CONSISTOMER OIL AND HEAT TRANSFER FLUID IS CURRENTLY UNDER DISCUSSION ON API MONOGRAM PROGRAM COMMITTEE. Spec 10 A and RP10B2

Size: px
Start display at page:

Download "CONSISTOMER OIL AND HEAT TRANSFER FLUID IS CURRENTLY UNDER DISCUSSION ON API MONOGRAM PROGRAM COMMITTEE. Spec 10 A and RP10B2"

Transcription

1 PLEASE FIND BELOW VARIOUS COMMENTS RECEIVED SINCE APRIL 2013 WHEN API RP 10B 2 WAS RELEASED. OTHER PROBLEMS OR ISSUES THAT HAVE BEEN UNCOVERED SINCE THE RELEASE ARE: CONSISTOMER OIL AND HEAT TRANSFER FLUID IS CURRENTLY UNDER DISCUSSION ON API MONOGRAM PROGRAM COMMITTEE. Spec 10 A and RP10B2 FLUID LOSS SCREEN REPLACEMENT ON STATIC FLUID LOSS TESTS OTHER ISSUES UNCOVERED IN Spec 10A that apply to 10B 2 OTHER ISSUES NOT YET DOCUMENTED OR PSSIBLE MISSED, ===================================================================================== DETERMINING GEL STRNGTH TOM DEALY API RP 10B, Twenty Second Edition, 1997 gives the following instructions for determining 10 sec and 10 min gel strength for cements: 12.7 Procedure for determination of gel strength The gel strength of a cement may be measured immediately after determining the rheological properties of the slurry sample, or as an independent observation. If increasing slurry gelation is observed during the rheological measurements a brief recondition of the slurry in the viscometer for 1 minute at 300 RPM may disperse the gels and allow better measurement of the gel strength. For all independent tests, the slurry should be prepared and loaded into the viscometer as outlined in through Shut off the viscometer for 10 seconds and record the slurry temperature Set the viscometer at the speed equivalent to 5.1 1/s and start rotation. Record the maximum observed reading immediately after turning on the instrument. Use this reading to calculate the 10 second gel strength per equation 20. Equation 20: Τ(lb/100 sq ft) = x F x θ Where: Τ =The shear stress in lb/100 ft 2 F = The instrument s torsion spring factor Θ = The viscometer reading in instrument degrees

2 Shut of the viscometer for 10 minutes and record the slurry temperature. Repeat the measurements as in to report the 10 minute gel strength After taking the readings, the temperature of the slurry in the viscometer cup should again be recorded The slurry gel strengths should be reported at the average of the recorded temperatures For improved reliability of the measurements, the entire procedure may be repeated several times using a freshly prepared sample of slurry each time. If the procedure is repeated several times, the gel strength values should then be reported as the average of all the acceptable measurements. At this time, there were no instructions for determining any gel strength tests at times other than 10 seconds and 10 minutes and the conditioning at 100 rpm was only performed before the 10 second gel strength determination. As Tom mentioned a 10 second gel strength measurement would not interfere with the 10 minute gel strength substantially so it was not reconditioned after the 10 second gel strength. In API RP 10B 2, First Edition, July 2005, the wording is basically unchanged. However, in the current version of API RP 10B 2, Second Edition, April 2013; it now contains information about performing the static gel strength at other times INSTEAD of 10 MINUTES. If another time is used (such as 30 or 60 minutes) the 10 second gel strength is not performed on the same slurry according to these procedures. If several gel strengths are desired, other tests will need to be performed for each static gel strength test. The rheological readings do not have to be performed on these other gel strength times, only condition the slurry at 300 rpm for 1 minute and then perform the 10 second and gel strength at the other time. It does not mention anything about determining several times on the same slurry Determination of Gel Strength The gel strength of a fluid may be measured immediately after determining the rheological properties of the sample or on a separate, freshly-prepared fluid. NOTE Gel strengths of a fluid are the measurements resulting of the application of the rheological property procedure described under These measurements are not comparable to static gel strength of a fluid measured under API 10B-6 or ISO and shall not be used for critical static gel strength determination Recondition the fluid in the viscometer for 1 min at 300 r/min to disperse the gels and allow better measurement of the gel strength. For tests on separate, freshly-prepared fluids, prepare, condition, and load the fluid into the viscometer as outlined in to and then condition the fluid for 1 min at 300 r/min Stop rotation of the rotor Set the viscometer at the speed equivalent to 5.1 s 1 (3 r/min with R1-B1 or 6 r/min with R1-B5) s after stopping the rotor, start rotation at 5.1 s 1 and record the maximum deflection of the dial observed immediately after starting rotation. Calculate the 10 s gel strength by multiplying the measured value by the factor for the applicable rotor, bob, and spring configuration found in Table Record the fluid temperature, then stop the viscometer for 10 min.

3 Start rotation at 5.1 s 1 10 min after stopping the rotor and record the maximum deflection of the dial observed immediately after starting rotation. Calculate the 10 min gel strength by multiplying the measured value by the factor for the applicable rotor, bob, and spring configuration found in Table Gel strength measurements may be taken at other static times, as desired. For other static times, follow the procedure of and , substituting the desired static time for 10 min. Dennis Comment: To me, indicates that only the 10 second and 10 minute time are performed on the same slurry. If a 30 minute gel strength is desired, the 30 minute gel strength can be substituted for the 10 minute gel strength (10 minute gel strength not performed) or a new slurry mixed if additional times are desired, the 10 second and the other desired gel strength time can be performed, but not the 10 minute plus another time After taking the maximum deflection dial reading, again record the temperature of the fluid in the viscometer cup Report the fluid gel strengths at the recorded temperatures (note the rotational velocity, the rotor/bob configuration, and spring factor used in the test) For improved reliability of the measurements, the entire procedure may be repeated two times or three times using freshly prepared fluid each time. Report the gel strength values as the average of the measurements. The drilling fluid API RP 13B 1 and is for drilling fluids. For drilling fluids, they do re stir the fluid for 10 seconds (instead of1 minute) at 600 rpm (not at 300 rpm but) before the 10 second static period and before the 10 minute static period Procedure Place the sample in a container and immerse the rotor sleeve exactly to the scribed line. Measurements in the field should be made with minimum delay (within 5 min, if possible) and at a temperature as near as practical to that of the drilling fluid at the place of sampling, but not differing by more than 6 C (10 F). The place of sampling should be stated on the test report. DANGER Maximum recommended operating temperature is 90 C (200 F). If it is necessary to test fluids above this temperature, a solid metal bob or a hollow metal bob with a completely dry interior should be used. Liquid trapped inside a hollow bob can vaporize when immersed in a high-temperature fluid and cause the bob to explode Record the temperature of the sample With the sleeve rotating at 600 r/min, wait for viscometer dial reading to reach a steady value (the time required is dependent on the drilling-fluid characteristics). Record the dial reading for 600 r/min Reduce the rotor speed to 300 r/min and wait for the viscometer dial reading to reach a steady value. Record the dial reading for 300 r/min Stir drilling fluid sample for 10 s at 600 r/min. Stop the rotor Allow drilling fluid sample to stand undisturbed for 10 s. Slowly and steadily turn the hand-wheel in the appropriate direction to produce a positive dial reading. The maximum reading is the initial gel strength. For instruments having a speed of 3 r/min, the maximum reading attained after starting rotation at 3 r/min is the initial gel strength. Record the initial gel strength (10-second gel) in pounds per 100 ft2.

