NITROGEN, TOTAL, Test N Tube * Method 10071

Size: px
Start display at page:

Download "NITROGEN, TOTAL, Test N Tube * Method 10071"

Transcription

1 NITROGEN, TOTAL, Test N Tube * Method TNT Persulfate Digestion Method (0 to 25.0 mg/l N) Scope and Application: For water and wastewater; digestion is required for determining total nitrogen. The digestion procedure is included in the method. 1. Turn on the COD Reactor. Heat to C (best temperature is 105 C). Place the plastic shield in front of the reactor. Note: Ensure safety devices are in place to protect the analyst from splattering should leakage occur. Note: For proof of accuracy, run a 20 mg/l NH 3 N standard through digestion and analysis. 2. Using funnels, add the contents of one Total Nitrogen Persulfate Reagent Powder Pillow to each of two Total Nitrogen Hydroxide Reagent vials. Note: Wipe off any reagent that may get on the lid or the tube threads. Note: One reagent blank is sufficient for each set of samples. 3. Add 2 ml of sample to one vial. Add 2 ml of organic-free water to another vial (the reagent blank). Cap both vials. Shake vigorously to mix (more than 30 seconds). Place the vials in the COD Reactor. Heat for 30 minutes. Note: The persulfate reagent may not dissolve completely after shaking. Note: Alternative water must be free of all nitrogen-containing species. 4. Using finger cots or gloves, remove the hot vials from the reactor and allow to cool to room temperature. Note: It is very important to remove the vials from the COD Reactor after exactly 30 minutes. * TenSette, Test N Tube, and Voluette are trademarks of Hach Company.

2 410 nm 5. Enter the stored program for Test N Tube Total Nitrogen. Press: READ/ENTER The display will show: Dial nm to 410 Note: DR/2000s with software version 3.0 and greater will not display Dial nm to message if the wavelength is already set correctly. The display will show the message in step 7. Proceed with step 8. Note: For DR/2000s without this stored program, see Instrument Setup following these steps. 6. Rotate the wavelength dial until the display shows: 410 nm Note: For DR/3000s, set the wavelength to 410 nm and press CLEAR. 7. Press: READ/ENTER The display will show: mg/l N TN TNT 8. Remove the caps from the cooled, digested vials and add the contents of one TN Reagent A Powder Pillow to each vial. Cap vials and shake for 15 seconds. Press: SHIFT TIMER after shaking. A three-minute reaction period will begin. Note: For DR/3000s, press 3 TIMER. 2

3 9. After the timer beeps, uncap the vials and add one TN Reagent B Powder Pillow to each vial. Cap vials and shake for 15 seconds. Press: SHIFT TIMER after shaking. A two-minute reaction period will begin. Note: The reagent will not completely dissolve. The solution will begin to turn yellow. 10. After the timer beeps, remove the caps from two TN Reagent C Vials and add 2 ml of digested, treated sample to one vial. Add 2 ml of the digested, treated reagent blank to the second TN Reagent C Vial. 11. Cap both vials and invert 10 times to mix. Use slow, deliberate inversions for complete recovery. The vials will be warm. Note: Hold the vial vertical with the cap up. Invert the vial and wait for all of the solution to flow to the cap end. Return the vial to the upright position and wait for all of the solution to flow to the vial bottom. Pause. This is one inversion (10 inversions = 30 seconds). 12. Press: SHIFT TIMER A 5-minute reaction period will begin. Do not shake the vial again. Note: The yellow color will intensify. Note: For DR/3000s, press 5TIMER. Note: For DR/3000s, press 2 TIMER. 13. Insert the COD Vial Adapter into the cell holder with the marker to the right. Note: For DR/3000s, the groove in the adapter faces the front of the instrument. 14. When the timer beeps, wipe the outside of both vials. Note: Wiping with a damp towel then a dry one will remove fingerprints or other marks. 15. Place the vial containing the reagent blank into the adapter with the Hach logo toward the front of the instrument. Place the cover on the adapter. Note: The reagent blank is stable when stored in the dark; see Blanks for Colorimetric Measurement on page 10 following this procedure. 16. Press: ZERO The display will show: then: WAIT 0.0 mg/l N TN TNT Note: For DR/3000s, press MANUAL PROGRAM. Then press ZERO. Press ZERO again if the display does not show Enter the concentration factor (50.05) and press CONC FACTOR. Then press 1 CONC. 3

4 17. Place the prepared sample into the cell holder with the Hach logo toward the front of the instrument and place the cover on the adapter. Note: Multiple samples may be read after zeroing on one reagent blank. 18. Press: READ/ENTER The display will show: WAIT then the result in mg/l total nitrogen will be displayed. Note: If the test overranges, repeat the digestion and measurement with diluted sample. The entire digestion must be repeated for accurate results. Note: For DR/3000s, pressing READ is not necessary. The instrument will automatically display the results. Instrument Setup DR/2000 with software versions 2.0 and 2.2. Enter the calibration as an update to Hach-stored programs. 1. Press I/O. 2. Press SHIFT CONFIG/METH. 3. Press: 4. Within 3 seconds, press SHIFT PROG/3 CONFIG/METH. The display will show: ENTER nm Note: If you make an error, press SHIFT CLEAR and reenter the number. When the number is correct, press READ/ENTER. 5. Press: The display will show: Decimal?

