COOKBOOK Book One AI Atomic Absorption Spectrometer

Size: px
Start display at page:

Download "COOKBOOK Book One AI Atomic Absorption Spectrometer"

Transcription

1 COOKBOOK Book One AI 12 Atomic Absorption Spectrometer Updated: Jan 22

2 AI 12 Cookbook Table of Contents AI 12 COOKBOOK Table of Contents BOOK ONE- FAAS Chapter 1: Theory of AAS Introduction 2 Flame Atomic Absorption Spectrometry (FAAS) 3 Graphite Furnace Atomic Absorption Spectrometry (GFAAS) 4 Vapor Hydride Generation Atomic Absorption Spectrometry (VG AAS) 5 Chapter 2: AAS Instrumentation Fundamentals 9 Light Source Atomizer Optics Detector Optics 1 Lenses Mirrors Monochromator Diffraction Grating Slit Width Atomizer 15 Flame Graphite Furnace Detector 18 Chapter 3: Background Correction Fundamentals 22 The Frequency of Measurement The Interval between Measurements The Function used to Calculate Net Absorption Spectral and Structured Backgrounds The Effect on the Linear Working Range Deuterium (D2) Background Correction 24 Smith-Hieftje (S-H) Background Correction 27 Zeeman Background Correction 28 Comparison of Background Correction Methods 29 i

3 AI 12 Cookbook Table of Contents Chapter 4: Comparison of Analytical Techniques Things to Consider 34 Applications 34 Expected Concentration Ranges 34 Elements 34 Atomization Efficiency 34 Interferences 35 Spectral Background Matrix Detection Limits 35 Sensitivity 35 Precision 35 Linear Working Range 35 Minimum Sample Volume 36 Sample Throughput 36 Sample Usage 36 Total Dissolved Solids 36 Method Development 36 Ease of Use 36 Automation/Unattended Operation 36 Costs 37 Initial Investment Running Costs Chapter 5: Standard and Sample Preparation Apparatus 41 Water 41 Standard and Blank Solutions 41 Sample Solutions 42 Storage of Solutions 42 Calibrations 43 Matrix Effects 43 Chemical Interferences 43 Incomplete dissociation of analyte compounds Ionization ii

4 AI 12 Cookbook Table of Contents Chapter 6: FAAS Analytical Data Sheets Introduction 47 THE ELEMENTS Aluminum, Al 49 Antimony, Sb 5 Arsenic, As 51 Barium, Ba 52 Beryllium, Be 53 Bismuth, Bi 54 Boron, B 55 Cadmium, Cd 56 Calcium, Ca (air/acetylene) 57 Calcium, Ca (nitrous oxide/acetylene) 58 Cesium, Cs 59 Chromium, Cr (air/acetylene) 6 Chromium, Cr (nitrous oxide/acetylene) 61 Cobalt, Co 62 Copper, Cu 63 Dysprosium, Dy 64 Erbium, Er 65 Europium, Eu 66 Gadolinium, Gd 67 Gallium, Ga 68 Germanium, Ge 69 Gold, Au 7 Hafnium, Hf 71 Holmium, Ho 72 Indium, In 73 Iridium, Ir 74 Iron, Fe 75 Lanthanum, La 76 Lead, Pb 77 Lithium, Li 78 Lutetium, Lu 79 Magnesium, Mg 8 Manganese, Mn 81 Mercury, Hg 82 Molybdenum, Mo 83 Neodymium, Nd 84 Nickel, Ni 85 Niobium, Nb 86 Osmium, Os 87 Palladium, Pd 88 Phosphorous, P 89 Platinum, Pt 9 Potassium, K 91 Praseodymium, Pr 92 Rhenium, Re 93 Rhodium, Rh 94 Rubidium, Rb 95 Ruthenium, Ru 96 Samarium, Sm 97 iii

5 AI 12 Cookbook Table of Contents Scandium, Sc 98 Selenium, Se 99 Silicon, Si 1 Silver, Ag 11 Sodium, Na 12 Strontium, Sr 13 Tantalum, Ta 14 Tellurium, Te 15 Thallium, Tl 16 Tin, Sn 17 Titanium, Ti 18 Tungsten, W 19 Uranium, U 11 Vanadium, V 111 Ytterbium, Yb 112 Yttrium, Y 113 Zinc, Zn 114 Zirconium, Zr 115 Chapter 7: Practical Applications - FAAS Marine 12 Water 121 Biological 122 Food 123 Agricultural 124 Petroleum 127 Miscellaneous 129 References 132 iv

6 1 - Theory of AAS Chapter 1: Theory of AAS Introduction Flame Atomic Absorption Spectrometry (FAAS) Graphite Furnace Atomic Absorption Spectrometry (GFAAS) Vapor Hydride Generation Atomic Absorption Spectrometry (VG AAS) 1

7 1 - Theory of AAS Introduction The essential elements of the theory behind the analytical technique of atomic absorption spectroscopy (AAS) are compacted into the following paragraphs. Atomic absorption spectroscopy (AAS) relies on the fact that the light absorption of free atoms [1-7]. All atoms can absorb light, but only at discrete wavelengths corresponding to the energy requirements of the particular atom. In other words, each element absorbs light at some specific and unique wavelength and does not absorb light at all on other wavelengths. For example, in a sample with multiple elements (say, copper, lead, iron, and nickel), only copper will absorb light that is at the characteristic wavelength for copper. Furthermore, the amount of light absorbed depends on the number of absorbing atoms that are present in the light path. All these factors enable AAS to be used as a tool for quantitative analysis. In practice, measuring the amount of light absorbed by several known standards allows a calibration curve to be constructed. Then, the unknown concentration of a sample can easily be determined based on the amount of light it absorbs. The amount of light energy absorbed at this wavelength depends on the concentration of the atoms in the medium (as dictated by Lambert s law and Beer s law). Lambert s law states that the portion of light absorbed by a transparent medium is independent of the intensity of the incident light and each successive unit layer of the medium absorbs an equal fraction of the light passing through it. Beer s law states that the light absorbed is proportional to the number of absorbing atoms in the medium. Mathematically, when light of intensity I o passes through a medium of length x with atom concentration of C, the intensity I of the light beam emerging from the medium is given by: I = I o e -kcx where k is a proportionality constant (the absorption coefficient). The absorption of the medium, A, is defined to be: A = lg (I o /I) = kcx This equation states that the absorbance, A, of the medium is linearly proportional to the concentration of the absorbing atoms. The absorption coefficient (or absorptivity), k, can be determined by constructing a calibration curve (i.e. plotting the observed absorbance versus the known sample concentration). The slope of the calibration curve is kx, and x is easily measurable or already known. Unknown sample concentrations may be determined from the calibration curve based on their measured absorbances. Any way that you look at, every AAS experiment can be broken down to the following procedure: A sample has an unknown amount of a known element (e.g. the sample is known to contain lead, but not how much lead). The sample must be made into a homogeneous, liquid solution (if it is not already). 2

8 1 - Theory of AAS A blank solution must be prepared. This blank must contain none of the element of interest. A series of standard solutions must be prepared. These standards have known (but varied) concentrations of the element of interest. These standards are used to prepare a calibration curve. Analyze the blank solution to determine the "blank" absorbance value. This is the absorbance value for a sample with a zero concentration of the element of interest. Individually analyze all of the standard solutions. Construct a calibration graph. For the blank and each standard solution, plot its absorbance value against its concentration. Analyze the unknown sample. The concentration of the unknown sample, based on its measured absorbance value, can be determined from the calibration curve. There are three main techniques of atomic absorption spectrometry: Flame atomic absorption spectrometry (FAAS), graphite furnace atomic absorption spectrometry (GFAAS), and vapor hydride generation atomic absorption spectrometry (VG AAS). Each has its own distinct advantages and disadvantages. Each has specific applications for which they are the superior AAS technique. For example, FAAS is suitable for analyses where the sample is above trace quantities, and where high sample throughput, ease of use, and low initial investment are required. GFAAS is ideal for samples that are in the parts per billion (ppb) range and where the sample volume is limited. VG AAS is useful for determining elements that form volatile hydrides at sub-trace levels. A brief outline of each technique is provided below. Flame Atomic Absorption Spectrometry (FAAS) The atomization process by which the atom population is generated is of primary importance in AAS because analysis depends entirely on the fact that free, uncombined atoms will absorb light of a particular wavelength. The key to successful operation of an atomic absorption spectrometer lies in generating a supply of free, uncombined atoms in the ground state and exposing this atom population to light at the characteristic absorption wavelength. The source of energy for free atom production is heat, most commonly in the form of an air-acetylene or nitrous oxideacetylene flame (Flame AAS). With this type of atomizer, the sample solution is introduced in the form of a spray of fine droplets. This is accomplished by a pneumatic nebulizer in most case. The spray of droplets is carried by a gas (usually the oxidant for the flame) through the spray chamber and burner head into the flame. The heat of the flame is sufficient to dry (desolvate) each of the sample droplets and (usually) to decompose chemical components from the resulting dried particles into their constituent atoms. Thus a population of ground state atoms is created in the flame and atomic absorption measurements can be made. Flame systems for AAS give excellent results, and they are simple, inexpensive, convenient and extremely useful. They permit rapid analytical measurements through a very simple sample introduction technique. The major limitation of flame AAS is that the burner-nebulizer system is a relatively inefficient sampling device. Only a small fraction of the sample that is taken up reaches the flame. Additionally, once atomized, the sample passes quickly through the light path. An improved sampling device would atomize the entire sample and retain it in the light path for an extended period of time to enhance the sensitivity of the 3

