SAXS and SANS facilities and experimental practice. Clement Blanchet

Similar documents
Small Angle X-Ray Solution Scattering of Biological Macromolecules

Chapter 10: Wave Properties of Particles

How Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source

Neutron Imaging at Spallation Neutron Sources

Exercise 1 Atomic line spectra 1/9

Unit title: Atomic and Nuclear Physics for Spectroscopic Applications

Light Source I. Takashi TANAKA (RIKEN SPring-8 Center) Cheiron 2012: Light Source I

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

The basics of structural biology. And Why we use synchrotron sources Sean McSweeney ESRF Structural Biology Group

Data Acquisition. What choices need to be made?

Neutron facilities and generation. Rob McQueeney, Ames Laboratory and Iowa State University

X-Ray Damage to Biological Crystalline Samples

PHYS Introduction to Synchrotron Radiation

Static and dynamic light scattering. Cy Jeffries EMBL Hamburg

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Physics 126 Practice Exam #4 Professor Siegel

Chapter 1 X-ray Absorption Fine Structure (EXAFS)

A2. Light Source. ( i ) Units of light intensity

Introduction to XAFS. Grant Bunker Associate Professor, Physics Illinois Institute of Technology. Revised 4/11/97

Synchrotron radiation: A charged particle constrained to move in curved path experiences a centripetal acceleration. Due to it, the particle radiates

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

Which of the following can be used to calculate the resistive force acting on the brick? D (Total for Question = 1 mark)

Collecting Macromolecular Crystallographic Data at Synchrotrons. Andrew Howard ACA Summer School 12 July 2007

How to judge data quality

Highenergy Nuclear Optics of Polarized Particles

The SAXS Guide. Getting acquainted with the principles. New Edition with special contributions

Nuclear Physics. (PHY-231) Dr C. M. Cormack. Nuclear Physics This Lecture

Biological Small Angle X-ray Scattering (SAXS) Dec 2, 2013

Neutron and x-ray spectroscopy

Data reduction and processing tutorial

Insertion Devices Lecture 2 Wigglers and Undulators. Jim Clarke ASTeC Daresbury Laboratory

Synchrotron Methods in Nanomaterials Research

Particle Detectors and Quantum Physics (2) Stefan Westerhoff Columbia University NYSPT Summer Institute 2002

2. X-ray Sources 2.1 Electron Impact X-ray Sources - Types of X-ray Source - Bremsstrahlung Emission - Characteristic Emission

Episode 506: Particles as waves

Synchrotron radiation: A charged particle constrained to move in curved path experiences a centripetal acceleration. Due to this acceleration, the

Research with Synchrotron Radiation. Part I

Characteristics and Properties of Synchrotron Radiation

Basics of Synchrotron Radiation Beamlines and Detectors. Basics of synchrotron radiation X-ray optics as they apply to EXAFS experiments Detectors

Shielding calculations for the design of new Beamlines at ALBA Synchrotron

Emphasis on what happens to emitted particle (if no nuclear reaction and MEDIUM (i.e., atomic effects)

Quantum and Atomic Physics - Multiple Choice

Scattering Lecture. February 24, 2014

Introduction to SAXS at SSRL

High Brilliance SAXS on Synchrotrons

Introduction to Accelerator Physics Part 1

LECTURE 4 PRINCIPLE OF IMAGE FORMATION KAMARUL AMIN BIN ABDULLAH

In a particular investigation the atomic spacing of the crystal is m and the electrons are accelerated through 3000 V.

CHAPTER 12 TEST REVIEW

Introduction to Dynamic Light Scattering with Applications. Onofrio Annunziata Department of Chemistry Texas Christian University Fort Worth, TX, USA

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

Examples for experiments that can be done at the T9 beam line

DUAL NATURE OF RADIATION AND MATTER

Small Angle Neutron Scattering in Different Fields of Research. Henrich Frielinghaus

X-ray Fluorescence Imaging Following Synchrotron Beam Excitation

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011

The wavefunction ψ for an electron confined to move within a box of linear size L = m, is a standing wave as shown.

