FG5 Absolute Gravimeter

Similar documents
To the best of our knowledge, the FG5 gravimeter represents the current state-of-the-art in the measurement of absolute gravity.

Absolute gravimetry. Gabriel Strykowski Geodynamics Department DTU-Space (formerly Danish National Space Center)

Absolute and Relative Gravity Measurements at the Istituto Nazionale di Ricerca Metrologica (INRIM) of Turin (Italy) in June 2006

ABSOLUTE GRAVITY MEASUREMENTS

Absolute Gravity Observation at the National Measurement Institute

Intermediary Report - August Prof. Dr. Olivier Francis and Prof. Dr. Tonie Van Dam

ABSOLUTE GRAVITY MEASUREMENTS IN LOMONOSOV, SAINT-PETERSBURG, RUSSIA

Final Report. January Prof. Dr. Olivier Francis

Results from a new interferometric ballistic differential gravity meter

Physics 476LW Advanced Physics Laboratory Michelson Interferometer

The Plasma Phase. Chapter 1. An experiment - measure and understand transport processes in a plasma. Chapter 2. An introduction to plasma physics

Experiment 3 1. The Michelson Interferometer and the He- Ne Laser Physics 2150 Experiment No. 3 University of Colorado

Intercomparison of Relative and Absolute Gravimeters. Olivier FRANCIS

Test and Measurements conference 2016 DESIGN AND DEVELOPMENT OF FREE FALL ABSOLUTE GRAVITY MEASURING SYSTEM USING PNEUMATIC ACTUATORS

GEOPHYSICS AND GEOCHEMISTRY Vol.III - Gravimetric Measurement Techniques - V. A. Childers GRAVIMETRIC MEASUREMENT TECHNIQUES

Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown and Twiss Photon Antibunching Setup

Istituto Nazionale di Ricerca Metrologica

Absolute Gravity Measurements at Maitri, Antarctica

Istituto Nazionale di Ricerca Metrologica

Gravimètre Quantique Absolu : une utilisation opérationnelle des atomes froids pour la mesure de gravité. Dr. Jean Lautier-Gaud Journée CNFGG

6WDWXVRI/,*2. Laser Interferometer Gravitational-wave Observatory. Nergis Mavalvala MIT IAU214, August 2002 LIGO-G D

Measurments with Michelson interferometers

PHY 111L Activity 2 Introduction to Kinematics

Saturated Absorption Spectroscopy (Based on Teachspin manual)

SYRTE - IACI. AtoM Interferometry dual Gravi- GradiOmeter AMIGGO. from capability demonstrations in laboratory to space missions

An Overview of Advanced LIGO Interferometry

One Year with Our Absolute Gravimeter

Quantum and Nano Optics Laboratory. Jacob Begis Lab partners: Josh Rose, Edward Pei

Characterization of an Atom Interferometer Gravimeter with Classical Sensors for the Use in Geodesy and Geophysics

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

Past and Future General Relativity Experiments: Equivalence Principle, Time Delay, and Black Holes. Irwin Shapiro 21 October 2005

Lab 2: Mach Zender Interferometer Overview

The Michelson Interferometer and the He-Ne Laser Physics 2150 Experiment No. 3 University of Colorado

Time of Flight Mass Spectroscopy and Velocity Map Imaging

The Stanford Gravitational Reference Sensor

The CLIC active prealignment studies H. MAINAUD DURAND, T. TOUZE CERN

THE ACTIVE PREALIGNMENT OF THE CLIC COMPONENTS H. MAINAUD DURAND, T. TOUZE CERN

DIODE LASER SPECTROSCOPY

Atomic Quantum Sensors and Fundamental Tests

Constant velocity and constant acceleration

PHY 221 Lab 9 Work and Energy

Atom interferometry applications in gravimetry

Experiment 6: Interferometers

ONE YEAR WITH OUR ABSOLUTE GRAVIMETER

Lab 4: Projectile Motion

Experiment O-2. The Michelson Interferometer

3 LIGO: The Basic Idea

The Michelson Interferometer

Towards compact transportable atom-interferometric inertial sensors

Lab 5: Projectile Motion

PS210 - Optical Techniques. Section VI

Michelson Interferometry and Measurement of the Sodium Doublet Splitting

Particle-Wave Duality and Which-Way Information

Reservoir Monitoring Using Hybrid Micro-Gravity Measurements in the Takigami Geothermal Field, Central Kyushu, Japan

CESIUM ATOMIC FOUNTAIN CLOCKS AT NMIJ

Some Topics in Optics

Midterm Solutions. 1 1 = 0.999c (0.2)

Gravitational tests using simultaneous atom interferometers

Introduction. Procedure. In this experiment, you'll use the interferometer to EQUIPMENT NEEDED: Lens 18mm FL. Component holder.

