DR/4000 PROCEDURE. HARDNESS, Total
|
|
- Magdalene Hawkins
- 5 years ago
- Views:
Transcription
1 Method 8374 DR/4000 PROCEDURE Calcium and Magnesium; Chlorophosphonazo Colorimetric Method ULR (0 to 1,000 µg/l Ca & Mg as CaCO 3 ) Scope and Application: For ultra pure water. The estimated detection limit for program number 2000 is 4 µg/l. HACH PROGRAM HACH PROGRAM: 2000 Hardness, Tot. ULR. ENTER 1. Press the soft key under HACH PROGRAM. Select the stored program for ultra low range hardness by pressing 2000 with the numeric keys. Press: ENTER Note: If samples cannot be analyzed immediately, see Sample Collection, Storage and Preservationfollowing these steps. Note: The Flow Cell and Sipper Modules cannot be used for this procedure. Plastic sample cells must be used. 2. The display will show: HACH PROGRAM: 2000 Hardness, Tot. ULR The wavelength (λ), 669 nm, is automatically selected. Note: : For best results, determine a reagent blank for each new lot of reagent as follows. Prepare a reagent blank by repeating steps 3 through 11, using deionized water as the sample. The value displayed after Step 11 will be the reagent blank value. Correct for the reagent blank by pressing the soft keys under OPTIONS, (MORE), and then BLANK: OFF. Enter the reagent blank value and press ENTER Repeat for each new lot of reagent. 3. Rinse a plastic sample cell and the cap three times with the water to be tested. Do not allow the underside of the cap to come in contact with surfaces that may contaminate it. Note: Plastic sample cells must be used. Glass will contaminate the sample. 4. Fill the plastic sample cell to the 25-mL mark with sample. HardnessTot_PP_ULow_CPZ_Eng_4000.fm Page 1 of 6
2 , continued ZERO 5. Add the contents of one Chlorophosphonazo Solution Pillow to the sample cell. Note: A small amount of solution may remain in the pillow. This will not affect results. Note: One ml of Chlorophosphonazo Solution (Cat. No ) may be used instead of the solution pillow. 6. Cap the cell and swirl to mix. 7. Place the sample cell into the cell holder. Close the light shield. 8. Press the soft key under ZERO. The display will show: 0 µg/l CaCO 3 Note: If you are using a reagent blank correction, the display will show the correction. Note: For alternate concentration units, press the soft key under OPTIONS. Then press the soft key under UNITS to scroll through the available options. Press ENTER to return to the read screen. 9. Remove the cell from the instrument. Add one drop of CDTA Reagent for Ultra Low Range Hardness. Note: Complete steps within 1-2 minutes. 10. Cap the cell and swirl to mix. 11. Place the sample cell into the cell holder. Close the light shield. The result in µg/l as CaCO 3 (or chosen units) will be displayed. Note: The results can be expressed as µg/l calcium (Ca 2+ ) or magnesium (Mg 2+ ). Press the soft key under OPTIONS, then FORM: to scroll through the available options. Page 2 of 6 HardnessTot_PP_ULow_CPZ_Eng_4000.fm
3 , continued Interferences Interference studies were conducted at various hardness levels between 0 and 500 µg/l as CaCO 3. Various cations and anions were evaluated at levels in the range appropriate for ultra pure water applications. An ion is said to interfere when the resulting concentration is changed by ±10%. Table 1 Interfering Substances and Suggested Treatments Interfering Substance Aluminum Ammonium Copper Formaldehyde Nitrate Potassium Silicon Sodium Interference Levels and Treatments Negative interference above 150 µg/l No interference at or below 1000 µg/l Positive interference above 250 µg/l No interference at or below 47,000 µg/l Positive interference above 250 µg/l No interference at or below 1,000 µg/l Positive interference above 1,000 µg/l Negative interference above 79,000 µg/l Sample Collection, Storage and Preservation Do not use glass containers. Collect samples in clean plastic containers, preferably with screw-type closures. Rinse containers several times with the water to be analyzed before collecting final sample. Seal to avoid contamination during transport. Analyze as soon as possible. Accuracy Check Standard Additions Method a. Leave the unspiked sample in the sample cell compartment. Verify that the units displayed are in µg/l. Select standard additions mode by pressing the soft keys under OPTIONS, (MORE) and then STD