5 6. Use the arrow keys to correctly position the decimal point. Press the down arrow key once. The display will show: DECIMAL? Press: READ/ENTER 7. The display will show: UNITS? 8. Use the arrow keys to select the appropriate unit of measure. Press the down arrow key twice. The display will show: mg/l 9. Press READ/ENTER when the correct unit of measure is displayed. This display will show: SYMBOL? 10. Construct the correct symbol display. N TN TNT a. Select letters and numbers by scrolling to the correct character with the arrow keys. b. To make a letter uppercase, press the SHIFT key. c. To enter a subscript number, enter the digit on the numeric keypad. d. To enter a superscript number, enter the digit on the numeric keypad then press SHIFT. e. The space is the character displayed after one press of the down arrow key. f. Make sure to enter the display line EXACTLY as shown, including the spaces. Do not enter trailing spaces. g. Accept each symbol by pressing READ/ENTER. h. To end symbol entry, press READ/ENTER a second time. 11. When the instrument is out of symbol entry mode, the display will read: TIMER? 12. This method has 3 timed steps, so press SHIFT TIMER. The display will show: MM:SS TIME 1? 13. Enter a timer value of 3 minutes. Press: 14. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 2? 15. Enter a timer value of 2 minutes. Press: 16. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 3? 17. Enter a timer value of 5 minutes. Press: 18. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 4? 5

6 19. Press READ/ENTER to complete the timer entry. The display will show: #1 Data 20. Enter the following 12 numbers as shown. Complete each number by pressing the READ/ENTER key. # 1 Data 0 # 2 Data 7968 # 3 Data 7968 # 4 Data 7968 # 5 Data 7967 # 6 Data 8447 # 7 Data # 8 Data # 9 Data #10 Data #11 Data 544 Checksum The final number is a check value which determines if the data sequence was correctly entered. If an error was made during number entry, the display will return to the prompt for data #1 and the entire sequence must be re-entered. If all numbers are correctly entered, the display will return to the method prompt and is ready for use. DR/2000 with software version 3.0 and higher 1. Turn the instrument on. Press SHIFT METHOD to enter the configuration mode. The display will show: MOMENTARY or CONSTANT ON. 2. Press the up arrow key twice to select HACH UPDATE. Press READ/ENTER. The display will show: ENTER #: Note: If you make an error, press SHIFT CLEAR and reenter the number. When the number is correct, press READ/ENTER. 3. Press: The display will show: P350 ENTER nm 4. Press: The display will show: P350 Decimal? Use the arrow keys to correctly position the decimal point. Press the down arrow key twice. The display will show: DECIMAL? Press: READ/ENTER. The display will show: P350 UNITS? 6. Use the arrow keys to select the appropriate unit of measure. Press the down arrow key twice. The display will show: P350 mg/l 6

7 7. Press READ/ENTER when the correct unit of measure is displayed. This display will show: P350 mg/l _ 8. Construct the display to read the correct symbol. The symbol must be entered exactly as shown including spaces between characters: N TN TNT a. Select letters and numbers by scrolling to the correct character with the arrow keys. b. To make a letter uppercase, press the SHIFT key. To make numbers and symbols superscript, subscript, or regular, press SHIFT until the display is correct. c. The space is the character displayed after one press of the down arrow key. d. Make sure to enter the display line EXACTLY as shown, including the spaces. Do not enter trailing spaces. e. Accept each symbol by pressing READ/ENTER. f. To end symbol entry, press READ/ENTER a second time. Alternatively, use this procedure: To enter N, press 7 8 READ/ENTER READ/ENTER (a space) 3 2 READ/ENTER READ/ENTER T N 8 4 READ/ENTER READ/ENTER 7 8 READ/ENTER READ/ENTER (a space) 3 2 READ/ENTER READ/ENTER T N T 8 4 READ/ENTER READ/ENTER 7 8 READ/ENTER READ/ENTER 8 4 READ/ENTER READ/ENTER 9. When the instrument is out of symbol entry mode, the display will read: P350 TIMER 10. This method has 3 timed steps, so press SHIFT TIMER. The display will show: MM:SS TIME 1? 11. Enter a timer value of 3 minutes. Press: 12. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 2? 13. Enter a timer value of 2 minutes. Press: 7

