IvanaJakonić 1, Nataša Todorović 2, JovanaNikolov 2, Ines KrajcarBronić 3, BranislavaTenjović 4, MiroslavVesković 2

Similar documents
Effect of liquid scintillating cocktail volume on 3 H and 14 C measurement parameters using a Quantulus spectrometer

COMPARATIVE STUDY BETWEEN EXTERNAL STANDARD METHOD AND INTERNAL STANDARD METHOD FOR LOW-LEVEL TRITIUM MEASUREMENTS

A New Measurement Method For Tritium Rapid Detection by LSC. IM.CAEP Xu Qinghua

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

COMPARISON OF TWO DIFFERENT METHODS FOR DETERMINATION OF BIOGENIC FRACTION IN LIQUID FUELS

Time-Resolved Liquid Scintillation Counting

Comparison of Different Methods of Environmental Radioactivity Measurements at the Zagreb Radiocarbon and Tritium Laboratory

Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM

TIME-RESOLVED LIQUID SCINTILLATION COUNTING

Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator

Jerde, Peterson and Stein (1967) indicated that high-energy radiation

Basics of a/b Discrimination for Liquid Scintillation Counting

GEMMA RAURET, J S MESTRES and J F GARCIA

Making sense of Limits of Detection (with respect to Tritium analysis) James Thomson Meridian Biotechnologies Ltd.

Unit 2. Instrumentation. Experts Teaching from Practical Experience

TECHNETIUM-99 IN WATER

Hidex 300 SL Automatic TDCR Liquid Scintillation Counter

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

Northwest University, Mafikeng, Centre of Applied Radiation Science & Technology (CARST), South Africa.

High Performance Liquid Scintillation Analyzers. For Any Research, Discovery or Environmental Application Challenging Your Lab

Liquid Scintillation Counter

LIQUID SCINTILLATION COUNTERS. {Beta Counters}

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting

The measurement of tritium in environmental water samples

Application of NaI(T1) Scintillation Detector to Measurement of Tritium Concentration*

COSMIC BACKGROUND REDUCTION IN THE RADIOCARBON MEASUREMENTS BY LIQUID SCINTILLATION SPECTROMETRY AT THE UNDERGROUND LABORATORY OF GRAN SASSO

Standardization of Tritium by CIEMAT/NIST Method and TDCR Method

Implementation of ISO/IEC in a low level liquid s cintillation tritium laboratory

Hyuncheol Kim*, Yoonhee Jung, Wanno Lee, Guen-Sik Choi, Kun Ho Chung, Mun Ja Kang

EVALUATION OF HIGH-PURITY SYNTHETIC SILICA VIALS IN ACTIVE AND PASSIVE VIAL HOLDERS FOR LIQUID SCINTILLATION COUNTING OF BENZENE

arxiv: v1 [physics.ins-det] 3 Feb 2011

Energy Calibration of Liquid Scintillation Counter to allow Semi-Qualitative Nuclide Identification in Water Samples

1/5 Specifications Hidex 300 SLL automatic TDCR liquid scintillation counter

RADON EQUILIBRIUM MEASUREMENT IN THE AIR *

CATALOG OR PART NUMBER SX SX18 4 SX20 5 SX20 SAMPLE SX20-4 SX SX21 5 SX21 SAMPLE SX22 5 SX22 SAMPLE SX23 5 SX23 SAMPLE SX24 5 SX24 SAMPLE

Determination of 210 Pb and 210 Po in Water Samples

RESULTS YOU CAN COUNT ON

ZICOS - New project for neutrinoless double beta decay experiment using zirconium complex in liquid scintillator -

Application Note. Abstract. Introduction. Experimental

TECHNETIUM-99 IN SOIL

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

The DarkSide-50 Outer Detectors

Procedure for Radio Dating of Water Using Nuclear Hydrological Isotopes Techniques

Primary Standardisation of I-125

DETERMINATION OF RADON USING SILICONE OIL SCINTILLATOR

Results of the 2015 national indoor radon intercomparison measurements in Serbia

NICKEL-63/59 IN WATER

W. Udo Schröder Departments of Chemistry & of Physics and Astronomy

Simultaneous determination of radium and uranium activities in natural water samples using liquid scintillation counting

Synthesis of Plastic Scintillation microspheres: composition evaluation and alpha/beta discrimination capabilities

