Handheld Miniature Ion Trap Mass Spectrometers

Size: px
Start display at page:

Download "Handheld Miniature Ion Trap Mass Spectrometers"

Transcription

1 Anal. Chem. 2009, 81, Handheld Miniature Ion Trap Mass Spectrometers Zheng Ouyang, Robert J. Noll, and R. Graham Cooks Downloaded via on September 23, 2018 at 19:01:28 (UTC). See for options on how to legitimately share published articles. Purdue University For field applications, miniature and rapid have become almost synonymous, yet these small mass spectrometers are not useful if performance is too severely compromised. (To listen to a podcast about this feature, please go to the Analytical Chemistry website at pubs. acs.org/journal/ancham.) MS is widely regarded as the gold standard for chemical analysis with respect to selectivity, detection limits, and broad applicability. However, mass spectrometers are comparatively delicate instruments, partly a consequence of the need for a vacuum system into which the sample is introduced. There have long been efforts at direct MS analysis of complex mixtures, 1 but extensive sample cleanup and increasingly sophisticated multianalyzers have been involved. These considerations have limited the applications of MS outside the laboratory. It is striking that although progress in many areas of technology is strongly associated with miniaturization, this correlation is barely evident in MS. Nevertheless, for many applications of MS, in situ experiments would have significant advantages over laboratory measurements, even if there are significant losses in analytical performance. Some applications of immediate interest include environmental monitoring (especially surveys that require large numbers of samples), quality control, food safety, forensics, security and public safety, and clinical diagnostics. The objective of this article is to summarize recent progress in developing handheld miniature mass spectrometers. Note that the focus is on complete systems, not particular components, even important ones like the mass analyzer. ON THE SMALL SIDE Interest in small mass analyzers stretches back several decades. For example, a hyperbolic ion trap analyzer the size of a quarter with a 2.5-mm radius and an m/z range of 70,000 was reported in A Mattauch-Herzog-type, nonscanning, double-focusing sector mass analyzer was reported in 1991, and a very small, double-focusing, crossed electric- and magnetic-field sector analyzer was described in ,4 Recent work by Ramsey, Austin, Short, and others has advanced small analyzers further, and the emphasis is increasingly turning to microelectromechanical systems (MEMS)-fabricated mass analyzers. 5-8 However, only in the past decadesand especially in the past 5 yearsshas significant progress been made in the development of small, low-power, portable, autonomous mass spectrometer systems; some have progressed to the point of commercialization. A review in 2000 summarized the mass analyzers used in miniature instruments, showing examples of virtually all the common types, including quadrupole mass filters, quadrupole ion traps, magnetic sector fields, TOF, and FT ion cyclotron resonance instruments. 9 More recent work has seen the commercialization of a novel toroidal ion trap system. 10 A website devoted to small instruments and a conference series on MS in harsh environments prominently features handheld mass spectrometers ( /ac900292w CCC: $ American Chemical Society Analytical Chemistry, Vol. 81, No. 7, April 1, Published on Web 03/12/2009

2 Table 1. Handheld mass spectrometers Instrument Weight (kg) Power (W) Mass analyzer MS/MS? Sample introduction or ionization mode Mass range (m/z)/ resolution (R) Reference Mini Rectilinear ion trap Yes MIMS, direct leak, ESI, DESI 550/ Mini Rectilinear ion trap Yes MIMS, direct leak, ESI, DESI 2000/ ChemCube/ChemPack Quadrupole mass No SPME, EI 400/ filter on chip Guardion Miniature toroidal Yes SPME, EI 500/500 10, 24 ion trap Suitcase TOF - - Miniature TOF No MALDI sample plate 50,000/30 25 Palm-Portable Miniature CIT No Pulsed gas leak, EI 300/ (without pump) Griffin Analytical Miniature CIT Yes MIMS, direct leak 425/400 Portable GC/MS Full-size QIT No Mini GC/preconcentration 100/ O. I. Analytical Mattauch-Herzog sector No Direct gas leak, EI 300/ MIMS, membrane inlet MS; SPME, solid-phase microextraction; QIT, quadrupole ion trap; EI, electron impact. Table 2. Characteristics of ion trap mass analyzers for use in miniature mass spectrometers Desired characteristic Reason for the characteristic Ion trap suitability High pressure tolerance Pumping limitations of small instruments Tolerates pressure best Miniaturization Minimize system size and weight Easily miniaturized, optimized geometry Tandem MS Analyze complex samples Single miniature ion trap can be used High ion currents Maximize signal strength Arrays of small traps can be multiplexed to increase ion current de). This article updates the 2000 review, and Table 1 lists some of the individual systems and their principal features, including power and weight. Most miniature (hand-portable) systems use conventionally fabricated components, including mass analyzers. MEMS-based analyzers and other components, including ion sources and detectors, are now appearing in conference presentations. Note that the entire mass spectrometer system, not the mass analyzer itself, is the focus of this article. Miniature mass spectrometers are of most interest in field applications, where speed of analysis is invariably a high priority. So important is this factor that miniature and rapid are almost synonymous in discussions of small instruments. In spite of this double requirement, miniature mass spectrometers are not useful if performance is too severely compromised. Analysis of complex mixtures must be possiblestherefore, if chromatography is not an option (it is often too slow for on-line work), then tandem MS, ion mobility, or exact mass measurements must be available. Of these choices, only tandem MS is currently available on small systems. The need for rapid in situ measurements in miniaturized systems places even more demands than usual on sample introduction. In laboratory instruments, high performance can be achieved by implementing slow but complete separations. But in portable miniature instruments, sample introduction systems need to be able to handle solid-, liquid-, and vapor-phase samples at rates compatible with in situ analysis. Membrane interfaces and solid-phase sorbent systems are useful in vapor-phase and some solution-phase analyses and can be implemented in conjunction with traditional electron ionization or chemical ionization methods, including atmospheric-pressure chemical ionization. However, the analysis of solid samples and nonvolatile compounds in solution requires modern desorption or spray ionization methods, which we discuss in this article. ION-TRAP-BASED MINIATURE INSTRUMENTS Many miniature mass spectrometers are based on ion trap technology. 11 The reasons for the choice of the ion trap as the mass analyzer in our program as well as at Brigham Young University, Oak Ridge National Laboratory, the University of North Carolina, the University of South Florida, and the Defence Evaluation and Research Agency (U.K.) are summarized in Table 2. 5,7,10,12 Two contrasting approaches to building small, capable mass spectrometer systems can be taken. In the bottom-up approach, the scale of interest is chosen, and components are built and assembled on that scale. This approach is being used by Syms and co-workers. 13 They concentrate on quadrupole mass filter analyzers and MEMS methods, although their current Chem- Cube/ChemPack instrument is a hybrid of MEMS and conventional technology. The top-down approach starts with a macroscale instrument and, in an iterative fashion, reduces component sizes while maintaining performance. We have used the relatively safe, but slow, top-down approach over the past decade We were able to produce small ion trap mass analyzers, first by using the simplified cylindrical ion trap (CIT) geometry and later by modifying the geometry of the high-ion-capacity linear ion trap to the simplified, easily miniaturized rectilinear ion trap. These capabilities allowed and facilitated shrinkage of the sizes of the other mass spectrometer components, especially the rf supply and the vacuum system. This inwardly spiraling characteristic of mass spectrometer miniaturization has a remarkable self-reinforcing aspect, with obvious parallels to Moore s law in computer science. The result has been a series of instruments of decreasing size, weight, and power consumption. REPRESENTATIVE MINIATURE ION TRAP INSTRUMENTS To illustrate the difficulties that remain in developing highly capable handheld mass spectrometers, as well as the capabilities 2422 Analytical Chemistry, Vol. 81, No. 7, April 1, 2009

