Address correspondence to M.C. Allgeier. HPLC effluent. Nebulizer. Nitrogen. Heated drift tube. Data acquisition. Analyte.

Similar documents
The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Evaporative Light Scattering Detector (ELSD)

Journal of Chromatography A, 1056 (2004) 67 75

HPLC Analysis of Nonvolatile Analytes Using Charged Aerosol Detection

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

HPLC Analysis of Non-volatile Analytes Using

Agilent 385-ELSD Evaporative Light Scattering Detector

Analysis of Weak UV-Absorbing Pharmaceutical Drug Substances Using CE with Condensation Nucleation Light-Scattering Detection

Evaluation of the chemiluminescent nitrogen detector for solubility determinations to support drug discovery

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Agilent 385-ELSD Evaporative Light Scattering Detector

Selecting Detectors for Compounds with No Optical Absorbance

Gas Chromatography. Introduction

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

with diode array detection

Luminescence transitions. Fluorescence spectroscopy

Varian 385-LC. ELSD for Low Molecular Weight Compounds

Agilent 1260 Infinity Evaporative Light Scattering Detector (ELSD)

Varian 380-LC. ELSD for Routine HPLC and GPC Analysis. Key Benefits

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Final Exam. Physical Constants and Conversion Factors. Equations

Application Note S/SL

Emission spectrum of H

Cork Institute of Technology. Summer 2005 Instrumental Analysis (Time: 3 Hours) Section A

Analytical techniques: Environmental samples. Lecture 2 Universidade do Algarve

HPLC Praktikum Skript

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

An Investigation into Detector Limitations Using Evaporative Light-Scattering Detectors For Pharmaceutical Applications

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Determination of Caffeine by HPLC

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

The Theory of HPLC. Quantitative and Qualitative HPLC

Comparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

3 - Atomic Absorption Spectroscopy

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

A Novel LC-ELSD Method for the Quantification of 2-Deoxy D-Glucose Using HILIC Chromatography

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

Evaluation of Ketone Containing Terpene Compounds using Evaporative Light Scattering Detection

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Better HPLC Methods Using Temperature Programming

2101 Atomic Spectroscopy

Far UV Absorbance Detector

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

Chromatography. Gas Chromatography

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Lecture 15: Introduction to mass spectrometry-i

LC and LC/MS Column Selection Flow Chart

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Mobile-Phase Cleanup Using Solid-Phase Extraction Disks

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

Expediting Achiral SFC Method Development Using a Multi-Channel SFC System with MS Detection

CHAPTER V ANALYTICAL METHODS ESTIMATION OF DICLOFENAC. Diclofenac (gift sample from M/s Micro Labs Ltd., Pondicherry)

Chemistry Instrumental Analysis Lecture 27. Chem 4631

EFAVIRENZ Final text for addition to The International Pharmacopoeia

An Investigation of the Impact of Common Experimental Parameters on Signal Intensity in SFC ESI MS

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

How to Troubleshoot a Failed System Suitability Test

ISSN X CODEN (USA): PCHHAX. Determination of captopril with potassium permanganate chemiluminescence system*

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Enhancement of Linearity and Response in Charged Aerosol Detection

[application note] ACQUITY UPLC/SQD ANALYSIS OF POLYMER ADDITIVES. Peter J. Lee, and Alice J. Di Gioia, Waters Corporation, Milford, MA, U.S.A.

for Acclaim Mixed-Mode HILIC-1 Column

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Instrumentation of HPLC. Detectors

NOMENCLATURE FOR CHROMATOGRAPHY NOMENCLATURE FOR SUPERCRITICAL FLUID CHROMATOGRAPHY AND EXTRACTION

Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST)

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

High Pressure/Performance Liquid Chromatography (HPLC)

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Transcription:

