THE UNCERTAINTY OF MEASUREMENT IN CALIBRATION USING A COMPARISON FORCE STANDARD MACHINE

Similar documents
Uncertainty of Calibration. Results in Force Measurements

INTERLABORATORY COMPARISON BETWEEN TUBITAK-UME TURKEY AND INM ROMANIA IN THE FIELD OF FORCE MEASUREMENTS

UNCERTAINTY SCOPE OF THE FORCE CALIBRATION MACHINES. A. Sawla Physikalisch-Technische Bundesanstalt Bundesallee 100, D Braunschweig, Germany

FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

UNCERTAINTY IN TORQUE CALIBRATION USING VERTICAL TORQUE AXIS ARRANGEMENT AND SYMMETRICAL TWO FORCE MEASUREMENT

Calibration and verification: Two procedures having comparable objectives and results

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal

< Final report > Report on the APMP.M.F-S1 supplementary comparison for 2 MN Force

THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY

ISO INTERNATIONAL STANDARD

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

R SANAS Page 1 of 7

ISO 376 Calibration Uncertainty C. Ferrero

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

Can there be a metrology for psychometricians?

Metrological Characterization of a Primary Vickers Hardness Standard Machine - NIS Egypt

Document No: TR 12 Issue No: 1

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

Evaluation of Force Standard Machines by using Build-up System

FORCE NATIONAL STANDARDS COMPARISON BETWEEN CENAM/MEXICO AND NIM/CHINA

Outline of the current status of measurement science: from the point of view of the International Vocabulary of Metrology (VIM)

Revision of the Guide to the expression of uncertainty in measurement impact on national metrology institutes and industry

Guidelines on the Calibration of Static Torque Measuring Devices

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

This document is a preview generated by EVS

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

Measurement uncertainty and legal limits in analytical measurements

A guide to expression of uncertainty of measurements QA4EO-QAEO-GEN-DQK-006

INVESTIGATION OF TRANSFER STANDARDS IN THE HIGHEST RANGE UP TO 50 MN WITHIN EMRP PROJECT SIB 63 Falk Tegtmeier 1, Michael Wagner 2, Rolf Kumme 3

Increased Requirements for Higher Nominal Forces, Necessities and Possibilities to Measure Them

A combinatorial technique for weighbridge verification

Calibration Traceability Guidelines

EA 4/02. Expression of the Uncertainty of Measurement in Calibration. Publication Reference

Essentials of expressing measurement uncertainty

Study for the requalification of Inmetro's Primary Hardness Standardization Machine for Vickers HV3 scale

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

ISO INTERNATIONAL STANDARD

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

Multi-Capacity Load Cell Concept

Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn. Aimo Pusa MIKES Finland

Experimental investigations of different force measuring systems

Reliable Test Results

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

An Introduction to the Differences Between the Two Most Recognized Force Standards

Guide to the Expression of Uncertainty in Measurement (GUM) and its supplemental guides

Introduction to the evaluation of uncertainty

Joint Committee for Traceability in Laboratory Medicine Terminology. R. Wielgosz and S. Maniguet

Guide to the Uncertainty of Force Measurement

Introduction To Metrology

ESTIMATION OF UNCERTAINTY IN HARDNESS MEASUREMENT OF RUBBER AND OTHER ELASTOPLASTIC MATERIALS

ISO INTERNATIONAL STANDARD. Metallic materials Knoop hardness test Part 1: Test method

COMPLETION AND MEASUREMENT UNCERTAINTY BUDGET OF THE MULTI-COMPONENT MEASURING DEVISE FOR FORCE UP TO 1 MN AND TORQUE UP TO 2 KN M

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

Measurement & Uncertainty - Basic concept and its application

Investigation of Piezoelectric Force Measuring Devices in Force Calibration and Force. Standard Machines

Creation of the π angle standard for the flat angle measurements

NIST ELECTROSTATIC FORCE BALANCE EXPERIMENT

OIML G GUIDE. Edition 2009 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

EVALUATION OF STATIC AND DYNAMIC PARASSITIC COMPONENTS ON THE INRIM 1 MN PRIMARY FORCE STANDARD MACHINE BY MEANS THE 500 kn SIX-COMPONENT DYNAMOMETER