4 NOTE To convert the dial reading to pascals: 1 Pa 0,511 lb/100 ft Re-stir the drilling fluid sample at 600 r/min for 10 s, stop the rotor and allow the drilling fluid to stand undisturbed for 10 min. Repeat the measurements as in and report the maximum reading as the 10-minute gel in pounds per 100 ft2. NOTE To convert the dial reading to pascals: 1 Pa 0,511 lb/100 ft2. Below are a few thoughts from Dennis Gray: 1. I do not see any place in API RP 10B 2 or earlier versions that the slurry is re stirred at 300 rpm after taking the 10 second gel strength indication or that more than two gel strength times are recommended to be performed on the same slurry. a. My thought would be if a second gel strength time is to be performed (in addition to the 10 second gel strength) it would be performed on a newly mixed slurry as the current API RP 10B 2 suggests. 2. Cements and drilling fluids have different properties. 3. The drilling fluids are probably more stable and have less of a tendency to settle or centrifuge at 600 rpm as they are conditioned for 10 seconds. 4. Drilling fluids do not set over time, but cements begin reacting as soon as they are mixed. 5. Allowing for multiple gel strength readings from the same slurry may allow the gel strength to get into a portion of the hydration period which will not be a gel strength, but an actual setting indication. a. This may be the reason cement testing does not have the slurry conditioned at 300 rpm after determining the 10 second gel strength and before determining the 10 minute gel strength (or other at another time). 6. If the longer static gel strength tests are to try to reproduce the SGS measurement performed in a MACS II, there will not be an exact correlation because of several things and some of them are: a. Although the slurry is preconditioned in an atmospheric or HPHT consistometer, the slurry must be static for a while during transferring from the conditioning unit to the Fann 35 and the test is not performed at higher temperatures and pressures. 7. Drilling fluids do not request gel strength tests be performed past the 10 minute gel strength. 8. I don t have records of API gel strength testing for drilling fluids, but suspect they have been around for a much longer time than the ones for cement. 9. Actually the 10 minute gel strength is performed 10 minutes after determining the 10 second gel strength, which will be about 10 minutes and 15 seconds for cements and about 10 minutes and 25 seconds for drilling fluids. 10. I am not sure why cement is re stirred for 1 minute and drilling fluids for only 10 seconds. 1 minute seems like it would give the slurry a chance to centrifuge out or settle and offset the gel strength measurement. If you set the timer for 30 minutes in step 9, the fluid would be sitting static for 40 total minutes before taking a reading. Therefore, you would have to pre stir the slurry at 300 RPM for 1 minute to break the gel strength before allowing it to sit static for 30 minutes.

5 The current API RP 10B 2 instructs the operator to substitute 30 minutes for 10 sec and 10 min, which is ambiguous. Initiating a global lab best practice to pre stir the slurry at 300 RPM for 1 minute prior to beginning all static periods would remove the ambiguity. Dennis Gray will be back in the office on Tuesday, November 5. I am copying him for his comments. I am also copying Robert Darbe for his comments. Robert has served as API Committee 10 Secretary for several years and is possibly most familiar with discussions on procedures. I do not know of any testing done to compare one method vs. the other, but it would be very useful. The question would be what effect does pre stirring before each static period have on final results. Please advise if the following could be classified as the correct procedure 1. Finish API rheologies 2. Complete a 1 min condition at 300 RPM (this is what recommended in API 10B for reproducibility) 3. Set rheometer to 3RPM before shutting off, simultaneously at shutoff start 10 sec timer. 4. After 10sec, turn rheometer back on and record the torque value at highest deflection point. 5. Recondition for 1 min at 300 RPM. 6. Set rheometer to 3RPM before shutting off, simultaneously at shutoff start 10 min timer. 7. After 10min, turn rheometer back on and record the torque value at highest deflection point. 8. If running another SGS test, recondition for 1 min at 300 RPM. 9. Set rheometer to 3RPM before shutting off, simultaneously at shutoff start 30 min timer. 10. After 30min, turn rheometer back on and record the torque value at highest deflection point. This will give me 10 sec, 10 min and 30 min readings. API does not specify 300 rpm conditioning between 10 sec, 10 min, 30 min, etc. readings. Below is the procedure from the latest edition, April 2013: MIXER RPM =/= ALAN DEAN SEE FOLLOWING PDF FROM MIXING As attached, the April 2013 Edition has an anomaly in the mixer allowable error. I have chosen to go with the +/-250rpm limit as it is specifically mentioned in two places in the body of the Standard. Hopefully these was considered more carefully than the entry in Table B.2

6 SEE PDF FROM RP CONVERSION FACTORS BERNARD FRABOULET as per 3) Equation 55 coefficient need to be rounded to ( or ) > Up today we found to errors with 2 occurrences in 10B-2 which need a > correction > 1) I report thermometer/thermocouple accuracy which need to be > corrected in clause and 8.2 at +/- 1 C (+/- 2 F ) as per annex > table B-2 > 2) Schlumberger/ Simon report Clause high curing temperature > which should be 88 C (190 F) and not as shown 90 C (194 F). > > 3: I found an error in Tables 8 and 9 : Conversion factor between pascal to lbf/100 ft² should be and not or if you prefer with R1B1 configuration -F1.0 spring per degree of dial deflection Pa is equivalent to lbf/100 ft² and not and so on. Table 8 and 9 should be corrected for USC values ( i hope that SI are right need to be checked) Is it all?? need to read again all clauses as Simon reported that the ballot file was right and the changes was done with the edition file ( and proof file), so something else can be found if I miss it during the last proof reading. Plan to do that but not before mid January. Coaxial cylinder viscosimeter Calculations of coefficients for R1B1 Spring factor 1 RP 13B 1, RP 13 B 2, RP13D, RP10B 2 provide the following coefficients : RP 13 D RP 13 B 1 RP13 B 2 RP 10 B 2 Shear rate coefficient ( k 3 ) s 1 Shear stress coefficient per degree lbf/100ft² Shear stress coefficient per degree * Pascal Torsion Spring F1.0 coefficient Dyne/cm and degree * calculated

7 Between these 4 recommended practices, discrepancies are observed. It is the same apparatus, coefficient should be the same one s ( USC unit shear rate coefficient is generally simplified and taken as 1). Main discrepancies is related to the shear rate coefficient, if the value is generally reported as Pa ( 5.11 dyne/cm) the conversion of this value into USC unit should be lbf/100 ft². Mario Zamora reports a value of lbf/100ft² based on a handnote written by J.G Savins ( 6/29/92) with a shear stress constant of Pa ( conversion to USC units : rounded to 1.066) The purpose of this note is to review calculations of coefficients for R1B1 configuration with spring F1.0. based on available data. Calculations introduce an experimental coefficient (k 2 ), the effective shear rate surface for the bob. 1) Shear rate coefficient per degree of dial deflection Shear rate are culated using the following formula and a k 3 coefficient based on the geometry R1B1 Viscosimeters suppliers are providing a value for k 3 of s 1 per r/min This value can be calculated using the formula used to calculate shear rate : 4π N 60 1 ² N where is the shear rate in s 1 N is the rotor speed in r/min ri is the bob external radius [B1 diameter : mm (1.358 in.)] re is the rotor internal radius [R1 diameter : mm(1.450 in. )] As dimensions are always provided in metric calculations is done with metric unit for radius ( with inches result is s 1 ) k 3 = rounded to s 1 per r/min However considering the tolerance on the diameters [ ±0.02 mm (±0.001 in.)] it should be recommended to take a value of : k 3 = 1.70 s 1 per r/min 2) Shear Stress coefficients per degree of dial deflection, k 1 k 2 In USC unit values are betwwen and lbf/100ft². Viscosimeters suppliers are providing a value of 5.11 dynes/cm ( Pa). This value expressed in lbf/100ft² will be