3 Figure 1. Performance characteristics of the Mini 10 for some simple samples. All data acquired using EI ionization. (a) The instrument. (b) Calibration curve for naphthalene dissolved in water, with a limit of detection of 3 ppb and a linear dynamic range of (c) Mass spectrum of toluene, showing unit resolution of the two main ion peaks at m/z 91 and 92. (d) Mass spectrum of perfluorotributylamine, a common m/z calibration compound, showing an upper m/z limit >500. (e) Single-stage mass spectrum of methyl salicylate, a chemical weapon simulant, followed by (f) isolation using stored waveform inverse FT of the molecular ion at m/z 152. (g) In turn, the parent ion is subjected to CID, creating characteristic ion fragments. (h) The ion fragment at 120 is in turn isolated and subjected to CID, resulting in the third-generation tandem mass spectrum. 14 of currently available instruments, we describe two ion trap instruments, the Mini 10 and Mini 11, which were built by our team in 2005 and ,17 For the Mini 10, data are acquired with preset scan functions for MS or MS n analysis. Peaks of interest are identified and used for compound recognition with custom software running on the local instrument computer or a remote central computer. A library containing peak lists and pre-acquired calibration curves for targeted analytes is used for compound matching. The Mini 10 uses a commercial turbomolecular high-vacuum pump (the Pfeiffer TPD 011) with a pumping speed of 11 L/s at ultimate vacuum, whereas the Mini 11 uses a much smaller, 500-g turbopump (Creare) with a pumping speed of 5 L/s. The Mini 10 has a small built-in Windows-capable computer; the Mini 11 is controlled wirelessly. Both instruments have GPS, can send and receive data, and can be controlled remotely (in the case of the Mini 11, via a personal digital assistant). Both use the same mm rectilinear ion trap and therefore display almost identical performance, which includes an m/z range of 550 with unit resolution and an extended range with lower resolution. Although a conventional electron impact (EI) ion source was used for early experiments with the Mini 10, it was later replaced by a glow discharge EI source. An atmospheric interface was installed recently to allow ESI and desorption ESI (DESI). Sample introduction can be accomplished in several ways: by direct injection of gas or vapor into the vacuum manifold, injection of gas-phase or solution-phase samples via a membrane introduction system, injection of gas-phase samples by a solid-phase sorption system, and direct sampling of liquid-phase samples via a simple atmospheric inlet system that consists of a long, thin, capillary tube. 14,18,19 The performance of the Mini 10 is summarized in Figure 1, which also illustrates its multistage (MS n ) capabilities, Figure 2. Cocaine on money. (a) Protonated cocaine molecule at m/z 304 shown on banknote by conventional geometry DESI. MS and (inset) MS/MS data provided by the Mini 10. (b) Mass spectrum from 10 ng cocaine on Teflon by using geometry-independent DESI. (Adapted from Ref. 19.) including ion isolation and product ion (MS/MS) spectra produced by collision-induced dissociation (CID). The instrument can be used to determine toxic compounds in air by using a solid sorbent that is selective for particular compound classes. For toxic gases, such as phosgene, ethylene oxide, sulfur dioxide, acrylonitrile, cyanogen chloride, hydrogen cyanide, acrolein, formaldehyde, and ethyl parathion, a 1-minute preconcentration time is required. Detection limits range from 800 ppt to 3 ppm, depending on the analyte. For these particular compounds, a linear dynamic range of 1-2 orders of magnitude was obtained over the concentration range (sub-parts per billion to parts per million) for all analytes. 18 Drugs and other compounds in solution can be examined by passing the analyte solution over a membrane that is then heated Analytical Chemistry, Vol. 81, No. 7, April 1,

