Determining Chemical Composition. Of Sputtered Uranium Oxide Thin Films. through X-Ray Photoelectron Spectroscopy

Similar documents
Advanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

Lecture 5. X-ray Photoemission Spectroscopy (XPS)

X-Ray Photoelectron Spectroscopy (XPS)

Electron Spectroscopy

The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements

Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis

Methods of surface analysis

X-Ray Photoelectron Spectroscopy (XPS)

Energy Spectroscopy. Excitation by means of a probe

X- ray Photoelectron Spectroscopy and its application in phase- switching device study

PHI 5000 Versaprobe-II Focus X-ray Photo-electron Spectroscopy

An Introduction to Diffraction and Scattering. School of Chemistry The University of Sydney

SOFT X-RAYS AND EXTREME ULTRAVIOLET RADIATION

Ion sputtering yield coefficients from In thin films bombarded by different energy Ar + ions

Energy Spectroscopy. Ex.: Fe/MgO

Characterization of Secondary Emission Materials for Micro-Channel Plates. S. Jokela, I. Veryovkin, A. Zinovev

Measurement of the Complex Index of Refraction for UO x in the Extreme Ultraviolet

Auger Electron Spectroscopy (AES)

Metal Deposition. Filament Evaporation E-beam Evaporation Sputter Deposition

Supporting Information. Re-Investigation of the Alleged Formation of CoSi Nanoparticles on Silica. Van An Du, Silvia Gross and Ulrich Schubert

1 Introduction COPYRIGHTED MATERIAL. 1.1 HowdoweDefinetheSurface?

Xps Study of the Oxidation State of Uranium Dioxide

PHOTOELECTRON SPECTROSCOPY (PES)

The Use of Synchrotron Radiation in Modern Research

Photoemission Spectroscopy

Lecture 23 X-Ray & UV Techniques

Electron Microprobe Analysis and Scanning Electron Microscopy

X-Ray Photoelectron Spectroscopy (XPS)-2

REMOVING CONTAMINANTS FROM SILICON WAFERS TO FACILITATE EUV OPTICAL CHARACTERIZATION. Ross Robinson

Thorium-Based Thin Films as Highly Reflective Mirrors in the EUV

PDF created with FinePrint pdffactory trial version

Auger Electron Spectroscopy (AES) Prof. Paul K. Chu

Measurement of the role of secondary electrons in EUV resist exposures

The design of an integrated XPS/Raman spectroscopy instrument for co-incident analysis

EEE4106Z Radiation Interactions & Detection

Fig 1: Auger Electron Generation (a) Step 1 and (b) Step 2

X-ray Photoelectron Spectroscopy (XPS)

Auger Electron Spectroscopy Overview

PHOTOELECTRON SPECTROSCOPY IN AIR (PESA)

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

An Introduction to Auger Electron Spectroscopy

MODERN TECHNIQUES OF SURFACE SCIENCE

Secondary Ion Mass Spectrometry (SIMS)

X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES)

Photoelectron Spectroscopy. Xiaozhe Zhang 10/03/2014

An introduction to X- ray photoelectron spectroscopy

Name: (a) What core levels are responsible for the three photoelectron peaks in Fig. 1?

X-Ray Photoelectron Spectroscopy (XPS)-2

Auger Electron Spectroscopy *

Combinatorial RF Magnetron Sputtering for Rapid Materials Discovery: Methodology and Applications

IV. Surface analysis for chemical state, chemical composition

XPS & Scanning Auger Principles & Examples

The Monte Carlo Simulation of Secondary Electrons Excitation in the Resist PMMA

Secondary ion mass spectrometry (SIMS)

Surface Chemistry and Reaction Dynamics of Electron Beam Induced Deposition Processes

ICTP School on Synchrotron Radiation and Applications 2008 Surface Science, Photoemission and Related Techniques Fadley, Goldoni

Construction of an extreme ultraviolet polarimeter based on highorder harmonic generation

Photon Interaction. Spectroscopy

X-ray photoelectron spectroscopic characterization of molybdenum nitride thin films

Review. Surfaces of Biomaterials. Characterization. Surface sensitivity

Developing Instrumentation for Fabricating and Characterizing Thin Film Aluminum Mirrors

5) Surface photoelectron spectroscopy. For MChem, Spring, Dr. Qiao Chen (room 3R506) University of Sussex.

