Characterization of the new scintillator Cs 2 LiYCl 6 :Ce 3+ *

Similar documents
Neutron Spectrometry Using a 7 Li Enriched CLYC Scintillation Detector

Investigation of fast neutron spectroscopy capability of 7 Li and 6. Li enriched CLYC scintillator for nuclear physics experiments

national security, nonproliferation, defense, border security, and homeland security applications.

Lawrence Berkeley National Laboratory Lawrence Berkeley National Laboratory

Gamma Spectroscopy. References: Objectives:

Design of a Lanthanum Bromide Detector for TOF PET

The Scintillation properties of Pr 3+ doped and Pr 3+, Ce 3+

A new scintillator detector for nuclear physics experiments: the CLYC scintillator

DELAYED COINCIDENCE METHOD FOR PICOSECOND LIFETIME MEASUREMENTS

M.J. CIESLAK, K.A.A. GAMAGE Department of Engineering, Lancaster University Lancaster, LA1 4YW United Kingdom ABSTRACT

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

The Characterization of new Eu 2+ doped TlSr2I5 Scintillator Crystals

SCINTILLATION DETECTORS AND PM TUBES

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Inorganic Scintillators

The fast light of CsI(Na) crystals

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

arxiv:physics/ v1 3 Aug 2006

DETECTORS. I. Charged Particle Detectors

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Investigation of Ce-doped Gd2Si2O7 as a scintillator. Author(s) Kurashige, Kazuhisa; Ishibashi, Hiroyuki; Furusaka, Instructions for use

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Alpha particle scintillation detector based on micro pixel avalanche photodiode and LYSO crystal

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

A new timing model for calculating the intrinsic timing resolution of a scintillator detector

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

Scintillation Detectors

OPPORTUNITY TO JOIN IEEE AND NPSS

Chapter 4 Scintillation Detectors

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

STUDY ON THE ENERGY RESPONSE OF PLASTIC SCINTILLATION DETECTOR TO MEV NEUTRONS ABSTRACT

arxiv: v1 [physics.ins-det] 29 Jun 2011

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

The NaI:Tl and CsI:Tl crystals for effective detection of X-rays and low energy charged particles

New Scintillators for the Border Monitoring Equipment

Week 6: Ch. 8 Scintillation Counters

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

28th Seismic Research Review: Ground-Based Nuclear Explosion Monitoring Technologies DESIGN OF A PHOSWICH WELL DETECTOR FOR RADIOXENON MONITORING

DIGITAL PULSE SHAPE ANALYSIS WITH PHOSWICH DETECTORS TO SIMPLIFY COINCIDENCE MEASUREMENTS OF RADIOACTIVE XENON

Europium-doped barium halide scintillators for x-ray and -ray detections

Characterization of the Lanthanum Chloride Scintillation Detector

arxiv: v1 [physics.ins-det] 3 Feb 2011

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

1 of :32

Characterization and Monte Carlo simulations for a CLYC detector

Enhanced α-γ Discrimination in Co-doped LaBr 3 :Ce

A Study on Radiation Damage in PWO-II Crystals

Scintillation Detector

Dual Isotope Imaging with LaBr3:Ce Crystal and H8500 PSPMT

Detector technology. Aim of this talk. Principle of a radiation detector. Interactions of gamma photons (gas) Gas-filled detectors: examples

Radionuclide Imaging MII Detection of Nuclear Emission

The Mössbauer Effect

PHYS 3446 Lecture #12

High-efficiency and fast-timing scintillators

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Scintillators General Characteristics

Pulse height non-linearity in LaBr 3 :Ce crystal for gamma ray spectrometry and imaging

BaF 2 :Ce POLYMER COMPOSITE GAMMA-RAY SCINTILLATORS M.B. Barta, J.H. Nadler, Z. Kang, B.K. Wagner

A method to measure transit time spread of photomultiplier tubes with Cherenkov light

arxiv:hep-ex/ May 2000

BSO Crystals for the HHCAL Detector Concept

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

Radioactivity and Ionizing Radiation

The Compton Effect. Martha Buckley MIT Department of Physics, Cambridge, MA (Dated: November 26, 2002)