4 Figure 3. The combination of ambient ionization by DESI with a miniature mass spectrometer by means of a suitable interface could result in an instrument for personal use. (Adapted from Ref. 28.) to release the analyte into the vacuum system. Short and co-workers have extensively explored membrane introduction in their work on trace chemicals in seawater. For analytes that are amenable to this technique, it is one of the simplest and most reliable sample introduction methods available. 20 The use of miniature mass spectrometers for the direct in situ study of solid environmental samples requires the implementation of DESI, direct analysis in real time, or a similar ambient ionization method. Such experiments have been undertaken with the Mini 10 for the detection of cocaine on banknotes, in which both important variants of DESI were used to obtain spectra (Figure 2). DISADVANTAGES AND FUTURE IMPROVEMENTS Experience with miniature mass spectrometers reveals a number of shortcomings. First, almost all quantitative measures of performance, including resolution and detection limits, are poorer than those obtained with the corresponding conventional laboratory instrumentation; this is due primarily to mechanical errors being relatively large compared to the scale of the miniaturized system. Less obviously, the need to conserve power means that the systems are typically not heated, and this can cause sample carryover with some types of sample introduction methods. Recent developments have revealed numerous ways in which these miniature systems can be improved and further reduced in size. The improvements include operation at higher pressure, routinely at 15 mtorr, which can be facilitated by using pressuretolerant electron multipliers. 19 By operating at a very low Mathieu parameter, the mass range can be extended by a factor of 100, which allows the mass spectra of intact proteins to be recorded with handheld mass spectrometers. 21 Multiplexed operation of several smaller rectilinear ion traps has been demonstrated. 22 Most important, the interface ( wand ) connecting the ion source and the mass spectrometer is now receiving much attention in an effort to increase efficiency and allow standoff detection. A PERSONAL MASS SPEC The combination of two existing technologies, the miniature mass spectrometer and DESI, could lead to a new instrument that allows direct analysis of samples of almost any type in the ambient environment. Both technologies have been reduced to practice separately, but their routine combination, which promises to be far more powerful than the individual systems, requires implementation of several new inventions. The instrument resulting from these combined techniques, effectively a personal mass spectrometer, could be a disruptive technology (Figure 3). Such a technology will have a variety of applications, including examination of tissue for disease; examination of biological fluids for temporal variations that might indicate disease; and detection of dangerous and toxic compounds on surfaces, in water, and elsewhere. Examining tissue for disease includes finding tumor margins by analyzing the lipid distributions in tissue sections sent to pathology during surgery, 23 noninvasive examination of skin lesions, and examination of needle biopsies in the doctor s office. The measurements made on biological fluids could include metabolomics studies, characterization of inborn errors of metabolism, and self-monitoring of biological fluids, an application for which regulatory permissions will have to develop along with the technology. Some of these applications will be used by medical professionals, but others will be performed by individuals who have a strong motivation to monitor their own health, in the context of obesity or diabetes, for example. In such cases, a personal mass spectrometer might become as useful and ubiquitous as the personal computer is today, however unlikely it may seem at present. ACKNOWLEDGMENT We are grateful for scientific collaboration with Griffin Analytical Technologies, a division of ICx, and Prosolia. We thank those experts in ion trap and miniature mass spectrometer technology who kindly commented on earlier drafts of this article. We also acknowledge support over the years of this general project by the Office of Naval Research; the Transportation Security Administration; the National Science Foundation; the Homeland Security Advanced Research Projects Agency; and the Department of Homeland Security, Science and Technology Division, Chem/Bio R&D Section. Thanks to Liang Gao for providing the figure that is the basis for the cover and the art on the opening page. Zheng Ouyang is an assistant professor of biomedical engineering at Purdue University who is interested in MS instrumentation and its application in chemical and biomedical analysis. Robert J. Noll is the managing director of the Center for Analytical Instrumentation Development and a research scientist in the Aston Laboratory for Mass Spectrometry at Purdue. His research interests include physical chemistry, MS, chemometrics, miniature instrumentation, and the Orbitrap analyzer. R. Graham Cooks is a professor at Purdue and is interested in all aspects of MS, from ion motion to ion chemistry, in small instruments and large, and in applications across the sciences. Address correspondence about this article to Cooks at Purdue University, Chemistry Department, 560 Oval Dr., West Lafayette, IN REFERENCES (1) Kondrat, R. W.; Cooks, R. G. Anal. Chem. 1978, 50, 81A. (2) Kaiser, R. E.; et al. Int. J. Mass Spectrom. Ion Process. 1991, 106, (3) Sinha, M. P.; Gutnikov, G. Anal. Chem. 1991, 63, (4) Diaz, J. A.; Giese, C. F.; Gentry, W. R. Field Anal. Chem. Technol. 2001, 5, (5) Pau, S.; et al. Phys. Rev. Lett. 2006, 96, (6) Austin, D. E.; et al. Anal. Chem. 2007, 79, (7) Chaudhary, A.; et al. Int. J. Mass Spectrom. 2006, 251, (8) Amerom, F. H. W. V.; et al. Chem. Eng. Commun. 2008, 195, (9) Badman, E. R.; Cooks, R. G. J. Mass Spectrom. 2000, 35, (10) Lammert, S. A.; et al. J. Am. Soc. Mass Spectrom. 2006, 17, (11) March, R. E.; Todd, J. F. J. Quadrupole Ion Trap Mass Spectrometry, 2nd ed.; Wiley: Hoboken, NJ, (12) Kornienko, O.; et al. Rapid Commun. Mass Spectrom. 1999, 13, (13) Geear, M.; et al. J. Microelectromech. Syst. 2005, 14, (14) Gao, L.; et al. Anal. Chem. 2006, 78, (15) Patterson, G. E.; et al. Anal. Chem. 2002, 74, (16) Riter, L. S.; et al. Anal. Chem. 2002, 74, (17) Gao, L.; et al. Anal. Chem. 2008, 80, (18) Keil, A.; et al. Anal. Chem. 2008, 80, (19) Keil, A.; et al. Anal. Chem. 2007, 79, (20) Short, R. T.; et al. J. Am. Soc. Mass Spectrom. 2001, 12, (21) Janfelt, C.; et al. Int. J. Mass Spectrom. 2008, 278, Analytical Chemistry, Vol. 81, No. 7, April 1, 2009