MEASUREMENT TRACEABILITY AND ERRORS RELATED TO FORCE MEASUREMENT

Annex xxx General guidelines for the assessment of uncertainties in solar thermal systems performance testing

VNA error models: Comments on EURAMET/cg-12/v.01

Force Key Comparison EUROMET.M.F-K2. (50 kn and 100 kn) EURAMET Project No 518. Final Report. 14 July Pilot: NPL, United Kingdom

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

Project 887 Force Uncertainty Guide. Andy Knott, NPL EUROMET TC-M Lensbury, Teddington, United Kingdom 1 March 2007

Uncertainty Propagation for Force Calibration Systems

Verso il VIM4? La situazione delle attività del JCGM / WG2 (VIM) Luca Mari.

FEEDBACK CONTROLLED DEADWEIGHT MACHINE

ISO/BIPM Guide: Uncertainty of measurement 1

Workshops. Thursday, November 21, 9:00 AM to 11:00 AM: Models of measurement: measuring systems and metrological infrastructure

Training Benefits Adding Values to Your Systems/ Organization through: Potential Attendees These Courses have been designed for;

SIM FORCE STANDARDS COMPARISON UP TO 10 kn

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS

Establishment of a Traceability Chain in Metrology for Materials

HOW TO EXPRESS THE RELIABILITY OF MEASUREMENT RESULTS. The role of metrological traceability and measurement uncertainty-

Guide to the Expression of Uncertainty in Measurement Supplement 1 Numerical Methods for the Propagation of Distributions

ISO INTERNATIONAL STANDARD. Metallic materials Vickers hardness test Part 3: Calibration of reference blocks

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

Uncertainty restrictions of transfer click torque wrenches

Measurement & Uncertainty - Concept and its application

Metallic materials Brinell hardness test. Part 2: Verification and calibration of testing machines

The Determination of Uncertainties in Bend Tests on Metallic Materials

Science of Stability 2017

Force/Torque measuring facility for friction coefficient and multicomponent

TEACHING GENERAL METROLOGY: WHY, WHAT, HOW?

Practical Interpretation of Metrological Traceability on Gauge Block as to VIM 3

Instrument types and performance characteristics

The calibration system of force measurement devices - conceptions and principles

CCM short note on the dissemination process after the proposed redefinition of the kilogram

Transcription:

THE UNCERTAINTY OF MEASUREMENT IN CALIBRATION USING A COMPARISON FORCE STANDARD MACHINE Adrian Gherasimov 1, Eugen Ghita 1 Romanian Bureau of Legal Metrology, Force Laboratory Timisoara, Romania POLITEHICA University of Timisoara, Romania ABSTRACT The paper presents a practical example for the evaluation of the uncertainty in calibration using a comparison force standard machine with a maximum load of 1000 kn, tension and compression mode. The uncertainty of measurement associated with the input estimates is evaluated. It is also presented a way to minimize the uncertainty of the applied force with the view to attaining the necessary value for the calibration of class 1 force transducers, in accordance with ISO 376:1999. For this purpose an evaluation of the loading regime of the force machine is presented. There was identified the main input components influencing the uncertainty in calibrations, such as the force stability during measurements and the metrological characteristics of the reference transducers. Further are presented the practical results of the investigation of the force comparison machine after the optimization, including the technical and economical benefits. 1. INTRODUCTION In order to provide for the traceability of the force measurements to the International System of Units (SI), Romania possesses, as national force standards, three deadweight calibration machines with the maximum range of 10 kn, 50 kn and 100 kn. The Romanian Bureau of Legal Metrology, through National Institute of Metrology, Force Laboratory Timisoara, keeps the national force standards in Romania [1]. The Romanian legislation of metrology [] stipulates that the national standards are used only for the reproduction of the SI measurement units as well for their dissemination to the immediate inferior standards. This stipulation represents an important limitation regarding the utilization of the national standards in the whole traceability chain of force measurements. These circumstances conduce to the necessity to enlarge the force standard machine basis. This is because in Romania the owner-laboratory of the national standards is, in the same time, a metrological laboratory where the working standards are calibrated. We consider that this situation is typical also for other national laboratories. In order to obtain the optimal arrangement for the development of a reference standard system, as a basis for the force measuring traceability, there is imposed the use of some alternatives, which have to be suitable and correct from the viewpoint of their technical and metrological parameters, as well from the viewpoint of a reasonable cost. The challenge is to keep the uncertainty of the applied force within an appropriate range, necessary for calibration of the force measurement devices, with minimum costs. For the optimization of the basis of standards of the Force Laboratory Timisoara, meeting the requests of the applicants is considered a very important factor. From this viewpoint, for the Force Laboratory Timisoara, using comparison force machines represents an acceptable solution. In order to perform the technical and metrological parameters in accordance with ISO 376:1999 [3] the measuring system of the comparison force machines is based on the characteristics of the component elements ( e. g. reference force transducers, force reproduction mechanical system ). The most important metrological parameters of the force transducers, which qualify them to be used as reference standards for the comparison force machines, are as follows: repeatability, 1

