On Signal to Noise Ratio Tradeoffs in fmri

Similar documents
7.3.A. The expression for signal recovery is similar to that derived under exercise 7.2 and is given by:

FREQUENCY SELECTIVE EXCITATION

Contrast Mechanisms in MRI. Michael Jay Schillaci

Introduction to MRI Acquisition

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution.

MRI Physics II: Gradients, Imaging. Douglas C. Noll, Ph.D. Dept. of Biomedical Engineering University of Michigan, Ann Arbor

Field trip: Tuesday, Feb 5th

EE225E/BIOE265 Spring 2013 Principles of MRI. Assignment 9 Solutions. Due April 29th, 2013

NMR and MRI : an introduction

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

Introduction to Biomedical Imaging

Background II. Signal-to-Noise Ratio (SNR) Pulse Sequences Sampling and Trajectories Parallel Imaging. B.Hargreaves - RAD 229.

Part III: Sequences and Contrast

Advanced Topics and Diffusion MRI

Basic MRI physics and Functional MRI

Physics of MR Image Acquisition

Outlines: (June 11, 1996) Instructor:

BMB 601 MRI. Ari Borthakur, PhD. Assistant Professor, Department of Radiology Associate Director, Center for Magnetic Resonance & Optical Imaging

Introduction to the Physics of NMR, MRI, BOLD fmri

Sequence Overview. Gradient Echo Spin Echo Magnetization Preparation Sampling and Trajectories Parallel Imaging. B.Hargreaves - RAD 229

} B 1 } Coil } Gradients } FFT

MRI in Review: Simple Steps to Cutting Edge Part I

Introduction to Magnetic Resonance Imaging (MRI) Pietro Gori

Lab 2: Magnetic Resonance Imaging

Tissue Parametric Mapping:

RAD229: Final Exam 2014/ SOLUTIONS You will have 3 hours to complete this Exam

Homework #2 Due date: 2/19/2013 (Tuesday) Translate the slice Two Main Paths in Lecture Part II: Neurophysiology and BOLD to words.

Apodization. Gibbs Artifact. Bioengineering 280A Principles of Biomedical Imaging. Fall Quarter 2013 MRI Lecture 5. rect(k x )

Midterm Review. EE369B Concepts Simulations with Bloch Matrices, EPG Gradient-Echo Methods. B.Hargreaves - RAD 229

Spatial encoding in Magnetic Resonance Imaging. Jean-Marie BONNY

K-space. Spin-Warp Pulse Sequence. At each point in time, the received signal is the Fourier transform of the object s(t) = M( k x

Spatial encoding in Magnetic Resonance Imaging. Jean-Marie BONNY

Course Review. Midterm Review: EE369B Concepts Simulations with Bloch Matrices, EPG SNR. B.Hargreaves - RAD 229. Section F1

MRI Physics I: Spins, Excitation, Relaxation

Bioengineering 278" Magnetic Resonance Imaging" " Winter 2011" Lecture 9! Time of Flight MRA!

Introductory MRI Physics

Exam 8N080 - Introduction to MRI

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Chapter 24 MRA and Flow quantification. Yongquan Ye, Ph.D. Assist. Prof. Radiology, SOM Wayne State University

EE591 Magnetic Resonance Imaging and Reconstruction Project Report. S i m u l a t i o n. Professor: Krishna Nayak ID: , Name: Yongjin Cho

Index. p, lip, 78 8 function, 107 v, 7-8 w, 7-8 i,7-8 sine, 43 Bo,94-96

How is it different from conventional MRI? What is MR Spectroscopy? How is it different from conventional MRI? MR Active Nuclei

Basic Pulse Sequences I Saturation & Inversion Recovery UCLA. Radiology

Master of Science Thesis. Development of a phantom for optimisation and quality control in functional MRI (fmri) Anders Nilsson

MR Advance Techniques. Flow Phenomena. Class I

BNG/ECE 487 FINAL (W16)