5 (22) Tabert, A. M.; et al. Anal. Chem. 2006, 78, (23) Wiseman, J. M.; et al. Angew. Chem., Int. Ed. 2006, 45, (24) Contreras, J. A.; et al. J. Am. Soc. Mass Spectrom. 2008, 19, (25) Ecelberger, S. A.; et al. J. Hopkins APL Tech. Digest 2004, 25, (26) Yang, M.; et al. J. Am. Soc. Mass Spectrom. 2008, 19, (27) Shortt, B. J.; et al. J. Mass Spectrom. 2005, 40, (28) Gao, L.; Cooks, R. G.; Ouyang, Z. Anal. Chem. 2008, 80, (29) Kibelka, G. CCD Detector Array for Direct Ion Measurement. Presented at the 5th Harsh Environment Mass Spectrometry Workshop, Sarasota, FL, September 22, 2005; Thursday/Kibelka.pdf. AC900292W Analytical Chemistry, Vol. 81, No. 7, April 1,

Griffin Analytical Technologies

Griffin Analytical Technologies Performance Characterization and Field Testing of a Portable Tandem Mass Spectrometer John Grossenbacher Mitch Wells, Garth Patterson, Brent Knecht, Dennis Barket, Jr., Brent Rardin, Mark Gregory, Jason

More information

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York Mass Spectrometry for the Novice John Greaves and John Roboz (^>, yc**' CRC Press J Taylor & Francis Group Boca Raton London New York CRC Press is an imprint of the Taylor & Francis Croup, an informa business

More information

TANDEM MASS SPECTROSCOPY

TANDEM MASS SPECTROSCOPY TANDEM MASS SPECTROSCOPY 1 MASS SPECTROMETER TYPES OF MASS SPECTROMETER PRINCIPLE TANDEM MASS SPECTROMETER INSTRUMENTATION QUADRAPOLE MASS ANALYZER TRIPLE QUADRAPOLE MASS ANALYZER TIME OF FLIGHT MASS ANALYSER

More information

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps ION ANALYZERS MASS ANALYSER sample Vacuum pumps Mass analysers - separate the ions according to their mass-to-charge ratio MASS ANALYSER Separate the ions according to their mass-to-charge ratio in space

More information

Types of Analyzers: Quadrupole: mass filter -part1

Types of Analyzers: Quadrupole: mass filter -part1 16 Types of Analyzers: Sector or double focusing: magnetic and electric Time-of-flight (TOF) Quadrupole (mass filter) Linear ion trap Quadrupole Ion Trap (3D trap) FTICR fourier transform ion cyclotron

More information

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics Mass spectrometry (MS) is the technique for protein identification and analysis by production of charged molecular species in vacuum, and their separation by magnetic and electric fields based on mass

More information

Development of Portable GC/MS System with Benchtop Performance for Critical DoD Applications. Mitch Wells, Ph.D. Philip Tackett, Ph.D.

Development of Portable GC/MS System with Benchtop Performance for Critical DoD Applications. Mitch Wells, Ph.D. Philip Tackett, Ph.D. Development of Portable GC/MS System with Benchtop Performance for Critical DoD Applications Mitch Wells, Ph.D. Philip Tackett, Ph.D. HEMS Workshop 2017 FLIR Systems, Inc. Industrial Commercial Government

More information

Translational Biomarker Core

Translational Biomarker Core Translational Biomarker Core Instrumentation Thermo Scientific TSQ Quantum Triple Quadrupole Mass Spectrometers. There are two TSQ Quantum Ultra AM instruments available in the TBC. The TSQ Quantum Ultra

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can be obtained from a MS spectrum?

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Applied Biosystems QStar Pulsar i Features of the API QSTAR Pulsar i The

More information

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University matthias.trost@ncl.ac.uk Previously: Resolution and other basics MALDI Electrospray 40 Lecture 2 Mass analysers Detectors

More information

Lecture 8: Mass Spectrometry

Lecture 8: Mass Spectrometry intensity Lecture 8: Mass Spectrometry Relative abundance m/z 1 Ethylbenzene experiment CH 2 CH 3 + m/z = 106 CH 2 + m/z = 91 C 8 H 10 MW = 106 CH + m/z = 77 + 2 2 What information can we get from MS spectrum?

More information

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

(Refer Slide Time 00:09) (Refer Slide Time 00:13) (Refer Slide Time 00:09) Mass Spectrometry Based Proteomics Professor Sanjeeva Srivastava Department of Biosciences and Bioengineering Indian Institute of Technology, Bombay Mod 02 Lecture Number 09 (Refer

More information

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez Mass Analyzers Ion Trap, FTICR, Orbitrap CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Last Update: Oct. 014 Some slides from Dr. Joel Kimmel (007) MS Interpretation

More information

LC-MS Based Metabolomics

LC-MS Based Metabolomics LC-MS Based Metabolomics Analysing the METABOLOME 1. Metabolite Extraction 2. Metabolite detection (with or without separation) 3. Data analysis Metabolite Detection GC-MS: Naturally volatile or made volatile

More information

Miniature Ambient Ionization Mass Spectrometry System For Analysis of Microorganisms

Miniature Ambient Ionization Mass Spectrometry System For Analysis of Microorganisms Miniature Ambient Ionization Mass Spectrometry System For Analysis of Microorganisms R. Graham Cooks 1 and Zheng Ouyang 2 1 Department of Chemistry and 2 Weldon School of Biomedical Engineering Purdue

