1 Chemometrics and Intelligent Laboratory Systems 106 (2011) Contents lists available at ScienceDirect Chemometrics and Intelligent Laboratory Systems journal homepage: EEMizer: Automated modeling of fluorescence EEM data Rasmus Bro a,, Maider Vidal b a Department of ood Science, Quality and Technology, aculty of Life, Sciences, University of Copenhagen, Rolighedsvej 30, DK-1958, rederiksberg C, Denmark b Departamento de Química Aplicada, acultad de Química, Universidad, del País Vasco, Apdo. 1072, San Sebastián, Spain article info abstract Article history: Received 14 ebruary 2010 Received in revised form 10 May 2010 Accepted 15 June 2010 Available online 25 June 2010 Keywords: luorescence EEM PARAAC Multi-way Tensor or many years it has been known that PARAAC offers a very attractive approach for modeling fluorescence excitation emission matrices. Due to the uniqueness of the PARAAC model and analogy between the structure of fluorescence data and the PARAAC model, it is apparent that PARAAC can resolve overlapping signals into pure spectra and relative concentrations under mild conditions. There are hundreds of applications exemplifying this, but still the use of PARAAC has not spread from chemometrics to more mainstream analytical chemistry. Many reasons can be offered to explain this, but one seems to be that in practice it is difficult for chemometric novices to make use of PARAAC. Selection of wavelengths, handling of scatter and of outliers are all issues that must be dealt with in order to build a good PARAAC model. In this paper, a new algorithm called EEMizer is developed that aims to automate the use of PARAAC. Through several examples it is shown how this algorithm can provide appealing PARAAC models of data that would otherwise be hard to model Elsevier B.V. All rights reserved. 1. Introduction luorescence excitation emission measurements are used in many different fields such as skin analysis [1], fermentation monitoring [2 4], environmental analysis [5 7],food[8] and clinical analysis [9].Common to many of these fields is that the spectroscopic measurements are performed directly on complex mixtures rather than on simple purified samples. This has posed severe problems in the subsequent analysis of the data, because the contributions of the individual fluorophores usually overlap. Multivariate methods have made it possible to develop calibration models for specific properties but not to extract the full information available in the data. The introduction of tensor-based modeling through the use of PARAAC has revolutionized the information-retrieval from such data [10 13]. Using PARAAC, it is possible under some circumstances, to perform so-called mathematical chromatography; that is, to separate the mixture measurements into the contributions from the underlying individual chemical analytes. or each analyte, the pure excitation and emission spectra are obtained as well as the relative concentration. Even though PARAAC has improved the possibilities for extracting information from EEM data, the use of PARAAC is by no means simple. Skilled operators are needed, that have a thorough understanding of fluorescence data and of tensor modeling. Naturally, this puts a limit on how widely applied PARAAC is. Corresponding author. address: rb@kvl.dk (R. Bro). There are several issues that data analysts have to take into account when analyzing a specific data set. Some of these issues are fairly trivial but a number of decisions on the route to a good model are taken on complex grounds and based on specific experience. Hence developing an automated tool for building a sound model is not a trivial task. Some of the issues that must be addressed is handling of scattering; especially Rayleigh scattering, possible exclusion of non-informative or misleading variables and decision on possible outlying samples. urthermore, the number of PARAAC components has to be determined. As all these issues affect each other, the end point; a sound PARAAC model, is not easy to achieve and it is likely that an inexperienced data analyst fails in building a good model even though the data supports it. Some papers describe how to achieve good PARAAC models of EEM data, but still leave it to the analyst to make the needed choices to produce a good model [14,15]. It is the purpose in this paper, to develop a tool that can automate the PARAAC model building in such a way that PARAAC can be more widely applied and especially so that PARAAC can be accessible to chemists with less background in data analysis. 2. Theory 2.1. Requirements In order to be able to automate the modeling, it is necessary to make certain assumptions on the data available. These are listed below: A sufficient number of adequate samples are measured A sufficient resolution is used both for excitation and emission /$ see front matter 2010 Elsevier B.V. All rights reserved. doi: /j.chemolab

2 R. Bro, M. Vidal / Chemometrics and Intelligent Laboratory Systems 106 (2011) Appropriate measures have been taken to ensure that the EEM data is meaningful in the context of Beers law. The above list seems somewhat fluffy in its requirements but this is purposely so. What constitutes a sufficient number of samples depends highly on the complexity of the problem. or very welldefined and simple problems where the fluorophores are reasonably different with respect to excitation and emission spectra and where their signal-to-noise ratio is high, as few as two samples can be sufficient. In other cases, many more samples may be required to be able to accurately estimate the signals from the underlying fluorophores. It is a premise for this paper and the developed algorithm that it is assumed that it is possible to estimate a meaningful PARAAC model and this is left for the experimenter to ensure. The term adequate samples refers to the fact that the samples should be such that the underlying fluorophores are varying independently in the samples. Imagine a situation where hundreds of samples are supplied but these samples are all identical. Such a data set would not allow PARAAC to estimate the underlying variation. Hence, adequate samples implies that the samples are chosen in such a way that all fluorophores vary independently. This may be difficult to ensure in practice, but if the conditions are not met it is expected that the outcome of the algorithm will not be chemically meaningful. It is then the responsibility of the experimenter to conclude that the requirements were not fulfilled and take proper actions. Proper actions could e.g. be to use spiking of samples to provide independent variation. As with the number and sort of samples needed, similar statements can be made about the spectral resolution. Ideally, even down to two excitation and two emission wavelengths can be sufficient, but in practice, more wavelengths will allow separating more signals. It is a general experience that having many more than ten wavelengths is often useful and that having many hundreds of wavelengths is typically not critical. Still, even within these limits, the nature and similarity of the fluorophores is what determines what a sufficient number of wavelengths is. A prerequisite for PARAAC to provide meaningful results is that Beers law is approximately obeyed, e.g. by assuring that no or little inner filter effect occurs, that ph and temperature is kept constant etc. [15]. It is assumed in the following that the data can be meaningfully described by a low number of PARAAC components. 3. Overall factors to investigate Given that the above requirements are fulfilled or at least approximately fulfilled, the decision on what is an appropriate PARAAC model includes at least the following decisions: Wavelengths to include How to handle Rayleigh scattering Number of components to use Samples to exclude. These four factors all depend on each other, so they cannot be chosen sequentially. or now, a fairly costly approach is adopted where many combinations of each are investigated. It may likely be possible to speed up computations by using more intelligent designs to investigate different combinations Included wavelengths It is characteristic of many fluorescent compounds that the spectra are quite broad relative to the spectral range measured. Therefore, it is mostly not necessary to consider variable selection in much detail. The only part of an EEM that can sometimes beneficially be removed is the low excitation wavelength area. Due to the characteristics of many instruments, excitations below approximately 240 nm can be very noisy and have a high variance which can disturb modeling. Therefore it is reasonable to try to eliminate the lowest excitations and verify if this improves the model. The parameter to optimize is called LOWEXREMOVE and indicates the number of excitations to remove (starting from the lowest) Rayleigh Rayleigh (and Raman) scattering is inconsistent with PARAAC and has to be dealt with. Many approaches have been developed for this [16 21]. Here we adopt a simple approach setting the scatter bands to missing and inserting zeros at emission below the excitation wavelength (below the ridge of missing values). The only choice to be made is then how wide the band of missing values should be. or the given paper, only Rayleigh is considered and the parameter to optimize is called SCATWIDTH which gives the extent in nanometers around the diagonal of equal excitation and emission which is set to missing. or many applications, any possible Raman scattering can be handled by simply subtracting a blank measurement from all samples Components In general PARAAC models are tested from smaller number of components and upwards. This is so because using excessive number of components in PARAAC can give meaningless results for mathematical reasons [22]. To determine the number of components, models with increasing number of components are tested and the one with the highest number that is still providing a good model is chosen. The parameter AC determines the number of components Outliers or any data set and choice of parameters (such as the above), outlying samples must be identified and handled in order for the model to be meaningful. The following approach is taken to identify outliers. or a particular model, samples with a high leverage or high sum-squared residual are removed one by one until no samples are assessed as outliers. Leverages for the samples are defined as the elements on the diagonal of the hat matrix of the score matrix. The removal is done by converting both leverage and sum-squared residuals into relative measures by dividing with the median value of these. or a sample set with e.g. 70 samples, there will be 70 leverages and these are all divided with the median of these 70 values. A level is defined (default five) and a sample with either a relative leverage or sum-squared residual above the level (five) is a candidate for removal. The one with the highest relative measure is removed and the analysis repeated. If there are very few samples in the data set (default 20) then the level is increased in order to allow more variability. Also, if more than a certain fraction of the samples (default 10%) have been removed, the level is increased to avoid too many samples being removed. These settings may seem fairly conservative in assessing when a sample is considered outlying, but in practice they ensure that very extreme, hence completely dissimilar, samples are identified. At the same time, samples are retained that would otherwise be deemed extreme e.g. when having a designed data set (where all samples may be extremes to some extent). More modest outliers are often not crucial to exclude as they help span similar variation, and hence just lead to extreme score values but an otherwise sound model. The main factor to optimize is LEVEL that defines when a sample is considered an outlier. In the following though, this factor is kept constant at five, but it is perfectly possible to optimize LEVEL. 4. Objective An objective is needed in order to automatically decide which possible models are good and which are bad. There are several

