Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

Similar documents
IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

A calibration facility for automatic weather stations

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

An experimental investigation of the thermal performance of an asymmetrical at plate heat pipe

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Uncertainty Workshop: Overview of uncertainty factors in HTGHPs

Argon Triple-Point Device for Calibration of SPRTs

INVESTIGATION OF VAPOR GENERATION INTO CAPILLARY STRUCTURES OF MINIATURE LOOP HEAT PIPES

Experimental Analysis of Wire Sandwiched Micro Heat Pipes

Thermo-Fluid Performance of a Vapor- Chamber Finned Heat Sink

Application of the Multi-current Transient Hot-Wire Technique for Absolute Measurements of the Thermal Conductivity of Glycols

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

AN EXPERIMENTAL INVESTIGATION OF BOILING HEAT CONVECTION WITH RADIAL FLOW IN A FRACTURE

An experimental investigation on condensation of R134a refrigerant in microchannel heat exchanger

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

MEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G.

Enhanced Boiling Heat Transfer by using micropin-finned surfaces for Electronic Cooling

Introduction to Blackbody Sources

EVALUATION OF THE THERMAL AND HYDRAULIC PERFORMANCES OF A VERY THIN SINTERED COPPER FLAT HEAT PIPE FOR 3D MICROSYSTEM PACKAGES

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

A NUMERICAL APPROACH FOR ESTIMATING THE ENTROPY GENERATION IN FLAT HEAT PIPES

ANALYSIS ON THERMAL AND HYDRAULIC PERFORMANCE OF A T-SHAPED VAPOR CHAMBER DESIGNED FOR MOTORCYCLE LED LIGHTS

Thermal Field in a NMR Cryostat. Annunziata D Orazio Agostini Chiara Simone Fiacco

CONTENTS. Introduction LHP Library Examples Future Improvements CARMEN GREGORI DE LA MALLA EAI. ESTEC, October 2004

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE

Chapter 1 INTRODUCTION AND BASIC CONCEPTS

Department of Mechanical Engineering ME 96. Free and Forced Convection Experiment. Revised: 25 April Introduction

EXPERIMENTAL INVESTIGATION OF IN SITU PRESSURE MEASUREMENT OF AN OSCILLATING HEAT PIPE

Received: 17 December 2009 / Accepted: 11 March 2010 / Published online: 10 April 2010 Springer Science+Business Media, LLC 2010

Steady and Transient Heat Transfer Characteristics of Flat Micro Heatpipe

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium?

Heat Exchangers for Condensation and Evaporation Applications Operating in a Low Pressure Atmosphere

Comparison tests of cellular glass insulation for the development of cryogenic insulation standards

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer

NUMERICAL INVESTIGATION OF STEADY STATE AND TRANSIENT THERMAL PERFORMANCE OF A MULTI-CHIP VAPOR CHAMBER MOHAMMAD PARHIZI

PERFORMANCE CHARACTERISTICS OF A LONG HEAT PIPE

HEAT AND MASS TRANSFER. List of Experiments:

Solar Flat Plate Thermal Collector

AN EXPERIMENTAL STUDY OF THE FROST FORMATION ON A COLD SURFACE IN FREE CONVECTIVE FLOW

He II Heat transfer through a Corrugated Tube - Test Report

Calibration of Temperature Block Calibrators

Experimental Assessment of Thermal Conductivity of a Brick Block with Internal Cavities Using a Semi-scale Experiment

Thermal Systems. What and How? Physical Mechanisms and Rate Equations Conservation of Energy Requirement Control Volume Surface Energy Balance

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium?