8 14. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 3? 15. Enter a timer value of 5 minutes. Press: 16. Press READ/ENTER to accept the timer value. The display will show: MM:SS TIME 4? 17. Press READ/ENTER to complete the timer entry. The display will show: #0 STANDARD 18. Press READ/ENTER to display the zero data pair. The display will show: Abs mg/l 19. Press READ/ENTER. The display will show: #1 STANDARD 20. Press READ/ENTER. The display will prompt for entry of the first concentration point: # mg/l 21. Enter concentration point #1 by pressing 0028 so that the display shows: #1028.0mg/l 22. Press READ/ENTER. The display will prompt for entry of the first absorbance point: # Abs 23. Enter the absorbance point #1 by pressing The display will show: # Abs 24. Press READ/ENTER. The display will show the first data pair: Abs mg/l 25. Press READ/ENTER to accept the first data pair. The display will show: #2 STANDARD 26. The data pair values from the table below are now entered. Standard Concentration Absorbance # mg/l Abs # mg/l Abs 27. The data pair values are now entered. The display will show: #2 STANDARD 28. Press SHIFT READ/ENTER to complete data point entry. The display will show: #: 29. Enter the validation number The display will show: #: Press READ/ENTER. The display will show: COMPLETED then: P350 mg/l N TN TNT Note: If the display shows INCORRECT #, then prompts for the validation number, a mistake may have been made during data entry. Make sure the validation number is correct. If so, the error occurred during some other portion of the method entry. Press METH and respond to the ABORT? message by pressing READ/ENTER, then re-enter the method. The instrument is now ready for use with method

9 Sampling and Storage Collect samples in clean plastic or glass bottles. Best results are obtained with immediate analysis. Preserve the sample by reducing the ph to 2 or less with concentrated sulfuric acid (at least 2 ml). Store at 4 C (39 F) or less. Preserved samples may be stored up to 28 days. Warm samples to room temperature and neutralize with 5 N sodium hydroxide before analysis. Correct the test result for volume additions; see Correcting for Volume Additions in Section 1 of the Procedures Manual. Accuracy Check This method generally yields % recovery on organic nitrogen standards. For proof of accuracy, Hach offers a set of 3 Primary Standards for Kjeldahl Nitrogen. 1. Prepare one or more of the following 3 solutions. Each preparation is for an equivalent 25 mg/l N standard. Use water that is free of all nitrogen-containing species. a. Weigh g of Ammonium p-toluenesulfonate(ptsa). Dissolve in a 1000-mL volumetric flask with deionized water. Add deionized water to the 1000-mL mark. b. Weigh g of Glycine p-toluenesulfonate. Dissolve in a 1000-mL volumetric flask with deionized water. Add deionized water to the 1000-mL mark. c. Weigh g of Nicotinic p-toluenesulfonate. Dissolve in a 1000-mL volumetric flask with deionized water. Add deionized water to the 1000-mL mark. 2. Analyze each of these solutions using the test procedure above. Calculate the percent recovery for each using this formula: measured concentration % recovery = The percent recovery is listed in Table 1: Hach analysts have found Ammonia-PTSA to be the most difficult to digest. Other compounds may yield different percent recoveries. Table 1 Recoveries For Primary Standards Compound Standard Solution Method For proof of accuracy, substitute 2 ml of a 20-mg/L ammonia nitrogen standard solution for the sample in the procedure. To prepare a 20-mg/L ammonia nitrogen standard, use a 20-mL Class A pipet to add 20 ml of a 100-mg/L Ammonia Nitrogen Standard (see Optional Reagents) to a 100-mL Class A volumetric flask. Dilute to the line with deionized water that is free of all nitrogen-containing species. A single analyst should obtain less than 5% variation on replicates. 9 Lowest Expected% Recovery Ammonia-PTSA 95% Glycine-PTSA 95% Nicotinic-PTSA 95%

10 Standard Additions Method a. Fill three, 25-mL graduated mixing cylinders with 25 ml of sample. b. Snap the neck off an Ammonia Nitrogen Standard Solution Voluette Ampule, 160 mg/l as NH 3 N. c. Use a TenSette Pipet to add 0.3, 0.6 and 0.9 ml of standard, respectively, to the three mixing cylinders. d. Stopper each cylinder and mix thoroughly. e. Add 2 ml of each prepared solution, respectively, to three Total Nitrogen Hydroxide Reagent Vials. f. Analyze each standard addition sample as described in the procedure. The nitrogen concentration should increase by about 1.9, 3.8 and 5.6 mg/l N, respectively. g. If these increases do not occur, see Standard Additions in Section 1 of the Procedures Manual for troubleshooting information. Blanks for Colorimetric Measurement The reagent blank may be used repeatedly for measurements using the same lots of reagents. Store the reagent blank in the dark at room temperature (18 25 C) for a maximum of 7 days. If a small amount of white floc appears prior to the end of one week, discard the reagent blank and prepare a new one. Interferences Table 2 lists substances that resulted in a concentration change of ± 10%: Table 2 Interferences that Result in a Concentration Change of ± 10% Substance Level Effect Cl - > 1000 ppm positive interference Br - > 60 ppm positive interference Table 3 lists substances that have been tested and do not interfere up to the indicated levels (in mg/l): Substance Table 3 Substances that Do Not Interfere Maximum Level Tested (mg/l) Barium 2.6 Calcium 300 Chromium (3+) 0.5 Iron 2 Lead 6.6 ppb Magnesium 500 Organic Carbon 150 ph 13 ph units Phosphorus 100 Silica 150 Silver 0.9 Tin