Hidex Liquid Scintillation Counters

Validation of detection efficiency of alpha particles in commercial liquid scintillation counters

ISOTOPES 3 H, δ 2 H AND δ 18 O IN GROUNDWATERS FROM VOJVODINA REGION

First tests of the big volume ultra low background gamma spectrometer

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

The Study for the Evaluation of the Method for 222 Rn in Water

TECHNETIUM-99 IN WATER

Luminescence free counting of 3 H facilitated by Hidex 300 SL/600 SL TDCR triple coincidence counters. Risto Juvonen

TRITIUM DETERMINATION IN WATER

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

PERFORMANCE AND DESIGN OF 03-ML TO 10-ML SYNTHETIC SILICA LIQUID SCINTILLATION VIALS FOR LOW-LEVEL 14C DETERMINATION

Plastic Scintillation and dryness evaporation: a new procedure for Alpha/Beta determination in water samples

ISO Water quality Determination of carbon 14 activity Liquid scintillation counting method

Depth Distribution of H-3, C-14 and Co-60 in Decommissioning of the Biological Shielding Concrete of KRR-2

BUREAU INTERNATIONAL DES POIDS ET MESURES

Radiation Detection and Measurement

Nuclear Engineering Seibersdorf GmbH

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second

SCINTILLATION DETECTORS AND PM TUBES

Dependence Of Gamma Ray Attenuation On Concentration Of Manganese (II) Chloride Solution

Survey of the performance of scintillation materials at low temperatures

THE POTENTIAL OF THE LONDON UNDERGROUND FOR LIQUID SCINTILLATION COUNTING

Scintillation detectors

Radioactivity Measurement Method for Environmental Monitoring Gross Alpha/beta Activities in Drinking Water in Turkey

Cherenkov Detector. Cosmic Rays Cherenkov Detector. Lodovico Lappetito. CherenkovDetector_ENG - 28/04/2016 Pag. 1

arxiv: v1 [physics.ins-det] 21 Dec 2015

PLASTIC SCINTILLATORS: A POWERFUL TOOL TO REDUCE MIXED WASTE.

Radio-HPLC Flow Cells

Quality Assurance. Purity control. Polycrystalline Ingots

Neutrino Physics. Neutron Detector in the Aberdeen Tunnel Underground Laboratory. The Daya Bay Experiment. Significance of θ 13

Simultaneous measurement of 89 Sr and 90 Sr in aqueous samples by liquid scintillation counting using the spectrum unfolding method

Hidex LSCs. LSC 2013 Barcelona

Study of absorption and re-emission processes in a ternary liquid scintillation system *

A Rapid Method for Measuring Strontium-90 Activity in Crops in China

Calibration & Use of a Capintec CAPTUS 3000 Portable Thyroid Uptake System for Iodine-125 Bioassay Measurements Todd W.

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Liquid Scintillator Time Projection Chamber Concept

Detection of 222 Rn and 226 Ra in environmental samples by scintillation method

LS6500 Scintillation Counting Systems

DISMANTLEMENT OF AN UNIT OF METABOLIC TREATMENTS WITH IODINE 131

IMPROVED TECHNIQUES FOR THE ACTIVITY STANDARDIZATION OF 109 Cd BY MEANS OF LIQUID SCINTILLATION SPECTROMETRY

AUTORADIOGRAPHY OF DEUTERATED Ti AND Pd TARGETS FOR SPATIALLY RESOLVED DETECTION OF TRITIUM PRODUCED BY COLD FUSION

Application of positrons in materials research

Preparatory experiments for cold-neutron induced fission studies at IKI

GAMMA RAY SPECTROSCOPY

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM)

Basic hands-on gamma calibration for low activity environmental levels

Journal of Environmental Radioactivity

2100TR Liquid Scintillation Counter

Renewed whole-body counting chamber in STUK

Transcription:

Optimization of low-levellevel LS counter Quantulus 1220 for tritium determination in water samples IvanaJakonić 1, Nataša Todorović 2, JovanaNikolov 2, Ines KrajcarBronić 3, BranislavaTenjović 4, MiroslavVesković 2 1 University of Novi Sad, Faculty of Technical Sciences, Novi Sad, Serbia 2 University of Novi Sad, Faculty of Sciences, Department of Physics, Novi Sad, Serbia 3 Laboratory for measurements of low-level radioactivity, Ruder Boskovic Institute, Zagreb, Croatia 4 University of Novi Sad, Faculty of Sciences, Department of Chemistry, Novi Sad, Serbia

