Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+
|
|
- May Morgan
- 5 years ago
- Views:
Transcription
1 INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ Applied'Analy+cal'and'Inorganic'Chemistry'Program' Department'of'Chemistry,'Faculty'of'Science' Mahidol'University' What+is+Plasma?+ Ar Ar + e - Plasma is an ionized gas that is macroscopically neutral. Source: FuncDon+of+Plasma+ Courtesy from Agilent desolva+on' 'vaporiza+on' 'atomiza+on' 'ioniza+on'
2 ICPMOES+or+ICPMMS+ Source: Today s Chemist at Work Volume 8, No. 10, 42-44, by Robert Thomas ionization excitation M M * -hν excitation M + M +* -hν Component+of+ICP+Spectrometer+ ICP-OES ICP-MS Aerosol+generaDon+and+sample+transport+in+plasma+ spectrometry+! Key'components'! Nebulizer'! Spray'Chamber'! Plasma'! Key'basic'parameters'! Drop'size'distribu+on'! Mass'transport'of'aerosol'! Loss'processes'occuring'in' spray'chamber'! Solvent/plasma' interac+on' Nebulizers+ Pneumatic nebulizer Cross flow nebulizer Concentric nebulizer sample solution gas aperture V-groove V-groove nebulizer
3 Ultrasonic+Nebulizer+ A piezoelectric crystal vibrates at ultrasonic frequencies (50 khz to 4 MHz), and then causes sample solution to break up into small aerosols PneumaDc+VS+Ultrasonic+Nebulizers:+ aerosol+size+distribudons+ Source: Inductively Coupled Plasma Spectrometry and Its Applications, edited by Steve J. Hill, CRC Spray+Chamber+ " Result'! Removal'of'larger'droplets'G'd'>'10'µm' " Goal'! Avoid'overloading'of'ICP'! Guarantee'efficient'atomiza+on'&'ioniza+on' " Phenomena'occurring'! Removal'of'larger'droplets'by'impac+on'on'chamber'walls'! Solvent'evapora+on'! Coagula+on'of'smaller'droplets' " Analyte'introduc+on'efficiency'! Reduced'to'1G2%'with'tradi+onal'PN' ScoRMType+DoubleMPass+ Spray+Chamber+ cooling water in to ICP (1 2 %) water jacket cooling water out to waste
4 Cyclonic+Spray+Chamber+ Source: to ICP side view top view ICP+Torch+Assembly+ Velmer Fassel ( ) Source: icp.tecmec.co.uk/whatisicp.html Source: Plasma+GeneraDon+ Source: icp.tecmec.co.uk/whatisicp.html AUXILIARY GAS NEBULIZER GAS Plasma+IgniDon+
5 Why+Argon+Plasma?+ " High'ioniza+on'energy'(15.76'eV)' " Chemically'inert'(cheapest'noble'gas)'! Simple'spectrum'! Capable'to'excite'and'ionize'most'of'the'elements' of'the'periodic'table'! No'stable'compounds'are'formed'between'argon' and'the'analytes' Plasma+Fire+Ball+ Ar Ar + e - Source: Component+of+ICP+Spectrometer+ ICP-OES Axial+or+Radial+ Axial Radial Source:
6 Axial+Plasma+CharacterisDcs+ Optical light path Recombination zone (severe matrix effect) Emission zone (no matrix effect) Courtesy from Spectro Analytical Instruments Excitation zone (slight matrix effect) OPI+operaDng+principle+ Adjustable argon jet Optical light path to spectrometer Recombination matrix effects Plasma Courtesy from Spectro Analytical Instruments Air+shear+gas+to+displace+the+cool+plume+ Source: DetecDon+System+ Sequential instrument Simultaneous instrument Source:
7 Component+of+ICP+Spectrometer+ ICP-MS Sample+IonizaDon+in+the+Plasma+ IonizaDon+in+ICP+ analyte ionization electron impact: M + e - M + + 2e - Penning ionization: Ar* + M M + + Ar + e - IonizaDon+Efficiency+in+ICP+ Degree of ionization (α) Ionization energy IE (ev)
8 IonizaDon+Efficiency+of+Elements+ InducDvely+Coupled+Plasma+ Ion+Source+for+Mass+Spectrometry+ PlasmaMSpectrometer+Interface+ " Sampling'cone' " Skimmer'cone'! Allows'introduc+on'of'ions'into'the'vacuum'chamber'! Material:''Nickel,'Pla+num'
9 " As'the'ion'beam'passes'through'the'sampler'cone'into'the' skimmer'cone,'expansion'will'take'place,'but'its'composi+on' will'be'maintained.' Natural+tendency+for+the+ions+to+repel+each+other+ " With'the'rapid'drop' in'pressure'in'the' lens'chamber,' electrons'diffuse' out'of'the'ion'beam.'