Thermal Interface Material Performance Measurement

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES

Experimental investigation on up-flow boiling of R1234yf in aluminum multi-port extruded tubes

Calibration of temperature sensors within Length Standard Section of NMIJ

SIB 52 - THERMO Stakeholder meeting May 16

Phone: , For Educational Use. SOFTbank E-Book Center, Tehran. Fundamentals of Heat Transfer. René Reyes Mazzoco

HEAT TRANSFER THERMAL MANAGEMENT OF ELECTRONICS YOUNES SHABANY. C\ CRC Press W / Taylor Si Francis Group Boca Raton London New York

Chapter 18 Temperature, Heat, and the First Law of Thermodynamics. Thermodynamics and Statistical Physics

New analysis for determination of hemispherical total emissivity by feedback-controlled pulse-heating technique

THERMOHYDRAULIC BEHAVIOUR OF HEII IN STRATIFIED CO-CURRENT TWO-PHASE FLOW AT HIGH VAPOR VELOCITIES

Experimental Study of Convective Heat Transfer and Thermal Performance in the Heat-Sink Channel with Various Geometrical Configurations Fins

ADVANCES IN BIPOROUS WICK DESIGN AND TESTING FOR THERMAL GROUND PLANES

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

Isothermal Technology Ltd

Towards SI traceable humidity measurements with radiosondes

Comparison of Heat Transfer rate of closed loop micro pulsating heat pipes having different number of turns

Effect of Nanofluid Concentration on the Performance of Circular Heat Pipe

Uncertainty Analysis of Flat Plate Oscillating Heat Pipe with different Working Fluids

Introduction to Heat and Mass Transfer. Week 14

Semiconductor Thermal Resistance Standards versus Real Life. Bernie Siegal Thermal Engineering Associates, Inc.

THERMO-MECHANICAL ANALYSIS OF A COPPER VAPOR LASER

Thermal Performance Characterization of Embedded Pulsating Heat Pipe Radiators by Infrared Thermography

LIQUID FILM THICKNESS OF OSCILLATING FLOW IN A MICRO TUBE

STRONG DOUBLE LAYER STRUCTURE IN THERMIONIC VACUUM ARC PLASMA *

Performance Analysis of Pulsating Heat Pipes Using Various Fluids

Al-Saudia Virtual Academy Online tuiton Pakistan Online Tutor Pakistan. Heat

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

CHAPTER 4 THERMAL CONDUCTIVITY AND VISCOSITY MEASUREMENTS

Boiling and Condensation (ME742)

Evaporation Heat Transfer Coefficients Of R-446A And R-1234ze(E)

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

Standard Test Method for Thermal Conductivity of Refractories 1

Experimental and Theoretical Evaluation of the Overall Heat Loss Coefficient of a Vacuum Tube Solar Collector

Ben Wolfe 11/3/14. Figure 1: Theoretical diagram showing the each step of heat loss.

Correlation to predict heat transfer of an oscillating loop heat pipe consisting of three interconnected columns

Principles of Food and Bioprocess Engineering (FS 231) Problems on Heat Transfer

Temperature and Length. Pieter Greeff October 2013

Corrections for Thermal Expansion in Thermal Conductivity Measurement of Insulations Using the High-Temperature Guarded Hot-Plate Method

Examination Heat Transfer

Isotech Journal of Thermometry Index Volumes 1 10, 1990 to 2000

TEMPERATURE MEASUREMENT. GROUP 3 Noah Beauchamp Kyle Camacho Jack Connolly Curtis Yu

A NEW HEAT PIPE COOLING DEVICE

DIRECT RADIOMETRIC TECHNIQUES

DETERMINATION OF R134A S CONVECTIVE HEAT TRANSFER COEFFICIENT IN HORIZONTAL EVAPORATORS HAVING SMOOTH AND CORRUGATED TUBES

Heat and Mass Transfer Unit-1 Conduction

HEAT TRANSFER 1 INTRODUCTION AND BASIC CONCEPTS 5 2 CONDUCTION

PAPER 2 THEORY QUESTIONS

Numerical Simulation on Flow and Heat Transfer in Oscillating Heat Pipes

Abstract. 1. Introduction

USING MULTI-WALL CARBON NANOTUBE (MWCNT) BASED NANOFLUID IN THE HEAT PIPE TO GET BETTER THERMAL PERFORMANCE *

Temperature. 3

XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil

EXPERIMENTAL ANALYSIS OF R-134a FLOW CONDENSATION IN A SMOOTH TUBE

Transcription:

