Accurate thickness measurement of graphene

Similar documents
Instrumentation and Operation

Supporting information

SUPPLEMENTARY INFORMATION

Instantaneous reduction of graphene oxide at room temperature

DETERMINATION OF THE ADHESION PROPERTIES OF MICA VIA ATOMIC FORCE SPECTROSCOPY

Module 26: Atomic Force Microscopy. Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM

Imaging Methods: Scanning Force Microscopy (SFM / AFM)

Electronic Supplementary Information

Imaging Carbon materials with correlative Raman-SEM microscopy. Introduction. Raman, SEM and FIB within one chamber. Diamond.

The Powerful Diversity of the AFM Probe

Basic Laboratory. Materials Science and Engineering. Atomic Force Microscopy (AFM)

Greg Andreev and Aravind Vijayaraghavan 9/27/13

Improving the accuracy of Atomic Force Microscope based nanomechanical measurements. Bede Pittenger Bruker Nano Surfaces, Santa Barbara, CA, USA

Friction Coefficient Mapping of 2D Materials via Friction-Induced Topographic Artifact in Atomic Force Microscopy

NIS: what can it be used for?

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur

Chapter 10. Nanometrology. Oxford University Press All rights reserved.

Correlative Raman Imaging of Polymeric Materials

Università degli Studi di Bari "Aldo Moro"

Jiang Deng, 1 Michael R. Nellist, 2 Michaela Burke Stevens, 2,# Christian Dette, 2 Yong Wang, 1 and Shannon W. Boettcher 2, *

The design of an integrated XPS/Raman spectroscopy instrument for co-incident analysis

Scanning Probe Microscopy. Amanda MacMillan, Emmy Gebremichael, & John Shamblin Chem 243: Instrumental Analysis Dr. Robert Corn March 10, 2010

GRAPHENE ON THE Si-FACE OF SILICON CARBIDE USER MANUAL

Supplementary Figure 1: Micromechanical cleavage of graphene on oxygen plasma treated Si/SiO2. Supplementary Figure 2: Comparison of hbn yield.

Atomic Force Microscopy Characterization of Room- Temperature Adlayers of Small Organic Molecules through Graphene Templating

Large Scale Direct Synthesis of Graphene on Sapphire and Transfer-free Device Fabrication

Introduction to Scanning Probe Microscopy Zhe Fei

Probing the Chemical Structure of Monolayer Covalent-Organic Frameworks Grown via Schiff-Base Condensation Reactions

Supporting Information

Shedding New Light on Materials Science with Raman Imaging

Lecture 4 Scanning Probe Microscopy (SPM)

And Manipulation by Scanning Probe Microscope

Surfactant-free exfoliation of graphite in aqueous solutions

Kavli Workshop for Journalists. June 13th, CNF Cleanroom Activities

Chapter 12. Nanometrology. Oxford University Press All rights reserved.

RHK Technology Brief

Supporting Information. Real-time nanoscale open-circuit voltage dynamics of perovskite solar cells

SUPPLEMENTARY NOTES Supplementary Note 1: Fabrication of Scanning Thermal Microscopy Probes

Outline Scanning Probe Microscope (SPM)

Optimal Design and Evaluation of Cantilever Probe for Multifrequency Atomic Force Microscopy

General concept and defining characteristics of AFM. Dina Kudasheva Advisor: Prof. Mary K. Cowman

SUPPLEMENTARY INFORMATION

Introductory guide to measuring the mechanical properties of nanoobjects/particles

Application Note #149 Improving the Accuracy of Nanomechanical Measurements with Force-Curve-Based AFM Techniques

Linker Dependent Bond Rupture Force Measurements in Single-Molecule Junctions

Supplementary Materials to Addressable Photo-Charging of Single Quantum Dots Assisted with Atomic Force Microscopy Probe

Oxford Scholarship Online

Graphene. Tianyu Ye November 30th, 2011

Combining High Resolution Optical and Scanning Probe Microscopy

SUPPORTING INFORMATION

Atomic Force Microscopy (AFM) Part I

Supporting Information s for

Electronic Supplementary Information

Scanning Tunneling Microscopy

Electrochemically Exfoliated Graphene as Solution-Processable, Highly-Conductive Electrodes for Organic Electronics

Wafer-scale fabrication of graphene

Characterization of Materials with a Combined AFM/Raman Microscope

Supplementary Figure 3. Transmission spectrum of Glass/ITO substrate.

