SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33

Size: px
Start display at page:

Download "SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33"

Transcription

1 Photometer-System SC400 Colorimeter SiO 2 Silica DE GB FR IT ES Bedienungsanleitung Seite 3 17 Instruction Manual Page d'emploi Page Instruction Manual Istruzioni d'uso Pagina Instrucciones Página 67 81

2 Requirements Method: Silicomolybdate Method Description: Under acidic conditions, silica and phosphate in the sample react with molydate ions in the reagent to form yellow colored chemical complexes. The addition of the citric acid reagent removes phosphate interferences. The level of the silica is then determined by measuring the remaining yellow color of the sample. Required Instrument: SC400 Colorimeter for Silica (P/N: SC400-90) Required Accessories: Sample Cell, Round Glass (Ø 24 mm) (P/N: L197620) Required Reagents: Range: Reagent Description: Reagent P/N: mg/l SiO 2 Silica No. 1 Tablet RT313-0 Silica No. 2 Tablet Silica PR Tablet RT314-0 RT mg/l SiO 2 Molybdate 3 Reagent Solution L Silica Citric Acid Powder Pack Silica Amino Acid Powder Pack RP171-0 RP mg/l SiO 2 Silica HR Molybdate Powder Pack RP172-0 Silica HR Acid Reagent Powder Pack Silica Citric Acid Powder Pack RP170-0 RP

3 Contents General notes...20 Guidelines for photometric measurements Method notes...20 Battery change Functional description...22 Operation...22 Backlight...23 Measuring the absorbance Methods Silicon dioxide, with tablet Silicon dioxide LR with Powder Reagent and Liquid Reagent Silicon dioxide HR with Powder Reagent Menu options...30 Menu selections Recall of stored data...30 Setting date and time Calibration...31 User calibration...31 Recommended calibration values Factory calibration reset Technical data...33 Operating messages Error codes CAUTION The accuracy of the instrument is only valid if the instrument is used in an environment with controlled electromagnetic disturbances according to DIN Wireless devices, e.g. wireless phones, must not be used near the instrument. 3

4 General notes Guidelines for photometric measurements 1. Vials, caps and stirring rods should be cleaned thoroughly after each analysis to prevent interferences. Even minor reagent residues can cause errors in the test result. 2. The outside of the vial must be clean and dry before starting the analysis. Clean the outside of the vials with a cleaning cloth to remove fingerprints and other marks. 3. calibration and the testing itself must be carried out with the same vial as there may be slight differences in optical performance between vials. 4. The vials must be positioned in the sample chamber for zeroing and test with the mark on the vial aligned with the mark on the instrument. 5. Always perform zeroing and test with the vial cap tightly closed. Only use the cap with a sealing ring. 6. Bubbles on the inside wall of the vial leads to incorrect measurements. To prevent this, remove the bubbles by swirling the vial before performing the test. 7. Avoid spilling water into the sample chamber because this can lead to incorrect test results. 8. Contamination of the lens in the sample chamber can result in errors. Check at regular intervals and if necessary clean the light entry surfaces of the sample chamber using a moist cloth or cotton buds. 9. Large temperature differences between the instrument and the environment can lead to errors due to the formation of condensation in the area of the lens or on the vial. 10. To avoid errors caused by stray light do not use the instrument in bright sunlight. 11. Always add the reagent tablets to the water sample straight from the foil without touching them with your fingers. 12. The reagents must be added in the correct sequence. Method notes Prior to measurement ensure that the sample is suitable for analysis (no major interferences) and does not require any preparation i.e. ph adjustment, filtration etc. Reagents are designed for use in chemical analysis only and should be kept out of the reach of children. Ensure proper disposal of reagent solutions. Material Safety Data Sheets: available on our web site, or upon request. 4

5 General notes Correct position of the vial (Ø 24 mm) Correct filling of the vial: Opening powder packs 1 2 B A 3 4 Battery change NOTE : If the battery is removed for more than one minute, the date and time menu starts automatically when the colorimeter is next turned on. 5

6 Functional description Operation On Off SIt Turn the unit on using the [ON/OFF] key. The display shows the following: Note: The unit is designed to measure Silica (SiO 2 ) either with tablet reagents or with powder packs. Please select the required method as described below. METHOD METHOD METHOD RESULT Select the required test using the [MODE] key: SIt (Silica tablet) SILr (Silica Low range) SIHr (Silica High range) AbS1 (Absorbance1) AbS2 (Absorbance2) SIt (Silica tablet)... (Scroll) The display shows the following: Fill a clean vial with the water sample up to the 10 ml mark, screw the cap on and place the vial in the sample chamber making sure that the marks are aligned. Press the [ZERO/TEST] key. The "Method" symbol flashes for approx. 8 seconds. The display shows the following: After zero calibration is completed, remove the vial from the sample chamber. When reagents are added, a color will develop if silica is present. Close the vial again and place in the sample chamber making sure that the marks are aligned. Press the [ZERO/TEST] key. The "Method" symbol flashes for approx. 3 seconds. The result appears in the display. The result is saved automatically. Repeating the test: Press the [ZERO/TEST] key again. New zero calibration: Press the [MODE] key until the desired "Method" symbol appears in the display again. 6