Int J Thermophys (2009) 30:257 264 DOI 10.1007/s10765-008-0495-9 Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements L. Rosso N. Koneva V. Fernicola Published online: 12 August 2008 Springer Science+Business Media, LLC 2008 Abstract The development of a flat heat-pipe-based hot plate operating in the temperature range from ambient to 200 C is presented here. The aim of the design was to improve the heat-transfer performance in order to provide almost ideal thermal uniformity of surface and volume temperature profiles, thereby ensuring a minimal temperature gradient and a small effective thermal resistance. The device was investigated to evaluate its performance in view of its potential use as a calibration apparatus for surface-temperature sensors. An accurate characterization was carried out to estimate the surface-temperature stability and uniformity under different thermal conditions. Measurement results showed temperature stability within 0.03 C and uniformity of the heat-pipe (HP) hot-plate surface better than 0.08 C. As a result of its high thermal conductivity, a small perturbation (<0.2 C) of the temperature field both on the surface and inside the device chamber, when a contact probe is applied on its surface, was obtained. The study of the HP hot plate and the performed tests suggest that such a device has potential as a calibration apparatus for surface-temperature sensors. Keywords Contact probe Heat pipe Hot plate Surface-temperature sensor 1 Introduction Several industrial and scientific applications call for practical and emissivity-independent measurements of the surface temperature of solid bodies. As contact-type thermometers are often used in these applications, a traceable calibration of such L. Rosso (B) V. Fernicola Thermodynamics Division, Istituto Nazionale di Ricerca Metrologica (INRIM), Strada delle Cacce 91, 10135 Torino, Italy e-mail: l.rosso@inrim.it N. Koneva Heat and Mass Transfer Insitute, Minsk, Belarus

258 Int J Thermophys (2009) 30:257 264 probes is required. Different calibration methods and facilities have been developed by the national metrology institutes (NMIs) over the last few years [1 4]. Calibration systems are generally based on a temperature-controlled hot plate that provides a stable and reproducible surface temperature. The temperature of the reference surface is obtained through extrapolation of the temperature values measured by two or more calibrated thermometers embedded into the plate at different depths. Other calibration setups ensure a uniform temperature field across the reference plate by partially immersing it in a thermostatic bath and maintaining its upper surface above the level of the liquid. In this case, a calibrated thermometer, embedded into the plate just under the calibration surface, enables the measurement of the surface temperature [5]. Where the above-mentioned calibration facilities are not available, temperature probes are calibrated by total immersion in a thermostatic bath. When a calibration by immersion is performed, the probe is brought into thermal equilibrium with the surrounding environment, i.e., the liquid bath. However, this condition is far from the actual operating conditions of the sensor, in which a sensorto-surface thermal interaction occurs. In this case, because of the presence of contact resistances at the interface between the sensor and the surface, thermal equilibrium is not established. As a consequence, the resulting temperature difference between the surface and the center of the sensor leads to a biased estimate of the temperature of the surface. On the other hand, the calibration of contact probes by means of hot plates takes place under conditions closer to those found in real applications, thus yielding a more realistic calibration correction for practical use. Of course, a contact thermometer cannot make surface-temperature measurements with the same accuracy as temperature measurements made in thermal equilibrium. In fact, the best measurement capabilities for surface-temperature measurements available at the European NMI facilities range approximately from 0.2 Cto2 C from ambient temperature to about 300 C. In order to establish the degree of equivalence of such national standards for surface temperature, a EUROMET intercomparison was recently carried out [6]. In order to improve the performance of the hot-plate systems, a flat heat spreader, based on heat-pipe (HP) technology [7], was designed and constructed at the HTMI and its potential application to surface thermometry for contact probe calibration was investigated. The operating principle of the device affords a very high thermal conductivity more than one order of magnitude greater than the thermal conductivity of metal blocks and potentially makes it a quasi-ideal heat source, thus resulting in a smaller thermal perturbation when the sensor to be calibrated contacts the reference surface. The high thermal conductivity enables estimation of the surface temperature through the measurement of a single thermometer embedded in the HP, thus avoiding the temperature extrapolation and the consequent errors. These features should improve the calibration uncertainty of surface-temperature sensors. The HP was integrated into a dedicated temperature control system, and the performance of the whole system was investigated at INRIM. An accurate characterization was carried out to evaluate the surface-temperature uniformity and stability over the operating range from ambient to about 200 C under different thermal conditions. A study of the perturbation effects introduced by the contact probe on the surface was also performed using an independent, remote measurement method based on phosphor thermography [8].

