Neue Möglichkeiten zur Überwachung der Umgebungsluft mit Thermodesorption und GC/MS

Size: px
Start display at page:

Download "Neue Möglichkeiten zur Überwachung der Umgebungsluft mit Thermodesorption und GC/MS"

Transcription

1 Neue Möglichkeiten zur Überwachung der Umgebungsluft mit Thermodesorption und GC/MS Moritz Hebestreit, PhD Applications Chemist Agilent Technologies Europe, Middle East, Africa and India

2 Monitoring Ambient Air by TD and GC/MS

3 What is Thermal Desorption? Sorbent Material Sample Matrix e.g. Chamber or micro-chamber air

4 What is Thermal Desorption? Sample passes onto the sorbent Compounds of interest are adsorbed on the sorbent surface

5 What is Thermal Desorption? Lighter gases such as nitrogen pass through the sorbent

6 What is Thermal Desorption? Sorbent is now heated in a reversed flow of clean carrier gas (back flushed)

7 What is Thermal Desorption? Compounds are released from the sorbent into the flow of carrier gas It is a simple extension of the technique of Gas Chromatography and is a sample introduction technology for difficult or real-world samples

8 2 Stage Thermal Desorption PROBLEM: Compounds are released SLOWLY from the sorbent tube Would lead to very wide chromatographic peaks and low sensitivity

9 2 Stage Thermal Desorption SOLUTION: Use a narrow secondary trap STAGE 1 Transfer compounds from tube to secondary trap Electrically cooled narrow bore cold trap

10 2 Stage Thermal Desorption STAGE 2 Rapid transfer of compounds from cold trap to GC Cold trap heated rapidly (100 C/sec) for sharp chromatographic peaks Backflush of cold trap for greater volatility range

11 Cold Trap Flow during trap desorption Flow during focusing / sampling onto trap Restricted narrow bore inlet / outlet end of trap Narrow design allows splitless injection Use sorbent(s) to suit specific application 15 standard traps + custom packed traps available from Markes

12 Key Component - Heated Valve The sample flow path through the valve assembly is uniformlyheated, short (< 10 cm total) and narrow to eliminate condensation Fused silica of column or transfer line inserted here Split filter or SecureTD-Q tube attaches here Sample tube connects here Inlet / outlet end of focusing trap connects here

13 Intensity Patented TD valve: All applications & sample re-collection Stage 1: Primary (tube) desorption with optional (inlet) split Heated valve To GC Time Patented heated valve is inert and low volume: Allows quantitative recovery of high & low volatility and reactive compounds The heated valve isolates the TD system allowing method compliance: leak testing, backflush trap desorption, purge to vent, overlap mode, etc.

14 Intensity Patented TD valve: All applications & sample re-collection Stage 2: Secondary (trap) desorption with optional (outlet) split Heated valve To GC Time Repeat analysis of re-collected samples makes it easy to validate analyte recovery through the TD flow path A change to the overall VOC profile indicates any bias

15 Using Re-collection (SecureTD-Q ) Validation of routine methods Demonstrating quantitative recovery of high boilers Repeat desorption of a mixed phthalate std di-ethyl- to di-n-decylphthalate Repeat analysis shows quantitative recovery without bias, across the analyte range Repeat 2 µl phthalate solution in methanol with 21:1 single split 20 re-collection, repeat analyses Good match between expected decay (lines) and observed decay (points) Demonstrates quantitative recovery Original DEHP DDP NB: ASTM Method D6196 references quantitative re-collection for validation

16 ULTRA 2 DiffLok TM caps Inert or Stainless steel outer body (shown transparent) Sample tube pushes into cap here Coned end for Auto-alignment with tube nozzle on ULTRA (the TD autosampler) O-ring for sealing tube into cap Threaded diffusion-locking insert in plain, cylindrical body. Schematic of DiffLok Cap as used to seal tubes on ULTRA Patented DiffLok caps provide a very effective seal for tubes on ULTRA 2 preventing artifact ingress and loss of analytes DiffLok caps allow gas to flow when pressure is applied & can stay on the tubes throughout TD operation This makes automatic TD simple and mechanically robust

17 Unique SafeLok sorbent tubes SafeLok tubes reduce risk of artefact ingress and loss of sampled analytes Threaded DiffLok inserts protect both ends of the sorbent tube They make it easier to handle tubes in the field (e.g. capping/uncapping) without risk of contamination They also facilitate collection and analysis of air samples from ultra-clean atmospheres. They allow pumped sampling at low flow rates over extended periods (< 1 ml/min) Same mass of sorbent and same external dimensions as standard tubes * Patent Numbers: GB , US 6, B1

18 TubeTAG : Tracking Air monitoring samples in transit A patented unique tool for tracking TD tubes in transit during field monitoring Consists of RFID tag assemblies for attaching onto sorbent tubes (various types and sizes), a read/write device (TAG SCRIBE ) and software Tags are re-usable indefinitely Patent # US 6.446,515 B2. Priority: 2002

19 The TubeTAG advantage for air monitoring Tube Information Sorbent packing # thermal cycles Impedance Leak test info Sample Information TubeTAG is a genuine breakthrough in TD technology. It allows the user to immediately identify the sorbent in each tube, when it needs repacking and whether or not it has a history of leak test failures, back-pressure anomalies, etc.