376 LCGC ORT AMERICA VOLUME 21 UMBER 4 APRIL 2003 www.chromatographyonline.com Comparison of an Evaporative Light-Scattering Detector and a Chemiluminescent itrogen Detector for Analyzing Compounds Lacking a Sufficient UV Chromophore The authors evaluated an evaporative light-scattering detector and a chemiluminescent nitrogen detector for high performance liquid chromatography analyses of compounds that lack a sufficient ultraviolet (UV) chromophore. They compared the detectors in three critical areas of performance: precision, linearity, and limit of detection. The four compounds used as test analytes included caffeine and three pharmaceutical compounds with weak UV chromophores. Based on these experiments, the authors discuss the advantages and disadvantages of these detectors. Matthew C. Allgeier, Mark A. ussbaum, and Donald S. Risley Eli Lilly and Co., Lilly Research Laboratories, Lilly Corporate Center, Indianapolis, Indiana 46285 Address correspondence to M.C. Allgeier. Ultraviolet (UV) detection has been the preferred detection technique for high performance liquid chromatography (PLC) analyses of pharmaceutical compounds because of its ease of use, dynamic linear range, precision, and high sensitivity. owever, one shortcoming of UV detection is the inability to detect compounds that lack UV chromophores. Chemists can use various derivatization methods to enhance UV absorption, but these manipulations tend to be tedious and introduce their own errors. To avoid the derivatization of these compounds, analysts need alternative detection schemes that are independent of radiative absorption by the compound yet still provide adequate linearity, precision, limit of detection, and ruggedness. Evaporative light-scattering detection (ELSD) and chemiluminescent nitrogen detection (CLD) are two commercially available detection techniques for PLC, both of which operate independently of radiative absorption. ELSD has been demonstrated successfully as an alternative to UV detection for weakly UV-absorbing pharmaceutical compounds within our laboratory (1 4). In ELSD, as depicted in Figure 1, the PLC effluent is nebulized and then vaporized in a heated drift tube, which results in a cloud of analyte particles that pass through a beam of light. The analyte particles scatter the light and generate a signal at a photodiode (or photomultiplier tube, depending upon the manu- ebulizer Analyte Laser light source PLC effluent Lightscattering cell eated drift tube itrogen Photodiode Data acquisition Amplifier Figure 1: Schematic design of an evaporative light-scattering detector.

378 LCGC ORT AMERICA VOLUME 21 UMBER 4 APRIL 2003 www.chromatographyonline.com Figure 2: O 2 C Figure 3: ebulizer Effluent O from 2 Ar PLC system Data acquisition A 2 CO 2 R O 2 PO 3 2 Structures of the test compounds. O 3 Reaction chamber Photomultiplier tube C luminescent nitrogen detector for PLC first was described in 1988 and subsequently refined and described in its current incarnation in 1992 (9,10). Since its introduction, CLD has been used successfully for PLC analyses of small organic molecules and peptides (11,12). Figure 2 depicts the principles of CLD. Oxidation of the nebulized PLC effluent by combustion in a hightemperature furnace converts all nitrogencontaining compounds, with the exception of 2, into nitric oxide. The vaporized water in the effluent is then removed in a membrane dryer. The dried gas is passed into a chamber where it reacts with ozone, which Furnace (1050 C) O 2 O other oxides Pyrotube Vacuum *O 2 O 3 O B Photon Schematic design of a chemiluminescent nitrogen detector. CO 2 Residual water O 2 CO 2 3 C O eedle valve O C 3 Caffeine Waste Membrane dryer C 3 facturer) that is proportional to the amount of light scattered. The amount of light scattered is dependent upon the particle size and quantity, which in turn are functions of the analyte, analyte concentration, nebulizer characteristics, mobile-phase flow rate, mobile-phase composition, nebulizing-gas flow rate, and drift-tube temperature (5 8). ELSD is limited to the use of volatile mobile-phase components and the detection of relatively nonvolatile analytes. In cases in which the weakly UVabsorbing compounds contain at least one nitrogen atom, CLD also can be used as an alternative to UV detection. The chemicauses the conversion of the nitric oxide in the dried gas to excited-state nitrogen dioxide, which subsequently produces photon emission upon relaxation. Thus, the signal generated is proportional to the moles of nitrogen present in a given molecule. Because of the nature of the detection mechanism, CLD is limited to detection of analytes that contain nitrogen. Conversely, the CLD detection mechanism also imposes another limitation all mobile-phase components must be free of nitrogen to keep baseline noise to a minimum. Common PLC organic modifiers that contain nitrogen, such as acetonitrile, thus are unsuitable. The volatility of the sample is of no concern with the chemiluminescent nitrogen detector, as is the case with the evaporative lightscattering detector, because the sample is completely oxidized in the chemiluminescent nitrogen detector chamber. As with ELSD, CLD is limited to analyzing volatile mobile-phase components. We compared the analytical capabilities of evaporative light-scattering and chemiluminescent nitrogen detectors using caffeine and three additional pharmaceutical compounds as test analytes that lacked sufficient UV chromophores (Figure 3) and provided the desired variation in the number and bonding of nitrogen atoms in the molecule. We will report the precision, linearity, and limit of detection results for each detector and each compound based upon these experimental studies. Experimental The PLC system included a model 1050 series autosampler and pump (Agilent Technologies, Wilmington, Delaware), a model 500 evaporative light-scattering detector (Alltech Associates Inc., Deerfield, Illinois), and a model 8060 chemiluminescent nitrogen detector (Antek Instruments Inc., ouston, Texas). We obtained a 150 mm 2.1 mm Intersil C4 column from GL Sciences (Tokyo, Japan) and used it for analyses of the pharmaceutical compounds (see Figure 3). We obtained a 150 mm 2.1 mm Zorbax RX-C8 column from Agilent Technologies and used it for the caffeine analysis. Caffeine was purchased from Sigma (St. Louis, Missouri) and the three pharmaceutical compounds were synthesized at Eli Lilly and Co. (Indianapolis, Indiana). We chose isocratic mobile phases to eliminate possible effects of the gradient (for example, eluent viscosity changes) in ELSD nebulization. The mobile phase for the pharmaceutical compounds A and B was