Motivation 3 H lowlevel intheenvironment special conditionshavetobe fulfilled in order to obtain accurate and reliable measurements by liquid scintillation counting(lsc). To optimize tritium analysis in waters by ultra-low background LS spectrometer Quantulus 1220 the influence of sample/scintillant ratio, choice of appropriate scintillation cocktail and comparison of their efficiency, background, minimal detectable activity (MDA) and FOM value, the effect of chemi- and photoluminescence and combination of scintillant/vial was investigated. Development and optimization of new direct method (faster and simpler than ASTM) [Pujol and Sanchez-Cabeza 1999] is presented 3 H level was determined in Danube river samples and several samples within intercomparison with Ruđer Bošković Institute(IRB).

Experimental setup Ultra Low Level Liquid Scintillation Spectrometer Wallac 1220 Quantulus (PerkinElmer, Finland, 2002) with EASYView and WinQ software background reduction system around the vial chamber - an active and passive shield two pulse analysis circuits accessible for the users: a pulse shape analysis(psa) and pulse amplitude comparator(pac) circuit a delayed coincidence circuit(dcos) two multichannel analyzers(mca) quenching was quantified with the external standard quenching parameter, SQP(E) 3 H configuration of the MCA s setting eliminates the random noise of phototubes, inhibits the coincidence pulse from the guard and the sample, and monitors the random coincidence by DCOS in a half of the MCA, the whole sample spectrum beingrecordedintheotherhalf of themca 3 Hwindowwasfixedtochannels1-250

Direct method of 3 H determination Pujol, Ll., Sanchez-Cabeza, J.A., 1999. Optimization of liquid scintillation counting conditions for rapid tritium determination in aqueous samples. J. Radioanal. Nucl. Chem. 242 (2), 391 398. Each water sample was filtrated through a slow depth filter (cellulose nitrate filter pore size 0.45 µm) 8 ml of sample was mixed with 12 ml of the cocktail in a 20 ml vial background preparation: 8 ml of distilled raw water (deep well water) was mixed with 12 ml of the scintillation cocktail. Two organic commercial standards made with 1-pentanol and 1,3-butanediol(Perkin Elmer), tritium activities of 52510 (+/- 3%) dpm/ml (reference date July 9, 2008) and 1.48 10 6 (+/- 3%)dpm/ml (referencedatejuly9, 2008)

Direct method of 3 H determination. Calculations thecountingefficiencyε t : V[ml] sample volume analyzed; T 2 [Bqml -1 ]thestandarddilutionactivity; R s [cps]standardcount-ratesfor 3 Hspectrum; 3 Hactivityconcentration[Bqml -1 ]: R a [cpm]samplecount-ratesfor 3 Hspectrum; R b [cpm] backgroundcount-ratesfor 3 Hspectrum; r q [cpm]count-ratesinchemiluminescencespectrum; The minimum detectable tritium activity (MDA) for a counting time T[min]:

Results and discussion MDA achieved The minimum detectable 3 H activity for total counting timeof 300min~ 2.0 Bq l -1 forbothcocktails 7 MDA 2 7 MDA 3 6 6 5 5 MDA [ Bq l -1 ] 4 3 2 MDA [ Bq l -1 ] 4 3 2 1 1 0 0 0 100 200 300 400 500 600 700 800 900 0 100 200 300 400 500 600 700 800 900 t [min] t [min] OptiPhase HiSafe 2 OptiPhase HiSafe 3

Results and discussion Chemiluminescencestudystudy Quantulus already has a built-in monitor for detection of chemiluminescence events (emission of single photon), Delayed Coincidence Circuitry, DCOS 1000 Counts per min 900 800 700 Cpm Data1Cpm No visible deviation on count rate for active samples during time that would come as consequence of chemical reactions if the sample was counted straightaway after its preparation 10 100 1000 t [min] Despite conclusions that 1 day waiting time is adequate for Quantulus performances, when measuring active samples, it is not necessary in our procedure if we monitor chemiluminescence spectrum