7 Functional description Display backlight! Press the [!] key to turn the display backlight on or off. The backlight turns off automatically during the measurement. Recall of stored data! If the instrument is turned on, press the [!] key for more than 4 seconds to access the recall menu. Measuring the absorbance AbS This function allows measurement of the absorbance at the specified wavelength. The result appears in the display in mabs (1000 mabs = 1 A (Absorbance unit)). Abs1: = 430 nm Abs2: = 660 nm 7

8 Methods Silicon dioxide with Tablet mg/l SiO SIt RESULT 1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample and perform zero calibration (see Operation ). 2. Add one SILICA No. 1 tablet (P/N: RT313-0) straight from the foil to the water sample and crush the tablet using a clean stirring rod. 3. Close the vial tightly with the cap and swirl several times until the tablet is dissolved. Wait for a reaction period of 5 minutes. After the reaction period is finished proceed as follows: 4. Add one SILICA PR tablet (P/N: RT315-0) straight from the foil to the same water sample and crush the tablet using a clean stirring rod. 5. Add one SILICA No. 2 tablet (P/N: RT314-0) straight from the foil to the same water sample and crush the tablet using a clean stirring rod. 6. Close the vial tightly with the cap and swirl several times until the tablet is dissolved. 7. Place the vial in the sample chamber making sure that the marks are aligned. Wait for a reaction period of 1 minute. 8. Press the [ZERO/TEST] key. The method symbol flashes for approx. 3 seconds. The result is shown in the display in mg/l SiO 2. Accuracy: ± 0.1 mg/l 8

9 Methods Notes: 1. The tablets must be added in the correct sequence. 2. Phosphate ions do not interfere under the given reaction conditions. 3. If Phosphate is known to be absent, the addition of the SILICA PR tablet may be omitted. 9

10 Methods Silicon dioxide LR with Powder Reagent and Liquid Reagent mg/l SiO 2 SILr Use two clean vials (24 mm Ø) and mark one as blank for zeroing. 1. Fill each vial with 10 ml of water sample. 2. Fill the vial with drops of the same size by holding the bottle vertical and squeeze slowly (Note 5): 3. Add 14 drops of Molybdate 3 Reagent Solution (P/N: L531730) into each vial. 4. Close the vials tightly with the caps and swirl the vials several times to mix the contents (Note 1). Wait for a reaction period of 4 minutes (Note 2). After the reaction period is finished proceed as follows: 5. Add the contents of one Silica Citric Acid F10 Powder Pack (P/N: RP171-0) straight from the foil into each vial. 6. Close the vials tightly with the caps and swirl the vials several times to mix the contents. Wait for a reaction period of 1 minute (Note 3). After the reaction period is finished proceed as follows: 7. Add the contents of one LR Silica Amino Acid F10 Powder Pack (P/N: RP160-0) straight from the foil into the sample vial. Wait for a reaction period of 2 minutes. 8. Place the vial (the blank) in the sample chamber making sure that the marks are aligned. SILr SILr RESULT 9. Press the [ZERO/TEST] key. The method symbol flashes for approx. 8 seconds. 10. Remove the vial from the sample chamber. 11. Place the vial (the sample) in the sample chamber making sure that the marks are aligned. 12. Press the [ZERO/TEST] key. The method symbol flashes for approx. 3 seconds. The result is shown in the display in mg/l SiO 2. Accuracy: ± 0.1 mg/l 10

11 Methods Notes: 1. Close the vials with the cap immediately after adding the Molybdate 3 Reagent Solution; otherwise low readings may result. 2. The given reaction time of 4 minutes refers to a water sample temperature of 68 F (20 C). At 86 F (30 C) a reaction time of 2 minutes, at 50 F (10 C) a reaction time of 8 minutes are required. 3. The given reaction time of 1 minute refers to a water sample temperature of 68 F 20 C. At 86 F (30 C) a reaction time of 30 seconds, at 50 F (10 C) a reaction time of 2 minutes are required. 4. Interferences: Substance Iron Phosphate Sulfide Interference large amounts interfere does not interfere at concentrations less than 50 mg/l PO 4 at 60 mg/l PO 4 the interference is approx. 2% at 75 mg/l PO 4 the interference is approx. 1% interferes at all levels Occasionally water samples contain forms of silica which react very slowly with Molybdate. A pre-treatment with Sodium hydrogencarbonate and then with Sulfuric Acid will make these forms reactive to Molybdate (pretreatment is given in Standard Methods for the Examination of Water and Wastewater under Silica Digestion with Sodium Bicarbonate ). 5. Differing drop size results can interfere with the stated accuracy. This can be minimized by using a pipette. (14 drops is equivalent to 0.5 ml). 11