11 Method Performance Hach chemists tested this chemistry on standard nitrogen solutions prepared from the following compounds with 95% or better recovery: Ammonium Acetate Ammonium Chloride Ammonium Sulfate Glycine Urea Ammonium chloride or nicotinic-ptsa spikes in domestic influent, effluent, and the ASTM standard specification for substitute wastewater (D ) also resulted in 95% recovery. The large amounts of nitrogen-free organic compounds in some samples may decrease digestion efficiency by consuming some of the persulfate present. Samples known to contain high levels of organics should be diluted and re-run to verify digestion efficiency. Precision A Hach chemist analyzed two independent nutrient standards. The lowest average percent recovery was 95% with a standard deviation of ± 2%. DR/2000: In a single laboratory, using a standard solution of 15 mg/l NH 3 N and two lots of reagent with a DR/2000 Spectrophotometer, a single operator obtained a standard deviation of less than 1.0 mg/l N. DR/3000: In a single laboratory, using a standard solution of 15 mg/l NH 3 N and two lots of reagent with a DR/3000 Spectrophotometer, a single operator obtained a standard deviation of less than 1.0 mg/l N. Summary of Method An alkaline persulfate digestion converts all forms of nitrogen to nitrate. Sodium metabisulfite is added after digestion to eliminate halogen oxide interferences. Nitrate then reacts with chromotropic acid under strongly acidic conditions to form a yellow complex with an absorbance maximum at 410 nm. 11

12 REQUIRED REAGENTS Description Cat. No. Test N Tube Total Nitrogen Reagent Set (50 tests) Includes: (4) *, (4) *, (1) , (1) , (1) Quantity Required Description Per Test Unit Cat. No. Total Nitrogen Hydroxide Reagent Vials, 0.1 N /pkg...* Total Nitrogen Persulfate Reagent Powder Pillows /pkg Total Nitrogen Reagent A Bisulfite Powder Pillows /pkg Total Nitrogen Reagent B Indicator Powder Pillows /pkg Total Nitrogen Reagent C Acid Solution Vials /pkg...* REQUIRED APPARATUS COD Reactor, 115/230 V, North American Plug each COD Reactor, 230 V, European Plug each COD Vial Adapter, DR/ each Funnel, micro each Pipet, TenSette, ml each Pipet Tips for /pkg Pipet, TenSette, ml each Pipet Tips for /pkg Safety Shield, for COD Reactor each Test Tube Cooling Rack each OPTIONAL REAGENTS Nitrogen, Ammonia, 100 mg/l NH 3 N ml Nitrogen, Ammonia, Voluette Ampules, 160 mg/l NH 3 N, 10 ml...16/pkg Primary Standards for Kjeldahl Nitrogen...3/pkg Sodium Hydroxide Standard Solution, 5.0 N ml MDB Sulfuric Acid, ACS ml Water, organic-free ml OPTIONAL APPARATUS Balance, analytical, 110/220 V...each Flask, volumetric, Class A, 1000 ml (3 required for accuracy check)...each * Not available separately. Hach Company, All rights reserved. Printed in the U.S.A /96 1ed aa/dp rev. 1 8/99

NITROGEN, TOTAL, HR, Test N Tube

NITROGEN, TOTAL, HR, Test N Tube Method 10072 NITROGEN, TOTAL, HR, Test N Tube 10 to 150 mg/l N TNT Persulfate Digestion Method For water and wastewater Digestion is required for determining total nitrogen. The digestion procedure is

More information

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials Nitrogen, Total DOC316.53.01086 Persulfate Digestion Method Method 10071 0.5 to 25.0 mg/l N (LR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific

More information

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials Nitrogen, Total DOC316.53.01085 Persulfate Digestion Method Method 10072 2 to 150 mg/l N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit

More information

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Nitrate, HR DOC316.53.01068 Chromotropic Acid Method Method 10020 0.2 to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows FILL LINE Method 8155 Salicylate Method 1 Powder Pillows Scope and Application: For water, wastewater, and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) (0.01 to 0.50 mg/l NH 3 N) Test Preparation

More information

NITROGEN, AMMONIA, High Range, Test N Tube

NITROGEN, AMMONIA, High Range, Test N Tube NITROGEN, AMMONIA, High Range, Test N Tube Method 10031 (0 to 50 mg/l NH 3 -N) For water, wastewater, and seawater Salicylate Method * 1. Enter the stored program number for nitrogen, ammonia, high range

More information

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE NITRATE. 2. The display will show: Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

Mercaptoacetic Acid Method*

Mercaptoacetic Acid Method* Method 8036 MOLYBDENUM, MOLYBDATE, High Range (0 to 40.0 mg/l) Mercaptoacetic Acid Method* Using Powder Pillows For water and wastewater 1. Enter the stored program number for high range molybdenumpowder

More information

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Method 8171 NITRATE, Mid Range (0 to 5.0 mg/l NO 3- -N) For water, wastewater and seawater* Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Using Powder Pillows 1. Enter the stored program