9 1 - Theory of AAS technique. Electrothermal atomization using a graphite furnace provides these features. Graphite Furnace Atomic Absorption Spectrometry (GFAAS) Graphite Furnace Atomic Absorption Spectrometry (GFAAS) has gained a reputation in the field of analytical chemistry as a routine technique for the determination of very low levels of trace metals in a variety of sample matrices. With GFAAS, the flame has been replaced by an electrothermally heated graphite tube. The sample is injected directly into the tube as a small liquid volume (5 to 1 µl), which is then heated in a programmed series of steps to remove the solvent and major matrix components. Free analyte atoms in the gaseous state are eventually produced inside the graphite furnace by rapidly heating with a strong electric current to temperatures between 15 and 3 C. All of the analyte is atomized, and the atoms are retained within the tube (and the light path, which passes through the tube) for an extended period (typically to.5 second). The performance of this technique relied on the stability of the temperature. The recently developed transversely heated integrated contact graphite furnace ensures the temperature over the entire length of the tube is very uniform. As a result, sensitivity and detection limits are significantly improved while matrix interferences and memory effects are reduced. After the measurement, the analyte vapor is then purged from the graphite furnace by helium or argon gas. The magnitude of the absorbance is recorded as a function of time by the readout system. The mechanism of atomization in a graphite furnace depends significantly on the chemical nature of the analyte element, the availability of active sites, the gaseous species within the graphite furnace, the atomization temperature, and the graphite furnace itself. After drying and pyrolysis the analyte atoms may be present in reduced, oxidized or complex form. Upon heating the graphite tube to the atomization temperature, free analyte atoms are generated by: (1) vaporization of the reduced form from the surface, (2) by dissociation of the oxide form into gaseous free analyte atoms as the graphite tube heats up, and (3) vaporization as oxides (or any other molecular species) and dissociation into free analyte atoms in the gaseous phase. Examples of these mechanisms are given below - where M is an analyte, C is carbon and O is oxygen and subscripts g and s refer to gaseous and solid phases: M (s) M (g) MO (s) + C (s) M (g) + CO (g) MO (s) MO (g) MO (g) M (g) + O (g) CO (g) + O (g) CO 2 (g). Analysis times for GFAAS are longer than those for FAAS, and fewer elements can be determined with this technique. Nonetheless, GFAAS s enhanced sensitivity, ability to analyze very small sample sizes, and ability to directly analyze certain types of solid samples significantly expand the capability of atomic absorption spectrometry. 4

10 1 - Theory of AAS Vapor Hydride Generation Atomic Absorption Spectrometry (VG AAS) For the determination of As, Bi, Ge, Pb, Sb, Se, Sn, Te and Hg with AAS, vapor/hydride generation (VG) techniques have been proven to provide very high sensitivities and reduced interferences. With VG AAS, analytes are first reduced to their corresponding volatile hydrides (or metallic form for Hg) by sodium borohydride in an acidic medium. The vapors are then transported by a carrier gas into the atomizer for atomization and AA measurement. Aurora Instrument s AI 12 uses an open ended, temperature controlled electrothermally heated quart tube for continuous flow vapor/hydride generation determinations. The heating unit can be installed and removed easily. 5

11 1 - Theory of AAS This page intentionally left blank. 6

12 1 - Theory of AAS This page intentionally left blank. 7

13 2 - Instrumentation Chapter 2: AAS Instrumentation Fundamentals Light Source Atomizer Optics Detector Optics Lenses Mirrors Monochromator Diffraction Grating Slit Width Atomizer Flame Graphite Furnace Detector 8

14 2 - Instrumentation Fundamentals There are four components that are essential to every AAS instrument: a light source, an atomizer, an optics system, and a detector. Light Source Usually a hollow cathode lamp (HCL) is used as the light source in AAS. A less common light source is the electrodeless discharge lamp (EDL). An HCL produces an intense, narrow line emission of light at a wavelength that is specific to the element that the HCL cathode is coated with. Most elements emit at multiple wavelengths, but all emissions are intense, sharp lines (called resonance lines). For example, a copper HCL emits at nm, nm, nm, nm, and nm. Generally, one line is more intense than the others and is therefore the most sensitive and useful line for AAS analysis. HCLs are coated with the element of interest to produce light of the resonance wavelength(s) that is/are specific to element. When this light is passed through a medium that contains atoms of the same element, the light will be partially absorbed. Atomizer The purpose of an atomizer is to create a population of free atoms that is suitable for absorption of light. The atomizer must have an energy source in order to do this. Usually the energy comes from heat, and the most common source of the heat is a flame (either an air/acetylene or a nitrous oxide/acetylene flame). An AAS instrument with a flame atomizer is called a flame atomic absorption spectrometer (FAAS). With a flame atomizer, the sample is introduced into the flame as an aerosol (a mist of tiny droplets). The flame burner head [8] is designed to be long, thin, and aligned with the light path. Such a design causes the aerosol atoms to be atomized in the flame while they are in the light path so that they can absorb the light. A very important part of the atomizer system is the nebulizer. The nebulizer is responsible for nebulizing a liquid sample into an aerosol. The sensitivity of a FAAS instrument depends heavily on how efficiently the nebulizer can convert a sample to an aerosol. Optics The optics system of an AAS instrument is responsible for getting the light from the light source to the detector. Along the way, the light must be passed through the atomized sample and through a monochromator. A monochromator is used to isolate specific wavelengths from the bulk light that it receives. For example, it may be necessary to isolate the analytical wavelength of interest from light that was emitted from the fill gas of the HCL, or from stray room light that entered the spectrometer. Detector The detector part of an AAS instrument measures how much light is transmitted through the spectrometer. Most commonly a photomultiplier tube (PMT) is employed for this purpose. While the above four components are the essential ones of an AAS instrument, there are still others that play important roles. 9

15 2 - Instrumentation There inevitably are electronic devices that convert the signals from the detector into something that is useful for the human researcher. Older instruments used to employ signal meters and plotters that would chart the strength of the absorbance signal on a moving strip of paper. Instruments of today have replaced the meter and chart recorder with computer software that has many more capabilities. Modern software provides real-time plots of absorbance versus time, constructs calibration curves, and calculates statistics such as RSD values. On the starting end of the sample analysis spectrum, computer software can keep track of samples that are running, that you will run, and that you did run. It can also be used to setup and run the instrument without operator intervention at all. Optics The ideal optics system will have the following characteristics: Have 1% efficient light throughput. In other words, if there is no atomized sample in the light path, then 1% of the light from the source will reach the detector. Allow zero stray light. Provide a high signal to noise ratio (S/N). Provide absolute selectivity and resolution for the wavelength being measured. Provide constant dispersion, regardless of wavelength. Have no optical aberrations. Provide unique performance over a wide wavelength range. Unfortunately, there is no such thing as an ideal optics system. The best that you can hope for is to have a system that is optimized for your needs. There are two ways to control the path of light within an optics system: with mirrors or with lenses. Most optics systems make use of both. Lenses Good quality lenses are made from silica glass and have good light transmission over a broad wavelength range (~19-9 nm). Transmission losses occur at both interfaces of the lenses (i.e. at the lens surface where the light enters and at the lens surface where the light exits). The losses are typically between 1-14% for each lens in the optical path. A feature of lenses that must be kept in mind is that the refractive index of the lens is dependent on the wavelength of the light being refracted. This means that the focal length of the lens will be different for every wavelength. Rather than moving a lens to keep the focal point in the same position for different wavelengths, optics systems will keep their lenses fixed and tolerate the relatively minor losses associated with the changing focal lengths. Most lenses used are designed for wavelengths in the UV region. This is because most analytical wavelengths are in this range and the median refractive index is about 25 nm. For an air/acetylene flame, where the path length is around 1 cm, the losses due to focal length differences are negligible. For a nitrous oxide/acetylene flame, however, the path length is only 5 cm and the losses can start to become noticeable. In GF AAS, where the sample atomization occurs 1

