Determination of Essential, Therapeutic. in Biological Materials; Applications of

Size: px
Start display at page:

Download "Determination of Essential, Therapeutic. in Biological Materials; Applications of"

Transcription

1 ESAC Copenhagen 15 th April 2008 Determination of Essential, Therapeutic and Toxic Elements and Their Compounds in Biological Materials; Applications of ICP-MS in the Clinical i l Laboratory Ed McCurdy, ICP-MS Specialist, Agilent Technologies Ltd

2 Scope Introduction to ICP-MS for clinical sample analysis Requirement for elemental analysis in clinical samples Routine trace element monitoring Analysis of toxic and harmful elements Elemental screening, poisoning, radionuclides Toxic organometallic compounds ICP-MS analysis for the measurement of organic compounds Identification and quantification of pesticide residues Quantification of Chemical Warfare Agent degradation products Forensic Applications of ICP-MS Laser Ablation ICP-MS Direct Analysis of Elemental Distribution in Tissues Page 2

3 What is ICP-MS? An inorganic (elemental) analysis technique ICP - Inductively Coupled Plasma MS - Mass Spectrometer high temperature electrical discharge, which decomposes, atomizes and ionizes samples forms ions, so compounds not measured directly quadrupole ( quad ) mass analyzer mass range from 5 to 260 amu (Li to U...) separates all elements in rapid sequential scan isotopic information available ions measured using dual mode detector ppt level LODs for most elements Calibration range up to 1000 s ppm Spectral interferences removed using collision/reaction cell Page 3

4 Agilent 7500cx ICP-MS System with Collision/Reaction Cell (CRC) Reaction Gas Inlet High temperature 27MHz plasma generator Multi-element interference removal by on-axis octopole reaction cell Fast simultaneous dual mode detector (9 orders dynamic range) Plasma Low flow sample introduction system Off-axis Lens Octopole High frequency hyperbolic quadrupole Page 4

5 Elements of Interest in Easily Extractable Fluids (Urine, Blood, Serum, Plasma) Highly toxic heavy metals As, Cd, Pb and Hg Potentially toxic elements Al, Sb, Ba, Be, Bi, Li, Ni, Sr and Tl Essential elements Cr, Co, Cu, Mg. Mn, Se, V and Zn Wide range of elements may be measured, including many, such as Be, As, Se, Hg, U, which are considered difficult by other techniques Page 5

6 Which Elements can be Measured Using ICP-MS? All elements in colour can be measured only those elements present in the plasma gas, <5amu, >260amu, and those which are not ionized, are inaccessible Page 6

7 Analytical Needs for Inorganic Measurements in Clinical Laboratories The technique must possess robust sample introduction To handle large sample numbers and differing matrices routinely monitored Measurement of many elements in the same fast acquisition Removal of interferences Screening applications Sample turnaround/productivity Reduced cost of analysis Quadrupole ICP-MS can meet all of these criteria Make measurements at trace levels (low detection limits) and at high concentrations in the same acquisition To reduce reruns, improve productivity and lower costs The ability to measure element species is also provided by ICP-MS Expands range of applications, and provides useful background research info Page 7

8 Collision/Reaction Cell to Remove Matrix-Based Spectral Interferences Apparent Cr5 52 Concentrat tion in Blank (ppb) Cr52 ClO/ClOH (mass 52) overlap in HCl matrix AC( ArC (mass 52) overlap in acetic acid matrix ArC and ClO overlaps in combined matrix 01% 0.1% 5% 5% HCl 1% 1% 200ppm 200ppm 500ppm Mixed HNO3 HNO3 H2SO4 AcOH Na Ca P Matrix H2 He NoGas Advantage of Collision/ i Reaction Cell ICP-MS Provides removal of spectral overlaps allows accurate trace element analysis in variable high-matrix samples Example shows the removal of ClO, ArC interferences in various matrix blanks He collision mode gives reliable removal of all polyatomic interferences regardless of the sample matrix composition Matrix Blank Page 8

9 Routine Analysis of Urine Sample Preparation and Calibration Direct analysis of urine samples following a 1/5 or 1/10 (v/v) dilution with deionized water and nitric acid No clogging of the nebulizer No particle deposition in the injector tube over 12 hours of analyses Calibration using Method of Standard Additions (MSA), so matrix-matched standards Standard addition calib is then converted to external calib and applied to subsequent samples 5 µg/l Tb internal standard added to all sample and calibration solutions Agilent Application Note: Rapid and reliable routine analysis of urine by Octopole Reaction Cell ICP- MS, EN, by Peter Heitland, Medical Laboratory Bremen, Germany Page 9