Electron spectroscopy Lecture Kai M. Siegbahn ( ) Nobel Price 1981 High resolution Electron Spectroscopy

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

Appearance Potential Spectroscopy

Studying Metal to Insulator Transitions in Solids using Synchrotron Radiation-based Spectroscopies.

Keywords: electron spectroscopy, coincidence spectroscopy, Auger photoelectron, background elimination, Low Energy Tail (LET)

Lecture 22 Ion Beam Techniques

AES - Auger Electron Spectrosopy

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2

Film Characterization Tutorial G.J. Mankey, 01/23/04. Center for Materials for Information Technology an NSF Materials Science and Engineering Center

X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka

Supporting Information s for

Vibrational Spectroscopies. C-874 University of Delaware

Table 1: Residence time (τ) in seconds for adsorbed molecules

MSE 321 Structural Characterization

Probing Matter: Diffraction, Spectroscopy and Photoemission

structure and paramagnetic character R. Kakavandi, S-A. Savu, A. Caneschi, T. Chassé, M. B. Casu Electronic Supporting Information

Electron Spettroscopies

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric

Secondary Ion Mass Spectroscopy (SIMS)

Chapter 7 Introduction to Spectroscopy

Contrasted strengths and weakness of EDS, WDS and AES for determining the composition of samples

Which of the following classes of electromagnetic waves will not ionise neutral atoms?

X-ray Energy Spectroscopy (XES).

XPS/UPS and EFM. Brent Gila. XPS/UPS Ryan Davies EFM Andy Gerger

Inelastic soft x-ray scattering, fluorescence and elastic radiation

EE 527 MICROFABRICATION. Lecture 5 Tai-Chang Chen University of Washington

Reduced preferential sputtering of TiO 2 (and Ta 2 O 5 ) thin films through argon cluster ion bombardment.

Repeatability of Spectral Intensity Using an Auger Electron Spectroscopy Instrument Equipped with a Cylindrical Mirror Analyzer

QUESTIONS AND ANSWERS

5.8 Auger Electron Spectroscopy (AES)

MS482 Materials Characterization ( 재료분석 ) Lecture Note 5: RBS

Multilayer Interference Coating, Scattering, Diffraction, Reflectivity

object objective lens eyepiece lens

Understanding electron energy loss mechanisms in EUV resists using EELS and first-principles calculations

Fig Photoemission process.

Transcription:

Determining Chemical Composition Of Sputtered Uranium Oxide Thin Films through X-Ray Photoelectron Spectroscopy Kristi Adamson Department of Physics & Astronomy Brigham Young University April 2004 Submitted to the Department of Physics and Astronomy in partial fulfillment of graduation requirements for the degree of Bachelor of Science Advisor: Dr. David Allred Signature: Thesis Coordinator: Dr. Justin Peatross Signature: Department Chair: Dr. Scott Sommerfeldt Signature:

Abstract This study involves the chemical composition of thin films (100-200? ) of uranium oxide, created through DC magnetron sputtering, as deposited as well after aging in the lab for a time lapse of over one year. The oxidation state of the uranium oxide films is investigated through x-ray photoelectron spectroscopy (XPS). The sample is placed under high vacuum and bombarded with x rays of known energies, producing photoelectrons. The energy of the ejected electrons is determined by an electron detector, and the number of electrons collected at each energy is counted. By this method, the bonding energy of the electrons at and near the surface is determined. The data shows that the surface oxidizes gradually over a period of months. By comparing our work with that of Teterin [1] ( A Study of Synthetic and Natural Uranium Oxides by X-Ray Photoelectron Spectroscopy. J. Phys. Chem. Minerals. 1981) it is concluded that the valence electrons of the uranium samples are intermediate between UO2 and UO3, with UO3 becoming more prominent as the sample has more time to oxidize. The composition below the surface is studied by using a beam of energetic argon ions to sputter off layers of the film, while periodically examining the sample with XPS. Nitrogen is also detected in the film, making up approximately 15% of the composition. It may have been incorporated into the samples during reactive sputtering. The amount of oxygen appears constant with depth and appears consistent with UO2. After over one year of exposure to atmosphere, the surface layer of the uranium trioxide is approximately 1nm thick.