INTRODUCTION Strained Silicon Monochromator Magnesium Housing Windows for Monochromator Shutter and Collimator Fission Detector HOPG Monochromator

China high-intensity accelerator technology developments for Neutron Sources & ADS

Small-Angle X-ray Scattering (SAXS)/X-ray Absorption Near Edge Spectroscopy (XANES).

Complementary use of SAXS and SANS. Jill Trewhella University of Sydney

Quantitative determination of the effect of the harmonic component in. monochromatised synchrotron X-ray beam experiments

λ = h = h p mv λ = h mv FXA 2008 Candidates should be able to :

Small-Angle X-ray Scattering (SAXS) SPring-8/JASRI Naoto Yagi

Electromagnetic spectra

GAMMA RAY OPTICS. Michael Jentschel Institut Laue-Langevin, Grenoble, France

tip conducting surface

Optical Systems Program of Studies Version 1.0 April 2012

PB I FEL Gas-Monitor Detectors for FEL Online Photon Beam Diagnostics BESSY

Selected Topics from Modern Physics

Name: Class: Date: Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Introduction to Accelerator Physics Part 1

PHYS Introduction to Synchrotron Radiation

Undulator Commissioning Spectrometer for the European XFEL

PARTICLE BEAMS, TOOLS FOR MODERN SCIENCE AND MEDICINE Hans-H. Braun, CERN

U n 3 n Ba Kr (D) Br (C) Kr (B) Rb (E) 94 37

Chapter Two. Energy Bands and Effective Mass

The Final Exam (Exam 4) will be on FRIDAY MAY 11 From 3 5 PM in LR1 VAN

Cambridge International Examinations Cambridge International Advanced Subsidiary and Advanced Level

Radiation Protection At Synchrotron Radiation Facilities

PHYS 4 CONCEPT PACKET Complete

Determination of Absolute Neutron Fluence to sub-0.1% uncertainty (and better)

3. Synchrotrons. Synchrotron Basics

A new method to suppress high-order harmonics for. synchrotron radiation soft x-ray beamline*

de Broglie Waves h p de Broglie argued Light exhibits both wave and particle properties

1 The Cathode Rays experiment is associated. with: Millikan A B. Thomson. Townsend. Plank Compton

2) Measure sample and empties. 3) Measure transmissions of sample and empties. 4) Normalize to exposure time and transmission. 5) Subtract the empties

EEE4106Z Radiation Interactions & Detection

WAVES AND PARTICLES. (c)

Neutron Physics. The De Broglie relation for matter is: h mv (1) λ =

Free-electron laser SACLA and its basic. Yuji Otake, on behalf of the members of XFEL R&D division RIKEN SPring-8 Center

Physical Science DCI Progression Chart

PHY492: Nuclear & Particle Physics. Lecture 25. Particle Detectors

Rb, which had been compressed to a density of 1013

Radiation and the Atom

Physics with Neutrons I, WS 2015/2016. Lecture 11, MLZ is a cooperation between:

A) n L < 1.0 B) n L > 1.1 C) n L > 1.3 D) n L < 1.1 E) n L < 1.3

3.17 Strukturanalyse mit Röntgenstrahlen nach Debye- Scherrer

The University of Hong Kong Department of Physics

Transcription:

SAXS and SANS facilities and experimental practice Clement Blanchet

SAS experiment Detector X-ray or neutron Beam Sample 2 s Buffer

X-rays Roengten, 1895

Electromagnetic wave

The electromagnetic spectrum

Why these wavelength? λ<<dimension of the object Transmission λ dimension of the object Diffraction λ >object, object invisible

Neutrons Matter wave (De Broglie, 1924) =h/mv

SAXS and SANS Common analysis methods but: The scattered particles are different Different interactions X-rays interact with electrons via electromagnetic forces Neutrons interact with nucleus via nuclear forces Different instruments

Outlines X-rays and neutrons sources SAXS and SANS instruments Optics Sample area Detectors Sample requirements and collection strategy

X-rays and neutron sources

X-rays: how are they produced? Principle Maxwell equation: accelerated charge -> radiation

X-ray sources Synchrotrons

X-ray sources Synchrotron radiation

Insertion devices Dipole bending magnet (APS) Undulator (PetraIII)

Synchrotrons around the world

X-ray sources

Lab sources Principle : electron beam send on a target Brehmstrahlung Fluorescence

X-ray sources Lab source (rotating anode, liquid jet)

Neutron production Nuclear reactor

Neutron sources Spallation source Accelerated protons hit a target.