3 88 R. Bro, M. Vidal / Chemometrics and Intelligent Laboratory Systems 106 (2011) possibilities for evaluating if a model is good but here the following three measures are chosen: fit-values, core consistency and split-half quality. IT values are given as a relative measure: IT =1 I i =1 I i =1 J j =1 J j =1 K e 2 ijk k =1 K x 2 ijk k =1 where e ijk is the residual of sample i at emission j and excitation k. Core consistency is calculated as described in the original article [22] and called COREC here. The core consistency is calculated as 0 COREC = d =1 e =1 f =1 1 2 g def t def C A another model on samples [C D]. The alternative split is to use samples [A C] and [B D]. These two alternative splits are both investigated and only the best one is used. This is reasonable because an unlucky split may lead to bad results due to splitting whereas it is unlikely that an otherwise bad model accidentally gives a good split-half result. The split-half quality is determined as the product of Tucker congruence [24] between the loadings of the two alternative models. or two competing models estimated on two halves of the data set, a set of loadings is obtained for each model in each mode. In the emission mode, the loadings are held in matrix B 1 for one half of the data and B 2 for the other half. The columns of the matrices are normalized and the Tucker congruence is calculated as the diagonal of B 1 T B 2 upon matching the components in the two models so that similar components appear as similar columns. The same is done for the excitations loadings (C 1 and C 2 ) and the element-wise product of these two congruence vectors is then multiplied to provide the splithalf quality. This measure is called SPLTH. All three measures vary between 0 and 1, where zero is bad and one is perfect. An overall measure is defined from the product of these three measures as EEMqual = IT*COREC*SPLTH: where is the number of components, t def is an element of the target core array of a PARAAC model which is a superdiagonal array with ones on the superdiagonal. The element g def is the core estimated in a least squares sense using the PARAAC components as loadings (see ref. [22] for more details). Split-half quality is determined from the similarity of the loadings obtained when fitting a PARAAC model to different subsets of the data. In practice this is done by splitting the data into the first half and last half of the samples and comparing the loadings [23]. If the correct number of components is used, the loadings should be the same in these two sample sets given that the samples reflect the same underlying variation. Note that this assumes that the samples represented in the first and the last half are approximately equally representative which may require some reshuffling of the samples depending on the design of the data. In practice, one alternative split is also investigated where the sample set is split into four parts ([A B C D]). The above split corresponds to using one model of [A B] and Thus, when EEMqual is close to one, all three measures are high and the model successful. In some cases, it may happen that a data set does not support e.g. splithalf validation for example because certain fluorophores occur only in one or a few samples. Then the overall measureeemqualmaybepersistently low in which case, suitable models can be found from scrutinizing the three individual measures that make up the final measure. 5. Algorithm EEMizer Having defined the parameters to optimize and an overall measure of the quality of the model, it is possible to devise an algorithm for finding a good PARAAC model. The algorithm takes as input the maximal number of components to test, maxac as well as the widths of windows where Rayleigh scatter is tentatively removed. Typically this is set to 5, 10, 15 and 25 nm, but others can be tested as well if more computational power is available or if the amount of scatter is deemed broader or more narrow. ig. 1. Typical examples of EEMs from each of the three data sets.

4 R. Bro, M. Vidal / Chemometrics and Intelligent Laboratory Systems 106 (2011) Above, the step of determining outliers proceeds as follows: UNCTION: Determine optimal model by removing outliers Estimate PARAAC model with AC factors and scatter removed Determine fit Determine core consistency Do split-half analysis and determine split-half quality. rom the estimated model, determine sample leverage and sumsquared residuals relative to respective median values. ind sample with highest leverage or sum-squared residual Adjust critical level If N10% of samples have been removed adjust the critical level using ig. 2. EEMQual for Dorrit data. fracfactor = Ncurrent!2 1 N original 0:10, where N current is the current number of ALGORITHM EEMizer or AC = 1 to maxac or SCATWIDTH = [5, 10, 15, 25] or LOWEXREMOVE = 1 to maxvalue Determine optimal model by removing outliers end end end samples and N original is the original number of samples. If original number of samples is lower than 20, then adjust the critical level using samplenumberfactor = 30 Ncurrent CRITICAL LEVEL = LEVEL*samplenumberfactor*fracfactor ind sample with highest leverage or sum-squared residual. If higher than the critical level, remove it. Repeat whole procedure until no more samples are removed. Upon running EEMizer, the best model is chosen for each number of components as the one yielding the highest EEMqual value. The ig. 3. Emission and excitation loadings for a four-component PARAAC model before and after EEMizer.