8 In conclusion shear rate coefficients should be recalculated based on B1 geometry and F1.0 spring torsional stiffness. 2.1 calculation using spring torsional stiffness and R1B1 effective geometry 2 ² where τ is the shear stress expressed in pascals (dynes/cm) for a dial deflection θ in degree k 1 is the spring F1.0 torsionnal stiffness : as per suppliers dynes/cm per degree of dial deflection or 3.86 Pascal per degree of dial deflection ri is the radius of the bob ( m) h is the height of the bob (0.038 m) ε is a correction factor on height for effective surface coefficient k 2 is the shear stress constant for effective bob surface. k 2 experimental value is provided by viscosimeter suppliers : cm 3 Then k 1 k 2 = x = dynes/cm rounded to Pa or Pa as given by suppliers. In USC units : k 1 k 2 = which is rounded to or lbf/100ft² 2.2 Calculation of Spring F1.0 torsional stiffness through dead weight calibration This value given at dynes/cm can be determined experimentally based on viscosimeter dead weight calibration measurement. It allows to verify the value k 1 torsional stiffness of spring F1.0 Where k 1 is the spring F1.0 torsional stiffness expressed in dynes/cm per degree of deflection M is the mass used for calibration in grams g is the gravitational constant : m/s² = cm/s² R is the radius of the calibration mandrel : cm : θ is the dial deflection in degree

9 With a mass of 100 g dial deflection should be 254 with a factory deviation of ± 0.5 k 1 = dynes/(cm. ) ±0.08 dynes/(cm. ) then shear stress coefficient per degree of dial deflection would be k 1 k 2 = ( ±0.08 ) x = dynes/cm ± dynes/cm or Pa ± Pa in USC k 1 k 2 = lbf/100ft² ± lbf/100ft² 2.3 Calculation based on overall viscosimeter coefficient Overall coefficient K is provided equal to it is used to determine viscosity with the relation θ N And K = 100 x k 1 k 2 / k 3 k 1 k 2 = ( x 300.0) /100 = dynes /(cm. ) or lbf/100ft² 2.4 Conclusion Considering the results and the incertainties of measurements the following coefficents should be recommended Shear rate coefficient per rpm : 1.70 s 1 Shear stress coefficient per degree of dial deflection : In SI units : Pa generally rounded to Pa In USC units :1.066 lbf/100 ft² to be rounded to lbf/100 ft² CALIBRATE DEFINITION DENNIS GRAY The first section below deals with the word calibrate. In one of the dictionaries I looked at, it states: to determine, check, or rectify the graduation of (any instrument giving quantitative measurements). Although calibrate can mean to make adjustments (rectify) in the definition; some are not including it can also mean check which will be to verify the accuracy. Since I and a few others mistook (and didn t look up) the actual definition of Calibrate, I would assume others may also. A suggestion is to make a statement in Annex B somewhere (possibly at the beginning or a note) that: Calibrate is a verification of an instrument or to actually make adjustments to the instrument if it is out of calibration. If a verification shows the instrument to be in calibration, the date and verification data will be the calibration. In this Calibration section, the word Verification pertains

10 to instruments that cannot be changed if they are inaccurate and the word Calibrate pertains to instruments that if they are out of calibration as seen by the Verification of the instrument may be adjusted to get them into calibration. 1. Paragraph B.2 and Table B.1 Equipment Calibration Requirements a. In this table there are a lot of components that show that they need to be calibrated. However after thinking about it, should these be verifications and if the verification is inaccurate then do the calibration? The way it is written, it does not indicate that if it is already calibrated that the previous calibration is acceptable with a new date, but the component must have the complete calibration performed. When we were going through this part of the document, I just took it for granted that if it was in calibration there was no need to calibrate it again. However, like many things there are more than 1 way to look at things and I think if it is unclear to some, it should be made clear. b. I visited with one person at the meeting and they do a verification of the calibration of the instrument and if it is still within the accuracy tolerance specifications, they consider it to be calibrated and use the old calibration for the calibration and date it was verified and in calibration at the new date. c. In the past we typically used the previous calibration but now some are asking if they need to do a new calibration although it is not needed. d. If the component is in calibration, I do not see any benefit in performing a new calibration. Sometimes a new calibration can get the component a little closer to the extreme for the tolerances than it already is (closer to being out of tolerance). e. I suggest all calibrations are changed to verification/calibration (V/C) in the table (with a note at the bottom stating V/C indicates the verification of calibration needs to be performed and only if it is not in calibration does the calibration need to be performed and the verification values are accepted at the time it is verified) and in B.2 one the calibration records contain a statement that if the component is in calibration during the as found verification, the as found verification be used as the calibration and a new calibration does not have to be performed. i. As well as in the two tables being changed to Verification/Calibration, I feel the written portion of Section B should have all the places that state shall be calibrated (or similar) be changed to verified and if out of calibration be calibrated. If in calibration, the date of the verification should be documented along with the verification values. ii. When we have a person in to test our load machine for breaking compressive strengths, they only check the accuracy of the load machine and do not make any adjustments if it is in calibration and certify it as is. I feel the same is done for our weighing scales also. f. A full calibration takes time and I feel it is a waste of time and money if the component is already in calibration as shown by the verification. 2. Paragraph B.3.5 Temperature Devices g. Basically this section recommends calibrations of the thermocouples and temperature instruments. There are some instruments (atmospheric consistometer and atmospheric water bath and possibly others) that I would suggest a different method that does not include verifying the calibration accuracy of the thermocouple. iii. Atmospheric consistometer 1. It is difficult, time consuming, can create a mess, possibly cause other operational problems and possible safety issues to calibrate the thermocouple. This is due to the bath typically being filled with oil and having to drain it, location of the thermocouple and difficulty getting to it and compactness of components in the electrical compartment where the indicator/controller resides. 2. My proposal is to allow the verification and calibration of this instrument to be accomplished by having one cell in place and rotating it to provide circulation within the bath to distribute the heat evenly, or to devise a paddle system that will fit one of the holes for the slurry cup to keep the bath at a uniform

11 temperature. Then a certified thermocouple be placed in the bath close to the tip of the atmospheric consistometer s thermocouple and connected to a certified temperature instrument calibrator to indicate the temperature of the bath. If changing the temperature results in temperatures on the certified temperature instrument calibrator and the indicator/controller on the unit are within the specified ±2 F it is in calibration. However if it is not within ±2 F the instrument/controller shall be calibrated so it reads within the specified ±2 F of the certified digital temperature calibrator. iv. Water bath with digital temperature indicator/controller and thermocouple/rtd. 3. The baths are made so it is extremely difficult to get to the thermocouple or RTD of a water bath. 4. My proposal is to allow the verification and calibration be performed with a stirring device in the water bath (if it doesn t already have one built in) to circulate the water for uniform temperature. Then a certified thermocouple be placed in the bath and connected to a certified temperature instrument calibrator to indicate the temperature of the bath. If changing the temperature results in temperatures on the certified temperature instrument calibrator and the indicator/controller on the unit within the specified ±2 F it is in calibration. However if it is not within ±2 F the instrument/controller shall be calibrated so it reads within the specified ±2 F. 3. Table B.2 Calibration and Verification of Well Cement Testing Equipment h. Digital Hand Held Thermometer v. I am not sure they have the capability to be calibrated or if they can only be verified, but it states in the table they are to be calibrated. If they can t be calibrated in the lab, I suggest they be verified for accuracy and if not within the ±2 F it be replaced. vi. I am also hearing some use this as an instrument to calibrate temperature for other instruments (such as those above). Since the accuracy of the digital thermocouple is typically no better than ±2F, I think there should a statement (possibly in the description area) stating it is only for checking temperature of a process and not an instrument to be used to calibrate another instrument. 4. Any changes that may be made, should be included when the calibration section is placed into the next version of API Specification 10A. DENNIS GRAY I was just sent an e mail about the new API RP 10B 2 where there may need to be some modifications. The problem is the weight sets are to have a tolerance of 0.1% of the stated weight for the weight (below). Table B.2 Calibration and Verification of Well Cement Testing Equipment Equipment and Calibration/ Check Points Tolerance Fr Component/Function Verification Weight sets Calibration Each piece 0.1 % of nominal value A Then when you go to the tolerances for the HPHT potentiometer weight sets they are all 0.1 gram. However if the weight sets are calibrated/verified to 0.1% of their weight, they do not meet the requirements for use in calibrating the potentiometer since at weights above 100 grams they will be out of tolerance for calibrating the potentiometer. At 150 grams the weight may be 1.5 grams at 0.1% tolerance and at 400 grams it may be at 0.1% tolerance. Does the info in Table B 2 need to have a tolerance of 0.1 grams or is there another way to make the two places so they are usable. The potentiometer weight information is below.