Figure 1. Decay Scheme for 60Co

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

Survey of the performance of scintillation materials at low temperatures

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Measurement of high energy gamma rays with large volume LaBr 3 :Ce scintillators

Material optimisation in dual particle detectors by comparing advanced scintillating materials using two Monte Carlo codes

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET

A POSITION SENSITIVE ALPHA PARTICLE DETECTOR BASED ON A LYSO CRYSTAL AND A MICRO-PIXEL AVALANCHE PHOTODIODE

Detection and measurement of gamma-radiation by gammaspectroscopy

Chapter Seven (Nuclear Detectors)

Improved scintillation proportionality and energy resolution of LaBr 3 :Ce at 80K

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations

Radioactivity. Lecture 6 Detectors and Instrumentation

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

ORTEC AN34 Experiment 10 Compton Scattering

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

Measurements of CsI(Tl) Crystals with PMT and APD. ipno.in2p3.fr Jean Peyré Milano - October 2006

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

Unit 2. Instrumentation. Experts Teaching from Practical Experience

Measurements of liquid xenon s response to low-energy particle interactions

High quantum efficiency S-20 photocathodes for photon counting applications

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Nuclear Instruments and Methods in Physics Research A

arxiv: v2 [physics.ins-det] 8 Feb 2013

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09

Chinese Physics C. Study of n-γ discrimination in low energy range (above 40. kevee) by charge comparison method with a BC501A liquid

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 55, NO. 4, AUGUST /$ IEEE

Timing and Energy Response of Six Prototype Scintillators

ISPA-Tubes with YAP:Ce Active Windows for X and Gamma Ray Imaging.

Special Nuclear Material Detection Studies With The SMANDRA Mobile System

EEE4106Z Radiation Interactions & Detection

Transcription:

Characterization of the new scintillator Cs 2 LiYCl 6 :Ce 3+ * Li Kui-Nian ( 李奎念 ) 1;1) Zhang Xian-Peng ( 张显鹏 ) 1, 3) Gui Qiang ( 桂强 ) 2) Jin Peng ( 金鹏 ) 1 ) Tian Geng ( 田耕 ) 1) 1 Northwest Institute of Nuclear Technology, Xi an 710024, China 2 Beijing Glass Research Institute, Beijing 101111, China 3 School of Nuclear Science and Technology, Xi an Jiaotong University, Xi an 710049, China Abstract: The first domestic inorganic scintillator Cs2LiYCl6:Ce 3+ (CLYC)was grown in Beijing Glass Research Institute using the vertical Bridgman method. In this work, we evaluated the performance of this new CLYC crystal in terms of gamma-ray energy resolution and pulse shape discrimination (PSD) capability between neutrons and gamma-rays. The decay times occurred through different scintillation mechanisms were achieved by fitting decay functions to the neutron and gamma-ray waveform structures. We found an energy resolution of ~4.5% for 662-keV gamma-rays and efficient neutron/gamma PSD with FoM ~2.6. Under gamma-ray excitation, there is ultrafast scintillation mechanism in CLYC, with a decay time about 2 ns, whereas there is no evidence of ultrafast decay under thermal neutron excitation. This work contributes to promote domestic development of CLYC. Keywords: Cs 2LiYCl 6:Ce 3+, energy resolution, pulse shape discrimination, decay time, thermal neutrons PACS:29.40.-n, 29.40.Mc, 81.10.Fq * Project supported by the National Natural Science Foundation of China (Grant No. 11575145) 1) Corresponding author. E-mail: lk.nian@foxmail.com