5 90 R. Bro, M. Vidal / Chemometrics and Intelligent Laboratory Systems 106 (2011) ig. 6. EEMQual for fermentation data. ig. 4. Result of EEMizer on DOM data. results are presented in terms of this value as a function of number of components and the user then simply has to pick the highest number of components for which the quality measure is sufficiently high. or very well-defined data, this will indeed be a simple task, but certain data sets can still provide somewhat ambiguous results. It may be that a PARAAC model needs five components, but the fifth component has a very low signal-to-noise. In such a case, the five component model may yield an intermediate value and it may be necessary that the user compares competing models with intermediate quality levels to find the appropriate model. The second data set consists of 65 EEMs which have been spectrally corrected for instrument biases, corrected for inner filter effects and Raman calibrated. As a result, all of the Rayleigh scatter and most of the Raman scatter has been removed. The data consists of water samples collected from a cruise on RV Gunnar Thorson in the Kattegat and Belt Sea region (at the entrance to the Baltic Sea) in August The data are part of a tutorial on doing PARAAC on samples for characterizing DOM (dissolved organic matter) fluorescence [14] which provides a practical guideline for achieving manually what this paper tries to do automatically. inally, the third data set is an example of low-resolution EEM data; in this case containing 338 samples of at-line measurements of fermentation broth [4] with 15 excitation and 15 emission wavelengths. 6. Materials and methods The EEMizer algorithm has been implemented in Matlab and is available at (May, 2010). The function depends on functions from PLS_Toolbox ver. 5.5 ( April 2010). Apart from the above-mentioned algorithmic settings, PARAAC is implemented using nonnegativity constraints in all modes in EEMizer. Three data sets have been analyzed with EEMizer. Detailed information on the data is available in the original papers. The first data set is called Dorrit and is a set of 27 samples containing various mixtures of hydroquinone, tryptophan, phenylalanine and dopa [25]. What is characteristic of this data set is that even though it seems like a simple data set, it is typically hard for inexperienced users to find a good model because of noisy variables as well as several outliers. A typical example of an EEM from this data set is shown in ig Results and discussion 7.1. Dorrit or the Dorrit data, EEMizer provides the overall result shown in ig. 2. The results may be interpreted in the following way. Up to four components are clearly supported and anything beyond five seems to be too many components. Looking into the five component model, it is clear that this model is not adequate (results not shown). One of the five components has spectral loadings that do not look sound and hence, the fourcomponent model is chosen. or this model, EEMizer has removed the seven lowest excitations, used a scatter width of 15 nm and retained all samples. As can be seen in ig. 3, the emission loadings look sound. urthermore, the scores are perfectly correlated to the ig. 5. Example of EEM (left) illustrating the low signal-to-noise level in these data and the PARAAC model estimate of the EEM (right) where a lot of the noise is effectively filtered off.

6 R. Bro, M. Vidal / Chemometrics and Intelligent Laboratory Systems 106 (2011) ig. 7. Emission and excitation loadings for a six-component PARAAC model before and after EEMizer. reference concentrations (not shown). It is noteworthy, that EEMizer was able to find a good model without removing any samples. This was not the case in the original analysis of the data [25] DOM The DOM data set is complicated because of very low signals and hence a large proportion of noise. The result of EEMizer on the DOM data set is shown in ig. 4. It seems that three components are supported. Looking into the results, the three-component model looks excellent from a chemical point of view. No samples or excitations are removed, so the result is essentially identical to what would be obtained immediately on the raw data (where Rayleigh scatter signal was already removed). It is interesting to see that the model of the data provides a significant noise-filtering of the original data (ig. 5) ermentation The fermentation data is different from the above examples in that the instrument used is a low-resolution process instrument. Due to the selected filters in the instrument, there is no scattering as all emissions are above the excitation wavelengths. Running EEMizer provides the results shown in ig. 6. It seems that definitely five and perhaps six components are supported. In the original publication, five components were suggested, but in fact the six-component model found by EEMizer appears valid and reasonable. Twelve samples are removed by EEMizer and the spectral loadings are reasonable and consistent with what was found in the original publication. This is not the case, when fitting the PARAAC model to the raw data (ig. 7). 8. Discussion EEMizer is an attempt at developing an automatic PARAAC model builder for EEM data. It provides good results for a number of applications of quite diverse nature which is very encouraging. The process of determining a good PARAAC model is significantly simplified compared to manual determination of optimal settings. Even for data as diverse as shown here, building a PARAAC model was simplified to at most assessing two competing models. urther refinements can definitely be envisioned and most of all a general speedup of the algorithm would be feasible. This can be achieved in various ways. or example, distributed computing seems like an obvious approach as the overall problem is easy to parallelize. Acknowledgments The VILLUM KANN RASMUSSEN foundation as well as the Danish Agency for Science, Technology and Innovations is gratefully thanked for financial support to the work reported here. Maider Vidal acknowledges financial support from Basque Government in the form of a scholar fellowship. The original contributors of the data used here are also thanked. The data sets are available at (March, 2010).

7 92 R. Bro, M. Vidal / Chemometrics and Intelligent Laboratory Systems 106 (2011) References [1] R.K. Lauridsen, H. Everland, L.. Nielsen, S.B. Engelsen, L. Norgaard, Exploratory multivariate spectroscopic study on human skin, Skin Res. Technol. 9 (2003) [2] A. Hagedorn, R.L. Legge, H. Budman, Evaluation of spectrofluorometry as a tool for estimation in fed-batch fermentations, Biotechnol. Bioeng. 83 (2003) [3] E. Morel, K. Santamaria, M. Perrier, S.R. Guiot, B. Tartakovsky, Application of multiwavelength fluorometry for on-line monitoring of an anaerobic digestion process, Water Res. 38 (2004) [4] P.P. Mortensen, R. Bro, Real-time monitoring and chemical profiling of a cultivation process, Chemom. Intell. Lab. Syst. 84 (2006) [5] R.M. Cory, D.M. McKnight, luorescence spectroscopy reveals ubiquitous presence of oxidized and reduced quinones in dissolved organic matter, Environ. Sci. Technol. 39 (2005) [6] G.J. Hall, J.E. Kenny, Estuarine water classification using EEM spectroscopy and PARAAC-SIMCA, Anal. Chim. Acta 581 (2007) [7] C.A. Stedmon, S. Markager, R. Bro, Tracing dissolved organic matter in aquatic environments using a new approach to fluorescence spectroscopy, Mar. Chem. 82 (2003) [8] J. Christensen, L. Nørgaard, R. Bro, S.B. Engelsen, Multivariate autofluorescence of intact food systems, Chem. Rev. 106 (2006) [9] E. Koller, O. Quehenberger, G. Jürgens, O.S. Wolfbeis, H. Esterbauer, Investigation of human plasma low density lipoprotein by three-dimensional fluorescence spectroscopy, EBS Lett. 198 (1986) [10] R.D. Jiji, G.A. Cooper, K.S. Booksh, Excitation emission matrix fluorescence based determination of carbamate pesticides and polycyclic aromatic hydrocarbons, Anal. Chim. Acta 397 (1999) [11] L. Moberg, G. Robertsson, B. Karlberg, Spectrofluorimetric determination of chlorophylls and pheopigments using parallel factor analysis, Talanta 54 (2001) [12] E. Alm, R. Bro, S.B. Engelsen, B. Karlberg, R. Torgrip, Vibrational overtone combination spectroscopy (VOCSY) a new way of using IRand NIR data, Anal. Bioanal. Chem. 388 (2007) [13] R. Bro, Review on multiway analysis in chemistry , Critical Reviews In Analytical Chemistry 36 (2006) [14] C.A. Stedmon, R. Bro, Characterizing dissolved organic matter fluorescence with parallel factor analysis: a tutorial, Limnol. Oceanogr. 6 (2008) [15] C.M. Andersen, R. Bro, Practical aspects of PARAAC modelling of fluorescence excitation emission data, J Chemom 17 (2003) [16] D.I.C. Kells, J.D.J. O'Neil, T. Hofmann, A method for eliminating Rayleigh scattering from fluorescence spectra, Anal. Biochem. 139 (1984) [17] R.D. Jiji, K.S. Booksh, Mitigation of Rayleigh and Raman spectral interferences in multiway calibration of excitation emission matrix fluorescence spectra, Anal. Chem. 72 (2000) [18] L.G. Thygesen, Å. Rinnan, S. Barsberg, J.K.S. Møller, Stabilizing the PARAAC decomposition of fluorescence spectra by insertion of zeros outside the data area, Chemom. Intell. Lab. Syst. 71 (2004) [19] A. Rinnan, C.M. Andersen, Handling of first-order Rayleigh scatter in PARAAC modelling of fluorescence excitation emission data, Chemom. Intell. Lab. Syst. 76 (2005) [20] A. Rinnan, K. Booksh, R. Bro, irst order Rayleigh as a separate component in the decomposition of fluorescence landscapes, Anal. Chim. Acta 537 (2005) [21] M. Bahram, R. Bro, C.A. Stedmon, A. Afkhami, Handling of Rayleigh and Raman scatter for PARAAC modeling of fluorescence data using interpolation, J Chemom 20 (2007) [22] R. Bro, H.A.L. Kiers, A new efficient method for determining the number of components in PARAAC models, J Chemom 17 (2003) [23] R.A. Harshman, W.S. De Sarbo, An application of PARAAC to a small sample problem, demonstrating preprocessing, orthogonality constraints, and split-half diagnostic techniques, in: H.G. Law, C.W. Snyder, J.A. Hattie, R.P. McDonald (Eds.), Research Methods for Multimode Data Analysis, Praeger Special Studies, New York, 1984, pp [24] G. Lorho,. Westad, R. Bro, Generalized correlation loadings extending correlation loadings to congruence and to multi-way models, Chemom. Intell. Lab. Syst. 84 (2006) [25] J. Riu, R. Bro, Jack-knife technique for outlier detection and estimation of standard errors in PARAAC models, Chemom. Intell. Lab. Syst. 65 (2003)