16 2 - Instrumentation only at a small point in the center of the graphite tube, the focal length from the light source is absolutely critical. As the light source wavelength is increased, the focal point will move away from its original position at the center of the graphite tube. Because of this, optics systems for GF AAS systems usually employ only mirrors, and not lenses. Mirrors Mirrors perform much better than lenses, both in terms of reflecting efficiency and focal length change. A mirror has a very thin (e.g. 2 µm) top coating of aluminum that reflects more than 9% of the light that strikes it (for the range 19-9 nm). Also, when light reflects off a mirror there is no change in the focal length. That is, the focal length of a mirror depends only on its shape and is independent of wavelength. Plane mirrors are used to fold light and curved mirrors (also called collimating mirrors) are used to focus light. For example, a plane mirror is needed to fold light around a 9 corner, and a curved mirror is needed to focus that light onto an entrance slit. While the use of mirrors does solve the problem of focal length differences, it raises the challenge of designing and manufacturing focusing mirrors that are free from other optical aberrations, such as astigmatism. Because of the thinness of the mirror coating, mirrors are extremely fragile and must be handled with care. Finger prints, and even soft tissues, can irreversibly damage a mirror s coating. Reactive liquids and gases can even cause harm. To increase their longevity, most optics mirrors are further coated with a transparent silica or magnesium fluoride film for protection. Even still, one should avoid any kind of direct contact with mirror surfaces. Monochromator There are several different designs of monochromators available. No matter which design is used in an AAS instrument, however, some fundamental principles remain the same. Light enters the monochromator through an entrance slit. The light is folded and focused in the monochromator by use of mirrors. The light is dispersed into its component wavelengths by some sort of diffracting element. The diffracted light then leaves the monochromator through an exit slit. The most common monochromator design is the Czerny-Turner design, which is used in the AI 12. A schematic of this type of monochromator is shown in Figure 2.1. The Czerny-Turner monochromator uses two separate mirrors to collimate and focus light. Mirror #1 receives the light that was focused through the entrance slit. The light that reflects off this mirror is collimated into parallel beams and then strikes the diffraction grating, which diffracts the light into a spectrum of wavelengths. This spectrum is dispersed at a variety of angles, depending on the wavelength of each component of the spectrum. The light then strikes Mirror #2, which focuses the light through the exit slit and into the detector. Some monochromators make use of only one mirror, but there is a definite advantage to using two mirrors. The two mirrors will each be smaller than a single mirror, so they are easier to manufacture. This means that there is less chance for surface aberrations and therefore allows optimum light throughput and resolution. 11

17 2 - Instrumentation Figure 2-1 Schematic diagram of a Czerny-Turner monochromator The detector is dumb it doesn t know whether the light it receives is an analytical signal or not. The detector merely counts the number of photons that it receives (the intensity of the light) and sends a signal to an amplifier then sends back to computer. So, it is the job of the monochromator to isolate a specific, narrow resonance line from the rest of the spectrum before the light is allowed to reach the detector. The monochromator ensures that only the analytical wavelength of light reaches the detector. This wavelength selectivity is achieved by rotation of the diffraction grating with respect to the incident light. Turning the grating moves the spectrum across the exit slit, and therefore changes the wavelength of light that passes through the slit. Another means of controlling the light that passes through the exit slit is by changing the width of the slit. A narrower slit allows two closely spaced wavelengths to be resolved, but it also decreases the light throughput of the optics system as a whole. If there are no interfering wavelengths close to the analytical wavelength, then a wider slit can be used to increase the light throughput, say, improve the signal noise ratio. Other, less common used monochromator designs are the Ebert-Fastie and the Littrow designs. The Ebert-Fastie monochromator design makes use of a single, large mirror to focus light. See Figure 2.2 for a schematic. One area of the mirror collimates incoming light onto the diffraction grating and then another area of the mirror focuses the light dispersed from the grating onto the exit slit. This single mirror does both of the jobs of the double mirrors in the Czerny-Turner design. A disadvantage of the Ebert-Fastie design is that the single mirror must be large. This increases the probability of surface imperfections during the manufacture of the mirror, which in turn impairs the light throughput and resolution of the monochromator. An Ebert-Fastie monochromator is, however, less expensive than a Czerny-Turner. 12

18 2 - Instrumentation Figure 2-2 Schematic diagram of an Ebert-Fastie monochromator The Littrow monochromator design is similar to the Ebert-Fastie design in that it also employs only one mirror. See Figure 2.3 for a schematic. The difference lies in that the Littrow monochromator uses the same area of the single mirror for collimating the light onto the grating as for focusing the dispersed light onto the exit slit. This fact increases the chances of optical aberrations (even more so than the Ebert-Fastie design). Figure 2-3 Schematic diagram of a Littrow monochromator Diffraction Grating There are two types of diffraction gratings that are commonly used today in AAS instruments: ruled gratings and holographic gratings. Ruled gratings have been around longer than holographic gratings, which were only introduced in the late 196s. A grating, in general, is a closely spaced series of grooves in a flat, reflecting surface. The grooves must be perfectly parallel and uniformly 13

19 2 - Instrumentation spaced from each other. The closer the grooves are spaced, the better the resolving capability of the grating. Gratings are produced with groove densities from 5 to 6 grooves/mm. The blaze wavelength of a diffraction grating is the wavelength of light that will be most efficiently diffracted by the grating. Generally, gratings can be used to diffract light that is 2/3 to 3/2 of the blaze wavelength. For example, consider a grating blazed at 4 nm. The grating can be used for wavelengths from 27 nm to 6 nm, but will diffract most efficiently wavelengths of 4 nm. Ruled grating are physically etched, groove-by-groove, by a machine. Basically, a mirror is mounted on a grooving machine, a diamond bit etches a straight groove, the mirror is moved a short distance and then the bit etches another straight groove parallel to the previous one. Holographic gratings, on the other hand, are manufactured with light, not a physical machine. A piece of glass is coated with a light-sensitive material, which is then exposed to two parallel beams of coherent light that produce an interference pattern on the coated glass. The bright areas of the interference pattern (where the light beams add constructively) form the grooves in the developed photoresist. A thin layer of aluminum is then applied onto the etched glass piece to form a mirrored, grooved surface a diffraction grating. The advantage of the holographic technique over the machine etching technique is that the holographic technique produces no systematic errors, since the grooves are the result of a perfect optical phenomenon. The machined technique, on the other hand, is only as good as the quality of the etching machine itself (which, for many purposes, is excellent, but will never be truly perfect ). Gratings that are used in monochromators are always copies of master gratings. A master grating is the original grating that was manufactured (either holographically or physically). Subsequent gratings can be made from the master by a process that essentially makes a molded copy of the master. There are advantages and disadvantages to both types of gratings. The appropriate one to use in a monochromator depends on several factors. Ruled gratings produce significant more stray light than holographic gratings, and this is especially true when groove density increases. For this reason, the maximum groove density of rules gratings is around 36 grooves/mm. Holographic gratings can have up to 6 grooves/mm. So, based on this factor, a holographic grating is better to use if a higher groove density is required to achieve a higher resolution and maintain a high signal to noise ratio. Ruled gratings do exhibit significantly better efficiency than holographic gratings. So, based on this factor, a ruled grating is more appropriate to use if light throughput is critical (for example, if the light source is very weak). If the light source is intense (as is the case in AAS instruments), then using a holographic grating is more advantageous than using a ruled grating. Slit Width The slit width affects how much light enters and exits the monochromator, and so is very important for light throughput. A wide slit width will allow more light to reach the detector and will improve the signal strength. But, if nonanalytical lines also reach the detector, then this will increase the noise and decrease the signal to noise ratio. Conversely, a narrow slit width will block out all non-analytical wavelengths, but may reduce the light throughput so much to make the signal to noise ratio unsatisfactory. So, the best slit width is the one that allows the most light to reach the detector and blocks out most of the noise. 14

20 2 - Instrumentation In other words, finding the optimum slit width is a compromise between maintaining high light throughput and maintaining a high signal to noise ratio. The maximum allowable slit width is generally determined by how closely spaced the analytical line of interest and its nearest neighbor in the spectrum are. The slit width must be narrow enough to block out any non-analytical lines, since they would increase the noise that the detector would see. Every spectrum is different, so the optimum slit width must be determined for each analysis. For elements whose analytical wavelengths are high (for example, rubidium at 78 nm), the spectra are usually not as dense as for elements with shorter analytical wavelengths (for example, zinc at 214 nm). Therefore, a larger slit width is usually permissible for elements like rubidium, whereas elements like zinc usually necessitate narrower slit widths. Atomizer The atomizer is arguably the most important part of an AAS instrument, since this is where the sample is converted into atoms that can absorb light. Having an efficient atomizer is essential. The absorbance signal is completely dependent on how many atoms there are to absorb the source light. So, a good atomizer will display good sensitivity to the sample being analyzed, and a poor atomizer will display poor sensitivity. The ideal atomizer has the following characteristics: Atomizes 1% of the sample delivered to it. No ionization occurs. No sample is left in the molecular, complexed form. Of course, no real atomizer is ideal, and the degree to which any sample is atomized depends to a large extent on the element being analyzed. Atomization is achieved by heating the sample to an extent where free ground state atoms are formed. In FAAS, atomization is done with a flame. In GFAAS, atomization is done with an electrically heated graphite tube furnace. In VG-AAS, atomization is done with an electrothermally heated furnace or a flame. Flame The atomizer in a FAAS instrument uses a nebulizer to convert a liquid sample into an aerosol. The sample is introduced into the nebulizer by aspiration though capillary tubing. The aspiration occurs pneumatically from the flow of fuel and oxidant gases through the nebulizer chamber. After the sample has traveled through the capillary tubing, it strikes a glass impact bead. This impact bead is designed so that when a stream of liquid strikes it, the liquid breaks apart into a mist of drops (an aerosol). This aerosol invariably contains drops of many sizes. The larger drops fall out, but the smaller drops remain suspended and are thoroughly mixed with the fuel and oxidant gases as the mixture is carried into the spray chamber. As the sample mist gets mixed with the gases, it moves along through the spray chamber and up towards the burner head. The sample entering the burner head is a uniform mixture of fuel gas, oxidant gas, and tiny sample droplets. Once the mixture enters the flame, the process of atomization by heat begins. The heat from the flame is usually sufficient to desolvate the sample droplets. Then, the solid particles that were formed (e.g. salts) are broken down, melted, or volatilized into gases. Finally, the molecules are thermally dissociated into atoms that are capable of absorbing their characteristic wavelength of light. Of course, the heat from the flame may be excessive and 15