10 Measured and Certified Concentrations in Urine Reference Material Lyphochek Element Concentration (µg/l - ppb) Lyphochek, level 1 Lyphochek, level 2 Measured certified Measured (n=10, external) (n=10, external) certified Cr 1.7 ± ± ± ± 4.1 Co 6.6 ± ± ± ± 4.2 Cu 24 ± ± ± ± 10 Se 56 ± ± ± ± 37 As 65 ± 6 67 ± ± ± 33 Cd 8.4 ± ± ± ± 3.1 Sb 6.9 ± ± ± ± 7 Tl 96± 9.6 ± ± 9.7 ± ± ± 40 Pb 13.5± ± ± 5 69 ± 14 Data: Medical Laboratory Bremen, Germany Page 10

11 High Throughput Analysis of Blood Samples using 7500ce ICP-MS Biomonitoring of trace elements in human blood samples - an important tool for occupational and environmental health Goals of this study Determine a high number of trace metals in blood of 130 unexposed subjects Develop a rapid routine method for the multi-element analyses of blood using collision/reaction cell-icp MS Blood samples were collected in lithium heparin monovettes 500 ul of the sample was diluted with 100 ul 0.1% (v/v) Triton-X-100 solution and 500 ul of the internal standard solution This solution was made up to 5 ml with a 0.5% (v/v) NH 4 OH solution in a 10 ml polypropylene autosampler tube More than 100 samples can be prepared in less than 1hour by one person Biomonitoring of 37 trace elements in blood samples from inhabitants of northern Germany by ICP MS, Peter Heitland, Helmut D. Koster, Journal of Trace Elements in Medicine and Biology 20 (2006) Page 11

12 Analytical Figures of Merit for Blood Analysis Limits of quantification (LOQs), calculated in undiluted blood, range from ug/l for 238 Uto01ug/L 0.1 for 69 Ga Spike recoveries of 1 ug/l (10 ug/l for B, Mn and Sr; 200 ug/l for Cu and Rb) from single element calibration solutions are in the range % Data: Peter Heitland, Helmut D. Koster, Medical Laboratory Bremen, Germany Page 12

13 Effects of Sample Matrix on the Sample Introduction System Sample Cone Skimmer Cone Photos of the interface and sample introduction ti system after a 90-sample run (whole blood). Both the sampler and skimmer cones show only minor matrix deposits none at the cone tips The standard 2.5mm injector torch used was virtually depositfree. The blood deposits on spray chamber and the nebulizer block were removed using a sodium hypochlorite solution Images: R. Wahlen et al., LGC Limited, UK Page 13

14 Screening for Toxic and Harmful Elements Elemental Screening of 1:10 diluted Urine (with interference removal in He mode) Screening to identify poisons (1:10 diluted urine scan) Unique Unknown capability element of spiked ICP-MS into acquire urine sample a scan across the entire mass range in about 2 minutes, screening elements from 1000 s ppm to sub-ppb levels [1 ] S p e c tru m N o.1 [ s e c ]: S M P L.D # / T u n e # 1 [C P S ] [L in e a r] 1.0 E 5 C Na Rb Mg 5.0 E 4 Unknown element Li CaC a Cu Fe Zn Br As Mo Sr I Sb Cs Sn Ba Pb m / z-> > Page 14

15 Analysis of Toxic and Harmful Elements Screening to identify poisons (1:10 diluted urine scan) Confirmation (from isotopic template) of presence of Thallium (2ppb spike) Can be quantified (semiquant) by reference to known concentration element Note 210 Po would also be seen in this mass region of the screening acquisition [1 ] S p e c tru m N o.1 [ s e c ]:0 005SMPL.D# 3 S M P L / T/ u n e # 1 [C P S ] [L in e a r] 1.0E5 Alexander Litvinenko - poisoning 5.0E4 205 Tl Tl 203 Tl 210 Po 208 Pb Pb m/z- > Page 15

16 Measurement of Radionuclides Left: Calibration for radionuclides is easily achieved at sub ng/l (ppt) levels, even with standard sample introduction (U used for illustration) Radionuclides are typically easily ionized and the spectrum is free from overlaps and backgrounds, so LOD s in the pg/l (ppq) range are achieved Scan of 1ppt 237 Np standard, showing high sensitivity and low random background Page 16