First order Rayleigh scatter as a separate component in the decomposition of fluorescence landscapes

First order Rayleigh scatter as a separate component in the decomposition of fluorescence landscapes Analytica Chimica Acta 537 (2005) 349 358 First order Rayleigh scatter as a separate component in the decomposition of fluorescence landscapes Åsmund Rinnan a,, Karl S. Booksh b, Rasmus Bro a a Quality

More information

Multivariate calibration What is in chemometrics for the analytical chemist?

Multivariate calibration What is in chemometrics for the analytical chemist? Analytica Chimica Acta 500 (2003) 185 194 Multivariate calibration What is in chemometrics for the analytical chemist? Rasmus Bro Department of Dairy and Food Science, The Royal Veterinary and Agricultural

More information

Maximum likelihood fitting through least squares algorithms

Maximum likelihood fitting through least squares algorithms Maximum likelihood fitting through least squares algorithms Rasmus Bro, Nicholaos Sidiropoulos & Age Smilde Chemometrics Group, The Royal Veterinary and Agricultural University, Denmark, rb@kvl.dk Department

More information

The Theory of HPLC. Quantitative and Qualitative HPLC

The Theory of HPLC. Quantitative and Qualitative HPLC The Theory of HPLC Quantitative and Qualitative HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Modeling multi-way data with linearly dependent loadings y

Modeling multi-way data with linearly dependent loadings y Special Issue Received: 17 August 2008, Revised: 18 October 2008, Accepted: 29 October 2008, Published online in Wiley InterScience: 2009 (www.interscience.wiley.com) DOI: 10.1002/cem.1206 Modeling multi-way

More information

Cross model validation and multiple testing in latent variable models

Cross model validation and multiple testing in latent variable models Cross model validation and multiple testing in latent variable models Frank Westad GE Healthcare Oslo, Norway 2nd European User Meeting on Multivariate Analysis Como, June 22, 2006 Outline Introduction

More information

Water-Quality Measurements with the HORIBA Jobin Yvon AquaLog

Water-Quality Measurements with the HORIBA Jobin Yvon AquaLog Water-Quality Measurements with the HORIBA Jobin Yvon AquaLog Adam M. Gilmore, Ph.D. HORIBA Scientific 3880 Park Avenue Edison, NJ 08822 TEL: 732-494-8660 135 Email: adam.gilmore@horiba.com Abstract Water-quality

More information

Spectroscopy tools for PAT applications in the Pharmaceutical Industry

Spectroscopy tools for PAT applications in the Pharmaceutical Industry Spectroscopy tools for PAT applications in the Pharmaceutical Industry Claude Didierjean Sr. Technology and Applications Consultant Real Time Analytics Mettler Toledo AutoChem, Inc. claude.didierjean@mt.com

More information

New method for the determination of benzoic and. sorbic acids in commercial orange juices based on

New method for the determination of benzoic and. sorbic acids in commercial orange juices based on New method for the determination of benzoic and sorbic acids in commercial orange juices based on second-order spectrophotometric data generated by a ph gradient flow injection technique (Supporting Information)

More information

Mixture Analysis Made Easier: Trace Impurity Identification in Photoresist Developer Solutions Using ATR-IR Spectroscopy and SIMPLISMA

Mixture Analysis Made Easier: Trace Impurity Identification in Photoresist Developer Solutions Using ATR-IR Spectroscopy and SIMPLISMA Mixture Analysis Made Easier: Trace Impurity Identification in Photoresist Developer Solutions Using ATR-IR Spectroscopy and SIMPLISMA Michel Hachey, Michael Boruta Advanced Chemistry Development, Inc.

More information

Chemometrics. 1. Find an important subset of the original variables.

Chemometrics. 1. Find an important subset of the original variables. Chemistry 311 2003-01-13 1 Chemometrics Chemometrics: Mathematical, statistical, graphical or symbolic methods to improve the understanding of chemical information. or The science of relating measurements

More information

SmartNotes. What is meant by the term interference?

SmartNotes. What is meant by the term interference? Interference Removal on ICP-OES icap 7000 Plus Series ICP-OES SmartNotes What is meant by the term interference? In ICP-OES we speak of interference when a result is biased either by other components in

More information

INDEPENDENT COMPONENT ANALYSIS (ICA) IN THE DECONVOLUTION OF OVERLAPPING HPLC AROMATIC PEAKS OF OIL

INDEPENDENT COMPONENT ANALYSIS (ICA) IN THE DECONVOLUTION OF OVERLAPPING HPLC AROMATIC PEAKS OF OIL INDEPENDENT COMPONENT ANALYSIS (ICA) IN THE DECONVOLUTION OF OVERLAPPING HPLC AROMATIC PEAKS OF OIL N. Pasadakis, V. Gaganis, P. Smaragdis 2 Mineral Resources Engineering Department Technical University

More information

Basics of Multivariate Modelling and Data Analysis

Basics of Multivariate Modelling and Data Analysis Basics of Multivariate Modelling and Data Analysis Kurt-Erik Häggblom 6. Principal component analysis (PCA) 6.1 Overview 6.2 Essentials of PCA 6.3 Numerical calculation of PCs 6.4 Effects of data preprocessing

More information

Drift Reduction For Metal-Oxide Sensor Arrays Using Canonical Correlation Regression And Partial Least Squares

Drift Reduction For Metal-Oxide Sensor Arrays Using Canonical Correlation Regression And Partial Least Squares Drift Reduction For Metal-Oxide Sensor Arrays Using Canonical Correlation Regression And Partial Least Squares R Gutierrez-Osuna Computer Science Department, Wright State University, Dayton, OH 45435,

More information

UV Quality Assurance. Dark Current Correction

UV Quality Assurance. Dark Current Correction UV Quality Assurance This document presents the Quality Assurance protocol used for the measurement of broad spectrum UV spectra according to the guidelines given by the United States Environmental Protection

More information

Introduction to Principal Component Analysis (PCA)

Introduction to Principal Component Analysis (PCA) Introduction to Principal Component Analysis (PCA) NESAC/BIO NESAC/BIO Daniel J. Graham PhD University of Washington NESAC/BIO MVSA Website 2010 Multivariate Analysis Multivariate analysis (MVA) methods

More information

ANALYZING ASTRONOMICAL OBSERVATIONS WITH THE NASA AMES PAH DATABASE

ANALYZING ASTRONOMICAL OBSERVATIONS WITH THE NASA AMES PAH DATABASE PAHs and the Universe C. Joblin and A.G.G.M. Tielens (eds) EAS Publications Series, 46 (2011) 117-122 www.eas.org ANALYZING ASTRONOMICAL OBSERVATIONS WITH THE NASA AMES PAH DATABASE J. Cami 1, 2 Abstract.