21 2 - Instrumentation may ionize some atoms (remove an outer electron from the atom), thus decreasing the absorption signal. As well, the heat from the flame may be insufficient and not atomize enough of the sample molecules. This also decreases the absorption signal. The process of thermally atomizing the sample in the flame is a complex equilibrium, and the actual chemistry inside the flame at any stage (especially the atomization stage) is not clear. There are often numerous side reactions that occur simultaneously. The degree to which a sample may be ionized in a flame depends on the element, since each element has different energy requirements for ionization. To reduce ionization of atoms, easily ionizable elements (EIEs), such as the Group I elements (Li, Na, K, Cs) can be added to the original sample solution. If the EIEs are much more easily ionized than the sample atoms, then they will create a large population of electrons in the flame and shift the atomization/ionization equilibrium of the sample atoms towards the atomization side. There are two types of flames that are commonly used for FAAS: air/acetylene and nitrous oxide/acetylene. The air/acetylene flame (air being the oxidant, acetylene being the reductant, or fuel) burns at around 23 C. The nitrous oxide/acetylene flame burns much hotter at around 3 C. So, the flame temperature is a factor when determining which type of flame is best suited for atomization of a given element. Because of its cooler burning temperature, the air/acetylene flame works well for elements that are relatively easily atomized, such as copper, iron, nickel, and gold. The nitrous oxide/acetylene flame, with its higher burning temperature, is needed to atomize elements that require more energy to atomize, such as aluminum, silicon, titanium, and tungsten. Another important factor in optimizing the atomization of an element in a flame is the stoichiometry of the oxidant and reductant gases. A lean flame is fuel poor, and is therefore an oxidizing flame. A rich flame has excess fuel, and is therefore a reducing flame. For each type of flame, certain elements are atomized best in reducing flames, and certain elements are atomized best in oxidizing flames. There is extensive data on the absorbance characteristics of all the elements in flames. For example, in a reducing flame there are excess carbon and hydrogen atoms present in the flame (from the acetylene molecules). These excess atoms help break down the strong oxide bonds that form with some elements, such as chromium. Other elements that are best atomized in a reducing flame are tin and molybdenum. On the other end of the spectrum, elements like silver, cadmium, gold, and nickel are best atomized in an oxidizing flame. Some elements, like iron and gallium, are best atomized in a stoichiometric flame (i.e. neither rich nor lean). Furthermore, some elements are satisfactorily atomized over a wide range of flame gas mixtures. Copper, for example, is atomized in both rich and lean flames. For this reason, copper is often used to test or validate the sensitivity of an AAS instrument. The major disadvantage of the flame atomization technique is that it is very inefficient at converting the original sample to atoms. Overall, the atomizer system of a FAAS instrument can convert less than.1% of the original sample to absorbing atoms. The nebulizer component usually transports only less than 1% of the aspirated sample into the aerosol, and the other 9% is lost as waste in the spray chamber and nebulizer chamber. Furthermore, the sample that does make it into the flame (in aerosol form) is already greatly diluted from its mixing with the flame gases. And once the sample gets atomized in the flame, the atoms residence times in the light path are extremely short. Atoms travel through the light path at great speeds (at least 1 cm/1 ms) as they exit the slit in the burner head and travel up the flame. 16

22 2 - Instrumentation Graphite Furnace The graphite furnace (GF) atomizer solves the two major problems of the flame atomizer: poor atomization efficiency and short residence times. A graphite furnace is a tube that is connected to two low voltage electrodes. When a current is forced through the tube, the tube heats up. The amount of heating caused by the current flow can be accurately controlled, and the atomization of samples in a graphite furnace is usually performed over several heating steps. The graphite furnace is mounted in the electrodes so that one open end of the tube faces the light source and the other open end faces the entrance to the optics. This allows the light to pass freely through the graphite furnace along the axis of the tube. The tube is aligned so that the light path travels down the graphite tube axis and right through the center of the tube. With a GF atomizer, a very small amount of liquid sample (between 5 and 1 µl) is placed inside the center of the graphite tube. A heating program to atomize the sample consists of at least three principal steps: 1. Drying Step The graphite tube is quickly heated to a temperature just below the boiling point of the solvent. Then the temperature is slowly ramped past the boiling point. This step gently evaporates the solvent (without causing splattering or sample ejection) and leaves the dried sample inside the tube. 2. Ashing Step (or Charring) This step removes any dry or semi-dry matrix that is left over from the drying step. Matrix modifiers can be added to the sample before the heating program to stabilize the sample during the ashing step. Modifier gases, such as oxygen or hydrogen, can be added to the graphite furnace workhead during the ashing step to help remove the matrix. Ashing temperatures depend on the element being analyzed in its matrix. For example, cadmium ashes at 3 C, arsenic at 14 C, iron at 6 C, and lead at 48 C. 3. Atomization Step At this stage, the sample is a dry solid at the bottom of the graphite tube. The sample is atomized by rapidly increasing the temperature of the tube. The AI 12 can heat at a rate upto 38 K/s. The required atomization temperature depends on the element being analyzed. For example, cadmium atomizes at 125 C, arsenic at 225 C, nickel at 225 C, and lead at 14 C. There are distinct advantages to the graphite furnace atomizer when its performance is compared to the flame atomizer. The graphite furnace atomizes 1% of the sample (compared to less than.1% for the flame). Also, the residence time of the atomized sample in the light path is much longer in the graphite furnace than in the flame. The residence time in the graphite tube can range from to.5 second, whereas in the flame it is only milliseconds. Both of these factors increase the sensitivity of the graphite furnace atomizer, and detection limits with GFAAS are typically one to two orders of magnitude better than with FAAS. 17

23 2 - Instrumentation Detector The detector is the part of the AAS instrument that receives the light output from the monochromator. The detector quantifies how much light it receives and creates an electrical current. That current is then amplified and converted into a digital signal that is recorded by a data acquisition system. By far the most common detector used in AAS instruments is the photomultiplier tube (PMT). Essentially, a PMT is a photon counter. Light from the monochromator enters the PMT through a quartz window. Photons (the quantum packets that make up light) strike the photocathode of the PMT. This converts the photon to a photoelectron via the photoelectric effect. However, the production of a single electron from a single photon won t generate a very strong signal, so an amplification 5 or 6 orders of magnitude is required. This amplification is achieved through the use of a series of 8 to 12 dynode plates. A voltage is applied between the photocathode and first dynode (on the order of 1 V). This voltage difference causes the photoelectron to accelerate from the photocathode to the dynode. When it strikes the dynode, several more electrons are produced. These electrons are in turn accelerated towards the second dynode, since there is also a voltage difference across the first and second dynodes. Each of the electrons striking the second dynode creates several more electrons. For example, if the electron collision into the first dynode created 5 electrons, then the collision of those 5 electrons into the second dynode will create 25 electrons. This chain reaction of electron production continues along the series of dynodes. If there are 12 dynodes in the chain, then the original single photon will produce 244,14,625 electrons from the final twelfth dynode. These final electrons are then collected by the PMT anode. The current that can result from the collection due to a single photon by the PMT can be as high as 1 ma! Clearly, PMTs are extremely sensitive detectors. Furthermore, the large amplification of the signal that is achieved by the PMT is achieved with very little increase in noise. 18

24 2 - Instrumentation This page intentionally left blank. 19

25 2 - Instrumentation This page intentionally left blank. 2

26 3 - Background Correction Methods Chapter 3: Background Correction Methods Fundamentals The Frequency of Measurement The Interval between Measurements The Function used to Calculate Net Absorption Spectral and Structure Backgrounds The Effect on Linear Working Range Deuterium (D2) Background Correction Smith-Hieftje (S-H) Background Correction Zeeman Background Correction Comparison of Background Correction Methods 21