Contents I. Introduction 4 II. X-Ray Photoelectron Spectroscopy 5 III. Background 8 IV. Experimental Procedure 12 V. Data 13 VI. Discussion 19 VII. Conclusion 22 Appendix 23 Bibliography 28

I. Introduction The study of thin films (100-200? ) as multilayer or single surface mirrors in the extreme ultraviolet (EUV) radiation range has been growing in the scientific community over the past years [1], [2]. Producing reliable mirrors at in this range will prove beneficial in many areas of technology, specifically lithography, medical equipment and space observation [3]. The physics department at Brigham Young University produced thin films containing uranium that were used in the IMAGE satellite project in 2000 for this purpose. It had been suggested that, theoretically, uranium should have very high reflectance in the desired range, and thus uranium/silicon multilayer reflectors were the initial product. But as the films sat open to atmosphere, the uppermost uranium layer bonded with the oxygen, chemically altering the surface of the mirrors. However, these altered films actually served better in their intended optical purpose than had been anticipated [4]. This has fostered more study of uranium oxide films as EUV mirrors. One particular difficulty with uranium dioxide is that it is chemically unstable, especially when in the form of a powder or thin film [5]. Thus the samples require continuous compositional studies to determine the rate of oxidation as the films are left open to atmosphere.

II. X-Ray Photoelectron Spectroscopy X-ray photoelectron spectroscopy (XPS) is a surface sensitive method that uses the photoelectric effect to determine the composition of thin films. Aluminum is bombarded with high-energy electrons (10keV) to create a stream of x rays. As these low energy x rays are quickly absorbed by air, the measurements must be conducted at extremely low pressures. In the system employed for this study, the base pressure was 10-6 torr. These x rays then impinge on our films. If the x ray has sufficient energy, an atom is ionized, and an electron is ejected from the sample. A position sensitive electron detector determines the kinetic energy of the electron. The hemispherical shape of the detector disperses the electrons across a resistive plate according to their energy. By analyzing the position of the electron after traversing the hemisphere, its energy can be determined. Figure 1. The X-Ray spectroscopy system used in this study. An ESCA SSX-100, located in the Brigham Young University Department of Chemistry and Biochemistry

By knowing the energy of the original x ray and the kinetic energy of the emitted electron (provided it experiences no energy loss from collisions), the energy with which the electron was bound to the atom, E b, can easily be determined by the following equation: E b =h?-e k where h? is the energy of the incident x ray and E k is the kinetic energy of the ejected electron. Each atom or molecule has characteristic energies with which their electrons are bound. Thus from this binding energy, information can be obtained about the relative abundance of different elements in a sample and its approximate chemical composition. The sampling depth from which the electrons escape without energy loss depends on two factors: the kinetic energy of the electron and the electron escape angle. Beyond 100eV, an electron with greater kinetic energy can travel through more material before undergoing inelastic scattering and thus losing its energy. Hence, weakly bound electrons come from deeper in the sample. As can be seen qualitatively from figure 2, the sampling depth is proportional to the cosine of the electron escape angle. e - e -?=0?=60? 2? Figure 2. The dependence of sampling depth on escape angle of electrons. [6]

Electrons escaping at an angle near normal can travel further within the sample before scattering and thus the sampling depth will be greater. [7] Electrons escaping at a near-grazing angle will give more information about the surface. While XPS reveals information only about the near-surface region of our sample, we can incorporate this technique with ion sputtering, which allows the composition to be determined at different depths within the film. The sample is bombarded with argon ions, which effectively sputter off layers of the film. As the sample is sputtered away, it is periodically scanned by XPS to ascertain the chemical composition. However, the depth at which we scan can only be determined qualitatively, as there are no determined sputtering rates for uranium oxides. A rough model, in which we assume constant density throughout the sample, is used to determine relative depth composition.

