Determination of chloride and sulfate in bio-ethanol by ion chromatography

Size: px
Start display at page:

Download "Determination of chloride and sulfate in bio-ethanol by ion chromatography"

Transcription

1 Indian Journal of Chemical Technology Vol. 23, January, 2016 pp Determination of chloride and sulfate in bio-ethanol by ion chromatography Tatjana Tomić 1, *, Martina Milčić 2, Nada Uzorinac Nasipak 1 & Sandra Babić 2 1 INA Industrija nafte d.d., Product Development Department, Central Testing Laboratory, Lovinčićeva b.b., Zagreb, Croatia 2 Department of Analytical Chemistry, Faculty of Chemical Engineering and Technology, University of Zagreb, Marulićev trg 19, Zagreb, Croatia tatjana.tomic@ina.hr Received 22 November 2013; accepted 10 April 2014 Bio-ethanol is produced from biomass or biodegradable waste and is used as a blending component in commercial gasoline fuel. Bio-ethanol can be contaminated with chloride and sulfate ions which can cause various unwanted side effects. European norm, EN defines the maximum permitted level of ions present and by that maximum chloride and sulfate content should not exceed 6.0 mg kg -1 and 4.0 mg kg -1, respectively. The optimization and validation of ion chromatography method for the quantitative determination of chloride and sulfate ions in bio-ethanol by direct injection ion chromatography using anion-exchange stationary phase, carbonate/hydrogen carbonate mobile phase and conductivity detection has been described. The method is optimized regarding mobile phase composition and flow rate as well as the length of the chromatographic column. Validation of the method has been performed in order to confirm its applicability. The developed method is applied for the determination of chloride and sulfate ions in real bio-ethanol samples used for inter-laboratory comparative testing. The obtained results are compared with the official tests results and z-score has been determined. For chloride determination higher z-scores is obtained (from 1.4 to >3) than for the determination of sulfate (from -0.9 to 0.1). Keywords: Bio-ethanol, Chloride, Ion chromatography, Sulfate Legislation and environmental awareness introduces an alternative to the increasing use of bio-fuels. Therefore, replacement of regular fuels by bio-ethanol blended fuels is endorsed and favored worldwide. Ethanol is one of the main components for future reformulated fuels and offers great advantages due to its physical chemical characteristic, raw materials availability, low production costs and beneficial environmental effects 1. From the environmental standpoint bio-ethanol as a fuel have considerable advantages. Many advantages are achieved by replacing the usual additives with ethanol. In the reformulated gasoline, bio-ethanol replaces additives which increase octane number, such as tetraethyl lead, benzene or methyl tetra-butyl ether (MTBE) and all of them are strong pollutants. The problem of using ethanol as a component of gasoline is due to their incompatibility with the materials that are used in the engines. Bio-ethanol added to bio-fuels may contain chloride and sulfate ions that can cause corrosion and deposit formation 2, 3. There are numerous data on the influence of chloride ions on the formation of pitting and stress corrosion 4, 5, 6. The sulfates can react with sodium in the ethanol distribution system to form insoluble sulfate salts and deposits. For the above reasons, the European standard EN determined the maximum allowed amount of chloride (6 mg kg -1 ) and sulfate (4 mg kg -1 ). If the amount of chloride and sulfate in bio-ethanol exceed a defined maximum allowed amount then it is unacceptable for blending of bio-fuels. A variety of methods have been employed for the analysis of inorganic ions: wet chemical methods such as colorimetry, gravimetry and titrimetry, electrochemical techniques such as use of an ion-selective electrode and amperometric titrations. Method for determination of chloride and sulfate in the bio-ethanol by X-ray fluorescence after a precipitation procedure was also described 8. Many of these methods suffer from interferences, limited sensitivity and time consumption; they are often difficult to automate. Some of the methods are based on previous reactions of analytes with reagents, sample burning, etc. For multiple ions determination use of several different methods is required. For such determination ion chromatography (IC) is favorable by European norms EN and EN and defined as the standard method for the determination of chloride and sulfate ions in bio-ethanol.