Keywords- Focused Ion Beams, Nanostructuring, Polymers, Functionalization, Electron Microscopy, Surface Modification

TEOS characterization of 2D materials from graphene to TMDCs

Electrospun UV-responsive Supramolecular Nanofibers from Cyclodextrin-Azobenzene Inclusion Complex

Epinephrine Electro-oxidation Highlights Fast Electrochemistry at the Graphite Basal Surface

Diamond-like-carbon (DLC) master creation for use in soft lithography using the Atomic Force Microscope (AFM)

SUPPLEMENTARY INFORMATION

Supporting Information for

Empowering today s nanoscale research with Peak Force Tapping

Toward Clean Suspended CVD Graphene

Carbon nanofibres from fructose using a light-driven hightemperature

Transparent Electrode Applications

The Controlled Evolution of a Polymer Single Crystal

Carbon Quantum Dots/NiFe Layered Double Hydroxide. Composite as High Efficient Electrocatalyst for Water

Crystalline Surfaces for Laser Metrology

Surface and Interface Analysis. Investigations of Molecular Depth Profiling with Dual Beam Sputtering. Journal: Surface and Interface Analysis

Nanoscale work function measurements by Scanning Tunneling Spectroscopy

SUPPLEMENTARY INFORMATION

Supramolecular electrospun nanofibers with high conductivity at. ultra-low carbon nanotube content

A. Optimizing the growth conditions of large-scale graphene films

AC Electrothermal Characterization of Doped-Si Heated Microcantilevers Using Frequency-Domain Finite Element Analysis

Supplementary Figure 1 a) Scheme of microfluidic device fabrication by photo and soft lithography,

A SCIENTIFIC APPROACH TO A STICKY PROBLEM

Supplementary Information

Room temperature synthesis of GaN driven by kinetic. energy be-yond the limit of thermodynamics

Supporting Information. Effect of Backbone Chemistry on the Structure of Polyurea Films Deposited by Molecular Layer Deposition

Scanning Probe Microscopy. L. J. Heyderman

Supplementary Information for. Effect of Ag nanoparticle concentration on the electrical and

SUPPLEMENTARY INFORMATION

Quantum Electrodynamics in Graphene

Precision Cutting and Patterning of Graphene with Helium Ions. 1.School of Engineering and Applied Sciences, Harvard University, Cambridge MA 02138

BioAFM spectroscopy for mapping of Young s modulus of living cells

Controlled Electroless Deposition of Nanostructured Precious Metal Films on Germanium Surfaces

Measurements of interaction forces in (biological) model systems

Figure S1 TEM image of nanoparticles, showing the hexagonal shape of the particles.

Graphene films on silicon carbide (SiC) wafers supplied by Nitride Crystals, Inc.

Nanoholes for leak metrology

Characterization of MEMS Devices

Two-dimensional homologous perovskites as light absorbing materials for solar cell applications

Graphene Fundamentals and Emergent Applications

Other SPM Techniques. Scanning Probe Microscopy HT10

The most versatile AFM platform for your nanoscale microscopy needs

Contents. What is AFM? History Basic principles and devices Operating modes Application areas Advantages and disadvantages

Transcription:

Accurate thickness measurement of graphene Cameron J Shearer *, Ashley D Slattery, Andrew J Stapleton, Joseph G Shapter and Christopher T Gibson * Centre for NanoScale Science and Technology, School of Chemical and Physical Sciences, Flinders University, Bedford Park, South Australia, 5042 Email: cameron.shearer@flinders.edu.au; christopher.gibson@flinders.edu.au Materials and methods Graphene samples were prepared from the mechanical exfoliation of graphite. Briefly, highly ordered pyrolytic graphite (HOPG, 1 x 1 x 0.1 cm, ZYH, Grade 2, NT-MDT) was pressed firmly onto adhesive tape (SCOTCH TM MAGIC TM Tape, 3M) and then plastic tweezers were pressed along the tape to ensure complete contact to the HOPG. The tape was then peeled slowly off the HOPG over 20 s. The now cleaved graphite on the adhesive tape was again pressed onto a clean piece of adhesive tape and pressed firmly with plastic tweezers followed by slow removal. This process was repeated 5 times until the graphite on the tape was partially transparent. The thin graphite film on adhesive tape was then pressed onto a silicon piece with 100 nm thermal oxide (CZ, 1 20 Ωcm, ABC GmbH, Germany) and peeled away slowly over 20 s transferring graphite onto the surface of the silicon. The SLG sections/areas on the Si wafer were then identified using the optical microscope on a confocal Raman spectrophotometer (Witec Alpha 300RS 100x, 0.9 numerical aperture, working distance 0.23 mm) to find areas of very thin graphite and then mapping the area by Raman spectral imaging with 532 nm laser excitation (with tuneable power up to 30 mw). Raman single spectra were acquired with integration times of 10 s and 3 accumulations. Presented single spectra are normalized to the intensity of the G-band. Raman spectral images were obtained by collecting a series of 50 x 50