More information

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,

More information

Nitrogen, Total Inorganic

Nitrogen, Total Inorganic Nitrogen, Total Inorganic DOC316.53.01090 Titanium Trichloride Reduction Method Method 10021 0.2 to 25.0 mg/l N Test N Tube Vials Scope and application: For water, wastewater and seawater. Test preparation

More information

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ ) DR/4000 PROCEDURE Method 8023 1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to 0.700 mg/l Cr 6+ ) Scope and Application: For water and wastewater;usepa accepted for reporting for

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827 Nitrogen, Total DOC316.53.01088 Persulfate Digestion Method Method 10208 5 to 40 mg/l N (HR) TNTplus 827 Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832 , TNTplus 832, 10205 DOC316.53.01083 Salicylate Method Method 10205 HR (2 to 47 mg/l NH 3 N) TNTplus 832 Scope and Application: For surface waters, municipal and industrial wastewaters. Test preparation

More information

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 , MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation

More information

DR/4000 PROCEDURE. IRON, Total

DR/4000 PROCEDURE. IRON, Total Method 8365 Powder Pillows DR/4000 PROCEDURE FerroMo Method* (0 to 1.800 mg/l) Scope and Application: For cooling water containing molybdate-based treatment; digestion is required for determining total

More information

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Chromium, Total DOC316.53.01034 Alkaline Hypobromite Oxidation Method 1 Method 8024 0.01 to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Scope and application: For water and wastewater.

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater. Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for

More information

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.

More information

NITRITE, Low Range (0 to mg/l NO 2- -N)

NITRITE, Low Range (0 to mg/l NO 2- -N) NITRITE, Low Range (0 to 0.350 mg/l NO 2- -N) Method 8507 For water, wastewater, seawater Diazotization Method * (Powder Pillows or AccuVac Ampuls); USEPA approved for reporting wastewater and drinking

More information

DPD Test N Tube Method *

DPD Test N Tube Method * CHLORINE, FREE (0 to 5.00 mg/l) DPD Test N Tube Method * Method 10102 For water, wastewater, and seawater 1. Enter the stored program number for Test N Tube free chlorine (Cl 2 ). Press: PRGM The display

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 1 to 60 mg/l COD (ULR) TNTplus 820 Scope and application: For wastewater, process water, surface water, and cooling water.

More information

Chloramine (Mono) and Nitrogen, Free Ammonia

Chloramine (Mono) and Nitrogen, Free Ammonia Chloramine (Mono) and Nitrogen, Free Ammonia DOC316.53.01016 Indophenol Method 1 Method 10200 0.04 to 4.50 mg/l Cl 2 0.01 to 0.50 mg/l NH 3 N Powder Pillows Scope and application: For the determination

More information

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water. Method 8337 Powder Pillows DR/4000 TT PROCEDURE QUATERNARY AMMONIUM COMPOUNDS Direct Binary Complex Method (0 to 5.00 mg/l as CTAB) Scope and Application: For cooling tower and pool/spa water. The estimated

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right. Barium DOC316.53.01315 Turbidimetric Method 1 Method 10251 2 to 100, 20 to 1000, 200 to 10,000 mg/l Ba (spectrophotometers) Powder Pillows 2 to 80, 20 to 800, 200 to 8000 mg/l Ba (colorimeters) Scope and

More information

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO Nitrate DOC316.53.01070 Dimethylphenol Method Method 10206 0.23 to 13. NO 3 N or 1.00 to 60.00 mg/l NO 3 (LR) TNTplus 835 Scope and application: For wastewater, drinking water, surface water and process

More information

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069 CHLORINE, FREE, Ultra-high Range (0.0 10.0 mg/l Cl 2 ) Method 10069 DPD Method USEPA accepted for reporting drinking water analyses * For testing higher levels of free chlorine (hypochlorous acid and hypochlorite)

More information

PAR Method Method to 2.0 mg/l Pb TNTplus 850

PAR Method Method to 2.0 mg/l Pb TNTplus 850 , TNTplus 850, 10216 DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and Application: For wastewater and process control Test preparation Before starting the test: FOR DR 2800:

More information

SILICA, Low Range (0 to 1.60 mg/l)

SILICA, Low Range (0 to 1.60 mg/l) SILICA, Low Range (0 to 1.60 mg/l) Heteropoly Blue Method * Method 8186 For water and seawater 1. Enter the stored program number for low range silica (SiO 2 ). Press: PRGM The display will show: PRGM?