27 3 - Background Correction Methods Fundamentals Background correction is a necessary part of any good AAS instrument. Basically, there are two types of backgrounds that need to be corrected for: non-specific radiation and non-specific absorption. Non-specific radiation was dealt with in Chapter 2. Non-specific radiation is the extra light (non-analytical wavelengths) that can pass through the optics of an AAS instrument, enter the monochromator, and have the chance of reaching the detector. If this non-specific radiation reaches the detector, it will result in a falsely signal. Nonspecific radiation can come from many sources, including the HCL fill gas, sunlight, room light, and the light emitted by the flame itself. As was discussed in Chapter 2 (Monochromator section), it is the job of the monochromator to effectively filter all entering light and allow only a specific, variable wavelength to exit and reach the detector. So, the monochromator ensures that the only light that reaches the detector is the wavelength of light that is being absorbed by the sample being analyzed. The other type of background that must be corrected for is non-specific absorption (and will be discussed in this chapter). This correction cannot be accomplished by the monochromator, since it involves the same analytical wavelength that is selected by the monochromator. Non-specific absorption (also called background absorption) has a broadband effect and occurs when the source light is prevented from reaching the detector by means other than absorption by analyte atoms, such as scattering and blocking of light by other species in the light path. Molecular species and solid particles present in the flame are the major causes of non-specific absorption. When the heat of the flame is not sufficient to fully break down all molecular species (for example, matrix compounds), there can be sufficient remaining molecules to absorb, block, or scatter the source light. These molecules can be thought of as causing the same interference as putting one s hand in the light path: the light gets blocked, less light reaches the detector, and a falsely high absorbance signal results (because the signal analysis software thinks that less light is reaching the detector because more light is being absorbed by the analyte atoms). In FAAS, the background absorption is relatively minor (usually less than.5 absorbance units). In GFAAS, on the other hand, background absorption is severe and can reach levels of 2. absorbance units. Therefore, effective background correction methods become essential for accurate GF analyses. There are three background correction methods currently being used by AAS instrument manufacturers: Deuterium (D2), Smith-Hieftje (S-H), and Zeeman. For all three methods, there are several common factors that determine the effectiveness of the method: The frequency at which the peaks are measured; The interval between total absorbance and background absorbance measurements; The mathematical function used to calculate net atomic absorption; The ability to correct for spectral or structured background; The effect of the method on the linear working range. The Frequency of Measurement How rapid and transient the absorption peaks are dictates how fast absorption measurements must be made [9, 1]. In GF analyses, signal durations are typically to.5 seconds. The temporally uniform, isothermal atomization of the AI 12 provides peak durations that are in the low end of this range. The non-specific absorbance normally exhibits similar peak durations. Furthermore, the change in the absorbance values of these peaks can often be as high as 1 absorbance units per second. Therefore, a very high sampling frequency is required to accurately measure the very rapid and narrow peaks and to determine peak shapes. 22

28 3 - Background Correction Methods For a GFAAS system, a 1 Hz sampling frequency can produce significant errors in the measurement of peak height and/or peak area. While the performance can be improved if a 3 Hz frequency is used, the errors still remain significant. An increase to a 6 Hz sampling frequency will provide acceptable results in a GFAAS system. The AI 12 utilizes a sampling frequency as high as 1Hz in single beam mode and 12 Hz in double mode, giving excellent peak definition and negligible errors in peak area and peak height measurements. The Interval between Measurements The Net Signal (NAS) is the analytical signal of interest. NAS is the absorption of the light that is attributed only to the analyte atoms. It is obtained by subtracting the Background Signal (BAS) from the Total Signal (TAS). Ideally, if BAS and TAS were measured simultaneously, then there would be no error in the NAS calculation. In reality, however, only one measurement can be made at a time. The best thing to do, then, is to make the time interval between successive measurements as short as possible, to approach simultaneous measurements. The shorter the time interval between a BAS measurement and a TAS measurement, the less error there will be in the calculation of NAS. This factor is especially important for GFAAS, where the background signal can change quite rapidly compared to the analyte signal. The AI12, which uses the D2 background correction method, uses a D2/HCL modulation frequency of 1 KHz, resulting in a time interval between successive measurements of less than.5 ms. With this system, the HCL and D2 lamps are alternately pulsed so that only one light source is passing through the sample at a time. When the HCL is pulsed, the D2 lamp is turned off and only the HCL light passes through the sample. The signal measured is the TAS. When the D2 lamp is pulsed, the HCL is turned off and only the D2 light passes through the sample. The signal measured is the BAS. The Function used to Calculate Net Absorption If a sufficiently fast switching frequency between the BAS and TAS measurements is employed (such as in the AI 12), then a simple subtraction of the BAS from the TAS can be performed to obtain the NAS. If relatively long time periods are elapsed between the BAS and TAS measurements, then interpolation techniques will be needed to approximate the NAS values. Furthermore, there are increased chances of significant changes in the background signals occurring over the elapsed time periods. Therefore, these interpolation techniques are much more susceptible to errors and inaccurate NAS values. Spectral and Structured Background Less than 1% of the samples encountered in the real world exhibit spectral or structured background interferences that cannot be overcome by optimizing the atomizer (for GFAAS, in particular, by optimizing the furnace heating program) and/or using an appropriate chemical modifier. The Effect on the Linear Working Range Due to complex splitting patterns, the TAS response of an instrument equipped with Zeeman background correction can be non-linear. As a result, when the BAS is subtracted from the TAS, a roll over point on a calibration curve can occur. That is, two different concentration values could correspond to a single NAS value. In such a case, the linear dynamic range has been reduced by the effects of the background correction method. In comparison, the D2 background correction method does not produce a roll over point in calibration curves and so has no negative effects on the linear dynamic range. 23

29 3 - Background Correction Methods Deuterium (D2) Background Correction The light source used in the Deuterium (D2) background correction method is the D2 lamp. The D2 lamp is a continuum source, rather than a sharp line emitter. That is, it emits a spectrum of radiation covering 18 nm to 425 nm. Even though this means that the use of the D2 method is limited to the UV range, it is not much of a detriment since the most significant background absorptions occur at low wavelengths anyway. See figure 3.1 for a graphical explanation of how D2 background correction works. The key thing about D2 background correction is the assumption that the absorption of radiation by the analyte atoms alone has a negligible effect on the D2 spectrum. The monochromator exit slit is relatively wide (at least nm), so a spectrum of D2 light of that width is allowed through. Compare this to the tiny width of spectrum that the analyte absorption occurs in (approximately.2 nm), and it s easy to see how absorption occurring in such small region of a wide spectrum will have a negligible effect on that spectrum s overall intensity. In other words, only a small fraction of the D2 spectrum is attenuated by the analyte absorption, and the rest is completely unaffected. Conversely, the atomic absorption due the analyte atoms has a significant effect on the intensity of the HCL radiation, since the HCL line is very sharp and narrow to begin with. But the absorption that occurs in real-life experiments is not due to analyte atoms alone. It is always a combination of analyte absorption and background absorption. Background absorption is broadband and so has an equal effect on the intensity of both the HCL line and the D2 band. So, when the individual effects of the analyte absorption and the background absorption are added together, the result is that the HCL line is attenuated cumulatively by both and the D2 band is only attenuated by the broadband background absorption. This fact forms the foundation of the D2 background correction method. When the D2 lamp is pulsed, the absorption of the D2 band is measured (the BAS). When the HCL lamp is pulsed, the absorption of the HCL line is measured (the TAS). The difference between the two (TAS - BAS) is the absorption due to the analyte atoms alone (the NAS), and this is the figure of analytical significance. 24

ICP/MS Multi-Element Standards

ICP/MS Multi-Element Standards Standards Ultra Pure Matrix Special Packaging Traceability to National Reference Materials AccuStandard s ICP/MS Standards are formulated to meet the needs of this very special instrument. As matrix effect

More information

Single-Element Standards for AAS

Single-Element Standards for AAS Single-Element Standards for AAS for AAS Flame Silver Ag in 2-5% HNO 3 Aluminium Al in 2-5% HCl Aluminium Al in 2-5% HNO 3 Arsenic As in 2-5% HCl Arsenic As in 2-5% HNO 3 Gold Au in 2-5% HCl Boron B in

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

Emission spectrum of H

Emission spectrum of H Atomic Spectroscopy Atomic spectroscopy measures the spectra of elements in their atomic/ionized states. Atomic spectrometry, exploits quantized electronic transitions characteristic of each individual

More information

2101 Atomic Spectroscopy

2101 Atomic Spectroscopy 2101 Atomic Spectroscopy Atomic identification Atomic spectroscopy refers to the absorption and emission of ultraviolet to visible light by atoms and monoatomic ions. It is best used to analyze metals.