single spectra (1.5 s integration per spectrum) over an area of 15 x 15 μm. The plot of the 2D/G band ratio was generated by exporting the values for maximum intensity for the 2D and G bands after using background subtraction (Witec Project 2.1), then computing the value for 2D/G for each position and plotting as a 3D surface matrix (Origin Pro 9.0). To exclude artefacts from the background area (where both 2D and G intensity are ~0 a.u), a minimum intensity value was set for calculation of 125 a.u., which was determined by the noise level of the spectra. Atomic force microscopy was performed using a Bruker Dimension FastScan AFM with Nanoscope V controller, operating in PeakForce Tapping (PFT) mode with the humidity controlled, in the immediate vicinity of the instrument, to less than 35 % using dehumidifying units (CLI-MATE TM model DH2500E). Images were acquired using ScanAsyst-Air probes (silicon tips on silicon nitride lever, Bruker) at a scan rate of 1 Hz at a resolution of 512 pixels and 512 lines. The nominal spring constant of the cantilevers is 0.4 N m -1 but the true value was determined experimentally prior to each scan using established calibration techniques [1, 2]. The deflection sensitivity was determined by measuring the slope in the contact regime of the PeakForce force curves on the hard silicon surface using an automated software function in the Nanoscope control software. The value for PeakForce set point was set prior to capturing an image and calculated by the software using the measured deflection sensitivity and calibrated cantilever spring constant. Engage settings were adjusted manually to ensure the minimal engage force, this was achieved by adjusting the engage setpoint and gain (starting at 0.05 V which equates to approx. 1 nn) until an engage was achieved. This timely process (due to the high number of false engage events) was carried out to ensure the initial engage was as soft as possible to ensure the quality of the AFM probe. Values for feedback gain and Z-limit were adjusted manually for each image. During a scan topography, adhesion, deformation and hardness were all captured concurrently with the calibration of the height sensor verified by scanning silicon calibration grids (Bruker model numbers PG: 1 μm pitch, 110 nm depth and VGRP: 10 μm pitch, 180 nm depth). Presented AFM topography images have been flattened and analysed using either the depth or step analysis features of Nanoscope Analysis 1.4. Single walled carbon nanotube (SWCNT) modified AFM probes were prepared following a previously reported procedure [3]. Briefly, a SWCNT film (arc-discharge, AP quality, Carbon

solutions) was mounted onto a micromanipulator (model MM3A, kleindiek) and an AFM probe (PFTUNA, Pt coated silicon tips on silicon nitride lever, Bruker) was mounted onto the sample stage within a scanning electron microscope (SEM) (Helios, D344 Dualbeam, FEI). An overhanging CNT, from the SWCNT film, was brought into contact with the AFM probe tip. The overhanging CNT was then cut by first injecting water vapour (magnesium sulfate heptahydrate) into the system via a needle located 200 μm from the electron beam and then line imaging the area to be cut with the electron beam at an accelerating voltage of 1 kv for 5 s. The now shortened CNT was then strongly adhered to the AFM probe by introducing a platinum precursor (Trimethyl [(1, 2, 3, 4, 5-ETA)- 1 Methyl 2, 4 Cyclopentadien-1-YL] Platinum) and imaging the CNT area in contact with the AFM probe at an accelerating voltage of 1 kv to deposit ~1 micron of platinum. The following imaging routine was followed for all experiments. The SLG area was cleaned to remove any surface adsorbates by slowly scanning the area with the laser (532 nm) during Raman spectral imaging [4]. The sample was then transferred to the AFM with the humidity set below 35 % and directly imaged with a new AFM probe. After finding the same SLG area, the PeakForce set point was systematically adjusted in intervals with a new image obtained for each peak force set point until the apparent height of the SLG was measured to be ~0.4 nm. Supplemental results To ensure the observed trend of measuring lower SLG height with increasing peak Force set point, the experiment was completed multiple times. Figure S1 and Figure S2 show the results for two further experiments with different AFM probes. The same trend is observed but the rate of change is different, likely due to different tip geometry.