More information

Talk Line-up. MassTech, Inc Columbia Gateway Drive Columbia, Maryland (443) Goals for MT Explorer 50 Development

Talk Line-up. MassTech, Inc Columbia Gateway Drive Columbia, Maryland (443) Goals for MT Explorer 50 Development A Field Deployable Ion Trap Mass Spectrometer with Atmospheric Pressure Interface Arnold Lee *, Alexander Misharin, Konstantin Novoselov, Victor Laiko, Vladimir M. Doroshenko Talk Line-up Goals for MT

More information

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Scientific LTQ Orbitrap Velos

More information

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Mass Analyzers

More information

Introduction to GC/MS

Introduction to GC/MS Why Mass Spectrometry? Introduction to GC/MS A powerful analytical technique used to: 1.Identify unknown compounds 2. Quantify known materials down to trace levels 3. Elucidate the structure of molecules

More information

MM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE

MM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE MM 2 The new Generation of Mobile Mass Spectrometers Innovation with Integrity Defence CBRNE Picture by: MOWAG GmbH Mobile CWA Detection and Identification Low weight and high performance detector Introducing

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 3 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Chemistry Instrumental Analysis Lecture 37. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 37 Most analytes separated by HPLC are thermally stable and non-volatile (liquids) (unlike in GC) so not ionized easily by EI or CI techniques. MS must be at

More information

Tendenze nell innovazione della strumentazione in spettrometria di massa:

Tendenze nell innovazione della strumentazione in spettrometria di massa: Tendenze nell innovazione della strumentazione in spettrometria di massa: Trappola lineare, orbitrap, ion mobility e nuova strumentazione René Magritte La Condition Humaine 1 Ion Trap Mass Spectrometry

More information

Choosing the metabolomics platform

Choosing the metabolomics platform GBS 748 Choosing the metabolomics platform Stephen Barnes, PhD 4 7117; sbarnes@uab.edu So, I have my samples what s next? You ve collected your samples and you may have extracted them Protein precipitation

More information

Mini review for final

Mini review for final Mini review for final NB: This does not cover all material! (very) Rough outline Units, Conversions, Powers of 10, Prefixes Errors: Sig Figs, Error propagation Statistics: Normal distribution, t-test,

More information

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies APPLICATION BENEFITS Regulatory guidelines and recommendations place a greater emphasis on the detection

More information

Finnigan LCQ Advantage MAX

Finnigan LCQ Advantage MAX www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Finnigan LCQ Advantage MAX The Finnigan LCQ Advantage MAX ion trap mass spectrometer

More information

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile LECTURE-11 Hybrid MS Configurations HANDOUT PREAMBLE As discussed in our previous lecture, mass spectrometry is by far the most versatile technique used in proteomics. We had also discussed some of the

More information

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401 Gas Chromatograph Mass Spectrometer No. GCMS-1401 Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS Introduction Development of methods for analysis of drugs of abuse has become a high

More information

Molecular Mass Spectrometry

Molecular Mass Spectrometry Molecular Mass Spectrometry Mass Spectrometry: capable of providing information about (1) Elemental composition of samples of matter: atomic mass (2) Structures of inorganic, organic, and biological molecules

More information

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai 303351 Instrumental Analysis Mass Spectrometry Lecturer:! Somsak Sirichai Mass Spectrometry What is Mass spectrometry (MS)? An analytic method that employs ionization and mass analysis of compounds in

More information

MS Goals and Applications. MS Goals and Applications

MS Goals and Applications. MS Goals and Applications MS Goals and Applications 1 Several variations on a theme, three common steps Form gas-phase ions choice of ionization method depends on sample identity and information required Separate ions on basis

More information

WADA Technical Document TD2003IDCR

WADA Technical Document TD2003IDCR IDENTIFICATION CRITERIA FOR QUALITATIVE ASSAYS INCORPORATING CHROMATOGRAPHY AND MASS SPECTROMETRY The appropriate analytical characteristics must be documented for a particular assay. The Laboratory must

More information

Mass Spectrometry in MCAL

Mass Spectrometry in MCAL Mass Spectrometry in MCAL Two systems: GC-MS, LC-MS GC seperates small, volatile, non-polar material MS is detection devise (Agilent 320-MS TQ Mass Spectrometer) Full scan monitoring SIM single ion monitoring

More information

Mass Spectrometry. Electron Ionization and Chemical Ionization

Mass Spectrometry. Electron Ionization and Chemical Ionization Mass Spectrometry Electron Ionization and Chemical Ionization Mass Spectrometer All Instruments Have: 1. Sample Inlet 2. Ion Source 3. Mass Analyzer 4. Detector 5. Data System http://www.asms.org Ionization

More information

Research Equipment in the UCA Chemistry Department

Research Equipment in the UCA Chemistry Department Research Equipment in the UCA Chemistry Department Varian 220 Atomic Absorption Spectrometer Atomic Absorption spectroscopy is used to measure the amount of a metal ion present in a solution. It is particularly

More information

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4 High-Resolution Mass spectrometry (HR-MS, HRAM-MS) (FT mass spectrometry) MS that enables identifying elemental compositions (empirical formulas) from accurate m/z data 9.05.2017 1 Atomic masses (atomic

More information

Analysis of Polar Metabolites using Mass Spectrometry

Analysis of Polar Metabolites using Mass Spectrometry Analysis of Polar Metabolites using Mass Spectrometry TransMed Course: Basics in Clinical Proteomics and Metabolomics. Oct 10-19, 2012 dd.mm.yyyy Vidya Velagapudi, Ph.D, Adjunct Professor Head of the Metabolomics

More information

MS/MS .LQGVRI0606([SHULPHQWV

MS/MS .LQGVRI0606([SHULPHQWV 0DVV6SHFWURPHWHUV Tandem Mass Spectrometry (MS/MS) :KDWLV0606" Mass spectrometers are commonly combined with separation devices such as gas chromatographs (GC) and liquid chromatographs (LC). The GC or

More information

Introduction. Chapter 1. Learning Objectives

Introduction. Chapter 1. Learning Objectives Chapter 1 Introduction Learning Objectives To understand the need to interface liquid chromatography and mass spectrometry. To understand the requirements of an interface between liquid chromatography

More information

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511 524) (Harris, Chapt.

