ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Similar documents
Study of channeling effects in silicon detectors for pulse-shape applications

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Correction of the electric resistivity distribution of Si wafers using selective neutron transmutation doping (SNTD) in MARIA nuclear research reactor

Nuclear Instruments and Methods in Physics Research A

Status Report: Charge Cloud Explosion

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

FAZIA detector for isospin φsics: status of the R&D program

Modeling of charge collection efficiency degradation in semiconductor devices induced by MeV ion beam irradiation

Impact of high photon densities on AGIPD requirements

Nuclear Instruments and Methods in Physics Research A

Electron transparency, ion transparency and ion feedback of a 3M GEM

Impact of high photon densities on AGIPD requirements

Mara Bruzzi INFN and University of Florence, Italy and SCIPP, UC Santa Cruz, USA

CMS Note Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

RD50 Recent Results - Development of radiation hard sensors for SLHC

A fast triple GEM detector for high-rate charged-particle triggering

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

A case study in nuclear physics: the FAZIA project. Giacomo Poggi (for the FAZIA Collaboration)

A new protocol to evaluate the charge collection efficiency degradation in semiconductor devices induced by MeV ions

arxiv:physics/ v2 [physics.ins-det] 18 Jul 2000

The european FAZIA initiative: a high-performance digital telescope array for heavy-ion studies

Development of New MicroStrip Gas Chambers for X-ray Applications

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

Charge Collection and Capacitance-Voltage analysis in irradiated n-type magnetic Czochralski silicon detectors

Control of the fabrication process for the sensors of the CMS Silicon Strip Tracker. Anna Macchiolo. CMS Collaboration

arxiv:physics/ v2 [physics.ins-det] 22 Nov 2005

PRE-EQUILIBIRUM α-particle EMISSION AS A PROBE TO STUDY α-clustering IN NUCLEI

Heavy ion fusion energy program in Russia

Controlled Si-Drift Detectors

THE spectroscopic performance of large volume CdZnTe

M. De Napoli, F. Giacoppo, G. Raciti, E. Rapisarda, C. Sfienti. Laboratori Nazionali del Sud (LNS) INFN University of Catania. IPRD Oct.

Characterization of Irradiated Doping Profiles. Wolfgang Treberspurg, Thomas Bergauer, Marko Dragicevic, Manfred Krammer, Manfred Valentan

Nuclear Instruments and Methods in Physics Research A

Advances in the Micro-Hole & Strip Plate gaseous detector

Position sensitive detection of thermal neutrons with solid state detectors (Gd Si planar detectors)

Exploring the limiting timing resolution for large volume CZT detectors with waveform analysis

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

Recent results from the ISOFAZIA experiment

Review of Semiconductor Drift Detectors

Analysis and design of a new SRAM memory cell based on vertical lambda bipolar transistor

Nuclear Instruments and Methods in Physics Research A

Non-traditional methods of material properties and defect parameters measurement

Pixels GaAs Detectors for Digital Radiography. M.E. Fantacci. and. Abstract

Detectors in Nuclear and High Energy Physics. RHIG summer student meeting June 2014

Pulse-shape shape analysis with a Broad-energy. Ge-detector. Marik Schönert. MPI für f r Kernphysik Heidelberg

THE USE OF GE DETECTORS FOR (N,XN) CROSS-SECTION MEASUREMENTS AT INTENSE AND LOW FREQUENCY PULSED NEUTRON BEAMS. Abstract

PRE-EQUILIBRIUM α-particle EMISSION AS A PROBE TO STUDY α-clustering IN NUCLEI

X-ray induced radiation damage in segmented p + n silicon sensors

Boron-based semiconductor solids as thermal neutron detectors

Radiation hardness of Low Gain Amplification Detectors (LGAD)

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

The ALICE Inner Tracking System Off-line Software

Boron-Proton Nuclear-Fusion Enhancement Induced in Boron-Doped Silicon Targets by Low-Contrast Pulsed Laser

