Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Similar documents
Final Report CCEM.RF-K8.CL COMPARISON CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

Document No: TR 12 Issue No: 1

R SANAS Page 1 of 7

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Comparison protocol EURAMET project

Comparison of V and 10 V DC Voltage References

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Euramet project 1187 Comparison of Instrument Current Transformers up to 10 ka. Technical protocol (March 2012)

Final Report August 2010

ANNEXE 8. Technical protocols for the interlaboratory comparisons

High pressure comparison among seven European national laboratories

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

FINAL REPORT ON THE KEY COMPARISON EUROMET.AUV.A-K3

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

Supplementary Comparison EUROMET.EM-S23 EURAMET PROJECT No 815 FINAL REPORT

Centre for Metrology and Accreditation, MIKES

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Measurement and Calibration of a High-Sensitivity Microwave Power Sensor with an Attenuator

High pressure comparisons between seven European National Laboratories

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

Electricity and Magnetism Supplementary Guide to the JCRB Instructions for Appendix C of MRA

A bilateral comparison of a 1 kg platinum standard

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK

EUROMET Project 702 EUROMET.M.D-K4

Modelling and Measurand Identification Required for Defining the Mise en Pratique of the kelvin

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

Intercomparison of Thermal Expansion Measurements. EUROMET Project 275. Final Report

Euramet EM-S40. Bilateral Comparison KIM-LIPI / LNE. Final Report

Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm nm

Ajchara Charoensook, Chaiwat Jassadajin. National Institute of Metrology Thailand. Henry Chen, Brian Ricketts and Leigh Johnson

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Force Key Comparison EUROMET.M.F-K2. (50 kn and 100 kn) EURAMET Project No 518. Final Report. 14 July Pilot: NPL, United Kingdom

Euramet project 1075 Bilateral comparison

Final Report On COOMET Vickers PTB/VNIIFTRI Key Comparison (COOMET.M.H- K1.b and COOMET.M.H- K1.c)

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

The Fundamentals of Moisture Calibration

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

A guide to expression of uncertainty of measurements QA4EO-QAEO-GEN-DQK-006

Report on EURAMET.M.M-S9

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

Final Report on the Torque Key Comparison CCM.T-K2 Measurand Torque: 0 kn m, 10 kn m, 20 kn m Dirk Röske 1), Koji Ogushi 2)

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

Uncertainty of Radionuclide Calibrator Measurements of 99

Electricity and Magnetism Supplementary Guide for the Submission of CMCs

BUREAU INTERNATIONAL DES POIDS ET MESURES

Physikalisch Technische Bundesanstalt

MASS AND VOLUME COMPARISONS AT MIKES

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

CCQM-K45: Sn in tomato paste key comparison

APMP supplementary comparison of absorbed dose rate in tissue for beta radiation

Force Key Comparison CCM.F-K1.a and CCM.F-K1.b 5 kn and 10 kn. Aimo Pusa MIKES Finland

Bureau International des Poids et Mesures, Sèvres, France. *KEBS, Kenya Bureau of Standards P.O. Box , Nairobi, Kenya

Measurement method for the proficiency testing program

Final report on CCQM-K36.1 with erratum

A Method for Verifying Traceability in Effective Area for High Pressure Oil Piston-Cylinders

Traceable Mass Determination and Uncertainty Calculation

Portuguese Institute for Quality. Contents

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

Advances in Radio Science

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock +

Calibration of Temperature Block Calibrators

Importance and estimation of mismatch uncertainty for RF parameters in calibration laboratories

Uncertainties in Acoustics

M3003 EDITION 2 JANUARY 2007

Power Splitter Characterisation EM Day

Estimation of Complex and Linear Uncertainities in S-Parameter Measurements for Metrology Applications

CCM short note on the dissemination process after the proposed redefinition of the kilogram

Final Report On RMO VICKERS KEY COMPARISON COOMET M.H-K1

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

Final Results of Bilateral Comparison between NIST and PTB for Flows of High Pressure Natural Gas

Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report

BUREAU INTERNATIONAL DES POIDS ET MESURES

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

CERTIFICATE OF CALIBRATION

CCT-K5: Comparison of local realizations of the ITS-90 between the silver point and 1700 C using vacuum tungsten strip lamps as transfer standards

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

BUREAU INTERNATIONAL DES POIDS ET MESURES. Comparison of Quantum Hall Effect resistance standards of the PTB and the BIPM

Appendix B1. Reports of SMU

VNA error models: Comments on EURAMET/cg-12/v.01

Analysis of interlaboratory comparison when the measurements are not normally distributed

The ITS-90 after definition of neon isotopic reference composition: extent of the isotopic effect tested on previous inter-comparison results

COOMET.M.V-S2 (587/RU-a/12) COOMET SUPPLEMENTARY COMPARISON IN THE FIELD OF MEASUREMENTS OF LIQUIDS KINEMATIC VISCOSITY

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

COMPARISON OF HUMIDITY MEASUREMENTS USING A DEW POINT METER AS A TRANSFER STANDARD APMP-IC-1-97 REPORT

Nanometrology and its role in the development of nanotechnology

Good practice guide containing experimental results and recommendations for the selection, preparation and calibration of the temperature sensors

Project 887 Force Uncertainty Guide. Andy Knott, NPL EUROMET TC-M Lensbury, Teddington, United Kingdom 1 March 2007

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

EA 4/02. Expression of the Uncertainty of Measurement in Calibration. Publication Reference

Technical protocol for the CCL-K11 key comparison of optical frequency/wavelength standards.