Spin Echo Imaging Sequence

Cambridge University Press MRI from A to Z: A Definitive Guide for Medical Professionals Gary Liney Excerpt More information

Optimized Gradient Waveforms for Spiral Scanning

Suppression of Static Magnetic Field in Diffusion Measurements of Heterogeneous Materials

Pulse Sequences: RARE and Simulations

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics

Fundamentals of MR Imaging

NMR/MRI examination (8N080 / 3F240)

Rad Tech 4912 MRI Registry Review. Outline of the Registry Exam: Certification Fees

Navigator Echoes. BioE 594 Advanced Topics in MRI Mauli. M. Modi. BioE /18/ What are Navigator Echoes?

Sketch of the MRI Device

Chapter 14:Physics of Magnetic Resonance

Tissue Characteristics Module Three

Post-Midterm Course Review

Principles of MRI EE225E / BIO265. Name That Artifact. RF Interference During Readout. RF Interference During Readout. Lecture 19

Spin-Echo MRI Using /2 and Hyperbolic Secant Pulses

MRS: IN VIVO SPECTROSCOPIC IMAGING MAIN POINTS

Velocity k-space analysis of Flow Effects in Echo-Planar, Spiral and Projection Reconstruction Imaging. Sangeetha Somayajula

Spin Echo Review. Static Dephasing: 1/T2 * = 1/T2 + 1/T2 Spin echo rephases magnetization Spin echoes can be repeated. B.Hargreaves - RAD 229

SENSE & SUSCEPTIBILITY: RESPIRATION-RELATED SUSCEPTIBILITY EFFECTS AND THEIR INTERACTIONS WITH PARALLEL IMAGING. John Sexton.

MRI beyond Fourier Encoding: From array detection to higher-order field dynamics

Nuclear Magnetic Resonance Imaging

The Basics of Magnetic Resonance Imaging

Chapter 26 Sequence Design, Artifacts and Nomenclature. Yongquan Ye, Ph.D. Assist. Prof. Radiology, SOM Wayne State University

Principles of Magnetic Resonance Imaging

Basic Concepts of MR Imaging, Diffusion MR Imaging, and Diffusion Tensor Imaging

RF Pulse Design. Multi-dimensional Excitation I. M229 Advanced Topics in MRI Kyung Sung, Ph.D Class Business

In vivo multiple spin echoes imaging of trabecular bone on a clinical 1.5 T MR scanner

Extended Phase Graphs (EPG)

Correction Gradients. Nov7, Reference: Handbook of pulse sequence

Correction of Physiologically Induced Global Off- Resonance Effects in Dynamic Echo-Planar and Spiral Functional Imaging

Towards Motion-Robust Magnetic Resonance Thermometry

Functional Magnetic Resonance Imaging (FMRI) is an imaging technique for

Quantitative/Mapping Methods

Computing FMRI Activations: Coefficients and t-statistics by Detrending and Multiple Regression

Disclosures. MR Physics. Recipe to Creating Images without Radiation. MRI Physics. Vector Math. What s In an Image 12/21/2012. Siemens Medical Systems

Part II: Magnetic Resonance Imaging (MRI)

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2008

RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES

The physics of medical imaging US, CT, MRI. Prof. Peter Bogner

SE Sequence: 90º, 180º RF Pulses, Readout Gradient e.g., 256 voxels in a row

Radiation Oncology Residents MR Physics Overview

The physics US and MRI. Prof. Peter Bogner

2.1.1 A Brief History of NMR The conception of NMR sprouted after the Pauli s prediction of nuclear spin in

MRI in Practice. Catherine Westbrook MSc, DCRR, CTC Senior Lecturer Anglia Polytechnic University Cambridge UK. John Talbot MSc, DCRR

HST.583 Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Fall 2006

BME I5000: Biomedical Imaging

BASIC MRI PHYSICS SPIN GYMNASTICS Don Plewes PhD, Walter Kucharczyk MD

RAD229: Midterm Exam 2015/2016 October 19, Minutes. Please do not proceed to the next page until the exam begins.