2 66 INDIAN J. CHEM. TECHNOL., JANUARY 2016 In this paper the IC method optimization and validation for chloride and sulfate content determination in bio-ethanol samples by anion exchange stationary phase, carbonate/bicarbonate puffer mobile phase and conductivity detection is presented. The possibility of analysis of bio-ethanol by direct injection of the sample onto the chromatographic column without tedious sample preparation step was investigated. EN describes a procedure for determination of ions in bio-ethanol indirectly with sample preparation procedure that involves the evaporation of bio-ethanol on a water bath, drying, and then dissolving in water. All these steps are possible sources of measurement error. The main advantage of the developed method is that it allows bio-ethanol analysis without previous sample preparation step thus significantly improving sample throughput and reduce sources of possible errors. Since the chloride content in bio-ethanol is usually less then levels set up in EN (from 4 mg L -1 to 30 mg L -1 ), the aim of our work was to examine the applicability of the proposed method with respect to the precision of EN method in the case when the chloride content is less than 4 mg L -1. The method was optimized regarding the mobile phase composition, flow rate and the length of the chromatographic column. After optimization of the method, validation was performed to verify the proposed method for a specific purpose. Experimental Section Materials As mobile phase in the system, mixture of sodium carbonate, Na 2 CO 3 (Kemika, Zagreb, Croatia) and sodium hydrogencarbonate, NaHCO 3 (Merck, Darmstadt, Germany) of chromatographic purity in deionised water (18 MΩ) obtained from TKA Pacific/Gen Pure water purification system (TKA Water Purification Systems GmbH, Germany) has been used. For chemical suppressor regeneration sulfuric acid (H 2 SO 4 ), p.a. (Kemika, Zagreb, Croatia) has been used. Organic modifier acetone (CH 3 COCH 3 ) (Gram-mol d.o.o., Zagreb, Croatia) was added for mobile phase optimization. For calibration purpose commercial chloride and sulfate standard solutions were used with concentration of 30 mg L -1 and 150 mg L -1 (Dionex, USA), respectively. Standard solution except chlorides and sulfates also contains fluoride, nitrite, bromide, nitrate and phosphate ions. Standard stock solutions were diluted to appropriate concentrations with deionized water and calibration curves were constructed. IC procedure Ion chromatography system (Metrohm IC 761, Switzerland) used for analysis is a complex modular system consisting of the injection system (766 IC Sample Processor, Metrohm, Switzerland), high pressure pump with flow rate ranged from 0.2 ml min -1 to 2.5 ml min -1 and maximum pressure of 15 MPa, peristaltic pump used for suppressor regeneration, tanks with solvents (Na 2 CO 3 /NaHCO 3 mobile phase, H 2 SO 4 for suppressor regeneration, deionized water for system purification), conductivity detector, chemical suppressor (100 mmol L -1 H 2 SO 4 at flow of 0.6 ml min -1 and deionized water at same flow rate were used for suppressor regeneration) and chromatographic columns Metrosep A Supp 5/150 ( mm I.D.) and Metrosep A Supp 5/250 ( mm I.D.). Both columns have same carrier material polyvinyl alcohol with quaternary ammonium groups and particle size of 5 mm. In a system for ion chromatography there is also suppressor module included when the conductivity detectors are used and the mobile phase is intensively conducting, saturating the detector's response. Injection volume was 20 µl. Chromatographic system is completely controlled by the Methrom IC Net 2.1 programme. The same computer program collects and processes analytical data. Instrument has been prepared by including all individual modules of the instrument and computer software to the computer. If necessary, the solvent tank was filled with fresh mobile phase and purge for 5 minutes. Run flow rate was set at 0.5 ml min -1 and then after 20 minutes gradually increasing to the working flow rate. After stabilization of the pump pressure and conductivity detector, the instrument is ready for the operation. Samples In this investigation bio-ethanol samples from inter-laboratory comparative testing were analyzed as unknown samples. Before analysis, the samples of bio-ethanol were filtered through a regenerated cellulose membrane filter with pore size of 0.45 µm, to remove undissolved particles from the sample. Volume of injected sample was 20 µl. Measured results (mg L -1 ) were converted to mg kg -1 by division with the bio-ethanol density (kg L -1 ).

3 TOMIC et al.: DETERMINATION OF CHLORIDE AND SULFATE IN BIO-ETHANOL 67 Results and Discussion The aim of this work was to develop ion chromatography method suitable for determination of chloride and sulfate ions in bio-ethanol at low concentration levels avoiding laborious sample preparation and with accuracy comparable to official EN method. For that purpose method was optimized, validated and tested by analysis of real bio-ethanol samples. Method optimization Method optimization included the examination of the effect of chromatographic column length, mobile phase composition and flow rate on retention and separation of tested components. Detection of components was performed by suppressed conductivity. Mobile phase composition During the separation process, ions present in the sample are separated in the column based on their size and charge. The mobile phase has generally a low to medium conductivity and choice of the mobile phase depends on the compatibility with the detection mode, nature and concentration of the competing ion, buffering capacity of the mobile phase and adding of organic modifiers 10. A mixture of sodium carbonate and sodium bicarbonate (Na 2 CO 3 /NaHCO 3 ) was used as the mobile phase. Effect of organic modifier addition to mobile phase was examined over the organic nature of ethanol. Organic modifier was added to reduce the impact of negative peak on the chloride peak due to their close retention times. Retention of chloride and sulfate ions was examined with mobile phase consisted of g Na 2 CO 3 and g NaHCO 3 per liter with the addition of organic modifier, CH 3 COCH 3, at concentrations of 20% (v/v) and 2.5% (v/v). When the acetone was added to the mobile phase analysis time was increased from 21 to 40 min. By adding an organic modifier in the system, expected improvements are not achieved. Moreover, significant reduction in resolution was obtained as well as overlapping of the chromatographic peaks (Fig. 1). Although organic modifier can increase the retention time of inorganic ions at the same time it reduces their mutual separation, i.e. decreases separation power of carbonate solution. Therefore, inorganic ions do not come out as single chromatographic curves but as a group of ions. Fig. 1 Chromatogram of anion standard solution (0.25 mg L -1 both chloride and sulfate concentration), mobile phase Na 2 CO 3 /NaHCO 3 with 20% (v/v) acetone