10 " Therefore,'there'is'a'clear'need'to'electrosta+cally'steer'the' ions'of'interest'back'into'the'center'of'the'ion'beam'by'placing' voltages'on'one'or'more'ion'lens'components' Lens+System+
11 quadrupole filter rf DC source Basic+Principle+of+ Quadrupole+Mass+Filter+ high mass filter low mass filter + bandpass mass filter = Quadrupole+Mass+Analyzer+ Ion+DetecDon:++ ConDnuous+Dynode+Electron+MulDplier+ multiplication effect multiplication factor: pulse counting mode vs. analog mode
12 Ion+DetecDon:++ Discrete+Dynode+Electron+MulDplier+ Dynodes Pulse multiplication factor: pulse counting mode vs. analog mode + From Mass Analyzer SAMPLING+TECHNIQUES+IN+ INDUCTIVELY+COUPLED+PLASMA+ SPECTROMETRY+ Hydride+GeneraDon+ Gaseous'sample'introduc+on'! Produc+on'of'hydride'of'target' element' Source: Hydrides'separated'from'solu+on'in' gas/liquid'separator'! Hydrides'transported'into'ICP'by' means'of'ar'carrier'gas' Hydride+GeneraDon+ " Advantages'! 100%'analyte'introduc+on'efficiency'! Separa+on'of'analyte'elements'from'matrix' # Reduc+on'of'spectral'&'nonGspectral'interferences' " Disadvantages'! Limited'applica+on'range' # As,'Bi,'Ge,'Pb,'Sb,'Se,'Sn'and'Te'! Specific'sample'preGtreatment'required' # ph'has'to'be'adjusted' # Correct'oxida+on'state'of'target'elements'required'
13 " combining'normal' pneuma+c'ac+on,' within'a'special'spray' chamber,'to'transport' both'normal'elements' and'hydride'elements' simultaneously,' without'adversely' affec+ng'either'' Source: Concomitant Metal Analyzers (CMA) Sampling+Accessories:+Laser+AblaDon+ Source: (ICP-MS Beginner s Guide) Electrothermal+Vaporizer+ CMA+Nebulizer+ by+jobinmyvon+horiba,+france+ ETV-ICP-MS Source: (ICP-MS Beginner s Guide)
14 ICPMOES+Measurement:+Procedure+ Development+ Line'selec+on' Tuning' Matrix'effects' Method'valida+on' Source: By Paul Gaines, Ph.D. Edited by Brian Brolin Precision Sensitivity Line+SelecDon+ Spectral Interferences Line+SelecDon:+SensiDvity+ " Make'sure'that'the'line'is'not'too'sensi+ve'(i.e.'G' going'outside'the'linear'working'range).'dilu+ons' are'easy'' " Have'more'than'one'line'that'meets'your'lower'limit' of'detec+on'and'quan+ta+ve'measurement' requirements' Line+SelecDon:+Precision+ " Try'to'keep'the'analyte'concentra+on'well'within'the'linear'working'range' " Avoid'lines'requiring'spectral'correc+on' " Avoid'lines'that'are'in'spectrally'complex'regions' " Increase'the'integra+on'+me'' " Use'an'all'glass'introduc+on'system'including'a'glass'concentric'nebulizer' " Be'aware'of'high'salt'containing'solu+ons' " If'your'samples'and'standards'have'the'same'background'then'skip'background'correc+on' " If'the'matrices'of'the'samples'and'standards'can'be'matched'then'eliminate'the'peristal+c' pump'and'go''free'flow''' " Allow'the'instrument'to'warm'up'for'at'least'an'hour'and'work'in'a'controlled'temperature' environment'' " Clean'your'introduc+on'system'and'replace'the'tubing'when'necessary'
15 Line+SelecDon:+ Spectral+Interferences+ " direct'spectral'overlap' " wing'overlap' " near'neighbors' Spectrum+of+a+high+concentraDon+of+Fe+(red+line)+showing+a+direct+ spectral+overlap+upon+the+b line+and+a+wing+overlap+on+a+b Spectrum+of+a+high+Ca+containing+matrix+(red+line)+causing+a+sloping+ background+for+the+cu or+ge lines+ Background+correc-on+is+difficult!+ A+wing+overlap+interference+of+Fe+(red+line)+upon+the++ Ba nm+line+
16 Background+Enhancement+from+ Ca(II) and nm+Broadening++ A+nearby+Fe+line+interferes+with+the+background+correcDon+upon+the+Al nm+line.+ Nega-ve+values+of+background+corrected+intensity!+ Plasma+Diagnosis:+OpDmizaDon+ IonicMtoMAtomic+Line+Intensity+RaDo++ " Mg(II,' 'nm)/Mg(I,' 'nm)'! closeness'of'the'excita+on'energies'of'the'atomic' and'ionic'lines'! independent'of'the'detector'condi+ons'! wavelength'response'correc+on'is'not'necessary'! ra+o'above'8'indicates'robust'condi+on'
17 Effect+of+Plasma+Power+ Effect+of+Carrier+Gas+Flow+Rate+ Matrix+Effects+ " Slight'differences'in'the'matrix'can'cause'a'considerable' systema+c'error'having'a'definite'bias' Standard Addition Internal Standardization Solving Matrix Effect Problems Matrix Removal Internal+StandardizaDon:+ SelecDon+Criteria+ " Is'the'internal'standard'(IS)'element'compa+ble'with'your' matrix?''' " Are'there'any'possible'spectral'interferences'upon'the'IS'line?' " Is'the'concentra+on'of'the'IS'sufficient'to'give'a'good'signal'to' noise'ra+o?' " Can'your'sample'possibly'contain'the'IS'element'as'a'natural' component?' " Is'the'IS'clean?'' " Is'your'method'of'addi+on'of'the'IS'very'precise?'' " Do'you'always'use'the'same'lot'of'IS'for'the'standards'and' samples?' " If'your'plasma'temperature'were'to'go'up'or'down,'is'the'IS' likely'to'follow'the'same'paiern'of'intensity'change'as'the' analyte?'
