Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

Similar documents
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 56, NO. 3, JUNE

Light yield non-proportionality and energy resolution of Lu 1.95 Y 0.05 SiO 5 :Ce and Lu 2 SiO 5 :Ce scintillation crystals

Measurements of CsI(Tl) Crystals with PMT and APD. ipno.in2p3.fr Jean Peyré Milano - October 2006

New Scintillators for the Border Monitoring Equipment

Light yield non-proportionality and intrinsic energy resolution of doped CsI scintillators

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

Scintillation Detector

Design of a Lanthanum Bromide Detector for TOF PET

Non-proportionality of organic scintillators and BGO

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Large Area Avalanche Photodiodes in X-rays and scintillation detection

Pulse height non-linearity in LaBr 3 :Ce crystal for gamma ray spectrometry and imaging

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Improved scintillation proportionality and energy resolution of LaBr 3 :Ce at 80K

arxiv:physics/ v1 3 Aug 2006

Non-proportional scintillation response of NaI:Tl to low energy X-ray photons and electrons

THE COMPTON EFFECT Last Revised: January 5, 2007

OPPORTUNITY TO JOIN IEEE AND NPSS

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Key words: avalanche photodiode, soft X-ray detector, scintillation γ-ray detector, imaging device PACS: 07.85;95.55.A;85.60.D

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data

Detection and measurement of gamma-radiation by gammaspectroscopy

Characterization and Monte Carlo simulations for a CLYC detector

Scintillation Properties of Lu 3 Al 5 O 12, Lu 2 SiO 5 and LaBr 3 Crystals Activated with Cerium

Lawrence Berkeley National Laboratory Lawrence Berkeley National Laboratory

Gamma Spectroscopy. References: Objectives:

Development of a High Precision Axial 3-D PET for Brain Imaging

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Sample Spectroscopy System Hardware

Positron Emission Tomography

A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector

Scintillation Detectors

Application of Nuclear Physics

A method to measure transit time spread of photomultiplier tubes with Cherenkov light

arxiv: v1 [physics.ins-det] 16 May 2017

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Characterization of the Lanthanum Chloride Scintillation Detector

A new timing model for calculating the intrinsic timing resolution of a scintillator detector

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Ultra High Quantum Efficiency PMT for energy resolution measurements of LaBr 3 (Ce) scintillation crystals

XENON Dark Matter Search. Juliette Alimena Columbia University REU August 2 nd 2007

GAMMA RAY SPECTROSCOPY

arxiv: v1 [physics.ins-det] 3 Feb 2011

arxiv: v1 [physics.ins-det] 29 Jun 2011

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

arxiv: v1 [physics.ins-det] 11 Mar 2010

Analysis of γ spectrum

Compton Camera. Compton Camera

EEE4106Z Radiation Interactions & Detection

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

arxiv: v2 [physics.ed-ph] 23 Jan 2018

Nuclear Instruments and Methods in Physics Research A

Journal of Asian Scientific Research, 1(3),PP

Non-proportional response and energy resolution of pure SrI 2 and SrI 2 :5%Eu scintillators

PHYS 3313 Section 001 Lecture #7

A liquid xenon PET camera - Simulation and position sensitive PMT tests

Readout of LYSO using a new silicon photodetector for positron emission tomography

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Radiation Detection and Measurement

Gamma Ray Spectroscopy

Charge collection in PET detectors

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

Experiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado

arxiv: v1 [physics.ins-det] 29 Jun 2011

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

Oslo Cyclotron Laboratory

AN34 Experiments in Nuclear Science Laboratory Manual Fourth Edition

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

Geant4 simulation of plastic scintillators for a prototype μsr spectrometer

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

Timing and Energy Response of Six Prototype Scintillators

a monte carlo simulation of photomultiplier resolution

SCINTILLATION DETECTORS AND PM TUBES

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations

Gamma ray coincidence and angular correlation

DETECTORS. I. Charged Particle Detectors

Hands on LUNA: Detector Simulations with Geant4

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

IN recent years, the Cherenkov effect for electrons at

A Liquid Argon Scintillation Veto for the GERDA Experiment

A novel design of the MeV gamma-ray imaging detector with Micro-TPC

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

ISPA-Tubes with YAP:Ce Active Windows for X and Gamma Ray Imaging.