Results and discussion Chemiluminescencestudystudy For samples with very low activity (blank samples) there is some indication for chemical reactions taking place during first 3 hours after its preparation 25 20 chemiluminescence spectrum Cou unt rate [cpm] 15 10 5 0 0 200 400 600 800 t [min] ] Count rate [min] 60 50 40 30 20 40 chemiluminescence spectrum 10 Count rate [cpm] 30 20 10 0 0 200 400 600 800 t [min] 8ml distilled water+12 ml OptiPhase HiSafe2 (up) 8ml distilled water+12 ml OptiPhaseHiSafe3 (down) counted immediately after its preparation 0 0 20 40 60 80 100 t [min] Half-life of chemiluminescence decay estimation: T 1/2 ~ 16.4 min

Results and discussion Photoluminescence study Reactions of photoluminescence due to activation of cocktail by ultraviolet light have a relatively rapid decay rate, these don t affect spectrum after ~2 hours 2200 2000 photo-luminescence 1800 counts per min 1600 1400 1200 1000 800 600 10 100 1000 t [min] Count rates of sample - 8ml tritiated water + 12 ml OptiPhase HiSafe 3, measured immediately after its exposure to sunlight.

Results and discussion Sample/scintillantratio optimization Measurements of samples with different ratios of tritiated water and cocktail (keeping the total volume 20 ml) were performed. The optimal water : cocktail ratio is 9 : 11 ml, but as environmental samples may induce phase separation, we decide to use 8ml sample+12 ml LS cocktail, as suggested by various authors. DWTS count rate [cps] DWTS count rate [cps] 30 30 25 25 count rate [ cps ] 20 15 10 5 count rate [cps] 20 15 10 5 0 0 2 4 6 8 10 12 14 16 18 20 0 2 4 6 8 10 12 14 16 18 V [ ml ] V [ml] OptiPhase HiSafe 2 OptiPhase HiSafe 3

Results and discussion Scintillant/vial combination crosscheck Chemical composition of scintillation cocktails used (manufacturer s Material Safety Data Sheet). The manufacturer of all is PerkinElmer. Cocktail Solvent and additives Scintillator OptiPhase HiSafe 2 DIN > 70% Sodium dioctyl-sulfosuccinate < 14% Poly(ethyleneglycol)mono(4-nonylphenyl)ether < 7% 2- (2-butoxyethoxy)ethanol < 5% N-lauroyl sarcosine < 5% Propylene glycol butyl ether < 3% PPO 2.5% bis-msb 2.5% Sodium borohydride < 1% OptiPhase HiSafe 3 Ultima Gold LLT Diethanolamine < 1% DIN > 60% poly(ethyleneglycol)mono(4-nonylphenyl)ether < 25 30% α-phenyl-ω-hydroxypoly(oxo-1,2-ethanediyl) phosphate < 10% DIN 40-60% Ethoxylated nonylphenol 20-40% 2-(2-butoxyethoxy)ethanol 2.5-10% Ethoxylated fatty alcohol 2.5% Nonylphenyl polyoxyethylene ether phosphate < 2.5% 3,6-dymethyl-4-octyne-3,6-diol 2.5% PPO 2,5 - diphenyloxazole DIN di-isopropylnaphtalene isomers bis-msb 1,4 -bis(2-methylstyryl)-benzene Pseudocumene 1,2,4-trimethylbenzene PPO 1% bis-msb 0.1% PPO 2.5% bis-msb 2.5%

Results and discussion Scintillant/vial combination crosscheck Two vial types: 20 ml glass vials(low potassium borosilicate glass), 20 ml plastic vials(low diffusion polyethylene anti static) Comparison of tritium standard count rates (A=5612 Bq l -1 ) for two different scintillation cocktails and two types of 20ml vials used. Water : scintillation cocktailratiowassetto8:12 ml. Cocktail P l a s t i c v i a l s G l a s s v i a l s cps Efficiency Efficiency FOM cps [%] [%] FOM OptiPhase HiSafe 2 15.75 (21) 35.1 (5) 1080.7 16.33 (29) 36.4 (6) 109.9 OptiPhase HiSafe 3 13.69 (25) 30.5 (6) 816.0 13.81 (26) 30.8 (6) 79.1 Background (and ~blank samples) should be prepared exclusively in plastic vials to avoid natural radioactivity contained inglassof thevial ( 40 K 40 Ar,whichemitsX-rays).