20 What can TD-GC/MS do? Any volatile or semi-volatile organic compounds which meet the following criteria: < n-c 40, bpt <525 C Can be easily gas chromatographed The sorbent or matrix containing the compounds is compatible with the high temperatures required Unsuitable compounds Inorganic compounds Most permanent gases exceptions include N 2 O, SF 6 & CS 2 Compounds with volatility > n-c 40 Compounds which don t work well with GC (including formaldehyde) Methane

21 Thermal Desorption systems UNITY 2 A universal platform for single tube thermal desorption TD-100 Automated TD system for up to 100 RFID-tagged or untagged tubes

22 Standard methods See TDTS 3: National and International Standard Methods Relating to Speciated Monitoring of Vapour-phase Organic Chemicals in Air

23 Online air monitoring applications CF 4 C 2 F 6 Extracted ion 69 C 2 F 6 SF 6 N 2 O Extracted ion 119 Extracted ion 127 Extracted ion 30 Acetylene up to 2L of sample taken without breakthrough Key: Blue (3 am) Red (12 noon) Green (5.30 pm) Green house gas analysis Diurnal profiling of ozone precursors

24 Canister analysis methods (e.g. US EPA TO-15) Method summary Grab sampling using canisters is easy, TWA monitoring is not Samples may be stored for up to 30 days A small volume of air from the canister (typically ~500 ml) can be introduced straight to the focusing trap of the desorber Trapping conditions are set such that water is selectively eliminated during the trapping process. Analysis by GC/MS CIA 8 UNITY 2 GC/MS Ambient Air

25 Air toxics in canisters: US EPA Method TO-15 1 L of a 1 ppb air toxics mix analysed splitless and cryogen-free using UNITY-CIA 8 Source: TDTS 81

26 Use canisters: Air monitoring: Canisters or tubes? 1. For non-polar compounds 2. Preferably at trace levels for ultravolatiles 3. When you have to Canister limitations 1. Expense ( each) 2. Poor recovery of anything higher boiling than Xylene 3. Cleaning needs expensive vacuum equipment, at least 3 cleaning cycles and verification with GC/MS

27 Canisters or tubes? Tubes Canisters Perception World-wide acceptance Gold standard for US ambient air market Applications Handling Ambient air, indoor air, vapor intrusion, industrial hygiene Material emissions Food & flavor Chemical weapons Light weight for personal monitoring and general ease of use Ambient air, indoor air, vapor intrusion, emergency response Larger and heavier; more costly to ship Sampling C 3 - C 40 Concentration range ppt to % C 2 -C 10 Concentration range ppt to low ppm Cleaning Analytical process automatically cleans tube for re-use Cost $50 $130 each $200 - $700 each Canister cleaning requires separate equipment as additional step prior to background certification and sampling. Profiles of soil gas contaminated with kerosene obtained using: a) canister sampling with TO-15 analysis (blue) b) sorbent tube sampling with TO-17 analysis (red)

28 Active (pumped) sampling TO-17 summary: Pre-screen area using 2 tubes in parallel (50 ml/min, 1 & 4 L) to identify type/level of VOCs. Confirm sorbent selection. Sample air Store capped tubes up to 30 days (below 4⁰C if multi-sorbent) TD-GC/MS analysis steps: Leak test then add gas phase IS and/or dry purge in sampling direction Sampling direction Weak Medium Strong Prepurge then desorb tube with carrier gas flowing in reverse direction Refocus analytes in electrically-cooled trap and desorb rapidly in backflush direction Analyse by GC/MS Pump Sampling Analysis Result

29 Air toxics on sorbent tubes: US EPA Method TO-17 10:1 Split desorption of Air Toxics tube loaded with 1 L of 1 ppb std Source: TDTS 86

30 Easy-VOC Pumped tube sampling made easy Grab sampling for sorbent tubes: Reliable sampling of 50 or 100 ml volumes (or multiples of same) Kit includes: hand pump, Safelok tubes and caps. Main features Ease of use Humidity effects Negligible Breakthrough? Minimised Use of SafeLok tubes and push on caps simplifies operation and prevents over tightening of storage caps no need for calibrated pumps Easy-VOC - For soil gas, workplace air & stack gas.

31 Comparisons of Easy-VOC with standard pumped sampling Comparison with FLEC constant flow pump (Black) and Hand pump (Red & Blue) equal performance

32 Extending the volatility range Smaller volume less chance of breakthrough, so lighter compounds can be retained. Isobutane Propane 2-butene

33 TD isn t just for trace levels: High-concentration industrial emission samples Pumped sampling of 1 L stack gas with TD-GC/MS analysis Sample splitting during both primary (tube) and secondary (trap) desorption. Total split ratio: 3000:1 Quantitative re-collection of both splits allows repeat analysis for confirmation

34 Passive (diffusive) sampling for workplace and environmental air monitoring Vapours migrate across the air gap at a constant uptake rate Diffusive sampling is a slow process, typically sample for complete shift (workplace/personal exposure) or days/weeks (environmental air monitoring) Brass cap Diffusion cap (fitted to sampling end of tube) Ambient conc. of vapours in the environment Zero conc. of vapours at the sorbent surface Note: Diffusive (passive) sampling won t work with glass or SafeLok tubes. Use standard stainless steel or Silcosteel tubes Air gap