Fourier Transform Spectrograph Development Project

Partner s Name: EXPERIMENT MOTION PLOTS & FREE FALL ACCELERATION

Newton s 2nd Law with demonstration track and Cobra4

Plasma Formation and Self-focusing in Continuum Generation

Michelson Interferometer

1

Characterization of Picometer Repeatability Displacement Metrology Gauges

Michelson Interferometry Hassan Mirza

Development of a cryogenic compact interferometric displacement sensor

BALKAN PHYSICS LETTERS Bogazici University Press 15 November 2016 BPL, 24, , pp , (2016)

arxiv: v1 [physics.atom-ph] 19 Jun 2014

David McIntyre. A slow beam of laser cooled rubidium atoms will be used as the matter-wave source. The atom

PHYS 229: Experiment 1 Expansion Coefficients of Copper and Invar Bars Through Laser Interferometry

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a).

Impact force measurement of a plastic sheet using drop ball test by the Levitation Mass Method (LMM)

MASS SPECTROMETRY. Topics

2 Each satellite will have two test masses, each being the end mirror for an interferometer.

Magnetic Field measurements with Csmagnetometers

Michelson Interferometer. crucial role in Einstein s development of the Special Theory of Relativity.

A PRELIMINARY ALIGNMENT PLAN FOR RIA AT MSU

Squeezed Light Techniques for Gravitational Wave Detection

UNIVERSITY OF SOUTHAMPTON

A longitudinal wave travels through a medium from left to right.

PHY 221 Lab 7 Work and Energy

Interference. Reminder: Exam 2 and Review quiz, more details on the course website

F.G. Major. The Quantum Beat. The Physical Principles of Atomic Clocks. With 230 Illustrations. Springer

Week 14 The Simple Pendulum

Gravitational Waves & Precision Measurements

Standard Small Angle Generator Using Laser Interferometer

Multiple charge states of titanium ions in laser produced plasma

OPSE FINAL EXAM Fall 2016 YOU MUST SHOW YOUR WORK. ANSWERS THAT ARE NOT JUSTIFIED WILL BE GIVEN ZERO CREDIT.

AN OVERVIEW OF LIGO Adapted from material developed by Brock Wells Robert L. Olds Junior High School, Connell, WA August 2001

Measuring the Universal Gravitational Constant, G

Progress on the Design of the Magnetic Field Measurement System for elisa

Optical Pumping of Rubidium

Zeeman Effect. Alex Povilus Physics 441- Fall 2003 December 20, 2003

Speed of Light in Air

PHYSICS ADMISSIONS TEST SAMPLE PAPER (2015 style, issued September 2015) Time allowed: 2 hours

Machine Positioning Uncertainty with Laser Interferometer Feedback

Waves Part 3B: Interference

Optical Interferometry

Transcription:

FG5 Absolute Gravimeter Micro-g LaCoste Derek van Westrum, Ph.D. www.microglacoste.com derek@microglacoste.com

FG5 Specifications Accuracy: 2 μgal (observed agreement between FG5 instruments) Precision: at a quiet site, 10s drop interval, 15μGal/sqrt(Hz) [eg. About 1 μgal in 3.75 minutes or 0.1microGal in 6.25 hours] Operating dynamic range: World-Wide Operating temperature range: 15 C to 30 C

FG-5 Principle of Operation Vacuum Chamber Freefalling Upper Mirror Interferometer A freely falling reflective test mass is dropped in a vacuum. This causes optical fringes to be detected at the output of an interferometer. This signal is used to determine the local gravitational acceleration. Interference Detector Stationary Lower Mirror

Interferometry MAX min λ/2 Laser B.S. Photodiode fringe signal sweeps in frequency as test mass falls under influence of gravity 5 4 Fringes 2 sin( 2πx 2 ) 0 2 4 Michelson s interferometer 5 0 1 2 3 4 5 0 x time recorded (w.r.t. rubidium oscillator) at each minimum creating (t,d) pairs at every λ/2 5

g Determination Fringe = l/2 x i For each x i, a measured time t i, The following function is fitted to the data x i,t i : ~ x = x + v t + i 0 0 i g ~ 0ti 2 2 ~ 2 0 i γx t + 2 1 ~ 3 1 ~ + γ v0ti + γg0t i 6 24 ~ ( xi x0) t = ti c γ is the vertical gravity gradient (~3 µgal/cm), c the speed of light x 0 the initial position v 0 the initial velocity g 0 the initial acceleration 4 x i, t i, i = 1,,700