CIND Pre-Processing Pipeline For Diffusion Tensor Imaging. Overview

Basic Pulse Sequences II - Spin Echoes. TE=12ms TE=47ms TE=106ms TE=153ms UCLA. Radiology

Quantitative Susceptibility Mapping and Susceptibility Tensor Imaging. Magnetization and Susceptibility

MR Spectroscopy: The Physical Basis and Acquisition Strategies

Transcription:

On Signal to Noise Ratio Tradeoffs in fmri G. H. Glover April 11, 1999 This monograph addresses the question of signal to noise ratio (SNR) in fmri scanning, when parameters are changed under conditions of constant total scan time. While the results are presented in terms of spiral methods, the analysis itself is general. I. Introduction When planning an fmri experiment, it is desirable to optimize the SNR to maximize the statistical return on scan time investment. At first glance, the tradeoffs may seem unclear. As more slices are acquired, the TR needed to fit in the slices increases linearly. With longer TR the flip angle and degree of magnetization recovery between pulses both increase, leading to higher signal. However, at constant scan time, fewer time frames can be acquired, leading to lower statistics in the post-processing analysis. It is not obvious which of these effects dominates: increasing signal with longer TR, or decreased overall SNR with fewer frames. Similarly, the spiral sequence allows interleaving to reduce the readout time, which can be beneficial in reducing blurring and distortion. However, the use of interleaves also reduces the number of time frames that can be acquired. Is there a SNR penalty? II. SNR in fmri The SNR in a single image acquired with an interleaved method is given by SNR o = C' x y z n Tad f(tr,te), (1) where C' is a constant, x, y, z are the dimensions of the imaging voxel, nl is the number of interleaves (which might be 1 for either EPI or spiral), Tad is the duration of the acquisition readout ("A/D time"), and f is a function that describes the NMR signal dependence on the relaxation characteristics of the tissue, given for gradient-recalled echo method by f(tr,te) = (1 E) sin(α) 1 cos(α) E e TE/T*, () where TR and TE are scanner parameters with the usual connotation, T* is the transverse relaxation time, α is the flip angle, and E is given by E = e TR/T1. (3) It is known that the signal is maximized when α is chosen as the Ernst angle, cos(α e ) = E. This can be shown by differentiating Eq. () with respect to α and setting the result equal to zero. With this choice, Eq. () can be written 1

f(tr,te) = Thus, Eqs. (1) and (4) describe the SNR in a MR image. 1 E 1 +E e TE/T*. (4) However, the real interest is in the BOLD signal, which is related to the sensitivity of the image intensity to small changes in the relaxivity of the tissue, R* = 1/T*, caused by BOLD modulation. Accordingly, the BOLD signal S is given by S = di/dr* R, (5) where I is the signal intensity, proportional to the SNR given by (1) and (4), and R* is the change in relaxivity that causes the BOLD contrast changes. It can be shown that S is maximized when the echo time TE is set equal to T* by differentiating Eq. (5) with respect to TE and setting the result equal to zero. With this choice of echo time, we have that the BOLD SNR is given by BSNR o = C'' x y z n Tad 1 E 1 +E, (6) where C'' is another constant. To recap, Eq. (6) is valid and the signal is maximized when the flip angle is chosen as αe and the echo time is chosen as T*. Now for interleaved spiral acquisitions, the equivalent resolution (matrix size) in the slew-rate limited case is given by (1) N = 3 N Tad γ DSo 1/3, (7) where γ is the gyromagnetic ratio for protons, D is the FOV, and So is the gradient slew rate. Thus, for a given resolution N Tad is a constant. Then, (6) becomes BSNR o = C x y z 1 E 1 + E. (8) III. Total SNR In an fmri experiment, Nf frames are acquired. While there are numerous statistical analysis methods that can be utilized to generate the activation map, most have the common feature that the final SNR in the post-processed activation map can be calculated as if the image frames are added incoherently. This is approximately true even if there is significant temporal autocorrelation between contiguous time samples (due to the slow hemodynamic filter function), since the background noise in unactivated pixels is suppressed incoherently and thereby reduces the type II errors (false positives). Thus the total SNR (or Z score) will increase approximately as the square root of Nf, assuming the background noise is indeed uncorrelated from frame to frame. (This assumption may fail at higher field strengths where noise from basal metabolic processes in the gray matter may dominate the RF "Johnson" noise; we will not consider this issue here.)

