Nuclear Magnetic Resonance Imaging

Similar documents
Magnetic Resonance Imaging (MRI)

Magnetic Resonance Imaging. Pål Erik Goa Associate Professor in Medical Imaging Dept. of Physics

The NMR Inverse Imaging Problem

The Basics of Magnetic Resonance Imaging

Introduction to Biomedical Imaging

Fundamental MRI Principles Module 2 N. Nuclear Magnetic Resonance. X-ray. MRI Hydrogen Protons. Page 1. Electrons

Topics. The concept of spin Precession of magnetic spin Relaxation Bloch Equation. Bioengineering 280A Principles of Biomedical Imaging

Magnetic Resonance Imaging in a Nutshell

MRI Physics I: Spins, Excitation, Relaxation

A Hands on Introduction to NMR Lecture #1 Nuclear Spin and Magnetic Resonance

Fundamental MRI Principles Module Two

V27: RF Spectroscopy

The Theory of Nuclear Magnetic Resonance Behind Magnetic Resonance Imaging. Catherine Wasko Physics 304 Physics of the Human Body May 3, 2005

Nuclear Magnetic Resonance Imaging

Physical fundamentals of magnetic resonance imaging

Biomedical Imaging Magnetic Resonance Imaging

NMR, the vector model and the relaxation

Principles of MRI. Vinyl Record. Last time: Today: Homework Due tonight! EE225E / BIO265. Transforms a temporal signal to a spatial signal

Lecture 12 February 11, 2016

Introduction to MRI. Spin & Magnetic Moments. Relaxation (T1, T2) Spin Echoes. 2DFT Imaging. K-space & Spatial Resolution.

Magnetic resonance imaging MRI

Principles of Magnetic Resonance Imaging

Basic MRI physics and Functional MRI

MR Fundamentals. 26 October Mitglied der Helmholtz-Gemeinschaft

Spectral Broadening Mechanisms

Chemistry 431. Lecture 23

Chem 325 NMR Intro. The Electromagnetic Spectrum. Physical properties, chemical properties, formulas Shedding real light on molecular structure:

10.4 Continuous Wave NMR Instrumentation

1 Magnetism, Curie s Law and the Bloch Equations

Topics. Spin. The concept of spin Precession of magnetic spin Relaxation Bloch Equation

Chapter 7. Nuclear Magnetic Resonance Spectroscopy

Sketch of the MRI Device

Spin Dynamics Basics of Nuclear Magnetic Resonance. Malcolm H. Levitt

Spectral Broadening Mechanisms. Broadening mechanisms. Lineshape functions. Spectral lifetime broadening

Part II: Magnetic Resonance Imaging (MRI)

Biophysical Chemistry: NMR Spectroscopy

Topics. The History of Spin. Spin. The concept of spin Precession of magnetic spin Relaxation

COPYRIGHTED MATERIAL. Production of Net Magnetization. Chapter 1

Index. p, lip, 78 8 function, 107 v, 7-8 w, 7-8 i,7-8 sine, 43 Bo,94-96

Basic p rinciples COPYRIGHTED MATERIAL. Introduction. Atomic s tructure

Measuring Spin-Lattice Relaxation Time

Biophysical Chemistry: NMR Spectroscopy

Bioengineering 278" Magnetic Resonance Imaging" Winter 2010" Lecture 1! Topics:! Review of NMR basics! Hardware Overview! Quadrature Detection!

G Medical Imaging. Outline 4/13/2012. Physics of Magnetic Resonance Imaging

Principles of Nuclear Magnetic Resonance Microscopy

NMR Spectroscopy of Polymers

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

Introduction to Magnetic Resonance Imaging (MRI) Pietro Gori

NUCLEAR MAGNETIC RESONANCE. The phenomenon of nuclear magnetic resonance will be used to study magnetic moments of nuclei.