4 68 INDIAN J. CHEM. TECHNOL., JANUARY 2016 Table 1 Retention times of ions at flow rate of 0.6 and 0.7 ml min -1 t R /min Flow rate 0.6 ml min ml min -1 Negative peak Cl (t R Cl -t R negative peak) SO Table 2 Retention times on 150 mm and 250 mm column length (flow rate, 0.7 ml min -1 ) t R /min Column length 150 mm 250 mm Negative peak Cl (t R Cl -t R negative peak) SO Mobile phase flow rate Tendency in changes of retention times and width of chromatographic peaks were monitored at two mobile phase flow rates: 0.6 and 0.7 ml min -1. It was noted that the time between negative peak and chloride ion peak is getting longer at flow rate of 0.6 ml min -1, while the total analysis time was about ten minutes longer. Extended time between negative peak and chloride ion peak is desirable from the analytical point of view because of the potential overlap and overlay of chromatographic peaks that can occur in real samples. Table 1 shows the retention times of chloride and sulfate ions and negative peaks at two investigated mobile phase flow rates. Column length Chromatographic retention of chloride and sulfate ions was examined on chromatographic columns of different lengths: 150 mm (Metrosep A Supp 5/150) and 250 mm (Metrosep A Supp 5/250). On the longer chromatographic column retention time of individual ions is higher. The obtained results are expected considering that the number of theoretical plates in a longer column is higher (Table 2). Total analysis time was also increased on the longer column from the same reason. The resolution between the negative peak and chloride ion peak is higher on the longer column and the possible impact of negative peak on chloride ion determination was reduced by using longer column. Calibration Based on results of preliminary tests, the optimal chromatographic conditions were defined as follows: carbonate/bicarbonate mobile phase without organic Table 3 Validation results Acceptance Cl - 2- SO 4 criteria Selectivity Information satisfies satisfies Linearity R 2 0,999 R 2 = R 2 = Accuracy RSD 10 % RSD = 6.4 % RSD = 1.7 % Repeatability RSD 5% RSD = 0.9 % RSD = 3.2 % LOD Information mg L mg L -1 LOQ Information mg L mg L -1 Robustness Information satisfies satisfies modifier added, flow rate 0.6 ml min -1 and 250 mm length columns. On these working conditions the calibration is performed by means of external standard method. Calibration curves were prepared by plotting the peak area versus the analyte concentration expressed in mg L -1. By means of calibration curves, individual ions are quantified as mass concentration. Each standard was measured in triplicate in seven calibration points in concentration range from 0.25 to 20 mg L -1. For chloride, calibration curve equation was y=13.68x with the correlation coefficient R 2 =0.9997, and for sulfate calibration curve equation was y=10.01x with the correlation coefficient R 2 = Method validation After optimization, validation of the method has been performed in order to confirm its validity and applicability. Prior to performing validation, validation parameters and criteria of their acceptability are defined (Table 3). Parameters were selected based on laboratory experience and according to good laboratory practice. Selectivity was tested by comparing the chromatograms of real samples of bio-ethanol with the chromatogram of standard. It was noted that the ions in both sample and standard solution have same retention times, so it was concluded that the method is selective. Linear calibration curves were obtained with coefficients of correlation higher than thus confirming the linearity of the method. Repeatability is calculated by measuring the ion content in a corresponding sample six times under identical conditions and performed by the same operator in the laboratory. Since the relative standard deviation is below 5% for both investigated anions, the method is considered repeatable. Limits of detection (LOD) and quantification (LOQ) were experimentally estimated from the injection of standard solutions serially diluted until

5 TOMIC et al.: DETERMINATION OF CHLORIDE AND SULFATE IN BIO-ETHANOL 69 the signal-to-noise ratio for analyte reached a value of ten for LOQ and three for LOD. Under the given conditions, noise of base line has been measured. The limits of detection of mg L -1 and mg L -1 were obtained for chloride and sulfate, respectively. The accuracy was checked as yield at three concentrations (1.67 mg L -1, 5.00 mg L -1 and mg L -1 ) by comparing the measured and actual values of concentration. Dividing the measured and actual values, yield is calculated for every concentration measured and expressed as a percentage value. RSD of the yield for chloride ion is significantly larger (6.4%) than for the sulfate ion (1.7%), although both are in the desired range. The goal of robustness examination was to determine the effect of small changes in operating parameters on the analytical results. Changes in operating parameters are usually appears as a result of human factor or conditions in the laboratory. Since the conditions in the laboratory are usually under control, robustness of the method is the mostly influenced by the analyst work. In the proposed method the greatest impact of analysts is in the preparation of the mobile phase, which involves weighing and dilution. Impact of changes in the strength of mobile phase on the final result of the quantitative analysis was tested by adding small portions of deionised water in mobile phase. By diluting mobile phase, retention time of each ion in the chromatographic column was extended. However, Fig. 2 Chromatogram of real sample of bio-ethanol (BioEtOH2) at the quantification there was no significant deviation. Based on the obtained results it can be concluded that a small error during dilution of the mobile phase didn t affect the performance of method. Analysis of bio-ethanol samples Given the lack of real samples of bio-ethanol on the market, method was applied to the analysis of the inter-laboratory samples. After method was validated, four real bio-ethanol samples from inter-laboratory comparative testing were analysed: BioEtOH1, BioEtOH2, BioEtOH3 and BioEtOH4. Inter-laboratory testing included the results obtained from ten laboratories. Chromatogram of BioEtOH2 sample as representative chromatogram is presented on Fig. 2. The content of chloride and sulfate ions in analyzed samples was determined. Measurement results obtained in units of mg L -1 (from calibration curves) were expressed in mg kg -1 taking into account density of each bio-ethanol sample. The obtained results were compared with the official tests results (OTR) and z-score was evaluated (Tables 4 and 5). Considering that baseline has not been stable at the start of the chromatogram as well as a wide negative peak, the results obtained for the determination of chloride have higher z-scores than the results of the determination of sulfate. Both, chloride and sulfate contents in interlaboratory samples are for a concentration decade