FG5 Schematic Laser is frequency-stabilized He-Ne laser (red light @ 633 nm) Interferometer splits beam into test and reference beams The test beam bounces off falling corner cube then off stationary spring corner cube The reference beam travels straight through interferometer. Beams are recombined and interference signal (fringes) is used to track falling test mass The time intervals between the occurrence of each fringe are measured by a Rubidium oscillator

FG5 Subsystems Dropping Chamber Superspring Interferometer Laser Electronics Software Real-Time Data Acquisition Post-Processing Data Analysis

FG5 Dropping Chamber Drag Free Cart Mechanical Drive Vacuum system (Ion Pump 10-6 Torr) Test Object (ball&vee contacts) Corner Cube Lock Mechanism

Drag-free Dropping Chamber Dropping chamber (Vacuum ~ 10-4 Pa 10-9 atm. ) Drag-free chamber Ion pump (always connected) Mach-Zender interferometer Reduces drag due to residual gas molecules Follows the dropped corner cube, gently arrest and lift it Shields the corner cube from external electrostatic forces

Corner Cube Retroreflectors Reflected ray parallel to input ray No phase change in wavefronts Insensitive to translation and rotation Used in both Dropping test mass and Stationary mass

Balls & Vees Re-orient dropper corner cube after each drop Tungsten parts (wear out). Typical lifetime ~250,000 drops (maximum ~500,000 drops, and depends on dropper tuning) Balls Corner Cube Vees Cart

Ion Pump Vacuum is maintained by an ion pump No moving parts, high voltage used to ionize molecules and plate them out Before pumping process begins, neutral atoms and molecules float in ion pump area +4kV When process starts, electrons are attracted to the positive voltage, and ions to the ground. Liberated electrons ionize other molecules, forming a pumping action ground

Ion Pump (cont.) Electrical current drawn by ion pump power supply is proportional to # of molecules ionized Therefore electrical current is proportional to pressure Vacuum level must be very good (< 10-5 Torr) before ion pump can be operated Vacuum started with mechanical turbo pump Limited life time (~5 years)

Mach-Zender Interferometer Michelson s Interferometer Falling mass Mach-Zender s interferometer: 2 beam splitters Falling mass Las er B.S. Fringes Laser Test beam B.S. Photodiode B.S. Fringes To other detectors Fringes Allows complete freedom to make the two interfering beams collinear (parallel and overlapped) for optimal fringes contrast

FG5 Interferometer Mach-Zender type Insensitive to rotations and translations Three optical outputs Main signal interferometer (APD) Telescope (verticality and/or beam alignment) Viewing port Two Electronic Signals Analog (Alignment) TTL (Timing)

FG5 Interferometer Adjustments Input beam fiber adjustment (test beam verticality) Twiddler (beams coincident) Final test beam mirror (beams parallel) Alignment of beams onto photodetector

Beam Path photodetector twiddler parallelizing mirror

Beam Verticality photodetector twiddler parallelizing mirror Note that the twiddler and the parallelizing mirror do nothing!

FG5 Superspring 60s Period Two Stage nested spring system Sphere Detector Coil transducer Lock Mechanism Temperature compensation Spring height adjustment Bubble level adjustments Delta rods Zeroing the sphere position (S-shaped response)

The Superspring The superspring: long period isolation spring that provides the inertial reference frame Inertial reference corner cube

The Superspring Pivots Aneroid 1 st Stage Spring 2 nd Stage Spring Corner Cube Sphere Coil Magnets

Measurement Scatter Without the superspring... and with the superspring

FG5 Electronics Computer Data acquisition & Reprocessing Main Power Supply Superspring Controller Dropping Chamber Controller Ion pump power supply Laser Controller Patch Panel Analog & Digital IO Computer Laser Controller System Interface Module Magma PCI Unit Power Supply

WEO Iodine Stabilized Laser Primary Standard (BIPM Certified) Stabilized to rotational states ( hyperfine splitting) of iodine Accuracy at 1 part in 10 11 Automatic peak locking Fiber launching system Faraday Isolator (prevents feedback into laser) 5-axis stage Polarized fiber Output collimation (~6mm) Operating Temperature: 15 25 C