www.chromatographyonline.com 99% water, 0.05% trifluoroacetic acid, and 1% methanol. The mobile phase used for compound C was 85% water, 0.1% trifluoroacetic acid, and 15% methanol. The mobile phase for caffeine was 60% water, 0.1% trifluoroacetic acid, and 40% methanol. The flow rate in all analyses was 0.2 ml/min with an injection volume of 10 L. The same mobile-phase preparation, column, and sample preparations were used for each detector and for each sample. The evaporative light-scattering detector settings were optimized for each mobile phase according to the manufacturer s recommendations (5). For compounds A, B, and C, the detector settings were as follows: 75 C drift-tube temperature, 3.1 standard liters per minute (SLPM) nitrogen gas flow, and 1 4 attenuation. For caffeine, the evaporative light-scattering detector conditions were as follows: 60 C drift-tube temperature, 2.4-SLPM nitrogen gas flow, and 1 8 attenuation. The chemiluminescent nitrogen detector settings for compounds A, B, and C were as follows: 10 C photomultiplier-tube temperature, 0.780-kV photomultiplier-tube voltage, high 10 sensitivity, 40-cm 3 /min argon makeup gas flow, 150-cm 3 /min inlet argon gas flow, 150-cm 3 /min inlet oxygen gas flow, 25-cm 3 /min ozone flow, 1050 C oven temperature, and 85 C membrane dryer temperature. The chemiluminescent nitrogen detector settings for caffeine were as follows: 0 C photomultiplier-tube temperature, 0.770-kV photomultiplier-tube voltage, high 1 sensitivity, 50-cm 3 /min makeup argon gas flow, 180-cm 3 /min inlet argon gas flow, 180-cm 3 /min inlet oxygen gas flow, 25-cm 3 /min ozone flow, 1050 C oven temperature, and 85 C membrane dryer temperature. Results and Discussion To evaluate the linearity of response for each detector, we prepared a stock solution in the range of 1 1.5 mg/ml for each compound. We performed serial dilutions of the stock until we found a concentration at the approximate limit of quantitation, a signalto-noise ratio of 5:1. We made duplicate injections of each concentration. The results of the linearity comparison are shown in Table I for the evaporative light-scattering detector and Table II for the chemiluminescent nitrogen detector. Figure 4 shows an example plot of the data for compound A. The data show that the chemiluminescent nitrogen detector exhibits a more linear response, with correlation coefficients closer to 1, y intercepts closer to 0, and a larger linear range compared with that obtained from ELSD. As expected, a nonlinear response in the ELSD data is evident for each compound, particularly at low concentrations: see Figure 4 (6,8). Unlike spectroscopic detectors, which operate within Beer s law (A bc), the ELSD response is described with the nonlinear equation A am x [1] where A is the peak area response, m is the sample quantity, and a and x are coefficients reflecting properties of the sample and the parameters of the chromatographic and detector systems. A log log plot of peak area APRIL 2003 LCGC ORT AMERICA VOLUME 21 UMBER 4 379 Table I: Linearity data for the test analytes analyzed using ELSD Concentration Range Response/ Correlation Compound Analyzed (mg/ml) mg Concentration y Intercept Coefficient A 0.0024 1.22 3.4 10 4 1020 0.9987 B 0.0033 1.16 3.1 10 4 1010 0.9984 C 0.0053 1.07 4.1 10 4 1460 0.9978 Caffeine 0.021 1.10 2.7 10 4 1990 0.9969 Table II: Linearity data for the test analytes analyzed using CLD Concentration Range Response/ Correlation Compound Analyzed (mg/ml) mg Concentration y Intercept Coefficient A 0.0012 1.22 2.6 10 4 184 0.9994 B 0.0009 1.66 1.6 10 4 199 0.9992 C 0.0008 1.07 4.2 10 4 104 0.9999 Caffeine 0.0008 0.44 1.0 10 5 256 0.9999