12 Methods Silicon dioxide HR with Powder Reagent mg/l SiO SIHr RESULT 1. Fill a clean vial (24 mm Ø) with 10 ml of the water sample and perform zero calibration (see Operation ). 2. Add the contents of one Silica HR Molybdate F10 Powder Pack (P/N: RP172-0) straight from the foil to the water sample. 3. Close the vial tightly with the cap and swirl several times to mix the contents. 4. Add the contents of one Silica HR Acid Rgt. F10 Powder Pack (P/N: RP170-0) straight from the foil to the same water sample. (Note 2) 5. Close the vial with the caps tightly and swirl the vials several times to mix the contents. Wait for a reaction period of 10 minutes. After the reaction period is finished proceed as follows: 6. Add the contents of one Silica Citric Acid F10 Powder Pack (P/N: RP171-0) straight from the foil to the water sample. (Note 3) 7. Close the vial tightly with the cap and swirl several times to mix the contents. 8. Place the vial in the zero chamber making sure that the marks are aligned. Wait for a reaction period of 2 minutes. 9. Place the vial in the sample chamber making sure that the marks are aligned. 10.Press the [ZERO/TEST] key. The method symbol flashes for approx. 3 seconds. The result is shown in the display in mg/l SiO 2. Accuracy: 0 20 mg/l: ± 1 mg/l > mg/l: ± 2 mg/l > mg/l: ± 3 mg/l > mg/l: ± 5 mg/l 12

13 Methods Notes: 1. Temperature of the sample should be 59 F 77 F (15 C 25 C). 2. If Silica or Phosphate is present a yellow color is developed. 3. In this step any yellow color due to Phosphate is removed. 4. Interferences: Substance Iron Phosphate Sulfide Interference large amounts interfere does not interfere at concentrations less than 50 mg/l PO 4 at 60 mg/l PO 4 the interference is approx. 2% at 75 mg/l PO 4 the interference is approx. 1% interferes at all levels Occasionally water samples contain forms of silica which reacts very slowly with Molybdate. A pre-treatment with Sodium hydrogencarbonate and then with Sulfuric Acid will make these forms reactive to Molybdate (pretreatment is given in Standard Methods for the Examination of Water and Wastewater under Silica Digestion with Sodium Bicarbonate ). 13

14 Menu options Store Time On Off! Date Cal Menu selections 1. Press the [MODE] key and hold. 2. Turn the unit on using the [ON/OFF] key. Hold (or depress) the [] key until the 3 decimal points are displayed. The [!] key allows for selection of the following menu points: 1 recall stored data 2 3 setting the date and time 4 user calibration The selected menu is indicated by an arrow in the display. Confirm the selection with the [MODE] key. 1 Recall of stored data! The photometer shows the last 16 data sets in the following format (automatically proceeds every 3 seconds until result is displayed): Number n xx (xx: ) Year YYYY (i.e. 2008) Date mm.dd (monthmonth:dayday) Time hh:mm (hourhour:minuteminute) Method Result x,xx The [ZERO/TEST] key repeats the current data set. The [MODE] key scrolls through all stored data sets. Quit the menu by pressing [!] key. SET DATE YYYY (2 sec.)! 2 3 Setting date and time (24-hour-format) After confirming the selection with the [MODE] key the value to be edited will be shown for 2 sec. The setting starts with the year (YYYY) followed by the actual value to be edited. Same applies for month (mm), day (dd), hour (hh) and minutes (mm). Set the minutes first in steps of 10, press the [!] key to continue setting the minutes in steps of 1. Increase the value by pressing the [MODE] key. Decrease the value by pressing [ZERO/TEST] key. Proceed to the next value to be edited by pressing [!] key. After setting the minutes and pressing the [!] key the display will show IS SET and instrument returns to the measurement mode. 14

15 Calibration 4 User calibration cal CAL CAL SIt METHOD CAL METHOD RESULT CAL CAL RESULT + x On Off : : Note: user calibration (Display in calibration mode) factory calibration (Display in calibration mode) After confirming the selection with the [MODE] key the instrument will show CAL/SIt. Scroll through methods using the [MODE] key. Fill a clean vial with the standard up to the 10 ml mark, screw the cap on and place the vial in the sample chamber making sure that the marks are aligned. Press the [ZERO/TEST] key. The method symbol flashes for approx. 8 seconds. The display shows the following in alternating mode: Perform calibration with a standard of known concentration (see "Operation"). Press the [ZERO/TEST] key. The method symbol flashes for approx. 3 seconds. The result is shown in the display, alternating with CAL. If the reading corresponds with the value of the calibration standard (within the specified tolerance), exit calibration mode by pressing the [ON/OFF] key. Changing the displayed value: Pressing the [MODE] key once increases the displayed value by 1 digit. Pressing the [ZERO/TEST] key once decreases the displayed value by 1 digit. Press the corresponding key until the reading equals the value of the calibration standard. By pressing the [ON/OFF] key, the new correction factor is calculated and stored in the user calibration software. Confirmation of calibration (3 seconds). Recommended calibration values Tablet: between 0.5 and 1.5 mg/l SiO 2 Powderpack LR: 1mg/L SiO 2 Powderpack HR: 50mg/L SiO 2 15

16 Calibration Factory calibration reset Resetting the user calibration to the original factory calibration will reset all methods and ranges. Store Time RESULT On Off Date Cal A user calibrated method is indicated by an arrow while the test result is displayed. To reset the calibration press both the [MODE] and [ZERO/TEST] key and hold. Turn the unit on using the [ON/OFF] key. Release the [MODE] and [ZERO/TEST] keys after approx. 1 second. The following messages will appear in turn on the display: SEL CAL The factory setting is active. (SEL stands for Select) or: SEL cal Calibration has been set by the user. (If the user calibration is to be retained, turn the unit off using the [ON/OFF] key). SEL CAL On Off Calibration is reset to the factory setting by pressing the [MODE] key. The following messages will appear in turn on the display: Turn the unit off using the [ON/OFF] key. 16