17 Radionuclide Analysis Quantification and Isotope Ratio Measurement Very high sensitivity and low background for 10ppt U in 1:10 diluted urine ~1.2 million cps/ppb U Natural U spike, so 235 U (0.72% abundance) concentration was 72ppq 2.0E4 [1] Spectrum No.1 [ sec]:10urine.d / Tune #1 [CPS] [Linear] 238 U (99.27%) U ICP-MS also provides isotopic information, so U isotopic pattern (isotope ratio) can be used to identify source of contamination Natural = 0.72% 235 U Waste depleted = % 235 U Pile depleted ~0.6% 235 U Enriched >0.72% 235 U 1.0E4 235 U (0.72%) m/z-> Page 17

18 Toxic Elemental Forms or Species For many elements, the level of toxicity is highly dependent on the chemical form of the element, so separation (chromatography) is required GC Laser Ablation ICP-MS LC Optional Conventional Detector(s) e.g. ESI-MS CE % P Respo onse (CPS) m/z Time (min) Page 18

19 Example: LC-ICP-MS for As Speciation Toxic! Less-Toxic Non-Toxic? Many As species exist the inorganic As species are known to be toxic and most organic species are relatively l harmless to humans. The potential toxicity of some species, such as the huge variety of arsenosugars, has not yet been established. Courtesy Ute Kohlmeyer GALAB, Germany Page 19

20 Chromatogram of As Standard (1.0 µg/l each) A new column has been developed to provide routine separation of the 5 most common As species in urine: Column G (4.6 x 250 mm) g pp Guard Column G Mobile Phase (Basic): 2 mm phosphate buffer solution (PBS) ph 11.0 adjusted with NaOH 02mMEDTA mm, CH 3 COONa 3.0 mm NaNO 3 1% ethanol Agilent Application Note: Routine Analysis of Toxic Arsenic Species in Urine Using HPLC with ICP-MS, EN, by Tetsushi Sakai and Steven Wilbur, Agilent Technologies Page 20

21 Determination of Organo-As Species Using HPLC with ESI-MS & ICP-MS ICP-MS Agilent 7500 HPLC Agilent % 85 % ESI-MS Agilent 1100 Elemental specific detection Molecular specific detection Single HPLC System with controlled split to provide sample flow to ESI-MS and ICP-MS. Provides simultaneous measurement of As-containing compounds and As concentration (ICP-MS) and concentration/structural information on the organic part of the As compounds (ESI-MS) Courtesy Jörg Feldmann et al, Aberdeen Univ. Page Page 21 21

22 Separation and Identification of Organo- Arsenic Species ph 5.3 ICP-MS ESI-MS Intensity Peak k1 Peak 2 Peak 3 % [M+H] DMAE Peak 1 (B) ICP-MS m/z 75 y m/z m/z 139 m/z 277 m/z 259 m/z 181 m/z 361 m/z Retention time (s) Chromatograms (above) for ICP-MS measurement of As (mass 75) and ESI- MS of various indicator masses. Mass spectra (ESI-MS) for the 3 peaks show characteristic fragmentation patterns for the identified species DMAE, DMA and (maybe) DMAA % DMA [M+H] + Peak 2 (C ) Peak 3 (D) [M+H] [2M+H] + [2M+H] [2M-H 2 O] m/z % m/z DMAA? m/z Hansen et al. J Anal At. Spectrom, 18, 474 Page Page 22 22

23 Which Elements can be Measured Using ICP- MS? ICP-MS can also measure non-metals used in highly toxic compounds, such as pesticides and chemical warfare agents, provided the backgrounds can be controlled e.g. using GC-ICP-MS Page 23

24 Agilent GC-ICP-MS Interface GC-ICP-MS System used: ICP-MS: Agilent 7500 GC: Agilent 6890 Interface: Agilent G3158A Fully heated and insulated GC transfer line Modified torch with heated injector replaces standard demountable torch Silicosteel transfer line and injector liner for inertness GC effluent injected directly into base of plasma Very high transport efficiency, high plasma temperature (no water vapour/aerosol) and no solvent-based interferences High plasma temperature means elemental response is high, even for poorly ionized elements. Also, elemental response is independent of compound, so compound independent calibration (CIC) is possible Courtesy Raimund Wahlen, LGC Teddington Page 24