ADD. b. Press ENTER to accept the default sample volume (ml), 25. c. Press ENTER to accept the default standard concentration (µg/l), 20,000. d. Press the soft key under ENTRY DONE. e. Obtain a Calcium Chloride Standard Solution, 20,000-µg/L as CaCO 3. f. Use the TenSette Pipet to add 0.1 ml, 0.2 ml and 0.3 ml of standard, respectively to three 25-mL samples and mix each thoroughly. g. Analyze each standard addition sample as described above. Accept the standard additions reading by pressing the soft key under READ each time. Each addition should reflect approximately 100% recovery. h. After completing the sequence, the display will show the extrapolated concentration value and the best-fit line through the standard additions data points, accounting for matrix interferences. i. See Section Standard Additions for more information. Standard Solution Method Use the Calcium Chloride Standard Solution 500-µg/L (as CaCO 3 ) listed under OPTIONAL REAGENTS AND STANDARDS. Perform the total hardness procedure as described above. HardnessTot_PP_ULow_CPZ_Eng_4000.fm Page 3 of 6
4 , continued Method Performance To adjust the calibration curve using the reading obtained with the 500-µg/L Calcium Chloride Standard Solution, press the soft keys under OPTIONS, (MORE) then STD: OFF. Press ENTER to accept the displayed concentration, the value of which depends on the selected units. If an alternate concentration is used, enter the actual concentration and press ENTER to return to the read screen. See Section Adjusting the Standard Curve for more information. Precision Standard: 100 µg/l Program For more information on determining precision data and method detection limits, refer to Section 1.5. Estimated Detection Limit For more information on derivation and use of Hach s estimated detection limit, see Section To determine a method detection limit (MDL) as defined by the 40 CFR part 136, appendix B, see Section Sensitivity Program Number: % Confidence Limits µg/l Program EDL µg/l Portion of Curve Abs Concentration Abs µg/l 500 µg/l µg/l 900 µg/l µg/l See Section Sensitivity Explained for more information. Summary of Method Calcium and magnesium combine equivalently with the chlorophosphonazo III indicator to form a complex which absorbs light very strongly at 669 nm. One drop of the CDTA reagent breaks up this complex, and the resultant decrease in absorbance is related to the amount of calcium and magnesium in the sample (as CaCO 3 ). Safety Good safety habits and laboratory techniques should be used throughout the procedure. Consult the Material Safety Data Sheet for information specific to the reagents used. For additional information, refer to Section 1. Pollution Prevention and Waste Management For information on pollution prevention and waste management, refer to Section 1. Page 4 of 6 HardnessTot_PP_ULow_CPZ_Eng_4000.fm
5 , continued REQUIRED REAGENTS AND STANDARDS Description Cat. No. ULR Hardness Reagent Set (100 tests) Includes: (1) , (1) , (1) , (1) ULR Hardness Reagent Set (500 tests) Includes; (1) , (2) , (1) , (1) Quantity Required Description per test Unit Cat. No. Chlorophosphonazo Indicator Solution Pillows... 1 pillow /pkg CDTA Solution... 1 drop ml SCDB REQUIRED EQUIPMENT AND SUPPLIES Clippers (Shears) for opening solution pillows each DR/ Inch Cell Adapter each Sample Cell, 1-inch, polystyrene w/ cap /pkg OPTIONAL REAGENTS AND STANDARDS Calcium Chloride Standard Solution, 20,000-µg/L as CaCO ml Calcium Chloride Standard Solution, 500-µg/L as CaCO ml Chlorophosphonazo Indicator Solution ml OPTIONAL EQUIPMENT AND SUPPLIES Dispenser, 1.0-mL, Repipet Jr... each DR/4000 Carousel Module Kit... each Pipet, TenSette, 0.1 to 1.0 ml... each Pipet Tips, for TenSette Pipet... 50/pkg HardnessTot_PP_ULow_CPZ_Eng_4000.fm Page 5 of 6
6 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web HACH COMPANY WORLD HEADQUARTERS Telephone: (970) FAX: (970) Hach Company, All rights reserved. Printed in the U.S.A. 6/03 11ed Page 6 of 6 HardnessTot_PP_ULow_CPZ_Eng_4000.fm
Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell
Hardness DOC316.53.01044 Chlorphosphonazo Colorimetric Method Method 8374 8 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow Scope and application: For boiler and ultrapure water. Test preparation
More informationDR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.