1 Introduction Cs2LiYCl6:Ce 3+ (CLYC[1] is a promising new inorganic scintillator. It was first discovered as a scintillator in 1999 by researchers from Delft University of Technology[2]. They reported that CLYC had the capability for thermal neutron detection based on the 6 Li(n,α) reaction and made further investigations on CLYC in next few years[3-5]. Van Loef revealed the scintillation mechanism of CLYC, known as Core-to-Valence-Luminescence (CVL), in 2005[6]. The CVL has an ultrafast decay time (~ns) and appears only under gamma-ray radiation[7]. CVL enabled CLYC to discriminate between neutron (alpha) particles and gamma-rays with pulse shape discrimination (PSD) methods[8]. The capability for PSD of CLYC drew great attentions from the nuclear science community[9-14]. Under gamma-rays irradiation, its light yield is about 20000 photons/mev[15], and it exhibits good energy resolution of ~4% FWHM @662keV due to excellent linearity[16]. Under thermal neutrons irradiation, the light yield is about 70500 photons per neutron. The thermal neutron peak resolution is typically 2.5~3.1%[9]. Radiation Monitoring Devices Inc. grew its first CLYC crystal in 2003. Since then, a number of CLYC crystals with different sizes and features have been grown with the development of crystal growth technology[17-19]. In 2012, CLYC became available through commercial approaches, but the price was still high. To establish the domestic growth technology of CLYC, Beijing Glass Research Institute (BGRI) grew a promising CLYC crystal in 2016. To our knowledge, it is the very first domestic CLYC crystal. To evaluate the performance of the new CLYC crystal, we focused on two important properties of it: gamma-ray energy resolution and PSD capability. Moreover, we achieved the decay times of different scintillation mechanism by fitting decay functions to the neutron and gamma-ray waveform structures. 2 Experimental 2.1 Crystal growth CLYC crystal was grown using the vertical Bridgman method in BGRI. In this

method, stoichiometric amounts of CsCl, LiCl, YCl3 and CeCl3 were mixed in a quartz crucible. Natural LiCl that contains about 7% of the 6 Li isotope was used. The dopant (Ce 3+ ) concentration was 0.5mol% as a standard doping level. After mixing powders in the quartz crucible, vacuum was pumped to 10-5 Pa to avoid oxidization. The quartz crucible was sealed, before fusing the mixed powders in a stove at a temperature above 640 C. Along with lowering the quartz crucible steadily with a velocity of 2 millimeter per hour, the CLYC crystal ingot was grown. After cutting and polishing, the CLYC crystal was packed in an aluminum casing with a reflective inner coating and sealed with a quartz window at one end. CLYC has density of 3.31g/cm 3 and its peak-emission wavelength is about 390nm, which is well matched to many commercially photomultiplier tubes. Fig.1 shows the picture of the CLYC crystal with 25.4 mm diameter and 15 mm thickness. Fig.1. Photograph of CLYC crystal with 25.4mm diameter and 15mm thickness grown in BGRI. The crystal is packaged in an aluminum enclosure. 2.2 Data acquisition The CLYC crystal was coupled with a 2 inch diameter 9815B blue-green sensitive bialkali photomultiplier (PMT) with optical grease. 9815B PMT s quantum efficiency is ~30% at peak-emission wavelength (390nm) of CLYC. To estimate gamma-ray energy resolution of CLYC crystal, a 137 Cs source was used as gamma-ray source. The signals from PMT anode with different heights were analyzed in a multi-channel spectroscopy by using a shaping time of 4 microseconds. An Am-Be source which emits approximately 1 10 4 neutrons per second was used as neutron and gamma-ray source to estimate the PSD capability of CLYC. The

schematic diagram of the experimental setup is shown in Fig.2. A Tektronix DPO7104 oscilloscope was set to FastFrame mode (1 G samples per second and 2 us recording length) to record the PMT anode signals from both neutrons and gamma-rays with a 50 Ohm terminator. The CLYC crystal coupled with 9815B PMT was placed in a polyethylene tube. A polyethylene block with 2.5cm thickness and a lead block with 2.5cm thickness were placed between the CLYC crystal and Am-Be source. Approximate 5000 signals were recorded, and analyzed off-line with PSD method (see section 3.2). 2.5cm Lead block 2.5cm Polyethylene block CLYC crystal Polyethylene tube High voltage supply Am-Be Source PMT Anode Signal CH1 DPO7104 Fig.2. Experimental setup in Am-Be source. The CLYC crystal was coupled with a 9815B PMT which connected directly to a digital oscilloscope (Tektronix DPO7104) 3 Results and analysis 3.1 Gamma-ray energy resolution Fig. 3 shows the energy spectra measured with CLYC crystal under irradiation from 137 Cs gamma-ray (662keV) source. The position of the 662keV energy peak was used to scale the horizontal axis in energy units (kev). The CLYC energy resolution at the 662keV full energy peak is ~4.5% (full width at half maximum, FWHM), which is better than those commonly used scintillators in gamma detection, such as NaI(Tl) and CsI(Tl), having 6.5% and 7-9% energy resolution @662keV respectively.