More information

Deconvolution of Overlapping HPLC Aromatic Hydrocarbons Peaks Using Independent Component Analysis ICA

Deconvolution of Overlapping HPLC Aromatic Hydrocarbons Peaks Using Independent Component Analysis ICA MITSUBISHI ELECTRIC RESEARCH LABORATORIES http://www.merl.com Deconvolution of Overlapping HPLC Aromatic Hydrocarbons Peaks Using Independent Component Analysis ICA V. Gaganis, N. Pasadakis, P. Smaragdis,

More information

RFI Mitigation for the Parkes Galactic All-Sky Survey (GASS)

RFI Mitigation for the Parkes Galactic All-Sky Survey (GASS) RFI Mitigation for the Parkes Galactic All-Sky Survey (GASS) Peter Kalberla Argelander-Institut für Astronomie, Auf dem Hügel 71, D-53121 Bonn, Germany E-mail: pkalberla@astro.uni-bonn.de The GASS is a

More information

Component Analysis of an Excitation-Emission Matrix of Water Samples Using PARAFAC Analysis

Component Analysis of an Excitation-Emission Matrix of Water Samples Using PARAFAC Analysis FP-2 Introduction Excitation-Emission Matrix (EEM) can be used in a wide variety of applications, especially in the analysis of environmental water. EEM provides the following information; (1) determining

More information

Elimination of variability sources in NIR spectra for the determination of fat content in olive fruits

Elimination of variability sources in NIR spectra for the determination of fat content in olive fruits Ref: C59 Elimination of variability sources in NIR spectra for the determination of fat content in olive fruits Natalia Hernández-Sánchez, Physical Properties Laboratory-Advanced Technologies In Agri-

More information

A Polynomial Curve Fitting Method for Baseline Drift Correction in the Chromatographic Analysis of Hydrocarbons in Environmental Samples

A Polynomial Curve Fitting Method for Baseline Drift Correction in the Chromatographic Analysis of Hydrocarbons in Environmental Samples Available online at www.sciencedirect.com ScienceDirect APCBEE Procedia 10 (2014 ) 2 6 ICESD 2014: February 19-21, Singapore A Polynomial Curve Fitting Method for Baseline Drift Correction in the Chromatographic

More information

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000) INTRODUCTION After a number of meetings and workshops, a group of representatives from 27 organizations

More information

Online NIR-Spectroscopy for the Live Monitoring of Hydrocarbon Index (HI) in Contaminated Soil Material

Online NIR-Spectroscopy for the Live Monitoring of Hydrocarbon Index (HI) in Contaminated Soil Material Online NIR-Spectroscopy for the Live Monitoring of Hydrocarbon Index (HI) in Contaminated Soil Material 20.12.2016 1 The Hydrocarbon Index (HI) Definition It is a cumulative parameter no absolute method

More information

Incompatibility Paradoxes

Incompatibility Paradoxes Chapter 22 Incompatibility Paradoxes 22.1 Simultaneous Values There is never any difficulty in supposing that a classical mechanical system possesses, at a particular instant of time, precise values of

More information

Optimal solutions to non-negative PARAFAC/multilinear NMF always exist

Optimal solutions to non-negative PARAFAC/multilinear NMF always exist Optimal solutions to non-negative PARAFAC/multilinear NMF always exist Lek-Heng Lim Stanford University Workshop on Tensor Decompositions and Applications CIRM, Luminy, France August 29 September 2, 2005

More information

AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-12 NATIONAL SECURITY COMPLEX

AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-12 NATIONAL SECURITY COMPLEX AUTOMATED TEMPLATE MATCHING METHOD FOR NMIS AT THE Y-1 NATIONAL SECURITY COMPLEX J. A. Mullens, J. K. Mattingly, L. G. Chiang, R. B. Oberer, J. T. Mihalczo ABSTRACT This paper describes a template matching

More information

CS 6375 Machine Learning

CS 6375 Machine Learning CS 6375 Machine Learning Decision Trees Instructor: Yang Liu 1 Supervised Classifier X 1 X 2. X M Ref class label 2 1 Three variables: Attribute 1: Hair = {blond, dark} Attribute 2: Height = {tall, short}

More information

BATCH PROCESS MONITORING THROUGH THE INTEGRATION OF SPECTRAL AND PROCESS DATA. Julian Morris, Elaine Martin and David Stewart

BATCH PROCESS MONITORING THROUGH THE INTEGRATION OF SPECTRAL AND PROCESS DATA. Julian Morris, Elaine Martin and David Stewart BATCH PROCESS MONITORING THROUGH THE INTEGRATION OF SPECTRAL AND PROCESS DATA Julian Morris, Elaine Martin and David Stewart Centre for Process Analytics and Control Technology School of Chemical Engineering

More information

Rejection of Data. Kevin Lehmann Department of Chemistry Princeton University

Rejection of Data. Kevin Lehmann Department of Chemistry Princeton University Reection of Data by Kevin Lehmann Department of Chemistry Princeton University Lehmann@princeton.edu Copyright Kevin Lehmann, 997. All rights reserved. You are welcome to use this document in your own

More information

The Role of a Center of Excellence in Developing PAT Applications

The Role of a Center of Excellence in Developing PAT Applications The Role of a Center of Excellence in Developing PAT Applications! Pittcon 2005 Orlando FL, March 2 nd 2005! Nils-Erik Andersson AstraZeneca R&D, Mölndal Sweden 1 Outline! Business needs and New ways of

More information

Advanced Spectroscopy Laboratory

Advanced Spectroscopy Laboratory Advanced Spectroscopy Laboratory - Raman Spectroscopy - Emission Spectroscopy - Absorption Spectroscopy - Raman Microscopy - Hyperspectral Imaging Spectroscopy FERGIELAB TM Raman Spectroscopy Absorption

More information

Functional Preprocessing for Multilayer Perceptrons

Functional Preprocessing for Multilayer Perceptrons Functional Preprocessing for Multilayer Perceptrons Fabrice Rossi and Brieuc Conan-Guez Projet AxIS, INRIA, Domaine de Voluceau, Rocquencourt, B.P. 105 78153 Le Chesnay Cedex, France CEREMADE, UMR CNRS

More information

Evaluating Labsphere s new UV-2000

Evaluating Labsphere s new UV-2000 Ciba Grenzach GmbH Optical Spectroscopy Sabrina Siegle Dr. Wolfgang Schlenker Tel. +49 76 24 / 12-28 13 E-mail wolfgang.schlenker@ciba.com May 26, 9 Evaluating Labsphere s new UV- Abstract Labsphere s

More information

Generalized Least Squares for Calibration Transfer. Barry M. Wise, Harald Martens and Martin Høy Eigenvector Research, Inc.