380 LCGC ORT AMERICA VOLUME 21 UMBER 4 APRIL 2003 www.chromatographyonline.com response versus sample quantity will be linear, which allows for quantitative results as described by the equation log A x log m log a [2] Likewise, all nonlinear curves have a region of linear response; if calibration standards and samples are within this linear range, a linear equation can be applied. The response of the chemiluminescent nitrogen detector ideally is equimolar with the amount of nitrogen in the sample. This response can be seen in the relative increase Response ( 10 4 ) 4.5 3.5 2.5 1.5 0.5 ELSD m 3.37 10 4, b 1020 R 2 0.9987 CLD m 2.60 10 4, b 184 R 2 0.9994 in slope for caffeine, which is caused by the large number of nitrogen atoms per molecule of caffeine (four nitrogens per molecule). owever, compound C, with five nitrogen atoms per molecule, did not show as large a slope as that demonstrated by caffeine. The likely reason for this is the presence of the R R bonding configuration in compound C. In this configuration, conversion to nitrogen oxide is less than optimal due to some conversion of the moiety to elemental nitrogen (13). Thus, the signal per mole of nitrogen for compound C is less than would be observed ELSD CLD Concentration (mg/ml) 0.00 0.07 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 Figure 4: The linearity results for compound A. The slope, intercept, and correlation coefficient of the linear fit also are shown in Tables I and II. The boxed region is an enlarged view of the lowconcentration range. Table III: Precision and limit of detection* results for test analytes using ELSD and CLD Compound Parameter ELSD CLD for an equal number of nitrogen atoms that were not in a R R configuration. To evaluate the precision of each detector, we determined the relative standard deviation (RSD) of the measured peak areas of 10 replicate injections (Table III). The results for the evaporative light-scattering detector generally were better than those obtained with the chemiluminescent nitrogen detector. The improved precision of the evaporative light-scattering detector is attributed partly to the superior peak shape obtained versus that of the chemiluminescent nitrogen detector (Figure 5). The superior peak shape might be attributed to the less complicated design and detection mechanism of the evaporative lightscattering detector versus the chemiluminescent nitrogen detector. The lower RSD for caffeine versus compounds A and B, when analyzed using the chemiluminescent nitrogen detector, resulted from the greater number of nitrogens in caffeine. This nitrogen content generates a larger signal and makes for more-precise peak integration. As expected for the chemiluminescent nitrogen detector, the RSD generally decreased as the number of nitrogens in a given molecule increased. We measured the limit of detection for the chemiluminescent nitrogen and evaporative light-scattering detectors by finding the lowest-concentration sample from serially diluted samples that yielded a peak roughly three times the baseline noise. The limit of detection is expressed as the mass of compound injected on the column, and, in the case of the chemiluminescent nitrogen detector, the mass of nitrogen injected on the column (Table III). The chemiluminescent nitrogen detector was determined to be more sensitive than the evaporative light-scattering detector by a factor of approximately four. The evaporative light-scattering detector demonstrated A Analyte (ng) 22 5.4 itrogen (ng) 0.4 RSD (%) 0.37 3.31 B Analyte (ng) 30 9.0 itrogen (ng) 0.4 RSD (%) 0.60 2.75 C Analyte (ng) 19 5.4 itrogen (ng) 1.3 RSD (%) 0.34 1.56 Caffeine Analyte (ng) 200 5.0 itrogen (ng) 1.5 RSD (%) 1.45 0.58 Response (mv) 180 124 68 12 44 100 (a) (b) 0 72 144 216 288 360 Time (s) * Limit of detection results are expressed as nanograms of analyte and nanograms of nitrogen injected on column and detected. Figure 5: Sample chromatograms of compound A as detected by (a) ELSD and (b) CLD. Conditions are described in the text.