Int J Thermophys (2009) 30:257 264 259 2 Description of the Device 2.1 Principle of the HP Hot Plate The HP hot plate is a high heat transfer device that enables quick changes and stabilization of the surface temperature. Consequently, it is not influenced by the application point of a heat loss element on the surface itself. The fast thermal response is due to the evaporation condensation process, because the heat transfer depends on the velocity of the vapor flow and not on the heat transfer rate through the spreading surface [7,9]. The HP can transfer heat with a small temperature gradient between the condenser and the evaporator surface, resulting in a small temperature difference between the condenser and the temperature probe in the vapor channel. The working fluid vaporizes in the evaporation section and flows toward the cold regions. The latent heat of vaporization is transferred to the condenser surface where the vapor is cooled by natural convection and by the temperature probe applied to its external surface. The heat transfer medium (water with a total mass of 7 mg) constantly circulates inside the device. At equilibrium, the liquid phase of the heat transfer fluid is assumed to follow the predictions of fluid mechanics. The liquid returns in the form of a thin liquid film to the evaporator along the vertical adiabatic section (by gravity and by the capillary force) under the action of the capillary pressure gradient. The vapor forms directly above the vapor liquid interface in the evaporator. The mass flow rate is a maximum at the opening of the liquid transport channel of the adiabatic section, decreasing to zero at the central point of the condenser and evaporator where the liquid film also becomes thinner. 2.2 Design of the HP Hot Plate The construction of the HP hot plate followed the principles of a gravitational heat pipe, where the condenser is placed above the evaporator as shown in Fig. 1. In such a case, the thermal performance of the device depends on the effective thermal conductivity of the wick, since its variable thermal resistance largely contributes to the overall thermal resistance. Modeling of the steady-state operation was carried out in the temperature range from ambient to about 200 C by approximating the temperature distribution through a network of thermal elements, with defined thermophysical properties, connected by constant thermal resistances [10]. The dimensions and geometry of the device were chosen to match the size of the hot plates used in the INRIM surface thermometry calibration system [11]. The HP is a vacuum-sealed cylinder, 100 mm outer diameter and 86 mm inner diameter, made of OFHC copper. Each inner wall (wall thickness 7 mm) is coated with a high thermal conductivity sintered copper powder medium. A porous medium was used for better temperature uniformity and to improve the heat transfer, even with a small temperature difference. The porous structure (metal particles with a dendritic shape and a porosity value of 0.55) has a uniform thickness of 2 mm and promotes capillary surface wetting in the condenser, the evaporator, and the adiabatic zones. At the same time, it avoids any increase in the thermal resistance at the inner walls due to film boiling. Such a porous structure

260 Int J Thermophys (2009) 30:257 264 2 Cooling 6 8 3 1 7 4 Heating Fig. 1 HP structure and principle: 1 evaporation condensation cycle; 2 surface temperature probe; 3 vapor channel; 4 porous sintered wick; 5 evaporator surface; 6 condenser surface; 7 filling line; 8 PRT probe well also allows uniform spreading of the liquid film in the condenser. Heat is supplied to the device by thermal conductance from a heat source at the bottom to the evaporation section. Inside the HP, a single axial vapor-flow channel is formed in which the water vapor pressure sets the internal pressure. A temperature well (55 mm in length and 6.5 mm in outer diameter) runs through the adiabatic zone in the middle of the vapor channel; it houses a short-stem PRT used for measurement of the equilibrium temperature. 5 3 Experiments 3.1 Temperature Control System The setup used to test the device is shown in Fig. 2. The reference surface temperature is generated on the condenser surface of the HP. The device was mounted on a flat, round heater and was surrounded by a micro-porous thermal insulator. The system was equipped with a guard-ring heater to reduce the radial heat loss, thereby achieving better temperature uniformity. A PID controller drove a voltage-controlled power supply to adjust the block temperature. A Pt-100 sensor was used as the control sensor. The equilibrium temperature was measured by a short-stem PRT embedded in the HP itself and connected toa81/2-digitmultimeter. 3.2 Characterization of the HP Hot Plate The tests on the HP hot plate were aimed at evaluating the performance of the device in view of its use as a calibration apparatus for surface temperature. To this end, its metrological characterization was addressed to estimate the temperature stability and surface-temperature uniformity for different thermal equilibrium conditions, i.e., with and without the contact probe to be calibrated. The temperature stability, obtained through measurements of the embedded PRT, resulted in a peak-to-peak value of less than 0.03 C over the whole operating range. A remote measurement method, based