reproducibility and stability [4]. There are available on the market some force transducers, which ensure the metrological requirements to be used as reference standards.. GENERAL VIEW The basis of standards, which ensure the traceability force measurements in Romania, is correlated with the requests of the customers. There is presented in figure 1a) the proportion of the class of accuracy, in accordance with ISO 376:1999, and, in figure 1b), the proportion of the nominal range of the force transducers calibrated in the Force Laboratory Timisoara, function of the applications of the customers. A statistics for three years is presented in figure 1 : 100 50 88,5 40 39 proportion (%) 10 8 proportion (%) 30 0 19 4 18 10 1 1,5 00 05 1 class ISO 376 0 1 10 10 100 100 1000 1000 nominal range (kn) a) b) Figure 1: Class and nominal ranges proportion from total calibration requests After the analysis of the data, the class 1 and (ISO 376) transducers in the nominal range up to 1000 kn represent a great share. This is the reason why, the Force Laboratory Timisoara promoted reference force standards in accordance with the customers requests. The uncertainty of the applied force requested by in order to ascertain the necessary class 1 (ISO 376) transducers, or less accurate, is easy to obtain by using a deadweight force calibration machine (the uncertainty of the applied force into the range of 5 x 10 5...1 x 10-4 ) and even by using a lever or hydraulic amplification machine ensuring an uncertainty of the applied force into the range of 1 x 10-4... 5 x 10-4. However these machines are expensive and sophisticated to be achieved, especially into the high forces ranges (over 100 kn). Taking into account the already extant facilities and the economic limitations, the Force Laboratory Timisoara succeeded the following directions: - keeping and developing the national force standards, the deadweight calibration machines into nominal ranges of 10 kn, 50 kn and 100 kn - developing reference standards for the calibration of dynamometers into the nominal ranges up to 1000 kn (compression and traction) and 3000 kn (compression) - assuming by the developed reference standards of some tasks of the national standards The tasks to be solved were to promote the extant reference force standards to the necessary metrological performances for calibration of the class 1 (ISO 376) force transducers.

For this purpose, the force calibration machines have to be able to ascertain a maximum relative uncertainty of 5 x 10-4. The difference between the uncertainty of the applied force and the uncertainty of calibration has been considered. The main difference is represented by the following reason: the uncertainty of calibration includes the behavior of the calibration standard during calibration, but the uncertainty of the applied force excludes these important component of the uncertainty. This is the reason why the uncertainty of the applied force is considered very similar with bmc the best measurement capability [5]. In order to achieve these purposes, the efforts have been concentrated to the comparison force standard machine of 1000 kn ( the most used in the force laboratory ). This machine is able to reproduce forces, between the limits of 10 kn and 1000 kn, in tensile and compression mode, by using a wide range of reference force transducers: 50 kn; 100 kn; 00 kn; 500 kn and 1000 kn, with high metrological performances. 3. EVALUATION OF THE UNCERTAINTY Generally, the measurand F FCM (the force reproduced by the standard machine FCM) is determined by the input parameters X 1, X...X n, according to the functional relation: F FCM = f(x 1, X,... X n ) (1) When the force is generated by mechanical devices, and the measurement of the force is performed by reference force transducers (comparison force standard machine), the input parameters are as follows: - result of the reference force transducer calibration (F Ref Tra ) - relative long-term drift of the reference force standard ( DriftRef ) - relative change in the calibration result due to the temperature change (c θ ) - relative deviation because of the variation of the applied force during a measurement by comparison process ( Stab ) The mathematical model followed the relation: F FCM = F ( 1 ) ( cθ ) ( ) () Re ftra Drift Re f Stab The force value reproduced by the standard machine has to be rectified with the results of the comparison by a deadweight calibration machine FSM. These comparisons conduce to the evaluation of some factors not yet completely defined (e. g. non-linearity, displacements pattern during calibration). The comparison is performed with transfer force standards. The rectification is carried out by the factor (1- Tras ), where Tras may be calculated in accordance with the accepted procedure [6]. The mathematical model of the reproduced force becomes: F FCM = F ( 1 ) ( cθ ) ( ) ( ) (3) Re ftra Drift Re f Stab Tras The evaluation of the standard measurement relative uncertainty of the reproduced force F FCM begins with the mathematical model, relation (3), assuming that the input values are not correlated. So, the combined relative variance is obtained from the random error propagation law: w( F FCM ) = Re ftra Drift Re f θ w ( F ) + w ( ) + w ( c ) + w ( ) + w ( ) (4) Stab Tras 3