35 Typical applications for air monitoring in industry Workplace air monitoring (Occupational hygiene) Industrial perimeter monitoring Stack and fugitive emissions Monitoring guidelines (sorbents, uptake rates, sampling volumes, etc.) available from standard methods (and from Markes)

36 Breath sampling for occupational hygiene (Bio-VOC ) Breath sampling is another application for TD-GC/MS It allows assessment of the total exposure inhalation, skin absorption, ingestion Key applications include: People working with chemicals absorbed by the skin People wearing protective equipment (How effective is it) Chronic exposure causing build up of chemicals in the body Environmental exposure studies Solvents in the breath of shoe workers

37 Specialist applications for air monitoring in industry Soil gas and vapour intrusion assessments Mapping criteria pollutants in ambient air, diurnal changes Atmospheric Research In situ monitoring of underground contamination

38 Specialist applications for air monitoring in industry Landfill gas monitoring Odorous industrial emissions Indoor air quality and tracer gases used for ventilation studies Biogenic emissions

39 Online monitoring of odorous sulphur compounds in ambient & industrial air (UNITY-Air Server ) Target compounds: H 2 S (hydrogen sulphide) CH 3 SH (methanethiol/me.mercaptan) C 2 H 6 S (dimethyl sulphide) C 2 H 6 S 2 (dimethyl disulphide) 3 channels with H 2 S focusing trap at - 15 C to -30ºC and flow path at 80ºC. GC-PFPD Performance in field operation: Detection limits: 0.15 ppb Retention time stability: <0.1% RSD across all compounds Gas standards at 20 ppb and 10 ppb levels Standard reproducibility: 0.5-5% RSD Recovery: >87% for all analytes

40 Input from Agilent? Sampling Thermal desorption GC/MS Data analysis Soil Gas samples MTS-32 no need for calibrated pumps Breath Samples Canisters and Bags Cryogen-free Method compliant 100 tubes Universal: C 2 to C 40 + reactive compounds Repeat analysis CFT (e.g. Backflush, Splitter, 2D-GC) Synchronous SIM/Scan 5 th generation of EPC EM Saver ChemStation, MassHunter RTL + DRS Libraries Sorbent Tubes Water management TubeTAG Normalized gain

41 Different Columns, Same GC Retention times vary from column-tocolumn by as much as 0.5 min... GC Column 1 Column 2 Column 3 Column 4

42 Retention Time Locking Retention times match from column-tocolumn AND instrument-to-instrument to min or better... GC1 Column 1 GC2 Column 2 GC3 Column 3 GC4 Column 4

43 Results for 5 Retention Time Locking Runs on GC/MS RTLOCK1.D -20 % pressure RTLOCK2.D -10 % pressure RTLOCK3.D nominal pressure RTLOCK4.D +10 % pressure RTLOCK5.D +20 % pressure

44 DRS - TICs of surface water extracts How many pesticides are in these samples?

45 Seventeen (17) surface water samples Targets Found (not counting ISTD) False Positives Processing Time CDFA* 37 1 ~8 hrs (ChemStation Only) Agilent DRS with 927 Compound Database Same more 0 34 min *CDFA is the California Department of Food and Agriculture Data files courtesy of Dr. Mark Lee and Steve Siegel

46 DRS = MSD + AMDIS + NIST Search MSD Chemstation RTLock and ID targets based on r.t. and 4 ions and quantitate AMDIS ID targets based on deconvoluted full spectra and qualify targets based on r.t. NIST Search search targets using deconvoluted full spectra from AMDIS against 191,000+ compounds Automate the above and produce an easy to read Report

47 What is AMDIS? Automatic Mass spectral Deconvolution and Identification Software Developed to automatically detect chemicals in violation of Chemical Weapons Convention Identify target compounds at low concentration levels in complex matrices Developed by National Institute of Standards and Technology Retention Time can be used as a qualifier, so RTL is a big advantage

48 A Single Chromatographic Peak May Contain Multiple Components TIC TIC & Spectrum Components and Mixed Spectra TIC Component 1 Component 2 Component 3

49 Components and Mixed Spectra AMDIS Deconvolution Deconvoluted components and spectra TIC matrix Component 1 Component 2 Component 3 interference Deconvolution target

50 DRS User Benefits Ease of use no need to learn deconvolution software Automation part of a method or part of a sequence, analyst can be doing other things Quality program results are not subject to changes in mood/attention of analyst Sensitivity AMDIS will find answers that an analyst might miss Confidence DRS will report the fewest false positive / negatives in the shortest time

51 DRS App Notes Available Comprehensive Pesticide Screening by GC/MSD using Deconvolution Reporting Software pub # New Tools for Rapid Pesticide Analysis in High Matrix Samples pub # A Blind Study of Pesticide Residues in Spiked and Unspiked Fruit Extracts Using Deconvolution Reporting Software pub # Building and Editing RTL Screener/Quant Databases and Libraries pub # Building Agilent GC/MSD Deconvolution Reporting Libraries for Any Application pub # (32 pages, so, use the videos!!!) Replacing Multiple 50-Minute FPD/ELCD/SIM Analyses with One 15-Minute Full-Scan Analysis for 10x Productivity Gain

52 DRS App Notes Available Screening for Hazardous Chemicals in Homeland Security and Environmental Samples Using GC/MS/ECD/FPD with a 731 Compound DRS Database pub # Improved Forensic Toxicology Screening Using A GC/MS/NPD System with a 725-Compound DRS Database pub # Semivolatiles Retention Time Locked (RTL) Deconvolution Databases for Agilent GC/MSD Systems pub # Screening for Pesticides in Food Using the Japanese Positive List Pesticide Method: Benefits of Using GC/MS with Deconvolution Reporting Software and a Retention Time Locked Mass Spectral Database pub # Can "Deconvolution" Improve GC/MS Detectability?