6 70 INDIAN J. CHEM. TECHNOL., JANUARY 2016 Table 4 Comparison of chloride concentrations obtained in laboratory and inter-laboratory testing (official tests results, OTR) Sample w (mg kg -1 ) OTR(mg kg -1 ) z-score BioEtOH BioEtOH BioEtOH >3 BioEtOH >3 Table 5 Comparison of sulfate concentrations obtained in laboratory and interlaboratory testing (official tests results, OTR) Sample w (mg.kg -1 ) OTR ( mg.kg -1 ) z-score BioEtOH BioEtOH BioEtOH BioEtOH smaller than the allowable limits of 6 mg kg -1 for chloride and 4 mg kg -1 for sulfate. Requirements of EN method for precision are defined in concentration range from 4 to 30 mg L -1 while for the lower chloride content there are no data available and therefore can not be applied. For sulfate content precision was not established yet in EN For both ions precision should be determined in much lower range comparing to their content in real samples. From January on the Croatian market sale of gasoline which contains bio-component is mandatory and bio-ethanol as a feedstock for their mixing is routinely analyzed in terms of determining the content of chloride and sulfate. The optimized method will be able to use as a quick and adequate method for this analysis. Conclusion Recent introduction of bio-ethanol as a component for the mixing into gasoline, requires the development of new fast and reliable analytical methods. The results obtained in this work indicate that the proposed ion chromatography method offers adequate solution for chloride and sulfate content determination in bio-ethanol. By method optimization, the impact of organic modifiers on improving the resolution has been tested and it is found that organic modifier causes a reduction in the resolution while lower flow rate and longer chromatographic column have a positive impact on method resolution. Results of method validation confirmed the applicability of the method for determination of chloride and sulfate in bioethanol since all validation parameters (selectivity, linearity, accuracy, repeatability, LOD, LOQ and robustness) met the established criteria. The main advantage of the developed method is that it allows bio-ethanol analysis without previous sample preparation step thus significantly improving sample throughput and reduce sources of possible errors. Furthermore, the method is able to determine low chloride and sulfate levels (LOD below 0.1 mg kg -1 ) expected in real bio-ethanol samples. References 1 Blottnitz H von & Curran M A, J Cleaner Prod, 15 (2007) Westbrook P, PhD thesis. Shell oil company. WSPA representative. State Water Resources Control Board s Advisory Panel; Lash R, The corrosion behaviour of metals, plated metals, and metal coatings in methanol/gasoline fuel mixtures. SEA technical paper no : Baenav L M, Gómez M & Calderón J A, Fuel, 95 (2012) Kulušić B, Krstulović Lj & Ivić J, Kem Ind, 53 (1) (2004) 1. 6 Fritz J D & Gerlock R J, Desalination, 135 (2001) EN Automotive fuels - Ethanol as a blending component of petrol - Requirements and test methods, 2011 (E). 8 Teixeira L S G, Chaves T J, Guimaraes P R B, Pontes L A M & Teixeira J S R, Anal Chim Acta, 640 (2009) EN Ethanol as a blending component for petrol -- Determination of inorganic chloride and sulfate content - Ion chromatographic method (EN 15492:2012). 10 Sawicki E, Mulik J D & Wittegenstein E (eds.), Ion Chromatographic Analysis of Environmental Pollutants, (Ann Arbor Science Publications, Ann Arbor,) MI, Vol. 2, 1979.

Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample...

Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample... I J P F A International Science Press Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample from Different Area of North Gujarat Region by Ion Chromatography Bharat

More information

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY As originally published in the SMTA Proceedings. DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY Christopher J. Pontius,

More information

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM IC Application Work AW CH6-1048-012011 Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM Branch: Chemical industry; Water, wastewater, environmental protection,

More information

Pelagia Research Library

Pelagia Research Library Available online at www.pelagiaresearchlibrary.com Der Pharmacia Sinica, 2011, 2 (2): 68-73 ISSN: 0976-8688 CODEN (USA): PSHIBD Method development and validation for determination of methane sulphonic

More information

Suppression in anion chromatography. More sensitive analysis of anions and organic acids

Suppression in anion chromatography. More sensitive analysis of anions and organic acids Suppression in anion chromatography More sensitive analysis of anions and organic acids What is suppression? 02 Suppression plays a key role in the analysis of anions and organic acids using ion-exchange

More information

Analysis of various clays with Metrohm Combustion IC

Analysis of various clays with Metrohm Combustion IC Branch General analytical chemistry Keywords IC 881 / Metrosep A Supp 5-150/4.0 / MMS 5000 / 920 / Absorber Module / Combustion Module / combustion ion chromatography / CIC / matrix elimination / fluoride

More information

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Determination of Tartaric cid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection Suparerk Tukkeeree, Chanita Chantarasukon, and Jeff Rohrer Thermo Fisher Scientific, angkok,

More information

Determination of trace anions in concentrated hydrofluoric acid

Determination of trace anions in concentrated hydrofluoric acid APPLICATION NOTE 78 Determination of trace anions in concentrated hydrofluoric acid Authors Archava Siriraks Thermo Fisher Scientific, Sunnyvale, CA Keywords HF, ICS-5000 +, IonPac AS10, IonPac AC10, ion

More information

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0 Metrosep A Supp 10-100 /4.0 Branch Environment, Food, Beverages Keywords IC; 850; 858; Metrosep A Supp 10-100/4.0; Cyanide; Sulfide; 2.850.9110; DC Mode Summary The determination of sulfide and cyanide

More information

Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography

Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography Author : Li Guoping Yu Rong Shui Rong Institute for the Control of Agrochemicals, Ministry of Agriculture, P.

More information

ION CHROMATOGRAPHY SYSTEM S 150

ION CHROMATOGRAPHY SYSTEM S 150 ION CHROMATOGRAPHY SYSTEM S 150 WATER ANALYSIS ENVIRONMENTAL ANALYSIS ANION & CATION ANALYSIS ION CHROMATOGRAPHY IIon Chromatography is an analytical separation technique based on ionic interactions. Dissolved

More information

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Trace Cations in Power Plant Waters Containing Morpholine Application Note 8 Determination of Trace Cations in Power Plant Waters Containing Morpholine INTRODUCTION Morpholine and ammonium are used as additives in power plant waters. Morpholine acts as a corrosion

More information

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production Thunyarat Phesatcha, Suparerk Tukkeeree, Jeff Rohrer 2 Thermo

More information

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography Yan Liu, Zhongqing Lu, and Chris Pohl, Thermo Fisher Scientific, Sunnyvale, CA USA Overview Purpose: In this work, new high-pressure

More information

"Analysis of Inorganic Anions in Potomac Water, Sediment and Floc by Ion Chromatography" FINAL REPORT. by James E. Girard, Ph.D.