where: w (F RefTra ): the relative variance of the calibration result of the reference force transducer w ( DriftRef ): the relative variance associated with the drft of the reference force transducer w (c θ ): the relative variance associated with the change in calibration result due to the temperature variation w ( Stab ) the relative variance associated with the variation of the force during the calibration w ( Tras ) the relative variance associated with the relative deviation determined for the generated force, due to traceing it back to the primary force standard machine The probability of distribution of the input parameters is presented in table 1: Table 1: The probability of distribution Input parameters Probability distribution Estimation of variance F Ref Tra normal u DriftRef rectangular a /3 c θ rectangular a /3 Stab triangular a /6 Tras normal u On the basis of the estimated relative uncertainty, in accordance with the relation (4), the uncertainty of the applied force stipulated by ISO 376, is calculated. In order to decrease the above-mentioned uncertainty for the comparison force machine of 1000 kn, the following opportunities were identified: - the use of reference standards with high metrological performances (repeatability, reproducibility and stability) - the accurate knowledge of the metrological characteristics of the reference transducers. The reference standards are periodically calibrated (usually every two years) in accordance with the deadweight calibration machines of PTB-Germany. The components of the hysteresis are evaluated in order to decrease their influence. - keeping adequate environmental conditions (especially the temperature) - ascertaining of an adequate stability of the applied force by the control of the loading. In order to minimize the influence on the uncertainty of the variation of the applied force, the identification of the system represented by the 1000 kn comparison force calibration machine was performed. There is presented in figures a and b the reply of the system in a free mode, to an input step of 100 kn for compression and respectively tensile force. A variation of the applied force of approximately 0,1% and a return of the loop of the pattern are observed. These will modify the metrological performances of the standard force machine. 10030 1000 v1 v v3 10030 1000 v1 v v3 10010 10010 force (dan) 10000 force (dan) 10000 9990 9990 9980 9980 9970 0 10 0 30 40 50 60 70 time (s) 9970 0,0 10,0 0,0 30,0 40,0 50,0 60,0 70,0 time (s) a) b) Figure : The reply of the FCM to an input of 100 kn in free mode 4