53 Commercial and RTL Databases

54 Free RTL Databases at RTL GC/MS Databases for Organotin Derivatives VOC MS Library and RTL Databases for GC and MS PCB Congener GC/MSD RTL Databases Forensic Toxicology GC/MSD RTL Databases Fatty Acid Methyl Ester GC and GC/MSD RTL Databases Flavors RTL Databases for GC/FID and GC/MS Allergens DRS Database for GC/MS Includes methyl, ethyl, and pentyl derivatives Includes the GC RT-locked database, the MS screener database, and full-spectra library of 65 volatile organic compounds Includes two RT-locked screener databases and MS library for 209 PCB congeners. Includes an RT-locked screener database and MS library for 277 drugs and other substances Includes RT-locked GC retention time databases and GC/MSD spectral libraries and screener databases for 37 common FAMEs Includes RT-locked GC/FID and GC/MS retention time databases and mass spectral library containing 409 compounds. Includes RT-locked GC/MS method, retention time databases and mass spectral library containing 24 regulated allergens and their isomers (total 32 target compounds and 1 internal standard), installation guide and related application note.

55 Who are monitoring Air Toxics? Contract laboratories Research laboratories and universities Government agencies in charge of environmental monitoring regulations Industry/manufacturing monitoring personal exposure Agencies involved in land remediation Health trusts interested in breath analysis for disease marker recognition

56 Summary Whichever vapour-phase organic chemicals are of interest Whatever your air monitoring application Whether it s VOCs in whale breath or the most polluted air on the planet Photograph courtesy of: Dr. Rei Rasmussen, Oregon Graduate Institute, USA

57 Further information Markes website Thermal Desorption Application notes Brochures Environmental applications guide

58 TDTS 16 Analysis of ozone precursors TDTS 80 Evaluation of a Soil Gas sorbent tube for improving the measurement of volatile and semi-volatile fuel vapors in soil contaminated land TDTS 81 EPA TO-15 canister analysis TDTS 86 EPA TO-17 tube analysis Further information TDTS 87 Analysis of Ultra volatile GHGs TDTS 78 Low level odorants in drinking water TDTS 29 VOC air monitoring technology and its application to contaminated land TDTS 47 The analysis of landfill gas compounds using thermal desorption GC/MS and a RTL Database TDTS 49 Fence-line and ambient monitoring of benzene and other hydrocarbons using diffusive sampling and thermal desorption analysis TDTS 37 Industrial air monitoring using pumped sampling onto sorbent tubes and the UNITY thermal desorber TDTS 77 Using thermal desorption for industrial (stack) emission testing

59 Thank you for your attention! Questions???

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist

Innovative Thermal Desorption From Agilent Technologies. Vivek Dhyani Application Chemist Innovative Thermal Desorption From Agilent Technologies Vivek Dhyani Application Chemist What is thermal desorption (TD)? How does thermal desorption compare to Head Space? Thermal desorption (TD) is the

More information

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software NEMC, Washington D.C. August 6 th 2012 Nicola Watson Environmental Specialist Markes International

More information

Application Note 032

Application Note 032 Application Note 3 Analysis of Sulfur Compounds Using On-line and Off line TD GC Application Note Abstract This application note shows how Markes TD technology is compatible with trace level sulfur compounds,

More information

Thermal Desorption Technical Support

Thermal Desorption Technical Support Thermal Desorption Technical Support Note 47: The Analysis of Landfill Gas Compounds using Thermal Desorption GC/MS and a Retention Time Locked Database Key Words: VOCs, directive, compliance, identifi

More information

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ Application Note Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ -3 Application Note Abstract This application note demonstrates the

More information

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs Vapor Intrusion Sampling Options: Performance Data for s, Badges, and Sorbent Tubes for VOCs Linda S. Coyne SKC Inc., 863 Valley View Road, Eighty Four, PA 1533 George Havalias, Maria C. Echarte American

More information

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated Lee Marotta, Field Application Scientist, PerkinElmer Miles Snow, Research Scientist, PerkinElmer

More information

TDTS. Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination

TDTS.   Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination Thermal Desorption Technical Support Note 75: Liquid standard injection, tube impedance and other factors which may cause discrimination during the calibration of thermal desorption methods Keywords TD

More information

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application Food Safety, Environmental Authors Chin-Kai Meng and Bruce Quimby Agilent Technologies, Inc. 2850 Centerville

More information

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc. Optimizing Sampling and Analytical Parameters for Soil Vapour Samples using Automated Thermal Desorption / Gas Chromatography / Mass Spectrometry (ATD/GC/MS) Stephen Varisco, Technical Manager, CARO Analytical

More information

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 77A mini TD and 70A GC Application Note Environmental Authors Tingting Bu, Xiaohua Li Agilent Technologies (Shanghai) Co.,

More information

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Replaces: Dated: Author: Date: AM-No.: New New Nils Arne Jentoft 18.06.2014 0 CHANGES This procedure is new. 1 SCOPE This document describes