Figure S1. Repeat measurement of graphene height with ScanAsyst-air probe. PeakForce tapping mode AFM images of graphene showing the change in measured height with peak force set point. Below each AFM image is the corresponding histogram depth plot from which the graphene height was determined. Scale bar = 200 nm. Figure S2. Repeat measurement of graphene height with ScanAsyst-air probe. PeakForce tapping mode AFM images of graphene showing the change in measured height with peak force set point. Below each AFM image is the corresponding histogram depth plot from which the graphene height was determined. Scale bar = 200 nm. To show that the observed trend in measured height is not related to imaging artefacts several control experiments were completed. The first was to compare the change in adhesion to the graphene and substrate with peak Force set point. Figure S3 shows adhesion maps with depth histograms obtained concurrently with the PFT topography images shown in Figure 2. The graphene layer does show greater adhesion than the substrate but the difference in adhesion does not vary significantly between peak Force set points. From this, the effect of changing adhesion can be ruled as a possible cause of the varying measured height.

Figure S3. (a, b) PeakForce tapping mode AFM adhesion images of graphene showing the change in measured adhesion with peak Force set point (as indicated in each AFM image). Images were obtained concurrently with the height images shown in Figure 2. (c) Graph of measured adhesion vs peak Force set point. Scale bar = 200 nm. HOPG was measured along with MLG to ensure that an accurate value for graphene layer spacing could be obtained with PFT AFM. Figure S4 shows the results for HOPG where the measured value for a single step was measured to be 0.255 ± 0.155 nm, which is very close to the theoretical value of 0.335 nm.

Figure S4. PeakForce tapping mode AFM topography images of HOPG showing the change in measured height with peak force set point (as indicated in each AFM image). Images were obtained sequentially from 1 nn up to 10 nn. Below each AFM image is the corresponding depth histogram from which the average graphite step height was determined. Scale bar = 200 nm. A second CNT modified AFM probe was used to measure SLG height with varying peak force set point. Shown in Figure S5, the same trend was observed to that reported in the main manuscript, the force required to achieve an accurate value was much lower than for a standard AFM probe. Figure S5. Repeat experiment of PeakForce tapping mode AFM topography images of graphene with a CNT modified AFM probe showing the change in measured height with peak Force set point (as indicated in each AFM image). Images were obtained sequentially for each different Force set point. Below each AFM image is the corresponding histogram depth plot from which the graphene height was determined.. Scale bar = 200 nm.

Figure S6 shows the adhesion maps obtained concurrently with Figure 5 where there is a difference in adhesion between graphene and the substrate but the magnitude of the difference does not change with peak force set point. Figure S6. PeakForce tapping mode AFM adhesion images of graphene using SWCNT-modified tip showing the change in measured adhesion with peak Force set point (as indicated in each AFM image). Images were obtained concurrently with the height images shown in Figure 5. (b) Graph of measured adhesion vs peak Force set point. Scale bar = 200 nm. Figure S7 shows the control experiment to show that imaging multiple graphene steps with a SWCNT modified probe (the probe as used in Figure 5) obtains an accurate value for graphene step height, regardless of applied force. Figure S7. PeakForce tapping mode AFM topography images with a CNT modified AFM probe of multilayer graphene showing the change in measured height with peak force set point (as indicated in each AFM image). Images were obtained sequentially for each Force set point. Below each AFM

image is the corresponding step height plot from which the average graphene step height was determined. Scale bar = 200 nm, grey boxes indicate area where step-height was determined. References [1]. Slattery A D, Blanch A J, Quinton J S, and Gibson C T 2013 Calibration of atomic force microscope cantilevers using standard and inverted static methods assisted by fib-milled spatial markers Nanotechnology 24 015710 [2]. Sader J E, Sanelli J A, Adamson B D, Monty J P, Wei X, Crawford S A, Friend J R, Marusic I, Mulvaney P, and Bieske E J 2012 Spring constant calibration of atomic force microscope cantilevers of arbitrary shape Rev. Sci. Instrum. 83 103705 [3]. Slattery A D, Blanch A J, Quinton J S, and Gibson C T 2013 Efficient attachment of carbon nanotubes to conventional and high-frequency afm probes enhanced by electron beam processes Nanotechnology 24 235705 [4]. Hosoya N, Tanimura M, and Tachibana M 2013 Effect of laser irradiation on few-layer graphene in air probed by raman spectroscopy Trans. Mater. Res. Soc. Jpn. 38 579-583