ORTEC. Review of the Physics of Semiconductor Detectors. Interaction of Ionizing Radiation with Semiconductor Detectors. Heavy Charged Particles

Modelling of Diamond Devices with TCAD Tools

First Results and Realization Status of a Proton Computed Radiography Device

BEGe Detector studies update

Proposed laser ion source for the Columbia University microbeam

Nuclear Instruments and Methods in Physics Research A

LOW TEMPERATURE PROPERTIES OF NTD GE: BEST CHOICE FOR CUORE EXPERIMENT

Method of active correlations in the experiment 249 Cf+ 48 Ca n

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

New solid state photomultiplier. Dmitry Shushakov and Vitaly Shubin

Measurement of high energy gamma rays with large volume LaBr 3 :Ce scintillators

GEM-based photon detector for RICH applications

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

Nuclear Instruments and Methods in Physics Research A

EEE4106Z Radiation Interactions & Detection

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors. Fabrication of semiconductor sensor

Charge Collection and Space Charge Distribution in Epitaxial Silicon Detectors after Neutron-Irradiation

Applied Radiation and Isotopes

The annealing of interstitial carbon atoms in high resistivity n-type silicon after proton irradiation

Development and characterization of 3D semiconductor X-rays detectors for medical imaging

Creep characteristics of piezoelectric actuators

Hard X- and g-ray measurements with a large volume coplanar grid CdZnTe detector

Progresses in FAZIA detection system and preliminary results from the ISO-FAZIA experiment

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

Status Report: Multi-Channel TCT

DESIGN OF A 10 NM ELECTRON COLLECTOR FOR A TRACK-

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Alpha-particle Stopping Powers in Air and Argon

Supplemental material for Bound electron nonlinearity beyond the ionization threshold

Wafer Charging in Process Equipment and its Relationship to GMR Heads Charging Damage

Review of the Shockley Ramo theorem and its application in semiconductor gamma-ray detectors

Preliminary measurements of charge collection and DLTS analysis of p + /n junction SiC detectors and simulations of Schottky diodes

Results from Silicon Photo-Multiplier neutron irradiation test

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH STUDIES OF THE RADIATION HARDNESS OF OXYGEN-ENRICHED SILICON DETECTORS

Lecture 2. Introduction to semiconductors Structures and characteristics in semiconductors

Radiation Effect Modeling

The Advanced Gamma Ray Tracking Array AGATA

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

A double junction model of irradiated silicon pixel sensors for LHC

Energy measurement and fragment identification using digital signals from partially depleted Si detectors

Pulse Formation on the Anode of Frisch-Gridded Ionization Chamber

Semiconductor Detectors

arxiv: v1 [physics.ins-det] 29 Jun 2011

1. Introduction. Etim I.P. 1, William E. S. 2, Ekwe S.O. 3. Licensed Under Creative Commons Attribution CC BY

High-Sensitivity Gamma-Ray Imaging With Double Sided Strip Detectors

Transcription:

Nuclear Instruments and Methods in Physics Research A 602 (2009) 501 505 Contents lists available at ScienceDirect Nuclear Instruments and Methods in Physics Research A journal homepage: www.elsevier.com/locate/nima A method for non-destructive resistivity mapping in silicon detectors L. Bardelli a,b,, G. Poggi a,b, G. Pasquali a,b, M. Bini a,b a University of Florence, Italy b I.N.F.N Sezione di Firenze, Italy For the FAZIA collaboration article info Article history: Received 12 December 2008 Accepted 13 January 2009 Available online 25 January 2009 Keywords: Silicon resistivity Solid state detectors Pulse-shape analysis Digital sampling abstract It is well known that the resistivity non-uniformity of silicon detectors is a crucial parameter when pulse-shape analysis is used to identify the charge and the mass of stopped heavy ions. In this work a method is described that allows a direct absolute resistivity measurement of the detector as a function of the position (mm resolution). The detector is used in a reverse-mount configuration and signals are collected for various applied voltages and for various ðx; yþ positions by using a point excitation. For each applied voltage-position combination, the average signal risetime is obtained via a digital pulse-shape analysis, finally allowing the extraction of the desired resistivity measurement as a function of the position. The method is non-destructive and can be applied to detectors with arbitrary shapes and readout geometries. Detectors can be fully tested in the laboratory before the actual experiments, possibly rejecting before beam time those not satisfying the resistivity uniformity requirements of the experiment. & 2009 Elsevier B.V. All rights reserved. 1. Introduction The use of pulse-shape analysis (PSA) methods for charged particle identification in silicon detectors is a promising technique for present and future Nuclear Physics experiments, under investigation by several international collaborations. Its main advantage with respect to the standard DE E technique is the lower energy threshold achievable for particle identification. It is well known (see for example Refs. [1 3]) that the PSA performances in silicon are severely influenced by the resistivity non-uniformity of the detector. In particular it has been shown that areas of the detector characterized by different resistivity produce signals having different shapes (e.g. different risetimes), thus jeopardizing the achievable performances when the full (i.e. non ad-hoc collimated) detector area is used. It is possible to roughly estimate the needed level of uniformity by considering existing experimental PSA data and comparing with, for example, an estimate of the effect of resistivity nonuniformity on the detector risetime. Using as an example the collimated NTD detector shown in Fig. 9 of Ref. [4], one sees that the pulse-shape-based isotopic separation between, e.g. 12 C and 13 C consists in a 0:2 ns difference in the two isotopes average risetimes. Given the 15 ns average value, this translates into a 1:3% difference. Therefore, since to a first-order approximation both the transit time and the plasma time (see for example the analytical estimate in Ref. [5]) depend linearly on the depletion voltage and hence on resistivity, a given requirement on signals risetime-spread directly translates into a non-uniformity requirement. The presently available quality of off-the-shelf silicon ingots and detectors (about 5 10%) is thus clearly not adequate and quality-check procedures should be employed. Moreover, despite the technological efforts of the semiconductor industry, presently it is quite difficult to measure (possibly in a non-destructive way) the non-uniformity of a detector at the percent level prior to its operation. It has also to be noted that this quantity may fluctuate widely from one detector batch to another, thus making a sampling-based test quite useless. In this work a non-destructive method for resistivity mapping is proposed that allows the determination of the detector resistivity map and thus the measurement of its non-uniformity. From the practical point of view, this allows testing of the detectors in the laboratory before the actual experiment, and rejection of those that do not match some predefined non-uniformity threshold. 2. The method Corresponding author at: I.N.F.N Sezione di Firenze, Italy. Tel.: +39 055 4572693. E-mail address: bardelli@fi.infn.it (L. Bardelli). Let us briefly recall a quite standard method for detector depletion-voltage (V depl ) or resistivity (r) determination, i.e. the 0168-9002/$ - see front matter & 2009 Elsevier B.V. All rights reserved. doi:10.1016/j.nima.2009.01.033