Let Ts be the total scan time. Then, the number of frames is N f = T s /n TR, (9) and the total BOLD SNR for the scan is given by (denoted simply as SNR) SNR = BSNR o N f = C x y z T s N TR 1 E 1 + E. (10) Note that the SNR decreases as the number of interleaves increases, and increases as the square root of the total observation time, Ts. Equation (10) allows us to determine the answer to the question originally posed: How does the SNR vary as the TR is changed? The answer is shown in Fig. 1. IV. Experiment Scans were performed using a single-shot spiral sequence with resolution of 96x96, 4 cm FOV, 5 mm slices, TE = 30 ms, Bo = 3T, with the subject performing bilateral finger tapping with a 0s on/ 0s off block paradigm, for 00 s. Four scans were obtained using the following values of TR/FA/Nf: TR Nf FA, deg 3000 67 83 000 100 75 1000 00 60 500 400 45 The flip angle was chosen as approximately the Ernst angle, with T1 = 1400 assumed for gray matter at 3T. The data were analyzed using cross correlation with a sine wave and displayed as z-scores. Results for 4 of 6 slices are shown in Fig. as raw z-score maps. The average z-scores in activated pixels were 1.97,.3,.80,.93, respectively. V. Discussion As may be seen in the Fig. 1, the SNR is optimized by making TR as short as possible, although the dependence is rather modest when TR is T1. What this means in practice is that only as many slices as are necessary should be chosen, because the larger the number of slices, the longer TR, and the smaller the SNR. This also implies that a 3D acquisition, which has the shortest TR (the minimum TR is independent of the number of slices and is the same as a 1 slice D acquisition), will have superior SNR to a D acquisition when compared at some number of slices. Unfortunately, at this time the inherent SNR advantage of 3D techniques may not be realized, because other noise effects having to do with motion, may sometimes dominate. A second consequence of (10) is that increasing the number of interleaves decreases the SNR. That is, because N Tad is a constant, Tad decreases, and (7) shows that the SNR per image is constant if the 3

resolution is fixed. With more interleaves, there are fewer time frames and the total SNR decreases. Thus, increasing the number of interleaves should only be done if it is needed to avoid T* dropouts or blurring from off-resonance. A third consequence is that longer T1 decreases the SNR. This unfortunately occurs at higher field strength. This may be seen by evaluating (10) in the limit of TR = 0: SNR = C x y z T s N T1. Thus, some of the SNR gain from higher field strength is offset by this effect. With very short TR, there may be disadvantages from in-flow effects and additional motion artifacts in end slices from saturation of the imaging volume. FIGURE 1. Plot of SNR for GRE acquisition under conditions of constant scan time. The SNR is monotone decreasing for increasing TR, which implies that the number of slices (and consequently TR) should be kept as small as is necessary to cover the desired region. This calculation assumes the SNR increases x for 3T and uses T1 values appropriate to the two field strengths. 4

FIGURE. Activation maps for 4 slices (columns) for TR = 3s, s, 1s, 0.5s (top row to bottom row, respectively). The advantage of shorter TR is manifest as higher z-scores (bottom rows). The effect of decreasing TR from 1s to 0.5 s is small, as expected from Fig. 1. Reference 1. Glover, G.H. Simple analytic spiral k-space algorithm. Magn. Reson. Med.4:41-415 (1999). 5