17 Technical data Operating messages Light source: Battery: Auto-OFF: Technical Data LED, Filter ( = 430 nm) LED ( = 660 nm) 9 V-block battery (Life approx. 600 tests), without display light. Automatic turn off 20 minutes after last keypress Ambient conditions: 41 F 104 F (5 40 C) 30 90% rel. humidity (non-condensing) CE: Certificate for Declaration of CE-Conformity at To ensure maximum accuracy of test results, always use the reagent systems supplied by the instrument manufacturer. Operating messages Hı Lo Measuring range exceeded or excessive turbidity. Result below the lowest limit of the measuring range. Replace battery. Error codes E 01 E 10 / E 11 E 20 /E 21 E 22 Light absorption too great. Reasons: e.g. dirty optics. Calibration factor "out of range" Too much light reaching the detector. Battery capacity was too low during measurement. Change battery. E 70 E 71 E 72 E 73 E 74 E 75 SIt: SIt: SILr: SILr: SIHr: SIHr: Factory calibration incorrect / erase User calibration incorrect / erase Factory calibration incorrect / erase User calibration incorrect / erase Factory calibration incorrect / erase User calibration incorrect / erase For assistance with Error Codes or other technical issues, please contact Customer Service. (Phone: or Toll-free U.S ) 17

18

19 Technical changes without notice FOR TECHNICAL ASSISTANCE, PRICE AND ORDERING INFORMATION: In the USA - Call or the distributor serving you Outside the USA: Call or the distributor serving you. Web: service@orbeco.com Fax: Orbeco-Hellige, Inc, All rights reserved. Printed in Germany. Updated March 2009, Edition 1

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters) Silica DOC316.53.01132 Heteropoly Blue Method 1 Method 8186 0.010 to 1.600 mg/l SiO 2 (LR, spectrophotometers) Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters) Scope and application: For boiler

More information

SC450. Single Parameter Colorimeter Chlorine & ph. Bedienungsanleitung Seite Instruction Manual Page Mode d'emploi Page 39 55

SC450. Single Parameter Colorimeter Chlorine & ph. Bedienungsanleitung Seite Instruction Manual Page Mode d'emploi Page 39 55 SC450 Single Parameter Colorimeter Chlorine & ph DE GB FR IT ES Bedienungsanleitung Seite 3 19 Instruction Manual Page 21 37 Mode d'emploi Page 39 55 Istruzioni d'uso Pagina 57 73 For use Instrucciones

More information

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated. Silicate LR VARIO PP 0.05 -.6 mg/l SiO 2 Heteropolyblue 35 SiLr Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption

More information

SILICA, Low Range (0 to 1.60 mg/l)

SILICA, Low Range (0 to 1.60 mg/l) SILICA, Low Range (0 to 1.60 mg/l) Heteropoly Blue Method * Method 8186 For water and seawater 1. Enter the stored program number for low range silica (SiO 2 ). Press: PRGM The display will show: PRGM?

More information

Photometer AL250. Scroll Memory (SM) N.I.S.T Traceability. Highlights

Photometer AL250. Scroll Memory (SM) N.I.S.T Traceability. Highlights Photometer AL250 10 The AL250 models each offer one test method, while the AL250+ units are programmed for a number of different parameters. The units supply accurate, reproducible results very quickly.

More information

DULCOTEST DT1B Photometer

DULCOTEST DT1B Photometer Operating manual DULCOTEST DT1B Photometer A0943 Please carefully read these operating instructions before use! Do not discard! The operator shall be liable for any damage caused by installation or operating

More information

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water,

More information

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Nitrate, HR DOC316.53.01068 Chromotropic Acid Method Method 10020 0.2 to 30.0 mg/l NO 3 N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832 , TNTplus 832, 10205 DOC316.53.01083 Salicylate Method Method 10205 HR (2 to 47 mg/l NH 3 N) TNTplus 832 Scope and Application: For surface waters, municipal and industrial wastewaters. Test preparation

More information

Chloramine (Mono) and Nitrogen, Free Ammonia

Chloramine (Mono) and Nitrogen, Free Ammonia Chloramine (Mono) and Nitrogen, Free Ammonia DOC316.53.01016 Indophenol Method 1 Method 10200 0.04 to 4.50 mg/l Cl 2 0.01 to 0.50 mg/l NH 3 N Powder Pillows Scope and application: For the determination

More information

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ ) DR/4000 PROCEDURE Method 8023 1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to 0.700 mg/l Cr 6+ ) Scope and Application: For water and wastewater;usepa accepted for reporting for

More information

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows DR/4000 PROCEDURE Method 8171 Powder Pillows or AccuVac Ampuls Cadmium Reduction Method MR (0 to 5.0 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit

More information

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows Nitrate DOC316.53.01067 Cadmium Reduction Method Method 8192 0.01 to 0.50 mg/l NO 3 N (LR) Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

NITROGEN, TOTAL, HR, Test N Tube

NITROGEN, TOTAL, HR, Test N Tube Method 10072 NITROGEN, TOTAL, HR, Test N Tube 10 to 150 mg/l N TNT Persulfate Digestion Method For water and wastewater Digestion is required for determining total nitrogen. The digestion procedure is

More information

Mercaptoacetic Acid Method*

Mercaptoacetic Acid Method* Method 8036 MOLYBDENUM, MOLYBDATE, High Range (0 to 40.0 mg/l) Mercaptoacetic Acid Method* Using Powder Pillows For water and wastewater 1. Enter the stored program number for high range molybdenumpowder

More information

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Aluminum DOC316.53.01002 Aluminon Method 1 Method 8012 0.008 to 0.800 mg/l Al 3+ (spectrophotometers) Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters) Scope and application: For water and wastewater.