III. Background Several XPS studies have been done of bulk uranium oxides in an attempt to understand oxidation rates and bonding parameters of oxygen and uranium [8], [9], [10]. An effort of special interest is determining which electrons in uranium are involved in the bonding process and the effects these bonds have on the energies of various uranium electron bands. As the uranium becomes more oxidized, the molecular orbitals shift in binding energy. These shifts are used to qualitatively understand the behavior of uranium sample oxidation. Teterin [8] extensively studied various uranium oxides using XPS. The studied samples were in the form of both finely dispersed powders and compact pressed powders on indium and titanium substrates. Spectra were taken three times for each sample. The data obtained from each type of sample was the varied only within experimental limits. The positions of the uranium 4f 5/2 and 4f 7/2 energy peaks are especially helpful in determining composition as it changes with the different amounts of oxygen. Figure 3 from Teterin shows how the peaks shift to higher binding energies with increased oxidation.

Figure 3. Relative intensity versus binding energy for electrons in the U4f 5/2 and 4f 7/2 orbitals [8] The O1s energy peak also shows great variance for different oxidation states. As can be seen from figure 4, it is actually composed of two closely positioned peaks that vary in intensity as oxidation changes. The higher energy peak becomes more prominent as O/U ratio changes from 2 to 3.

Figure 4. Relative intensity versus binding energy for electrons in the O1s orbital. [8] Comparison to published data can provide qualitative information as to the behavior of the films, but cannot necessarily give quantitative indications. Thus a more exact determination of the composition can be established from a relative comparison of electron energy peak areas. The number of photoelectrons per second (N) in a specific spectra peak can be determined the equation: N = n f σ θ γ λ A T where n represents the number of atoms of the element per cm 3 in the sample, f is the x- ray flux in photons/cm 2 /sec, s is the photoelectric cross-section for the atomic orbital of

interest in cm 2,? is a factor of angular efficiency of the instrumental arrangement in radians, based on the angle between the photon path and the detected electron,? represents the efficiency in the photoelectric process for formation of photoelectrons of the normal photoelectron energy,? is the mean free path of the photoelectrons in the sample in cm, A is the area of the sample in cm 2, and T is the detection efficiency for electrons emitted from sample[11]. The above equation can be solved for n and a ratio of atoms per element can be determined by the equation: n n 1 2 = N N 1 2 S S 2 1 where S is defined as the atomic sensitivity factor. Values of S can be determined experimentally and are known for most elements [12]. In the ion sputtering process, it is valuable to know the depth at which each scan is being taken. Without a definitive knowledge of the sample s composition, it is not possible to determine an exact sputtering rate. However, it is known that the sputtering rate will be proportional to the density of the sample as well as the sputtering time. As our results show fairly constant composition, we will assume uniform density and determine a relative scanning depth based solely on the sputtering time.

IV. Experimental Procedure The samples of uranium oxide are produced through DC magnetron sputtering. The uranium is deposited on silicon wafers in the presence of oxygen (partial pressure of 3*10-4 torr) creating a uranium oxide film. The base vacuum pressure is maintained below 4*10-6 torr, while the partial pressure of argon is 4*10-3 torr. The uranium is allowed to deposit to a thickness of approximately 30nm [15]. The thicknesses are determined with atomic force microscopy, low-angle x-ray diffraction and ellipsometry. The reflectance of the samples is measured by the monochromator at Brigham Young University as well as by the synchrotron at the University of California Berkeley. An SSX-100 ESCA (Electron Spectroscopy for Chemical Analysis) Spectrometer is used to gather electron binding energies. It is calibrated monthly by the Au4f line, which is located at a binding energy of 83.98eV, as well as the Cu2p and 3p lines at 75.13eV and 932.66eV, respectively, to ensure consistently precise results. The pressure within the chamber is maintained at 2-6*10-9 torr for high resolution scans. All scans shown in this paper are done with a detector angle of 35, unless otherwise noted. Compositions are determined through comparison to other works [8] as well as computed using the analysis program, ESCAVB [13], which employs the peak area comparison method. In sputtering the sample, a SPECS ion gun system (PU-IQE 12/38) was used. A spot of 1.5 mm 2 was rastered for 20-120 seconds between each XPS scan. Between tests and scans, samples are stored in plastic containers which are left open to atmosphere.