502 L. Bardelli et al. / Nuclear Instruments and Methods in Physics Research A 602 (2009) 501 505 (x, y) collimated light pulse y x measurement of the detector capacitance C as a function of the applied voltage (V appl ). Due to the semiconductor properties of the detector, the capacitance is very large for V appl 5V depl, it lowers as a function of the applied voltage as long as V appl ov depl, finally saturating for V appl 4V depl this behavior experimentally defines the quantity V depl. The detector resistivity can be finally obtained using the following formula [6], valid for a planar configuration and constant resistivity in the whole detector active volume: d 2 z r ¼ 2V depl R m rear contact silicon detector not depleted region z 0 front contact Vbias volume involved in charge collection Fig. 1. Reverse-mount silicon detector irradiated with a collimated light source. Only a small fraction of the active volume is involved in the charge collection process (for the sake of clarity the detector thickness has been greatly exaggerated in the figure). where d is the detector thickness, R is the silicon dielectric constant, m is the majority-carrier mobility in the detector bulk. It is worth noting that both V depl and r extracted in this way must be regarded as average quantities over the whole detector. In this work we first extend the basic ideas behind the standard C V appl method by defining the local depletion voltage V depl ðx; yþ, i.e. the detector depletion voltage as a function of the position ðx; yþ the discussed configuration is shown in Fig. 1. A collimated light pulse (or a collimated a particle source), having very low penetration depth inside silicon, produces a cloud of electrons and holes that are collected by the detector. 1 It has to be noted that, due to the collimation, only a small part of the detector is involved in this process, namely the volume defined by the source collimation in the ðx; yþ plane, by the lateral diffusion processes during charge collection, and by the detector thickness. As shown in Fig. 1 the detector is used in a reverse-mount configuration, i.e. the used light pulse enters the detector from the low field side this means that, when V appl ov depl, the carriers are produced in a region of nominally zero electric-field (that is the area marked as not depleted region in Fig. 1). For a given position ðx; yþ, by examining the collected signals as a function of the applied voltage V appl, we expect some abrupt change in the signal properties (for example the risetime) in passing from the regime of non-completely depleted detector to 1 For the method to work the dead-layer thickness of the relevant detector face (low field side) must be much smaller that the penetration depth of the used source (either laser pulse or alpha particle). depleted region (1) the over-depleted one (i.e. absence of any non-depleted region). In analogy with the standard C V appl method, taking into account that the signal is governed by the properties of the interested active volume only, we thus define the local depletion voltage V depl ðx; yþ as the voltage where this abrupt change occurs (later we will discuss how to practically determine it). For typical collimations (mm diameter) and detector thicknesses (hundreds of mm) one can approximate the involved active volume as a planar geometry, and thus, assuming that no appreciable change in the detector resistivity occurs along the z direction, it is possible to measure the desired absolute detector resistivity rðx; yþ as a function of position by application of Eq. (1). In the case of significantly varying resistivity along the z direction (not expected for thin detectors), the method can be still applied in order to find the average resistivity of the material along z as a function of ðx; yþ position. The proposed method thus somewhat resembles the well established Transient Current Techniques (TCT, [7]) but, at variance from those methods, in this work almost no hypothesis on the detector properties is made. In Fig. 2 the used experimental setup is shown. The silicon detector under test is mounted on a x y movement with remotecontrol, and irradiated with a collimated pulsed laser source (sub-nanosecond pulse width). It would be also possible to use a collimated a source, although the need to operate in vacuum makes this solution somewhat more complex to set-up. A collimator of 1 mm diameter has been used during the tests. The detector output is fed into a preamplifier [8] whose charge output is connected to a digital sampling system (employing a 125 MS/s 12 bit digitizer, see Ref. [9]). Both the x y movement and the applied voltage V appl are remotely controlled by the acquisition system. For each ðx; yþ position, a V appl scan is performed by the control system, while recording the detector pulse shapes the procedure is repeated until the whole detector surface has been explored. 2 After the desired scan parameters have been introduced, no manual intervention is needed to operate the system. By means of a digital PSA it is possible, for each event, to extract the signal risetime (the algorithms of Ref. [4] have been used). After a scan has been completed (for example about 2.5 h are needed for a 20 20 mm 2 detector, with 1 mm x y resolution, 12 different values for V appl, and 100 Hz laser pulse rate), it is possible to extract for each ðx; yþ position an average risetime vs. V appl plot an example of such plot is shown in Fig. 3. For V appl ov depl the measured signal risetime is very long (hundreds of ns), due to the presence of a non-depleted region in the detector (see Fig. 1). For V appl 4V depl the signal risetime reaches the nominal value for the used detector type (practically the detector transit time) and, when compared with the V appl ov depl regime, only a minimal dependence on V appl remains. The local depletion voltage can be extracted from a plot like Fig. 3 by fitting the experimental points with a function describing the detector s signals risetime as a function of the applied voltage. For V appl ov depl, the signal is composed of two components: (1) a slow component due to the diffusion-driven motion of carriers in the zero-electric-field volume of the detector (i.e. zoz 0 in Fig. 1), (2) a fast one due to carriers flying in the depleted area (i.e. z4z 0 in Fig. 1). 2 In practice, in order to avoid excessive voltage stress to the detector, the two scans are performed in the opposite order, i.e. for a given V appl the full detector is scanned in the ðx; yþ plane. In the used operating conditions, this makes it possible to change the applied voltage on the detector every 0.2 h instead of every few seconds.