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right. Aluminum DOC316.53.01003 Eriochrome Cyanine R Method 1 Method 8326 0.006 to 0.250 mg/l Al 3+ Powder Pillows Scope and application: For water. 1 Adapted from Standard Methods for the Examination of Water

More information

Chlorine, Free and Total, High Range

Chlorine, Free and Total, High Range Chlorine, Free and Total, High Range DOC316.53.01490 USEPA DPD Method 1 Method 10069 (free) 10070 (total) 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of free

More information

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample. Method 8033 DR/4000 PROCEDURE Dithizone Method* (0 to 300 µg/l) Scope and Application: For water and wastewater; USEPA accepted for reporting for wastewater analysis (digestion is required).** See Section

More information

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Cadion Method Method to 0.30 mg/l Cd TNTplus 852 Cadmium DOC316.53.01013 Cadion Method Method 10217 0.02 to 0.30 mg/l Cd TNTplus 852 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows

More information

Chlorine, Free and Total, High Range

Chlorine, Free and Total, High Range Chlorine, Free and Total, High Range DOC316.53.01449 USEPA DPD Method 1 Method DPD 0.1 to 8.0 mg/l Cl 2 Powder Pillows Scope and application: For testing residual chlorine and chloramines in water, wastewater,

More information

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows Silver DOC316.53.01134 Colorimetric Method Method 8120 0.02 to 0.70 mg/l Ag Powder Pillows Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

CHLORINE, TOTAL (0 to 4.00 mg/l)

CHLORINE, TOTAL (0 to 4.00 mg/l) CHLORINE, TOTAL (0 to 4.00 mg/l) DOC316.53.01261 For water, wastewater, and seawater Method 10250 DPD Method Powder Pillows USEPA accepted for reporting water and wastewater analyses * Note: This product

More information

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows , 8326 Eriochrome Cyanine R Method 1 Scope and Application: For water 1 Adapted from Standard Methods for the Examination of Water and Wastewater. DOC316.53.01003 Method 8326 (0.006 to 0.250 mg/l Al 3+

More information

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ). Method 8194 DR/4000 PROCEDURE Diaminobenzidine Method* (0 to 1.000 mg/l) Scope and Application: For water and wastewater; distillation is required for determining total selenium. See the Distillation procedure

More information

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848 Aluminum DOC316.53.01004 Chromazurol S Method Method 10215 0.02 to 0.50 mg/l Al TNTplus 848 Scope and application: For drinking water, surface water, swimming pool water, wastewater and process analysis.

More information

PAR Method Method to 2.0 mg/l Pb TNTplus 850

PAR Method Method to 2.0 mg/l Pb TNTplus 850 Lead DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows all

More information

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),

More information

p-dimethylaminobenzaldehyde Method*

p-dimethylaminobenzaldehyde Method* HYDRAZINE (0 to 500 µg/l) p-dimethylaminobenzaldehyde Method* Using Reagent Solution Method 8141 For boiler water/feedwater, water and seawater 1. Enter the stored program number for hydrazine (N 2 H 4

More information

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 0.1 to 25.0 mg/l Cl Reagent Solution Scope and application: For water and wastewater. Test preparation Instrument-specific information Table

More information

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Scope and application: For wastewater, seawater, drinking water, surface water and process water. Cyanide DOC316.53.01495 USEPA 1 Pyridine Barbituric Acid Method 2 Method 10265 0.01 to 0.60 mg/l CN TNTplus 862 Scope and application: For wastewater, seawater, drinking water, surface water and process

More information

Quaternary Ammonium Compounds

Quaternary Ammonium Compounds , 8337 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB Powder Pillows Scope and Application: For cooling tower water and pool/spa water Test

More information

Chlorine, Free and Total, Low Range

Chlorine, Free and Total, Low Range Chlorine, Free and Total, Low Range DOC316.53.01450 USEPA DPD Method 1 Method 8021 (free) 8167 (total) 0.02 to 2.00 mg/l Cl 2 (LR) Powder Pillows or AccuVac Ampuls Scope and application: For testing residual

More information

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters) Silica DOC316.53.01132 Heteropoly Blue Method 1 Method 8186 0.010 to 1.600 mg/l SiO 2 (LR, spectrophotometers) Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters) Scope and application: For boiler

More information

Manganese III Digestion Method * (with chloride removal)

Manganese III Digestion Method * (with chloride removal) Method 10067 OXYGEN DEMAND, CHEMICAL (20 to 1,000 mg/l) For water and wastewater Manganese III Digestion Method * (with chloride removal) 1. Enter the stored program number for Manganese III COD. Press:

More information

TPH (Total Petroleum Hydrocarbons)

TPH (Total Petroleum Hydrocarbons) TPH (Total Petroleum Hydrocarbons) DOC316.53.01475 Immunoassay 1 Method 10050 Scope and application: For water. 1 This test is semi-quantitative. Results are shown as more or less than the threshold value

More information

DR/4000 PROCEDURE. HARDNESS, Total

DR/4000 PROCEDURE. HARDNESS, Total Method 8374 DR/4000 PROCEDURE Calcium and Magnesium; Chlorophosphonazo Colorimetric Method ULR (0 to 1,000 µg/l Ca & Mg as CaCO 3 ) Scope and Application: For ultra pure water. The estimated detection

More information

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Chlorine Free, Indophenol, 10241 Indophenol 1 1 Patent pending. DOC316.53.01256 Method 10241 (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Scope and Application: For determining residual free chlorine levels

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right. Nitrite DOC316.53.01074 USEPA Diazotization Method 1 Method 8507 0.002 to 0.300 mg/l NO 2 N (LR, spectrophotometers) 0.005 to 0.350 mg/l NO 2 N (LR, colorimeters) Scope and application: For water, wastewater