Essentials of expressing measurement uncertainty

Surface Mount Chip Capacitors

Transcription:

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS Jan P.M. de Vreede Department of Electricity, Radiation and Length NMi Van Swinden Laboratorium Thijsseweg 11, 2629 JA Delft, the Netherlands e-mail: JdeVreede@nmi.nl January 2007 Abstract: In this bilateral comparison between NMi-VSL and MIKES the calibration factor and reflection coefficient of thermistor mounts were measured at 8 frequencies between 10 MHz and 18 GHz. The maximum comparison uncertainty for the calibration factor ranges from 0.8 % at 50 MHz to 2.2 % at 18 GHz. All results are consistent within the comparison uncertainty, using k=2. The uncertainty stated for the reflection coefficient was up to 0.02, consistent within the claimed uncertainty. Taking these facts into account, the results show a good agreement for both the calibration factor and the reflection coefficient. Also the DoE of MIKES is calculated for entry in the KCDB as part of the key comparison EUROMET.EM.RF-K8.CL Final report Euromet.EM.RF-K8.1.CL.doc 1/19

Table of Contents 1. Introduction p. 3 2. Participants and schedule p. 3 3. Transfer standard and required measurements p. 3 4. Behaviour of the transfer standard p. 4 5. Measurement methods p. 4 6. Technical protocol p. 5 7. Measurement results p. 5 7.1. General results p. 5 7.2. Determining the Reference Value p. 8 7.2.1 Linking to SI: p. 9 7.3 Degrees of Equivalence for MIKES p. 10 7.4. Uncertainty budgets p. 12 8.1 Reflection coefficient p. 12 8.2 Torque wrench p. 12 9. Conclusions p. 12 10. References p. 13 App. A Degrees of equivalence for calibration factor at 10 MHz and 18 GHz p. 14 App. B Participant uncertainty budget for a thermistor mount p. 15 App. C Comparison protocol and schedule p. 17 C1: Technical Protocol p. 17 C2: Contact Persons p. 19 Final report Euromet.EM.RF-K8.1.CL.doc 2/19

1. Introduction In recent years European national metrology institutes had the possibility to participate in two key comparisons in the field of power measurements using thermistor mounts as device under test [1,2]. The first one (CCEM.RF-K8.CL) was under the auspices of the Comite Consultatif d Ēlectricité et Magnetisme (CCEM) and its working group for radiofrequency quantities (GT-RF). The second one (EUROMET.EM.RF-K8.CL; Euromet project 633) was a follow-up exercise of the first one to investigate some problems found in the first one, but also to provide some laboratories with a direct link to the other national laboratories. NMi Van Swinden Laboratorium (NMi-VSL) acted as pilot laboratory for both comparisons. A few years ago MIKES, the national metrology institute of Finland, built facilities in the radio frequency field. After obtaining experience in practical work it now wants to verify formally the performance in this field. As power is one of the most important quantities, MIKES expressed the wish to participate in a bilateral comparison with one of the participants in the previous comparisons. The most logical candidate would be the pilot laboratory of these comparisons. The bilateral comparison is formalized as Euromet project 818 and has been entered in the KCDB under the code EUROMET.EM.RF-K8.1.CL. In essence the same technical procedure is followed as in CCEM.RF-K8.CL and Euromet 633. 2. Participants and schedule After some discussion about the most convenient way to carry out the comparison it was decided to use DUTs from both laboratories. Each laboratory would measure its own device at the beginning and the end of the exercise and measure the other DUT while its own DUT was still in the laboratory. The measurements are considered to be of a routine nature. Table 1 gives the basic information on the comparison. Acronym NMi-VSL MIKES Table 1. List of participants and measurement dates. National Metrology Institute NMi Van Swinden Laboratorium - Pilot Centre for Metrology and Accreditation Country The Netherlands Finland Standard at the laboratory December 2004 February 2005 December 2004 February 2005 Date of submission of report July 2005 June 2005 3. Transfer Standard and required measurements Because the project is a follow-up of previous comparisons it was decided to use a set of similar devices. Hence the following DUTs were used: - Hewlett Packard model 8478B Opt.H48 (sn.3318 A 25650), owner NMi-VSL; the code DUT1 is used in this report; - Hewlett Packard model 8478B (sn.2106 A 23988), owner MIKES; the code DUT2 is used in this report. All devices are equipped with a Type-N connector. Final report Euromet.EM.RF-K8.1.CL.doc 3/19