Analysis of Inorganic Anions in Potomac Water, Sediment and Floc by Ion Chromatography FINAL REPORT. by James E. Girard, Ph.D. DC WRRC Report No. 37 "Analysis of Inorganic Anions in Potomac Water, Sediment and Floc by Ion Chromatography" FINAL REPORT by James E. Girard, Ph.D. The work upon which this publication is based was supported

More information

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems Yan Liu, Zhongqing Lu, and Chris Pohl; Thermo Fisher Scientific, Sunnyvale, CA USA Overview

More information

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography Application Note 75 Determination of Sulfate and Chloride in Ethanol by Ion Chromatography Ethanol used as a blending agent in gasoline can be contaminated with chloride and sulfate that form plugging

More information

Dionex IonPac AS28-Fast-4µm column

Dionex IonPac AS28-Fast-4µm column CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS-Fast-4µm Columns Product Specifications The Thermo Scientific Dionex IonPac AS-Fast-4µm column is a high-capacity, hydroxide-selective anionexchange column

More information

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection APPLICATION NOTE 72686 Determination of underivatized aflatoxins,, G2, and in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection Authors Sylvia Grosse, Mauro De Pra, Frank

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

IDENTIFICATION AND DETERMINATION OF HYDROQUINONE IN COSMETIC PRODUCTS 2 14/11/17 ACM 003 BY TLC AND HPLC

IDENTIFICATION AND DETERMINATION OF HYDROQUINONE IN COSMETIC PRODUCTS 2 14/11/17 ACM 003 BY TLC AND HPLC A. IDENTIFICATION BY TLC 1. SCOPE AND FIELD OF APPLICATION The method describes the identification of hydroquinone in cosmetic products. 2. PRINCIPLE Hydroquinone is identified by thin layer chromatography

More information

Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography

Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography L. Canella, C. Austin, A. Beavis, P. Maynard, M. Dawson, C Roux, and P. Doble Centre for Forensic Science, University of Technology,

More information

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS anion-exchange column is designed for the fast analysis of inorganic

More information

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION Andrew J. Aubin and Tanya L. Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Benzocaine (4-Aminobenzoic

More information

This method describes the identification of the following prohibited colorants in cosmetic products:

This method describes the identification of the following prohibited colorants in cosmetic products: A. IDENTIFICATION BY TLC 1. SCOPE AND FIELD OF APPLICATION This method describes the identification of the following prohibited colorants in cosmetic products: Names C I number Pigment Orange 5 12075 Metanil

More information

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device Determination of Bromate in Bottled Mineral Water Using the CRD 00 Carbonate Removal Device Thunyarat Phesatcha, Weerapong Worawirunwong, and Jeff Rohrer Thermo Fisher Scientific, Bangkok, Thailand; Thermo

More information

Air Monitoring. Semi-continuous determination of ambient air quality

Air Monitoring. Semi-continuous determination of ambient air quality Air Monitoring Semi-continuous determination of ambient air quality The Particle Into Liquid Sampler a simple solution for the determination of ions in aerosol particles 02 Combustion of fossil fuels for

More information

BRIEFING. (EXC: K. Moore.) RTS C Propylparaben C 10 H 12 O Benzoic acid, 4 hydroxy, propyl ester; Propyl p hydroxybenzoate [ ].

BRIEFING. (EXC: K. Moore.) RTS C Propylparaben C 10 H 12 O Benzoic acid, 4 hydroxy, propyl ester; Propyl p hydroxybenzoate [ ]. BRIEFING Propylparaben. The European Pharmacopoeia is the coordinating pharmacopeia for the international harmonization of the compendial standards for the Propylparaben monograph, as part of the process

More information

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID Asian Journal of Chemistry Vol. 21, No. 3 (2009), 1739-1746 Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID PRAVEEN KUMAR BALIYAN*, R.P.

More information

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Customer Application Note: 114 Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Sébastien N. Ronkart, Ph.D.; Société wallonne des eaux, rue de la

More information

Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography 1

Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography 1 Designation: D 4327 03 An American National Standard Standard Test Method for Anions in by Chemically Suppressed Ion Chromatography 1 This standard is issued under the fixed designation D 4327; the number

More information

Ion Chromatography Validation for the Analysis of Anions in Gunshot Residue

Ion Chromatography Validation for the Analysis of Anions in Gunshot Residue Ion Chromatography Validation for the Analysis of Anions in Gunshot Residue Student Report Written for: Dr. Samide CH 424 November 22, 2010 Abstract The purpose of this study is to determine whether or

More information

DISCLAIMER: This method:

DISCLAIMER: This method: Inorganic arsenic determination in fresh mussels using water bath extraction and anion exchange chromatography-inductively coupled plasma mass spectrometry DISCLAIMER: This method: - has to be considered

More information

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Determination of Inorganic Anions in Drinking Water by Ion Chromatography Determination of Inorganic s in Drinking Water by Ion Chromatography Peter Jackson Thermo Fisher Scientific Inc. Application Note Introduction The determination of common inorganic anions in drinking water

More information

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers. Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers. 1 Scope This document describes a chromatographic method which allows the identification as well as the determination of

More information

Theory of Ion Chromatography

Theory of Ion Chromatography Theory of Ion Chromatography The ideal method for analyzing anions, cations and polar substances is ion chromatography. It can be used to reliably quantify substances throughout a wide concentration range.

More information

H 3 CO H 3 CO S CH 3

H 3 CO H 3 CO S CH 3 FENITROTHION 35 H 3 CO P H 3 CO S O CH 3 NO 2 ISO common name Chemical name Fenitrothion O,O-Dimethyl O-4-nitro-m-tolyl phosphorothioate (IUPAC) O,O-Dimethyl O-(3-methyl-4-nitrophenyl)- phosphorothioate

More information

Personalised Learning Checklists AQA Chemistry Paper 2

Personalised Learning Checklists AQA Chemistry Paper 2 AQA Chemistry (8462) from 2016 Topics C4.6 The rate and extent of chemical change Calculate the rate of a chemical reaction over time, using either the quantity of reactant used or the quantity of product

More information

groups I and II cations, amines, divalent transition metals

groups I and II cations, amines, divalent transition metals IC Columns Introduction Alltech IC Column Specifications Column Composition ph Range Applications Anion Exchange Columns Allsep polymer-based anion ph 0 inorganic anions, weak and strong acid ions, exchanger,

More information

columns IonPac AS22 Anion-Exchange Column

columns IonPac AS22 Anion-Exchange Column columns IonPac AS Anion-Exchange Column The IonPac AS is a carbonate based anion-exchange column designed for the determination of inorganic anions and low-molecular-weight organic acids including fluoride,

More information

anion-exchange column

anion-exchange column columnsionpac ASA Anion-Exchange Column The IonPac ASA is a carbonateselective anion-exchange column designed for the fast separation of inorganic anions, including fluoride, chlorite, bromate, chloride,

More information

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS CH.MOUNIKA*, M.KINNERA Research Article SIR.C.R.REDDY COLLEGE OF PHARMACEUTICAL SCIENCES, ELURU.