V. Data Surface Composition The surface of a sample of uranium oxide was measured through XPS over the course of one year. Figures 5 and 6 below show the survey scans of the sample (for more graphs, see appendix). 3.8K 3.6K 3.4K 3.2K 3K 2.8K 2.6K 2.4K 2.2K 2K 1.8K 1.6K 1.4K 1.2K 1K 800 600 400 200 1099 1049 999 949 899 849 799 749 699 649 599 549 499 449 399 349 299 249 199 149 99 49 Figure 5. Survey scan of initial uranium oxide sample

1.6K 1.5K 1.4K 1.3K 1.2K 1.1K 1K 900 800 700 600 500 400 300 200 100 1099 1049 999 949 899 849 799 749 699 649 599 549 499 449 399 349 299 249 199 149 99 49 Figure 6. Survey scan of uranium oxide 52 weeks after deposition Identifying Peaks: 99eV U5d 5/2 393eV U4f 5/2 107eV U5d 3/2 531eV O1s 153eV Si2s 740eV U4d 5/2 198eV U5p 3/2 783eV U4d 3/2 285eV C1s 975eV O Auger 382eV U4f 7/2 By comparing areas underneath the major peaks of uranium, oxygen and carbon, the relative composition of the surface was determined and is displayed in figure 7. Surface Composition 70 60 Percent Composition 50 40 30 20 10 0 0 10 20 30 40 Time (weeks) 50 60 Uranium Oxygen Carbon Figure 7. Percentage composition with time determined by peak area comparison.

To see more clearly the behavior of the O1s and U4f peaks, high resolution scans were performed over small energy ranges. Figures 8-10 show the U4f 7/2 and U4f 5/2 peaks, and figures 11 and 12 show the O1s peak. (For more graphs, see appendix.) 12K 11K 10K 9K 8K 7K 6K 5K 4K 3K 2K 1K 401.9 399.9 397.9 395.9 393.9 391.9 389.9 387.9 385.9 383.9 381.9 379.9 377.9 375.9 373.9 Figure 8. U4f 5/2 and U4f 7/2 peaks of initial sample. 45K 40K 35K 30K 25K 20K 15K 10K 411.9 406.9 401.9 396.9 391.9 386.9 381.9 376.9 Figure 9. U4f 5/2 and U4f 7/2 peaks of sample 12 weeks after deposition.

18K 17K 16K 15K 14K 13K 12K 11K 10K 9K 8K 7K 6K 5K 4K 3K 411.9 406.9 401.9 396.9 391.9 386.9 381.9 376.9 Figure 10. U4f 5/2 and U4f 7/2 peaks of sample 52 weeks after deposition. 8K 7.5K 7K 6.5K 6K 5.5K 5K 4.5K 4K 540.9 538.9 536.9 534.9 532.9 530.9 528.9 526.9 524.9 522.9 Figure 11. O1s peak of initial sample.

8.5K 8K 7.5K 7K 6.5K 6K 5.5K 5K 4.5K 537.9 535.9 533.9 531.9 529.9 527.9 525.9 523.9 521.9 519.9 Figure 12. O1s peak scanned 52 weeks after deposition. After allowing time (over a year) for the sample to come to a fairly stable state, the surface was sputtered with an argon ion gun and the composition was determined with depth. Figure 13 below shows the composition calculated through relative peak areas. Relative Composition Composition (Percentage) 80 70 60 50 40 30 20 10 0 0 50 100 Relative Depth Uranium Oxygen Nitrogen Silicon Carbon Figure 13. Relative composition with relative depth determined through peak area comparison.