The DUT power detectors are thermistor mounts that must be used in connection with a thermistor bridge that keeps the thermistor resistance to a fixed value of 200 Ω. Several commercial thermistor bridges are available to determine the DC substitution power P DC when RF power is applied to the thermistor mount. The mount has an available compensation scheme that allows the detection of power, even when the ambient temperature is not constant. The two signals (V RF and V comp ) from two separated bridges inside of commercial thermistor bridges may be detected separately to determine P DC. It is also possible to use the recorder output, which is proportional to P DC because of an internal electronic manipulation with the V RF and V comp signals. The quantity under investigation in this comparison is the calibration factor K, which is defined by: K= P DC /P inc with: P DC - the DC substitution power determined by the thermistor bridge of the participant and P inc - the RF power incident to the thermistor mount (DUT) at the measurement frequency. The participants were asked to submit measurement results on each thermistor mount at 8 frequencies (10 MHz, 50 MHz, 1 GHz, 4 GHz, 8 GHz, 12 GHz, 15 GHz and 18 GHz) concerning its calibration factor and also its reflection coefficient, both with an uncertainty (coverage factor k =1). To substantiate the technical performance the technical protocol put emphasis on the uncertainty statements and the consistency of the measurement results. Hence, a detailed uncertainty budget, containing sources and magnitudes, was requested, as well as the traceability of the standards, in order to take into account the possibility of correlation between the results. The quantity reflection is necessary for the uncertainty calculations. In this comparison it is not the quantity under investigation. In the Technical Protocol no requirements are given concerning the ambient conditions. 4. Behaviour of the transfer standard As the DUTs are property of the participants, the in-house stability is known. Based upon the experience obtained in previous comparisons and the small number of participants, no additional checks concerning stability for transport have been performed. The two measurements done at each laboratory on its own device are used to obtain an idea about possible drift and transport influences. No significant influence is found. Hence, the official results are based upon the average result for the laboratory s own device. The VSL-mount had a little dent on the outer connector contact end. Some instabilities (informal information) in that mount were found at MIKES, but not at NMi-VSL. 5. Measurement methods As indicated in the guidelines each laboratory should use the same measurement instrumentation as used for high level calibration for external customers. All systems are based on a (in)direct comparison between a (working) standard and the DUT. For each laboratory the measurement procedure (including traceability) is briefly described here. Also information about the measurement of the reflection is given. NMi-VSL pilot laboratory: A substitution system is used, where the signal comes from a stable signal generator, with a 10 db attenuator to improve the VSWR of the output port. The standard and DUT are placed alternatively on Final report Euromet.EM.RF-K8.1.CL.doc 4/19

the output port of the generator, and are of similar design (thermistor mounts). The response of the thermistor mounts is obtained using the recorder output of a self-balancing bridge, HP 432A. The recorder output has been characterised during normal maintenance using V RF and V comp readings. The DUT is measured against at least two standards in different orientations. Traceability is based on the primary VSL power facility (microcalorimeter): the working standards are calibrated in the microcalorimeter every half year. The reflection coefficients are measured using Vector Network Analysers (HP 8753E with external test set and Wiltron 360A). MIKES: A power splitter system is used with an Agilent 8481A power sensor permanently attached to one arm. The DUT and the standard are attached alternatively to the other arm of the splitter. An HP 432B bridge and an Agilent 34970A Data Acquisition system is used for reading three signals: V RF, V comp and their difference. From these values the power reading of the DUT and the standard is calculated. The DUT and the standard were measured in three different positions 120 degrees apart. The standard (another HP 8478B thermistor mount) is traceable to NPL, UK. The reflection coefficients are measured using Agilent 8510C and E8357A VNAs. 6. Technical protocol In the protocol ("Technical Protocol for the comparison", see Appendix C1) participants were asked to present their measurement results in the format of the mean of the calibration factor and the magnitude of the reflection coefficient at the 8 frequencies, including a statement of uncertainty with a coverage factor of k = 1. In addition they were requested to give a detailed uncertainty budget that would allow the pilot laboratory to determine whether important contributions might have been overlooked and to allow for drafting a common agreed basis for uncertainty calculation in this field. Reference was made to the EA document on uncertainty (EA-04/02) [3] that gives guidance for providing such an uncertainty budget, and to the uncertainty budgets as given the Euromet project 393. Also the traceability for the standards used should be provided to ascertain that correlation between measurement results would not be overlooked. 7. Measurement results 7.1. General results The participants were asked to submit measurement results on each thermistor mount at 8 frequencies (10 MHz, 50 MHz, 1 GHz, 4 GHz, 8 GHz, 12 GHz, 15 GHz and 18 GHz) concerning its calibration factor and also its reflection coefficient, both with an uncertainty (coverage factor k =1). After receiving the measurement data (including uncertainty statement) the coordinator has compiled these results in an Excel spreadsheet for further analysis. The pilot laboratory has used the k=2 uncertainties as an analysis tool throughout this document. Figures 1 and 2 give an overview over all data relevant for the comparison. Figure 1 contains the three series of data concerning the calibration factor and Figure 2 contains the results on the reflection coefficient as measured by the two laboratories. The uncertainty bars are the k=2 values based upon the information given by each of the laboratories. In general the pilot laboratory gave larger uncertainties for its measurements. Final report Euromet.EM.RF-K8.1.CL.doc 5/19

