OPERATING INSTRUCTIONS

Similar documents
Photoluminescence Spectrometer (FLS980)

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

THE DIFFRACTION GRATING SPECTROMETER

NORTHERN ILLINOIS UNIVERSITY PHYSICS DEPARTMENT. Physics 211 E&M and Quantum Physics Spring Lab #9: Diffraction Spectroscopy

CHEM Outline (Part 15) - Luminescence 2013

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

Microplate Spectrophotometer

FLUORESCENCE APPLICATIONS

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

39th International Physics Olympiad - Hanoi - Vietnam Experimental Problem

The Quantum Model of the Hydrogen Atom

Atomic Spectra. d sin θ = mλ (1)

Chapter 15 Molecular Luminescence Spectrometry

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers

Fluorescence Spectrophotometry

INSTRUCTION MANUAL ANALOG COLORIMETER

Millikan Oil Drop Experiment

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Singlet. Fluorescence Spectroscopy * LUMO

Rydberg constant from atomic spectra of gases

Experiment 3: Radiative Lifetime Determination by Laser-Induced Fluorescence

APAS Laboratory { PAGE } Spectroscopy SPECTROSCOPY

PHYS 3324 Lab Millikan Oil Drop Experiment: Demonstration of the Quantization of Charge

Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI

FLS980 Series Reference Guide

Experiment 24: Spectroscopy

An Ultra Low Resistance Continuity Checker

Brown University PHYS 0060 Physics Department LAB B -190

SPECTROPHOTOMETERS. Visible and UV-Visible

Department of Physics and Astronomy University of Iowa 29:137 Astronomical Laboratory Fall 2011 Lab 4: Stellar Spectroscopy

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Evaluating Labsphere s new UV-2000

To successfully square an HD series machine, you will need the following tools:

X-ray practical: Crystallography

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO

Ph 3455/MSE 3255 Experiment 2: Atomic Spectra

Clinical Chemistry (CHE221) Professor Hicks Week 7. Endpoint vs Kinetic Techniques. Instrumentation and Assay techniques

Optics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P

Practical 1P4 Energy Levels and Band Gaps

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

Practical 1P4 Energy Levels and Band Gaps

C101-E145 TALK LETTER. Vol. 17

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms

Atomic Spectra HISTORY AND THEORY

high temp ( K) Chapter 20: Atomic Spectroscopy

Protocol for Use of BCMP Jasco J-815 Circular Dichroism Spectropolarimeter

Sample Alignment (2D detector) Part

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

Physics 1C OPTICAL SPECTROSCOPY Rev. 2-AH. Introduction

Pearson Edexcel GCSE Physics/Science Unit P1: Universal Physics

Instruction Manual for EP-20 e/m of the Electron Apparatus

Any first year text, sections on atomic structure, spectral lines and spectrometers

Experiment #9. Atomic Emission Spectroscopy

Jasco V-670 absorption spectrometer

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 FRAUNHOFER DIFFRACTION

Chem Homework Set Answers

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

25 Instruments for Optical Spectrometry

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

Photoelectric Photometry of the Pleiades Student Manual

Brooklyn College Department of Chemistry

University of Massachusetts, Amherst

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

n(λ) = c/v(λ). Figure 1: Dispersion curves for some common optical glass types.

Physics 24, Spring 2007 Lab 2 - Complex Spectra

Optical Pumping of Rubidium

Investigation #9 OBSERVATION OF THE PHOTOELECTRIC EFFECT

The Spectrophotometer and Atomic Spectra of Hydrogen Physics 246

UV-Vis optical fiber assisted spectroscopy in thin films and solutions

4. Dispersion. The index of refraction of the prism at the input wavelength can be calculated using

Analyzing Line Emission Spectra viewed through a Spectroscope using a Smartphone

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I

Experimental Design and Data Collection

Experiment 6a: Excited-state properties of 2-naphthol (the acidity constants) (Dated: October 29, 2009)

and Environmental Science Centre

Atomic Spectroscopy. Absorption and Emission Spectra. Lodovico Lappetito. SpettroscopiaAtomica - 15/07/2015 Pag. 1