Lecture 02 Nuclear Magnetic Resonance Spectroscopy Principle and Application in Structure Elucidation

INTRODUCTION TO NMR and NMR QIP

K-space. Spin-Warp Pulse Sequence. At each point in time, the received signal is the Fourier transform of the object s(t) = M( k x

Physical Background Of Nuclear Magnetic Resonance Spectroscopy

Classical Description of NMR Parameters: The Bloch Equations

Spin. Nuclear Spin Rules

Magnetization Gradients, k-space and Molecular Diffusion. Magnetic field gradients, magnetization gratings and k-space

Chapter 14:Physics of Magnetic Resonance

Classical Description of NMR Parameters: The Bloch Equations

Ferdowsi University of Mashhad

NMR Spectroscopy Laboratory Experiment Introduction. 2. Theory

Introductory MRI Physics

Magnetic Resonance Spectroscopy EPR and NMR

The Physical Basis of Nuclear Magnetic Resonance Part I ESMRMB. Jürgen R. Reichenbach

Pulsed Nuclear Magnetic Resonance

Spin. Nuclear Spin Rules

Part III: Sequences and Contrast

Spin Relaxation and NOEs BCMB/CHEM 8190

NMR BMB 173 Lecture 16, February

NMR and MRI : an introduction

With that first concept in mind, it is seen that a spinning nucleus creates a magnetic field, like a bar magnet

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Magnetic Resonance Imaging

BASIC MRI PHYSICS SPIN GYMNASTICS Don Plewes PhD, Walter Kucharczyk MD

1.b Bloch equations, T 1, T 2

Physics of MR Image Acquisition

Shimming of a Magnet for Calibration of NMR Probes UW PHYSICS REU 2013

VIII. NUCLEAR MAGNETIC RESONANCE (NMR) SPECTROSCOPY

We have seen that the total magnetic moment or magnetization, M, of a sample of nuclear spins is the sum of the nuclear moments and is given by:

MRI Homework. i. (0.5 pt each) Consider the following arrangements of bar magnets in a strong magnetic field.

The physics US and MRI. Prof. Peter Bogner

Nuclear Magnetic Resonance Spectroscopy

More NMR Relaxation. Longitudinal Relaxation. Transverse Relaxation

Rochester Institute of Technology Rochester, New York. COLLEGE of Science Department of Chemistry. NEW (or REVISED) COURSE:

Relaxation times in nuclear magnetic resonance

NMR Imaging in porous media

Magnetic Resonance Imaging

Apodization. Gibbs Artifact. Bioengineering 280A Principles of Biomedical Imaging. Fall Quarter 2013 MRI Lecture 5. rect(k x )

Geology 228/378 Applied & Environmental Geophysics Lecture 8. Induced Polarization (IP) and Nuclear Magnetic Resonance (NMR)

Nuclear magnetic resonance spectroscopy

COPYRIGHTED MATERIAL. Basic Principles. 1.1 Introduction

Lecture #6 NMR in Hilbert Space

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Introduction of Key Concepts of Nuclear Magnetic Resonance

Electromagnetics in Medical Physics

Physics of Imaging Systems Basic Principles of Magnetic Resonance Imaging II

NMR Spectroscopy. Structural Analysis. Hans Wienk & Rainer Wechselberger 2011

Magnetic Resonance Imaging in Medicine

Medical Imaging Physics Spring Quarter Week 9-1

How does this work? How does this method differ from ordinary MRI?

RADIOLOGIV TECHNOLOGY 4912 COMPREHENSEIVE REVIEW/MRI WORSHEET #1- PATIENT CARE AND SAFETY/PHYSICAL PRINCIPLES

Transcription:

Nuclear Magnetic Resonance Imaging Simon Lacoste-Julien Electromagnetic Theory Project 198-562B Department of Physics McGill University April 21 2003 Abstract This paper gives an elementary introduction to NMR imaging, with emphasis on the electromagnetic content of the theory. The basic physical principles of NMR are presented, followed by the image encoding principles and an overview of the EM hardware used. Contents 1 Introduction 2 2 Nuclear Magnetic Resonance 2 2.1 Basic Principles............................ 2 2.2 Larmor Precession.......................... 3 2.3 Bloch model and relaxation time.................. 4 3 NMR Imaging 5 3.1 Slice selection............................. 5 3.2 Gradient Encoding and Fourier Transform............. 6 4 The Hardware 7 4.1 The Magnet.............................. 7 4.2 RF Coils................................ 7 4.3 Gradient Coils............................ 8 1