L. Bardelli et al. / Nuclear Instruments and Methods in Physics Research A 602 (2009) 501 505 503 pulsed light source collimation control system si detector preamp XY movement digital sampling high voltage Fig. 2. Experimental setup: the detector is mounted on a x y movement and detector signals are collected by a digital sampling system. Detector position and applied voltage scans are remotely controlled by the acquisition/control system. Det.70965: scan results for x = 2.0 y = 7.0 800 700 Exp. data Av. risetime (ns) 600 500 400 300 200 V depl = 112.4 V Fit wih Eq. (2) 100 0 90 100 110 120 130 140 150 160 V appl (V) Fig. 4. Typical case of an absolute resistivity map obtained with the proposed method (a 0.5 mm scanning step has been used). Circular striations in the silicon resistivity are apparent. Fig. 3. Typical example of a V appl scan in a given ðx; yþ detector position. The solid curve represents the fit to the data using Eq. (3). to the signal fast component (those details are not considered here). We can estimate the slow component by using the following simple model: the charge density at t ¼ 0 is a zero width distribution at z ¼ 0, one carrier type (electrons for n-type detectors) is instantaneously collected at t ¼ 0 and does not provide any contribution to the signal, the charge density rðz; tþ of the remaining carriers (holes for n- type detectors) changes as a function of the time t as a Gaussian packet subject to diffusion: " # rðz; tþ ¼ 2Q 0 pffiffiffiffiffiffi exp z2 (2) s D 2p 2s 2 D p having standard deviation s D ¼ ffiffiffiffiffiffiffiffi 2Dt (D is the diffusion coefficient). Q 0 is the total charge released in the detector (zx0), a carrier entering the depleted zone (i.e. the detector volume where the electric field is non-zero z4z 0 in Fig. 1) starts drifting towards the electrode, thus giving a contribution Under these hypotheses the slow component of the charge signal is given by Z 1 z 0 Q slow ðtþ ¼ rðz; tþ dz ¼ Q 0 1 erf p z 0 2 ffiffiffiffiffi Dt The 10 90% risetime of the signal can thus be obtained by numerical solution of the equations Q slow ðt 90% Þ¼0:90 Q 0 and Q slow ðt 10% Þ¼0:10 Q 0, finally yielding the estimate t rise;slow /ðz 0 Þ 2 (valid for z 0 40 that means V appl ov depl ). The measured risetime can thus be approximated with the sum of t rise;slow and of the fast-component risetime, that is of the order of the transit time in the depleted detector, and practically not dependent on V appl (as compared with the slow component). The local depletion voltage can be extracted from a plot like that of Fig. 3 by fitting the data with the function ( t rise ðv appl Þ¼ t pffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi p 0 þ að V depl ðx; yþ ffiffiffiffiffiffiffiffiffiffiffi V appl t 0 Þ 2 for V appl ov depl ðx; yþ for V appl 4V depl ðx; yþ (3)