Results and discussion Scintillant/vial combination crosscheck Environmental samples (Danube samples) with low activity expected (close to background). Water:cocktail ratio was set to 8:12 ml. Counting time was 600min. C o c k t a i l OptiPhase HiSafe 2 OptiPhase HiSafe 3 Sample Plastic vials Cpm Glass vials Cpm Plastic vials Cpm Glass vials cpm Blank 1.14 (4) 12.06 (14) 1.14 (4) 12.00 (14) Karlovci 1.68 (5) 12.72 (15) 1.80 (6) 12.66 (15) Oficirac 1.50 (5) 12.42 (15) 1.56 (5) 12.42 (14) Becarac 1.44 (5) 12.66 (15) 1.74 (5) 12.48 (15) Strand 1.68 (5) 12.96 (15) 1.86 (6) 12.66 (15) Sample Karlovci Oficirac Becarac Strand OptiPhase HiSafe 2 OptiPhase HiSafe 3 Activity concentration 2.1 (4) Bq l -1 2.9 (5) Bq l -1 18 (4) TU 24 (4) TU 2.5 (4) Bq l -1 2.9 (4) Bq l -1 21 (4) TU 24 (4) TU < MDA 2.5 (4) Bq l -1 20 (4) TU 3.2 (4) Bq l -1 4.5 (5) Bq l -1 27 (4) TU 38 (4) TU Calculated activity concentrations of Danube samples (plastic vials). Comparison of adequacy and behavior of two cocktails. The upper limit for tritium in drinking water is 100 Bql -1 (European Commission)

Results and discussion Intercomparisonwith IRBfor IAEAsamples Sample *IRB result IAEA ref. IAEA TRIC2008 Intercomp. T17 TRIC2008 Intercomp. T19 14.44±0.40 TU 14.46±0.08 1.70±0.04 Bq l -1 TU OptiPhase HiSafe 3 used OptiPhase HiSafe 2 used NS result z-score NS result z-score <МDА (electrolitic enrichment recommended) <МDА (electrolitic enrichment recommended) IAEA 0.9 9.7 576±13 TU 568.7±2.3 584±17 TU 724±16 TU 68. 0±1.5 Bq l -1 TU 68.9±2.0 Bq l -1 85.5±1.9 Bq l -1 Sample 1 1.365 10 8 TU (±12957) 1.61 10 7 Bq l -1 (±1530) very 3 H active 1.389 10 8 TU (±2768880) 1.639 10 7 Bq l -1 (±326750) (for IAEA ref) 0.47 (for IRB ref) 0.87 1.31 10 8 TU (±1866610) 1.546 10 7 Bq l -1 (±220275) (for IAEA ref) 9.25 (for IRB ref) -2.95 Sample 2 47563±214 TU 5612±25 Bq l -1-47079±1202 TU 5556±142 Bq l -1-0.40 46284±1016 TU 5462±120 Bq l -1-1.26 *Two IAEA samples were measured using OptiPhase HiSafe 3 (for the first one electrolytic enrichment was performed), the other two were measured with Ultima Gold LLT

Conclusions Optimization of detector system in order to apply direct method without distillation led to the following statements: Optimal sample/scintillant ratio- confirmed to be 8:12 ml Chemiluminescence is suppressed 3 hours after sample preparation. T 1/2 ~ 16.4 min MDA (t=300 min )= 2.0 Bq l -1 both for OptiPhase HiSafe 2 and OptiPhase HiSafe 3 Backgrounds and samples with low-level activity close to background are to be put strictly in plastic vials (to avoid natural radioactivity of glass vials which becomes dominant source of radiation)

Conclusions 3 H level of Danube samples is low as expected. Calculated activities obtained with OptiPhase HiSafe 2 are slightly higher compared to the ones measured with OptiPhase HiSafe 3. OptiPhase HiSafe 2 has better FOM values and higher efficiency (35.1±0.5%) than OptiPhase HiSafe 3 but failed during intercomparison with IRB and IAEA. OptiPhase HiSafe 3 achieved modest FOM parameter and worse efficiency(30.5±0.6%) but obtained excellent results in all measurements within intercomparison. Ultima Gold LLT remains the best choice for scintillation cocktail with highest efficiency (up to 56%) giving the mostpreciseresultsasshown inirband IAEAdata.

Acknowledgements The authors acknowledge the financial support of the Ministry of Education, Science and Technological Development of Serbia, within the project No. 43002. Research was conducted at the Laboratory for Nuclear Physics, Department of Physics of the University of Novi Sad DP-UNS (Serbia) and Laboratory for Low-level Radioactivities at IRB (Croatia). Thank you for your attention!