1 Introduction Before Nuclear Magnetic Resonance (NMR), many scientists were taught that you cannot image an object smaller than the wavelength than the energy used to image it. It could explain why radio waves have been very lately thought to be used to image the inside of the human body, even if they are quite more inoffensive than the usual x-rays which can ionize atoms (it is worth noting that the only penetrating radiations in the human body are radio waves and lower frequencies; as well as x-rays and higher frequencies). The NMR phenomenon comes from resonance absorption of radiation from nuclear spins in an external magnetic field. The frequency of resonance is very well defined and depends on the magnetic field, thus a spatial encoding of a sample can be made by a spatial variation of the magnetic field and then measuring the different resonant frequencies observed for specific excitations. The usual scattering method to image an object is replaced by the magnetic field encoding and resonance signal. We will start this paper with a brief overview of the physics of nuclear magnetic resonance, which is the basic phenomenon behind NMR. We will then describe how NMR can be used to image an object. We will finish with an overview of some of the EM hardware used in NMR. 2 Nuclear Magnetic Resonance NMR was discovered independently by two groups of physicists headed by F. Bloch and E.M. Purcell in 1946 (see [2] for the historical references). It is fundamentally a quantum mechanical effect which arises when particles with a non-zero magnetic moment are put in a strong constant magnetic field superposed with a weaker orthogonal oscillating magnetic field with the correct frequency. 2.1 Basic Principles We will consider here the spin- 1 2 case, as it is the one used in NMR imaging. Let s say that we have a constant field B 0 oriented in the z-direction. We know that the energy of a magnetic dipole µ in an external magnetic field B 0 is given by: H = µ B 0 (1) and the magnetic moment of a particle with spin S is: µ = γs (2) where γ is a constant called the gyromagnetic ratio for the particle. In the case where we apply a small oscillating magnetic field B 1 in the x-direction, it is shown in [5, p.104] that the probability of transition from a spin up to a spin down state goes like a Lorentzian shape with resonant frequency w 0 = 2

µb 0 / h 1 and width characterized by the quotient B 1 /B 0 (and so very narrow when B 1 B 2 0 ). This means that only radiation with a specific frequency ν 0 = w 0 /2π (characterized by B 0 and called the Larmor frequency) can excite the spins of the particles. In NMR imaging, the particles considered are the single protons of the hydrogen atom 1 H, and those have a gyromagnetic ratio of γ = 2.7 10 8 s 1 T 1 and thus a resonant frequency of 42.6 MHz (radio frequency) for a field B 0 of 1 Tesla. 2.2 Larmor Precession In quantum mechanics, the spin in the x-direction or the y-direction can be considered as a linear combination of spin up and spin down (in z-direction), and so the intermediate states in the spin up to spin down transition in a sample with lots of nuclei could be seen as the appearance of a transverse spin (and thus transverse magnetization). Instead of going in the quantum mechanic treatment of this phenomenon, we can simply look at the classical theory of electromagnetism which explains some of its features. Since the magnetic moment µ is related to the angular momentum (spin) S by equation (2) and that the torque N exerted on a magnetic dipole is given by N = µ B [3, p.174], then we have the classical motion equation for the dipole: dµ dt = γµ B 0 (3) which has as solution, in the case of a static B 0 field, that the dipole µ precesses about B 0 on the surface of a cone at a Larmor frequency w 0 = γb 0 (see figure 1(a)). In the case where we replace B 0 by kb 0 + ib 1 cos wt, then equation (3) expressed in a rotating coordinate system around the z-axis with angular frequency w becomes: dµ dt = γµ k(b 0 w/γ) + ib rf (4) where B rf = B 1 /2 and is a constant in time, and the unit vectors are in the direction of the primed coordinate axis. By comparing this equation with equation (3), we see that the motion in the rotating frame is simply a precession around the effective static magnetic field B eff = k(b 0 w/γ) + ib rf (see figure 1(b)). At magnetic resonance, w = γb 0 and thus µ precesses around ib rf with angular frequency w 1 = γb rf. This means that with the correct timing (which would only depend on B rf ), we could induce a rotation of 90 o to µ in the rotating coordinate frames and thus obtained a pure transverse dipole. If then the B rf would be turned off, we would be left with a transverse dipole precessing around the z-axis with Larmor frequency w 0 (and this could induce an oscillating electric field in a coil with characteristic frequency which could be measured; this is the signal which is measured in NMR imaging). 1 We note that µb 0 is simply the energy difference between the spin up and spin down state. 2 B 0 is of the order of 1 Tesla in NMR imaging and B 1 is the order of 10 mt. 3