More information

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer Application Note By: Anne Jurek Abstract This application note presents

More information

Series 2 UNITY A universal TD platform for tagged and untagged tubes

Series 2 UNITY A universal TD platform for tagged and untagged tubes Series 2 UNITY Thermal desorber A universal TD platform for tagged and untagged tubes Series 2 UNITY Thermal Desorption GC(MS) Thermal desorption (TD) is a highly versatile, sensitive and labour-saving

More information

Raw Material Finished product By product Exposure standards Length of shift Physical environment

Raw Material Finished product By product Exposure standards Length of shift Physical environment Measurement &Sampling Part 1 Introduction to Sampling What to sample? - Hazard Raw Material Finished product By product Exposure standards Length of shift Recognition Physical environment To Determine

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Automated Handling Techniques for the Analysis of Elevated Volatile Organic Compound (VOC) Concentrations in Soils Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek

More information

TD-100. Automated thermal desorber. Universal TD system for up to 100 RFID-tagged or untagged tubes

TD-100. Automated thermal desorber. Universal TD system for up to 100 RFID-tagged or untagged tubes TD-100 Automated thermal desorber Universal TD system for up to 100 RFID-tagged or untagged tubes TD-100 Simply the best available dedicated system for automated tube desorption Thermal desorption (TD)

More information

TO-15 Checklist Determination of VOCs in Air by GC-MS

TO-15 Checklist Determination of VOCs in Air by GC-MS LAB ID: DATE: LAB NAME: ASSESSOR NAME: Method Number: TO-15 Checklist Determination of VOCs in Air by GC-MS SOP Number: Revision Number: SOP Date: Personnel records observed: Data records observed: Revision

More information

Real Time On-Site Odor and VOC Emission Measurements Using a znose

Real Time On-Site Odor and VOC Emission Measurements Using a znose Real Time On-Site Odor and VOC Emission Measurements Using a znose Edward J. Staples, Electronic Sensor Technology, EST@ESTCAL.COM Remediation Site Description Remediation of contaminated soil from where

More information

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Chemistry Instrumental Analysis Lecture 27. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 27 Gas Chromatography Introduction GC covers all chromatographic methods in which the mobile phase is gas. It may involve either a solid stationary phase (GSC)

More information

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which

More information

The Application of Method TO-15 to Naphthalene Measurements in Indoor Air

The Application of Method TO-15 to Naphthalene Measurements in Indoor Air The Application of Method TO-15 to Naphthalene Measurements in Indoor Air Extended Abstract #13 Heidi C. Hayes and Diane J. Benton Air Toxics Ltd. 180 Blue Ravine Rd. Ste. B Folsom, CA 95630 INTRODUCTION

More information

Thermal Desorption Technical Support

Thermal Desorption Technical Support Thermal Desorption Technical Support Note 86a: US EPA Method TO-17 for Monitoring Air Toxics in Ambient Air Using Sorbent Tubes and Automated, Cryogen-free Thermal Desorption Application Note Environmental,

More information

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS International Gas Union Research Conference 14 THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS Main author Hironori IMANISHI Tokyo Gas Co., Ltd. JAPAN himanishi@tokyo-.co.jp

More information

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS Application Note Food Authors Nobuo Ochiai and Kikuo Sasamoto GERSTEL K.K. 2-13-18 Nakane

More information

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds Atmospheric Analysis Gases Sampling and analysis of gaseous compounds Introduction - External environment (ambient air) ; global warming, acid rain, introduction of pollutants, etc - Internal environment

More information

Highly Sensitive and Rugged GC/MS/MS Tool

Highly Sensitive and Rugged GC/MS/MS Tool Highly Sensitive and Rugged GC/MS/MS Tool For Pesticide Multiresidue Analysis in Food Samples Agilent 7 Series Triple Quadrupole GC/MS. The world s first MS/MS designed specifically for GC Analysis Introduction

More information

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method

More information

Thermal Desorption Technical Support. Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing.

Thermal Desorption Technical Support. Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing. Thermal Desorption Technical Support Note 56: TD-GC/MS analysis of vapour-phase organic compounds for materials emissions testing Keywords certification protocols, (T)(V)(S)VOCs, ULTRA-UNITY, field/laboratory

More information

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES JPACSM 127 AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES Trace Analytical Inc. Menlo Park, CA ABSTRACT GC based gas analyzers with Reduction Gas Detector (RGD) and Flame Ionization

More information

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting Application Note Authors Angela Smith Henry and Bruce Quimby Agilent

More information

Agilent 5977A Series GC/MSD System

Agilent 5977A Series GC/MSD System Agilent 5977A Series GC/MSD System Data Sheet 5977A GC/MSD Overview The Agilent 5977A Series GC/MSD builds on a 45-year tradition of leadership and innovation, bringing together the technologies of the

More information

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract. Application Note 8 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-5 Application Note Environmental, Canister, TO-5, Air Monitoring, Air Toxics Abstract This application

More information

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical INNOVATIVE PRODUCTS, SUPERIOR SUPPORT Presenter: Anne Jurek, Senior Applications Chemist,, pp, EST Analytical Air pollution is a growing problem due to the global economy and industrial development in

More information

Series 2. Continuous monitoring of trace chemical vapours

Series 2. Continuous monitoring of trace chemical vapours Series 2 TT24-7 Continuous monitoring of trace chemical vapours Series 2 TT24-7 TT24-7 Series 2 Continuous monitoring of trace chemical vapours The TT24-7 Series 2 is an advanced thermal desorption (TD)