More information

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure Methylene Blue Method 1 (0 0.70, 0 7.00 and 0 70.00 mg/l) DOC316.53.01320 Method 10254 Scope and Application: For oil and gas field waters. 1 Adapted from Standard Methods for the Examination of Water

More information

MERCURY (0.1 to 2.5 µg/l) Method 10065

MERCURY (0.1 to 2.5 µg/l) Method 10065 MERCURY (0.1 to 2.5 µg/l) Method 10065 Cold Vapor Mercury Preconcentration Method * Scope and Application: For water, wastewater and seawater Cat. No. 26583-88 Phase 1 Sample Digestion - must be done in

More information

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 ) Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 ) DOC316.53.01523 DPD Method Method 10290 0.10 to 10.00 mg/l PAA / 0.05 to 5.00 mg/l H 2 O 2 Powder Pillows Scope and application: For testing peracetic

More information

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced. Method 10065 DR/4000 PROCEDURE Scope and Application: For water, wastewater and seawater. * Adapted from Analytical Chemistry, 25 (9) 1363 (1953) Cold Vapor Mercury Preconcentration Method* (0.1 to 2.5

More information

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator Alkalinity DOC316.53.01166 Phenolphthalein and Total Alkalinity Method 8203 10 to 4000 mg/l as CaCO 3 Digital Titrator Scope and application: For water, wastewater and seawater. Test preparation Before

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution , drinking water, 8359 DOC316.53.01239 Direct Measurement ISE Method Method 8359 0.04 to 4.00 mg/l NO 3 N TISAB Solution Scope and Application: Drinking water Test preparation How to use instrument-specific

More information

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell Hardness DOC316.53.01044 Chlorphosphonazo Colorimetric Method Method 8374 8 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow Scope and application: For boiler and ultrapure water. Test preparation

More information

Ammonia Gas Sensing Combination Electrode

Ammonia Gas Sensing Combination Electrode 51940-88 Ammonia Gas Sensing Combination Electrode Model 50250 Hach Company, 1999. All rights reserved. Printed in the U.S.A. hm/dk 4/99 1ed TRADEMARKS OF HACH COMPANY 2 TABLE OF CONTENTS SPECIFICATIONS...

More information

Sample Preparation LCW 909. Total Kjeldahl Nitrogen, TKN NB: NEW! LCW 909 GB

Sample Preparation LCW 909. Total Kjeldahl Nitrogen, TKN NB: NEW! LCW 909 GB Principle Organically bound nitrogen is digested by hot fuming sulphuric acid in the presence of catalysts and converted to ammonium, which, after steam distillation, can be determined as so-called Total

More information

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys. DR/4000 PROCEDURE Method 8027 Pyridine-Pyrazalone Method* Powder Pillows (0 to 0.240 mg/l CN ) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

Nitrogen total LR VARIO TT mg/l N b) Persulphate Digestion

Nitrogen total LR VARIO TT mg/l N b) Persulphate Digestion Nitrogen total LR VARIO TT 0.5-25 mg/l N b) Persulphate Digestion 280 Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption

More information

Trihalomethane Formation Potential (THMFP)

Trihalomethane Formation Potential (THMFP) Trihalomethane Formation Potential (THMFP) DOC316.53.01147 THM Plus Method 1 Method 10224 Scope and application: To determine the potential of potable source waters that form trihalomethanes and other

More information

Polychlorinated Biphenyls (PCB) in Soil

Polychlorinated Biphenyls (PCB) in Soil Polychlorinated Biphenyls (PCB) in Soil DOC316.53.01107 Immunoassay 1 Method 10050 Scope and application: For soil. 1 This test is semi-quantitative. Results are shown as more or less than the threshold

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+ Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+ Many substances absorb light. When light is absorbed, electrons in the ground state are excited to higher energy levels. Colored

More information

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated. Nitrogen total HR VARIO TT 5-150 mg/l N b) 81 Persulphate Digestion Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption

More information

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting. Chlorine, Total DOC316.53.01497 USEPA DPD Method 1 Method 10260 0.04 to 10.00 mg/l as Cl 2 Chemkey Reagents Scope and application: For drinking water. This product has not been evaluated to test for chlorine

More information

Ammonia Gas Sensing Combination Electrode

Ammonia Gas Sensing Combination Electrode 51927-88 Ammonia Gas Sensing Combination Electrode Model 51927-00 Hach Company, 2001. All rights reserved. Printed in the U.S.A. te/dk 08/01 1ed 2 TABLE OF CONTENTS SPECIFICATIONS...5 SAFETY PRECAUTIONS...7

More information

RAPID LIQUID TESTING. Hach Company, 1996 All rights reserved. Printed in the U.S.A dld/dp ed

RAPID LIQUID TESTING. Hach Company, 1996 All rights reserved. Printed in the U.S.A dld/dp ed 26578-88 RAPID LIQUID TESTING Hach Company, 1996 All rights reserved. Printed in the U.S.A. 7040 dld/dp 02-02-96 1ed TABLE OF CONTENTS OPERATION...1 INTRODUCTION...3 Rapid Liquid Testing...3 Rapid Liquid