12 We had a comment from one of the users of the Apr 2013 edition of 10B 2 about an inconsistency. I have listed it below. Simon followed up and checked the ballot edition which was actually correct. Is there a way to correct the published version? From the user: I am looking at the new API 10B 2 second edition that just came and noticed this inconsistency: Curing at Pressures Above Atmospheric For samples cured at or below 88 C (190 F), maintain test temperature and pressure until 45 min ( 5 min) prior to testing. For test temperatures above 90 C (194 F), discontinue heating and allow samples to cool at such a rate that the sample temperature is 90 C (194 F) or less 45 min prior to testing. Maintain test pressure on the curing vessel during the cooling process. Release the pressure gradually and remove the molds from the curing vessel. Immediately remove the samples from the molds and immerse them in a water cooling bath at 27 C 3 C (80 F 5 F) for 45 min ( 5 min) until the samples are tested. Ref: API 10B-2 2 nd Edition Seems like in the whole document there is a switch from 90degC (194degF) temp reference to 88 degc (190degF) temp reference. My opinion this is due to SI vs. Imperial unit rounding. It was more convenient to reference 190degF than 194degF for Imperial unit user and easier for SI user to quote 90degC rather than 88 degf. However the gap on the quote above still stands (although the gap is not big ).

13 From: Simon James Sent: Friday, April 19, :08 AM To: Gunnar DeBruijn Subject: FW: API 10B-2 second edition Gunnar, We need to raise this to the API. The ballot version was correct: For samples cured at or below 88 C (190 F), maintain test temperature and pressure until 45 min (±5 min) prior to testing. For test temperatures above 88 C (190 F), discontinue heating 45 min (±5 min) before the end of the required curing period and cool samples as fast as possible to 88 C (190 F) or less. Maintain test pressure on the curing vessel during the cooling process. Release the pressure gradually and remove the moulds from the curing vessel. Immediately remove the samples from the moulds and immerse them in a water cooling bath at 27 C ±3 C (80 F ±5 F) for 45 min (±5 min) until the samples are tested. Simon From: Labinot Mahmuti [mailto:lmahmuti@slb.com] Sent: Thursday, April 18, :40 PM To: cembb@slb.com Subject: API 10B-2 second edition Dear Community, I am looking at the new API 10B 2 second edition that just came and noticed this inconsistency: Curing at Pressures Above Atmospheric For samples cured at or below 88 C (190 F), maintain test temperature and pressure until 45 min ( 5 min) prior to testing. For test temperatures above 90 C (194 F), discontinue heating and allow samples to cool at such a rate that the sample temperature is 90 C (194 F) or less 45 min prior to testing. Maintain test pressure on the curing vessel during the cooling process. Release the pressure gradually and remove the molds from the curing vessel. Immediately remove the samples from the molds and immerse them in a water cooling bath at 27 C 3 C (80 F 5 F) for 45 min ( 5 min) until the samples are tested. Ref: API 10B-2 2 nd Edition Seems like in the whole document there is a switch from 90degC (194degF) temp reference to 88 degc (190degF) temp reference. My opinion this is due to SI vs. Imperial unit rounding. It was more convenient to reference 190degF than 194degF for Imperial unit user and easier for SI user to quote 90degC rather than 88 degf. However the gap on the quote above still stands (although the gap is not big ).

On the Rheological Parameters Governing Oilwell Cement Slurry Stability

On the Rheological Parameters Governing Oilwell Cement Slurry Stability ANNUAL TRANSACTIONS OF THE NORDIC RHEOLOGY SOCIETY, VOL. 12, 2004 On the Rheological Parameters Governing Oilwell Cement Slurry Stability Roni Gandelman, Cristiane Miranda, Kleber Teixeira, André L. Martins

More information

Manual. Low Shear Viscometer LS 300. prorheo GmbH Bahnhofstr Althengstett Tel

Manual. Low Shear Viscometer LS 300. prorheo GmbH Bahnhofstr Althengstett Tel Manual Low Shear Viscometer LS 300 prorheo GmbH Bahnhofstr. 38 75382 Althengstett Tel. +49-7051-92489-0 Office@proRheo.de www.prorheo.de 1. Introduction... 3 1.1. Measuring system... 3 1.2. Principle of

More information

INDIANA DEPARTMENT OF TRANSPORTATION OFFICE OF MATERIALS MANAGEMENT. VERIFYING THERMOMETERS ITM No T

INDIANA DEPARTMENT OF TRANSPORTATION OFFICE OF MATERIALS MANAGEMENT. VERIFYING THERMOMETERS ITM No T 1.0 SCOPE. INDIANA DEPARTMENT OF TRANSPORTATION OFFICE OF MATERIALS MANAGEMENT VERIFYING THERMOMETERS ITM No. 909-08T 1.1 This test method covers the procedure for a verification of scale accuracy of liquid-in-glass

More information

Rheometer: Procedure: Part A: Viscosity v Time

Rheometer: Procedure: Part A: Viscosity v Time Rheometer A fluid is defined as a substance that deforms continuously under the action of a shear stress, no matter how small the shear stress may be. Without shear stress, there will be no deformation.

More information

Standard Test Method for Coefficient of Linear Thermal Expansion of Plastics Between 30 C and 30 C with a Vitreous Silica Dilatometer 1

Standard Test Method for Coefficient of Linear Thermal Expansion of Plastics Between 30 C and 30 C with a Vitreous Silica Dilatometer 1 Designation: D 696 08 Standard Test Method for Coefficient of Linear Thermal Expansion of Plastics Between 30 C and 30 C with a Vitreous Silica Dilatometer 1 This standard is issued under the fixed designation

More information

DOLOMITE AS AN ALTERNATIVE WEIGHTING AGENT IN DRILLING FLUIDS

DOLOMITE AS AN ALTERNATIVE WEIGHTING AGENT IN DRILLING FLUIDS Journal of Engineering Science and Technology Vol. 2, No. 2 (2007) 164-176 School of Engineering, Taylor s University College DOLOMITE AS AN ALTERNATIVE WEIGHTING AGENT IN DRILLING FLUIDS M. J. BADRUL*,

More information

GG170L: RHEOLOGY Handout (Material to read instead of a Lab Book chapter you might want to print this out and bring it to lab)

GG170L: RHEOLOGY Handout (Material to read instead of a Lab Book chapter you might want to print this out and bring it to lab) GG170L: RHEOLOGY Handout (Material to read instead of a Lab Book chapter you might want to print this out and bring it to lab) Rheology is the study of how materials flow, and includes two main properties.

More information

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

2.1. Accuracy, n- how close the indication of the thermometer is to the true value. AASHTO Accreditation Policy and Guidance on Thermometer Selection and Records Revised: January 25, 2018 Revision #: 0 1. Objective 1.1. The purpose of this document is to clearly define acceptance criteria

More information

Project PAJ2 Dynamic Performance of Adhesively Bonded Joints. Report No. 3 August Proposed Draft for the Revision of ISO

Project PAJ2 Dynamic Performance of Adhesively Bonded Joints. Report No. 3 August Proposed Draft for the Revision of ISO NPL Report CMMT(A)81 Project PAJ2 Dynamic Performance of Adhesively Bonded Joints Report No. 3 August 1997 Proposed Draft for the Revision of ISO 11003-2 Adhesives - Determination of Shear Behaviour of

More information

Chemistry Basic Science Concepts. Observations: are recorded using the senses. Examples: the paper is white; the air is cold; the drink is sweet.