500 662 kev 400 Counts 300 200 E~4.5% 100 0 0 200 400 600 800 Energy /kev Fig.3. 137 Cs energy spectra measured with CLYC crystal. An energy resolution of ~4.5% FWHM @662keV was acheived. 3.2 Neutron and gamma-ray discrimination PSD was applied to separate gamma-ray from neutron waveforms based on the traditional charge integration method. In this method, two integration windows of the signal pulses, named as prompt integration window and delays integration window individually, were selected and integrated. The prompt window includes the sharp rise and peak portion, and the delayed window includes the decaying portion of the signals. Generally, there are infinite combinations of possible integration window widths. It is essential for achieving the best possible PSD by optimizing window widths. The automated waveform analysis algorithm can optimize window width through the following procedure: (1) Aligning the waveforms to a common zero. The point where the amplitude was 20% of the peak value was chosen as the common zero. (2) Charge integration according to two integration windows. The width of the prompt integration window ranged from 5 ns to 300 ns by 5 ns. The delayed integration window started right after the end point of the prompt window and its width ranged from 10 ns to 1300 ns by 10 ns. (3) Calculating the PSD ratio. For each unique combination of the prompt and delayed integration windows, PSD ratio was calculated for each waveform based on charge integrals Qprompt and Qdelayed. The PSD ratio was defined as Qdelayed /Qprompt.

(4) Calculating the Figure-of-Merit (FoM). FoM was utilized to evaluate the performance of the neutron and gamma-ray discrimination. FoM was defined as FoM = D n FWHM +γ FWHM (1) Where D is the distance between the centroids of the neutron and the gamma peak in the PSD ratio histogram, and n FWHM and γ FWHM are the corresponding FWHMs. To obtain the centroid positions and FWHMs, Gaussian fits were performed to the neutron and gamma-ray peaks in the histogram. The result of the optimization for PSD is shown in Fig.4. This figure shows that FoM is more sensitive to the prompt window width compared with the delayed window width. With increasing prompt window width, FoM shows a fast and short increase, followed by a sharp, long decrease. Change of delayed window width has a minimal effect on the FoM. For the CLYC crystal in this study, the best FoM we found was 2.6, in a prompt window width of 20 ns and delayed window width of 900ns. 3.0 2.610 2.5 2.285 2.0 1.473 1.5 FOM 0.4975 1.0 0 50 prompt width /ns 100 150 200 250 300 1400 1200 1000 800 600 400 200 0 0.5 Delayed width /ns Fig.4. (color online) FoM values under different prompt and delayed integration windows combinations. The prompt width ranges from 5 ns to 300 ns by 5 ns. The delayed width range from 10 ns to 1300 ns by 10 ns.