Experiment 12: SPECTROSCOPY: EMISSION & ABSORPTION

RED. BLUE Light. Light-Matter

Physical Chemistry Advanced laboratory course. Excimer formation

FOCUS 30/FOCUS 35 Field Calibration with Survey Pro Field Software

EXPERIMENT #3 A Beer's Law Study

4. Dispersion. The index of refraction of the prism at the input wavelength can be calculated using

Chapter 6 Photoluminescence Spectroscopy

LC-4: Photoelectric Effect

CEE 697K Bromide Lab

Lab 5 - ELECTRON CHARGE-TO-MASS RATIO

1901 Application of Spectrophotometry

Atomic Absorption Spectroscopy (AAS)

High photostability and enhanced fluorescence of gold nanoclusters by silver doping-supporting information

LAB 10: OPTICAL MATERIALS AND DISPERSION I

High Resolution Optical Spectroscopy

Raman and stimulated Raman spectroscopy of chlorinated hydrocarbons

Radioactivity APPARATUS INTRODUCTION PROCEDURE

Spectrum 22ED Spectrophotometer #FOT Absorbency 0A to 1.99A Photometric accuracy 2.0% T Power requirement 115/230 VAC; Hz automatic Size

Measurement method for the proficiency testing program

ATOMIC PHYSICS PHOTOELECTRIC EFFECT Practical 2 DETERMINATION OF PLANCK S CONSTANT BY MEANS OF THE STOPPING POTENTIAL

CH 112 Special Assignment #4 Chemistry to Dye for: Part C

Transcription:

OPERATING INSTRUCTIONS for the Perkin Elmer Model 650-10S Fluorescence Spectrophotometer

Re: Operating Instructions PE 650 2 The following flow diagram should help to guide you through the steps necessary to run a fluorescence spectrum. For routine fluorescence spectra sections A, B, C, and E should be consulted. The less routine instructions are left for later sections; these may usually be skipped for an introductory laboratory using the instrument. Turn on and warm up spectrophotometer, Xenon lamp and recorder, see Section A Do you know the excitation or emission wavelengths of your sample? No Yes Choose excitation and emission wavelengths See Section F Choose monochromator slit widths See section G Prepare recorder See Section B Emission or excitation spectrum Emission Excitation See Section C See Section D Set Excitation Wavelength Set Emission Wavelength Find wavelength of maximum emission And set sensitivity controls for full Scale deflection Find wavelength on excitation, which gives maximum emission intensity and set sensitivity controls for full scale deflection Set starting wavelength Set starting excitation wavelength Run Spectrum Run Spectrum Turn off lamp and allow to cool for 5 minutes with fan running see Section E Turn off xenon lamp power supply, instrument and recorder. Place the pen on a piece of plastic.