Chemistry Basic Science Concepts. Observations: are recorded using the senses. Examples: the paper is white; the air is cold; the drink is sweet. Note Packet # 1 1 Chemistry: the study of matter. Chemistry Basic Science Concepts Matter: anything that has mass and occupies space. Observations: are recorded using the senses. Examples: the paper is

More information

Rotational Motion. Figure 1: Torsional harmonic oscillator. The locations of the rotor and fiber are indicated.

Rotational Motion. Figure 1: Torsional harmonic oscillator. The locations of the rotor and fiber are indicated. Rotational Motion 1 Purpose The main purpose of this laboratory is to familiarize you with the use of the Torsional Harmonic Oscillator (THO) that will be the subject of the final lab of the course on

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 8426 Second edition 2008-02-01 Hydraulic fluid power Positive displacement pumps and motors Determination of derived capacity Transmissions hydrauliques Pompes et moteurs volumétriques

More information

AERO 214. Lab II. Measurement of elastic moduli using bending of beams and torsion of bars

AERO 214. Lab II. Measurement of elastic moduli using bending of beams and torsion of bars AERO 214 Lab II. Measurement of elastic moduli using bending of beams and torsion of bars BENDING EXPERIMENT Introduction Flexural properties of materials are of interest to engineers in many different

More information

Drilling Fluid Thixotropy & Relevance

Drilling Fluid Thixotropy & Relevance ANNUAL TRANSACTIONS OF THE NORDIC RHEOLOGY SOCIETY, VOL. 13, 2005 Drilling Fluid Thixotropy & Relevance Richard Jachnik1, 1Baker Hughes INTEQ, Stoneywood Park North, Dyce, Aberdeen, Scotland, UK ABSTRACT

More information

AADE-03-NTCE-35. Once the yield stress has been exceeded, the fluid displays viscous flow characteristics.

AADE-03-NTCE-35. Once the yield stress has been exceeded, the fluid displays viscous flow characteristics. AADE-03-NTCE-35 Drilling Fluid Yield Stress: Measurement Techniques for Improved Understanding of Critical Drilling Fluid Parameters David Power and Mario Zamora, M-I L.L.C. Copyright 2003 AADE Technical

More information

Scientific Literacy & the Scientific Method

Scientific Literacy & the Scientific Method Scientific Literacy & the Scientific Method What does it mean to be? You ve probably hear that term before, and you might be thinking that literate means the ability to and. But what does it mean to be

More information

Investigation #2 TEMPERATURE VS. HEAT. Part I

Investigation #2 TEMPERATURE VS. HEAT. Part I Name: Investigation #2 Partner(s): TEMPERATURE VS. HEAT These investigations are designed to help you distinguish between two commonly confused concepts in introductory physics. These two concepts, temperature

More information

EXPERIMENT 9 ENTHALPY OF REACTION HESS S LAW

EXPERIMENT 9 ENTHALPY OF REACTION HESS S LAW EXPERIMENT 9 ENTHALPY OF REACTION HESS S LAW INTRODUCTION Chemical changes are generally accompanied by energy changes; energy is absorbed or evolved, usually as heat. Breaking chemical bonds in reactants

More information

GUIDELINE FOR HAND HELD SHEAR VANE TEST

GUIDELINE FOR HAND HELD SHEAR VANE TEST GUIDELINE FOR HAND HELD SHEAR VANE TEST NZ GEOTECHNICAL SOCIETY INC August 2001 CONTENTS Page 1.0 Introduction 2 2.0 Background 2 3.0 Recommended Practice 3 4.0 Undrained Shear Strength 3 5.0 Particular

More information

Designation: D (Reapproved 2009)

Designation: D (Reapproved 2009) Designation: D3702 94 (Reapproved 2009) Standard Test Method for Wear Rate and Coefficient of Friction of Materials in Self- Lubricated Rubbing Contact Using a Thrust Washer Testing Machine 1 This standard

More information

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE Concepts: Density Equipment Calibration Approximate time required: 90 minutes for density 90 minutes for two thermometers

More information

CH5716 Processing of Materials

CH5716 Processing of Materials CH5716 Processing of Materials Ceramic Thick Film Processing Lecture MC5 Slurry Characterisation Specific Surface Area Powder size & specific surface area (area per unit wt) closely related As particle

More information

Donald P. Shiley School of Engineering ME 328 Machine Design, Spring 2019 Assignment 1 Review Questions

Donald P. Shiley School of Engineering ME 328 Machine Design, Spring 2019 Assignment 1 Review Questions Donald P. Shiley School of Engineering ME 328 Machine Design, Spring 2019 Assignment 1 Review Questions Name: This is assignment is in workbook format, meaning you may fill in the blanks (you do not need

More information

BIO Lab 3: Measurements

BIO Lab 3: Measurements Measurements All Wisdom is from the Lord God and has been always with Him and is before all time. Who has numbered the sand of the sea, and the drops of rain, and the days of the world? Who has measured

More information

Standard Test Method for Measuring Apparent Viscosity at High-Temperature and High-Shear Rate by Multicell Capillary Viscometer 1

Standard Test Method for Measuring Apparent Viscosity at High-Temperature and High-Shear Rate by Multicell Capillary Viscometer 1 Designation: D 5481 04 An American National Standard Standard Test Method for Measuring Apparent Viscosity at High-Temperature and High-Shear Rate by Multicell Capillary Viscometer 1 This standard is issued

More information

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web:  sales: Morehouse 1 Morehouse Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA 17403 PH: 717-843-0081 web: www.mhforce.com sales: edlane@mhforce.com 2 This presentation will cover the calibration

More information

Certificate of Accreditation

Certificate of Accreditation PERRY JOHNSON LABORATORY ACCREDITATION, INC. Certificate of Accreditation Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of: (Hereinafter called the Organization) and hereby declares

More information

Sliding Bearings. Fig.(1) (a) Full-journal bearing and (b) partial-journal bearing

Sliding Bearings. Fig.(1) (a) Full-journal bearing and (b) partial-journal bearing Sliding Bearings The goal of a bearing is to provide relative positioning and rotational freedom while transmitting a load between two parts, commonly a shaft and its housing. The object of lubrication

More information

Chemistry and Measurement

Chemistry and Measurement Chapter 1 Chemistry and Measurement Concept Check 1.1 Matter can be represented as being composed of individual units. For example, the smallest individual unit of matter can be represented as a single

More information

CHM201 General Chemistry and Laboratory I Laboratory 7 Thermochemistry and Hess s Law May 2, 2018

CHM201 General Chemistry and Laboratory I Laboratory 7 Thermochemistry and Hess s Law May 2, 2018 Purpose: CHM201 General Chemistry and Laboratory I Laboratory 7 Thermochemistry and Hess s Law May 2, 2018 In this laboratory, you will measure heat changes arising from chemical reactions. You will use

More information

In fact, we are going to be sneaky and use Hess s Law to determine the heat of magnesium combustion indirectly. Go to the website:

In fact, we are going to be sneaky and use Hess s Law to determine the heat of magnesium combustion indirectly. Go to the website: Chemistry 161 Please prepare your notebook though the data tables before class on Wednesday, October 27. Write an abstract and place it at the front of your individual report. The abstract and the report

More information

Required Items. ACTIVE PARTICIPATION in class and lab. Use of iclicker

Required Items. ACTIVE PARTICIPATION in class and lab. Use of iclicker Welcome to Chem103 Required Items Textbook: Chemistry: The Molecular Nature of Matter and Change, Martin S. Silberberg, 5 th Edition, McGraw-Hill, 2009 iclicker Scientific calculator ACTIVE PARTICIPATION

More information

Basic Aggregates Study Guide

Basic Aggregates Study Guide Basic Aggregates Study Guide General Conversions 1) There are pounds in one ton. 2) There are grams in one pound. 3) One kilogram consists of grams. 4) The linear distance that one station covers is feet.