With the optimal prompt and delayed integration windows, a two-dimensional PSD scatter plot was generated for the neutron and gamma-ray events from an Am-Be source (see Fig.5). In Fig.5, gamma-ray events and neutron events show us two disparate distribution patterns. Gamma-ray events have lower PSD ratio across the wide energy range while neutron events are isolated with higher PSD ratio. The neutron events can be divided into three groups based on different reaction: fast neutrons ( 35 Cl(n,p) reaction), thermal neutrons ( 6 Li(n,α) reaction) and fast neutrons ( 6 Li(n,α) reaction), as indicated in Fig.5. Fig.6 (dash line) shows the one-dimensional projection of neutron and gamma-ray events corresponding to the Y-axis in Fig.5. A Gaussian function was fitted to the neutron and gamma-ray peaks in Fig.6 (solid line). The centroids and FWHMs obtained from the Gaussian fits were used to calculate FoM value. 60 50 Fast 35 Cl(n,p) Fast 6 Li(n,α) Q delayed /Q prompt 40 30 20 Thermal 6 Li(n,α) Neutron 10 γ 0 0 20 40 60 80 100 120 140 Q total Fig.5. Two-dimensional PSD scatter plot based on the data collected under irradiation from an Am-Be source. X-axis is the PSD ratio (Q delayed /Q prompt) and Y-axis is the total integral Q total corresponding to the energy of events. Gamma-ray events (on the bottom) and neutron events (on the top) are separated clearly. Fast neutrons ( 35 Cl), thermal neutrons ( 6 Li) and fast neutrons ( 6 Li) are indicated.

40 30 counts 20 10 γ-ray neutron 0 10 20 30 40 50 60 Q delayed /Q prompt Fig.6. One-dimensional projection of neutron and gamma-ray events to the Y-axis in Fig.5. A Gaussian function was fitted to the neutron and gamma-ray peaks (solid line). 3.3 Pulse shape analyses According to the PSD optimal results, neutron signals and gamma-ray signals were averaged respectively to produce the standard-pulses. These standard-pulses contribute to eliminate noise fluctuations in individual signals and reveal the details of the prompt and decayed pulse shape. Neutron standard-pulse and gamma-ray standard-pulse were inverted and normalized to their peaks, as shown in Fig. 7. Significantly different characteristics can be found between neutron pulse and gamma-ray pulse due to different scintillation mechanisms. Amplitude /A.U 1.0 0.8 0.6 0.4 0.2 gamma neutron gamma fit neutron fit 0.0 0 300 600 900 1200 1500 1800 Time /ns Fig. 7. (color online) Overlay of neutron and gamma-ray standard-pulse. Exponential decay

curves fitting of the standard pulses are shown. The red solid line is the gamma-ray standard-pulse exponential fitting curve and the green solid line is the neutron standard-pulse exponential fitting curve. Generally, for Ce3+-doped CLYC crystal, up to four mechanisms contribute to the scintillation[10,16]: Ce 3+ (direct electron-hole capture by Ce 3+ ), Vk (binary Vk-electron diffusion), STE (self-trapped exciton emission), and CVL (core-to-valence luminescence). The first scintillation mechanism Ce 3+ may occur when an electron that is created in the valence band is excited into the conduction band of the crystal by the incident gamma-ray. The formed free electron-hole pair may be captured directly by a Ce 3+ ion leading to 4f-5d excitation and 5d-4f emission which produce a prompt emission with very fast rise time. If the hole gets trapped by two anions, a molecular complex named as Vk center will be formed. From this point, two scintillation mechanisms will proceed. The Vk center can migrate from one site to another in the crystal lattice through thermal activation until it is captured by a Ce 3+ ion to form Ce 4+ or a Ce 3+ -Vk associated complex. The free electron is captured subsequently, leading to luminescence. This process is referred to as binary Vk-electron diffusion. Before trapped by Ce 3+, the Vk center may capture an electron from the conduction band to form a STE which itself is luminescent with a decay time on the order of microseconds. Finally, CVL may occur when the incident high-energy photon excites an electron from the upper core band (3pCl ) into the conduction band (5pCl ). To verify the scintillation mechanisms under neutron and gamma-ray excitation, we used the exponential decay curves to fit the standard pulses (see Fig.7). Each decay time constant of exponential component represents the decay time of different scintillation mechanism. The fitting result of gamma waveform reveals four exponential decay components. Their decay times are 2 ns, 50 ns, 420 ns, and 3400 ns, corresponding to CVL, Ce 3+, Vk, and STE respectively. For neutron waveforms, only Vk, and STE are observed: 390 ns and 1500 ns (see Table I).