25 Pesticide Analysis by GC-ICP-MS Single ion chromatograms for C, P and S (right) and Cl, Br and I (below), extracted from multi-element GC-ICP-MS acquisition Low backgrounds (due to absence of solvent) and good ionization (due to high temperature of dry plasma), leads to excellent signal to background and low LOD Chlorine Ion (34.70 to 35.70): CICCAL3.D Bromine Ion (78.70 to 79.70): CICCAL3.D Iodine Ion ( to ): CICCAL3.D Carbon Ion (11.70 to 12.70): CICCAL3.D Ion (30.70 to 31.70): CICCAL3.D Phosphorus Ion (33.70 to 34.70): CICCAL3.D Sulphur Compound Conc (pg/ul - ppb) Calib Elements Elemental % Dichlobenil 610 Cl ,4,6-TBA 287 Br 72.5 Ethoprop 39 P, S 12.8, 26.4 DBOB 100 Br 35.1 Phorate 210 P, S 11.9, 36.9 PCNB 169 Cl 60.1 Terbufos 745 P, S 10.8, Diazinon 976 P, S 10.2, Malathion 107 P, S 9.37, Dursban 569 Cl, P, S 30.3, 8.82, 9.15 Table (right) shows components, concentrations and elemental l weight % in 1/10 diluted d CIC pesticide id mix. Page 25 Ioxynil (methyl ester) 50 I 66 TPP 158 P 50.3

26 CIC - Sulphur in Pesticide Mix Sulphur elemental response is independent of the compound Sulphur Response Area R 2 = Malathion Ethoprop Phorate Dursban Diazinoni Compound Terbufos Compound Concentration RT S conc S pg/ul (min) (ppb) response Ethoprop Phorate Terbufos Diazinon Malathion Dursban Concentration (ppb) Pesticide compound LoD s typically single ppb or sub-ppb Page 26

27 Chemical Warfare Agent (CWA) Regulation Chemical Weapons Convention of January 1993 Enforcement began April 1997 August 24 th, 2006; Meeting of 180 countries (representing 98% of World Population) which are members of the OPCW. Organization i for the Prohibition i of Chemical Weapons ~70,000 Metric Tons of Chemical Weapons Declared (24/8/2006) ~14,000 Metric Tons Destroyed (24/8/2006) 2006 Budget $96 Million "Determined for the sake of all mankind, to exclude completely the possibility of the use of chemical weapons..." From Doug Richardson, Univ Cincinnati Page 27

28 Chemical Warfare Agent Analysis by ICP-MS Ner ve Agent s G-Type V-Type H 3 C O CH P O C H 3 CH3 F H 3 C CH 3 Soman (GD) H 3 C O CH 3 P O C H CH F 3 Sarin (GB) O O H 3 C P O CH 2 CH 3 C P H 3 S S O CH 3 CH 3 N O O O P CN H 3 C P O N F VX TAbun (GA) Cycl osar in (GF) N Russian VX (RVX) 31 P Selective Detection All these agents contain a P atom, so ICP-MS can be used to identify and quantify the concentration of agent, based on the consistent (compound independent) response for 31 P. Nerve agents mostly decompose in the environment to MPA (via EMPA, IMPA, CMPA ) From Doug Richardson, Univ Cincinnati Page 28

29 CWA Analysis in Natural Samples by LC-ICP-MS Right: Standards Below: Unspiked and spiked Apple Juice Column: Hamilton PRP-X100 Anion Exchange P Respo onse (CPS) P 1 Resp ponse (CPS S) Time (min) Apple Juice + Spike (3ppm) Apple Juice Time (min) Elution Order 1. MPA 2. H 2 PO EPA 4. DMHP 5. PPA 6. EMPA 7. IMPA 8. DEHP 9. IPHEP 10. IBHMP From Doug Richardson, Univ Cincinnati Page 29

30 Forensic Applications of ICP-MS Glass Fragment Analysis Almost any solid fragment collected from a suspect individual or location may be suitable for analysis using laser ablation ICP-MS sample size as small as 50um diameter can be measured routinely Glass samples can be analysed using simple screening scan (qualitative or semi-quantitative) or calibrated against well-characterized reference glasses. NIST 600 series Trace Elements in Glass eg NIST 612 ~ 50ppm Images courtesy of New Wave Research