18 Technique+of+Standard+AddiDons+ tedious but more reliable approach fastest approach for unknown matrices " At'least'two'spectral'lines'should'be'used'and'the' spectral'region'carefully'scanned'and'studied' " Aiempt'to'make'the'addi+ons'approximately'2x'and' 3x,'where'x'is'the'analyte'concentra+on'' ICPMMS+Measurement:+Procedure+ Development+ Tuning' Interferences' Matrix'Effects' Source: By Paul Gaines, Ph.D. Edited by Brian Brolin Why+OpDmize+(Tune)?+ " Op+mize'sensi+vity'! Maximize'signal'! Minimize'noise' " Verify'correct'mass'calibra+on' " Verify'correct'ion'ra+o'response' " Minimize'interferences'! Oxides'! Doubly'charged'ions'! Argides' Ion+Signal,+Plasma+Power,+ Nebulizer+Flow+Rate,+Sampling+Distance+
19 Ion+Signal,+Plasma+Power,+ Nebulizer+Flow+Rate,+Sampling+Distance+ Interferences+in+ICPMMS+ Isobaric Interferences Polyatomic Interferences Doubly Charged Ion Interferences Isobaric+Interferences+ Mass Element (% abundance) Interfering Isotope (% abundance) 46 Ti (7.95) Ca (0.0032) 48 Ti (73.45) Ca (0.179) 54 Fe (5.9) Cr (2.38) 58 Ni (67.76) Fe (0.33) 64 Zn (48.89) Ni (1.16) 114 Cd (28.81) Sn (0.65) 138 Ba (71.66) Ce (0.25) 152 Sm (26.63) Gd (0.2) Correction is possible. I (114Cd) = I (114) - (0.65/24.01)*I (118Sn) Polyatomic+(Molecular)+Interferences+ Source Interfering species Analyte isotope Support gas 40 Ar 80 2 Se Reagent acids 35 Cl 16 O 40 Ar 35 Cl 51 V 75 As Water, solvent 40 Ar 16 O 56 Fe Matrix elements 40 Ar 35 Cl 44 Ca 16 O 75 As 60 Ni
20 Even+high+purity+ water+can+lead+to+ spectral+ interferences!+ Spectral Interferences Due to The Presence of Cl Spectral Interferences Due to The Presence of C Spectral Interferences Due to The Presence of S
21 Spectral Interferences Due to The Presence of P Doubly+Charged+Ion+Interference+ Mass Interfering species Analyte isotope Ba 2+ Ga Nd 2+, 150 Sm 2+ As Pb 2+ Rh+ Matrix+Effects+ " Shil'of'ioniza+on'equilibria'in'plasma' " Ambipolar'diffusion'in'plasma' " Space'charge'repulsion'in'ion'op+cal'system,' ion'beam'dispersion,'space'charge'induced' scaiering' Shih+in+IonizaDon+Equilibrium+ in+presence+of+eies+ Ionization equilibrium in ICP M M + + e Ionization constant K M ni. n = n Ionization efficiency i a ni α = n + n a e KM = K + n M e constant constant With addition of easily ionizable elements (EIEs), n e increases with subsequently leading to decrease of ionization efficiency suppression most pronounced for high IP analyte elements
22 Ambipolar+Diffusion+Effect+ ICP M + ions electrons diffusion of electrons & M + ions mobility of electrons >> mobility of M + further diffusion of electrons causes M + ions to diffuse leading to lower M + density in center Space+Charge+Effects+ During+Ion+ExtracDon+ interface beam becomes enriched in postively charged particles extraction lens Matrix+Effects:+on+Lithium+ Signal B Zn Cd Au high ev Na Rb Cs U low ev Mass of matrix element Space charge effect Ionization effect Solvent+Effect In+ signal intensity (arbitrary units) 0.14 mol/l HNO % EtOH nebulizer gas flow rate (l/min) At lower plasma temperature, ionization requires longer residence time in ICP At higher temperature, ionization after shorter residence time in ICP
23 Intensity 1% HNO 3 5% HNO 3 20% HNO 3 Concentration of analyte QuanDtaDve+Analysis+Measurement+ Techniques+ External'Calibra+on' Standard'Addi+on' Isotope'Dilu+on' External+CalibraDon+ Intensity (cps) Concentration (ppb) External+CalibraDon+with+the+Use+of+ Internal+Standard+ " Avoid'M 2+ 'interferences' " Avoid'MO'and'other'molecular'interferences' " Any'naturally'occurring'internal'standard'element'in'the' sample'must'be'insignificant'in'comparison'to'the'amount' added' " Use'internal'standard'elements'as'close'as'possible'to'the' masses'of'the'analyte'elements' " Make'sure'the'matrix'doesn't'react'with'the'internal'standard' to'alter'(lower)'it's'concentra+on' " Common'internal'standard'elements'are'Li,'Be,'Sc,'Ga,'Ge,'Y,' Rh,'In,'Cs,'Pr,'Tb,'Ho,'Re,'Bi,'and'Th'
