Raman spectroscopy to improve the understanding of the kinetics, thermodynamics 6 Understanding the colloidal stability of protein therapeutics using dynamic light scattering

and mechanism(s) of aggregation and the correlation of specific protein structural motifs to aggregation to improve product knowledge and formulation stability. About Malvern's Bioscience Development Initiative Malvern Instruments' Bioscience Development Initiative (BDI) was established to accelerate innovation, development and the promotion of new technologies, products and capabilities to address unmet measurement needs in the biosciences markets. References 1. Diffusion and sedimentation interaction parameters for measuring the second virial coefficient and their utility as predictors of protein aggregation. Saluja A. et al., Biophysical Journal. 2010. 99, 2657-2665. 2. Specific interactions in high concentration antibody solutions resulting in high viscosity. Yadav S. et al., Journal of Pharm. Sci. 2010, 99, 1152-1168. 7 Understanding the colloidal stability of protein therapeutics using dynamic light scattering

Malvern Instruments Limited Grovewood Road, Malvern, Worcestershire, UK. WR14 1XZ Tel: +44 1684 892456 Fax: +44 1684 892789 www.malvern.com Malvern Instruments is part of Spectris plc, the Precision Instrumentation and Controls Company. Spectris and the Spectris logo are Trade Marks of Spectris plc. All information supplied within is correct at time of publication. Malvern Instruments pursues a policy of continual improvement due to technical development. We therefore reserve the right to deviate from information, descriptions, and specifications in this publication without notice. Malvern Instruments shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Malvern Instruments owns the following registered trademarks: Bohlin, FIPA, Insitec, ISYS, Kinexus, Malvern, Malvern 'Hills' logo, Mastersizer, Morphologi, Rosand, 'SEC-MALS', Viscosizer, Viscotek, Viscogel and Zetasizer. 2014 Malvern Instruments Limited