The U4f 5/2, U4f 7/2 and O1s peaks were also looked at with higher resolution and the peak energies are charted below. Peak Energy (ev) 381.6 381.4 381.2 381 380.8 380.6 Uranium Peak 1 380.4 380.2 380 0 20 40 60 Relative Depth 80 100 Figure 14. Position of U4f 7/2 peak with depth. Uranium Peak 2 Peak Energy (ev) 392.8 392.6 392.4 392.2 392 391.8 391.6 391.4 391.2 391 390.8 0 20 40 60 80 100 Relative Depth Figure 15. Position of U4f 5/2 peak with depth

Oxygen Peak 531.8 531.6 Peak Energy (ev) 531.4 531.2 531 530.8 530.6 530.4 530.2 530 0 20 40 60 80 100 Relative Depth Figure 16. Position of O1s peak with depth.

VI. Discussion From the two observed peaks, we see two different trends. On the surface the O1s peak indicates that the sample is moving to a higher oxidation state. Figure 17 shows how the composition changes with time, with composition ascertained by comparison to Teterin[8]. It appears as though the surface of the sample will completely oxidize to UO 3 with time. Composition Percent Composition 100 80 60 40 20 0 0 10 20 30 Time (weeks) Figure 17. Oxidation states of uranium oxide over time. The top line represents the percentage of UO 3 while the bottom represents that of UO 2. The U4f peaks, however, show a different trend. If the sample is shifting to a higher oxidation state, the uranium peaks should move to a higher energy. Yet, we can see from the charts that the peaks moved to a lower energy, indicating that the sample is moving to a lower oxidation state. This trend occurred for the first 18 weeks, after which the peaks began to shift to higher energies. This is due to the fact that, as noted previously, the sampling depth of XPS is a function of the electron s kinetic energy. The kinetic energy for the ejected O1s electrons is approximately 950eV while that of the U4f

electrons is 1110eV. From studying figure 18 [14], the difference in sampling distance can be determined. (The y-axis is measured in nanometers, while the x-axis is in electron volts.) Figure 18. The sampling distance of an electron as a function of energy [14] Thus, the difference in kinetic energy equates to a difference in sampling depth. Hence, the O1s peak is somewhat more sensitive to oxidation near the surface, while the U4f peaks probe the oxidation condition to a somewhat greater depth. It is obvious that the two trends are not identical. The scans run at larger angles are in agreement with what was determined for the U4f and O1s peaks at 30º. The scan done at 55º (which samples a greater depth than 75º) shows a slightly lower oxidation state than the scan done at 75º, but a higher oxidation state than that at 30º. The depth profiling data indicates non-uniform composition throughout the sample. The uranium at the surface is in a higher oxidation state than the rest of the sample. As we penetrate deeper into the sample, the uranium to oxygen ratio is fairly consistent with UO 2. As the silicon wafer is approached, the amount of oxygen goes up.

This may be due to bonding of oxygen with silicon; however, the shift in the uranium peaks indicates that the uranium is also bonding with more oxygen. Thus the top and bottom of the sample both have higher oxidation states than the bulk of the sample. Nitrogen is detected below the surface of the sample, increasing to and leveling off at approximately 15% of the composition. The position of the nitrogen peak is at 398eV, while the U4f 7/2 peak is at 391eV. The proximity to the uranium peak creates difficulty in detecting the presence of nitrogen, as the uranium peak will overlap the nitrogen peak. Thus it is more likely that the amount of nitrogen is consistently 15% throughout the sample, it is simply not detected until the uranium peaks begin to decrease.