31 Trace Element Distribution Patterns (sum to 100%) Synthetic ti Glass and Unknown Samples Data courtesy of New Wave Research Page 31

32 Analysis of Bic Black Pen Inks 15 black Bic pen inks different sources Non-ablated Ink analyzed in triplicate 24 elements / 26 isotopes 105 comparisons taken as pairs 86/105 (82%) pairs were discriminated by Pb Of 19 remaining, 17 pairs were discriminated by Co Of 2 remaining, 1 pair was discriminated by Ba Both Zn and Cu have similar discriminating power to Co Only 1 pair (<1%) could not be discriminated Most useful discriminators: Pb, Co, Ba, Cu, Zn Ablated Ink Images courtesy of FBI Academy Page 32

33 Eight Bic Black Pen Inks Unknown #2b is 02 Bic (high W and low Mo) Unknown #8b is 08 Bic (High Pb and low W) Al/Cu Pb W/Mo Mo/Co Zn/Ba Bic Mean 02 Bic Mean 03 Bic Mean 04 Bic Mean 05 Bic Mean 06 Bic Mean 07 Bic Mean 08 Bic Mean Unk (#2b) Unk (#8b) Data courtesy of FBI Academy Page 33

34 Metal Imaging Mass Spectrometry (MIMS) Slice frozen tissue, e.g. brain 10 µm tissue sections LA-ICP-MS Image reconstruction Raster Laser Data acquisition iti Images courtesy of Dominic Hare, UTS Page 34

35 MIMS maps of various elements in brain section of Parkinson ss disease rat model system Intact side of brain EM Lesion side of brain (Parkinson-like) Intensity increases from blue green yellow red 31 P 56 Fe 57 Fe Images courtesy of Dominic Hare, UTS Page Page 35 35

36 Conclusions ICP-MS offers a unique combination of: Wide elemental coverage (almost all elements can be measured) Low limits of detection (typically 10 s ppq for easily ionized elements) Wide dynamic range (from sub-ppt to 1000 s ppm) Very rapid analysis (<3 minutes for full elemental screening, much less for consistent matrices and fewer elements) Freedom from spectral interferences (using collision/reaction cell) Tolerance of a wide range of matrices (due to high plasma temperature) Facility to link to chromatographic and direct solid sampling accessories, to further extend the range of applications Page 36

Agilent Technologies at TIAFT 2013

Agilent Technologies at TIAFT 2013 Agilent Technologies at TIAFT 2013 Analytical approaches for the measurement of trace metals in forensic samples Funchal, Madeira September 2-6 CRIMINALISTIC TOXICOLOGY DNA INVESTIGATION DOCUMENT EXAMINATION

More information

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,

More information

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry

More information

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Application note Semiconductor Authors Junichi Takahashi Agilent Technologies, Japan Introduction

More information

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction. Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA

More information

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson,

More information

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Application note Environmental Authors Steve Wilbur, Craig Jones Agilent Technologies,

More information

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Application Note Semiconductor Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Japan Introduction Hydrochloric

More information

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS

More information

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide

More information

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific, Omaha NE, USA Steve Wilbur,

More information

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS Technical Overview Abstract Inductively coupled plasma mass spectrometry

More information

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction

More information

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki

More information

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction

More information

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry APPLICATION NOTE Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry AN43227 Antonella Guzzonato 1, Shona McSheehy Ducos

More information

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Authors Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS Application Note Food safety Craig Jones and Jenny Nelson Agilent Technologies, USA Introduction In the U.S., marijuana remains

More information

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis Agilent ICP-MS Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis High Sample Matrix Tolerance Superior & Simple Interference Removal Ultra Wide Linear

More information

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS Tomoko Vincent 1, Simon Lofthouse 2, Daniel Kutscher 1 and Shona McSheehy Ducos

More information

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode Lead isotope analysis: Removal of Hg isobaric interference from using ICP-QQQ in MS/MS mode Application note Authors Geochemistry and isotope analysis Glenn Woods Agilent Technologies, LDA UK Ltd., Stockport,

More information

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS Accurate analysis of neptunium in a uranium matrix, using ICP-QQQ with MS/MS Application note Nuclear, environmental Authors Garry Duckworth, Springfields Fuels Ltd, K, Glenn Woods, Agilent Technologies,

More information

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords

More information

Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages

Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages National Environmental Monitoring Conference 2013 Austin TX Steve Wilbur and Amir Liba Agilent Technologies Let s Define

More information

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically

More information

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical

More information

ICP-OES Application Note Number 35

ICP-OES Application Note Number 35 ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable

More information

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity

More information

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Thermo Scientific icap TQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively

More information

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Author Application Environmental Tetsuo Kubota Agilent Technologies, Japan Introduction Regulations concerning activities such

More information

Samples compliant with the WATER

Samples compliant with the WATER ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions

More information

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio NEMC August 2018 Craig Jones Applications Specialist ICP-MS, ICP-MS/MS The Agilent Atomic Spectroscopy Lineup An Instrument for Any Application!