More information

CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis Updated: 10 December 2014 Print version CEE 772: Instrumental Methods in Environmental Analysis Lecture #21 Mass Spectrometry: Mass Filters & Spectrometers (Skoog, Chapt. 20, pp.511-524) (Harris, Chapt.

More information

Planar Electrode Quadrupole Ion Traps

Planar Electrode Quadrupole Ion Traps Planar Electrode Quadrupole Ion Traps Brett Hansen¹, Hannah Quist¹, Aaron Hawkins¹, Daniel Austin² ¹Electrical and Computer Engineering ²Chemistry Brigham Young University, Provo, UT Abstract We present

More information

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching s by Removal of Isobaric Interferences Differential Mobility Used as a Tool to Address Selectivity Challenges Adrian M. Taylor

More information

1. The range of frequencies that a measurement is sensitive to is called the frequency

1. The range of frequencies that a measurement is sensitive to is called the frequency CHEM 3 Name Exam 1 Fall 014 Complete these problems on separate paper and staple it to this sheet when you are finished. Please initial each sheet as well. Clearly mark your answers. YOU MUST SHOW YOUR

More information

Mass Spectrometry. Quantitative Mass Spectrometry Chiral Mass Spectrometry

Mass Spectrometry. Quantitative Mass Spectrometry Chiral Mass Spectrometry Mass Spectrometry Quantitative Mass Spectrometry Chiral Mass Spectrometry Quantitation by MS Goal is to develop methodology to sensitively, specifically, accurately and rapidly measure one or more compounds

More information

Introduction to LC-MS

Introduction to LC-MS Wednesday April 5, 2017 10am Introduction to LC-MS Amy Patton, MS Laboratory Manager, Pinpoint Testing, LLC Little Rock, AR DESCRIPTION: Amy Patton, laboratory manager for Pinpoint Testing, will begin

More information

Thermo Finnigan LTQ. Specifications

Thermo Finnigan LTQ. Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Thermo Finnigan LTQ Specifications

More information

Università degli Studi di Bari CHIMICA ANALITICA STRUMENTALE

Università degli Studi di Bari CHIMICA ANALITICA STRUMENTALE Università degli Studi di Bari Corso di Laurea Magistrale in Scienze Chimiche Corso di CHIMICA ANALITICA STRUMENTALE II semestre, 5 + 1 crediti (40 + 15 ore) Prof. Ilario Losito E-mail: illosdid@hotmail.com

More information

Mass Spectrometry Course

Mass Spectrometry Course Mass Spectrometry Course Árpád Somogyi Mass Spectrometry Laboratory, Department of Chemistry and Biochemistry University of Arizona, Tucson, AZ Eötvös University, Budapest April 11-20, 2012 1 2 UA Chemistry

More information

The Claviature of Gas Analysis

The Claviature of Gas Analysis The Claviature of Gas Analysis Mass spectrometric gas analysis for quality assurance of gases and gas mixtures High purity and special gases are becoming increasingly important in the development of modern

More information

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS Mass Spectrometry Hyphenated Techniques GC-MS LC-MS and MS-MS Reasons for Using Chromatography with MS Mixture analysis by MS alone is difficult Fragmentation from ionization (EI or CI) Fragments from

More information

Selecting and planning for the right mass spectrometer

Selecting and planning for the right mass spectrometer Selecting and planning for the right mass spectrometer MSACL US 2018 Palm Springs, CA Wednesday, January 24th Deborah French Ph.D., DABCC, FACB University of California San Francisco Learning Objectives

More information

All of us engaged in science face exciting opportunities

All of us engaged in science face exciting opportunities Ion Trap Mass Spectrometry: A Personal Perspective FOCUS: QUADRUPOLE ION TRAPS ACCOUNT AND PERSPECTIVE George Stafford, Jr. LC and LC/MS Division, Thermo Finnigan, San Jose, California, USA This paper

More information

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Chemistry Instrumental Analysis Lecture 34. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 34 From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry Mass spectrometry X-ray spectrometry

More information

The Planar Electrode Linear Ion Trap

The Planar Electrode Linear Ion Trap The Planar Electrode Linear Ion Trap Brett Hansen 1, Hannah Quist 1, Aaron Hawkins 1, Daniel Austin 2 1 Department of Electrical and Computer Engineering, Brigham Young University, Provo, Utah 2 Department

More information

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 DEPARTMENTS OF CHEMISTRY AND BIOCHEMISTRY GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2 Mass Analysers Shabaz Mohammed October 20, 2015 High Vacuum System Turbo pumps Diffusion pumps Rough pumps Rotary

More information

Overview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC

Overview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC AQualEER Chemistry faculty (Belgrade, Serbia, December 2017) Overview of NETCHEM MSc & PhD courses: mass spectrometry in EFSC Emeritus Prof. Dr Jean-Claude Tabet 1,2,3 1 Institut Parisien de Chimie Moléculaire

More information

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY: Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY: I. BASIC ELEMENTS OF A MASS SPECTROMETER Inlet System or Chromatograph Ion Source Mass Analyzer Detector Computer II. ION SOURCES A. Electron