24 Method+of+Standard+AddiDon+ Intensity (cps) Concentration (ppb) Isotope+DiluDon+Technique+ Intensity Sample m/z Known quantities: - Sample amount - Isotope ratio Isotope diluted sample Intensity m/z Measure quantity: Isotope ratio of the isotope diluted sample Intensity Isotope spike m/z Known quantities: - Spike amount - Isotope ratio - Spike concentration " Concentra+on'of'Hg'in'sample'is'calculated'from'the' rela+ve'abundance'of' 201 Hg'and' 202 Hg'in'both'the'sample' and'spike' [Hg] = (A sp - RB sp ) x W sp (RB s A s ) x W s! '[Hg] '=''''concentra+on'in'ppm'! 'R' '=''''the' 201 Hg'to' 202 Hg'ra+o'! 'A sp '=''''the'spike' 201 Hg'abundance''! 'B sp '=''''the'spike' 202 Hg'abundance'! 'W sp '=''''the'spike'weight'in'µg'! 'A s '=''''the'natural' 201 Hg'abundance'! 'B s '=''''the'natural' 202 Hg'abundance'! 'W s '=''''the'sample'weight'in'g' Isotope+DiluDon+Technique+ " Advantages'! Excellent'correc+on'for'matrix'effects'! Aler'isotopic'equilibra+on:'analyte'losses'do'not'affect'result' # WellGsuited'in'case'of'matrix/trace'separa+on'! Isotope'ra+os'barely'affected'by'changes'in'sensi+vity'! Most'reliable'calibra+on'method' ' " Disadvantages'! Not'suited'for'monoGisotopic'elements' # Be,'Na,'Al,'P,'K,'Sc,'Mn,'Co,'As,'Y,'Nb,'Rh,'I,'Cs,'Pr,'Ho,'Tm,'Au,'Bi,'Th' # Some+mes':'use'of'longGlived'radionuclide'(e.g.,' 129 I)'! High'purchase'price'of'spikes' # '500G1000'US'$'/'100'mg' # Isotopic'enrichment'&'nuclide'(natural'isotopic'abundance)'! Analyte'concentra+on'has'to'be'approximately'known'! Less'suited'for'mul+Gelement'analysis'
25 Analyte+element:++InformaDon+on+the+web+ Source: Analyte+element:++InformaDon+on+the+web+ Source: Drawbacks+of+ICP+Emission+Spectrometry+ Typical+ICPMMS+Full+Mass+Spectrum+
26 ComparaDve+Spectral+Complexity:+ ICPMMS+vs.+ICPMOES+ Element Emission Lines Isotopes (natural) Alkali metals Lithium 30 2 Cesium Alkali earths Magnesium Calcium Transition metals Chromium Iron Cerium Volume 8, No. 10, 42-44, Robert Thomas AAS,+AES,+or+AMS+ SelecDon+Process+! Define'the'requirement''! Establish'criteria''! Define'the'task'in'detail''! Compare'techniques''
27 Comparison+of+AnalyDcal+Techniques+ Detection Limits Analytical Working Range Sample Throughput Purchasing Costs Interferences Skill Levels and Methodology DetecDon+Limit+Ranges+ AnalyDcal+Working+Ranges+ Sample+Throughput+ Flame AAS 3-10 s per element per sample GFAAS 2-3 min per element per sample ICP-OES elements in a few minutes ICP-MS elements in a few minutes
28 Interferences+in+Atomic+Spectroscopy+ Technique Type of Interference Method of Compensation Flame AAS GFAAS ICP-OES ICP-MS Ionization Chemical Physical Physical and chemical Molecular absorption Spectral Spectral Matrix Spectral Matrix Ionization buffers Releasing agent or nitrous oxide-acetylene flame Dilution, matrix matching, or method of additions Standard Temperature Platform Furnace (STPF) conditions, standard additions Zeeman or continuum source background correction Zeeman background correction Background correction or the use of alternate analytical lines Internal standardization Interelement correction, use of alternate masses, higher resolution systems or reaction/collision cell technology Internal standardization Skill+Levels+and+Methodology+ Three'criteria:' # Ease'of'use' # Operator'skill'level'' # Availability'of'applica+on'method' ApplicaDon+Examples+ " Copper'in'an'Electropla+ng'Bath'' " Lead'and'Cadmium'in'Blood'' " 22'Elements'in'Wastewater'' " 18'Elements'in'HighG'Purity'Hydrogen'Peroxide'' Summary:+ Choosing+AnalyDcal+Technique+ " No'one'technique'suitable'for'all'spectroscopic' applica+ons' " Each'has'strengths'and'weaknesses'that'must'be' evaluated'to'make'the'right'choice' " As'an'analy+cal'chemist,'you'are'commiied'to'supplying' data'of'the'highest'quality' " Your'success'could'depend'on'selec+ng'the'right'trace' element'technique'to'achieve'this'goal'
Lecture 7: Atomic Spectroscopy
Lecture 7: Atomic Spectroscopy 1 Atomic spectroscopy The wavelengths of absorbance and emission from atoms in the gas phase are characteristic of atomic orbitals. 2 In the lowest energy transition, the
More informationPartial Energy Level Diagrams
Partial Energy Level Diagrams 460 nm 323 nm 610 nm 330 nm 819 nm 404 nm 694 nm 671 nm 589 / 590 nm 767 / 769 nm Lithium Sodium Potassium Gas Mixtures Maximum Temperatures, C Air-Coal Gas 1825 Air-Propane
More informationSample Analysis Design PART III
Sample Analysis Design PART III Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis
More informationApplications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry
Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS
More informationSources of Errors in Trace Element and Speciation Analysis
Sources of Errors in Trace Element and Speciation Analysis Zoltan Mester, National Research Council of Canada, Institute for National Measurement Standards Outline Definitions Sources of errors in the
More informationFundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company
Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company Why ICP Mass Spectrometry? Ultra-trace multi-element analytical
More informationSample Analysis Design PART II
Sample Analysis Design PART II Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis
More informationhigh temp ( K) Chapter 20: Atomic Spectroscopy
high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment
More informationOES - Optical Emission Spectrometer 2000
OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the
More informationCharacterization of Catalysts and Surfaces. Elemental Analysis (ICP, AAS etc.) Fall Semester 2016 Bodo Hattendorf HCI G105
Outline Characterization of Catalysts and Surfaces Elemental Analysis (ICP, AAS etc.) Fall Semester 2016 Bodo Hattendorf HCI G105 bodo@inorg.chem.ethz.ch Instrumental Methods for Determination of the Elements
More informationThermo Scientific icap RQ ICP-MS: Typical limits of detection
TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry
More informationSample Analysis Design Polyatomic Interferences
Sample Analysis Design Polyatomic Interferences More serious than isobaric interferences Result from possible, short-lived combination of atomic species in the plasma or during ion transfer Common recombinants
More informationChapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques
Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are
More informationAnalysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS
TECHNICAL NOTE 4322 Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS Keywords Argon gas dilution, AGD, High matrix samples, Seawater Goal To critically
More informationDetermination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS
APPLICATION NOTE ICP - Mass Spectrometry Author Kenneth Ong PerkinElmer, Inc. Singapore Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS Introduction The control of impurity
More informationTrace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell
Trace elemental analysis of distilled alcoholic beverages using the Agilent 77x ICP-MS with octopole collision/ reaction cell Application note Food testing Author Glenn Woods Agilent Technologies Cheadle
More informationAnalysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ
Application Note Semiconductor Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Japan Introduction Hydrochloric
More informationSamples compliant with the WATER
ENVIRONMENTAL analysis A routine method for the Quantitative Measurement of trace metals in WATER Samples compliant with the WATER Quality Standard EN ISO 17294 on the Agilent 7000 Series ICP-MS. Solutions
More informationCHAPTER 4: ANALYTICAL INSTRUMENTATION
CHAPTER 4: ANALYTICAL INSTRUMENTATION 4.1 INTRODUCTION In this section, a review of the analytical instrumentation used during sample preparation and analysis is presented which includes an overview of
More informationDetermination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode
Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode Application note Semiconductor Authors Michiko Yamanaka, Kazuo Yamanaka and Naoki
More informationDirect Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application
Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application Semiconductor Author Junichi Takahashi Koichi Yono Agilent Technologies, Inc. 9-1, Takakura-Cho, Hachioji-Shi,
More informationSpectrometric Methods of Analysis. OCN 633 Fall 2013
Spectrometric Methods of Analysis OCN 633 Fall 2013 Plasma Emission and Plasma Mass Spectroscopy Two fields of elemental analysis undergoing the most study Myriad analytical applications Three categories
More informationAtomic Spectroscopy AA/ICP/ICPMS:
Atomic Spectroscopy AA/ICP/ICPMS: A Comparison of Techniques VA AWWA/VWEA Lab Practices Conference July 25, 2016 Dan Davis Shimadzu Scientific Instruments AA/ICP/ICPMS: A Comparison of Techniques Topics
More informationFergus Keenan Shona McSheehy Julian Wills
Novel QCell Technology for Inference Removal in ICP-MS - Combining Low Mass Filtration with Kinetic Energy Discrimination Fergus Keenan Shona McSheehy Julian Wills 1 The world leader in serving science
More informationKey Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company
Key Analytical Issues: Sample Preparation, Interferences and Variability Tim Shelbourn, Eli Lilly and Company Presentation Outline Sample preparation objectives and challenges Some common interferences
More informationDirect Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS
Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS Application Note Semiconductor Authors Junichi Takahashi Agilent Technologies Tokyo, Japan Abstract Ammonium hydroxide
More informationDirect Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ
Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction
More informationThe New Agilent 7700 Series ICP-MS