More information

TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air

TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air Application TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air Summary This Application Note describes validation of a thermal

More information

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Sample Throughput In Purge And Trap Analysis Maximizing Sample Throughput In Purge And Trap Analysis LINDSEY PYRON ANNE JUREK INTRODUCTION There are several demands and requirements imposed on chemists performing volatile organic analysis (VOC) in

More information

Series 2 Air Server Automated canister and on-line air/gas analysis

Series 2 Air Server Automated canister and on-line air/gas analysis Series 2 Air Server Thermal Desorption Options Automated canister and on-line air/gas analysis Series 2 Air Server A range of accessories for series 2 UNITY thermal desorbers offering automated canister

More information

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry The end of mass-speculation Pre-cleaned and certified vials for mass spectrometry Bonded, ultra-pure closures LC/MS and GC/MS Manufacturing the Ultimate Vial for Mass Spectrometry More than a quarter century

More information

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible

More information

Air Monitoring Sorbent Pens. For Environmental and IH Applications

Air Monitoring Sorbent Pens. For Environmental and IH Applications Air Monitoring Sorbent Pens For Environmental and IH Applications Sorbent Pens for Quantitative Air Monitoring Sorbent Pens mark the latest innovation in quantitative air monitoring from Entech Instruments.

More information

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek Requirements of an Automated Sample Delivery System in Today s Realm of Ever Increasing Sensitivity Demands Utilizing the Atomx Concentrator/Multimatrix Autosampler. Application Note By: Anne Jurek Abstract

More information

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation Introduction Fire investigation involves many different types of analyses from crime scene

More information

Useful GC/MS Sensitivity for Improved Environmental Analyses. GC/MS Agilent Restricted

Useful GC/MS Sensitivity for Improved Environmental Analyses. GC/MS Agilent Restricted Useful Sensitivity for Improved Environmental Analyses Page 1 Definitions Sensitivity Analytical definition: slope of the plot of amount of analyte (x-axis) versus signal response (y-axis) Commonly used

More information

Volatile Organic Compounds in Every Day Food

Volatile Organic Compounds in Every Day Food Volatile Organic Compounds in Every Day Food Application Note Abstract Volatile Organic Compounds (VOCs) have been analyzed in drinking water and pharmaceuticals extensively in recent years. Now, VOCs

More information

Understanding Gas Chromatography

Understanding Gas Chromatography Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X ?? K? Page 2 Typical GC System Gas supply Injector

More information

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK Multi-residue GC-MS analysis Richard Fussell CSL York, UK e-mail: r.fussell@csl.gov.uk Outline of Presentation Background Gas Chromatography Mass Spectrometry - Options Capabilities and Applications Summary

More information

Optimizing Conditions for Volatiles Analysis

Optimizing Conditions for Volatiles Analysis July 2016 SPEXCertiPrep.com/ECS Optimizing Conditions for Volatiles Analysis by Mark A. Ferry GC/MS Technical Consultant MFerry@SPEX.com A publication by Environmental Consulting & Supplies a Division

More information

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Page 1 of 13 CIPAC/4105/R GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY Prepared for CIPAC by Dr M J Tandy*, P M Clarke and B White (UK) The rapid

More information

Introduction to IH Analytical Chemistry

Introduction to IH Analytical Chemistry Introduction to IH Analytical Chemistry Sampling and Analysis Manning, Eide, Van Etten (Apr 2013) Foreword Thank you for purchasing the Introduction to IH Analytical Chemistry self-study workbook. Course

More information

Headspace Technology for GC and GC/MS: Features, benefits & applications

Headspace Technology for GC and GC/MS: Features, benefits & applications Headspace Technology for GC and GC/MS: Features, benefits & applications Karima Baudin Oct 2015 Why use Headspace? Very Simple no to minimum sample prep Robust enhance uptime Non-detectable carry-over

More information

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003 ALCOHOLS IV 1403 (1) HOCH 2 CH 2 OCH 3 MW: (1) 76.09 CAS: (1) 109-86-4 RTECS: (1) KL5775000 (2) HOCH 2 CH 2 OCH 2 CH 3 (2) 90.12 (2) 110-80-5 (2) KK8050000 (3) HOCH 2 CH 2 O(CH 2 ) 3 CH 3 (3) 118.17 (3)

More information

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

Fast and Accurate EPA 8270 Quantitation Using Deconvolution Fast and Accurate EPA 8270 Quantitation Using Deconvolution Application Note Environmental Authors Mike Szelewski and Chinkai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808

More information

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Application Note Environmental Author Suli Zhao, Andy Zhai Agilent Technologies Co., Ltd. 412 Yinglun Road

More information

Determination of Volatile Substances Proof of Food Adulteration

Determination of Volatile Substances Proof of Food Adulteration ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of Volatile Substances Proof of Food Adulteration (method: gas chromatography with mass spectrometric detection) Exercise guarantor:

More information

HYDROCARBONS, AROMATIC 1501

HYDROCARBONS, AROMATIC 1501 HYDROCARBONS, AROMATIC 1501 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1501, Issue 3 EVALUATION: Full Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table 2