Generalized Least Squares for Calibration Transfer. Barry M. Wise, Harald Martens and Martin Høy Eigenvector Research, Inc. Generalized Least Squares for Calibration Transfer Barry M. Wise, Harald Martens and Martin Høy Eigenvector Research, Inc. Manson, WA 1 Outline The calibration transfer problem Instrument differences,

More information

PAT/QbD: Getting direction from other industries

PAT/QbD: Getting direction from other industries PAT/QbD: Getting direction from other industries IFPAC Cortona 2010 Poll of Process Users 1. Analytical failure prediction 2. Result validation 3. More process-specific information (timely, higher quality,

More information

7 Principal Components and Factor Analysis

7 Principal Components and Factor Analysis 7 Principal Components and actor nalysis 7.1 Principal Components a oal. Relationships between two variables can be graphically well captured in a meaningful way. or three variables this is also possible,

More information

Data Preprocessing Tasks

Data Preprocessing Tasks Data Tasks 1 2 3 Data Reduction 4 We re here. 1 Dimensionality Reduction Dimensionality reduction is a commonly used approach for generating fewer features. Typically used because too many features can

More information

MULTIVARIATE STATISTICAL ANALYSIS OF SPECTROSCOPIC DATA. Haisheng Lin, Ognjen Marjanovic, Barry Lennox

MULTIVARIATE STATISTICAL ANALYSIS OF SPECTROSCOPIC DATA. Haisheng Lin, Ognjen Marjanovic, Barry Lennox MULTIVARIATE STATISTICAL ANALYSIS OF SPECTROSCOPIC DATA Haisheng Lin, Ognjen Marjanovic, Barry Lennox Control Systems Centre, School of Electrical and Electronic Engineering, University of Manchester Abstract:

More information

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters ) September 17, 2018 Reference literature (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters 13-14 ) Reference.: https://slideplayer.com/slide/8354408/ Spectroscopy Usual Wavelength Type of Quantum

More information

Karhunen-Loéve (PCA) based detection of multiple oscillations in multiple measurement signals from large-scale process plants

Karhunen-Loéve (PCA) based detection of multiple oscillations in multiple measurement signals from large-scale process plants Karhunen-Loéve (PCA) based detection of multiple oscillations in multiple measurement signals from large-scale process plants P. F. Odgaard and M. V. Wickerhauser Abstract In the perspective of optimizing

More information

Automated Modal Parameter Estimation For Operational Modal Analysis of Large Systems

Automated Modal Parameter Estimation For Operational Modal Analysis of Large Systems Automated Modal Parameter Estimation For Operational Modal Analysis of Large Systems Palle Andersen Structural Vibration Solutions A/S Niels Jernes Vej 10, DK-9220 Aalborg East, Denmark, pa@svibs.com Rune

More information

Multimodal multiplex Raman spectroscopy optimized for in vivo chemometrics

Multimodal multiplex Raman spectroscopy optimized for in vivo chemometrics Multimodal multiplex Raman spectroscopy optimized for in vivo chemometrics S. T. McCain, M. E. Gehm, Y. Wang, N. P. Pitsianis, and D. J. Brady Duke University Fitzpatrick Center for Photonics and Communication

More information

The Canonical Tensor Decomposition and Its Applications to Social Network Analysis

The Canonical Tensor Decomposition and Its Applications to Social Network Analysis The Canonical Tensor Decomposition and Its Applications to Social Network Analysis Evrim Acar, Tamara G. Kolda and Daniel M. Dunlavy Sandia National Labs Sandia is a multiprogram laboratory operated by

More information

Agilent MassHunter Quantitative Data Analysis

Agilent MassHunter Quantitative Data Analysis Agilent MassHunter Quantitative Data Analysis Presenters: Howard Sanford Stephen Harnos MassHunter Quantitation: Batch Table, Compound Information Setup, Calibration Curve and Globals Settings 1 MassHunter

More information

2.5 Multivariate Curve Resolution (MCR)

2.5 Multivariate Curve Resolution (MCR) 2.5 Multivariate Curve Resolution (MCR) Lecturer: Dr. Lionel Blanchet The Multivariate Curve Resolution (MCR) methods are widely used in the analysis of mixtures in chemistry and biology. The main interest

More information

FERMENTATION BATCH PROCESS MONITORING BY STEP-BY-STEP ADAPTIVE MPCA. Ning He, Lei Xie, Shu-qing Wang, Jian-ming Zhang

FERMENTATION BATCH PROCESS MONITORING BY STEP-BY-STEP ADAPTIVE MPCA. Ning He, Lei Xie, Shu-qing Wang, Jian-ming Zhang FERMENTATION BATCH PROCESS MONITORING BY STEP-BY-STEP ADAPTIVE MPCA Ning He Lei Xie Shu-qing Wang ian-ming Zhang National ey Laboratory of Industrial Control Technology Zhejiang University Hangzhou 3007

More information

Spectroscopy in Transmission

Spectroscopy in Transmission Spectroscopy in Transmission + Reflectance UV/VIS - NIR Absorption spectra of solids and liquids can be measured with the desktop spectrograph Lambda 9. Extinctions up to in a wavelength range from UV

More information

Process Analytical Technology Diagnosis, Optimization and Monitoring of Chemical Processes

Process Analytical Technology Diagnosis, Optimization and Monitoring of Chemical Processes FRAUNHOFER INSTITUTe FoR Chemical Technology ICT Process Analytical Technology Diagnosis, Optimization and Monitoring of Chemical Processes Process Analytical Technology Diagnosis, Optimization and Monitoring

More information

Water Quality Measurements with HORIBA Jobin Yvon Fluorescence Instrumentation

Water Quality Measurements with HORIBA Jobin Yvon Fluorescence Instrumentation F e a t u r e A r t i c l e Feature Article Water Quality Measurements with HORIBA Jobin Yvon Fluorescence Instrumentation Adam M. Gilmore Water quality is one of the most significant global environmental

More information

Raman spectroscopy of phthalocyanines and their sulfonated derivatives

Raman spectroscopy of phthalocyanines and their sulfonated derivatives Journal of Molecular Structure 744 747 (2005) 481 485 www.elsevier.com/locate/molstruc Raman spectroscopy of phthalocyanines and their sulfonated derivatives B. Brożek-Płuska*, I. Szymczyk, H. Abramczyk

More information

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES Chemistry 524--Final Exam--Keiderling May 4, 2011 3:30 -?? pm -- 4286 SES Please answer all questions in the answer book provided. Calculators, rulers, pens and pencils are permitted. No open books or

More information

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive Chapter 9: End-of-Chapter Solutions 1. The following comparison provides general trends, but both atomic absorption spectroscopy (AAS) and atomic absorption spectroscopy (AES) will have analyte-specific

More information

Atomic Spectra. d sin θ = mλ (1)

Atomic Spectra. d sin θ = mλ (1) Atomic Spectra Objectives: To measure the wavelengths of visible light emitted by atomic hydrogen and verify that the measured wavelengths obey the empirical Rydberg formula. To observe emission spectra

More information

Application of Raman Spectroscopy for Detection of Aflatoxins and Fumonisins in Ground Maize Samples

