POSITIONING THE SPRING-8 MAGNETS WITH THE LASER TRACKER

Similar documents
LONG-TERM VARIATION OF THE MAGNET ALIGNMENT IN SPRING8 STORAGE RING

ALIGNMENT RESULT FOR MAGNET UNITS OF THE SPRING-8 STORAGE RING

SURVEY AND ALIGNMENT FOR THE SYNCHROTRON LIGHT SOURCE ELETTRA

Fine Alignment of the ATF Damping Ring

The Alignment of BEPCII RING

Survey & Alignment of Pohang Light Source

STATUS REPORT ON STORAGE RING REALIGNMENT AT SLRI

THE LASER TRACKER : A MAJOR TOOL FOR THE METROLOGY OF THE LHC

THERMALLY DRIVEN VERTICAL DISPLACEMENT OF IP QUADRUPOLE MAGNET

THE SPEAR3 UPGRADE PROJECT AT SLAC *

A NEW STANDARD METHOD FOR THE SURVEYING AND ALIGNMENT OF BEAMLINE FACILITIES AT GSI

BEPCII Pre-alignment and Hold Level Adjustment. Accelerator Center of IHEP Xiaolong Wang

MAINTENANCE OF THE GEOMETRY OF LEP IMPROVEMENT & OPTIMIZATION OF THE VERTICAL AND RADIAL SMOOTHING

INSTRUMENTATION AND SURVEY NETWORKS AT THE ESRF

THE GSI FUTURE PROJECT: AN INTERNATIONAL ACCELERATOR FACILITY FOR BEAMS OF IONS AND ANTIPROTONS

THE SPEAR3 UPGRADE PROJECT AT SLAC * Abstract

3-D METROLOGY APPLIED TO SUPERCONDUCTING DIPOLE MAGNETS FOR LHC

The Alignment Concept For The Fermilab Recycler Ring Using Permanent Magnets

Ultra-Low Emittance Storage Ring. David L. Rubin December 22, 2011

Interferometer for Squareness measurement

Status Report on CSR Survey and Alignment at IMP

AN ULTRA-HIGH RESOLUTION PULSED-WIRE MAGNET MEASUREMENT SYSTEM. Alex D Audney Thesis Defense Colorado State University

PRELIMINARY FULL SIZE TEST OF A SIMPLIFIED VERSION OF THE CMS LINK ALIGNMENT SYSTEM AT THE ISR HALL

Survey and Alignment of the Fermilab Electron Cooling System

Simulation of Optics Correction for ERLs. V. Sajaev, ANL

STATUS OF KEKB. Ryuhei Sugahara, Yasunobu Ohsawa High Energy Accelerator Research Organization (KEK) Oho 1-1, Tsukuba, Ibaraki, Japan

Status Report on the Survey and Alignment efforts at DESY

On-axis injection into small dynamic aperture

Feasibility Study of Short-Wavelength and High-Gain FELs in an Ultimate Storage Ring. Introduction Ultimate Storage Ring Analysis Simulation Summary

David Martin High Precision Beamline Alignment at the ESRF IWAA, Grenoble 3-7 October 2016

SPEAR3 Alignment Chapter For Nuclear Instrumentation and Methods in Physics Research, Section A June 25, 2004

Lattice Cell/Girder Assembly

Free electron lasers

Eigenfrequency Wire Alignment System for Magnet Fiducialization

The 10th International Workshop on Accelerator Alignment, KEK, Tsukuba, February 2008 ALIGNMENT OF S-LSR

Survey and Alignment at the Advanced Light Source

Status Report Relativistic Heavy Ion Collider

Long Path Industrial OCT High-precision Measurement and Refractive Index Estimation

ALIGNMENT OF CAVITIES AND MAGNETS AT J-PARC LINAC

Dimensional metrology and positioning operations: basics for a spatial layout analysis of measurement systems

Graduate Written Examination Spring 2014 Part I Thursday, January 16th, :00am to 1:00pm

Analysis of the influence of the Spherically Mounted Reflector on the measurements done with Laser Tracker

GEODETIC MEASUREMENTS FOR THE HERA PROTON RING, THE NEW EXPERIMENTS AND THE TESLA TEST FACILITY - A STATUS REPORT

Accelerator Physics. Tip World Scientific NEW JERSEY LONDON SINGAPORE BEIJING SHANGHAI HONG KONG TAIPEI BANGALORE. Second Edition. S. Y.