The New Agilent 7700 Series ICP-MS the best ICP-MS just got better! Presenter: Uwe Noetzel and Jerome Darrouzes ICP-MS Product Specialist Life Sciences and Chemical Analysis Global Footprint Liquid Separation
More informationatomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive
Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific
More informationInductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur
Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur Introduction What is ICP-MS? Inductively Coupled Plasma Mass Spectrometry Mass spectrometry
More informationELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4
ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4 Tatiana Trejos, M.Sc Florida International University Department of Chemistry and Biochemistry International Forensic Research Institute Outline
More informationDetermination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS
Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS Application note Semiconductor Authors Junichi Takahashi Agilent Technologies, Japan Introduction
More informationCH. 21 Atomic Spectroscopy
CH. 21 Atomic Spectroscopy 21.1 Anthropology Puzzle? What did ancient people eat for a living? Laser Ablation-plasma ionization-mass spectrometry CH. 21 Atomic Spectroscopy 21.2 plasma In Atomic Spectroscopy
More informationHands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment
: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment Max Planck Institute for Nuclear Physics, Heidelberg, Germany E-mail: mykola.stepaniuk@mpi-hd.mpg.de Stefano Nisi E-mail: stefano.nisi@lngs.infn.it
More informationINDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY Edited by AKBAR MONTASER George Washington University Washington, D.C. 20052, USA WILEY-VCH New York Chichester Weinheim Brisbane Singapore Toronto CONTENTS
More informationTrace Analyses in Metal Matrices Using the ELAN DRC II
www.perkinelmer.com Trace Analyses in Metal Matrices Using the ELAN DRC II Introduction Analyses of matrices containing high levels of metals present a challenge for ICP-MS. First, the concentrations of
More informationEnhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS
Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris,
More informationa. An emission line as close as possible to the analyte resonance line
Practice Problem Set 5 Atomic Emission Spectroscopy 10-1 What is an internal standard and why is it used? An internal standard is a substance added to samples, blank, and standards. The ratio of the signal
More informationUNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.
UNIVERSITI SAINS MALAYSIA Second Semester Examination Academic Session 2004/2005 March 2005 KAA 502 Atomic Spectroscopy Time: 3 hours Please make sure this paper consists of FIVE typed pages before answering
More informationChapter 8: An Introduction to Optical Atomic Spectrometry
Chapter 8: An Introduction to Optical Atomic Spectrometry Sample is atomized (gaseous atoms/ions) absorption or emission measured Optical Atomic Spectra Atomization Methods Sample Introduction Methods
More informationAtomic Absorption & Atomic Fluorescence Spectrometry
Atomic Absorption & Atomic Fluorescence Spectrometry Sample Atomization Atomic Absorption (AA) Atomic Fluorescence (AF) - Both AA and AF require a light source - Like Molecular Absorption & Fluorescence,
More informationAutenticity control, provenance testing and fraud detection using mass spectrometry
Autenticity control, provenance testing and fraud detection using mass spectrometry Abdal-Azim Al-Terkawi Ersan Özelci Melissa-Jane Monks Narendra Lagumaddepalli Venkatareddy Seyed Mohsen Jebreiil Khadem,
More informationDetermination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS
Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS Application note Pharmaceutical Authors Miao Jing, Yingping Ni, Yanping Wang and Zhixu Zhang Agilent Technologies, China Introduction
More informationAnalysis Repeatability of Trace and Major Elements in a Water Sample
Analysis Repeatability of Trace and Major Elements in a Water Sample Agnès COSNIER HORIBA Scientific Longjumeau, France Keywords: environment Elements: Al, As, B, Ba, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Li,
More information9/13/10. Each spectral line is characteristic of an individual energy transition
Sensitive and selective determination of (primarily) metals at low concentrations Each spectral line is characteristic of an individual energy transition 1 Atomic Line Widths Why do atomic spectra have
More informationRapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.
Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA
More informationAgilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology
Agilent s New 8800 ICP-QQQ Transforming ICP-MS Technology 1 8800 ICP-QQQ Key Product Hardware High matrix introduction (HMI) technology Dual conical Extraction and Omega lens focuses ions across the mass
More informationOptimizing Analytical Performance in ICP-OES Applications
application note Optimizing Analytical Performance in ICP-OES Applications Introduction To obtain the best possible performance from an analytical instrument, it is necessary to optimize the operating
More informationICP/MS Multi-Element Standards
Standards Ultra Pure Matrix Special Packaging Traceability to National Reference Materials AccuStandard s ICP/MS Standards are formulated to meet the needs of this very special instrument. As matrix effect
More informationToday s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy
Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy Evrim Kilicgedik Product Specialist, Atomic Spectroscopy Agilent Technologies 04.11.2011 2011 The Atomic Spectroscopy
More informationHigh Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast
High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast Application Note Environmental Authors Austin Schultz and Jake Unnerstall, Elemental Scientific, Omaha NE, USA Steve Wilbur,
More informationCh. 8 Introduction to Optical Atomic Spectroscopy
Ch. 8 Introduction to Optical Atomic Spectroscopy 8.1 3 major types of Spectrometry elemental Optical Spectrometry Ch 9, 10 Mass Spectrometry Ch 11 X-ray Spectrometry Ch 12 In this chapter theories on
More informationENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques
ENVG 60500 FALL 2013 ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques HISTORY In the 1940s, arc and high-voltage spark spectrometry became widely utilized for metal analysis
More informationDetermination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS
APPLICATION NOTE 43374 Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS Authors Tomoko Vincent, Product Specialist, Thermo Fisher Scientific Keywords
More informationOpening the analytical black box: Insights into interferences, corrections, and data quality
Opening the analytical black box: Insights into interferences, corrections, and data quality Dr. Anthony J. Bednar Research Chemist USACE ERDC-EL 29 March 2012 D.M. Sirkis, W.E. Harris, T.J. Kelly USACE-NAP
More informationAgilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis
Agilent ICP-MS Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis High Sample Matrix Tolerance Superior & Simple Interference Removal Ultra Wide Linear
More informationThermo Scientific icap TQ ICP-MS: Typical limits of detection
TECHNICAL NOTE 43287 Thermo Scientific icap TQ ICP-MS: Typical limits of detection Authors Tomoko Vincent Keywords BEC, LOD, SQ-KED, TQ mass shift, TQ on mass, typical performance Introduction Inductively
More informationSimple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS
Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS Application note Environmental Authors Steve Wilbur, Craig Jones Agilent Technologies,
More informationICP-OES Application Note Number 35
ICP-OES Application Note Number 35 Rapid measurement of major, minor and trace levels in soils using the Varian 730-ES Vincent Calderon Varian, Inc. Introduction As part of the global strategy for sustainable
More informationCh. 9 Atomic Absorption & Atomic Fluorescence Spectrometry
Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry 9.1 9A. Atomization Most fundamental for both techniques. Typical types 1. flame - burner type 2. Electrothermal graphite furnace 3. Specialized
More informationICP-Mass Spectrometer
ICP-Mass Spectrometer New Mass Spectrometers The main issue: sequential vs. simultaneous Scanning, peak hopping are sequential Like viewing a photo through a peephole One pixel at a time Other parts of
More informationMulti-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS
Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS Application note Energy and fuels Authors Jenny Nelson, Agilent Technologies, USA Ed McCurdy, Agilent Technologies, UK Introduction
More informationTrace elemental analysis solutions for your application. June 6, 2018
Trace elemental analysis solutions for your application June 6, 2018 Understanding how each technique works Components of instrument Selection Criteria Application Fields OUTLINE All these techniques can
More informationOptical Atomic Spectroscopy
Optical Atomic Spectroscopy Methods to measure conentrations of primarily metallic elements at < ppm levels with high selectivity! Two main optical methodologies- -Atomic Absorption--need ground state
More informationSample Analysis Design Internal Standard
Internal Standard Advantages: Fluctuations are monitored in each sample/ calibration / blank Disadvantages: Assume that behavior of IS is the same as the analyte Instrumental Drift Correct for instrument
More informationThe 30-Minute Guide to ICP-MS
TECHNICAL NOTE The 30-Minute Guide to ICP-MS ICP - Mass Spectrometry A Worthy Member of the Atomic Spectroscopy Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been
More informationThe Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES
The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Authors Christine M. Rivera Doug Shrader
More informationPRINCIPLE OF ICP- AES
INTRODUCTION Non- flame atomic emission techniques, which use electrothermal means to atomize and excite the analyte, include inductively coupled plasma and arc spark. It has been 30 years since Inductively
More informationHandling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology
Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology Shona McSheehy Ducos The world leader in serving science Elemental Analysis What are you trying to accomplish?
More informationThe Agilent 7700x ICP-MS Advantage for Drinking Water Analysis
The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction
More informationAN INTRODUCTION TO ATOMIC SPECTROSCOPY
AN INTRODUCTION TO ATOMIC SPECTROSCOPY Atomic spectroscopy deals with the absorption, emission, or fluorescence by atom or elementary ions. Two regions of the spectrum yield atomic information- the UV-visible
More informationNucleus. Electron Cloud
Atomic Structure I. Picture of an Atom Nucleus Electron Cloud II. Subatomic particles Particle Symbol Charge Relative Mass (amu) protons p + +1 1.0073 neutrons n 0 1.0087 electrons e - -1 0.00054858 Compare
More informationImplementing EPA Method 6020 with the Agilent ICP-MS Portfolio
Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio NEMC August 2018 Craig Jones Applications Specialist ICP-MS, ICP-MS/MS The Agilent Atomic Spectroscopy Lineup An Instrument for Any Application!
More informationAgilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)
www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +1.847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS) The Agilent
More informationChina Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS
Application Note Environmental China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS Fast, accurate quantitative analysis of 16 elements using an Agilent 7800 ICP-MS with
More information2101 Atomic Spectroscopy
2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.
More informationAnalysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS
ICP-MS Environmental Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS Application Note Steve Wilbur Analysis of arsenic and selenium in seawater at trace
More informationU.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP
Prodigy ICP Application Note: 1035 U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP Introduction This Application Note describes the capability of the Teledyne Leeman Labs Prodigy High
More informationChemistry Instrumental Analysis Lecture 17. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 17 Introduction to Optical Atomic Spectrometry From molecular to elemental analysis there are three major techniques used for elemental analysis: Optical spectrometry
More informationComparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS
Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS Technical Overview Abstract Inductively coupled plasma mass spectrometry
More informationIMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES
DoD Environmental Data Quality Workshop, St. Louis, MO IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES Jamie Fox Brooks Applied Labs www.brooksapplied.com
More informationUltra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES
Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES Application note Geochemistry, metals, mining Authors John Cauduro Agilent Technologies, Mulgrave, Australia Introduction
More informationFebruary 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:
February 20, 2012 Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Subject: ICP-MS Report Job Number: S0CHG688 Dear Joe: Please find enclosed the procedure report for the analysis
More information5. Is Chlorine important (far UV option for some ICP-OES spectrometers)?
CRITERIA for method selection ATOMIC SPECTROSCOPY 1 Introduction Inductively coupled plasma-optical emission spectrometry (ICP-OES) is an attractive technique that has led many analysts to ask whether
More informationChemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy
Atomic line widths: Narrow line widths reduce the possibility of spectral overlap and thus interferences. The band width at half height is used to indicate width. This is also sometimes called the effective
More informationCurrent State of Extraction Don t Be Deceived! Sharon F. Webb, Ph.D. Director of Quality Program
Current State of Extraction Don t Be Deceived! Sharon F. Webb, Ph.D. Director of Quality Program Overview Factors Purpose of Dissolution Quality Objectives of Program Effectiveness of Dissolution Technique
More informationAccurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS
APPLICATION NOTE 434 Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS Authors Introduction Marcus Manecki, Daniel Kutscher, Shona McSheehy
More informationLead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode
Lead isotope analysis: Removal of Hg isobaric interference from using ICP-QQQ in MS/MS mode Application note Authors Geochemistry and isotope analysis Glenn Woods Agilent Technologies, LDA UK Ltd., Stockport,
More informationUltra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode
Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode Application note Semiconductor Author Katsuo Mizobuchi Masakazu
More informationBasic and Application Inductively coupled plasma mass spectrometry (ICPMS)
Basic and Application Inductively coupled plasma mass spectrometry (ICPMS) 2 What is ICP MS Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry which is capable of detecting
More informationSingle-Element Standards for AAS
Single-Element Standards for AAS for AAS Flame Silver Ag in 2-5% HNO 3 Aluminium Al in 2-5% HCl Aluminium Al in 2-5% HNO 3 Arsenic As in 2-5% HCl Arsenic As in 2-5% HNO 3 Gold Au in 2-5% HCl Boron B in
More informationLow Level Detection of Priority Inorganic Pollutants in Water. Anthony Neumann, MSc Darren Crichton, BSc
Low Level Detection of Priority Inorganic Pollutants in Water Anthony Neumann, MSc Darren Crichton, BSc Priority List of Pollutants Consists of 16 metals and/or transitional elements. Enter the human food
More information7) Applications of Nuclear Radiation in Science and Technique (1) Analytical applications (Radiometric titration)
7) Applications of Nuclear Radiation in Science and Technique (1) (Radiometric titration) The radioactive material is indicator Precipitation reactions Complex formation reactions Principle of a precipitation
More informationICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer
Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer Inductively Coupled Plasma Optical Emission Spectrometer is powerful simultaneous full
More informationUse of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark
Use of ICP-MS in analysing radioisotopes Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark Inductively Coupled Plasma Mass Spectrometry (ICP-MS) History ICP-AES
More informationMulti Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE
1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for
More informationSmartNotes. What is meant by the term interference?
Interference Removal on ICP-OES icap 7000 Plus Series ICP-OES SmartNotes What is meant by the term interference? In ICP-OES we speak of interference when a result is biased either by other components in
More informationRapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry
Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry Zhichao Lin, Kathryn Emanuele, Stephanie Healey, and Patrick Regan Analytical Branch Winchester Engineering and
More informationCapillary Electrophoresis Gas Chromatography
Agilent 7500 Inductively Coupled Plasma Mass Spectrometry Course Number H8974A ChemStation Revision 01.XX NT Operating System Student Manual Revision 1 Mass Spectrometry Data Systems Liquid Chromatography
More informationThis work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under contract
This work was performed under the auspices of the U.S. Department of Energy by under contract DE-AC52-7NA27344. Lawrence Livermore National Security, LLC The ITER tokamak Tungsten (W) is attractive as
More information