More information

Atomization. In Flame Emission

Atomization. In Flame Emission FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic by its monatomic particles in gaseous state in the

More information

OES - Optical Emission Spectrometer 2000

OES - Optical Emission Spectrometer 2000 OES - Optical Emission Spectrometer 2000 OES-2000 is used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the OES-2000 as an aerosol that is carried into the

More information

Chapter 2 Atoms and the Periodic Table

Chapter 2 Atoms and the Periodic Table Chapter 2 1 Chapter 2 Atoms and the Periodic Table Solutions to In-Chapter Problems 2.1 Each element is identified by a one- or two-letter symbol. Use the periodic table to find the symbol for each element.

More information

Note that the protons and neutrons are each almost 2,000 times more massive than an electron; What is the approximate diameter of an atom?

Note that the protons and neutrons are each almost 2,000 times more massive than an electron; What is the approximate diameter of an atom? Atomic Structure and the Periodic Table Evolution of Atomic Theory The ancient Greek scientist Democritus is often credited with developing the idea of the atom Democritus proposed that matter was, on

More information

ATOMIC SPECROSCOPY (AS)

ATOMIC SPECROSCOPY (AS) ATOMIC ABSORPTION ANALYTICAL CHEMISTRY ATOMIC SPECROSCOPY (AS) Atomic Absorption Spectroscopy 1- Flame Atomic Absorption Spectreoscopy (FAAS) 2- Electrothermal ( Flame-less ) Atomic Absorption Spectroscopy

More information

Writing Chemical formula with polyatomic groups

Writing Chemical formula with polyatomic groups Writing Chemical formula with polyatomic groups 1. Use the Periodic table to determine the combining powers of single elements. Eg. Magnesium is in Group 2 and has a combining power of 2. 2. Use Table

More information

Chemistry Instrumental Analysis Lecture 18. Chem 4631

Chemistry Instrumental Analysis Lecture 18. Chem 4631 Chemistry 4631 Instrumental Analysis Lecture 18 Instrumentation Radiation sources Hollow cathode lamp Most common source Consist of W anode and a cathode sealed in a glass tube filled with Ne or Ar. Hollow

More information

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Atomic Absorption Spectrophotometry Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon Defination In analytical chemistry, Atomic absorption spectroscopy is a

More information

Full file at

Full file at 16 Chapter 2: Atoms and the Periodic Table Solutions to In-Chapter Problems 2.1 Each element is identified by a one- or two-letter symbol. Use the periodic table to find the symbol for each element. a.

More information

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry 9.1 9A. Atomization Most fundamental for both techniques. Typical types 1. flame - burner type 2. Electrothermal graphite furnace 3. Specialized

More information

novaa 800 D Atomic Absorption Spectrometer

novaa 800 D Atomic Absorption Spectrometer Technical Data Atomic Absorption Spectrometer Cpt : +27 (0) 21 905 0476 Jhb : +27 (0) 11 794 Dbn : +27 (0) 31 266 2454 1/7 General The is a compact atomic absorption spectrometer with deuterium background

More information

Chemistry: A Molecular Approach, 2e (Tro) Chapter 2 Atoms and Elements. Multiple Choice Questions

Chemistry: A Molecular Approach, 2e (Tro) Chapter 2 Atoms and Elements. Multiple Choice Questions Chemistry: A Molecular Approach, 2e (Tro) Chapter 2 Atoms and Elements Multiple Choice Questions 1) In a chemical reaction, matter is neither created or destroyed. Which law does this refer to? A) Law

More information

Partial Energy Level Diagrams

Partial Energy Level Diagrams Partial Energy Level Diagrams 460 nm 323 nm 610 nm 330 nm 819 nm 404 nm 694 nm 671 nm 589 / 590 nm 767 / 769 nm Lithium Sodium Potassium Gas Mixtures Maximum Temperatures, C Air-Coal Gas 1825 Air-Propane

More information

Using the Periodic Table

Using the Periodic Table MATH SKILLS TRANSPARENCY WORKSHEET Using the Periodic Table 6 Use with Chapter 6, Section 6.2 1. Identify the number of valence electrons in each of the following elements. a. Ne e. O b. K f. Cl c. B g.

More information

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber CH217 Fundamentals of Analytical Chemistry Module Leader: Dr. Alison Willows Electromagnetic spectrum Properties of electromagnetic radiation Many properties of electromagnetic radiation can be described

More information

Electron Configurations

Electron Configurations Section 3 Electron Configurations Key Terms electron configuration Pauli exclusion principle noble gas Aufbau principle Hund s rule noble-gas configuration Main Ideas Electrons fill in the lowest-energy

More information

Chapter 2 Lecture Notes: Atoms

Chapter 2 Lecture Notes: Atoms Educational Goals Chapter 2 Lecture Notes: Atoms 1. Describe the subatomic structure of an atom. 2. Define the terms element and atomic symbol. 3. Understand how elements are arranged in the periodic table

More information

high temp ( K) Chapter 20: Atomic Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy high temp (2000-6000K) Chapter 20: Atomic Spectroscopy 20-1. An Overview Most compounds Atoms in gas phase high temp (2000-6000K) (AES) (AAS) (AFS) sample Mass-to-charge (ICP-MS) Atomic Absorption experiment

More information

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters ) September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum

More information

25 Instruments for Optical Spectrometry

25 Instruments for Optical Spectrometry 25 Instruments for Optical Spectrometry 25A INSTRUMENT COMPONENTS (1) source of radiant energy (2) wavelength selector (3) sample container (4) detector (5) signal processor and readout (a) (b) (c) Fig.

More information

COMPARISON OF ATOMIZERS

COMPARISON OF ATOMIZERS COMPARISON OF ATOMIZERS FOR ATOMIC ABSORPTION SPECTROSCOPY Introduction Atomic spectroscopic methods are all based on the interaction of light and analyte atoms in the gas phase. Thus, a common component

More information

An Introduction to Atoms

An Introduction to Atoms An Introduction to Atoms Matter (stuff) is made of atoms. John Dalton (1776-1884) Check your current model: Draw a carbon atom Model of the Atom Atoms are made of subatomic particles. There are three types

More information

ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light

ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of elements. It makes use of the absorption of light by these elements in order to measure their concentration.

More information

Prof. Dr. Biljana Škrbić, Jelena Živančev

Prof. Dr. Biljana Škrbić, Jelena Živančev 5 th CEFSER Training Course Analysis of chemical contaminants in food and the environment Faculty of Technology, University of Novi Sad, Novi Sad, Republic of Serbia 7-11 May 2012 Analysis of heavy elements

More information

Principles of Chemistry: A Molecular Approach 2e (Tro) Chapter 2 Atoms and Elements

Principles of Chemistry: A Molecular Approach 2e (Tro) Chapter 2 Atoms and Elements Principles of Chemistry: A Molecular Approach 2e (Tro) Chapter 2 Atoms and Elements 1) Which of the following is an example of the law of multiple proportions? A) A sample of chlorine is found to contain

More information

Title: Dec 5 8:12 AM (1 of 29)

Title: Dec 5 8:12 AM (1 of 29) Title: Dec 5 8:12 AM (1 of 29) Title: Dec 5 8:12 AM (2 of 29) Section 5.5, pages 184 187 Metals and Nonmetals Two major groups of elements are the metals and the nonmetals. Look at the examples below.

More information

1 Arranging the Elements

1 Arranging the Elements CHAPTER 12 1 Arranging the Elements SECTION The Periodic Table BEFORE YOU READ After you read this section, you should be able to answer these questions: How are elements arranged on the periodic table?

More information

Lecture 7: Atomic Spectroscopy

Lecture 7: Atomic Spectroscopy Lecture 7: Atomic Spectroscopy 1 Atomic spectroscopy The wavelengths of absorbance and emission from atoms in the gas phase are characteristic of atomic orbitals. 2 In the lowest energy transition, the

More information

Atomic absorption spectroscopy

Atomic absorption spectroscopy Atomic absorption spectroscopy Modern atomic absorption spectrometers Atomic absorption spectroscopy (AAS) is a spectroanalytical procedure for the quantitative determination of chemical elements using

More information

Principles of Chemistry: A Molecular Approach, 3e (Tro) Chapter 2 Atoms and Elements

Principles of Chemistry: A Molecular Approach, 3e (Tro) Chapter 2 Atoms and Elements Principles of Chemistry: A Molecular Approach, 3e (Tro) Chapter 2 Atoms and Elements 1) Which of the following is an example of the law of multiple proportions? A) A sample of chlorine is found to contain

More information

Chemistry/Additional Science

Chemistry/Additional Science Write your name here Surname Other names Pearson Edexcel GCSE Centre Number Candidate Number Chemistry/Additional Science Unit C2: Discovering Chemistry Foundation Tier Wednesday 14 June 2017 Morning Time:

More information

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT. CHEM 322 Name Exam 3 Spring 2013 Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT. Warm-up (3 points each). 1. In Raman Spectroscopy, molecules are promoted

More information

Optical Atomic Spectroscopy

Optical Atomic Spectroscopy Optical Atomic Spectroscopy Methods to measure conentrations of primarily metallic elements at < ppm levels with high selectivity! Two main optical methodologies- -Atomic Absorption--need ground state