Re: Operating Instructions PE 650 3 A. Preparatory Steps 1. Turn the POWER switch on the xenon lamp power supply to ON. 2. Make certain that the fan in the light source compartment is operating properly and ten or fifteen seconds after turning the POWER switch to ON depress the START switch on the xenon lamp power supply. 3. Turn on the POWER switch on the spectrophotometer. 4. Turn the recorder POWER switch from the OFF position to the AMP position. This allows the recorder to warm up. Set the recorder switch box to fluorescence. WARNING: High voltage is applied when lighting the xenon lamp and the lamp is kept at a high temperature when lit. Do not touch the lamp compartment during operation of the lamp. 5. If stable analytical data is to be obtained, wait about 20 minutes before making sample runs. 6. Preliminary Control Settings: Confirm that the switches and knobs listed below are set as indicated. Wavelength SCAN switch: STOP RESPONSE: NORM MODE switch: NORMal SCAN SPEED: 120 nm/min ZERO SUPPRESSION Knob: OFF On recorder: PM GAIN Control: NORMal POLARITY: + Shutter: Closed. (Out) PHOTOMETRIC RANGE: 1 (100% T) SCAN SPEED: 120 nm/min B. Recorder Preparation 1. Turn the recorder POWER switch from the AMP position to the SERVO position. 2. Set the MEAS/ZERO switch at the ZERO position and adjust the recorder pen to the zero position on the chart by turning the ZERO control. (The zero level is normally located on the left side of the chart at a convenient line.) 3. Release the recorder MEAS/ZERO switch to the MEAS position and adjust the recorder pen exactly to the zero position on the chart by turning the ZERO ADJUST Knob on the instrument. Note: Once the zero level has been adjusted properly, zero adjustment is not required for subsequent measurements. The recorder is now ready to operate. The recorder chart drive will operate when the instrument SCAN switch is set to EX or EM, and the chart will move at a speed set with the recorder CHART SPEED control. C. Measuring Procedures, Emission Spectra The ratio mode (MODE switch in NORMAL position) should be used for all spectra and data collection. (The ratio mode provides stable performance for extended periods, permitting quantitative determinations.) The energy mode (MODE switch in ENERGY position) is primarily for focus and alignment of the source.

Re: Operating Instructions PE 650 4 Note: To check zero, pull out (close) the sample shutter. (When it is not required to pass light through the sample, the shutter should be closed to prevent photo decomposition of the sample, which could occur because of excessive irradiation.) 1) Perform the preparatory steps as instructed in Sec. A and B. 2) Make certain that the MODE switch is in the NORMAL position. 3) Fill a cell with sample and place the cell in the cell holder. Make certain that the cell sits properly in the holder. 4) Set the PM GAIN switch to the NORMAL position. 5) Set the EXCITATION and EMISSION WAVELENGTH controls to the optimum wavelengths position for the sample (see Sec. F). 6) Push in (open) the sample shutter. 7) With the SENSITIVITY RANGE and SENSITIVITY FINE controls, bring the recorder pen to between 90 and 100%. 8) Rotate the EMISSION WAVELENGTH control to the shortest wavelength in the wavelength range in which spectral recording is to take place. 9) Set the instrument SCAN SPEED switch and the recorder chart speed selector switch for the desired speeds. Setting both to the same speed gives a presentation of 10 nm per division. See Sec. H. 10) Bring the recorder pen in contact with the chart paper. 11) Turn the wavelength SCAN switch from the STOP to the EM position. 12) After recording the spectrum, set the wavelength SCAN switch at the STOP position. 13) Raise the recorder pen. 14) Pull out (close) the sample shutter. D. Excitation Spectra 1) Perform steps 1-8 of Sec. C. 2) Set the EXCITATION WAVELENGTH control for the shortest wavelength range intended for recording. 3) Set the instrument SCAN SPEED switch and the recorder chart speed selector switch for the desired speeds. 4) Bring the recorder pen in contact with the chart paper. 5) Turn the wavelength SCAN switch from the STOP to the EX position. 6) After recording the spectrum, set the wavelength SCAN switch at the STOP position. 7) Raise the recorder pen and marker. 8) Pull out (close) the sample shutter. E. Shutdown Procedure To shutdown the instrument proceed as follows. 1) Set the Wavelength SCAN switch at STOP. 2) Close the shutter. 3) Turn off the xenon lamp power supply and then quickly turn it back on again. This turns the lamp off but allows the lamp compartment fan to continue running to cool the xenon lamp. Let the fan run for at least 5 minutes. 4) Set the recorder POWER switch to OFF.