Neutron Sources

SAXS and SANS Instruments

Optics Prepare the beam coming from the source Monochromatic beam Focus/collimated beam

Monochromatic X-ray Bragg diffraction on a crystal n = 2dsin

Monochromator Before After Polychromatic One wavelength + harmonics

Focusing/low divergence 2 Small beam at the detector position Small beam at the sample position

Focusing X-ray Focussing mirror Reflectivity 1,0 0,8 Transmission 0,6 0,4 0.15 Degree 0.25 Degree 1 Degree 0,2 0,0 10000 Energy [ev]

Focussing mirror harmonics filter

Monochromatic neutrons De Broglie equation λ=h/mv The wavelength of a neutron is related to its velocity. Velocity selector

Collimation neutrons A collimator is used to obtain a parallel beam

Sample environment As many sample environment as there is sample For biological macromolecules in solution: Liquid containing cell Preferably in vacuum thermostated

Sample cell Cell material, low absorption and low scattering Mica, polycarbonate Cell thickness: compromise between absorption and scattering

Sample environment On dedicated beamline, Sample handling is now automated: Faster measurement Better cleaning Unattended operation

Flight tube

Beamstop Prevent the direct beam from hitting the detector Big enough to stop the direct beam Small enough to collect the small angle Measure transmitted beam

Detectors

Single photon counting detector principle

Single photon counting detector Pilatus High dynamic range No background noise Fast framing Ideal for SAXS

Neutron detection He3 detector: n + 3 He 3 H + 1 H + 0.764 MeV

Experimental practice

Experiment SAS applicable to many type of samples. Biological macromolecules in solution Isotropic scattering Weakly scattering

SAS Experiment

Buffer subtraction Biological sample scatters very weakly Care should be taken for the buffer subtraction Exactly matching buffer (dialysis, elution buffer) Sample and buffer measured in the same cell

Monodispersity SAS is very sensible to aggregation, the sample should be monodisperse

Monodispersity Check the monodispersity of your sample before coming to the beamline. (native gel, dynamic light scattering, ultracentrifugation, )

Monodispersity Improving monodispersity: online size exclusion column

SEC + SAXS Defined buffer region

Inter-particle interactions

Inter-particle interactions Change solution (ph, salt concentration) to limit interactions Measure different concentrations and extrapolate

Measure also water and/or standard protein to estimate the molecular mass of your sample using the forward scattering For data on an absolute scale (water measurement) M=I(0)*N A /(C* Using a protein standard M=M BSA *I(0)/I BSA (0)

X-rays - Radiation damage!!! With intense third generation synchrotons: creation of free radicals in solution, that degrades protein and causes aggregation. Monitor radiation damage: collect several frames and compare them. Limit the radiation damage Use of scavengers DTT, Glycerol. Flow measurements.

Contrast in neutron Neutrons interact with the nucleus of atoms Each atoms has its own scattering length: H D C N O P S -.3742 0.6671 0.6651 0.940 0.5804 0.517 0.2847

Labeling Deuteration of sample Expression of the protein in deuterated medium

Collection strategy Solvent matching Find the solvent composition that match the contrast of the component you want to hide Measure the sample in this solvent Contrast variation Measure the sample in solvent with different contrast Using Stuhrmann analysis you can access the curves of the different components.

Conclusion - Instruments SAXS and SANS facilities: great high tech instruments. Apply for beamtime (dedicated biosaxs beamline in PetraIII/EMBL, ESRF, Soleil, SSRL, Australian synchrotron, spring8, diamond) Lab source also available for SAXS

SAS sample Protein concentration: 1-10 mg/ml Volume: 5-50 microliter (SAXS), 200-300 microliter (SANS) Time: lab source: 5-60 min Synchrotron: seconds Neutrons: 30 minutes - hours

SAS sample Pure and monodisperse sample Exactly matching buffer Measure concentration series For SAXS: Be aware of radiation damage For SANS: Carefully design your experiment, think of your collection strategy.