Table1. Measured decay times for gamma and neutron induced CLYC emission Types CVL /ns Ce 3+ /ns Vk /ns STE /ns Gamma-ray 2 50 420 3400 Neutron - - 390 1500 4 Conclusion CLYC crystal has excellent properties, and can be a replacement of 3 He material in homeland security and nuclear safeguards. The capability of robust discrimination between neutrons and gamma-rays makes CLYC a radiation detector with dual neutron and gamma-ray sensitivity[20-21]. Also, based on 6 Li(n,α) reaction and 35 Cl(n,p) reaction, CLYC can discriminate between thermal neutrons and fast neutrons through pulse height discrimination. In this study, we evaluated the gamma-ray energy resolution and PSD capability of CLYC crystal grown in BGRI. We also achieved the decay times of different scintillation mechanism by fitting decay functions to the neutron and gamma-ray waveform structures. The CLYC crystal has a gamma-ray energy resolution of ~4.5% FWHM @662keV, and has good PSD capability with FoM ~2.6. There is ultrafast scintillation decay (about 2 ns) in CLYC under gamma-ray excitation whereas there is no evidence of ultrafast decay under thermal neutron excitation. The CLYC crystal can be used in neutron-gamma dual mode radiation detector, and can be applied to detect radiation in space. In future work, we plan to grow CLYC crystal doped with enriched 6 Li, larger size and better properties by optimizing crystal growth technology. References [1] Glodo J, Hawrami R, Shah K S. J. Cryst. Growth, 2013, 379: 73 [2] Combes C M, Dorenbos P, VanEijk C W E et al. J. Lumin., 1999, 82: 299 [3] VanLoef E V D, Dorenbos P, VanEijk C W E et al. J. Phys. Condens. Mat., 2002, 14:8481

[4] Bessiere A, Dorenbos P, VanEijk C W E et al. IEEE Trans. Nucl. Sci. 2004, 51: 2970 [5] Bessiere A, Dorenbos P, VanEijk C W E et al. Nucl. Instrum. Methods Phys. Res. Sect. A, 2005, 537:242 [6] VanLoef E V D, Glodo J, Higgins W M et al. IEEE Trans. Nucl. Sci., 2005, 52: 1819 [7] Rodnyi P A. Radiat. Measur. 2004, 38: 343 [8] LI Kui-Nian, LI Bin-Kang, ZHANG Mei et al. Acta Phys. Sin, 2014, 63: 202901 (in Chinese) [9] Olympia N D, Chowdhury P, Guess C J et al. Nucl. Instrum. Methods Phys. Res. Sect. A, 2012, 694: 140 [10] Budden B S, Stonehill L C, Terry J R et al. IEEE Trans. Nucl. Sci., 2013, 60: 946 [11] Machrafi R, Khan N, Miller A. Radiat. Measur., 2014, 70: 5 [12] Bourne M M, Mussi C, Miller E C et al. Nucl. Instrum. Methods Phys. Res. Sect. A, 2014, 736:124 [13] Yang K, Menge P R. Nucl. Instrum. Methods Phys. Res. Sect. A, 2015, 784: 74 [14] Giaz A, Blasi N, Boiano C et al. Nucl. Instrum. Methods Phys. Res. Sect. A, 2016, 825: 51 [15] Lee D W, Stonehill L C, Klimenko A et al. Nucl. Instrum. Methods Phys. Res. Sect. A, 2012, 664: 1 [16] 19Olympia N D, Chowdhury P, Lister C J et al. Nucl. Instrum. Methods Phys. Res. Sect. A, 2013, 714: 121 [17] Glodo J, Higgins W M, VanLoef E V D et al. IEEE Trans. Nucl. Sci., 2008, 55: 1206 [18] Glodo J, Higgins W M, VanLoef E V D et al. IEEE Trans. Nucl. Sci., 2009, 56: 1257 [19] Higgins W M, Glodo J, Shirwadkar U et al. J. Cryst. Growth, 2010, 312: 1216 [20] Giaz A, Pellegri L, Camera F et al. Nucl. Instrum. Methods Phys. Res. Sect. A, 2016, 810:

132 [21] Whitney C M, Soundara-Pandian L, Johnson E B et al. Nucl. Instrum. Methods Phys. Res. Sect. A, 2015, 784: 346