It comes out that the variation of the applied force depends on the rate of proximity to the ascertain value (kn/s), as well to the deflection of the transducer to be calibrated. The first dependence is direct proportionally, but the second is reversing proportionally. Taking into account these behaviors and using a suitable automated system, the variation of the applied force decreased to approximately 0,01% and the return of the loop of the loading pattern became negligible, (figure 3). force (dan) 10010 c1 10008 c 10006 c3 10004 1000 10000 9998 9996 9994 999 9990 50 55 60 65 70 75 80 time (s) Figure 3: The reply of the FCM to an input of 100 kn in the controlled mode The uncertainty of the standard comparison calibration machine of 1000 kn has been evaluated in accordance with EA 4/0 [7]. Transfer standard, which were previously calibrated by deadweight force calibration machines of 100 kn and 1000 kn in PTB, have been used. The results of the evaluation of the uncertainty of the applied force of the loading step of 100 kn, are presented in table. Some of the input parameters are estimated as a zero value, but have been considered for the uncertainty calculation: - relative drift of the transfer transducers ( DriftTra ), - relative deviation of the force realization by standard machine FSM ( Realization ), - relative hysteresis of FCM influenced by the hysteresis of FSM ( HysFCM ), - relative drift of the reference transducer ( DriftRef ), - relative deviation of the calibration result due to the temperature variation (c θ ), - relative variation of the applied force during the measurement by comparison ( Stab ), - relative deviation of the mean force values determinated between FCM and FSM ( RelDev ) Table. Uncertainty of the applied force for the comparison machine of 1000 kn for the 100 kn force step Quantity Estimation Relative halfwidth value a Distribution of probability Relative standard uncertainty Sensitivity coefficient Relative uncertainty contribution F RefTra 100 kn normal 5 x 10-5 1 5 x 10-5 DriftRef 0 kn,0 x 10-5 rectangular 1, x 10-5 1 1, x 10-5 c θ 0 kn 5,0 x 10-5 rectangular,9 x 10-5 1,9 x 10-5 Stab 0 kn 1,0 x 10-4 triangular 4,1 x 10-5 1 4,1 x 10-5 Realization 0 kn normal 1,0 x 10-5 1 1,0 x 10-5 F FSM 100 kn normal,5 x 10-5 1,5 x 10-5 F FCM 100 kn normal 4,5 x 10-5 1 4,5 x 10-5 HysFCM 0 kn 1, x 10-5 rectangular 1,0 x 10-5 1 1,0 x 10-5 DriftTra 0 kn,0 x 10-5 rectangular 1, x 10-5 1 1, x 10-5 RelDev 0 kn 1,8 x 10-4 triangular 7,3 x 10-5 1 7,3 x 10-5 F FCM 100 kn normal 1, x 10-4 Expanded relative uncertainty W = k. w (F FCM ) for k =,4 x 10-4 Specification of the uncertainty of the applied force 5 x 10-4 5

4. CONCLUSIONS The comparison force calibration machine with the nominal range of 1000 kn, tensile and compression mode, is used in the Force Laboratory Timisoara as a reference standard to calibrate sevaral force measurement devices. The requests of the applicants impose to achieve a relative uncertainty of the applied force of maximum 5 x 10-4, which is necessarry to calibrate the class 1 force transducers in accordance with ISO 376:1999. The main factors of influence about the uncertainty of the applied force which may be optimized are: the stability of the applied force and the associated uncertainty of the reference transducers. Through an appropriate control system of the loading force and using a wide range of reference force transducers having high metrological performances, an uncertainty of 5 x 10-4 has been attained. The main advantages regarding the use for calibration of a comparison force standard machine are: a low cost, a relative convenient maintenance and calibration, the sparing of the national force standards for usual calibration purposes. REFERENCES [1] A.Gherasimov,E.Ghita, The state of the art in the Romanian traceability system of force measurements, Proceedings of the 8 th International Conference on Flexible Technologies, Novi Sad/Serbia and Montenegro, 003, pp. 63-64. [] Romanian Law 178/003, Romanian Governmental Order 0/199, Romanian Low 11/1994. [3] ISO 376:1999. Metallic materials. Calibration of force-proving instruments used for the verification of the uniaxial testing machines. [4] BIPM, CEI, IFCC, ISO, IUPAC, IUPAP, OIML, International vocabulary of basic and general terms in metrology (VIM), the second edition, 1993. [5] EA-10/04 (EAL G):1996. Uncertainty of Calibration Results in Force Measurements. [6] A.Savla, Uncertainty scope of the force calibration machines, Proceedings of the XVI IMECO Congress, Vienna/Austria, 000, pp. 53-58. [7] EA-4/0 (EAL-R):1997. Expression of Uncertainty of Measurement in Calibration. Addresses of the Authors: Gherasimov Adrian, Romanian Bureau of Legal Metrology, Force Laboratory Timisoara, 64 Crizantemelor Str., Timisoara, Romania. E-mail: agherasimov@yahoo.com drmltm@yahoo.co.uk Conf.dr.eng. Ghita Eugen, POLITEHNICA University of Timisoara, Mechanical Engineering Faculty, Department Strength of Materials, Bd. M.Viteazul nr.1, Timisoara, Romania. E-mail: ghita@mec.utt.ro eghita63@yahoo.com 6