VII. Conclusion Based on the data collected through x-ray photoelectron spectroscopy, it appears that the uranium oxide samples continue to oxidize after deposition, approaching a stable composition after a period of several weeks. In comparing to Teterin [8], it appears that the surface is initially UO 2 and slowly oxidizes to UO 3. However, the depth profiling of the film indicates that bulk of the sample remains at the initial composition of UO 2. Near the silicon wafer, the oxidation state may be higher. The surface easily collects high amounts of carbon which may cause problems in the reflectance of the mirrors. This problem has been under investigation by members of the BYU thin films research group. It has been found that UV cleaning effectively removes all carbon from the surface of silicon oxide films with a relatively short amount of exposure time (approximately 5 minutes). However it is possible that this may cause further oxidation or other damage to the surface. More studies are needed to determine the effect of this cleaning on other surfaces, such as uranium oxide. Although the uranium was believed to be sputtered only in the presence of oxygen, nitrogen may also have been present as it is found consistently throughout the films. It is unknown how this nitrogen may be bonded to the uranium oxide. Further investigation is needed to determine what role the nitrogen plays in our compound. Pressures of nitrogen in the DC sputtering chambers were not recorded. Additional investigation of the initial depth profile is required to determine if this is indeed the source.

Appendix A. Survey scans of uranium oxide sample 3.8K 3.6K 3.4K 3.2K 3K 2.8K 2.6K 2.4K 2.2K 2K 1.8K 1.6K 1.4K 1.2K 1K 800 600 400 200 1099 1049 999 949 899 849 799 749 699 649 599 549 499 449 399 349 299 249 199 149 99 49 Figure 19. Survey scan of initial uranium oxide sample. 2.2K 2K 1.8K 1.6K 1.4K 1.2K 1K 800 600 400 200 1099 1049 999 949 899 849 799 749 699 649 599 549 499 449 399 349 299 249 199 149 99 49 Figure 20. Survey scan of uranium oxide sample 12 weeks after deposition.

1.7K 1.6K 1.5K 1.4K 1.3K 1.2K 1.1K 1K 900 800 700 600 500 400 300 200 100 1099 1049 999 949 899 849 799 749 699 649 599 549 499 449 399 349 299 249 199 149 99 49 Figure 21. Survey scan of uranium oxide sample 25 weeks after deposition. 1.6K 1.5K 1.4K 1.3K 1.2K 1.1K 1K 900 800 700 600 500 400 300 200 100 1099 1049 999 949 899 849 799 749 699 649 599 549 499 449 399 349 299 249 199 149 99 49 Figure 22. Survey scan of uranium oxide sample 52 weeks after deposition.

B. High-resolution scans of U4f 7/2 and U4f 5/2 peaks 12K 11K 10K 9K 8K 7K 6K 5K 4K 3K 2K 1K 401.9 399.9 397.9 395.9 393.9 391.9 389.9 387.9 385.9 383.9 381.9 379.9 377.9 375.9 373.9 Figure 23. U4f 7/2 and U4f 5/2 peaks of initial uranium oxide sample. 45K 40K 35K 30K 25K 20K 15K 10K 411.9 406.9 401.9 396.9 391.9 386.9 381.9 376.9 Figure 24. U4f 7/2 and U4f 5/2 peaks of uranium oxide sample 12 weeks after deposition.

30K 28K 26K 24K 22K 20K 18K 16K 14K 12K 10K 8K 6K 411.9 406.9 401.9 396.9 391.9 386.9 381.9 376.9 Figure 25. U4f 7/2 and U4f 5/2 peaks of uranium oxide sample 25 weeks after deposition. 18K 17K 16K 15K 14K 13K 12K 11K 10K 9K 8K 7K 6K 5K 4K 3K 411.9 406.9 401.9 396.9 391.9 386.9 381.9 376.9 Figure 26. U4f 7/2 and U4f 5/2 peaks of uranium oxide sample 52 weeks after deposition.