More information

EXPERIMENT 7: THE LIMITING REACTANT

EXPERIMENT 7: THE LIMITING REACTANT EXPERIMENT 7: THE LIMITING REACTANT PURPOSE To find the ratio of moles of a reactant to moles of a product of a chemical reaction. To relate this ratio to the coefficients of these substances in the balanced

More information

Lab 6 Forces Part 2. Physics 225 Lab

Lab 6 Forces Part 2. Physics 225 Lab b Lab 6 Forces Part 2 Introduction This is the second part of the lab that you started last week. If you happen to have missed that lab then you should go back and read it first since this lab will assume

More information

RheolabQC. Rotational Rheometer for Quality Control. ::: Intelligence in Rheometry

RheolabQC. Rotational Rheometer for Quality Control. ::: Intelligence in Rheometry RheolabQC Rotational Rheometer for Quality Control ::: Intelligence in Rheometry RheolabQC The Powerful QC Instrument Viscosity measurement and rheological checks for quality control are made routine and

More information

Superconductivity. Never store liquid nitrogen in a container with a tight fitting lid.

Superconductivity. Never store liquid nitrogen in a container with a tight fitting lid. Superconductivity 1 Introduction In this lab we will do some very simple experiments involving superconductors. You will not have to take much data; much of what you do will be qualitative. However, in

More information

Talk Science Professional Development

Talk Science Professional Development Talk Science Professional Development Transcript for Grade 4 Scientist Case: The Heavy for Size Investigations 1. The Heavy for Size Investigations, Through the Eyes of a Scientist We met Associate Professor

More information

ISO 376 Calibration Uncertainty C. Ferrero

ISO 376 Calibration Uncertainty C. Ferrero ISO 376 Calibration Uncertainty C. Ferrero For instruments classified for interpolation, the calibration uncertainty is the uncertainty associated with using the interpolation equation to calculate a single

More information

The Effect of High Pressures and High Temperatures on the Properties of Water Based Drilling Fluids

The Effect of High Pressures and High Temperatures on the Properties of Water Based Drilling Fluids Energy Science and Technology Vol. 4, No. 1, 2012, pp. 27-33 DOI:10.3968/j.est.1923847920120401.256 ISSN 1923-8460[PRINT] ISSN 1923-8479[ONLINE] www.cscanada.net www.cscanada.org The Effect of High Pressures

More information

PHYSICS 15a, Fall 2006 SPEED OF SOUND LAB Due: Tuesday, November 14

PHYSICS 15a, Fall 2006 SPEED OF SOUND LAB Due: Tuesday, November 14 PHYSICS 15a, Fall 2006 SPEED OF SOUND LAB Due: Tuesday, November 14 GENERAL INFO The goal of this lab is to determine the speed of sound in air, by making measurements and taking into consideration the

More information

ISO 844 INTERNATIONAL STANDARD. Rigid cellular plastics Determination of compression properties

ISO 844 INTERNATIONAL STANDARD. Rigid cellular plastics Determination of compression properties INTERNATIONAL STANDARD ISO 844 Fifth edition 2007-04-15 Rigid cellular plastics Determination of compression properties Plastiques alvéolaires rigides Détermination des caractéristiques de compression

More information

SCI 265 Investigating Density, Volume, Mass and Weight

SCI 265 Investigating Density, Volume, Mass and Weight SCI 265 Investigating Density, Volume, Mass and Weight Density, Volume and Mass Play around with the PhET simulation Density. Choose: Custom and My Block 1. If you change the mass of the block, how does

More information

STANDARD PROCEDURE: NWSP R2 (15) Polyacrylate Superabsorbent Powders Determination of the Particle Size Distribution by Sieve Fractionation

STANDARD PROCEDURE: NWSP R2 (15) Polyacrylate Superabsorbent Powders Determination of the Particle Size Distribution by Sieve Fractionation 1. Scope STANDARD PROCEDURE: Polyacrylate Superabsorbent Powders Determination of the Particle Size Distribution by Sieve Fractionation This test method specifies a method for measuring particle size distributions

More information

Measurement Error PHYS Introduction

Measurement Error PHYS Introduction PHYS 1301 Measurement Error Introduction We have confidence that a particular physics theory is telling us something interesting about the physical universe because we are able to test quantitatively its

More information

Thermal & Fluids PE Exam Technical Study Guide Errata

Thermal & Fluids PE Exam Technical Study Guide Errata Thermal & Fluids PE Exam Technical Study Guide Errata This product has been updated to incorporate all changes shown in the comments on the webpage and email comments as of October, 30 2017. If you have

More information

Energy and Energy Conversion Minneapolis Community and Tech. College Principles of Chemistry 1 v q water = m water C water T water (Equation 1)

Energy and Energy Conversion Minneapolis Community and Tech. College Principles of Chemistry 1 v q water = m water C water T water (Equation 1) Energy and Energy Conversion Minneapolis Community and Tech. College Principles of Chemistry 1 v.6.13 Energy Energy is defined by most textbooks as the capacity to do work. However, the true usefulness

More information

Available online Journal of Scientific and Engineering Research, 2016, 3(6): Research Article

Available online   Journal of Scientific and Engineering Research, 2016, 3(6): Research Article Available online www.jsaer.com, 2016, 3(6):318-323 Research Article ISSN: 2394-2630 ODEN(USA): JSERBR omparative Study of the Rheological Properties of Niger Delta rude Oil Bright Bariakpoa Kinate*, Kingdom

More information

MEASUREMENT IN THE LABORATORY

MEASUREMENT IN THE LABORATORY 1 MEASUREMENT IN THE LABORATORY INTRODUCTION Today's experiment will introduce you to some simple but important types of measurements commonly used by the chemist. You will measure lengths of objects,

More information

Specific Accreditation Guidance. Infrastructure and Asset Integrity. Measurement Uncertainty in Geotechnical Testing

Specific Accreditation Guidance. Infrastructure and Asset Integrity. Measurement Uncertainty in Geotechnical Testing Specific Accreditation Guidance Infrastructure and Asset Integrity Measurement Uncertainty in Geotechnical Testing January 2018 Copyright National Association of Testing Authorities, Australia 2018 This

More information

BGD 611 Falling Rod Viscometer

BGD 611 Falling Rod Viscometer BGD 611 Falling Rod Viscometer INSTRUCTION MANUAL BIUGED LABORATORY INSTRUMENTS(GUANGZHOU)CO.,LTD -----------------------------------------------------------------------------------------------------------------------------------------

More information

ISO 178 INTERNATIONAL STANDARD. Plastics Determination of flexural properties. Plastiques Détermination des propriétés en flexion

ISO 178 INTERNATIONAL STANDARD. Plastics Determination of flexural properties. Plastiques Détermination des propriétés en flexion INTERNATIONAL STANDARD ISO 178 Fifth edition 2010-12-15 Plastics Determination of flexural properties Plastiques Détermination des propriétés en flexion Reference number ISO 178:2010(E) ISO 2010 PDF disclaimer

More information

Lecture - 2A Instruments-I

Lecture - 2A Instruments-I Engineering Metrology Prof. J. Ramkumar Department of Mechanical Engineering & Design Programme Indian Institute of Technology, Kanpur Dr. Amandeep Singh Oberoi Department of Industrial & Production Engineering

More information

Lab 11 Simple Harmonic Motion A study of the kind of motion that results from the force applied to an object by a spring

Lab 11 Simple Harmonic Motion A study of the kind of motion that results from the force applied to an object by a spring Lab 11 Simple Harmonic Motion A study of the kind of motion that results from the force applied to an object by a spring Print Your Name Print Your Partners' Names Instructions April 20, 2016 Before lab,

More information

Properties of Liquids

Properties of Liquids Experiment: Properties of Liquids Many of the organic compounds you will be studying this year will be liquids, and in lab, you will frequently have to identify unknown liquids and confirm the identity