ACCURACY OF WALL STATION SURVEYS

1 BROOKHAVEN SCIENCE ASSOCIATES

Multiparameter optimization of an ERL. injector

HIRFL STATUS AND HIRFL-CSR PROJECT IN LANZHOU

Analysis of KEK-ATF Optics and Coupling Using Orbit Response Matrix Analysis 1

CESR CREATING NEW FORMS OF MATTER. Meet CESR in Action. To Increase Luminosity

Status Report on Survey and Alignment Efforts at DESY

Lattice Design and Performance for PEP-X Light Source

Optimization of the SIS100 Lattice and a Dedicated Collimation System for Ionisation Losses

On the possibility to create a prototype of laser system for space debris movement control on the basis of the 3-meter telescope.

First commissioning of the HLS-II storage ring*

Status Report on the Survey and Alignment Activities at Fermilab

Pupil matching of Zernike aberrations

CEPC partial double ring magnet error effects

REFERENCING COORDINATE SYSTEMS MAP PROJECTIONS GEOREFERENCING

2 Closed Orbit Distortions A dipole kick j at position j produces a closed orbit displacement u i at position i given by q i j u i = 2 sin Q cos(j i ;

STATUS REPORT ON SURVEY AND ALIGNMENT OF J-PARC AFTER THE EARTHQUAKE

TORSION OF SURFACE PLATE OF THE ACTIVE SUPPORT TABLE FOR THE ATF DAMPING RING

Rotational Kinetic Energy

Survey, Alignment and Beam Stability at the Advanced Light Source

NIST ELECTROSTATIC FORCE BALANCE EXPERIMENT

ALIGNMENT OF THE SPARC LINEAR ACCELERATOR. M. Esposito, M. Paris, F. Sgamma, S. Tomassini, M. Troiani, (INFN LNF) Abstract

METROLOGY AND SURVEYING AT CERN : RECENT NEWS AND FACTS

MAGNET INSTALLATION AND ALIGNMENT FOR THE FUJI TEST BEAM LINE AT KEKB

Alignment-Aspects for the HERA-Experiments ZEUS and H1

A Laser-Compass Magnet Probe for Field Measurements in Solenoids

MAGNETIC MEASUREMENTS OF QUADRUPOLE FOCUSING MAGNETS AT SLAC*

3.2.2 Magnets. The properties of the quadrupoles, sextupoles and correctors are listed in tables t322_b,_c and _d.

A PRELIMINARY ALIGNMENT PLAN FOR RIA AT MSU

Accelerator Design and Construction Progress of TPS Project

Enhanced Performance of the Advanced Light Source Through Periodicity Restoration of the Linear Lattice

Auto collimator. Introduction. Objectives: Apparatus: Theory:

Diagnostics at the MAX IV 3 GeV storage ring during commissioning. PPT-mall 2. Åke Andersson On behalf of the MAX IV team

Development of the concentric reticle Baum for optical surveying instruments

Machine Positioning Uncertainty with Laser Interferometer Feedback

1

Survey and Alignment of the J-PARC

ILC Beam Dynamics Studies Using PLACET

Magnet Alignment Sensitivities in ILC DR Configuration Study Lattices. Andy Wolski. US ILC DR Teleconference July 27, 2005

PERFORMANCE OF WIRE POSITION SENSORS IN A RADIATION ENVIRONMENT. CERN, Geneva, Switzerland

Measurements in Optics for Civil Engineers

HE-LHC Optics Development

Amplitude dependent shift of betatron oscillation center

Conceptual design of an accumulator ring for the Diamond II upgrade

ALIGNMENT OF INSERTION UNITS IN NSRL RING

Electricity & Magnetism Study Questions for the Spring 2018 Department Exam December 4, 2017

Update on Optics Modeling for the ATF Damping Ring at KEK Studies for low vertical emittance

IMPROVING BEAM QUALITY NEW TELESCOPE ALIGNMENT PROCEDURE

VIBRATION ANALYSIS AND REPAIR PROCESS FOR THE VENTILATION SYSTEM FOR SMOKE DRAIN IN THE THERMAL POWER PLANT

Laser Optics-II. ME 677: Laser Material Processing Instructor: Ramesh Singh 1

Experimental study of nonlinear laser-beam Thomson scattering

and another with a peak frequency ω 2

Alignment of the Dipole Magnets in the Cornell Electron-Positron Storage Ring

Knot undulator to generate linearly polarized photons with low on-axis power density. Abstract

HLS online monitoring during beam assisted girder re-alignment at the accelerator SLS (Swiss Light Source)

Polarised e ± at HERA

Transcription:

II/185 POSITIONING THE SPRING-8 MAGNETS WITH THE LASER TRACKER C.Zhang, S.Matsui, J.Ohnishi, X.Ouyan, Y.Sasaki, K.Hasegawa, and K.Tsumaki SPring-8, Akoh-gun, Hyogo, Japan 1. INTRODUCTION The SPring-8 storage ring has a circumference of 1436m with a Chasman-Green lattice of 48 cells. Each cell consists of three units of two bending magnets, ten quadrupoles and seven sextupoles. The storage ring of third-generation synchrotron radiation source is much sensitive to magnet misalignment, especially quadrupole misalignment. To reduce this sensitivity, the magnet alignment for the Spring-8 storage ring is divided into two stages: magnets are aligned within unit of about 5m with RMS error less than +50pm, and the units are aligned along the storage ring with relative accuracy of ±0.2mm. The two-stage alignment method will reduce the amplitudes of the orbit distortions induced by the quadrupole misalignment [1]. The tasks face to us are: 1) For sake of Mihara-Kuriyama hill, It is impossible for us to set up a monument at the center of the storage ring, the central monument usually plays a very important role in controlling the absolute position of the beam elements. We have to find a solution that make the magnet alignment has not only a high relative precision but also a small displacement accumulation rate along the 1.4 km circumference. Two commonly used alignment networks of the quadrilateral and the triangulation are investigated. They are both have about 0.1mm or more relative error in controlling magnet positions, and the deviation accumulated along the ring is also enormous. 2) The storage ring installation is divided into five sections: in the first section four cell magnets are installed, second section eight cells, third section eight cells and so on. All the units are recommended to be put to the positions in one step before executing the overall survey and smoothing of the ring. This unclosed network is easily biased by systematic errors such as the two end datum points, atmospheric condition, calibration etc. Alignment of Spring-8 storage ring magnets is composed of monument survey and magnet survey and alignment. 10 monuments are set outside the storage ring, 88 monuments are buried at the intersection points of the both side straight lines of the bending magnet. The monuments are used as the references for individual unit installation, also are used as the guideposts for magnet alignment and path smoothing. The magnets are surveyed and aligned with the laser tracker Smart 310, a 3-dimension measurement system. This system is for the first time employed in the measurement of storage ring networks. The laser tracker SMART 310 is a dynamic measurement system which consists of a laser interferometer, a rotating mirror on two axes with two angle encoders and servo motors, a position diode etc.(fig.1). Returning beam from retroreflector is partly divided to the position diode. Through a servo loop the offset on the diode corrects the angles of the mirror to keep the beam on the target. The tracker gives 3D spherical coordinates of a target in space with a distance resolution of 1.26pm, an angular resolution of 0.7. Many factors may influence measurement accuracy of the laser tracker: the calibration, atmospheric conditions, and the way laser tracker been used, To realize the accuracy it can achieve, the laser tracker was checked under various Fig.1 Principle of the laser tracker condition such as different measurement angles

II/186 and encoder positions, as well as two faces of its measuring head. Also it is compared with HP 5527A interferometer. Experiments show the laser tracker has a distance accuracy of 0.001+0.2ppmxL(mm) and an angular accuracy of about 10 I..trad, which is four times the error of theodolite T3000. It is evident that using Smart for network survey, the precision is mainly restricted by the angular resolutions of the encoders. This accuracy will result at least 7mm traverse unclose for the storage ring orbit if the laser tracker is simply used for coordinate measurement. Magnet alignment network for the Spring-8 storage ring is a distance-only trilateral network. The laser tracker has different accuracy for distance measurement when its position changes with respect to measuring targets. To reduce the influence of angular error, the laser tracker s positions within network are chosen by checking the distance accuracy it results(fig.2): Fig.2 Smart distance measurement between two points where I is the length between two measuring points, a,zl, Z2 are angle and lengths from laser tracker to these points, ma, ml], ml2 are their measurement accuracy respectively. The laser tracker gives the most precise distance when it is placed at the extension line of two measuring points, where it can eliminate the calibration error of absolute distance and reflector eccentricity. The distance from the mirror to the target is also precise on the condition that the laser tracker is well calibrated. Because this distance is measured directly by the interferometer. 2. MONUMENT SURVEY In January to February, 1993, 28 monuments around the storage ring were surveyed with an error of 0.5mm using mekometer ME5000 and theodolite T3000(Fig.3). In the some time the building construction began. After that, second time monument survey increased monument number in the tunnel. These monuments are lately used as monitors for the monument survey inside the tunnel after building construction. After building construction all 88 monuments in the tunnel are surveyed with the Smart 310(Fig.4). Survey network includes 192 points and 480 distances. Because the tunnel is very narrow, if simply use the quadrilateral network, monument deviation is proved to be large: error ellipse of maximum deviation of 0.8mm in radial direction. For the following reasons the deviation increases: parts of the tunnel has completed while 3/4 parts are being constructed. Air temperature and refraction are quite different along the ring; After building construction the monument datums have about 0.8mm movements. Systematic error may increase deviation of survey to 1.5mm in radial direction. To reduce deviation the solutions are: 1) using T3000 measure 24 angles in every two cells, 2) from previous monument datum chose two fixed points which are believed to have least movements, 3) monument coordinates are adjusted by best-fitting to selected 16 monuments. These monuments are convinced to have little movement from three time surveys. Simulation study shows after taking above steps the monument survey will reduce the RMS error to 0.5mm, peak to peak error ±1.3mm in radial direction. Because using Smart 310 in survey causes only 0.3mm deviation in the beam direction, what we much concerned is the displacement in radial direction for the long and narrow tunnel. The survey results are compared with theodolite