C. High-resolution scans of O1s peak of uranium oxide sample 8K 7.5K 7K 6.5K 6K 5.5K 5K 4.5K 4K 540.9 538.9 536.9 534.9 532.9 530.9 528.9 526.9 524.9 522.9 Figure 27. O1s peak of initial uranium oxide sample. 22K 21.5K 21K 20.5K 20K 19.5K 19K 18.5K 18K 17.5K 17K 16.5K 16K 15.5K 15K 14.5K 14K 13.5K 13K 540.9 538.9 536.9 534.9 532.9 530.9 528.9 526.9 524.9 522.9 Figure 28. O1s peak of uranium oxide sample 12 weeks after deposition.

20K 19.5K 19K 18.5K 18K 17.5K 17K 16.5K 16K 15.5K 15K 14.5K 14K 13.5K 13K 12.5K 12K 11.5K 11K 10.5K 540.9 538.9 536.9 534.9 532.9 530.9 528.9 526.9 524.9 522.9 Figure 29. O1s peak of uranium oxide sample 25 weeks after deposition. 8.5K 8K 7.5K 7K 6.5K 6K 5.5K 5K 4.5K 537.9 535.9 533.9 531.9 529.9 527.9 525.9 523.9 521.9 519.9 Figure 30. O1s peak of uranium oxide sample 52 weeks after deposition.

Bibliography [1] R. A. M. Keski-Kuha, Layered synthetic microstructure technology considerations for the extreme ultraviolet. Appl. Opt. 23, 3534-3537 (1984). [2] D. Attwood, Soft X Rays and Ultraviolet Radiation. (Cambridge University Press, Cambridge, 1999), pp. 59-61, 371. [3] D. Attwood, Soft X Ray Microscopy and EUV Lithography: Imaging in the 20-50nm Region colloquium given Jan. 9, 2002 at Brigham Young University, Dept. Physics & Astronomy. [4] M. B. Squires, On Determining the Optical Constants of Sputtered U and a-si at 304 and 584?. Honors Thesis, Brigham Young University, 1999. [5] E. H. P. Cordfunke, The Chemistry of Uranium. (Elsevier Publishing Company, Amsterdam, 1969), pp.64-97. [6] R. Painter, http://goliath.inrs-ener.uquebec.ca/surfsci/arxps/sampling.html, March 16, 2002. [7] J. R. Baird and C. S. Fadley, X-Ray Photoelectron Angular Distributions with Dispersion-Compensating X-Ray and Electron Optics. J. Elec. Spec. Relat. Phenom. 11, 39-65 (1977). [8] Y. A. Teterin, V. M. Kulakov, A. S. Baev, N.B. Nevzorov, I.V. Melnikov, V. A. Streltsov, L. G. Mashirov, D. N. Suglabov, A. G. Zelenkov. A Study of Synthetic and Natural Uranium Oxides by X-Ray Photoelectron Spectroscopy. J. Phys. Chem. Minerals. 7, 151-158 (1981). [9] G. C. Allen and N. R. Holmes. Characterization of Binary Uranium Oxides by Infrared Spectroscopy. Appl. Spec. 48, 525-530 (1994). [10] Y. A. Teterin, Structure of X-Ray Photoelectron, Emission and Conversion Spectra and Molecular Orbitals of Uranium Compounds. J. Struc. Chem. 39, 850-857 (1998). [11] J. F. Moulder,, W. F. Stickle, P. E. Sobol, K. D. Bomben, Handbook of X-ray Photoelectron Spectroscopy. (Physical Electronics, Inc., Eden Prairie, MN, 1992). [12] C. D. Wahner et al, Surf. Interface Anal. 3, 211 (1981). [13] ESCA NT Software for Surface Science Instruments. Version 3.0, (Service Physics Inc, Bend OR, 2000).

[14] M. P. Seah, W. A. Dench. Compilation of Experimental Data Determined with Various Electron Energies for a Large Variety of Materials. Surf. Interface. Anal. I (1979). [15] S. Lunt, Determining the Indices of Refraction of Reactively Sputtered Uranium Dioxide Thin Films From 46 to 584 Angstroms. M. S. Thesis, Brigham Young University, 2002.