More information

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State

More information

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Application Note Inorganic Ions, Water Testing, Minerals, Metals, Basic Chemicals Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography Authors Anne Mack, Adam Bivens

More information

International Journal of Pharmacy and Pharmaceutical Sciences Vol 2, Issue 1, 2010

International Journal of Pharmacy and Pharmaceutical Sciences Vol 2, Issue 1, 2010 International Journal of Pharmacy and Pharmaceutical Sciences Vol 2, Issue 1, 2010 RP HPLC ESTIMATION OF EZETIMIBE IN TABLET DOSAGE FORMS NAGARAJU. P *, KRISHNACHAITHANYA. K, CHANDRABABU. D, SRINIVAS.

More information

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

Available online  Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2012, 4(6):3275-3279 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Development and validation of a matrix solid-phase

More information

Inline sample preparation for the determination of anions in sodium hydroxide

Inline sample preparation for the determination of anions in sodium hydroxide APPLICATION UPDATE 72331 Inline sample preparation for the determination of anions in sodium hydroxide Authors Hua Yang and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Sample preparation,

More information

637. Thiamethoxam. HPLC method

637. Thiamethoxam. HPLC method 637. Thiamethoxam HPLC method CIPAC Collaborative Trial according to CIPAC Information Sheet N o 293 Dr. Sven Adolph Syngenta Crop Protection Münchwilen AG CH-4333 Münchwilen Switzerland May 212 page 1

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Expanded capabilities in ammonium and amine detection

Expanded capabilities in ammonium and amine detection PRODUCT SPECIFICATIONS Thermo Scientifi c Dionex SC-CERS 500 Salt Converter-Cation Electrolytically Regenerated Suppressor Expanded capabilities in ammonium and amine detection Benefits Broadened ammonium

More information

The analysis of organic acid content of additives, premix, feed, and water.

The analysis of organic acid content of additives, premix, feed, and water. The analysis of organic acid content of additives, premix, feed, and water. Contents Foreword Introduction Warnings 1. Scope 2 1.1 LOD and LOQ 3 2. Normative References 3 3. Definitions 3 3.1 Feed (or

More information

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC TECHNICAL NOTE 122 Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC Authors Terri Christison and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA,

More information

Determination of Adsorbable Organic Halogen in Wastewater

Determination of Adsorbable Organic Halogen in Wastewater Determination of Adsorbable Organic Halogen in Wastewater Jingli Hu, Jeff Rohrer, and Kirk Chassaniol Thermo Fisher Scientific, Sunnyvale, CA The world leader in serving science Outline AOX (Adsorbable

More information

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection Barbara Shao, Terri Christison, Fei Pang, Cathy Tanner, and Frank Hoefler, Thermo Fisher Scientific, Sunnyvale,

More information

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components APPLICATION UPDATE 157 Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components Authors Sumate Pengpumkiat, Weerapong Worawirunwong,

More information

Sulfur speciation in mining leachates by ion exchange with direct UV detection, using a perchlorate mobile phase.

Sulfur speciation in mining leachates by ion exchange with direct UV detection, using a perchlorate mobile phase. direct UV detection, using a perchlorate mobile phase. Branch Branch 15: Mineral resources Branch 2: Water, wastewater, air, environmental protection Keywords IC / Metrohm 944 Professional UV/VIS Detector

More information

Ion Chromatography (IC)

Ion Chromatography (IC) Ion Chromatography (IC) Purpose: This module provides an introduction to Ion Chromatography (IC). In this module the basic theory and applications of IC will be presented at a level that assumes a basic

More information

Ester Synthesis And Analysis: Aspirin and Oil of Wintergreen. Vanessa Jones November 19, 2015 Thursday 8:30 Lab Section Lab Partner: Melissa Blanco

Ester Synthesis And Analysis: Aspirin and Oil of Wintergreen. Vanessa Jones November 19, 2015 Thursday 8:30 Lab Section Lab Partner: Melissa Blanco Ester Synthesis And Analysis: Aspirin and Oil of Wintergreen Vanessa Jones November 19, 2015 Thursday 8:30 Lab Section Lab Partner: Melissa Blanco INTRODUCTION For this lab, students attempted to synthesize

More information

Antimony leaching from contaminated soil under manganese- and iron-reducing conditions: column experiments

Antimony leaching from contaminated soil under manganese- and iron-reducing conditions: column experiments Environ. Chem. 2014, 11, 624-631 CSIRO 2014 Supplementary material Antimony leaching from contaminated soil under manganese- and iron-reducing conditions: column experiments Kerstin Hockmann, A,C Susan

More information

Validated RP-HPLC Method for Estimation of Cefprozil in Tablet Dosage Form

Validated RP-HPLC Method for Estimation of Cefprozil in Tablet Dosage Form International Journal of PharmTech Research CDEN (USA): IJPRIF ISSN : 0974-4304 Vol.4, No.3, pp 1228-1232, July-Sept 2012 Validated RP-HPLC Method for Estimation of Cefprozil in Tablet Dosage Form Manzoor

More information

Determination of Inorganic Anions in Wastewater by Ion Chromatography

Determination of Inorganic Anions in Wastewater by Ion Chromatography Application Note 5 Determination of Inorganic Anions in Wastewater INTRODUCTION The determination of common inorganic anions in environmental waters, such as wastewater and drinking, ground, and surface

More information

TERTIARY BUTYLHYDROQUINONE

TERTIARY BUTYLHYDROQUINONE SYNONYMS TBHQ, INS No 319 DEFINITION TERTIARY BUTYLHYDROQUINONE Prepared at the 49th JECFA (1997), published in FNP 52 Add 5 (1997) superseding specifications prepared at the 44th JECFA (1995), published