More information

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS APPLICATION NOTE 43326 Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS Authors Introduction Tomoko Vincent1, Simon Lofthouse2,

More information

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Application Note Consumer products Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ Effective removal of tungsten-based interferences on five Hg isotopes using MS/MS Authors

More information

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS) Application Note Environmental Authors Steve Wilbur Agilent Technologies,

More information

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode The ultratrace determination of iodine in aqueous samples using the 7700x ICP-MS with oxygen reaction mode Application note Nuclear Authors Kazumi Nakano, Yasuyuki Shikamori, Naoki Sugiyama and Shinichiro

More information

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction

More information

Analysis Repeatability of Trace and Major Elements in a Water Sample

Analysis Repeatability of Trace and Major Elements in a Water Sample Analysis Repeatability of Trace and Major Elements in a Water Sample Agnès COSNIER HORIBA Scientific Longjumeau, France Keywords: environment Elements: Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Li,

More information

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy

More information

speciation in water Page 1

speciation in water Page 1 Hyphenated LC-ICP-MS Arsenic speciation in water Page 1 Introduction Speciation Why and How? Page 2 Why Speciation Rather than Total Concentration ti of Metals? Almost all elements form species, with the

More information

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS Application note Semiconductor Author Naoki Sugiyama Agilent Technologies, Tokyo,

More information

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS) Basic and Application Inductively coupled plasma mass spectrometry (ICPMS) 2 What is ICP MS Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry which is capable of detecting

More information

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,

More information

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS Application note Metals Analysis & Production Authors Naoki Sugiyama and Mineko Omori Agilent Technologies,

More information

The New Agilent 7700 Series ICP-MS

The New Agilent 7700 Series ICP-MS The New Agilent 7700 Series ICP-MS the best ICP-MS just got better! Presenter: Uwe Noetzel and Jerome Darrouzes ICP-MS Product Specialist Life Sciences and Chemical Analysis Global Footprint Liquid Separation

More information

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences

More information

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin Elemental Scientific prepfast IC: Inline Autodilution Method to Eliminate Species Interconversion for LC-ICPMS Based Applications Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan

More information

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent

More information

DISCLAIMER: This method:

DISCLAIMER: This method: Inorganic arsenic determination in fresh mussels using water bath extraction and anion exchange chromatography-inductively coupled plasma mass spectrometry DISCLAIMER: This method: - has to be considered

More information

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology Agilent s New 8800 ICP-QQQ Transforming ICP-MS Technology 1 8800 ICP-QQQ Key Product Hardware High matrix introduction (HMI) technology Dual conical Extraction and Omega lens focuses ions across the mass

More information

Fergus Keenan Shona McSheehy Julian Wills

Fergus Keenan Shona McSheehy Julian Wills Novel QCell Technology for Inference Removal in ICP-MS - Combining Low Mass Filtration with Kinetic Energy Discrimination Fergus Keenan Shona McSheehy Julian Wills 1 The world leader in serving science

More information

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS Application Note Environmental China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS Fast, accurate quantitative analysis of 16 elements using an Agilent 7800 ICP-MS with

More information

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis Application note Authors Environmental, Food, Chemical Wim Proper 1, Ed McCurdy 2 and Junichi Takahashi 3 1 Eurofins Analytico,

More information

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment : ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment Max Planck Institute for Nuclear Physics, Heidelberg, Germany E-mail: mykola.stepaniuk@mpi-hd.mpg.de Stefano Nisi E-mail: stefano.nisi@lngs.infn.it

More information

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application Metals Analysis Authors Ed McCurdy, Glenn Woods and Don Potter Agilent

More information

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors. Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology Part 2 of a 3 part series on Environmental Analysis Application Environmental Authors