Method 8337 Powder Pillows DR/4000 TT PROCEDURE QUATERNARY AMMONIUM COMPOUNDS Direct Binary Complex Method (0 to 5.00 mg/l as CTAB) Scope and Application: For cooling tower and pool/spa water. The estimated
More informationDR/4000 PROCEDURE. IRON, Total
Method 8365 Powder Pillows DR/4000 PROCEDURE FerroMo Method* (0 to 1.800 mg/l) Scope and Application: For cooling water containing molybdate-based treatment; digestion is required for determining total
More information1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )
DR/4000 PROCEDURE Method 8023 1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to 0.700 mg/l Cr 6+ ) Scope and Application: For water and wastewater;usepa accepted for reporting for
More informationDR/4000 PROCEDURE NITRATE. 2. The display will show:
Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number
More informationDR/4000 PROCEDURE NITRATE. Using Powder Pillows
DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit
More informationDR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.
Method 8033 DR/4000 PROCEDURE Dithizone Method* (0 to 300 µg/l) Scope and Application: For water and wastewater; USEPA accepted for reporting for wastewater analysis (digestion is required).** See Section
More informationTetraphenylborate Method Method to 7.0 mg/l K Powder Pillows
Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information
More informationInstrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.
Barium DOC316.53.01315 Turbidimetric Method 1 Method 10251 2 to 100, 20 to 1000, 200 to 10,000 mg/l Ba (spectrophotometers) Powder Pillows 2 to 80, 20 to 800, 200 to 8000 mg/l Ba (colorimeters) Scope and
More informationCadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows
Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationChromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials
Nitrate, HR DOC316.53.01068 Chromotropic Acid Method Method 10020 0.2 to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information
More informationQuaternary Ammonium Compounds
, 8337 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB Powder Pillows Scope and Application: For cooling tower water and pool/spa water Test
More informationNitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows
FILL LINE Method 8155 Salicylate Method 1 Powder Pillows Scope and Application: For water, wastewater, and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) (0.01 to 0.50 mg/l NH 3 N) Test Preparation
More informationCadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)
Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater
More informationPowder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)
Silica DOC316.53.01132 Heteropoly Blue Method 1 Method 8186 0.010 to 1.600 mg/l SiO 2 (LR, spectrophotometers) Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters) Scope and application: For boiler
More informationDR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).
Method 8194 DR/4000 PROCEDURE Diaminobenzidine Method* (0 to 1.000 mg/l) Scope and Application: For water and wastewater; distillation is required for determining total selenium. See the Distillation procedure
More informationCadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
More informationChromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)
Chromium, Total DOC316.53.01034 Alkaline Hypobromite Oxidation Method 1 Method 8024 0.01 to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Scope and application: For water and wastewater.
More informationColorimetric Method Method to 0.70 mg/l Ag Powder Pillows
Silver DOC316.53.01134 Colorimetric Method Method 8120 0.02 to 0.70 mg/l Ag Powder Pillows Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
More informationPersulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials
Nitrogen, Total DOC316.53.01085 Persulfate Digestion Method Method 10072 2 to 150 mg/l N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information
More informationPersulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials
Nitrogen, Total DOC316.53.01086 Persulfate Digestion Method Method 10071 0.5 to 25.0 mg/l N (LR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific
More informationMercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution
Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 0.1 to 25.0 mg/l Cl Reagent Solution Scope and application: For water and wastewater. Test preparation Instrument-specific information Table
More informationInstrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.
Aluminum DOC316.53.01003 Eriochrome Cyanine R Method 1 Method 8326 0.006 to 0.250 mg/l Al 3+ Powder Pillows Scope and application: For water. 1 Adapted from Standard Methods for the Examination of Water
More informationMethod to 0.50 mg/l NH 3 N Powder Pillows
, 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows
More informationPowder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)
Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.