More information

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure Methylene Blue Method 1 (0 0.70, 0 7.00 and 0 70.00 mg/l) DOC316.53.01320 Method 10254 Scope and Application: For oil and gas field waters. 1 Adapted from Standard Methods for the Examination of Water

More information

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows FILL LINE Method 8155 Salicylate Method 1 Powder Pillows Scope and Application: For water, wastewater, and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) (0.01 to 0.50 mg/l NH 3 N) Test Preparation

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right. Barium DOC316.53.01315 Turbidimetric Method 1 Method 10251 2 to 100, 20 to 1000, 200 to 10,000 mg/l Ba (spectrophotometers) Powder Pillows 2 to 80, 20 to 800, 200 to 8000 mg/l Ba (colorimeters) Scope and

More information

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Nitrate, MR DOC316.53.01069 Cadmium Reduction Method Method 8171 0.1 to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters) Scope and application: For water, wastewater

More information

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 , MR, 8171 DOC316.53.01069 Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3 N) Powder Pillows or AccuVac Ampuls Scope and Application: For water, wastewater and seawater Test preparation

More information

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Chromium, Total DOC316.53.01034 Alkaline Hypobromite Oxidation Method 1 Method 8024 0.01 to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters) Scope and application: For water and wastewater.

More information

SILICA, Low Range SiLR

SILICA, Low Range SiLR SILICA, Low Range SiLR SP-910 Test Method Test Program Description:SP-910 Silica Low Range Method (0.02 1.60 ppm SiO 2 ) (Heteropoly Blue Method) Instruments and Reagents: Program 1. SP-910 Portable Water

More information

Nitrogen, Total Inorganic

Nitrogen, Total Inorganic Nitrogen, Total Inorganic DOC316.53.01090 Titanium Trichloride Reduction Method Method 10021 0.2 to 25.0 mg/l N Test N Tube Vials Scope and application: For water, wastewater and seawater. Test preparation

More information

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows , 8155 Salicylate Method 1 Scope and Application: For water, wastewater and seawater 1 Adapted from Clin. Chim. Acta., 14, 403 (1966) DOC316.53.01077 Method 8155 0.01 to 0.50 mg/l NH 3 N Powder Pillows

More information

Chlorine, Free and Total, High Range

Chlorine, Free and Total, High Range Chlorine, Free and Total, High Range DOC316.53.01490 USEPA DPD Method 1 Method 10069 (free) 10070 (total) 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of free

More information

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 ) Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 ) DOC316.53.01523 DPD Method Method 10290 0.10 to 10.00 mg/l PAA / 0.05 to 5.00 mg/l H 2 O 2 Powder Pillows Scope and application: For testing peracetic

More information

CHLORINE, TOTAL (0 to 4.00 mg/l)

CHLORINE, TOTAL (0 to 4.00 mg/l) CHLORINE, TOTAL (0 to 4.00 mg/l) DOC316.53.01261 For water, wastewater, and seawater Method 10250 DPD Method Powder Pillows USEPA accepted for reporting water and wastewater analyses * Note: This product

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right. Nitrite DOC316.53.01074 USEPA Diazotization Method 1 Method 8507 0.002 to 0.300 mg/l NO 2 N (LR, spectrophotometers) 0.005 to 0.350 mg/l NO 2 N (LR, colorimeters) Scope and application: For water, wastewater

More information

Photometer-System MD100

Photometer-System MD100 Lovibond Water ing Tintometer Group Photometer-System MD100 Cl ph CyA TA CaH DE GB FR Bedienungsanleitung Seite 4 27 Instruction Manual Page 28 51 Mode d emploi Page 52 75 IT ES PT Istruzioni d uso Pagina

More information

NITROGEN, AMMONIA, High Range, Test N Tube

NITROGEN, AMMONIA, High Range, Test N Tube NITROGEN, AMMONIA, High Range, Test N Tube Method 10031 (0 to 50 mg/l NH 3 -N) For water, wastewater, and seawater Salicylate Method * 1. Enter the stored program number for nitrogen, ammonia, high range

More information

Operating Manual. Turb 355 IR/T. Portable turbidimeter. ba75454e03 09/

Operating Manual. Turb 355 IR/T. Portable turbidimeter. ba75454e03 09/ Operating Manual Turb 355 IR/T Portable turbidimeter ba75454e03 09/2010 41 Accuracy when going to press The use of advanced technology and the high quality standard of our instruments are the result of

More information

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials Nitrogen, Total DOC316.53.01086 Persulfate Digestion Method Method 10071 0.5 to 25.0 mg/l N (LR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific

More information

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials Nitrogen, Total DOC316.53.01085 Persulfate Digestion Method Method 10072 2 to 150 mg/l N (HR) Test N Tube Vials Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Phosphate HR T 1-80 mg/l P Vanadomolybdate

Phosphate HR T 1-80 mg/l P Vanadomolybdate Phosphate HR T 1-80 mg/l P Vanadomolybdate 321 Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer

More information

Chlorine, Free and Total, High Range

Chlorine, Free and Total, High Range Chlorine, Free and Total, High Range DOC316.53.01449 USEPA DPD Method 1 Method DPD 0.1 to 8.0 mg/l Cl 2 Powder Pillows Scope and application: For testing residual chlorine and chloramines in water, wastewater,

More information

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows Potassium DOC316.53.01127 Tetraphenylborate Method Method 8049 0.1 to 7.0 mg/l K Powder Pillows Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific information

More information

PAR Method Method to 2.0 mg/l Pb TNTplus 850

PAR Method Method to 2.0 mg/l Pb TNTplus 850 , TNTplus 850, 10216 DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and Application: For wastewater and process control Test preparation Before starting the test: FOR DR 2800:

More information

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE NITRATE. 2. The display will show: Method 8192 Powder Pillows DR/4000 PROCEDURE Cadmium Reduction Method LR (0 to 0.50 mg/l NO 3 N) Scope and Application: For water, wastewater and seawater. The estimated detection limit for program number

More information

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated. Nitrogen total HR VARIO TT 5-150 mg/l N b) 81 Persulphate Digestion Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption

More information

Instruction Manual C401/301/201/102/101. Portable Colorimeter. 99 Washington Street Melrose, MA Phone Toll Free

Instruction Manual C401/301/201/102/101. Portable Colorimeter. 99 Washington Street Melrose, MA Phone Toll Free Instruction Manual Portable Colorimeter 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 68X357704 Rev. 0 08/04 Technology Made

More information

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator Alkalinity DOC316.53.01166 Phenolphthalein and Total Alkalinity Method 8203 10 to 4000 mg/l as CaCO 3 Digital Titrator Scope and application: For water, wastewater and seawater. Test preparation Before

More information

Instruction Manual. Waterproof Portable Colorimeter C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105. Part of Thermo Fisher Scientific

Instruction Manual. Waterproof Portable Colorimeter C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105. Part of Thermo Fisher Scientific Instruction Manual C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105 Waterproof Portable Colorimeter Technology Made Easy... 68X357704 Rev 2 05/08 Part of Thermo Fisher Scientific Preface This manual functions

More information

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827 Nitrogen, Total DOC316.53.01088 Persulfate Digestion Method Method 10208 5 to 40 mg/l N (HR) TNTplus 827 Scope and application: For water and wastewater. Test preparation Instrument-specific information

More information

Nitrogen total LR VARIO TT mg/l N b) Persulphate Digestion

Nitrogen total LR VARIO TT mg/l N b) Persulphate Digestion Nitrogen total LR VARIO TT 0.5-25 mg/l N b) Persulphate Digestion 280 Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption

More information

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows , 8326 Eriochrome Cyanine R Method 1 Scope and Application: For water 1 Adapted from Standard Methods for the Examination of Water and Wastewater. DOC316.53.01003 Method 8326 (0.006 to 0.250 mg/l Al 3+

More information

DR/4000 PROCEDURE. IRON, Total

DR/4000 PROCEDURE. IRON, Total Method 8365 Powder Pillows DR/4000 PROCEDURE FerroMo Method* (0 to 1.800 mg/l) Scope and Application: For cooling water containing molybdate-based treatment; digestion is required for determining total

More information

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right. Aluminum DOC316.53.01003 Eriochrome Cyanine R Method 1 Method 8326 0.006 to 0.250 mg/l Al 3+ Powder Pillows Scope and application: For water. 1 Adapted from Standard Methods for the Examination of Water

More information

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution Chloride DOC316.53.01017 Mercuric Thiocyanate Method Method 8113 0.1 to 25.0 mg/l Cl Reagent Solution Scope and application: For water and wastewater. Test preparation Instrument-specific information Table

More information

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated. Chlorine L a) 0.02-4.0 mg/l Cl 2 DPD 101 CL6 Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer

More information

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater. Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for

More information

Rate Law Determination of the Crystal Violet Reaction. Evaluation copy

Rate Law Determination of the Crystal Violet Reaction. Evaluation copy Rate Law Determination of the Crystal Violet Reaction Computer 30 In this experiment, you will observe the reaction between crystal violet and sodium hydroxide. One objective is to study the relationship

More information

Chematest 25. Operator s Manual A Swan Analytical Instruments AG CH-8340 Hinwil/Switzerland ANALYTICAL INSTRUMENTS

Chematest 25. Operator s Manual A Swan Analytical Instruments AG CH-8340 Hinwil/Switzerland ANALYTICAL INSTRUMENTS Chematest 25 Operator s Manual A70.065.611 ANALYTICAL INSTRUMENTS Swan Analytical Instruments AG CH-8340 Hinwil/Switzerland Table of Contents Photometry 1.1 Basic Rules for Photometric Determinations 5

More information

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows Silver DOC316.53.01134 Colorimetric Method Method 8120 0.02 to 0.70 mg/l Ag Powder Pillows Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Method 8171 NITRATE, Mid Range (0 to 5.0 mg/l NO 3- -N) For water, wastewater and seawater* Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls) Using Powder Pillows 1. Enter the stored program