FG5 Setup* Check Ion Pump Voltage Turn on WEO laser Level Superspring Tripod Measure first reference height Lock Superspring in tripod, level SS bubbles using feet Attach interferometer to top of Superspring Place dropper tripod on top of interferometer Lock dropping chamber in dropper tripod Decouple dropper from interferometer Verticalize the dropper using feet Measure second reference height Adjust beam verticality using alcohol pool Center Superspring position Optimize fringe amplitude Fill in parameters to software *See FG5 User s Manual for details.

g Gravity Acquisition and Processing Software Windows Based Graphics package Gravity corrections Earth Tide Models Ocean Load Correction Statistical analysis Real time data acquisition Post processing

g Software control Site Specification Instrument Parameters Data Acquisition Parameters Gravity Corrections Graphics Reports

g Input Parameters Site Specification Latitude Longitude Elevation (std pressure) Gradient (-3.1 μgal/cm) Polar Motion Data Acquisition Parameters Number of drops/set Number of sets Interval between drops (normally 1s) Start time of data acquisition Projects (sets of sets)

Gravity Corrections & Error Sources Gravity Corrections Earth Tides Ocean Loading Barometer Polar motion Gradient Speed of Light Error Sources Verticality: 9 arcsec = 1μGal 1 spot = 4μGal Water Table: 2.5 cm = 1μGal T.M. Niebauer et al, Metrologia, 1995, 32, 159-180

Prescaling & g Fit Example Prescale*Multiplexor = 1000 #Fringes = 700 recorded fringe # actual # of fringes time (s) T distance (mm) X 1 1 0.00025 0.0003 2 1001 0.0078 0.300 3 2001 0.0111 0.600........ 700 700001 0.207 210.000

Residuals 100 100 80 ( ) t 2 + runif( 1, 4, 4) 0 + 100 t 2 + 100 ( ) t 2 + t 2 + runif( 4, 4, 1) 0 + 50 60 40 20 Measurements Best Fit Residuals 0 0 0 2 4 6 8 10 0 t Note: vertical scale exaggerated, normal residuals are approximately 1nm. 10

Simple Statistics: How much data should I take? First, some definitions: σ = drop scatter (standard deviation of measurements) δ stat = statistical uncertainty δ sys = systematic uncertainty ( built in system uncertainty and model uncertainties) δ total = sum, in quadrature, of statistical and systematic uncertainties δ stat =σ / N drops δ = δ + δ total 2 sys 2 stat Measure drop scatter, σ Pick your desired statistical uncertainty, δ stat This determines N drops Spread this N drops over a convenient number of sets (12 hours or 24 hours). Remember the balls & vees: only run as long as you need to!

Simple Statistics (cont) FG5: ~2μGal Systematic Uncertainty Example: Drop scatter = 15μGals 2μGals statistical uncertainty => ~100 drops For 100μGal scatter (noisy site!) => 2500 drops total Lifetime ~250,000 drops => 100 site occupations

FG5 Electronics Trigger to Dropper TIA TRIG TIA FRINGE TIA CLOCK A2D Analog In Computer Magma PCI A2D Digital TRIG OUT TTL Fringes From Interferometer Laser Controller Superspring SIM Magma PCI Unit 10MHz Power Supply

Regular Maintenance Regular maintenance of the system at Micro-g LaCoste is necessary Typically after about 250,000 drops (maximum ~500,000 drops) Dropper belt wear Optics Cleaning Ferrofluidic feedthrough replacement Ion pump degradation (plating) Ball & Vee wear (Micro-g) Laser tube degradation (Micro-g)

FG5 Results (1) Below are the results from a Comparison of Absolute Gravimeters in Luxembourg, 2004* 15 gravimeters, independent operators, 5 days Standard Deviation of FG5s: 2.3µGal * O. Francis, et al., Results of the Intercomparison of Absolute Gravimeters in Walferdange, Luxembourg of November 2003, International Association of Geodesy Symposia, Vol 129, 2004.

FG5 Results (2) Shown below are the results of absolute gravimeter measurements at Churchill, Canada*. The slow reduction in gravity over 12 years is due to postglacial rebound (uplift in the crust as the earth recovers from the weight of the ice in the last ice age). This type of long-term study is only possible with the inherent stability of an absolute gravimeter The blue squares are from JILA-g meter measurements, and the red squares are FG5 measurements. * A. Lambert et al., New constraints on Laurentide postglacial Rebound from Absolute Gravity measurements, Geophysical Research letters, Vol 28, No. 10, pp. 2109-2112, May 15, 2001.