More information

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No Validation of Tetrahydrofuran Using SKC Passive Sampler Cat. No. 575-002 Page 1 of 10 Publication 1841 Issue 1407 Research Report Validation of Tetrahydrofuran Using the SKC Cat. No. 575-002 Passive Sampler

More information

CIA Advantage-xr. Cryogen-free automated canister and whole-air sampling system

CIA Advantage-xr. Cryogen-free automated canister and whole-air sampling system CIA Advantage-xr Cryogen-free automated canister and whole-air sampling system CIA Advantage-xr Introducing the CIA Advantage-xr a cryogen-free system for the automated GC and GC MS analysis of trace-level

More information

ChromTech options for PAL systems

ChromTech options for PAL systems ChromTech options for PAL systems SPDE features full automation for dynamic sample extraction, desorption and analysis using the CTC Combi PAL. Most of our customers use SPDE for food and flavor research,

More information

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS Application Note Optimization of Method Parameters for the Evaluation of USEPA Method 524.2 Using a Purge and Trap with GC/MS Objective By: Glynda Smith In this paper, an evaluation of different methods

More information

high performance OPTIC 3 and applications a total analytical solution

high performance OPTIC 3 and applications a total analytical solution high performance OPTIC 3 and applications a total analytical solution Excellent by nature Excellent effectivity, efficient performance and great flexibility. OPTIC 3: a fully programmable injector for

More information

Dynatherm Thermal Desorption Systems

Dynatherm Thermal Desorption Systems Leaders In Sample Introduction Technology Dynatherm Thermal Desorption Systems A complete suite of sample collection and concentration systems FA S T RELIABLE FLEXIBLE UPGRADEABLE CDS Analytical is the

More information

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS Application Note #1820230 Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS GC-MS Introduction USEPA Method 8270 [1] for semivolatiles is is used by by laboratories to measure

More information

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column Analyze Hydrocarbon Impurities in,3-butadiene with an Agilent J&W GS-Alua PT Column Application Note Energy and Chemicals Authors Yun Zou and Chunxiao Wang Agilent Technologies Shanghai Ltd Abstract The

More information

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES Mihriban Yılmaz Civan, Öznur Kuntasal and Gürdal Tuncel 1 1 OUTLINE Introduction Importance and

More information

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY AppNote 6/2004 Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons Edward A. Pfannkoch, Jacqueline A. Whitecavage, and Jeffrey Christenson Gerstel, Inc., 701 Digital Drive,

More information

METHOD 3600B CLEANUP

METHOD 3600B CLEANUP METHOD 3600B CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Method 3600 provides general guidance on selection of cleanup methods that are appropriate for the target analytes of interest. Cleanup methods are applied

More information

SELECTED ION FLOW TUBE MASS SPECTROMETRY. Interscience Expert Center: VIP Meeting 15 Sept. 2009

SELECTED ION FLOW TUBE MASS SPECTROMETRY. Interscience Expert Center: VIP Meeting 15 Sept. 2009 SELECTED ION FLOW TUBE MASS SPECTROMETRY Interscience Expert Center: VIP Meeting 15 Sept. 2009 PROGRAM 14.00 u: The IS. X strategy 14.15 u: What exactly is SIFT-MS? 14.30 u: Developing methods on the Voice

More information

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES J Kouzaki 1*, S Sato 1, S Nakai 1, Y Shirasuna 2, K Hirano 2 1 Graduate School of Environmental and Information

More information

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS PAGE: 1 of 14 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY CONTENTS 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors Trace-Level Analysis of Melamine in Milk Products on Agilent 789A/5975C GC/MSD Using a ew Agilent J&W DB-5ms Ultra Inert Column and SampliQ SCX Cartridges Application ote Gas Chromatography/Mass Spectrometry/Food

More information

Application Note S/SL

Application Note S/SL Analysis of Ultra low Sulfur Compounds in Natural Gas and Gaseous Fuels by Gas Chromatography and Chemiluminescence according to ASTM D Ultra low detection limits Excellent Sensitivity, Repeatability &

More information

Rapid Catalyst Screening Reactors

Rapid Catalyst Screening Reactors Rapid Catalyst Screening Reactors Rapid screening of catalysts Analysis of a variety of sample types Multi-modes of operation Rx-3050TR Single µ-reactor Rx-3050SR Overview Two types of rapid catalyst screening

More information

A Micro-GC Based Chemical Analysis System

A Micro-GC Based Chemical Analysis System A Micro-GC Based Chemical Analysis System D E F I ANT TECHNL G IES B TH #25 1 8 PATRICK L E W I S, P R L E W IS@DEFIANT - TEC H. C M, 50 5-3 0 7-3576 D UG ADKINS, PHD, G E RGE D U L L E C K, G ARY F U

More information

Petrochemical. Transformer Oil Gas Analysis - TOGA

Petrochemical. Transformer Oil Gas Analysis - TOGA Petrochemical Transformer Oil Gas Analysis - TOGA www.dps-instruments.com The DPS TOGA GC Systems are designed to analyze oil from electrical insulation materials that may have decomposed under thermal,

More information

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application Environmental Analysis Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville

More information

Analytical Trap Comparison for USEPA Method 8260C

Analytical Trap Comparison for USEPA Method 8260C Analytical Trap Comparison for USEPA Method 8260C Application Note Abstract Purge and trap concentration is a technique that is used for the analysis of Volatile Organic Compounds (VOCs). The major component