Int J Thermophys (2009) 30:257 264 261 Guard heater Reference surface DVM PRT Insulating material Control system PID Fig. 2 Block diagram of the experimental system used to test the HP hot-plate device on fluorescent decay-time temperature sensors, was exploited to evaluate the surfacetemperature uniformity [8]. A two-dimensional array of fluorescence sensing dots was applied to the HP hot plate by means of a thick-film coating technique. An optical fiber, close to the surface, conveyed the excitation light from a laser diode to the various point sensors and transmitted the fluorescent response to the detection stage. In such a way, a thermal map of the HP hot-plate surface was obtained. Figure 3 shows the surface temperature as a function of radial position as measured by the fluorescence sensors at various temperatures (50 C, 100 C, 150 C, and 185 C) along two orthogonal radial directions depicted by filled and open symbols, respectively. For each temperature, the broken line represents the best fit to the resulting profile. A temperature uniformity of better than 0.08 C was obtained over the whole temperature range. Further investigations were carried out to evaluate the perturbation of the surfacetemperature profile, and the ability of the device to recover its equilibrium temperature, when a probe is put in contact with its surface. The temperature along a radial direction was measured by means of the fluorescent point sensors when the probe was applied and then removed from the center of the HP hot-plate surface. Figure 4 shows the resulting surface-temperature change detected at 100 C, 150 C, and 180 Casa function of the radial position. As expected, the effect due to the contact probe is a maximum near the probe (at x = 0) and decreases with distance. The thermal perturbation introduced by the contact probe was also investigated by simultaneously comparing the surface temperature and the inner equilibrium temperature before and after applying a probe to its surface. Figure 5a represents the surface-temperature change at 150 C as obtained from fluorescence decay-time measurements of the sensors placed near the application point of the contact probe, while Fig. 5b shows the inner temperature as measured by the embedded PRT. A comparison of Fig. 5a and b highlights that the temperature change is slightly higher on the surface than inside the device (0.2 C vs. 0.14 C). This is consistent with the fact that the disturbing effect of the probe is higher on the surface near the contact point. However, it should be noted that the disturbance is comparatively small in magnitude, both on the surface and inside the HP chamber, thanks to the high thermal conductivity of the device.

262 Int J Thermophys (2009) 30:257 264 Surface temperature, C 186.10 186.00 152.10 152.00 Surface temperature, C 102.10 102.00 50.10 50.00 49.90 0 10 20 30 40 50 Radial position, mm Fig. 3 HP hot-plate surface temperature along two radial directions (filled and open symbols) at various temperatures. Broken lines are best fit lines to show the temperature profile Surface temperature change, C 0.00-0.05-0.10-0.15-0.20-0.25-0.30 0 10 20 30 40 Radial position, mm 100 C 150 C 180 C Fig. 4 Surface-temperature variation of the HP hot plate due to the contact probe applied to the center of its surface at various temperatures 4 Discussion The investigations of the developed HP hot plate were useful in evaluating its characteristics under different thermal conditions. The actual operating range of the device, due to the use of water as the working fluid, is limited to 200 C because of the internal pressure limits; however, such a range can be easily extended by using other working fluids. The measurement results, obtained by fluorescence decay-time thermometry, showed the temperature uniformity of the HP hot plate surface to be better than 0.08 C.