More information

Trace Analyses in Metal Matrices Using the ELAN DRC II

Trace Analyses in Metal Matrices Using the ELAN DRC II www.perkinelmer.com Trace Analyses in Metal Matrices Using the ELAN DRC II Introduction Analyses of matrices containing high levels of metals present a challenge for ICP-MS. First, the concentrations of

More information

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid Application Semiconductor Authors Abe G. Gutiérrez Elemental Scientific 2440 Cuming St Omaha, NE 68131 USA abe@icpms.com

More information

Sample Analysis Design PART II

Sample Analysis Design PART II Sample Analysis Design PART II Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis

More information

Analysis of domestic sludge using the Agilent 4200 MP-AES

Analysis of domestic sludge using the Agilent 4200 MP-AES Analysis of domestic sludge using the Agilent 4200 MP-AES Application note Environmental Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction Managing the treatment and disposal

More information

Determination of iopromide in environmental waters by ion chromatography-icp-ms

Determination of iopromide in environmental waters by ion chromatography-icp-ms Determination of iopromide in environmental waters by ion chromatography-icp-ms Application note Environmental Authors Armando Durazo, Tarun Anumol, and Shane A. Snyder Department of Chemical & Environmental

More information

Multi-residue analysis of pesticides by GC-HRMS

Multi-residue analysis of pesticides by GC-HRMS An Executive Summary Multi-residue analysis of pesticides by GC-HRMS Dr. Hans Mol is senior scientist at RIKILT- Wageningen UR Introduction Regulatory authorities throughout the world set and enforce strict

More information

Perfluorinated compounds detection using non-target analysis of RP-HPLC-ICPMS/MS

Perfluorinated compounds detection using non-target analysis of RP-HPLC-ICPMS/MS Perfluorinated compounds detection using non-target analysis of RP-HPLC-ICPMS/MS Nor Laili Azua Jamari, Jan Frederik Dohmann, Andrea Raab, Eva M. Krupp and Jörg Feldmann Trace Element Speciation Laboratory

More information

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Petrochemical Authors Ed McCurdy & Don Potter Agilent Technologies Ltd. Lakeside Cheadle Royal

More information

Arsenic Speciation in Chicken. Mary Carson, Ph.D. FDA/CVM/OR/ Division of Residue Chemistry

Arsenic Speciation in Chicken. Mary Carson, Ph.D. FDA/CVM/OR/ Division of Residue Chemistry Arsenic Speciation in Chicken Mary Carson, Ph.D. FDA/CVM/R/ Division of Residue Chemistry Elements of Clinical or Public Health Significance Elemental Speciation The chemical species are specific forms

More information

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs Application note Authors Food safety Kazuhiro Sakai 1, Junichi

More information

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Application Note Clinical Research Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS Authors Yanan Yang 1, Victor Mandragon 2, and Peter Stone 1 1

More information

Speciation Measurement using Hyphenated ICP-MS

Speciation Measurement using Hyphenated ICP-MS Speciation Measurement using Hyphenated ICP-MS Ed McCurdy European ICP-MS Specialist Agilent Technologies What is Speciation? Elemental Speciation - the analyses that lead to determining the distribution

More information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry Application Note Food Authors Jianqiu Mi, Zhengxiang Zhang, Zhixu Zhang,

More information

Sample Analysis Design PART III

Sample Analysis Design PART III Sample Analysis Design PART III Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis

More information

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction

More information

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Application note Semiconductor Author Katsuo Mizobuchi Masakazu

More information

ICP-Mass Spectrometer

ICP-Mass Spectrometer ICP-Mass Spectrometer New Mass Spectrometers The main issue: sequential vs. simultaneous Scanning, peak hopping are sequential Like viewing a photo through a peephole One pixel at a time Other parts of

More information

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4 ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4 Tatiana Trejos, M.Sc Florida International University Department of Chemistry and Biochemistry International Forensic Research Institute Outline

More information

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System Application Note Foods Author Glenn Woods ICP-MS Specialist Agilent Technologies UK Ltd. Lakeside Business

More information

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software Alain Desprez Agilent Technologies 1 The Total Solution Choice of Mass Spectrometers 7800 quadrupole ICP-MS cost-effective, 3

More information

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB. Quantitative analysis of small molecules in biological samples 100 Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB % 0 300 500 700 900 1100 1300 1500 1700 m/z Class Overview Introduction