More information

FLAME PHOTOMETRY AIM INTRODUCTION

FLAME PHOTOMETRY AIM INTRODUCTION FLAME PHOTOMETRY AIM INTRODUCTION Atomic spectroscopy is based on the absorption, emission or fluorescence process of light by atoms or elementary ions. Information for atomic scale is obtained in two

More information

Graphite Furnace Tubes

Graphite Furnace Tubes Graphite Furnace Tubes Quality Raw Materials for Long Life and Performance CPI graphite tubes are produced from one of the highest grades of spectrally pure graphite available today. Prior to use, the

More information

6.3 Classifying Elements with the Periodic Table

6.3 Classifying Elements with the Periodic Table 6.3 Classifying Elements with the Periodic Table The Periodic Table was developed by scientists to organize elements in such a way as to make sense of the growing information about their properties. The

More information

AN INTRODUCTION TO ATOMIC SPECTROSCOPY

AN INTRODUCTION TO ATOMIC SPECTROSCOPY AN INTRODUCTION TO ATOMIC SPECTROSCOPY Atomic spectroscopy deals with the absorption, emission, or fluorescence by atom or elementary ions. Two regions of the spectrum yield atomic information- the UV-visible

More information

materials and their properties

materials and their properties materials and their properties macroscopic properties phase state strength / stiffness electrical conductivity chemical properties color / transparence spectroscopical properties surface properties density

More information

B Gross properties of atoms and solids. B.1 The periodic table of elements

B Gross properties of atoms and solids. B.1 The periodic table of elements Introduction to Cluster Dynamics Paul-Gerhard Reinhard, Eric Suraud 2004 WILEY-VCH Verlag GmbH & Co. B Gross properties of atoms and solids Some basic properties and characteristics of atoms and bulk have

More information

Atomic Emission Spectra. and. Flame Tests. Burlingame High School Chemistry

Atomic Emission Spectra. and. Flame Tests. Burlingame High School Chemistry Atomic Structure Atomic Emission Spectra and Flame Tests Flame Tests Sodium potassium lithium When electrons are excited they bump up to a higher energy level. As they bounce back down they release energy

More information

Essential Chemistry for Biology

Essential Chemistry for Biology 1 Chapter 2 Essential Chemistry for Biology Biology and Society: More Precious than Gold A drought is a period of abnormally dry weather that changes the environment and one of the most devastating disasters.

More information

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY Introduction The purpose of this experiment is to determine the concentration of zinc in a sample of hair. You will use both the calibration

More information

8. Relax and do well.

8. Relax and do well. CHEM 1314 3;30 pm Theory Exam III John III. Gelder November 13, 2002 Name TA's Name Lab Section INSTRUCTIONS: 1. This examination consists of a total of 8 different pages. The last page include a periodic

More information

Candidate Name. Number. A.M. MONDAY, 17 January minutes

Candidate Name. Number. A.M. MONDAY, 17 January minutes Candidate Name Centre Number Candidate Number 0 GCSE 236/01 SCIENCE FOUNDATION TIER CHEMISTRY 1 A.M. MONDAY, 17 January 2011 45 minutes ADDITIONAL MATERIALS Question 1. 2. 3. 4. 5. 6. 7. 8. Total For s

More information

Gamma and X-Ray Standards

Gamma and X-Ray Standards Gamma and X-Ray Standards A wide range of gamma and x-ray standards for research and educational use are available in the energy range of 5.9 to 2614 kev. Many nuclides are available up to 100 µci (3.7

More information

a. An emission line as close as possible to the analyte resonance line

a. An emission line as close as possible to the analyte resonance line Practice Problem Set 5 Atomic Emission Spectroscopy 10-1 What is an internal standard and why is it used? An internal standard is a substance added to samples, blank, and standards. The ratio of the signal

More information

Ch. 3 Answer Key. O can be broken down to form two atoms of H and 1 atom of O. Hydrogen and oxygen are elements.

Ch. 3 Answer Key. O can be broken down to form two atoms of H and 1 atom of O. Hydrogen and oxygen are elements. Ch. 3 Answer Key 1. The Greeks believed that all matter is made of elements. We currently believe the same thing. However, the Greeks believed that there were 4 elements: earth, water, air and fire. Instead,

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Atomic line widths: Narrow line widths reduce the possibility of spectral overlap and thus interferences. The band width at half height is used to indicate width. This is also sometimes called the effective

More information

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy Topic 2b: X-ray Fluorescence Spectrometry Text: Chapter 12 Rouessac (1 week) 4.0 X-ray Fluorescence Download, read and understand EPA method 6010C ICP-OES Winter 2009 Page 1 Atomic X-ray Spectrometry Fundamental

More information

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy A. Evaluation of Analytical Parameters in Atomic Absorption Spectroscopy Objective The single feature that contributes most to making atomic

More information

Chemistry/Additional Science Unit C2: Discovering Chemistry

Chemistry/Additional Science Unit C2: Discovering Chemistry Write your name here Surname Other names Edexcel GCSE Centre Number Candidate Number Chemistry/Additional Science Unit C2: iscovering Chemistry Higher Tier Thursday 13 June 2013 Morning Time: 1 hour You

More information

Atomic Absorption Spectrometer ZEEnit P series

Atomic Absorption Spectrometer ZEEnit P series Atomic Absorption Spectrometer ZEEnit P series Technical Data ZEEnit series Update 07/2014 OBue 1/ 5 ZEEnit P series Variable high-end AA Spectrometer with Deuterium and Zeeman Background Correction with

More information

1 Arranging the Elements

1 Arranging the Elements CHAPTER 11 1 Arranging the Elements SECTION The Periodic Table BEFORE YOU READ After you read this section, you should be able to answer these questions: How are elements arranged on the periodic table?

More information

4.1 Atomic structure and the periodic table. GCSE Chemistry

4.1 Atomic structure and the periodic table. GCSE Chemistry 4.1 Atomic structure and the periodic table GCSE Chemistry All substances are made of atoms this is cannot be chemically broken down it is the smallest part of an element. Elements are made of only one

More information

What is Internal Energy?

What is Internal Energy? Thermal Energy What is Internal Energy? Individual molecules making up a body have energy Even in a solid, the vibration of the molecules give them kinetic energy The electromagnetic force between the

More information

Chemistry/Additional Science

Chemistry/Additional Science Write your name here Surname Other names Edexcel GCSE Centre Number Candidate Number Chemistry/Additional Science Unit C2: Discovering Chemistry Foundation Tier Monday 21 May 2012 Morning Time: 1 hour

More information

B. X : in phase; Y: out of phase C. X : out of phase; Y: in phase D. X : out of phase; Y: out of phase

B. X : in phase; Y: out of phase C. X : out of phase; Y: in phase D. X : out of phase; Y: out of phase 2015 April 24 Exam 3 Physics 106 Circle the letter of the single best answer. Each question is worth 1 point Physical Constants: proton charge = e = 1.60 10 19 C proton mass = m p = 1.67 10 27 kg electron

More information

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency. Light We can use different terms to describe light: Color Wavelength Frequency Light is composed of electromagnetic waves that travel through some medium. The properties of the medium determine how light

More information

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Chapter 9 Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques Emission Spectroscopy Flame and Plasma Emission Spectroscopy are based upon those particles that are

More information

EMISSION SPECTROSCOPY

EMISSION SPECTROSCOPY IFM The Department of Physics, Chemistry and Biology LAB 57 EMISSION SPECTROSCOPY NAME PERSONAL NUMBER DATE APPROVED I. OBJECTIVES - Understand the principle of atomic emission spectra. - Know how to acquire

More information

Pearson Edexcel GCSE Chemistry/Additional Science Unit C2: Discovering Chemistry

Pearson Edexcel GCSE Chemistry/Additional Science Unit C2: Discovering Chemistry Write your name here Surname Other names Pearson Edexcel GCSE Centre Number Candidate Number Chemistry/Additional Science Unit C2: Discovering Chemistry Higher Tier Wednesday 15 June 2016 Afternoon Time:

More information

1 Electrons and Chemical Bonding

1 Electrons and Chemical Bonding CHAPTER 13 1 Electrons and Chemical Bonding SECTION Chemical Bonding BEFORE YOU READ After you read this section, you should be able to answer these questions: What is chemical bonding? What are valence

More information

Thank you for choosing AIMS!