More informationMethod to 0.50 mg/l NH 3 N Powder Pillows
, 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows
More informationPersulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827
Nitrogen, Total DOC316.53.01088 Persulfate Digestion Method Method 10208 5 to 40 mg/l N (HR) TNTplus 827 Scope and application: For water and wastewater. Test preparation Instrument-specific information
More informationMethod 8326 (0.006 to mg/l Al 3+ ) Powder Pillows
, 8326 Eriochrome Cyanine R Method 1 Scope and Application: For water 1 Adapted from Standard Methods for the Examination of Water and Wastewater. DOC316.53.01003 Method 8326 (0.006 to 0.250 mg/l Al 3+
More informationPhosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation
Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,
More informationCadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3
, MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation
More informationCadion Method Method to 0.30 mg/l Cd TNTplus 852
Cadmium DOC316.53.01013 Cadion Method Method 10217 0.02 to 0.30 mg/l Cd TNTplus 852 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows
More informationNitrogen, Total Inorganic
Nitrogen, Total Inorganic DOC316.53.01090 Titanium Trichloride Reduction Method Method 10021 0.2 to 25.0 mg/l N Test N Tube Vials Scope and application: For water, wastewater and seawater. Test preparation
More informationChloramine (Mono) and Nitrogen, Free Ammonia
Chloramine (Mono) and Nitrogen, Free Ammonia DOC316.53.01016 Indophenol Method 1 Method 10200 0.04 to 4.50 mg/l Cl 2 0.01 to 0.50 mg/l NH 3 N Powder Pillows Scope and application: For the determination
More informationScope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.
Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for
More informationPAR Method Method to 2.0 mg/l Pb TNTplus 850
Lead DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows all
More informationSalicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832
, TNTplus 832, 10205 DOC316.53.01083 Salicylate Method Method 10205 HR (2 to 47 mg/l NH 3 N) TNTplus 832 Scope and Application: For surface waters, municipal and industrial wastewaters. Test preparation
More informationPAR Method Method to 2.0 mg/l Pb TNTplus 850
, TNTplus 850, 10216 DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and Application: For wastewater and process control Test preparation Before starting the test: FOR DR 2800:
More informationDimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO
Nitrate DOC316.53.01070 Dimethylphenol Method Method 10206 0.23 to 13. NO 3 N or 1.00 to 60.00 mg/l NO 3 (LR) TNTplus 835 Scope and application: For wastewater, drinking water, surface water and process
More informationInstrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.
Nitrite DOC316.53.01074 USEPA Diazotization Method 1 Method 8507 0.002 to 0.300 mg/l NO 2 N (LR, spectrophotometers) 0.005 to 0.350 mg/l NO 2 N (LR, colorimeters) Scope and application: For water, wastewater
More informationChromazurol S Method Method to 0.50 mg/l Al TNTplus 848
Aluminum DOC316.53.01004 Chromazurol S Method Method 10215 0.02 to 0.50 mg/l Al TNTplus 848 Scope and application: For drinking water, surface water, swimming pool water, wastewater and process analysis.
More informationCHLORINE, TOTAL (0 to 4.00 mg/l)
CHLORINE, TOTAL (0 to 4.00 mg/l) DOC316.53.01261 For water, wastewater, and seawater Method 10250 DPD Method Powder Pillows USEPA accepted for reporting water and wastewater analyses * Note: This product
More informationOxygen Demand, Chemical
Oxygen Demand, Chemical DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 1 to 60 mg/l COD (ULR) TNTplus 820 Scope and application: For wastewater, process water, surface water, and cooling water.
More informationDimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
More informationMercaptoacetic Acid Method*
Method 8036 MOLYBDENUM, MOLYBDATE, High Range (0 to 40.0 mg/l) Mercaptoacetic Acid Method* Using Powder Pillows For water and wastewater 1. Enter the stored program number for high range molybdenumpowder
More informationFerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV
Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),
More informationOxygen Demand, Chemical
Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.
More informationSulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure
Methylene Blue Method 1 (0 0.70, 0 7.00 and 0 70.00 mg/l) DOC316.53.01320 Method 10254 Scope and Application: For oil and gas field waters. 1 Adapted from Standard Methods for the Examination of Water
More informationPeracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )
Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 ) DOC316.53.01523 DPD Method Method 10290 0.10 to 10.00 mg/l PAA / 0.05 to 5.00 mg/l H 2 O 2 Powder Pillows Scope and application: For testing peracetic
More informationChlorine, Free and Total, High Range
Chlorine, Free and Total, High Range DOC316.53.01490 USEPA DPD Method 1 Method 10069 (free) 10070 (total) 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of free
More informationBy Authority Of THE UNITED STATES OF AMERICA Legally Binding Document
By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has
More informationChlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.
Chlorine, Total DOC316.53.01497 USEPA DPD Method 1 Method 10260 0.04 to 10.00 mg/l as Cl 2 Chemkey Reagents Scope and application: For drinking water. This product has not been evaluated to test for chlorine
More informationChlorine, Free and Total, High Range
Chlorine, Free and Total, High Range DOC316.53.01449 USEPA DPD Method 1 Method DPD 0.1 to 8.0 mg/l Cl 2 Powder Pillows Scope and application: For testing residual chlorine and chloramines in water, wastewater,
More informationChloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA
, 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine
More informationPhenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator
Alkalinity DOC316.53.01166 Phenolphthalein and Total Alkalinity Method 8203 10 to 4000 mg/l as CaCO 3 Digital Titrator Scope and application: For water, wastewater and seawater. Test preparation Before
More informationDR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.
DR/4000 PROCEDURE Method 8027 Pyridine-Pyrazalone Method* Powder Pillows (0 to 0.240 mg/l CN ) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number
More informationSILICA, Low Range (0 to 1.60 mg/l)
SILICA, Low Range (0 to 1.60 mg/l) Heteropoly Blue Method * Method 8186 For water and seawater 1. Enter the stored program number for low range silica (SiO 2 ). Press: PRGM The display will show: PRGM?
More informationTrihalomethane Formation Potential (THMFP)
Trihalomethane Formation Potential (THMFP) DOC316.53.01147 THM Plus Method 1 Method 10224 Scope and application: To determine the potential of potable source waters that form trihalomethanes and other
More informationMethod (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows
Chlorine Free, Indophenol, 10241 Indophenol 1 1 Patent pending. DOC316.53.01256 Method 10241 (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Scope and Application: For determining residual free chlorine levels
More informationTPH (Total Petroleum Hydrocarbons)
TPH (Total Petroleum Hydrocarbons) DOC316.53.01475 Immunoassay 1 Method 10050 Scope and application: For water. 1 This test is semi-quantitative. Results are shown as more or less than the threshold value
More informationChlorine, Free and Total, Low Range
Chlorine, Free and Total, Low Range DOC316.53.01450 USEPA DPD Method 1 Method 8021 (free) 8167 (total) 0.02 to 2.00 mg/l Cl 2 (LR) Powder Pillows or AccuVac Ampuls Scope and application: For testing residual
More informationScope and application: For wastewater, seawater, drinking water, surface water and process water.
Cyanide DOC316.53.01495 USEPA 1 Pyridine Barbituric Acid Method 2 Method 10265 0.01 to 0.60 mg/l CN TNTplus 862 Scope and application: For wastewater, seawater, drinking water, surface water and process
More informationNITROGEN, TOTAL, HR, Test N Tube
Method 10072 NITROGEN, TOTAL, HR, Test N Tube 10 to 150 mg/l N TNT Persulfate Digestion Method For water and wastewater Digestion is required for determining total nitrogen. The digestion procedure is
More informationNITROGEN, AMMONIA, High Range, Test N Tube
NITROGEN, AMMONIA, High Range, Test N Tube Method 10031 (0 to 50 mg/l NH 3 -N) For water, wastewater, and seawater Salicylate Method * 1. Enter the stored program number for nitrogen, ammonia, high range
More informationCadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)
Method 8171 NITRATE, Mid Range (0 to 5.0 mg/l NO 3- -N) For water, wastewater and seawater* Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Using Powder Pillows 1. Enter the stored program
More informationNITRITE, Low Range (0 to mg/l NO 2- -N)
NITRITE, Low Range (0 to 0.350 mg/l NO 2- -N) Method 8507 For water, wastewater, seawater Diazotization Method * (Powder Pillows or AccuVac Ampuls); USEPA approved for reporting wastewater and drinking
More informationNITROGEN, TOTAL, Test N Tube * Method 10071
NITROGEN, TOTAL, Test N Tube * Method 10071 TNT Persulfate Digestion Method (0 to 25.0 mg/l N) Scope and Application: For water and wastewater; digestion is required for determining total nitrogen. The
More informationDR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.