504 L. Bardelli et al. / Nuclear Instruments and Methods in Physics Research A 602 (2009) 501 505 Fig. 5. Reconstruction of the original wafer resistivity map by measuring the resulting detectors. The circular concentric structure of the striations is apparent. The resistivity color scale spawns 4% with respect to the wafer mean value. where t 0 and a represent the asymptotic risetime (for V appl bv depl ) and the coefficient of the behavior apparent for V appl ov depl, respectively. 3 In Fig. 3 the value of V depl ðx; yþ obtained from such a fit is also shown. Once V depl ðx; yþ has been extracted for all the scanned positions of the detector it is possible to apply Eq. (1) and obtain rðx; yþ in the next section some results obtained with this procedure will be shown. 3. Results By using the procedure outlined in the previous section it is possible to build a bidimensional resistivity map of the used detector in Fig. 4 a typical example is shown. The most striking feature of the shown resistivity landscape is the presence of circular structures, or striations (very similar structures are present, with varying importance, in all the tested detectors). This is in fair agreement with the well known striations (see for example Ref. [10]) present in silicon material due to the silicon ingot growing processes this has been verified by measuring several detectors 4 belonging to the same wafer. Knowing the mask employed during the manufacturing process and how the placement of the bare silicon chips with respect to the mounting frame relates to their original placement on the wafer, it is thus possible to reconstruct the original wafer resistivity map. The result of such a procedure is shown in Fig. 5. In the figure a partial resistivity map of three wafers has been reconstructed starting from the resistivity measurements of eight detectors black solid lines refer to the processing mask, while an additional circle (dashed line) has been drawn to guide the eye. The picture clearly confirms the origin of the striations, concentric with the original wafer center. In Table 1 the obtained results for some of the tested detectors are shown. In the first four columns the detector serial number, the nominal thickness, nominal depletion voltage and nominal resistivity are reported. The fifth column ( Meas. av. V depl ) reports the average measured depletion voltage, i.e. the average of 3 Eq. (3) assumes that for V appl 4V depl the fast signal component has no appreciable change in the measured risetime (as compared with the slow one) like, for example, when a slow preamplifier/electronics is used. In this case the asymptotic risetime is only due to the electronics, and does not depend on V appl (like in Fig. 3). We have verified that the formula (and hence the overall method) applies also in the case of an appreciable change in measured risetime, provided that the region V appl 4V depl is fitted with a voltage dependent formula like, for example, a linear dependence t 0 þ bðv appl V depl Þ. 4 Unfortunately we did not have access to all detectors produced starting from a given wafer but only to a few of them, as apparent from Fig. 5. Table 1 Collected results for some of the tested detectors. Det. Thickness (mm) Nominal V depl (V) Nominal resistivity (O cm) Meas. av. V depl (V) Meas. av. resistivity (O cm) 70 964 315 110 3100 121.4 2896.8 3.8 70 965 315 110 3100 116.4 3015.0 2.1 70 966 315 110 3100 121.3 2886.1 4.3 70 967 516 220 4200 229.4 4130.0 0.2 70 968 516 220 4200 230.1 4127.9 0.3 70 969 713 150 11900 163.0 11386.9 3.9 73 310 315 110 3100 113.9 3094.1 4.4 73 311 315 110 3100 110.3 3159.6 4.6 73 312 516 220 4200 230.5 4142.7 0.4 73 313 516 220 4200 230.2 4140.0 0.6 73 314 713 150 11900 163.5 11126.6 7.0 73 315 713 150 11900 168.2 10816.2 4.2 73 316 713 150 11900 170.8 10589.2 4.1 73 946 318 120 2900 112.3 3168.5 0.9 73 947 318 120 2900 109.7 3241.3 1.3 Meas. nonuniformity (%) In the first columns nominal values (i.e. the manufacturer s declared ones) for the detectors are given. The measured depletion voltage and resistivity after averaging over the detector surface are given as well. The detector non-uniformity (see text for the definition) is given in the last column. the measured V depl ðx; yþ, while in the sixth column ( Meas. av. resistivity ) the average of rðx; yþ is shown. In the last column the measured non-uniformity is reported (in percent), defined as the FWHM of the resistivity variation across the detector surface. In extracting the reported non-uniformity we decided to exclude a border of 1 mm from the detector edges, where a non-perfect alignment of the measurement setup may influence the results due to a non-correct irradiation of the very border of the detector. While silicon manufacturers usually quote the maximum minimum variation of the resistivity, we have preferred to use the FWHM since it is directly related to the experimental use of such detectors, when the full active area is used for particle detection (and identification). As a rule of thumb, we found that typically Dr max2min 3 Dr FHWM. As shown in Table 1, the manufacturer declared values are in fair agreement with the obtained results (for example a maximum of 10 V discrepancy on the depletion voltage is present). The striking feature is the last column, where wide variations of the measured non-uniformity are present, even between detectors having similar characteristics. All the tested detectors are Neutron Transmutation Doped (NTD), i.e. the ingot has been obtained by neutron irradiation of a high resistivity Float-Zone material. This type of material is known to provide, at least in principle, the best resistivity homogeneity available to date.