Thank you for choosing AIMS! TM Thank you for choosing AIMS! Please use this free activity in your classroom, and watch your students begin to experience the "Aha!" moments of real learning. We like hearing from you. Like us and share

More information

Atomic Absorption Spectroscopy

Atomic Absorption Spectroscopy CH 2252 Instrumental Methods of Analysis Unit IV Atomic Absorption Spectroscopy Dr. M. Subramanian Associate Professor Department of Chemical Engineering Sri Sivasubramaniya Nadar College of Engineering

More information

Atomic Physics. Chapter 6 X ray. Jinniu Hu 24/12/ /20/13

Atomic Physics. Chapter 6 X ray. Jinniu Hu 24/12/ /20/13 Atomic Physics Chapter 6 X ray 11/20/13 24/12/2018 Jinniu Hu 1!1 6.1 The discovery of X ray X-rays were discovered in 1895 by the German physicist Wilhelm Roentgen. He found that a beam of high-speed electrons

More information

Investigation of Nutrient Elements in Cucurbita pepo Using Atomic Absorption Spectrometry

Investigation of Nutrient Elements in Cucurbita pepo Using Atomic Absorption Spectrometry Available online at www.ilcpa.pl International Letters of Chemistry, Physics and Astronomy 2 (2013) 11-17 ISSN 2299-3843 Investigation of Nutrient Elements in Cucurbita pepo Using Atomic Absorption Spectrometry

More information

Lecture 16 Instrumentation for ICP AES-VIII-Instruments

Lecture 16 Instrumentation for ICP AES-VIII-Instruments Inductive Couple Plasma Atomic Emission Spectrometry (ICP-AES) for Pollution Monitoring Dr. J R Mudakavi Department of Chemical Engineering Indian Institute of Science, Bangalore Lecture 16 Instrumentation

More information

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry 13A Measurement Of Transmittance and Absorbance Absorption measurements based upon ultraviolet and visible radiation

More information

Nucleus. Electron Cloud

Nucleus. Electron Cloud Atomic Structure I. Picture of an Atom Nucleus Electron Cloud II. Subatomic particles Particle Symbol Charge Relative Mass (amu) protons p + +1 1.0073 neutrons n 0 1.0087 electrons e - -1 0.00054858 Compare

More information

GOZO COLLEGE BOYS SECONDARY SCHOOL

GOZO COLLEGE BOYS SECONDARY SCHOOL GOZO COLLEGE BOYS SECONDARY SCHOOL Embracing Diversity Half Yearly Exams 2013-2014 FORM 4 CHEMISTRY TIME: 1h 30min Name: Class: Useful Data: One Faraday is equivalent to 96500 C. 1 2 3 4 5 6 7 0 1 H 1

More information

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures Villanova University Date: Oct 2011 Page 1 of 9 Villanova University Villanova Urban Stormwater Partnership Watersheds Laboratory Standard Operating Procedure VUSP F Zinc Metal Determination Perkin Elmer

More information

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy. UNIVERSITI SAINS MALAYSIA Second Semester Examination Academic Session 2004/2005 March 2005 KAA 502 Atomic Spectroscopy Time: 3 hours Please make sure this paper consists of FIVE typed pages before answering

More information

Atomic Absorption Spectroscopy (AAS)

Atomic Absorption Spectroscopy (AAS) Atomic Absorption Spectroscopy (AAS) Alex Miller ABC s of Electrochemistry 3/8/2012 Contents What is Atomic Absorption Spectroscopy? Basic Anatomy of an AAS system Theory of Operation Practical Operation

More information

Name: SCH3U Worksheet-Trends

Name: SCH3U Worksheet-Trends PERIODIC TRENDS WORKSHEET 1. Choose which statement about the alkali metals lithium and cesium is correct. a) as the atomic number increases, the Electronegativity of the elements increases b) as the atomic

More information

Principles of Chemistry: A Molecular Approach (Tro) Chapter 2 Atoms and Elements

Principles of Chemistry: A Molecular Approach (Tro) Chapter 2 Atoms and Elements Principles of Chemistry: A Molecular Approach (Tro) Chapter 2 Atoms and Elements 1) Which of the following is an example of the law of multiple proportions? A) A sample of chlorine is found to contain

More information

Part I Assignment: Electron Configurations and the Periodic Table

Part I Assignment: Electron Configurations and the Periodic Table Chapter 11 The Periodic Table Part I Assignment: Electron Configurations and the Periodic Table Use your periodic table and your new knowledge of how it works with electron configurations to write complete

More information

CH. 21 Atomic Spectroscopy

CH. 21 Atomic Spectroscopy CH. 21 Atomic Spectroscopy 21.1 Anthropology Puzzle? What did ancient people eat for a living? Laser Ablation-plasma ionization-mass spectrometry CH. 21 Atomic Spectroscopy 21.2 plasma In Atomic Spectroscopy

More information

This book makes learning easier through the following features:

This book makes learning easier through the following features: PREFACE You will find this book interesting: Chemistry concepts presented in a diagrammatic form. Specially written to ease learning and to stimulate interest in Chemistry, this book will help students

More information

Matter and Atomic Structure

Matter and Atomic Structure Matter and Atomic Structure Chemistry is the study of matter and reactions between types of matter. What, exactly, is matter? Matter is anything that has mass and occupies space. The world contains lots

More information

7. Relax and do well.

7. Relax and do well. CHEM 1215 Exam II John II. Gelder October 7, 1998 Name TA's Name Lab Section INSTRUCTIONS: 1. This examination consists of a total of 5 different pages. The last page includes a periodic table and a solubility

More information

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Application Note Semiconductor Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ Authors Kazuo Yamanaka and Kazuhiro Sakai Agilent Technologies, Tokyo, Japan Introduction

More information

Chapter 1 -- Introduction to Qualitative Analysis. NAME: Lab Section: Date: Sign-Off:

Chapter 1 -- Introduction to Qualitative Analysis. NAME: Lab Section: Date: Sign-Off: Chapter 1 -- Introduction to Qualitative Analysis NAME: Lab Section: Date: Sign-Off: Chapter 1 -- Introduction to Qualitative Analysis Introduction Qualitative analysis is a way inorganic chemists have

More information

Atomic Absorption & Atomic Fluorescence Spectrometry

Atomic Absorption & Atomic Fluorescence Spectrometry Atomic Absorption & Atomic Fluorescence Spectrometry Sample Atomization Atomic Absorption (AA) Atomic Fluorescence (AF) - Both AA and AF require a light source - Like Molecular Absorption & Fluorescence,

More information

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I SCCH 211: Analytical Chemistry I Analytical Techniques Based on Optical Spectroscopy Course Details September 22 October 10 September 22 November 7 November 17 December 1 Topic Period Introduction to Spectrometric

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Lab Day and Time: Instructions. 1. Do not open the exam until you are told to start.

Lab Day and Time: Instructions. 1. Do not open the exam until you are told to start. Name: Lab Day and Time: Instructions 1. Do not open the exam until you are told to start. 2. This exam is closed note and closed book. You are not allowed to use any outside material while taking this

More information

The Periodic Table of the Elements

The Periodic Table of the Elements The Periodic Table of the Elements All matter is composed of elements. All of the elements are composed of atoms. An atom is the smallest part of an element which still retains the properties of that element.

More information

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection TECHNICAL NOTE 43427 Thermo Scientific icap RQ ICP-MS: Typical limits of detection Author Tomoko Vincent Keywords BEC, interference removal, KED, LOD Introduction Inductively Coupled Plasma Mass Spectrometry

More information

Chapter 4 Scintillation Detectors

Chapter 4 Scintillation Detectors Med Phys 4RA3, 4RB3/6R03 Radioisotopes and Radiation Methodology 4-1 4.1. Basic principle of the scintillator Chapter 4 Scintillation Detectors Scintillator Light sensor Ionizing radiation Light (visible,

More information

CertiPUR. A class of its own. Merck s ICP Laboratory now with ISO accreditation

CertiPUR. A class of its own. Merck s ICP Laboratory now with ISO accreditation CertiPUR A class of its own. Merck s ICP Laboratory now with ISO 17025 accreditation CertiPUR We are setting new standards What are your benefits of ISO 17025 accreditation? An interview with Dr. Dietmar

More information

Chapter 3: Stoichiometry

Chapter 3: Stoichiometry Chapter 3: Stoichiometry Chem 6A Michael J. Sailor, UC San Diego 1 Announcements: Thursday (Sep 29) quiz: Bring student ID or we cannot accept your quiz! No notes, no calculators Covers chapters 1 and

More information

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry Fundamentals and Applications to the Petrochemical Industry Outline Some background and fundamentals of ICPMS

More information

PE summary from last class:

PE summary from last class: PH300 Modern Physics SP11 Today: Finish photoelectric effect Photons Next Week: Atomic spectra/balmer series Lasers learned very early the difference between knowing the name of something and knowing something.!

More information

PRINCIPLES OF AAS atomization flame furnace atomization absorption

PRINCIPLES OF AAS atomization flame furnace atomization absorption INTRODUCTION Atomic absorption spectroscopy (AAS) currently is the most widely used of atomic spectroscopic techniques. AAS is a quantitative method of elemental analysis that is applicable to many metals

More information

II. Spectrophotometry (Chapters 17, 19, 20)

II. Spectrophotometry (Chapters 17, 19, 20) II. Spectrophotometry (Chapters 17, 19, 20) FUNDAMENTALS (Chapter 17) Spectrophotometry: any technique that uses light to measure concentrations (here: U and visible - ~190 800 nm) c = 2.99792 x 10 8 m/s

More information