Figure 1: (a) Precession of the magnetic moment µ about B 0 in the laboratory coordinate system; (b) precession of µ about the effective magnetic field k(b 0 w/γ) + ib rf in the rotating coordinate system. Taken from [4, p.13]. 2.3 Bloch model and relaxation time We now need to make some macroscopic treatment of the process in order to link it with the real world. At thermal equilibrium, there will be more nuclei with spin up than spin down since they have lower energy in the magnetic field B 0 (by equation 1). The probability of finding a nucleus in a particular spin state with energy E m at thermal equilibrium is given by the Boltzmann statistics: P m e E m/k B T (5) where k B is the Boltzmann constant and T is the temperature. The difference in population of the different energy levels (spins) is small (e.g. (n + n )/n 10 5 at room temperature and at 1 Tesla); but the large number of hydrogen protons in a human sample yields a macroscopic difference between the number of spin up and down and thus a macroscopic magnetization M = ρ µ where ρ and µ are the concentration and average magnetic moment of the nuclei, respectively. According to the results of the preceding section, we could rotate the magnetization vector by 90 o with a r.f. radiation pulse at the Larmor frequency w 0 with the magnetic polarization in the x-y plane and with the correct duration determined by B rf. After the pulse, the average magnetization would keep precessing around the z-axis at the Larmor frequency w 0, and according to Faraday s law, it would induce an alternating current with the same frequency in a coil with a normal in the x-y plane. But according to the Bloch model (see [4, p.15]), the transverse magnetization decays exponentially to 0 with a characteristic time of T 3 2 known as spin-spin relaxation time. The measured signal is thus a decaying oscillating current, called the Free Induction 3 This parameter depends on the environment in which the hydrogen atoms are and thus can be used to differentiate different type of tissues in NMR imaging. 4

Current (FID), which will be proportional to ρ. The decay is due to small interactions between the microscopic dipoles, which makes small perturbation amongst the Larmor frequencies of the dipoles and cause them to dephase and average to zero after a certain time. Also, the magnetization in the z- direction gradually relaxes to its initial value due to thermal perturbations and internuclear interactions. In the Bloch model, the relaxation is also given as an exponential law: M z = M 0 (1 e t/t1 ) (6) where T 1 is called the spin-lattice relaxation time. This sets the scale of time after which we can repeat a r.f. pulse to excite again the spins. 3 NMR Imaging We are now ready to understand how we can do NMR imaging. A simplified version of it could be described with the following steps: 1. The patient is put in a strong uniform magnetic field B 0. 2. RF coils are used to select a slice of the sample to map (see section 3.1) by rotating the nuclear moments by 90 o only in a specific slice. 3. Gradient coils are used to create a precise linear spatial variation of the magnetic field to encode the position of the excitation in the slice (see section 3.2). 4. RF resonant coils measure the FID produced by the decaying precession of the nuclear moment in the spatially encoded magnetic field. 5. The signal can be inverse Fourier transformed to obtain a spatial distribution of the hydrogen nuclei density. 3.1 Slice selection To select a thin slice of the sample to map, a small magnetic field gradient in the z-direction is superposed (with component parallel to the main field) onto the main uniform field. This means that B 0 will be constant only in x- y planes and thus that the nuclei have constant Larmor frequency planes by planes. Since we saw in section 2.1 that the width of the magnetic resonance effect was quite narrow when the exciting field is very small compared to the constant field, only a specific slice will be excited if the exciting radiation has only one narrow band of frequencies. The time variation of the B rf field usually used (as for equation (4)) is a sinusoidal with frequency w 0 multiplied by a sincshaped pulse (shaped like sin x/x). This signal is used because it is a pulse with a narrow square frequency distribution (as can be seen from its Fourier transform), and thus it will only excite a narrow slice. The r.f. pulse is made so that it has just the right duration to rotate the spins by 90 o. 5