More information

LOW-RANGE TOTAL CHLORINE

LOW-RANGE TOTAL CHLORINE LOW-RANGE TOTAL CHLORINE Thermo Orion Method AC2072 Revision 4 3/14/2002 Accepted by U.S. EPA on March 20, 2002 for wastewater compliance monitoring Accepted by U.S. EPA on September 31, 2002 for drinking

More information

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV Iron, Total DOC316.53.01338 FerroZine Method 1 Method 10264 1 to 100 µg/l Fe (10-cm cell) Reagent Solution Scope and application: For ultrapure water. 1 Adapted from Stookey, L.L., Anal. Chem., 42(7),

More information

Manganese III Digestion Method * (with chloride removal)

Manganese III Digestion Method * (with chloride removal) Method 10067 OXYGEN DEMAND, CHEMICAL (20 to 1,000 mg/l) For water and wastewater Manganese III Digestion Method * (with chloride removal) 1. Enter the stored program number for Manganese III COD. Press:

More information

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO Nitrate DOC316.53.01070 Dimethylphenol Method Method 10206 0.23 to 13. NO 3 N or 1.00 to 60.00 mg/l NO 3 (LR) TNTplus 835 Scope and application: For wastewater, drinking water, surface water and process

More information

NITROGEN, TOTAL, Test N Tube * Method 10071

NITROGEN, TOTAL, Test N Tube * Method 10071 NITROGEN, TOTAL, Test N Tube * Method 10071 TNT Persulfate Digestion Method (0 to 25.0 mg/l N) Scope and Application: For water and wastewater; digestion is required for determining total nitrogen. The

More information

Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer

Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for

More information

Chlorine, Free and Total, Low Range

Chlorine, Free and Total, Low Range Chlorine, Free and Total, Low Range DOC316.53.01450 USEPA DPD Method 1 Method 8021 (free) 8167 (total) 0.02 to 2.00 mg/l Cl 2 (LR) Powder Pillows or AccuVac Ampuls Scope and application: For testing residual

More information

Vet Photometer DP 700 Operating Manual. Version V5.9

Vet Photometer DP 700 Operating Manual. Version V5.9 1 Vet Photometer DP 700 Operating Manual Version V5.9 2 Dear customer, Thank you for selecting Diaglobal s Vet Photometer and for the confidence you have placed in us. The Vet Photometer belongs to a new

More information

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA , 10255 DOC316.53.01322 Direct Measurement ISE Method Method 10255 3.55 g/l to 35 g/l Cl Powder Pillow ISA Scope and Application: For the determination of high concentrations (1 M) of chloride in brine

More information

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter Water and Lab Products, Thermo Fisher Scientific Technical Note 505 Key Words Thermo Scientific Orion

More information

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell Hardness DOC316.53.01044 Chlorphosphonazo Colorimetric Method Method 8374 8 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow Scope and application: For boiler and ultrapure water. Test preparation

More information

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water. Method 8337 Powder Pillows DR/4000 TT PROCEDURE QUATERNARY AMMONIUM COMPOUNDS Direct Binary Complex Method (0 to 5.00 mg/l as CTAB) Scope and Application: For cooling tower and pool/spa water. The estimated

More information

PRELIMINARY ACTIVITY FOR

PRELIMINARY ACTIVITY FOR PRELIMINARY ACTIVITY FOR Beer s Law Investigations Guided Inquiry Version Experiment 11 The primary objective of this Preliminary Activity is to determine the concentration of an unknown copper (II) sulfate

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01104 USEPA Reactor Digestion Method Method 10212 250 to 15,000 mg/l COD (UHR) TNTplus 823 Scope and application: For wastewater and process waters; digestion is required.

More information

Cuvette Test LCK 381 TOC Total organic carbon

Cuvette Test LCK 381 TOC Total organic carbon Cuvette Test Principle Total carbon () and total inorganic carbon () are converted to carbon dioxide (CO 2 ) by, respectively, oxidation and acidification. The CO 2 passes from the digestion cuvette through

More information

aa + bb cc + dd Equation 1

aa + bb cc + dd Equation 1 Experiment: The Determination of K eq for FeSCN 2+ Introduction For any reversible chemical reaction at equilibrium, the concentrations of all reactants and products are constant or stable. There is no

More information

Phosphate total LR TT mg/l P b) Phosphomolybdenum Blue

Phosphate total LR TT mg/l P b) Phosphomolybdenum Blue Phosphate total LR TT 0.07-3 mg/l P b) Phosphomolybdenum Blue 317 Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption

More information

HI REDUCING SUGARS ISM for wine analysis

HI REDUCING SUGARS ISM for wine analysis Instruction Manual HI 83746 REDUCING SUGARS ISM for wine analysis www.hannainst.com This Instrument is in Compliance with the CE Directives 1 Dear Customer, Thank you for choosing a Hanna product. This

More information

DHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual

DHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual DHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual WD1063 Rev B Revised 02/23/15 Wahl Instruments Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free: 800-421-2853 Phone: 828-658-3131

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law Vernier Spectrometer 1 The primary objective of this experiment is to determine the concentration of an unknown copper (II) sulfate solution. You