A novel method to calibrate DOI function of a PET detector with a dual-ended-scintillator readout

Status report on parity violation in the (1232) resonance

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

arxiv:physics/ v1 [physics.ins-det] 14 Feb 2005

Particles and Waves Particles Waves

Lucite Hodoscope for SANE

PoS(TIPP2014)093. Performance study of the TOP counter with the 2 GeV/c positron beam at LEPS. K. Matsuoka, For the Belle II PID group

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

Measurements of Liquid Scintillator Light Yield for Future Neutrino Experiments

Transcription:

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data Harold Rothfuss a,b, Larry Byars c, Michael E. Casey a, Maurizio Conti a, Lars Eriksson a, Charles L. Melcher b, Christian Michel a, Merry Spurrier b, Piotr Szupryczynski a,b a Siemens Medical Solutions, Molecular Imaging, 810 Innovation Drive, Knoxville, TN 37932, USA b Scintillation Materials Research Center, University of Tennessee, Knoxville, TN c Byars Consulting Oak Ridge, TN 37831 USA

Basics of Radiation Spectroscopy with scintillators. Ideal Energy Spectrum e - Photoelectric effect and Compton scattering seen in spectrum s structure as energy deposited in the scintillator. Compton Scatter region, continuous due to energy dependence on theta γ γ # hv hv' = hv 1+ (1 cosθ ) 2 m0c Peak also includes scatter + absorption of all initial energy Photoelectric absorption of incident gamma E γ e -

Basics of Radiation Spectroscopy with scintillators. Departure from Ideal Energy Spectrum Interaction effects from ideal energy spectrum blur due to physical effects such as transport efficiency, non proportional effects, statistical effects and electrical effects. Measure of blurring of the full energy peak is done with the energy resolution of the of the scintillator. Counts ch 0 E s = Re FWHM FWHM ch 0 Channel

Introduction Monte Carlo simulations of Positron Emission Tomographs (PET) often overestimate the experimental sensitivity, and blurring functions are also needed to properly represent spatial and energy resolution. In order to understand and overcome these issues, we introduced in the Monte Carlo simulation light yield nonproportionality and transport of optical photons in scintillating material, in addition to the conventional electromagnetic interaction. The simulation results were compared with fundamental experiments on single LSO crystals and blocks. The simulation code used was Geant4.

Crystals and geometry Crystal size : LSO 1cc cube, saw cut. Reflectant : Teflon sheet over crystal and tube Light guide : no, crystal directly on PMT with optical grease PMT : XP2020Q (Photonics) Source : 137 Cs ( 0.404 MBq, 662 kev γ) at 20 cm from the detector

Polished and saw cut LSO crystals

Experimental set-up

Electronics set-up High Voltage : Ortec 556 with -2000V Pre-Amplifier : Ortec Model 113, input capacitance 20pF Amplifier : Ortec 855 Dual Spec Amp with x25 gain MCA : Ortec Maestro 32MCB25 Acquisition time : 24 hours (and 24h background subtraction)

Experimental set-up 137 Cs LSO PMT Pre-Amp Amp Teflon sheet PC MCA reflector LSO PMT Glass window Simulation detail

Simulation performed with GEANT4 Geant4 Low Energy Library is used Source was a 137 Cs point source in vacuum with 4*π emission of a 662 kev Gamma Energy deposit in LSO is scored Non proportionality is applied to modify the number of light photons emitted (below 1 MeV E dep ) Scintillation photons are generated and transported in the crystal