Re: Operating Instructions PE 650 5 5) Turn the spectrophotometer OFF. 6) When the lamp compartment is cool turn off the lamp power supply. 7) Sign out in the log book. Place the recorder pin on a piece of plastic. IMPORTANT Notes for Correct Measurement 1) When setting wavelengths, always turn the WAVELENGTH dials from shorter to longer wavelengths. 2) When setting slit widths, always turn the SLIT controls from narrower to wider slit widths. 3) When a cell is put into the cell holder, it should always be oriented in the same direction. 4) The shutter should be closed (pulled out) as long as it is not necessary to open it for sample irradiation. 5) When a cell is filled with a volatile sample, cover it with the cover plate furnished. F. Determining Excitation and Emission Wavelengths When either the excitation or the emission wavelength, or both, are unknown, they can be determined by following the instructions given below. 1) Excitation Wavelength Record an excitation spectrum of the unknown sample with the emission monochromator set at zero order on the EMISSION WAVELENGTH dial. The zero order setting lets all the fluorescence fall on the detector. The wavelengths at which peaks appear correspond to the excitation wavelengths for the sample. Choose the excitation wavelength that gives the maximum fluorescence emission signal. 2) Emission Wavelength Record an emission spectrum of the unknown sample with the excitation monochromator set at the wavelength chosen in Section F1. The zero order setting allows white light to fall on the sample. The wavelengths at which peaks appear correspond to the emission wavelengths of the sample. Notes: For measuring samples where the excitation wavelength and emission wavelength are very close to each other, it will be advantageous to shift the excitation wavelength to the shorter wavelength side of the band to avoid overlap and minimize the effect of Rayleigh scattering on the wavelength of interest. The monochromators may be turned by hand if the SCAN switch is in the STOP position. Scanning by hand will save time. G. How to Select Proper Slit Width Analytical results can be optimized by selecting the proper slit width or spectral bandpass. The factors, which are influenced by slit width, are: a) the narrower the slits the better the resolution. b) the wider the slits the better the signal to noise, and therefore the more sensitive the method.

Re: Operating Instructions PE 650 6 1. If the optimum slit widths are not known, set the EXCITATION and EMISSION SLIT dials at 10 nm. 2. The table following includes recommended slit widths for various analytical purposes. RECOMMENDED SLIT WIDTHS FOR VARIOUS ANALYTICAL PURPOSES Analytical Excitation Emission Remarks Purpose Slit Slit Excitation Narrow Wide For measuring broad Spectra band spectra, both the Emission Wide Narrow slits can be wide. Spectra Quantitative Wide Wide For obtaining high S/N Ratio, slit width should be As wide as possible. Analysis of Select excitation light Photochem- Narrow Wide as weak as possible and ically reactive open the shutter only while samples the samples are to be irradiated. 3. To test if the slit widths are narrow enough for adequate resolution run a spectrum then, decrease the slit width and run the spectrum again. If there is no significant change, the slit widths are adequate. H. Determining Wavelength Scanning Speed It is generally advantageous to select fast scanning speeds for samples showing broad spectra, and slow scanning speeds for resolution. A high scanning speed is possible for quick analyses. Care should be taken that a fast enough pen response is used at fast speeds to follow bands accurately. USE OF RESPONSE SWITCH The RESPONSE switch should be set as specified below for the conditions listed: FAST - for recording sharp spectra, measuring phosphorescence lifetime, and quickly tracing spectra. NORM for (most) general measurements. SLOW - for quantitative analysis and analysis requiring exceptional stability.

Re: Operating Instructions PE 650 7 The table below summarizes settings of the SLIT control, SCAN SPEED selector, and RESPONSE switch (A * indicates a rarely used combination of settings.) GENERAL GUIDE FOR RESPONSE SETTING Scan Slit 3 nm or 3 nm 6 nm 10 nm or Speed narrower ~ 6 nm ~10 nm -wider 15 nm/min SLOW SLOW SLOW SLOW NORM 30 nm/min NORM NORM NORM NORM SLOW SLOW SLOW 60 nm/min NORM NORM NORM NORM FAST 120 nm/min FAST NORM NORM NORM FAST 240 nm/min * FAST FAST NORM FAST 480 nm/min * * FAST FAST