More information

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath APPLICATIN NTE 9 Determination of an anionic fluorochemical surfactant in a semiconductor etch bath Authors Mark Laikhtman and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography,

More information

Application Note. An Optimum Solution for the Automatic Control of Cutting Oils

Application Note. An Optimum Solution for the Automatic Control of Cutting Oils An Optimum Solution for the Automatic Control of Cutting Oils Cutting oils are widely used in many mechanical machining processes such as cutting and grinding. The chemical and physical parameters of the

More information

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column Chromatography Thermo Scientific Dionex IonPac AS Anion-Exchange Column Product Specifications The Thermo Scientific Dionex IonPac AS high-capacity, carbonate based anion-exchange column is designed for

More information

A Simple, Novel Validated Stability Indicating RP-HPLC method for estimation of Duloxetine HCl in Capsule Pharmaceutical Formulation

A Simple, Novel Validated Stability Indicating RP-HPLC method for estimation of Duloxetine HCl in Capsule Pharmaceutical Formulation Pharmaceutical Research A Simple, Novel Validated Stability Indicating RP-HPLC method for estimation of Duloxetine HCl in Capsule Pharmaceutical Formulation Manisha Puranik* a, Sailesh Wadher b and Kritika

More information

DEVELOPMENT OF HPLC METHOD FOR ANALYSIS OF NITRITE AND NITRATE IN VEGETABLE

DEVELOPMENT OF HPLC METHOD FOR ANALYSIS OF NITRITE AND NITRATE IN VEGETABLE Journal of Agricultural, Food and Environmental Sciences UDC 635.546.173/.175]:543.544.5.068.7 Original scientific paper DEVELOPMENT OF HPLC METHOD FOR ANALYSIS OF NITRITE AND NITRATE IN VEGETABLE A. Najdenkoska*

More information

Department of Civil and Environmental Engineering, Stanford University, 473 Via Ortega, Stanford, California 94305, United States

Department of Civil and Environmental Engineering, Stanford University, 473 Via Ortega, Stanford, California 94305, United States Supporting Information for Capture and Reductive Transformation of Halogenated Pesticides by an Activated Carbon-Based Electrolysis System for Treatment of Runoff Yuanqing Li 1 and William A. Mitch 1,

More information

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water Thermo Scientific Dionex Ion Chromatography Solutions Global water safety bromate analysis in drinking water Safe drinking water public health assurance As a vital limited resource required for survival,

More information

2017 Reaction of cinnamic acid chloride with ammonia to cinnamic acid amide

2017 Reaction of cinnamic acid chloride with ammonia to cinnamic acid amide 217 Reaction of cinnamic acid chloride with ammonia to cinnamic acid amide O O Cl NH 3 NH 2 C 9 H 7 ClO (166.6) (17.) C 9 H 9 NO (147.2) Classification Reaction types and substance classes reaction of

More information

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS Issn No: 976-39 RESEARCH ARTICLE STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS CHETLAPALLI SATYA SRINIVAS 1, P.RENUKA DEVI 2 and GAMPA VIJAYAKUMAR*

More information

columns IonPac AS17-C Anion Exchange Column

columns IonPac AS17-C Anion Exchange Column columns IonPac AS-C Anion Exchange Column The IonPac AS-C is a hydroxide-selective anion exchange column designed for fast gradient separation of inorganic anions. The key application for the AS-C column

More information

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application Pharmaceuticals Author Cliff Woodward and Ronald Majors Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808

More information

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples Application Note 79 Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples INTRODUCTION Cow s milk is of particular dietary value to infants,

More information

Cu-Creatinine- Metol system

Cu-Creatinine- Metol system Quantification of Creatinine in Human Serum using Metol as a Chromogenic Probe Materials and methods 6.1. Reagents 6.1.1. N-methyl-p-aminophenol sulfate N-methyl-p-aminophenol sulfate also denoted as Metol

More information

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow 1. Application Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow Parameters and Codes: Nitrogen, ammonia, dissolved, I-2522-90 (mg/l as N): 00608 Nitrogen, ammonia, total-in-bottom-material,

More information

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8032 is used to determine trace amounts of acrylamide monomer (CAS No. 79-06-1) in aqueous matrices. This method may be

More information

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE 1.0 SCOPE AND APPLICATION 1.1 This method can be used for measuring total solubilized nitrate in drinking

More information

Overcoming Challenging Matrices in Ion Chromatography

Overcoming Challenging Matrices in Ion Chromatography Overcoming Challenging Matrices in Ion Chromatography Presented by: Kirk Chassaniol Gulf Coast Conference 2014 October 14, 2014 Thermo Fisher Scientific NA IC Tech Support 1 The world leader in serving

More information

INDIGOTINE. Disodium 3,3'-dioxo-[delta 2,2' -biindoline]-5,5'-disulfonate (principal component) (principal component)

INDIGOTINE. Disodium 3,3'-dioxo-[delta 2,2' -biindoline]-5,5'-disulfonate (principal component) (principal component) INDIGOTINE Prepared at the 73rd JEFA (2010) and published in FAO Monographs 10 (2010), superseding specifications prepared at the 28 th JEFA (1984) and published in the ombined ompendium of Food Additive

More information

LUMEFANTRINUM LUMEFANTRINE

LUMEFANTRINUM LUMEFANTRINE July 2008 LUMEFANTRINE: Final text for addition to The International Pharmacopoeia (July 2008) This monograph was adopted at the Forty-second WHO Expert Committee on Specifications for Pharmaceutical Preparations

More information

IC Columns Introduction

IC Columns Introduction IC Columns Introduction Alltech IC Column Specifications Column Composition ph Range Applications Anion Exchange Columns Allsep polymer-based anion ph 0 inorganic anions, weak and strong acid ions, exchanger,