Application of Raman Spectroscopy for Detection of Aflatoxins and Fumonisins in Ground Maize Samples Application of Raman Spectroscopy for Detection of Aflatoxins and Fumonisins in Ground Maize Samples Kyung-Min Lee and Timothy J. Herrman Office of the Texas State Chemist, Texas A&M AgriLife Research

More information

Schedule for a proficiency test

Schedule for a proficiency test Schedule for a proficiency test Once a laboratory haegistered the year s programme the interaction between the laboratory and Eurofins Miljø A/S will be according to the schedule described below. 1 Reminder

More information

Chapter 15 Molecular Luminescence Spectrometry

Chapter 15 Molecular Luminescence Spectrometry Chapter 15 Molecular Luminescence Spectrometry Two types of Luminescence methods are: 1) Photoluminescence, Light is directed onto a sample, where it is absorbed and imparts excess energy into the material

More information

Inferential Analysis with NIR and Chemometrics

Inferential Analysis with NIR and Chemometrics Inferential Analysis with NIR and Chemometrics Santanu Talukdar Manager, Engineering Services Part 2 NIR Spectroscopic Data with Chemometrics A Tutorial Presentation Part 2 Page.2 References This tutorial

More information

2p or not 2p: Tuppence-based SERS for the detection of illicit materials

2p or not 2p: Tuppence-based SERS for the detection of illicit materials SUPPLEMENTARY INFORMATION 2p or not 2p: Tuppence-based SERS for the detection of illicit materials Figure S1. Deposition of silver (Grey target) demonstrated on a post-1992 2p coin. Figure S2. Raman spectrum

More information

Automated Unmixing of Comprehensive Two-Dimensional Chemical Separations with Mass Spectrometry. 1 Introduction. 2 System modeling

Automated Unmixing of Comprehensive Two-Dimensional Chemical Separations with Mass Spectrometry. 1 Introduction. 2 System modeling Automated Unmixing of Comprehensive Two-Dimensional Chemical Separations with Mass Spectrometry Min Chen Stephen E. Reichenbach Jiazheng Shi Computer Science and Engineering Department University of Nebraska

More information

Consistent Histories. Chapter Chain Operators and Weights

Consistent Histories. Chapter Chain Operators and Weights Chapter 10 Consistent Histories 10.1 Chain Operators and Weights The previous chapter showed how the Born rule can be used to assign probabilities to a sample space of histories based upon an initial state

More information

METHOD 3600B CLEANUP

METHOD 3600B CLEANUP METHOD 3600B CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Method 3600 provides general guidance on selection of cleanup methods that are appropriate for the target analytes of interest. Cleanup methods are applied

More information

Linear Least-Squares Data Fitting

Linear Least-Squares Data Fitting CHAPTER 6 Linear Least-Squares Data Fitting 61 Introduction Recall that in chapter 3 we were discussing linear systems of equations, written in shorthand in the form Ax = b In chapter 3, we just considered

More information

IDENTIFICATION AND QUANTITATION OF HARD-TO-RESOLVE INDIVIDUAL COMPONENTS IN MIXTURES WITHOUT CHROMATOGRAPHY, CHEMOMETRICS, AND REFERENCE STANDARDS

IDENTIFICATION AND QUANTITATION OF HARD-TO-RESOLVE INDIVIDUAL COMPONENTS IN MIXTURES WITHOUT CHROMATOGRAPHY, CHEMOMETRICS, AND REFERENCE STANDARDS WHITE PAPER IDENTIFICATION AND QUANTITATION OF HARD-TO-RESOLVE INDIVIDUAL COMPONENTS IN MIXTURES WITHOUT CHROMATOGRAPHY, CHEMOMETRICS, AND REFERENCE STANDARDS Aleksandr V. Mikhonin, Matthew T. Muckle,

More information

An indicator for the number of clusters using a linear map to simplex structure

An indicator for the number of clusters using a linear map to simplex structure An indicator for the number of clusters using a linear map to simplex structure Marcus Weber, Wasinee Rungsarityotin, and Alexander Schliep Zuse Institute Berlin ZIB Takustraße 7, D-495 Berlin, Germany

More information

Quantum Information Types

Quantum Information Types qitd181 Quantum Information Types Robert B. Griffiths Version of 6 February 2012 References: R. B. Griffiths, Types of Quantum Information, Phys. Rev. A 76 (2007) 062320; arxiv:0707.3752 Contents 1 Introduction

More information

Semi-Quantitative Analysis of Analytical Data using Chemometric Methods. Part II.

Semi-Quantitative Analysis of Analytical Data using Chemometric Methods. Part II. Semi-Quantitative Analysis of Analytical Data using Chemometric Methods. Part II. Simon Bates, Ph.D. After working through the various identification and matching methods, we are finally at the point where

More information

Spectroscopy II Introduction: Spectrophotometry and Fluorometry

Spectroscopy II Introduction: Spectrophotometry and Fluorometry Introduction: Spectrophotometry and Fluorometry In this experiment you will use the SpectroVis Plus (a small computer controlled Spectrophotometer + Fluorometer) to: Measure the Transmission and Absorbance

More information

Estimation of IMFP except for a constant factor using only XPS background-optimized peak intensities and cross sections

Estimation of IMFP except for a constant factor using only XPS background-optimized peak intensities and cross sections Paper Estimation of IMFP except for a constant factor using only XPS background-optimized peak intensities Masatoshi Jo,* National Metrology Institute of Japan, AIST AIST Tsukuba Central 5, 1-1-1 Higashi,

More information

Dimitri Nion & Lieven De Lathauwer

Dimitri Nion & Lieven De Lathauwer he decomposition of a third-order tensor in block-terms of rank-l,l, Model, lgorithms, Uniqueness, Estimation of and L Dimitri Nion & Lieven De Lathauwer.U. Leuven, ortrijk campus, Belgium E-mails: Dimitri.Nion@kuleuven-kortrijk.be

More information

PLS. theoretical results for the chemometrics use of PLS. Liliana Forzani. joint work with R. Dennis Cook

PLS. theoretical results for the chemometrics use of PLS. Liliana Forzani. joint work with R. Dennis Cook PLS theoretical results for the chemometrics use of PLS Liliana Forzani Facultad de Ingeniería Química, UNL, Argentina joint work with R. Dennis Cook Example in chemometrics A concrete situation could

More information

Designing Information Devices and Systems I Fall 2018 Lecture Notes Note Introduction to Linear Algebra the EECS Way

Designing Information Devices and Systems I Fall 2018 Lecture Notes Note Introduction to Linear Algebra the EECS Way EECS 16A Designing Information Devices and Systems I Fall 018 Lecture Notes Note 1 1.1 Introduction to Linear Algebra the EECS Way In this note, we will teach the basics of linear algebra and relate it

More information

IntelliCal FAQ Revised 9/27/12

IntelliCal FAQ Revised 9/27/12 IntelliCal FAQ Revised 9/27/12 IntelliCal Intelligent spectral calibration routine from Princeton Instruments, available with 64-bit LightField data acquisition software. Q. What is IntelliCal? Figure

More information

Vector Space Models. wine_spectral.r

Vector Space Models. wine_spectral.r Vector Space Models 137 wine_spectral.r Latent Semantic Analysis Problem with words Even a small vocabulary as in wine example is challenging LSA Reduce number of columns of DTM by principal components

More information

Measuring Colocalization within Fluorescence Microscopy Images

Measuring Colocalization within Fluorescence Microscopy Images from photonics.com: 03/01/2007 http://www.photonics.com/article.aspx?aid=39341 Measuring Colocalization within Fluorescence Microscopy Images Two-color fluorescence-based methods are uncovering molecular

More information

Journal of Luminescence

Journal of Luminescence Journal of Luminescence 142 (2013) 103 109 Contents lists available at SciVerse ScienceDirect Journal of Luminescence journal homepage: www.elsevier.com/locate/jlumin Study of the ph influence on the optical

More information

Moving towards a technical specification for fluorescence excitation-emission mapping and absorbance analysis of colored dissolved organic matter

Moving towards a technical specification for fluorescence excitation-emission mapping and absorbance analysis of colored dissolved organic matter Design and Nature VI 295 Moving towards a technical specification for fluorescence excitation-emission mapping and absorbance analysis of colored dissolved organic matter A. Gilmore1, R. Hurteaux2, S.