More information

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+ INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ Applied'Analy+cal'and'Inorganic'Chemistry'Program' Department'of'Chemistry,'Faculty'of'Science' Mahidol'University' What+is+Plasma?+ Ar Ar + e - Plasma is an ionized

More information

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED Application Note: 40852 Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED - Interference-free analysis of 29 elements in environmental

More information

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals Validating performance of an Agilent 7800 ICP MS for USP / & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals Authors Lindsey Whitecotton, Ed McCurdy, Craig Jones and Amir Liba

More information

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS ICP-MS Environmental Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS Application Note Steve Wilbur Analysis of arsenic and selenium in seawater at trace

More information

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Application note Petrochemical Author Lieve Balcaen, Frank Vanhaecke 1, Glenn Woods 2, Martín

More information

Low Level Detection of Priority Inorganic Pollutants in Water. Anthony Neumann, MSc Darren Crichton, BSc

Low Level Detection of Priority Inorganic Pollutants in Water. Anthony Neumann, MSc Darren Crichton, BSc Low Level Detection of Priority Inorganic Pollutants in Water Anthony Neumann, MSc Darren Crichton, BSc Priority List of Pollutants Consists of 16 metals and/or transitional elements. Enter the human food

More information

Florida State University Libraries

Florida State University Libraries Florida State University Libraries Electronic Theses, Treatises and Dissertations The Graduate School 2013 Accurate and Precise Determination of Low Concentration Iron, Arsenic, Selenium, Cadmium, and

More information

Agilent 8900 Triple Quadrupole ICP-MS

Agilent 8900 Triple Quadrupole ICP-MS Agilent 8900 Triple Quadrupole ICP-MS Technical Overview Introduction Agilent is the worldwide market leader in quadrupole ICP-MS, and the only supplier of triple quadrupole ICP-MS (ICP-QQQ). Introduced

More information

OES - Optical Emission Spectrometer 2000

OES - Optical Emission Spectrometer 2000 OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the

More information

ICP/MS Multi-Element Standards

ICP/MS Multi-Element Standards Standards Ultra Pure Matrix Special Packaging Traceability to National Reference Materials AccuStandard s ICP/MS Standards are formulated to meet the needs of this very special instrument. As matrix effect

More information

Autenticity control, provenance testing and fraud detection using mass spectrometry

Autenticity control, provenance testing and fraud detection using mass spectrometry Autenticity control, provenance testing and fraud detection using mass spectrometry Abdal-Azim Al-Terkawi Ersan Özelci Melissa-Jane Monks Narendra Lagumaddepalli Venkatareddy Seyed Mohsen Jebreiil Khadem,

More information

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Elemental analysis of river sediment using the Agilent 4200 MP-AES Elemental analysis of river sediment using the Agilent 4200 MP-AES Application note Environmental: Soils, sludges & sediments Authors Neli Drvodelic Agilent Technologies, Melbourne, Australia Introduction

More information

Application Note. Abstract. Authors. Environmental

Application Note. Abstract. Authors. Environmental Determination of Acidic Herbicides using an Agilent 6460 Triple Quadrupole LC/MS Equipped with Agilent Jet Stream Technology and Direct Aqueous Injection, for Potable and Environmental Samples Application

More information

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity Janet Hammond Waters Corporation, Wilmslow, UK APPLICATION BENEFITS High

More information

Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements

Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements Practicum Spectroscopy Fall 2010 LIM Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements Jorge Ferreiro, study degree Chemistry, 5 th semester,

More information

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS Application note Food safety Authors Bing Yue China National Center For Food Safety Risk Assessment, Beijing, China Juane

More information

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System DiscovIR-GPC DiscovIR-GC Deposition and Detection System Application Note 41 Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System ABSTRACT The DiscovIR-GC

More information

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Prodigy ICP Application Note: 1035 U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Introduction This Application Note describes the capability of the Teledyne Leeman Labs Prodigy High

More information

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors Determination of major, minor and trace elements in rice flour using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Application note Food testing Authors John Cauduro Agilent Technologies,

More information

Supporting Information

Supporting Information 1 Supporting Information 2 3 Discovery and implications of C 2 and C 3 perfluoroalkyl sulfonates in aqueous film forming foams (AFFF) and groundwater 4 Krista A. Barzen-Hanson a and Jennifer A. Field b*

More information

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application Environmental Author Neal Julien Midwest Research Institute

More information

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full

More information