DUT1: VSL-8478B Calibration Factor 1.01 1.00 0.99 0.98 0.97 0.96 0.95 0.94 0.93 0.92 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Frequency (GHz) VSL-1 MIKES VSL-2 DUT2: MIKES-8478B Calibration Factor 1.01 1.00 0.99 0.98 0.97 0.96 0.95 0.94 0.93 0.92 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Frequency (GHz) MIKES-1 VSL MIKES-2 Figure 1: The calibration factor as function of frequency as determined by the two participants. Each laboratory has measured its own thermistor mount at the beginning and at the end of the comparison. The uncertainty bars refer to the expanded standard deviation (k=2) in the results of the participants. Final report Euromet.EM.RF-K8.1.CL.doc 6/19

DUT1: VSL-8478B 0.14 0.12 0.1 Refl.Coeff 0.08 0.06 0.04 0.02 0 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Frequency (GHz) VSL-1 MIKES DUT2: MIKES-8478B 0.1400 0.1200 0.1000 Refl.Coeff 0.0800 0.0600 0.0400 0.0200 0.0000 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Frequency (GHz) VSL MIKES-1 Figure 2: The reflection coefficient as function of frequency as determined by the two participants. The uncertainty bars refer to the standard deviation (k=1) in the results of the participants. Final report Euromet.EM.RF-K8.1.CL.doc 7/19

7.2. Determining the Reference Value For this bilateral comparison the result from the pilot laboratory is taken as the reference value. In the analysis process the difference in results between the two laboratories is taken to be the official result of the comparison. If a laboratory has measured the device twice, the average value of the two measurements are used. For the associated uncertainty the average value of the two uncertainties is used, as the uncertainty of the Type-B evaluation is dominant in almost all cases and the measurement set-ups were not changed in between. As can be seen in Table 2, the differences between the two DUTs is small compared to the uncertainty. Hence, a mathematically more accurate model is not applied, as we expect only negligible changes will occur. The result presented is defined as: Result = dcf = CF(MIKES) - CF(NMi-VSL), where CF is the calibration factor measured at a specific frequency. In Table 2 and Figure 3 the results for DUT1 and DUT2 are given. The error bars are based on the RSS value of the individual uncertainty budgets. A slight systematic deviation as function of frequency seems to be present, but still within the combined uncertainty for k=2. Difference in CF 0.030 0.020 CF(MIKES)-CF(VSL) 0.010 0.000-0.010-0.020-0.030-0.040-0.050 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Frequency (GHz) DUT1 DUT2 Figure 3: The difference in measured calibration factor as function of frequency. The calibration factor determined by the pilot is taken as reference. The error bars are based on the RSS value of the individual uncertainty budgets, using k=2. Final report Euromet.EM.RF-K8.1.CL.doc 8/19

Table 2: Differences in measured Calibration Factor per DUT and averaged to one result per frequency Frequency DUT1 DUT2 Average GHz dcf Unc dcf Unc dcf Unc 0.01 0.0026 0.0151 0.0011 0.0099 0.0019 0.0125 0.05 0.0011 0.0096 0.0010 0.0073 0.0010 0.0085 1.00 0.0004 0.0112 0.0015 0.0092 0.0009 0.0102 4.00-0.0007 0.0129 0.0001 0.0113-0.0003 0.0121 8.00-0.0081 0.0189-0.0059 0.0144-0.0070 0.0167 12.00-0.0035 0.0179-0.0038 0.0159-0.0036 0.0169 15.00-0.0079 0.0203-0.0103 0.0176-0.0091 0.0190 18.00-0.0173 0.0238-0.0173 0.0211-0.0173 0.0225 7.2.1. Linking to SI: As already mentioned above MIKES is traceable to NPL, UK, for the quantity of power. Whether part of the systematic deviation is due to the different traceability chains, can be investigated using the results of NPL and NMi-VSL in previous comparisons. As this comparison should be linked directly to Euromet project 633 (EUROMET.EM.RF-K8.CL) it is logical to use this comparison for a potential systematic deviation in the realisation of the quantity power at NPL and NMi-VSL. In Table 3 this information is given (obtained from [4]), using the average of the measured values on the devices TM4 and TM6. The column Spread refers to the standard deviation for the two results obtained in project 633 (their uncertainties are not used here). In Figure 4 the difference between the results from NPL and NMi-VSL is given as function of frequency, together with the results from Figure 3 by averaging the results for DUT1 and DUT2 and using the average of the uncertainty statements. Table 3: Differences between NPL and NMi-VSL as measured in EUROMET.EM.RF-K8.CL NPL-VSL Frequency [GHz] Average: Spread: 0.01 0.0021 0.0011 0.05 0.0004 0.0001 1.00 0.0011 0.0004 4.00-0.0002 0.0012 8.00-0.0028 0.0013 12.00-0.0019 0.0030 15.00-0.0083 0.0034 18.00-0.0059 0.0020 Final report Euromet.EM.RF-K8.1.CL.doc 9/19