Method 10065 DR/4000 PROCEDURE Scope and Application: For water, wastewater and seawater. * Adapted from Analytical Chemistry, 25 (9) 1363 (1953) Cold Vapor Mercury Preconcentration Method* (0.1 to 2.5
More informationDirect Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution
, drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific
More informationDPD Test N Tube Method *
CHLORINE, FREE (0 to 5.00 mg/l) DPD Test N Tube Method * Method 10102 For water, wastewater, and seawater 1. Enter the stored program number for Test N Tube free chlorine (Cl 2 ). Press: PRGM The display
More informationPolychlorinated Biphenyls (PCB) in Soil
Polychlorinated Biphenyls (PCB) in Soil DOC316.53.01107 Immunoassay 1 Method 10050 Scope and application: For soil. 1 This test is semi-quantitative. Results are shown as more or less than the threshold
More informationManganese III Digestion Method * (with chloride removal)
Method 10067 OXYGEN DEMAND, CHEMICAL (20 to 1,000 mg/l) For water and wastewater Manganese III Digestion Method * (with chloride removal) 1. Enter the stored program number for Manganese III COD. Press:
More informationCHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069
CHLORINE, FREE, Ultra-high Range (0.0 10.0 mg/l Cl 2 ) Method 10069 DPD Method USEPA accepted for reporting drinking water analyses * For testing higher levels of free chlorine (hypochlorous acid and hypochlorite)
More informationALLOWAY METHOD OUTLINE
ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:
More informationStandard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.
Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter Missouri State University and Ozarks Environmental and Water Resources Institute (OEWRI) Prepared
More informationMERCURY (0.1 to 2.5 µg/l) Method 10065
MERCURY (0.1 to 2.5 µg/l) Method 10065 Cold Vapor Mercury Preconcentration Method * Scope and Application: For water, wastewater and seawater Cat. No. 26583-88 Phase 1 Sample Digestion - must be done in
More informationp-dimethylaminobenzaldehyde Method*
HYDRAZINE (0 to 500 µg/l) p-dimethylaminobenzaldehyde Method* Using Reagent Solution Method 8141 For boiler water/feedwater, water and seawater 1. Enter the stored program number for hydrazine (N 2 H 4
More informationHach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater
Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric
More informationSILICA, Low Range SiLR
SILICA, Low Range SiLR SP-910 Test Method Test Program Description:SP-910 Silica Low Range Method (0.02 1.60 ppm SiO 2 ) (Heteropoly Blue Method) Instruments and Reagents: Program 1. SP-910 Portable Water
More informationHach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis
Hach Method 1061 Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis Hach Company Method 1061 Revision 1. December 015 Organic Carbon in Finished
More informationRAPID LIQUID TESTING. Hach Company, 1996 All rights reserved. Printed in the U.S.A dld/dp ed
26578-88 RAPID LIQUID TESTING Hach Company, 1996 All rights reserved. Printed in the U.S.A. 7040 dld/dp 02-02-96 1ed TABLE OF CONTENTS OPERATION...1 INTRODUCTION...3 Rapid Liquid Testing...3 Rapid Liquid
More informationRevision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:
ALLOWAY METHOD OUTLINE Standard Laboratory Method: SM Parameter: Methylene Blue Method: Colorimetric Reporting Level: Reference: 0.05 mg/l Standard Methods for the Examination Of Water and Wastewater;
More informationUSGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7
Troy, NY 12180 Date: 03/16/2012 Page 1 of 7 USGS District Laboratory, Troy, NY Inductively Coupled Plasma Optical Emission Spectrometry Standard Operating Procedure 1. Scope and Application 1.1 Analytes
More informationHach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water
Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished
More informationL Mott SALT WATER AQUACULTURE. Model AQ-4/AQ-5 Code / test kit instruction manual
L Mott SALT WATER AQUACULTURE test kit instruction manual Model AQ-4/AQ-5 Code 3635-03/3636-03 This booklet provides step-by-step detailed instructions for the Model AQ-4 (Code 3635-03) and AQ-5 (Code
More informationAmmonia Gas Sensing Combination Electrode
51940-88 Ammonia Gas Sensing Combination Electrode Model 50250 Hach Company, 1999. All rights reserved. Printed in the U.S.A. hm/dk 4/99 1ed TRADEMARKS OF HACH COMPANY 2 TABLE OF CONTENTS SPECIFICATIONS...