More information

Infrared Experiments of Thermal Energy and Heat Transfer

Infrared Experiments of Thermal Energy and Heat Transfer Infrared Experiments of Thermal Energy and Heat Transfer You will explore thermal energy, thermal equilibrium, heat transfer, and latent heat in a series of hands-on activities augmented by the thermal

More information

Readings and Exercises

Readings and Exercises Nebraska Wesleyan University Math 4800: Research Experience Section 1 Spring 02016 Readings and Exercises Tuesday, January 19 Section 1 Exercise 1 Section 2 Exercise 1, 4ce Thursday, January 21 Section

More information

CHM 130LL: The Metric System

CHM 130LL: The Metric System CHM 130LL: The Metric System In this experiment you will: Determine the volume of a drop of water using a graduated cylinder Determine the volume of an object by measuring its dimensions Determine the

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 13318-2 Second edition 2007-09-15 Determination of particle size distribution by centrifugal liquid sedimentation methods Part 2: Photocentrifuge method Détermination de la distribution

More information

Online International Interdisciplinary Research Journal, {Bi-Monthly}, ISSN , Volume-IV, May 2014 Special Issue

Online International Interdisciplinary Research Journal, {Bi-Monthly}, ISSN , Volume-IV, May 2014 Special Issue An improving Rheological Properties and Filter Loss of Water-Base Drilling Fluids Using Carboxymethyle Cellulose (CMC) Abusabah.E.Elemam a, Rashid.A.M.Hussien b, Ahmed.A.Ibrahim b, Sumya.A.Mohamed ba College

More information

Handle Food Samples with Care for Reliable Rheological Results

Handle Food Samples with Care for Reliable Rheological Results Handle Food Samples with Care for Reliable Rheological Results Dr. Klaus Oldörp The world leader in serving science Overview Food and rheology Sample handling before the measurement The right measuring

More information

To use calorimetry results to calculate the specific heat of an unknown metal. To determine heat of reaction ( H) from calorimetry measurements.

To use calorimetry results to calculate the specific heat of an unknown metal. To determine heat of reaction ( H) from calorimetry measurements. Calorimetry PURPOSE To determine if a Styrofoam cup calorimeter provides adequate insulation for heat transfer measurements, to identify an unknown metal by means of its heat capacity and to determine

More information

ME345 Modeling and Simulation, Spring 2018 Case Study 3 Assigned: Friday April 20

ME345 Modeling and Simulation, Spring 2018 Case Study 3 Assigned: Friday April 20 ME345 Modeling and Simulation, Spring 2018 Case Study 3 Assigned: Friday April 20 Due Date 1 (for email confirmation of final grade): Thursday May 10 (11:59 pm) Due Date 2 (absolute latest possible submission):

More information

Wisdom is not measured by age, Intelligence is not measured by grades, Personality is not measured by what others say. O.

Wisdom is not measured by age, Intelligence is not measured by grades, Personality is not measured by what others say. O. Wisdom is not measured by age, Intelligence is not measured by grades, Personality is not measured by what others say. O.Z "Osama Alzoubi" I. What you should already know Exact Numbers counting numbers;

More information

Establishing an Experimental Preconditioning Procedure for Rheological Characterization of Oil-Based Drilling Fluids

Establishing an Experimental Preconditioning Procedure for Rheological Characterization of Oil-Based Drilling Fluids ANNUAL TRANSACTIONS OF THE NORDIC RHEOLOGY SOCIETY, VOL. 23, 205 Establishing an Experimental Preconditioning Procedure for Rheological Characterization of Oil-Based Drilling Fluids Dias Assembayev, Velaug

More information

Density of Brass: Accuracy and Precision

Density of Brass: Accuracy and Precision Density of Brass: Accuracy and Precision Introduction Density is a measure of a substance s mass-to-volume ratio. For liquids and solids, density is usually expressed in units of g/ml or g/cm 3 ; these

More information

ENTHALPY OF FORMATION OF MgO

ENTHALPY OF FORMATION OF MgO ENTHALPY OF FORMATION OF MgO ELECTRONIC LABORATORY NOTEBOOK (ELN) INSTRUCTIONS All work for this experiment must be recorded, attached, or answered in the ELN. Create a pre & inlab page in the Experiment

More information

Chapter 3 Scientific Measurement

Chapter 3 Scientific Measurement Chapter 3 Scientific Measurement Measurements 2 types: Qualitative measurements (words) Heavy, hot, or long Quantitative measurements (# s) & depend on: 1) Reliability of measuring instrument 2) Care w/

More information

How to Perform a Time Study

How to Perform a Time Study How to Perform a Time Study Since the emergence of Taylorism in the 1880 s (a system of scientific management developed by Frederick W. Taylor), industrial production has been broken down into its most

More information

EF 152 Exam 1 Fall 2018 Page 1 Copy 165

EF 152 Exam 1 Fall 2018 Page 1 Copy 165 EF 152 Exam 1 Fall 2018 Page 1 Copy 165 Name: Seat Assignment: Specify your EXAM ID on the right. Use 000 if you don t know your exam ID. Circle your TEAM SECTION 11:10 12:40 2:10 216 217 218 A216 Matt

More information

Review of Anemometer Calibration Standards

Review of Anemometer Calibration Standards Review of Anemometer Calibration Standards Rachael V. Coquilla rvcoquilla@otechwind.com Otech Engineering, Inc., Davis, CA Anemometer calibration defines a relationship between the measured signals from

More information

Accuracy and Precision of Laboratory Glassware: Determining the Density of Water

Accuracy and Precision of Laboratory Glassware: Determining the Density of Water Accuracy and Precision of Laboratory Glassware: Determining the Density of Water During the semester in the general chemistry lab, you will come into contact with various pieces of laboratory glassware.

More information

Introduction to Fluid Mechanics - Su First experiment: Flow through a Venturi

Introduction to Fluid Mechanics - Su First experiment: Flow through a Venturi 530.327 - Introduction to Fluid Mechanics - Su First experiment: Flow through a Venturi 1 Background and objectives. In this experiment, we will study the flow through a Venturi section using both flow

More information

Analyses of Field Measured Data With Rheology and Hydraulics Models

Analyses of Field Measured Data With Rheology and Hydraulics Models International Journal of Fluids Engineering. ISSN 0974-3138 Volume 8, Number 1 (2016), pp. 1-12 International Research Publication House http://www.irphouse.com Analyses of Field Measured Data With Rheology

More information

Standard Test Method for Thermal Flow, Cure, and Behavior Properties of Pourable Thermosetting Materials by Torque Rheometer 1

Standard Test Method for Thermal Flow, Cure, and Behavior Properties of Pourable Thermosetting Materials by Torque Rheometer 1 Designation: D 3795 00a Standard Test Method for Thermal Flow, Cure, and Behavior Properties of Pourable Thermosetting Materials by Torque Rheometer 1 This standard is issued under the fixed designation

More information

Measurement Error PHYS Introduction

Measurement Error PHYS Introduction PHYS 1301 Measurement Error Introduction We have confidence that a particular physics theory is telling us something interesting about the physical universe because we are able to test quantitatively its

More information

Measurements, Sig Figs and Graphing

Measurements, Sig Figs and Graphing Measurements, Sig Figs and Graphing Chem 1A Laboratory #1 Chemists as Control Freaks Precision: How close together Accuracy: How close to the true value Accurate Measurements g Knowledge Knowledge g Power

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics Spring Experiment 5: Faraday s Law

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics Spring Experiment 5: Faraday s Law MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2009 OBJECTIVES Experiment 5: Faraday s Law 1. To become familiar with the concepts of changing magnetic flux and induced current

More information

Standard Practices for Air Speed Calibration Testing

Standard Practices for Air Speed Calibration Testing Standard Practices for Air Speed Calibration Testing Rachael V. Coquilla Bryza Wind Lab, Fairfield, California Air speed calibration is a test process where the output from a wind measuring instrument

More information

Choose the right equipment for lab work. Following Rules for Precision and Accuracy. Following Significant Figure Rules