Int J Thermophys (2009) 30:257 264 263 Fluorescence decay-time, ms 2.5150 2.5145 2.5140 2.5135 2.5130 0.2 C (a) Temperature, C 150.40 150.30 150.20 (b) 150.10 100 200 300 400 500 600 Time, s Fig. 5 (a) Change of the surface temperature and (b) the inner equilibrium temperature of the HP hot plate before and after applying a probe to its surface It is interesting to note that a continuous change in the slope of the surface-temperature profile occurs as the temperature increases from ambient to 200 C (see Fig. 3). The surface temperature and its stability are perturbed when a probe is put in contact with the surface, although the measurements showed a comparatively small change in the surface temperature (about 0.2 C) and an even smaller change inside the device (about 0.14 C). The measurements also pointed out that the disturbance introduced by a probe on the temperature field inside the device is much lower than that obtained under the same conditions with a hot-plate system [11]. Thanks to its high thermal conductivity, this device is less sensitive to thermal perturbation when calibrating a probe. As a result, the equilibrium temperature of the chamber, as measured by the embedded PRT, is a good estimate of the actual reference surface temperature, even when the latter is perturbed by the presence of the probe to be calibrated. This feature suggests a different approach to the calibration procedure compared to that generally followed for hot-plate-based calibrations. When a probe to be calibrated is applied to its surface, the reference temperature can be determined through the measurement of a single temperature using the embedded PRT. The above results show that such a device has the potential to be used as a calibration apparatus for contact surface-temperature sensors. In principle, it could also be used for the calibration of radiation thermometers, although an accurate determination of the surface emissivity would also be required. Acknowledgments The authors would like to acknowledge that parts of the work were supported by the INTAS program of European Commission within the project INTAS Ref. No. 3-55-1757. Thanks also go to the reviewer for the useful suggestions and comments given to the authors.

264 Int J Thermophys (2009) 30:257 264 References 1. IMEKO TC12, in Proceedings of the Workshop on Surface Thermal Measurements, Budapest, Hungary, (1995) 2. F. Edler, M. Gorgieva, J. Hartmann, M. Wagner, in Proceedings of TEMPMEKO 2001, 8th International Symposium on Temperature and Thermal Measurements in Industry and Science, ed. by B. Fellmuth, J. Seidel, G. Scholz (VDE Verlag, Berlin, 2002), pp. 1105 1110 3. R. Morice, E. Andras, E. Devin, T. Kovacs, in Proceedings of TEMPMEKO 2001, 8th International Symposium on Temperature and Thermal Measurements in Industry and Science, ed. by B. Fellmuth, J. Seidel, G. Scholz (VDE Verlag, Berlin, 2002), pp. 1111 1116 4. EA Temperature and Humidity Expert Group, Questionnaire Results on Surface Contact Thermometer, European Cooperation for Accreditation (1999) 5. M.J. De Groot, High Temp.-High Press. 29, 591 (1997) 6. E. Andras, Comparison of the reference surface temperature apparatus at NMIs by comparison of transfer surface temperature standards, EUROMET Project P635 Final Report (Budapest, Hungary, 2003) 7. P.D. Dunn, D.A. Reay, Heat Pipes (Pergamon Press, Oxford, 1994) 8. L. Rosso, V.C. Fernicola, A. Tiziani, in Temperature: its Measurement and Control in Science and Industry, vol. 7, part 2, ed. by D.C. Ripple (AIP, Melville, New York, 2003), pp. 1045 1050 9. S.W. Chi, Heat Pipe Theory and Practice (McGraw-Hill, New York, 1976) 10. N.S. Koneva, L.S. Domorod, in Proceedings of the 6th International Seminar on Heat Pipes, Heat Pumps, Refrigerators, Minsk, Belarus, (2005) 11. F. Arpino, M. Dell Isola, V.C. Fernicola, A. Frattolillo, L. Rosso, in Proceedings of TEMPMEKO2004, 9th International Symposium on Temperature and Thermal Measurements in Industry and Science,ed. by D. Zvizdic (FSB/LPM, Zagreb, Croatia, 2004), pp. 635 641