Difference in CF 0.0200 0.0100 CF(MIKES)-CF(VSL) 0.0000-0.0100-0.0200-0.0300-0.0400-0.0500 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Frequency (GHz) Project 818 NPL-VSL Figure 4: The data of project 818 refer to the measured difference in Calibration Factor between MIKES and NMi-VSL (using the data from Table 2). The data marked as NPL-VSL refer to similar results in project 633 between NPL and NMi-VSL (using the information from Table 3). Both comparisons shows a similar trend, but of slightly different magnitude. Between NMi-VSL and MIKES a small systematic deviation might exist. Informal information indicates that the difference in results between NPL and VSL might be somewhat larger than found in Table 3, especially at 18 GHz. 7.3 Degrees of Equivalence for MIKES The final step in linking the present comparison to the previous one (project 633) can now be done using the Degrees of Equivalence (DoE) of NMi-VSL in the previous comparison (see Table 4). The DoE consists of the difference of the results of VSL from the KCRV and the associated uncertainty. Table 4: Degrees of Equivalence for NMi-VSL in EUROMET.EM.RF-K8.CL Frequency [GHz] Value Uncertainty (k=2) 0.01-0.0009 0.0224 0.05-0.0002 0.0055 1.00-0.0005 0.0051 4.00 0.0004 0.0069 8.00 0.0035 0.0114 12.00 0.0033 0.0106 15.00 0.0077 0.0129 18.00 0.0035 0.0157 The value for the final result is obtained as: CF(MIKES)= dcf + DoE(NMi-VSL: value) where dcf is obtained from Table 2 and DoE from Table 4. Final report Euromet.EM.RF-K8.1.CL.doc 10/19

For the uncertainty we have to look to the correlation in the measurement results of NMi-VSL, and especially the uncertainty part of the DoE. The uncertainty in the KCRV itself is in general less than 0.001 and at 18 GHz less than 0.004. This means that this part is negligible compared to the expanded uncertainties given by NMi-VSL. In Table 5 the uncertainty contributions given by NMi-VSL are compared. The contribution due to statistical spread in the measurements (Type A) is usually small compared to the systematic contributions. Except for 10 MHz, the latter contributions in this comparison are larger than in the previous comparison. This is mainly due to the variations found in the calibration of the working standards in the microcalorimeter (dominant term in the uncertainty: see Table B1). The power ratio drift (or variation) between DUT and standard is much smaller. This indicates that the reproducibility of the measurement system is small enough to neglect effects due to the time interval between the two comparisons 633 and 818. As the systematic contributions for NMi-VSL should be taken into account only once, it is decided to use the uncertainty evaluation results obtained in Table 2. Table 5: Uncertainties relevant in linking CF (MIKES) to EUROMET.EM.RF-K8.CL Uncertainties (k=2) for measurements at NMi-VSL Frequency [GHz] Typical values for Type A Expanded values in 633 Expanded values in 818 Expanded uncertainty in measurement result (Table 2) 0.01 0.001 0.0224 0.0112 0.0125 0.05 0.001 0.0055 0.0080 0.0085 1.00 0.002 0.0051 0.0094 0.0102 4.00 0.002 0.0069 0.0112 0.0121 8.00 0.004 0.0114 0.0164 0.0167 12.00 0.006 0.0106 0.0159 0.0169 15.00 0.008 0.0129 0.0182 0.0190 18.00 0.010 0.0157 0.0196 0.0225 Comparing the data in the last two columns it is clear that the uncertainty statement of the linking laboratory plays a significant role in the overall uncertainty. The result is presented in Figure 5 and Table 6. Table 6: Final result of comparison: MIKES relative to reference value of EUROMET.EM.RF-K8.CL Frequency [GHz] Deviation in CF Uncertainty (k=2) 0.01 0.0009 0.0125 0.05 0.0008 0.0085 1.00 0.0004 0.0102 4.00 0.0001 0.0121 8.00-0.0035 0.0167 12.00-0.0003 0.0169 15.00-0.0015 0.0190 18.00-0.0138 0.0225 Final report Euromet.EM.RF-K8.1.CL.doc 11/19