www.chromatographyonline.com very poor sensitivity for caffeine. We used various instrument settings as well as methanol water and acetonitrile water mobile-phase systems to try to decrease the relatively large limit of detection obtained with the evaporative light-scattering detector for caffeine without success. We believe that solute properties affected this relative increase in limit of detection. Cebolla and co-workers (14) demonstrated inaccuracy in ELSD responses using only molecular weight and boiling points of polyaromatic compounds. Kildew and colleagues (15) illustrated the importance of the sample vapor pressure with variations in response factors for ELSD. Using an aerosol splitter for the evaporative light-scattering detector would have enabled lower drift tube temperature, which could have resulted in improved particle formation, increased sensitivity, and better sensitivity for caffeine. In general, ELSD response is proportional to concentration. owever, these results confirm some nonuniversal response characteristics of the evaporative light-scattering detector with some variation to the chemical structure and physical properties of the analyte; therefore, ELSD is not a true mass detector in certain situations (7). The limit of detection of compounds A and B for the chemiluminescent nitrogen detector agree well with other literature reports (11,16). The presence of the R R configuration in compound C likely explains the high limit of detection, as measured in nanograms of nitrogen, observed for compound C relative to compounds A and B (13). Caffeine was not run on our chemiluminescent nitrogen detector until after significant maintenance was performed on the detector, including rebuilding the nebulizer and replacing the pyrotube, membrane dryer, and ozone generator. These changes resulted in a higher limit of detection, an increase by nearly a factor of four, as measured by comparing data obtained before and after maintenance. If caffeine had been evaluated on the system before these changes, we expect that the limit of detection would have been comparable to the values obtained for compounds A and B. lower limit of detection than the evaporative light-scattering detector. The precision of the chemiluminescent nitrogen detector generally was poorer than that of the evaporative light-scattering detector, likely because of the more complicated mechanism of detection of the chemiluminescent nitrogen detector. The evaporative lightscattering detector required no maintenance during the course of these experiments, but the chemiluminescent nitrogen detector required significant maintenance and troubleshooting. The chemiluminescent nitrogen detector currently is much less rugged and not readily amenable for routine use, but it will become a powerful analytical tool when the reliability issues are resolved. owever, both detectors are extremely valuable analytical tools compared with UV detectors for analyzing compounds that lack chromophores. References (1) D.S. Risley and M.A. Strege, Anal. Chem. 72(8), 1736 1739 (2000). (2) D.S. Risley, K.F. ostettler, and J.A. Peterson, LCGC 16(6), 562 568 (1998). (3) A.G. Guisbert, V.S. Sharp, J.A. Peterson, and D.S. Risley, J. Liq. Chromatogr. Rel. Technol. 23(7), 1019 1028 (2000). APRIL 2003 LCGC ORT AMERICA VOLUME 21 UMBER 4 381 (4) D.S. Risley, J.A. Peterson, K.L. Griffiths, and S. McCarthy, LCGC 14(12), 1040 1047 (1996). (5) Alltech Model 500 ELSD Operating Manual, Alltech Associates, Inc. (Deerfield, Illinois, May 1997). (6) G. Guiochon, A. Moysan, and C. olley, J. Liq. Chromatogr. 11(12), 2547 2570 (1988). (7) M. Righezza and G. Guiochon, J. Liq. Chromatogr. 11(9-10), 1967 2004 (1988). (8) T.. Mourey and L.E. Oppenheimer, Anal. Chem. 56(13), 2427 2434 (1984). (9) A. Robbat Jr.,.P. Corso, and T.Y. Liu, Anal. Chem. 60, 173 174 (1988). (10) E.M. Fujinari and L.O. Courthaudon, J. Chromatogr. 592(1-2), 209 214 (1992). (11) E.W. Taylor, M.G. Qian, and G.D. Dollinger, Anal. Chem. 70, 3339 3347 (1998). (12) R. Bizanek, J.D. Manes, and E.M. Fujinari, Pep. Res. 9(1), 40 44 (1996). (13) M. oman, personal communication, 2000. (14) V.L. Cebolla, L. Membrado, J. Vela, and C.A. Ferrando, Semin. Food Anal. 2, 171 189 (1997). (15) B.R. Kildew, R. Saari-ordhaus, M.J. Wilcox, D.A. Young, Determining Sample Volatility for an Evaporative Light Scattering Detector, paper presented at the 50th Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy Pittcon 1999, Orlando, Florida, 7 12 March 1999. (16) D. Brannegan, M. Ashraf-Khorassani, and L.T. Taylor, J. Chromatogr. Sci. 39, 217 221 (2001). Conclusion In terms of performance, the chemiluminescent nitrogen detector provides a complementary tool to the evaporative lightscattering detector for detecting compounds that lack a sufficient chromophore. The chemiluminescent nitrogen detector clearly demonstrated a more linear response and