Choose the right equipment for lab work. Following Rules for Precision and Accuracy. Following Significant Figure Rules Chemistry is a Quantitative Science Part I Measuring Matter At the end of Part I you should be able to: Choose the right equipment for lab work Make accurate measurements Following Rules for Precision

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 22768 First edition 2006-07-15 Rubber, raw Determination of the glass transition temperature by differential scanning calorimetry (DSC) Caoutchouc brut Détermination de la température

More information

Effect of Temperature and Pressure on Rheological Measurements of Cement Slurries

Effect of Temperature and Pressure on Rheological Measurements of Cement Slurries ANNUAL TRANSACTIONS OF THE NORDIC RHEOLOGY SOCIETY, VOL. 22, 2014 Effect of Temperature and Pressure on Rheological Measurements of Cement Slurries Kristján Friðrik Alexandersson 1 and Sunna Ólafsdóttir

More information

Measurement and Measurement Errors

Measurement and Measurement Errors 1 Measurement and Measurement Errors Introduction Physics makes very general yet quite detailed statements about how the universe works. These statements are organized or grouped together in such a way

More information

CHM101 Lab Measurements and Conversions Grading Rubric

CHM101 Lab Measurements and Conversions Grading Rubric CHM101 Lab Measurements and Conversions Grading Rubric Name Team Name Criteria Points possible Points earned Lab Performance Printed lab handout and rubric was brought to lab 3 Safety and proper waste

More information

Protocol Particle size distribution by centrifugal sedimentation (CPS)

Protocol Particle size distribution by centrifugal sedimentation (CPS) Protocol Particle size distribution by centrifugal sedimentation (CPS) 1. Method The CPS Disc Centrifuge separates particles by size using centrifugal sedimentation in a liquid medium. The sedimentation

More information

PHYSICS 206b HOMEWORK #3 SOLUTIONS

PHYSICS 206b HOMEWORK #3 SOLUTIONS HYSICS 06b HOMEWORK # SOLUTIONS M A= cm V=0 cm = a 1. Consider the sample of ideal, monatomic gas in a cylinder sealed by a piston shown above. Neglect any pressure of atmosphere outside of the cylinder.

More information

Theory & Practice of Rotor Dynamics Prof. Rajiv Tiwari Department of Mechanical Engineering Indian Institute of Technology Guwahati

Theory & Practice of Rotor Dynamics Prof. Rajiv Tiwari Department of Mechanical Engineering Indian Institute of Technology Guwahati Theory & Practice of Rotor Dynamics Prof. Rajiv Tiwari Department of Mechanical Engineering Indian Institute of Technology Guwahati Module - 8 Balancing Lecture - 1 Introduce To Rigid Rotor Balancing Till

More information

LABORATORY II DESCRIPTION OF MOTION IN TWO DIMENSIONS

LABORATORY II DESCRIPTION OF MOTION IN TWO DIMENSIONS LABORATORY II DESCRIPTION OF MOTION IN TWO DIMENSIONS This laboratory allows you to continue the study of accelerated motion in more realistic situations. The cars you used in Laboratory I moved in only

More information

C q T q C T. Heat is absorbed by the system H > 0 endothermic Heat is released by the system H < 0 exothermic

C q T q C T. Heat is absorbed by the system H > 0 endothermic Heat is released by the system H < 0 exothermic PLEASE REORD ALL DATA DIRETLY INTO YOUR LAB NOTEBOOKS Introduction Heating a substance is one of the simplest processes carried out in the chemical laboratory, and is usually accompanied by a rise in the

More information

Dynamics of Machines. Prof. Amitabha Ghosh. Department of Mechanical Engineering. Indian Institute of Technology, Kanpur. Module No.

Dynamics of Machines. Prof. Amitabha Ghosh. Department of Mechanical Engineering. Indian Institute of Technology, Kanpur. Module No. Dynamics of Machines Prof. Amitabha Ghosh Department of Mechanical Engineering Indian Institute of Technology, Kanpur Module No. # 07 Lecture No. # 01 In our previous lectures, you have noticed that we

More information

Table 6, replace the Key with: Table 6, change SBX 151 for dimension G, from: 5.59 mm to 5.56 mm mm to mm

Table 6, replace the Key with: Table 6, change SBX 151 for dimension G, from: 5.59 mm to 5.56 mm mm to mm Date of Issue: Ocber 2015 Affected Publication: ANSI/API Specification 17D/ISO 13628-4, Design and Operation of Subsea Production Systems Subsea Wellhead and Tree Equipment, Second Edition, May 2011 ERRATA

More information

Fluid Mechanics Introduction

Fluid Mechanics Introduction Fluid Mechanics Introduction Fluid mechanics study the fluid under all conditions of rest and motion. Its approach is analytical, mathematical, and empirical (experimental and observation). Fluid can be

More information

Gravitational Constant Brett Waite and Daniel Klinge

Gravitational Constant Brett Waite and Daniel Klinge Gravitational Constant Brett Waite and Daniel Klinge The universal gravitational constant was discovered by Henry Cavendish during the years of 1797-1798 His goal was to determine the density of the earth

More information

OHD L-14 METHOD OF TEST FOR DETERMINING THE SPECIFIC GRAVITY AND UNIT WEIGHT OF COMPACTED BITUMINOUS MIXTURES

OHD L-14 METHOD OF TEST FOR DETERMINING THE SPECIFIC GRAVITY AND UNIT WEIGHT OF COMPACTED BITUMINOUS MIXTURES Page 1 METHOD OF TEST FOR DETERMINING THE SPECIFIC GRAVITY AND UNIT WEIGHT OF COMPACTED BITUMINOUS MIXTURES I. SCOPE. This method of test covers the procedures for determining the bulk specific gravity

More information

Chapter 17 Temperature and heat

Chapter 17 Temperature and heat Chapter 17 Temperature and heat 1 Temperature and Thermal Equilibrium When we speak of objects being hot and cold, we need to quantify this by some scientific method that is quantifiable and reproducible.

More information

Chapter 3 Scientific Measurement

Chapter 3 Scientific Measurement Chapter 3 Scientific Measurement Measurements We make measurements every day: buying products, sports activities, and cooking Qualitative measurements are words, such as heavy or hot Quantitative measurements

More information

Millikan Oil Drop Experiment

Millikan Oil Drop Experiment Millikan Oil Drop Experiment Introduction The electronic charge, or electrical charge carried by an electron, is a fundamental constant in physics. During the years 1909 to 1913, R.A. Millikan used the

More information

ME 105 Mechanical Engineering Laboratory Spring Quarter INTRODUCTION TO TRANSIENT CONDUCTION AND CONVECTION

ME 105 Mechanical Engineering Laboratory Spring Quarter INTRODUCTION TO TRANSIENT CONDUCTION AND CONVECTION ME 105 Mechanical Engineering Lab Page 1 ME 105 Mechanical Engineering Laboratory Spring Quarter 2003 2. INTRODUCTION TO TRANSIENT CONDUCTION AND CONVECTION This set of experiments is designed to provide

More information

Lesson Plan Book-stacking Activity

Lesson Plan Book-stacking Activity T o g o d i r e c t l y t o a l e s s o n, c l i c k o n e o f t h e f o l l o w i n g l i n k s : B o o k - s t a c k i n g A c t i v i t y B a l l o o n A c t i v i t y H y d r o g e n G a s L a b F

More information

Measuring Viscosity. Advanced Higher Physics Investigation. Jakub Srsen SCN: Boroughmuir High School Center Number:

Measuring Viscosity. Advanced Higher Physics Investigation. Jakub Srsen SCN: Boroughmuir High School Center Number: Measuring Viscosity Advanced Higher Physics Investigation Jakub Srsen SCN: 050950891 Boroughmuir High School Center Number: 0 09/02/2009 Contents Introduction 2 Summary 2 Underlying Physics 2 Procedures

More information