Calibration Factor (MIKES) relative to EUROMET.EM.RF-K8.CL Difference in CF 0.025 0.020 0.015 0.010 0.005 0.000-0.005-0.010-0.015-0.020-0.025-0.030-0.035-0.040 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Frequency (GHz) Figure 5: The average difference between the results from MIKES and the reference value obtained in the EUROMET.EM.RF-K8.CL 7.4. Uncertainty budgets For the uncertainty budget the EA guidance document is followed. Compared to the previous comparisons NMi-VSL has changed the lay-out of the budget to be in line with that of MIKES. 8.1. Reflection coefficient The two laboratories agree very well in their results. 8.2. Torque wrench In the previous comparisons the use of a torque wrench was requested. In this comparison the technical protocol did not ask for it explicitly, as DUT2 (owned by MIKES) did not have a suitable surface. Both laboratories have used a torque wrench, when possible. 9. Conclusions The comparison uncertainty for the calibration factor ranges from 0.8 % at 50 MHz to 2.2 % at 18 GHz, using k=2. The overall uncertainty is largely dominated by the measurement uncertainty of NMi-VSL. All results are consistent within the claimed uncertainty. The uncertainty stated for the reflection coefficient was up to 0.02, consistent within the claimed uncertainty. Taking these facts into account, the results show a good agreement for both the calibration factor and the reflection coefficient. Final report Euromet.EM.RF-K8.1.CL.doc 12/19

10. References [1] Jan P M de Vreede, Final Report of the comparison CCEM.RF-K8.CL: Calibration factor of thermistor mounts, 2005, Metrologia 42 01008 [2] Jan P.M. de Vreede, Final Report: EUROMET.EM.RF-K8.CL COMPARISON - CALIBRATION FACTOR OF THERMISTOR MOUNTS - September 2006 [3] Expression of the Uncertainty in Measurement in Calibration document EA-04/02, December 1999, available on the EA-website (www.european-accreditation.org) [4] Jan P.M. de Vreede, Final Report LINKING EUROMET PROJECT 633 To CCEM.RF-K8.CL (CALIBRATION FACTOR OF THERMISTOR MOUNTS) September 2006 Final report Euromet.EM.RF-K8.1.CL.doc 13/19

APPENDIX A Degrees of equivalence for calibration factor of MIKES Regional Key comparison: Euromet.EM.RF-K8.1.CL Measurand: calibration factor in coaxial 7 mm transmission line Nominal value: 1.00 Pilot laboratory: NMi-VSL Travelling standards: two thermistor mounts identified as DUT1 and DUT2; both have a male type N 50 ohm connector As the actual calibration factors of the DUTs are not relevant for the quality of the measurement results, the comparison reference value is shifted to zero. As the comparison is aimed at linking the results from MIKES to the key comparison EUROMET.EM.RF-K8.CL, the results from the pilot laboratory are adjusted in such a way that they agree in value with the DoE of the key comparison mentioned above. The uncertainty in the DoE of the pilot laboratory is not taken into account, as otherwise the contributions of systematic nature to the uncertainties would be counted twice. D i = the difference from the reference value for laboratory i U i = the uncertainty of D,i taken into account the uncertainty of the bilateral comparison. Frequency GHz Laboratory D i U i 0.01 MIKES 0.0009 0.0125 NMi-VSL -0.0009 0.0224 0.05 MIKES 0.0008 0.0085 NMi-VSL -0.0002 0.0055 1.0 MIKES 0.0004 0.0102 NMi-VSL -0.0005 0.0051 4.0 MIKES 0.0001 0.0121 NMi-VSL 0.0004 0.0069 8.0 MIKES -0.0035 0.0167 NMi-VSL 0.0035 0.0114 12.0 MIKES -0.0003 0.0169 NMi-VSL 0.0033 0.0106 15.0 MIKES -0.0015 0.0190 NMi-VSL 0.0077 0.0129 18.0 MIKES -0.0138 0.0225 NMi-VSL 0.0035 0.0157 Degrees of Equivalence 0.03 Deviation from Reference 0.02 0.01 0-0.01-0.02-0.03-0.04 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Frequency (GHz) MIKES NMi-VSL Final report Euromet.EM.RF-K8.1.CL.doc APPENDIX A 14/19