L. Bardelli et al. / Nuclear Instruments and Methods in Physics Research A 602 (2009) 501 505 505 4. Conclusions In this work a method for resistivity mapping of silicon detectors has been presented. By using a collimated pulsed light source and a reverse-mount configuration of the silicon detector, surface scans of the detector are performed for different applied voltages, recording the resulting charge signal with a digital sampling system. After extraction of the average signal risetime for each position and applied-voltage, it is possible to extract a local depletion voltage (as defined in Section 2) and to finally obtain the desired absolute resistivity measurement as a function of the position, i.e. rðx; yþ. No additional special hypothesis is needed on the detector resistivity behavior in order to apply the method. These measurements make it possible to experimentally characterize the non-uniformity of the used silicon detectors, quantified by the FWHM variation of the resistivity across the detector area. In Table 1 results for some of the tested detectors have been reported. This kind of non-destructive detector characterization is very important for the purpose of pulse-shape-based Nuclear Physics projects (like FAZIA [11]) where, as it is well known (see for example Refs. [1,2]), the detector non-uniformity plays a fundamental role in the attainable particle identification performances. The detectors tested in this work have been used by the FAZIA collaboration in a R&D test at Laboratori Nazionali di Legnaro (LNL) in December 2007, providing particle identification performances in qualitative agreement with the measured nonhomogeneities (i.e. good uniformity provides good particle identification and viceversa). The details and the results of that experiment will be reported in a forthcoming paper. Acknowledgments The authors would like to thank the staff of machine shop of the University of Florence for the realization of the mechanics we have used during these test. The authors would like also to thank L. Lavergne for useful discussions and E. Scarlini for technical assistance with the silicon detectors. We would like also to thank CANBERRA for providing many informations and feedback about the detectors employed in these measurements. References [1] G. Pausch, et al., IEEE Trans. Nucl. Sci. NS-44 (3) (1997). [2] F.Z. Henari, et al., Nucl. Instr. and Meth. A 288 (1990) 439. [3] H.A. Rijken, et al., IEEE Trans. Nucl. Sci. NS-40 (4 Part 1 2) (1993) 349. [4] L. Bardelli, et al., Nucl. Instr. and Meth. A 521 (2004) 480. [5] W. Seibt, et al., Nucl. Instr. and Meth. 113 (1973) 317. [6] G.F. Knoll, Radiation Detection and Measurement, third ed., Wiley, New York, 1999. [7] Several papers are available in the literature on TCT methods few of them are: V. Eremin, Nucl. Instr. and Meth. A 372 (1996) 388; D. Menichelli, Nucl. Instr. and Meth. A 476 (2002) 614; J. Fink, Nucl. Instr. and Meth. A 565 (2006) 227. [8] H. Hamrita, et al., Nucl. Instr. and Meth. A 531 (2004) 607. [9] G. Pasquali, et al., Nucl. Instr. and Meth. A 570 (2007) 126. [10] Topsil, Application Note NTD silicon for power electronics. [11] FAZIA project hhttp://fazia.in2p3.fri.