More informationAmmonia Gas Sensing Combination Electrode
51927-88 Ammonia Gas Sensing Combination Electrode Model 51927-00 Hach Company, 2001. All rights reserved. Printed in the U.S.A. te/dk 08/01 1ed 2 TABLE OF CONTENTS SPECIFICATIONS...5 SAFETY PRECAUTIONS...7
More informationSC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33
Photometer-System SC400 Colorimeter SiO 2 Silica DE GB FR IT ES Bedienungsanleitung Seite 3 17 Instruction Manual Page 19 33 d'emploi Page 35 49 Instruction Manual Istruzioni d'uso Pagina 51 65 Instrucciones
More information1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction
1. Application 1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction Parameters and Codes: EDB and DBCP, whole water recoverable, O-3120-90 Parameter (µg/l)
More informationCat. No
2430-90 Cat. No. 2430-03 Hach Company 1989, 1993, 1999. aa/dk 11/99 3ed All rights reserved. Printed in the U.S.A. TABLE OF CONTENTS Introduction...3 Acidity...5 Alkalinity...7 Ammonia...9 Carbon Dioxide...12
More informationZ/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:
sop-c-102 Determination of Chemical Oxygen Demand Revision 6 Approval: Laborato ry Manager/LQAO/RS O Date Z/zs Date Z-z* I A Renewat date: 7lt:,: Effective date: 3, tnitiats: wu\ Texas lnstitute for Applied
More informationGENERAL INSTRUCTIONS GENERAL PREPARATION
GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Ammonium Ion Selective Electrode is used to quickly, simply, accurately, and economically measure potassium in aqueous solutions. Required
More informationExperimental Procedure Lab 402
Experimental Procedure Lab 402 Overview One set of solutions having known molar concentrations of FeNCS 2+ is prepared for a calibration curve, a plot of absorbance versus concentration. A second set of
More informationCHEMISTRY DEPARTMENT
POINT LEPREAU GENERATING STATION GADOLINIUM ARSENAZO III Issued By: Date: APPROVED BY: of 6 REVISION RECORD AUTHOR REV DESCRIPTION DATE 0 Initial Issue W. Mawhinney Not Tracked Oct./9 W. T. Underhill 2
More informationSupernatant: The liquid layer lying above the solid layer after a precipitation reaction occurs.
Limiting Reagent Introduction The quantities of substances involved in a chemical reaction represented by a balanced equation are often referred to as stoichiometric amounts. Solution stoichiometry is
More informationSIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE
SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER 10265 COLORIMETRIC DETERMINATION OF CYANIDE TOPIC EPA METHOD NUMBER 335.2 (1980) Direct Photometric Method 0.04 0.80 mg cyanide/l SCOPE AND APPLICATION
More informationHach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater
Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater Hach Company TNTplus TM Phosphorus Method 10209/10210/843/844/845 November 2015 Revison 2 Spectrophotometric
More informationDirect Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry
Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry NOTICE: This method has been developed by the EURL-CEFAO using the facilities available in its laboratories.
More informationMETHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY
METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY 1.0 SCOPE AND APPLICATION 1.1 This digestion procedure is used for the preparation
More informationFormaldehyde Fluorescent Detection Kit
K-ASSAY KAMIYA BIOMEDICAL COMPANY KAMIYA BIOMEDICAL COMPANY Formaldehyde Fluorescent Detection Kit For detection in human urine and tissue culture media Cat. No. KT-723 For Research Use Only. Not for Use
More informationExperiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+
Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+ Many substances absorb light. When light is absorbed, electrons in the ground state are excited to higher energy levels. Colored
More information