More information

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin Elemental Scientific prepfast IC: Inline Autodilution Method to Eliminate Species Interconversion for LC-ICPMS Based Applications Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan

More information

Reprinted from February Hydrocarbon

Reprinted from February Hydrocarbon February2012 When speed matters Loek van Eijck, Yokogawa, The Netherlands, questions whether rapid analysis of gases and liquids can be better achieved through use of a gas chromatograph or near infrared

More information

Designing Information Devices and Systems I Spring 2018 Lecture Notes Note Introduction to Linear Algebra the EECS Way

Designing Information Devices and Systems I Spring 2018 Lecture Notes Note Introduction to Linear Algebra the EECS Way EECS 16A Designing Information Devices and Systems I Spring 018 Lecture Notes Note 1 1.1 Introduction to Linear Algebra the EECS Way In this note, we will teach the basics of linear algebra and relate

More information

GS Analysis of Microarray Data

GS Analysis of Microarray Data GS01 0163 Analysis of Microarray Data Keith Baggerly and Bradley Broom Department of Bioinformatics and Computational Biology UT M. D. Anderson Cancer Center kabagg@mdanderson.org bmbroom@mdanderson.org

More information

ECE521 week 3: 23/26 January 2017

ECE521 week 3: 23/26 January 2017 ECE521 week 3: 23/26 January 2017 Outline Probabilistic interpretation of linear regression - Maximum likelihood estimation (MLE) - Maximum a posteriori (MAP) estimation Bias-variance trade-off Linear

More information

Spectroscopy Problem Set February 22, 2018

Spectroscopy Problem Set February 22, 2018 Spectroscopy Problem Set February, 018 4 3 5 1 6 7 8 1. In the diagram above which of the following represent vibrational relaxations? 1. Which of the following represent an absorbance? 3. Which of following

More information

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction II. Quality Control A. Introduction ENCO adheres to strict quality control practices in order to assure our clients that the data provided are accurate and reliable. We are required by the EPA to analyze

More information

Learning Outcomes Framework

Learning Outcomes Framework Learning Outcomes Framework May 2004 Mathematics Grades 7 9 Learning Outcomes Framework Mathematics Grades 7 9 GRADE 7 Grade 7 GCO A: Students will demonstrate number sense and apply number-theory concepts.

More information

APPLICATION OF INDEPENDENT COMPONENT ANALYSIS TO CHEMICAL REACTIONS. S.Triadaphillou, A. J. Morris and E. B. Martin

APPLICATION OF INDEPENDENT COMPONENT ANALYSIS TO CHEMICAL REACTIONS. S.Triadaphillou, A. J. Morris and E. B. Martin APPLICAION OF INDEPENDEN COMPONEN ANALYSIS O CHEMICAL REACIONS S.riadaphillou, A. J. Morris and E. B. Martin Centre for Process Analytics and Control echnology School of Chemical Engineering and Advanced

More information

Third-Order Tensor Decompositions and Their Application in Quantum Chemistry

Third-Order Tensor Decompositions and Their Application in Quantum Chemistry Third-Order Tensor Decompositions and Their Application in Quantum Chemistry Tyler Ueltschi University of Puget SoundTacoma, Washington, USA tueltschi@pugetsound.edu April 14, 2014 1 Introduction A tensor

More information

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection Role of Validation Method Validation An overview Confirms the fitness for purpose of a particular analytical method. ISO definition: Conformation by examination and provision of objective evidence that

More information

Exploratory Factor Analysis and Principal Component Analysis

Exploratory Factor Analysis and Principal Component Analysis Exploratory Factor Analysis and Principal Component Analysis Today s Topics: What are EFA and PCA for? Planning a factor analytic study Analysis steps: Extraction methods How many factors Rotation and

More information

Geog 210C Spring 2011 Lab 6. Geostatistics in ArcMap

Geog 210C Spring 2011 Lab 6. Geostatistics in ArcMap Geog 210C Spring 2011 Lab 6. Geostatistics in ArcMap Overview In this lab you will think critically about the functionality of spatial interpolation, improve your kriging skills, and learn how to use several

More information

GS Analysis of Microarray Data

GS Analysis of Microarray Data GS01 0163 Analysis of Microarray Data Keith Baggerly and Kevin Coombes Section of Bioinformatics Department of Biostatistics and Applied Mathematics UT M. D. Anderson Cancer Center kabagg@mdanderson.org

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

Lab 11: Must what goes in be the same as what comes out? Spectroscopy & Fluorescence in Chlorophyll.

Lab 11: Must what goes in be the same as what comes out? Spectroscopy & Fluorescence in Chlorophyll. Lab 11: Must what goes in be the same as what comes out? Spectroscopy & Fluorescence in Chlorophyll. Introduction to Fluorescence: Fluorescence is one of the possible mechanisms for emission of light by

More information

GS Analysis of Microarray Data

GS Analysis of Microarray Data GS01 0163 Analysis of Microarray Data Keith Baggerly and Kevin Coombes Section of Bioinformatics Department of Biostatistics and Applied Mathematics UT M. D. Anderson Cancer Center kabagg@mdanderson.org

More information

Advancing from unsupervised, single variable-based to supervised, multivariate-based methods: A challenge for qualitative analysis

Advancing from unsupervised, single variable-based to supervised, multivariate-based methods: A challenge for qualitative analysis Advancing from unsupervised, single variable-based to supervised, multivariate-based methods: A challenge for qualitative analysis Bernhard Lendl, Bo Karlberg This article reviews and describes the open

More information

Linear Least Squares Problems

Linear Least Squares Problems Linear Least Squares Problems Introduction We have N data points (x 1,y 1 ),...(x N,y N ). We assume that the data values are given by y j = g(x j ) + e j, j = 1,...,N where g(x) = c 1 g 1 (x) + + c n

More information

Exploratory Factor Analysis and Principal Component Analysis

Exploratory Factor Analysis and Principal Component Analysis Exploratory Factor Analysis and Principal Component Analysis Today s Topics: What are EFA and PCA for? Planning a factor analytic study Analysis steps: Extraction methods How many factors Rotation and

More information

Introduction to Machine Learning

Introduction to Machine Learning 10-701 Introduction to Machine Learning PCA Slides based on 18-661 Fall 2018 PCA Raw data can be Complex, High-dimensional To understand a phenomenon we measure various related quantities If we knew what

More information

Physics 476LW Advanced Physics Laboratory Atomic Spectroscopy

Physics 476LW Advanced Physics Laboratory Atomic Spectroscopy Physics 476LW Atomic Spectroscopy 1 Introduction The description of atomic spectra and the Rutherford-Geiger-Marsden experiment were the most significant precursors of the so-called Bohr planetary model

More information