II/187 T3000 angle measurement results. Difference between them is small. Peak to peak value of 0.7-0.8 is convinced(fig.5). 3. SURVEY AND ALIGNMENT OF MAGNET UNITS Alignment network for the storage ring magnet units is composed of 288 quadrupole fiducial points, 88 monuments and 96 auxiliary brackets on inner wall. Total distance observation is 1440 in the ring(fig.6). Several aspects are particular in this magnet alignment network. Quadrupoles and the laser tracker positions directly compose the knots of network. It make magnet positioning more directly, error translation of the reference points smaller. Among distance observations, over 50 percent distances are measured directly by the laser tracker interferometer, Accuracy of these distances is 4pm in 10 m range. The network precision depend on both the accuracy of laser tracker and the ratio of measurement length to the width of the net. Measurement lengths are optimized by simulation study of error accumulation. The study shows that measurement length shorter than 15 m, the ratio of measurement length to width of 5:1, has least error accumulation rate along the ring. Distance longer than one cell is no beneficial to control radial deviation of the magnets. Error ellipse analysis shows this network has a sufficient precision for controlling magnet displacement: Maximum magnet displacement of less than ±1mm is expected with respect to geodetic coordinate. The relative displacement between adjacent units is expected to be ±0.04mm(Fig.6). Up to now over 60 percent of magnet units are surveyed and aligned. Survey is carried out three rounds: two rounds for adjustments, one round for check purpose. Cell 04 to cell 25 gives an example of magnet alignment process(fig.7). In the first round, the magnets are adjusted to a smooth curve of polynomial fitting to the monuments. Monuments are the guideposts to control absolute positions. Magnets in the second round are adjusted to a local weighted curve which fits to magnets only. After two round of adjustment, the relative displacement between units achieves a precision ofi 0.04mm of RMS value. Four 50-meter long straight sections provide good spots to examine the accuracy of magnet positioning by using Taylor-hobson telescope to examine the straightness. This result shows magnet deviations are less than 0.1mm within 40 meters. Fig.8 gives the straightness of short straight sections. In short straight sections four magnets in 14 meters are measured by the telescope. The result well convinces the relative precision of magnet alignment is better than ±0.04mm.

II/188 Precise calibration is essential to the measurement of the Smart 310. Smart s absolute distance error includes home point distance error, interferometer error, and reflector eccentricity. To calibrate it precisely, a 20-meter calibration stand is made and set up in the tunnel. Calibration method can be referred to Fig.2. On the stand HP 5527A interferometer is used to provide a standard length. the HP reflector and the Smart reflector are in the same moving stage. Precise leveler ensures the two laser reflector moving the same distance(fig.9). To eliminate angle measurement error the Smart is set up most nearly to the stand. Smart measures the two point distances II, Zz and angle a, Distance correction value is deduced by the cosine law, referring the HP interferometer length. Through this process both the home point distance error and reflector eccentricity error are corrected. Distance measurement accuracy of 3pm is obtained in 10 meters. 4. CONCLUSION Third-generation synchrotron radiation source is very sensitive to magnet misalignment. To reduce this sensitivity magnet alignment for the SPring-8 storage ring adopts two-stage alignment method, and the relative displacement is more concerned. Although the Smart 310 has a large angle measurement error which is not match for its distance accuracy, it is also can be successfully used in network survey for positioning magnets as long as take its strong point and avoid its shortcoming. The magnet alignment results are gratifying and reliable. We can adjust the magnet to wherever we want to a lo range of 0.04mm. ACKNOWLEDGMENTS The authors would like to give thanks to Dr. A.Ando for his helpful advice and Mr. Nakashima, Mr. M.Kawakami, and Ms. Hu for their useful work. REFERENCES [1] H. Tanaka, N. Kumagai, and K. Tsumaki, Nucl. Instr. and Meth. in Phys. Res. A313 (1992) 529-545.

II/189

II/190

II/191

II/192

II/193