More information

ELECTRONICALLY REPRINTED FROM. Volume 34 Number 3 MARCH

ELECTRONICALLY REPRINTED FROM. Volume 34 Number 3 MARCH ELECTRONICALLY REPRINTED FROM Volume 34 Number 3 MARCH 2016 2016 www.chromatographyonline.com Extending the Hydrocarbon Range for the Analysis of Soil Gas Samples Using Automated Thermal Desorption Coupled

More information

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES CONTENTS 1. An overview of SIFT-MS 2. Principles of SIFT-MS a. Why chemical ionization? b. Why multiple reagent

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No. G12871B_02

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No. G12871B_02 Polyseam A/S Ravneveien 7 Linnestad Næringsområde 3174 Revetal Norway Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing Date 09 October 2014

More information

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Analysis of Trace (mg/kg) Thiophene in Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application Petrochemical Authors James D. McCurry and Bruce D. Quimby Agilent Technologies

More information

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation Technical Overview Introduction Inert flow path technology Modern GC and GC/MS instrumentation is an important analytical tool

More information

Gas Chromatography (GC)

Gas Chromatography (GC) Gas Chromatography (GC) Ahmad Aqel Ifseisi Assistant Professor of Analytical Chemistry College of Science, Department of Chemistry King Saud University P.O. Box 2455 Riyadh 11541 Saudi Arabia Office: AA53

More information

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application Static Headspace Blood Alcohol Analysis with the G Network Headspace Sampler Application Forensics Author Roger L. Firor and Chin-Kai Meng Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-0

More information

Thermal Desorption: A Practical Applications Guide

Thermal Desorption: A Practical Applications Guide www.markes.com Thermal Desorption: A Practical Applications Guide III. Defence and Forensic Introduction to Markes International Ltd. Formed in 1997, Markes International Ltd. is one of the world s leading

More information

High-Pressure Volumetric Analyzer

High-Pressure Volumetric Analyzer High-Pressure Volumetric Analyzer High-Pressure Volumetric Analysis HPVA II Benefits Dual free-space measurement for accurate isotherm data Free space can be measured or entered Correction for non-ideality

More information

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface Application Note No. 033 Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface Ryoichi Sasano*, Takayuki

More information

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Application Note Environmental Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Authors Ye Kong and Zhe Cao Agilent Technologies,

More information

Test Report in accordance CDPH-IAQ

Test Report in accordance CDPH-IAQ Hilti Entwicklungsgesellschaft GmbH Hiltistraße 6 86916 Kaufering Germany Test Report in accordance CDPH-IAQ Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark voc@eurofins.com www.eurofins.com/voc-testing

More information

We don t have anything against mass spectrometry. We just think it s time for a worthy alternative.

We don t have anything against mass spectrometry. We just think it s time for a worthy alternative. We don t have anything against mass spectrometry. We just think it s time for a worthy alternative. gas chromatography detector Everything you need in one GC detector Universal and selective detector with

More information

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology Gas Chromatography/Mass Spectrometry Author Harry Prest 1601 California Avenue

More information

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note Environmental Author Anne Jurek Applications

More information

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note. Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID Application Note Volatile Organic Compounds (VOCs) Author Andrew Jones Activated

More information

Analytical Instrumentation

Analytical Instrumentation Gas Chromatography and Mass Spectrophotometry Environmental, food safety, toxicology, and forensic gas chromatography applications are more demanding than ever. Our portfolio of GC, GC- MS, GC-MS/MS and

More information

SIFT-MS. technology overview

SIFT-MS. technology overview SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY technology overview contents 1. An overview of SIFT-MS 2. Principles of SIFT-MS a. Why chemical ionization? b. Why three reagent ions? c. How does SIFT-MS

More information

Test Report. Glava AS. Emission test of Ecophon Venus according to M1 classification. October / November Smedeskovvej 38, DK-8464 Galten

Test Report. Glava AS. Emission test of Ecophon Venus according to M1 classification. October / November Smedeskovvej 38, DK-8464 Galten Test Report Glava AS Emission test of Ecophon Venus according to M1 classification October / November 2006 Client: Glava AS Nybråtveien 2 1832 Askim Norge Date: 8 th of November 2006 Testing Laboratory:

More information

Benzene Contamination in BabyFood and Beverages by New Generation of Static Headspace Autosampler coupled to Fast GC-TOFMS

Benzene Contamination in BabyFood and Beverages by New Generation of Static Headspace Autosampler coupled to Fast GC-TOFMS Benzene Contamination in BabyFood and Beverages by New Generation of Static Headspace Autosampler coupled to Fast GC-TOFMS Authors: Daniela Cavagnino DANI Instruments SpA viale Brianza, 87 Cologno Monzese

More information

COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation

COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation SAMPLE DESCRIPTION & TESTING PARAMETERS Tesoro Woods submitted exemplars of their

More information

VOC TEST REPORT M1. 23 October Regulation or protocol Conclusion Version of regulation or protocol

VOC TEST REPORT M1. 23 October Regulation or protocol Conclusion Version of regulation or protocol Würth Oy Würthintie 1 11710 Riihimäki Finland Eurofins Product Testing A/S Smedeskovvej 38 8464 Galten Denmark VOC@eurofins.com www.eurofins.com/voc-testing VOC TEST REPORT M1 23 October 2017 1 Sample

More information