More information

Rapid RH Portable ph Meter. User's Manual

Rapid RH Portable ph Meter. User's Manual Rapid RH Portable ph Meter User's Manual This page intentionally left blank. 2 Contents 1. Introduction... 4 2. Included with the Rapid RH Portable ph Meter... 4 3. Understanding ph... 5 4. Calibrating

More information

DR/4000 PROCEDURE. HARDNESS, Total

DR/4000 PROCEDURE. HARDNESS, Total Method 8374 DR/4000 PROCEDURE Calcium and Magnesium; Chlorophosphonazo Colorimetric Method ULR (0 to 1,000 µg/l Ca & Mg as CaCO 3 ) Scope and Application: For ultra pure water. The estimated detection

More information

Thermo Scientific Orion AQUAfast AQ3070 Colorimetry Meter. User Guide

Thermo Scientific Orion AQUAfast AQ3070 Colorimetry Meter. User Guide Thermo Scientific Orion AQUAfast AQ3070 Colorimetry Meter User Guide ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787. AQUAfast, Cahn, ionplus, KNIpHE,

More information

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL HI96821 Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using this

More information

Quaternary Ammonium Compounds

Quaternary Ammonium Compounds , 8337 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB Powder Pillows Scope and Application: For cooling tower water and pool/spa water Test

More information

Oxygen Demand, Chemical

Oxygen Demand, Chemical Oxygen Demand, Chemical DOC316.53.01103 USEPA Reactor Digestion Method Method 10211 1 to 60 mg/l COD (ULR) TNTplus 820 Scope and application: For wastewater, process water, surface water, and cooling water.

More information

DIAGNOSTICS NO TOUCH FOREHEAD THERMOMETER

DIAGNOSTICS NO TOUCH FOREHEAD THERMOMETER DIAGNOSTICS NO TOUCH FOREHEAD THERMOMETER nu-beca NO TOUCH THERMOMETER : Thank you for purchasing the nu-beca No Touch Forehead Thermometer. Please read this instruction manual first so you can use this

More information

HI Hardness ISM

HI Hardness ISM Instruction Manual Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com

More information

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Cadion Method Method to 0.30 mg/l Cd TNTplus 852 Cadmium DOC316.53.01013 Cadion Method Method 10217 0.02 to 0.30 mg/l Cd TNTplus 852 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Chemistry 102 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

Mouse Glutathione S Transferase Alpha 1 (GSTa1)

Mouse Glutathione S Transferase Alpha 1 (GSTa1) Mouse Glutathione S Transferase Alpha 1 (GSTa1) For the quantitative determination of mouse (GSTa1) in serum, plasma and other biological fluids Cat. No. KT-16414 For Research Use Only. Not for use in

More information

Determining the Concentration of a Solution: Beer s Law

Determining the Concentration of a Solution: Beer s Law Determining the Concentration of a Solution: Beer s Law LabQuest 11 The primary objective of this experiment is to determine the concentration of an unknown nickel (II) sulfate solution. You will be using

More information

Eppendorf UVette. Instructions for use

Eppendorf UVette. Instructions for use Instructions for use Instructions Eppendorf see Fig. Tab. p. p. UVette for use Copyright 2014 Eppendorf AG, Hamburg, Germany. All rights reserved, including graphics and images. No part of this publication

More information

Phosphate total HR TT mg/l P b) Phosphomolybdenum Blue

Phosphate total HR TT mg/l P b) Phosphomolybdenum Blue Phosphate total HR TT 1.5-20 mg/l P b) Phosphomolybdenum Blue 318 Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption

More information

PAR Method Method to 2.0 mg/l Pb TNTplus 850

PAR Method Method to 2.0 mg/l Pb TNTplus 850 Lead DOC316.53.01056 PAR Method Method 10216 0.1 to 2.0 mg/l Pb TNTplus 850 Scope and application: For wastewater and process control. Test preparation Instrument-specific information Table 1 shows all

More information

DPD Test N Tube Method *

DPD Test N Tube Method * CHLORINE, FREE (0 to 5.00 mg/l) DPD Test N Tube Method * Method 10102 For water, wastewater, and seawater 1. Enter the stored program number for Test N Tube free chlorine (Cl 2 ). Press: PRGM The display

More information

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Chlorine Free, Indophenol, 10241 Indophenol 1 1 Patent pending. DOC316.53.01256 Method 10241 (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Scope and Application: For determining residual free chlorine levels

More information

Trihalomethane Formation Potential (THMFP)

Trihalomethane Formation Potential (THMFP) Trihalomethane Formation Potential (THMFP) DOC316.53.01147 THM Plus Method 1 Method 10224 Scope and application: To determine the potential of potable source waters that form trihalomethanes and other

More information

Chemical Reactions: The Copper Cycle

Chemical Reactions: The Copper Cycle 1 Chemical Reactions: The Copper Cycle ORGANIZATION Mode: pairs assigned by instructor Grading: lab notes, lab performance and post-lab report Safety: Goggles, closed-toe shoes, lab coat, long pants/skirts

More information

InfraRed Thermometer with Laser Pointer

InfraRed Thermometer with Laser Pointer User Manual 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com InfraRed Thermometer with Laser Pointer MODEL IR400 Introduction Congratulations

More information