More information

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein. Mass Analyzers After the production of ions in ion sources, the next critical step in mass spectrometry is to separate these gas phase ions according to their mass-to-charge ratio (m/z). Ions are extracted

More information

Explosive Screening by Vapour Detection: Operational Aspects

Explosive Screening by Vapour Detection: Operational Aspects Explosive Screening by Vapour Detection: Operational Aspects G. Fernandez de la Mora SEDET, Boecillo, Spain, gfdelamora@sedet.com November 14 th, 2013 1 ACES OVERVIEW Product: Air Cargo Explosives Screener

More information

Mass spectrometry gas phase transfer and instrumentation

Mass spectrometry gas phase transfer and instrumentation Objectives of the Lecture spectrometry gas phase transfer and instrumentation Matt Renfrow January 15, 2014 1. Make ions 2. Separate/Analyze 3. Detect ions 4. What is mass resolution and mass accuracy?

More information

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC INTRODUCTION Principle: LC/MS is a technique that combines physical separation capabilities of liquid

More information

Extrel is widely respected for the quality of mass spectrometer systems that are

Extrel is widely respected for the quality of mass spectrometer systems that are Extrel is widely respected for the quality of mass spectrometer systems that are available to the world's top research scientists. In response to increasing requests for complete turn-key systems built

More information

Mass Spectrometry. What is Mass Spectrometry?

Mass Spectrometry. What is Mass Spectrometry? Mass Spectrometry What is Mass Spectrometry? Mass Spectrometry (MS): The generation of gaseous ions from a sample, separation of these ions by mass-to-charge ratio, and measurement of relative abundance

More information

Harris: Quantitative Chemical Analysis, Eight Edition

Harris: Quantitative Chemical Analysis, Eight Edition Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 21: MASS SPECTROMETRY CHAPTER 21: Opener 21.0 Mass Spectrometry Mass Spectrometry provides information about 1) The elemental composition of

More information

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications Doug Stevens Waters Corp. Introduction The range of GC/MS instrument types has expanded

More information

CHROMATOGRAPHY AND MASS SPECTROMETER

CHROMATOGRAPHY AND MASS SPECTROMETER 22 CHROMATOGRAPHY AND MASS SPECTROMETER 22.1 INTRODUCTION We know that the biochemistry or biological chemistry deals with the study of molecules present in organisms. These molecules are called as biomolecules

More information

Mass Spectrometry. A truly interdisciplinary and versatile analytical method

Mass Spectrometry. A truly interdisciplinary and versatile analytical method Mass Spectrometry A truly interdisciplinary and versatile analytical method MS is used for the characterization of molecules ranging from small inorganic and organic molecules to polymers and proteins.

More information

Chemistry 311: Topic 3 - Mass Spectrometry

Chemistry 311: Topic 3 - Mass Spectrometry Mass Spectroscopy: A technique used to measure the mass-to-charge ratio of molecules and atoms. Often characteristic ions produced by an induced unimolecular dissociation of a molecule are measured. These

More information

Multi-stage Mass Spectrometry up to MS 4 on a QTof System

Multi-stage Mass Spectrometry up to MS 4 on a QTof System Valerie Koppen, 1 Russell Mortishire-Smith, 2 Filip Cuyckens 1 1 Janssen R&D, Beerse, Belgium 2 Waters Corporation, Wilmslow, United Kingdom APPLICATION BENEFITS This application note illustrates how multi-stage

More information

Introduction to Mass Spectrometry

Introduction to Mass Spectrometry Introduction to Mass Spectrometry Table of Contents What is mass spectrometry? Where can mass spectrometry be used? How did mass spectrometry originate? A history of mass spectrometry What is a mass spectrometer?

More information

Study of Non-Covalent Complexes by ESI-MS. By Quinn Tays

Study of Non-Covalent Complexes by ESI-MS. By Quinn Tays Study of Non-Covalent Complexes by ESI-MS By Quinn Tays History Overview Background Electrospray Ionization How it is used in study of noncovalent interactions Uses of the Technique Types of molecules

More information

Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory

Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory Steven L. Mullen, Ph.D. Associate Director SCS Mass Spectrometry Laboratory Contact Information 31 oyes Laboratory (8:00-5:00

More information

Ion Sponge: A 3 Dimentional Array of Quadrupole Ion Traps for Trapping and Mass-Selectively Processing Ions in Gas Phase

Ion Sponge: A 3 Dimentional Array of Quadrupole Ion Traps for Trapping and Mass-Selectively Processing Ions in Gas Phase This is an open access article published under an ACS AuthorChoice License, which permits copying and redistribution of the article or any adaptations for non-commercial purposes. pubs.acs.org/ac Ion Sponge:

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

Miniaturized Sensor for the Detection of Environmental Pollutants Alexander Graf 1, Ronald Stübner 1, Christian Kunath 1, Sebastian Meyer 1, Harald Schenk 1 1 Fraunhofer Institute for Photonic Microsystems

More information

Mass Spectrometry for Chemists and Biochemists

Mass Spectrometry for Chemists and Biochemists Erasmus Intensive Program SYNAPS Univ. of Crete - Summer 2007 Mass Spectrometry for Chemists and Biochemists Spiros A. Pergantis Assistant Professor of Analytical Chemistry Department of Chemistry University

More information

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry Introduction The oil industry requires detailed information on the composition of

More information

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS Marian Twohig, Michael D. Jones, Dominic Moore, Peter Lee, and Robert Plumb Waters Corporation, Milford, MA, USA APPLICATION

More information

Ionization Techniques Part IV

Ionization Techniques Part IV Ionization Techniques Part IV CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Presented by Prof. Jose L. Jimenez High Vacuum MS Interpretation Lectures Sample Inlet Ion Source Mass Analyzer Detector