Non Proportionality At low energy (below 1 MeV) the number of scintillation photons is not proportional to the deposited energy We developed a model from experimental data [4] to decrease the effective deposited energy in LSO (eq.1) Effect is applied after the electromagnetic interaction and prior to the scintillation light transport Eq.1 E eff = E dep (1 e ( 2.13E ( MeV ) 0.625) dep )

Optical photons transport We assumed 30000 scintillation photons per MeV and 25% PMT quantum efficiency The simulation components were the LSO crystal, a glass window of the PMT, the detecting photocathode of the PMT and some reflector. Surfaces of the crystal were modeled as a ground surface, index of refractions were N LSO = 1.8 N Window = 1.45 When scintillation photons cross through the crystal surface, has probability of 0.99 to be reflected on reflector. Most scintillation photons undergo internal reflection due to mismatch of index of refractions. N air = 1.0

Deposited energy spectra Experimental energy spectrum (solid line) from 137 Cs source with background subtracted, 24h acquisition, simulated spectrum (Δ). Left plot has simulation with simulated EM processes only. Right plot has simulation that includes Non-proportional effect and light transport.

Energy Resolution/Sensitivity E Resolution sim(w/lt & NP) = 10.48 +- 0.04% E Resolution exp = 10.2 +- 0.4% Energy Resolution was obtained with a Gaussian fit of the curves between 540 kev and 775 kev, and defined as the FWHM of the Gaussian function. Separate simulations allowed to separate the contribution to the energy resolution of the non proportionality (5.2%) and optical transport (9.3%), assuming that: E Resolution2 =E nonprop2 +E trans 2

Sensitivity Sensitivity Sim = (4.36 ± 0.13) 10-5 Sensitivity Exp = (4.39 ± 0.05) 10-5 Sensitivity was computed adding detected events with energy from 540 kev to 775 kev divided by the total number of photons emitted from the source. Experimental emission of source was calculated with Emission rate = activity(ci)* 3.7*10 10 (counts/sec)* acquisition time(sec)* 0.851

Timing Resolution Timing was simulated by tracking the path length and making the relationship of path length to transit time. The Timing resolution is the FWHM of the histogram of the difference of the start and stop signals. Simulated timing resolution only includes transit time of scintillation photons. Timing Resolution Exp (1cc) = 254 ± 12 psec Timing Resolution Sim (1cc) = 52 ± 7 psec Timing Resolution Exp (2cc) = 296 ± 7 psec Timing Resolution Sim (2cc) = 108 ± 4 psec

Conclusions Introducing light yield non proportionality and optical transport in Monte Carlo simulation we could estimate energy resolution and absolute sensitivity that well agree with experimental data. Information on the non proportionality of scintillating materials is available in literature and must be introduced in any simulation that aims to describe accurately detection of photons below 1MeV. Proper modeling of optical transport properties of the scintillating crystals is a key element for the correctness of the simulation, and requires a considerable work of fine tuning and experimental testing. This work will continue with simulation of crystals and blocks used in commercial PET scanners.

References [1] C.L Melcher and J.S Schweitzer, IEEE Trans. Nucl. Sci. NS-39 (1992) 503. [2] P. Dorenbos, J.T.M de Haas, C.W.E van Eijk, IEEE Trans. Nucl. Sci. NS-42 (1995) 2190. [3] M. Balcerzyk, M. Moszynski, M. Kapusta, D. Wolski, J. Pawelke, C.L. Melcher, IEEE Trans. Nucl. Sci., NS-47 (2000) 1319. [4] M. Kapusta, P. Szupryczynski, C. L. Melcher, M. Moszynski, M. Balcerzyk, W. Czarnacki, A. A. Carey, M. A. Spurrier, IEEE Trans. Nucl. Sci., NS-52 (2005) 1098. [5] S. Agostinelli et al., Geant4 - A Simulation Toolkit, Nucl. Instr. and Meth. A 506 (2003) 250. [6] J. Allison et al., Geant4 Developments and Applications, IEEE Trans. Nucl. Sci. 53 (1) (2006) 270. [7] GEANT4 website http://geant4.web.cern.ch/geant4/