More information

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures. User Manual IonPac NS2 Columns 065501 Revision 02 October 2015 For Research Use Only. Not for use in diagnostic procedures. Product Manual for IonPac NS2 Columns IonPac NS2, 5µm, 4x250mm Analytical Column

More information

Determination of Ions F -, Cl -, NO 3 -, SO 4 2-, PO 4 3- In The Blood Serum of Women With Breast Cancer Using Ion Chromatography Techniques

Determination of Ions F -, Cl -, NO 3 -, SO 4 2-, PO 4 3- In The Blood Serum of Women With Breast Cancer Using Ion Chromatography Techniques International Journal of Chemistry and Applications. ISSN 0974-3111 Volume 7, Number 1 (2015), pp. 25-31 International Research Publication House http://www.irphouse.com Determination of Ions F -, Cl -,

More information

columns IonPac AS18 Anion-Exchange Columns

columns IonPac AS18 Anion-Exchange Columns columns IonPac AS Anion-Exchange Columns The IonPac AS is a hydroxideselective anion-exchange column designed for the determination of inorganic anions and low-molecularweight organic acids including fluoride,

More information

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS Application Note Energy and Fuels - Biofuels Authors Ron Honnold, Ph.D. and Robert Kubas Agilent Technologies, Inc. Santa Clara, CA

More information

SIMPLE AND SENSITIVE VALIDATED REVERSE PHASE HPLC-UV METHOD FOR THE DETERMINATION OF LYMECYCLINE IN PHARMACEUTICAL DOSAGE FORMS

SIMPLE AND SENSITIVE VALIDATED REVERSE PHASE HPLC-UV METHOD FOR THE DETERMINATION OF LYMECYCLINE IN PHARMACEUTICAL DOSAGE FORMS IJPSR (2012), Vol. 3, Issue 07 (Research Article) Received on 26 March, 2012; received in revised form 25 June, 2012; accepted 28 June, 2012 SIMPLE AND SENSITIVE VALIDATED REVERSE PHASE HPLC-UV METHOD

More information

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV*

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV* Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV* EPA Method 8315A UCT Part Number: EUC1812M15 (Unendcapped C18-2000 mg/15 ml cartridge) March 2013 Method

More information

Glyphosate in all its forms

Glyphosate in all its forms Glyphosate in all its forms S. Goscinny and V. Hanot Scientific Institute for Public Health Pesticides Unit A breakthrough molecule Since its first appearance on the market in 1974, glyphosate has become

More information

Data Pack. Ion Chromatography Methods. June 2009

Data Pack. Ion Chromatography Methods. June 2009 Data Pack Ion Chromatography Methods June 2009 Table of Contents Introduction. 2 Ion Chromatography Methods List. 3 Inorganic Anions in Waters..7 Inorganic Anions in High Purity Waters..15 Inorganic Disinfection

More information

MEDROXYPROGESTERONE INJECTION

MEDROXYPROGESTERONE INJECTION September 2011 RESTRICTED DRAFT PROPOSAL FOR The International Pharmacopoeia MEDROXYPROGESTERONE INJECTION (September 2011) DRAFT FOR COMMENT Should you have any comments on the attached draft proposal,

More information

Ion Analysis of Hydraulic Fluids in Fracking Process

Ion Analysis of Hydraulic Fluids in Fracking Process Ion Analysis of Hydraulic Fluids in Fracking Process Presented by Jay Gandhi *, PhD Metrohm USA NEMC-2013 1 NEMC-2013 2 What we will cover today Background Hydraulic Fracking Ion Analysis Applications

More information

Anion Analysis of Ophthalmic Irrigation Solutions with the Shimadzu Prominence IC System Craig Young, PhD, Sr. Tech Support Specialist

Anion Analysis of Ophthalmic Irrigation Solutions with the Shimadzu Prominence IC System Craig Young, PhD, Sr. Tech Support Specialist High Performance Liquid Chromatography SSI HPLC 5 Anion Analysis of Ophthalmic Irrigation Solutions with the Shimadzu Prominence IC System Craig Young, PhD, Sr. Tech Support Specialist Introduction Ophthalmic

More information

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur Combustion Ion Chromatography Fast and reliable determination of halogens and sulfur Combustion digestion and ion chromatography combined in one system 02 Combustion Ion Chromatography (CIC) extends the

More information

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection Terri Christison, Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Overview

More information

Analysis of pulping liquors by suppressed ion chromatography

Analysis of pulping liquors by suppressed ion chromatography T 699 om-87 PROVISIONAL METHOD 1983 OFFICIAL TEST METHOD 1987 1987 TAPPI The information and data contained in this document were prepared by a technical committee of the Association. The committee and

More information

4. Ion chromatography (IC) 4.1. Educational aims and objectives

4. Ion chromatography (IC) 4.1. Educational aims and objectives 4. Ion chromatography (IC) 4.1. Educational aims and objectives Principles of ion chromatography method and instrumentation. Methods for ionic compounds analysis including ion chromatography and its subtypes.

More information

Preparation and characterisation of a sorbent suitable for technetium separation from environmental matrices

Preparation and characterisation of a sorbent suitable for technetium separation from environmental matrices Preparation and characterisation of a sorbent suitable for technetium separation from environmental matrices A. Bartosova, P. Rajec, M. Reich Faculty of Natural Sciences, Department of Nuclear chemistry,

More information

Ion Chromatography. Anion Exchange. Chromatography Ion Exchange Theory. Dr. Shulamit Levin

Ion Chromatography. Anion Exchange. Chromatography Ion Exchange Theory. Dr. Shulamit Levin Ion Exchange Chromatography Chromatographic Process BA Mobile phase Stationary Phase A Shula Levin Bioforum B Distribution: K = C s/c m B shulal@zahav.net.il http://shulalc.co.il/ A Elution through the

More information

Chapter 4: Verification of compendial methods

Chapter 4: Verification of compendial methods Chapter 4: Verification of compendial methods Introduction In order to ensure accurate and reliable test results, the quality control laboratory (QCL) needs to use analytical methods (and accompanying

More information