Figure 2: Axial head NMR image with pathology. Can you see the pathology? Image taken from [2]. 3.2 Gradient Encoding and Fourier Transform Just before recording the FID of the precessing spins, gradient coils are turned on to encode the x-y position of the spin volumes. A frequency of precession encoding is made by adding a small steady linear variation in the z-field along the x direction (the frequency of precession is determined by the magnetic field, as seen in section 2.2). Also, a phase encoding is made by adding for a specific interval of time a small linear variation in the z-field along the y direction (the difference of frequency along the y direction will cause a phase shift between the different dipole precessions, which will come back to the same frequency when the y-gradient will be turned off). After those two quick encoding, the FID signal is recorded using r.f. resonant coils. Because of the phase and frequency spatial encoding, the density distribution of the nuclei in a slice can be found back by invert Fourier transforming the recorded current signal. The density of nuclei can then be depicted on a gray-scale image (an example of typical NMR image is given in figure 2). Of course, this rapid turning on and off of gradient coils could induce eddy currents in the conducting structures which in turn induce both temporally and spatially varying magnetic field which could create distortions in the image (which will be one kind of image artifact) and attenuation of the signal. Advanced techniques try to reduce those effects (as well as the inherent inhomogeneities of the field by specially designed gradient waveforms which 6

Figure 3: Major components. Image taken from [2]. compensate for the created eddy currents see [4] and [2] for more information about image artifacts. 4 The Hardware We end this paper with a brief presentation of some of the EM hardware used in NMR. Figure 3 gives more details about the general setting. 4.1 The Magnet It is important to have a strong uniform main magnetic field in order to have a stable magnetic resonance effect and also because the position of the nuclei is encoded with the field, so that any undesired variation in the field causes geometric distortions of the image (variation in the field can also be caused by metallic objects on (or in!) the patient). The typical strength of the fields used in NMR imaging is between 0.5 and 2 Tesla [1]. Such strong fields cannot be obtained with resistive magnets (coils with electric current), so superconducting magnets are used instead. The magnet is bathed in liquid helium to keep it in the superconducting regime. 4.2 RF Coils The main functions of r.f. coils are to excite the magnetization in a sample and to receive the signal produced by the excited magnetization (they can sometimes act both as a receiver and as a transmitter). They are simply coils which are in series with an inductor and a capacitor in order to have a resonance behaviour with angular frequency w = 1/ LC (7) 7

Figure 4: Galey coils to produce gradient in the transverse direction. Image taken from [2]. where L is the inductance and C, the capacitance. They are sometimes shaped according to specific parts of the body of the patient (leg, wrist, head, etc.) in order to have a closer measurement of the signal. 4.3 Gradient Coils The field gradient in the z-direction is simply made with a normal cylindrical coil. The field gradient in the x and y-directions can be made using Galey coils as is shown in figure 4. References [1] Gould, Todd A., How MRI Works. Retrieved April 21, 2003, from http://electronics.howstuffworks.com/mri.htm [2] Hornak, Joseph P., The Basics of MRI. Retrieved April 21, 2003, from http://www.cis.rit.edu/htbooks/mri/ [3] Jackson, John David, Classical Electrodynamics, 3rd Ed., John Wiley & Sons, Berkeley, 1999. [4] Kuperman, Vadim, Magnetic Resonance Imaging, Academic Press, Chicago, 2000. [5] Townsend, John S., A Modern Approach to Quantum Mechanics, University Science Books, California, 2000. 8