Appendix B Participant uncertainty budget for a thermistor mount Pilot laboratory: NMi-VSL The uncertainty budget refers to an individual measurement and not to the average of a series of measurements. The data are indicative for the frequency range and the type of DUT. The calibration factor in each measurement series is determined as: M X K X = K S * PR * pc * M S where K X = calibration factor of the DUT K S = calibration factor of the standard M X = mismatch factor for the DUT M S = mismatch factor for the standard p c = linearity correction factor P R = Ratio of the power readings for the DUT and standard. Table B.1 shows a typical example for a measurement on DUT2, where ukx indicates the calculated uncertainty for this measurement. Table B.1: Uncertainty budget for a specific measurement during EUROMET.EM.RF-K8.1.CL Frequency GHz P R (%) p c (%) M S (%) M X (%) K S (%) K X (%) ukx (%, k=2) 0.01 0.09 0.20 0.24 0.21 0.41 96.79 1.13 0.05 0.00 0.20 0.07 0.06 0.21 99.46 0.61 1 0.03 0.20 0.03 0.03 0.31 99.24 0.74 4 0.13 0.20 0.05 0.10 0.41 98.15 0.98 8 0.39 0.20 0.17 0.16 0.56 97.55 1.49 12 0.38 0.20 0.27 0.04 0.66 96.49 1.67 15 0.08 0.20 0.47 0.27 0.65 96.11 1.75 18 0.19 0.20 0.21 0.11 0.90 95.16 1.93 Final report Euromet.EM.RF-K8.1.CL.doc APPENDIX B 15/19

MIKES The calibration factor was determined as where K X K S δk D M X M S p C p X p S M K X = ( K S + δk D ) M X S p C p p unknown sensor calibration factor standard sensor calibration factor drift of standard sensor calibration factor mismatch factor for unknown power sensor mismatch factor for standard power sensor linearity correction factor power ratio of standard sensor to reference power ratio of unknown sensor to reference X S Note that that here the calibration factor is defined with respect to the DC substitution power, unlike in EA-4/02 where the calibration factor was defined with respect to the 50 MHz reference value. Table 1 shows a typical example of the uncertainty budget breakdown. Table 1. HP 8478B SN 2106A 23988 first round Frequency u(p S ) u(p X ) u(p C ) u(m S ) u(m X ) u(δk D ) u(k S ) K X u(k X ) ν eff [GHz] [abs. %] [abs. %] [abs. %] [abs. %] [abs. %] [abs. %] [abs. %] [abs. %] [abs. %] 0,01 0,0119 0,0076 0,0097 0,1532 0,1575 0,1005 0,2514 97,12 0,35 142 0,05 0,0211 0,0096 0,0100 0,0296 0,0298 0,1000 0,1500 99,59 0,19 100 1,00 0,0214 0,0143 0,0099 0,0237 0,0249 0,1003 0,2005 99,37 0,23 79 4,00 0,0257 0,0141 0,0098 0,0926 0,0968 0,1004 0,2007 98,16 0,26 128 8,00 0,0304 0,0177 0,0097 0,0835 0,0682 0,1007 0,2518 97,08 0,29 89 12,00 0,0333 0,0153 0,0096 0,0707 0,0363 0,1007 0,3022 96,41 0,33 70 15,00 0,0366 0,0167 0,0095 0,0489 0,1336 0,1005 0,3014 95,25 0,35 87 18,00 0,0294 0,0666 0,0094 0,2233 0,3408 0,1001 0,3504 93,82 0,55 180 Note: the stated u(k X ) refers to a coverage factor of k=1. Final report Euromet.EM.RF-K8.1.CL.doc APPENDIX B 16/19

APPENDIX C C1) Technical Protocol Technical Protocol for Euromet project 818 Scope: This project is a follow-up of Euromet project 633. This project is a bilateral comparison to link MIKES to the participants of Euromet projects 393 and 633. In essence the same procedure will be followed as in Euromet 393 and its international extensions CCEM.RF-K8.CL and Euromet 633 (EUROMET.EM.RF-K8.CL). In this way a firm link may be obtained with the other comparisons. Measuring quantity: Power sensors are usually calibrated in terms of calibration factor. In most cases a reference frequency of 50 MHz is used to obtain the frequency dependence of a power sensor. Thermistor mounts are considered to be the most fundamental power measuring device for traceability to the fundamental SI units. Therefore they are used as primary standards in most of the national standards laboratories. Also high level calibration laboratories use these devices as their highest internal standard. The purpose of the exercise is to determine the level of consistency of calibration results as given by different national standards laboratories. The main measuring quantity therefore is the calibration factor as determined at a number of prescribed frequencies, together with the appropiate uncertainty statement. Also the value of the reflection coefficient has to be determined, as it is, at least, necessary for the uncertainty calculation. Travelling standards A set of two thermistor mounts is used, both with type-n male connector: 1: travelling standard - Hewlett Packard - model 8478b Option H48 - sn. 3318 A 25650 - connector: Type-N male - Ident: H48.25 2: travelling standard - Hewlett Packard - model 8478b - sn. 2106 A 23988 - connector: Type-N male Measurement procedure As already indicated, the normal laboratory procedure for high level calibration of power sensors should be used. Hence, no attempt should be made to improve facilities just for this comparison. Usually customers expect to be served within a couple of weeks. This is also the main reason for allowing a relative short turn-around time for the measurements. The two travelling standards are to be calibrated at the following frequencies (in GHz): 0.010, 0.050, 1.00, 4.00, 8.00, 12.00, 15.00 and 18.00. Submission of results Each laboratory is expected to submit its report to the coordinator within 6 weeks after the end of its measuring period. Anyway, the pilot laboratory needs sufficient information to make a first evaluation of the results before a general discussion can take place on the draft report as prepared by the pilot laboratory. A breakdown of the uncertainty budget is an essential part of evaluating measurement results. According to the CIPM guidelines the ISO Guide on the Calculation of Uncertainties in Measurements (GUM) should be Final report Euromet.EM.RF-K8.1.CL.doc APPENDIX C 17/19