More information

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee Module - 6 Spectroscopic Techniques Lecture - 6 Atomic Spectroscopy

More information

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada) Quantitation and Identification of Pharmaceuticals and Personal Care Products (PPCP) in Environmental Samples using Advanced TripleTOF MS/MS Technology André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Natural Products. Innovation with Integrity. High Performance NMR Solutions for Analysis NMR

Natural Products. Innovation with Integrity. High Performance NMR Solutions for Analysis NMR Natural Products High Performance NMR Solutions for Analysis Innovation with Integrity NMR NMR Spectroscopy Continuous advancement in Bruker s NMR technology allows researchers to push the boundaries for

More information

Powerful Scan Modes of QTRAP System Technology

Powerful Scan Modes of QTRAP System Technology Powerful Scan Modes of QTRAP System Technology Unique Hybrid Triple Quadrupole Linear Ion Trap Technology Provides Powerful Workflows to Answer Complex Questions with No Compromises While there are many

More information

Lecture 15: Introduction to mass spectrometry-i

Lecture 15: Introduction to mass spectrometry-i Lecture 15: Introduction to mass spectrometry-i Mass spectrometry (MS) is an analytical technique that measures the mass/charge ratio of charged particles in vacuum. Mass spectrometry can determine masse/charge

More information

Chip-Scale Mass Spectrometers for Portable Gas Analyzers Luis Fernando Velásquez-García. A. I. Akinwande, K. Cheung, and L.-Y Chen.

Chip-Scale Mass Spectrometers for Portable Gas Analyzers Luis Fernando Velásquez-García. A. I. Akinwande, K. Cheung, and L.-Y Chen. Chip-Scale Mass Spectrometers for Portable Gas Analyzers Luis Fernando Velásquez-García. A. I. Akinwande, K. Cheung, and L.-Y Chen. Microsystems Technology Laboratories (MTL) lfvelasq@mit.edu November

More information

Reactions of Vinyl Chloride and Methanol in a Quadrupole Ion Trap Mass Spectrometer during VOC Analysis

Reactions of Vinyl Chloride and Methanol in a Quadrupole Ion Trap Mass Spectrometer during VOC Analysis Environ. Sci. Technol. 1999, 33, 1528-1533 Reactions of Vinyl Chloride and Methanol in a Quadrupole Ion Trap Mass Spectrometer during VOC Analysis LIANGQIAO BIAN, EARL G. ALLEY,, AND BERT C. LYNN, JR.*,,

More information

Ion-mobility mass spectrometry of polyoxometalate Keplerate clusters and their supramolecular assemblies

Ion-mobility mass spectrometry of polyoxometalate Keplerate clusters and their supramolecular assemblies SUPPLEMENTARY INFORMATION Table of Contents Summary 1. General materials and methods 2. High-resolution mass spectra 3. Ion-mobility mass spectra 4. Fragmentation of clusters in the TWIMS drift tube 5.

More information

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph SCIETIFIC ISTRUMET EWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Introduction of Hitachi Chromaster 5610 MS Detector for High Performance

More information

Introduction to the Q Trap LC/MS/MS System

Introduction to the Q Trap LC/MS/MS System www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment ABI Q Trap LC/MS/MS Introduction to the Q Trap LC/MS/MS System The Q Trap

More information

L. Seda Mut Neslihan Ötük

L. Seda Mut Neslihan Ötük L. Seda Mut 20970802 Neslihan Ötük 20622809 Beytepe Ankara 12.04.2012 Outline Historical Background of MS and SIMS What is MS and SIMS? Working Principle of MS and SIMS Instrumental Structures What properties

More information

Carbon Carbon Bond Activation in Saturated Hydrocarbons by Field-Assisted Nitrogen Fixation Guangtao Li, Xin Li, Zheng Ouyang, and R.

Carbon Carbon Bond Activation in Saturated Hydrocarbons by Field-Assisted Nitrogen Fixation Guangtao Li, Xin Li, Zheng Ouyang, and R. Carbon Carbon Bond Activation in Saturated Hydrocarbons by Field-Assisted Nitrogen Fixation Guangtao Li, Xin Li, Zheng Ouyang, and R. Graham Cooks* Department of Chemistry, Purdue University560 Oval Drive,

More information

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Jonathan Williams Waters Corporation, Milford, MA, USA A P P L I C AT

More information

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability UV Hound Series Portable UVDOAS Multi-gas Analyzers Multi-Gas Capability Dependable High Quality Air Monitoring Individual VOCs PPB Sensitivity Non-Contact Optical Measurement Automated Reporting Configurable

More information

UV3000. Accurate, precise, and portable ambient gas point analyzer

UV3000. Accurate, precise, and portable ambient gas point analyzer UV3000 Accurate, precise, and portable ambient gas point analyzer The Cerex UV3000 is a multifunction analyzer designed to detect part per billion (ppb) to percent level concentrations of multiple gases

More information

Agilent G3212 GC-APCI Source

Agilent G3212 GC-APCI Source Agilent G3212 GC-APCI Source Quick Start Guide Where to find information 2 Getting Started 3 Step 1. Start the Data Acquisition program for the GC and the Q-TOF 3 Step 2. Prepare the GC and Q-TOF for data

More information

SIFT-MS. technology overview

SIFT-MS. technology overview SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY technology overview contents 1. An overview of SIFT-MS 2. Principles of SIFT-MS a. Why chemical ionization? b. Why three reagent ions? c. How does SIFT-MS

More information

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative

Autoresonant Ion Trap Mass Spectrometer The RGA Alternative Autoresonant Ion Trap Mass Spectrometer The GA Alternative 1 What is a Mass Spectrometer? Ionizer Mass Spectrometer Mass Separator Magnetic Sector, Quadrupole or Ion Trap Detector Partial Pressure Measurement

More information