More information

RAPID KJELDAHL BENCHNOTES

RAPID KJELDAHL BENCHNOTES RAPID KJELDAHL BENCHNOTES Methodology for the Determination of Alcohol in a Mixture by Direct Distillation with the RapidStill II Principle: This method covers the determination of percent alcohol in distilled

More information

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water Hach Method 1041 Spectrophotometric Measurement of Free Chlorine (Cl ) in Finished Drinking Water Hach Company Method 1041 Revision 1. November 015 Spectrophotometric Measurement of Free Cl in Finished

More information

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample. METHOD 9200 NITRATE 1.0 SCOPE AND APPLICATION 1.1 This method is applicable to the analysis of ground water, drinking, surface, and saline waters, and domestic and industrial wastes. Modification can be

More information

Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University

Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University Adapted from "Experiment 3: Analysis of Phosphate in Water," Laboratory Experiments

More information

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,: sop-c-102 Determination of Chemical Oxygen Demand Revision 6 Approval: Laborato ry Manager/LQAO/RS O Date Z/zs Date Z-z* I A Renewat date: 7lt:,: Effective date: 3, tnitiats: wu\ Texas lnstitute for Applied

More information

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Nitrate Method 10206 Spectrophotometric Measurement of Nitrate in Water and Wastewater Hach Company TNTplus 835/836 Method 10206 Revision 2.2 January 15, 2013 Spectrophotometric

More information

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Tex-620-J, Determining Chloride and Sulfate Contents in Soil Contents in Soil Contents: Section 1 Overview...2 Section 2 Sample Preparation...3 Section 3 Ion Chromatography Method...5 Section 4 Wet Chemical Method...9 Section 5 Archived Versions...15 Texas Department

More information

ALLOWAY METHOD OUTLINE

ALLOWAY METHOD OUTLINE ALLOWAY METHOD OUTLINE Standard Laboratory Method SM4500-Cl -G Parameter Residual Chlorine & Free Chlorine Method DPD Colorimetric Test Kit Date Issued Originator: Section Supervisor: QA Manager Date:

More information

METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation)

METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation) METHOD #: 340.1 Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) TITLE: Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation) ANALYTE: CAS # F Fluoride 7782-41-4 INSTRUMENTATION: Spectrophotometer

More information

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER 10265 COLORIMETRIC DETERMINATION OF CYANIDE TOPIC EPA METHOD NUMBER 335.2 (1980) Direct Photometric Method 0.04 0.80 mg cyanide/l SCOPE AND APPLICATION

More information

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC) 1.0 SCOPE AND APPLICATION 1.1 Method 7196 is used to determine the concentration of dissolved hexavalent chromium [Cr(VI)] in EP/TCLP characteristic extracts

More information

SILICA, Low Range SiLR

SILICA, Low Range SiLR SILICA, Low Range SiLR SP-910 Test Method Test Program Description:SP-910 Silica Low Range Method (0.02 1.60 ppm SiO 2 ) (Heteropoly Blue Method) Instruments and Reagents: Program 1. SP-910 Portable Water

More information

The Synthesis and Analysis of Aspirin

The Synthesis and Analysis of Aspirin The Synthesis and Analysis of Aspirin Computer 22 Aspirin, the ubiquitous pain reliever, goes by the chemical name acetylsalicylic acid. One of the compounds used in the synthesis of aspirin is salicylic

More information

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION Written by Pamela Doolittle, University of Wisconsin- Madison, pssemrad@wisc.edu 2014 In this experiment you will determine the phosphorus (P) concentration

More information

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE) METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE) 1.0 SCOPE AND APPLICATION 1.1 This automated method is applicable to ground water, drinking and surface waters, and domestic and industrial wastes

More information

EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY

EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY University of Puget Sound Department of Chemistry Chem 230 EXP. 9 EVALUATION OF METHODS FOR DETECTING AMMONIUM: ION SELECTIVE ELECTRODES AND SPECTROSCOPY INTRODUCTION Nitrogen is an essential nutrient

More information

Determination of an Equilibrium Constant

Determination of an Equilibrium Constant Last updated 1/29/2014 - GES Learning Objectives Students will be able to: Determine the numerical value of an equilibrium constant from measured concentrations of all reaction species. Use an absorption

More information

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15 Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15 Abstract: This lab was performed to synthesize acetyl salicylic acid or aspirin from a carboxylic acid and an alcohol. We had learned in class

More information

L Mott SALT WATER AQUACULTURE. Model AQ-4/AQ-5 Code / test kit instruction manual

L Mott SALT WATER AQUACULTURE. Model AQ-4/AQ-5 Code / test kit instruction manual L Mott SALT WATER AQUACULTURE test kit instruction manual Model AQ-4/AQ-5 Code 3635-03/3636-03 This booklet provides step-by-step detailed instructions for the Model AQ-4 (Code 3635-03) and AQ-5 (Code

More information