followed. A practical implementation of this document within the European Accreditation bodies is the EA- 04/02 (1999) document. (this document is available on the EA website: www.european-accreditation.org). The report should also contain at least a short description of the measurement set-up, preferably with some schematic drawing, the relevant statistical information on the individual measurement results and traceability chain. An example of presenting a summary of the basic results is given in the table below. Results of Euromet project XXX Laboratory: Frequency [GHz] Calibration factor Uncertainty (k=1) Reflection coefficient Uncertainty in refl. coefficient (k=1) 0.01 0.05 1.00 4.00 8.00 12.00 15.00 18.00 Contributions to the uncertainty In Euromet project 393 and 633 the following contributions were considered mainly to be responsible for the uncertainty in the determination of the calibration factor: - reflection coefficient of the source - reflection coefficient of the internal working standard - reflection coefficient of the travelling standard - uncertainty in the calibration factor of the internal working standard. The term internal working standard is used here to describe the power sensor used directly in calibrating the travelling standards. The BIPM Guideline, which is also used for Euromet comparisons with minor changes, requests an uncertainty evaluation at the level of one standard uncertainty, giving also the number of degrees of freedom. Discussion of results It is expected that an open discussion will take place as quickly as possible after distributing a draft report containing a compilation of the results and a first attempt of interpretation. Afterwards the final result can be published in Metrologia (in short form) and preferably as a full paper in the open literature as well. Problems during the exercise: If technical and/or other problems arise, it is of the utmost importance to contact immediately the coordinator to discuss the matter and to inform the laboratory next in line about this fact. If the problem can not be solved within the allowed time frame, it will be necessary to adapt the schedule by shifting a few laboratories to a latter time slot. It is assumed that the participating laboratory takes care of insurance of the package during the stay at the laboratory and the transportation to the next participant. Transport and customs Final report Euromet.EM.RF-K8.1.CL.doc APPENDIX C 18/19

The travelling standards can be sent using regular package mail. The devices are stored in a plastic container, which is provided by the coordinator. Additional packaging as protection is suggested. Inside the European Union no customs papers are necessary, but a pro-forma invoice is provided in case of questioning. For all participants outside the Union, an ATA-carnet will be provided, if applicable. Circulation time schedule The circulation schedule is agreed upon between the original participants. Updates of the schedule will be sent when and where necessary. A turn-around time between laboratories of 3 weeks is used. It is the responsibility of each participating laboratory to inform the next participant in advance to arrange the transportation of the standards, and to inform the coordinator about the date of transportation. The time schedule is as follows: Institute Measuring Period DUT Contact person NMi VSL November / December 2004 VSL Jan de Vreede MIKES December 2004 / January 2005 MIKES/VSL Kari Ojasalo NMi VSL January 2005 MIKES/VSL Jan de Vreede MIKES February 2005 MIKES Kari Ojasalo Coordinator The pilot laboratory for this comparison is NMi Van Swinden Laboratorium (VSL). The coordinator for this comparison is: Dr. Jan P.M. de Vreede NMi Van Swinden Laboratorium Schoemakerstraat 97 P.O. Box 654 2600 AR Delft The Netherlands Telephone: +31-15 269 1500 Fax: +31-15 261 2971 E-mail: JdeVreede@nmi.nl Finland: MIKES C2) Contact Persons Mr. Kari Ojasalo MIKES (Centre of Metrology and Accreditation) Electricity and Acoustics P.O.Box 9 (Tekniikantie 1) FIN-02151 ESPOO FINLAND The Netherlands: NMi VSL Dr. Jan P.M. de Vreede Department of Electricity, Radiation and Length Postbus 654 2600 AR DELFT THE NETHERLANDS Tel.: + 358 010 6054 423 Fax: + 358 010 6054 299 e-mail: kari.ojasalo@mikes.fi Tel.: +31 15 269 1500 Fax: +31 15 261 2971 e-mail: JdeVreede@nmi.nl Final report Euromet.EM.RF-K8.1.CL.doc APPENDIX C 19/19