Monte Carlo modelling of a NaI(Tl) scintillator detectors using MCNP simulation code

Similar documents
Response Function of the BGO and NaI(Tl) Detectors Using Monte Carlo Simulations

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data

Radiation Detection and Measurement

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Gamma Spectroscopy. References: Objectives:

Characterization and Monte Carlo simulations for a CLYC detector

Researchers at the University of Missouri-Columbia have designed a triple crystal

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

Chapter 4 Scintillation Detectors

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

Mass of the electron m 0

III. Energy Deposition in the Detector and Spectrum Formation

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

Sample Spectroscopy System Hardware

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate

Determination of the activity of radionuclides

Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods. Abstract

Scintillation Detector

Efficiency. Calculations for Selected Scintillators. > Detector Counting. Efficiency. > Photopeak Efficiency of Various Scintillation Materials

Chemical Engineering 412

MCNP analysis and optimization of a double crystal phoswich detector

Developing DRF for 1 X1 LaBr 3 Detector. Sangkyu Lee

OPPORTUNITY TO JOIN IEEE AND NPSS

Radiation Signals and Signatures in a Detector (Gamma spectroscopy) Sangkyu Lee

THE COMPTON EFFECT Last Revised: January 5, 2007

"Neutron Flux Distribution"

Analysis of γ spectrum

SCINTILLATION DETECTORS AND PM TUBES

An Optimized Gamma-ray Densitometry Tool for Oil Products Determination

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

Comparative Analysis of Nuclear Cross Sections in Monte Carlo Methods for Medical Physics Applications

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices

Experiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado

in Cross-Section Data

MC simulation of a PGNAA system for on-line cement analysis

Compton suppression spectrometry

Slides by: Prof. Abeer Alharbi

ORTEC AN34 Experiment 10 Compton Scattering

Compton Camera. Compton Camera

Calculating standard captured spectra of formation elements

ON THE OMNIPRESENT BACKGROUND GAMMA RADIATION OF THE CONTINUOUS SPECTRUM

Detection and measurement of gamma-radiation by gammaspectroscopy

Nuclear Physics and Astrophysics

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

K 40 activity and Detector Efficiency

arxiv:physics/ v1 3 Aug 2006

Gamma Ray Spectroscopy

EEE4106Z Radiation Interactions & Detection

New Scintillators for the Border Monitoring Equipment

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Gamma-Spectrum Generator

Monte-Carlo Simulation of Response Functions for Natural Gamma-Rays in LaBr 3 Detector System with Complex Borehole Configurations

anti-compton BGO detector

RESULTS AND DISCUSSION ON EFFECT OF COLLIMATOR SIZE ON THE ATTENUATION OF GAMMA RAY IN PURE ELEMENTS

The Best Gamma-Ray Detector

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

DETECTORS. I. Charged Particle Detectors

Monte Carlo simulations as a tool for the development of a new reference ionisation chamber

MCSHAPE: A MONTE CARLO CODE FOR SIMULATION OF POLARIZED PHOTON TRANSPORT

Hands on LUNA: Detector Simulations with Geant4

Determination of Photon Ambient Dose Buildup Factors for Radiological Applications for Points and Plaque Source Configurations Using MCNP5

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

Gamma ray coincidence and angular correlation

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

Figure 1. Decay Scheme for 60Co

COMPTON SCATTERING OF GAMMA RAYS

Modern Physics. Laboratory Experiment. Compton Scattering. Boston University International Program. Technische Universität Dresden

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

Simulation of 4π HPGe Compton-Suppression spectrometer

Radiation Detection. 15 th Annual OSC Readiness Training Program.

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure

Quality Assurance. Purity control. Polycrystalline Ingots

PHYS 3650L - Modern Physics Laboratory

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

Multi Channel Analyzer (MCA) Analyzing a Gamma spectrum

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

GAMMA RAY SPECTROSCOPY

Calibration of EJ-301, a Liquid Scintillator

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Gamma Radiation Absorption Characteristics of Concrete with Components of Different Type Materials

Radionuclide Imaging MII Detection of Nuclear Emission

On the energy deposition by electrons in air and the accurate determination of the air-fluorescence yield

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Semiconductor Detector

MEASUREMENT OF THE D-D NEUTRON GENERATION RATE BY PROTON COUNTING

Validation of the MCNP computational model for neutron flux distribution with the neutron activation analysis measurement

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

Transcription:

Journal of Materials and Environmental Sciences ISSN : 2028-2508 Copyright 2017, University of Mohammed Premier Oujda Morocco J. Mater. Environ. Sci., 2017 Volume 8, Issue 12, Page 4560-4565 http://www.jmaterenvironsci.com Monte Carlo modelling of a NaI(Tl) scintillator detectors using MCNP simulation code Received 22 Sep 2016, Revised 03 Jan 2017, Accepted 13 Jan 2017 Keywords Monte Carlo simulation; MCNP code; NaI detector; Absolute efficiency; Gamma-ray spectrometry A Elanique elanique01@yahoo.com I. Mouhti 1, A. Elanique 1,*, M. Y. Messous 2 1. Faculty of Sciences, Laboratory of Condensed Matter Physics and Nanomaterials for Renewable Energy, BP 8106, University Ibn Zohr, 80000 Agadir, Morocco 2. National Center for Energy, Sciences and Nuclear Techniques -CNESTEN-B.P 1382 R.P 10001, Rabat, Morocco Abstract Due to their attractive performances, inorganic scintillators such as NaI(Tl) detectors are widely used in the field of gamma-ray spectrometry where knowledge of detector efficiency is a key factor for successful quantitative ray analysis. Monte Carlo simulations using MCNP code have been performed in order to evaluate absolute efficiency of NaI detectors. Two mathematical models were constructed for the NaI crystal: 7.62cm x 7.62cm cylindrical NaI and 7.62cm diameter spherical NaI. Detector efficiency was calculated for both NaI detectors in the photon energy range from 60 kev to 2750 kev. Simulation results were compared to experimental efficiency data. 1. Introduction Inorganic scintillators such as NaI(Tl) crystals are widely used for the detection of ionizing radiations, especially gamma rays. This is firstly due to the high atomic number of its iodine constituent (Z=53) which assures that photoelectric absorption will be a relatively important process [1]. Other interesting performances of such attractive material are their simplicity of use at room temperature (no cryogenic system required), robustness and relatively low cost. Knowledge of detector efficiency is absolutely essential when performing quantitative ray analysis using a gamma spectrometry. Detector efficiency is described either as absolute, a (E), or intrinsic, i (E), where E is the absorbed photon energy. Absolute efficiency is the product of ( and the intrinsic efficiency of the detector, whereas ( denotes the fractional solid angle subtended by the detector at point source position [2]. Intrinsic efficiency is the probability that a photon incident on the detector will produce a recorded pulse. For NaI scintillators, i (E) mostly depends on the crystal dimensions, the thickness of Aluminium housing material and other absorbing layers like MgO reflector. The detector efficiency is commonly measured by calibration sources or computed by Monte Carlo (MC) simulations. Since the advent of fast and low-cost personal computers, mathematical modelling of radiation detectors using Monte Carlo (MC) programs is becoming a powerful and convenient tool to fully characterize detector parameters by simulating the transport of radiation through matter. The Monte Carlo methods make it possible to calculate the response function for detectors with good accurate results [3, 4]. In the work reported here, the MCNP code [5], based on the Monte Carlo method, was used to evaluate the photon detection efficiency for a 7.62cm x 7.62cm NaI(Tl) scintillator detector in the 60-2750 kev photon energy range. We simulated two geometrical configurations for the NaI crystal: cylindrical and spherical. In both cases, the detector was exposed to gamma rays from 4 radioactive isotropic point like sources: 241 Am, 137 Cs, 22 Na, 60 Co and 54 Mn. 2. Monte Carlo simulations 2.1. Monte Carlo code The simulations were performed using the Monte Carlo (MC) code MCNP version 4C to calculate the NaI detector efficiency. This code is a general purpose radiation transport code developed at the Los Alamos National Laboratory. It is used to calculate coupled neutron-photon-electron transport. For photons, MCNP accounts for Compton scattering (incoherent) and Thomson scattering (coherent), the possibility of fluorescent Mouhti et al., JMES, 2017, 8 (12), pp. 4560-4564 4560

emission after photoelectric absorption, and pair production with local emission of annihilation radiation, and bremsstrahlung. The code obtains the solution of the problem by simulating individual particle trajectories and recording some aspects of their average behaviour (it does not solve the Boltzmann particle transport equation). The individual probabilistic events that comprise a process of interaction of nuclear particle with material are sequentially simulated. Each particle may generate additional particles (daughters) or tracks at a collision site. Each history (birth to death) is considered an independent random event. MCNP is immediately suitable for modelling the detector response, since it contains a special tally, F8, which is specific for pulse height determination, i.e. energy deposited in the detector. Thus, the pulse-height tally per photon emitted from the source gives the absolute efficiency a (E) for each spectral peak in the chosen geometry. In the MCNP code, the tracking of each photon, as well as each photon and electron resulting from interactions in the detector, is followed until its energy in the detector is low enough for it to be considered killed (1 kev cutoff). 2.2. Accuracy versus precision of the MCNP results To understand the precision and acceptability of the tally results produced by the Monte Carlo run, we should distinguish between precision and accuracy of MC results. The Accuracy is a measure of how close the result is to the true physical quantity being estimated. When running a MC calculation, many factors can significantly affect the accuracy of the results. This includes inaccuracies introduced by MCNP in its use of (1) physics models, (2) uncertainties in the photon cross sections (<2%) over the considered energy range [6] and (3) eventually poor modelling of the actual detector geometry by the MC simulations. Indeed, the detector specifications, generally provided by the manufacturer, are often not sufficient and sometimes need to be optimised by radiographic imaging [3, 4, 7]. By contrast, precision of the MC method is the statistical counting error caused by the statistical fluctuations ( 1 N ), where N is the number of particles histories considered. The more histories run, the better will be the precision of the tallies [8]. In the present work, the number of particles histories simulated in each run was 10 6 so that the precision of the MC results was always kept below 0.2%. MC simulations were performed using MCNP with default physics features enabled and detailed tracking of both photons and electron (i.e. mode p e). 2.3. Modelling the NaI detector Two mathematical models for the NaI detectors were constructed using MCNP4C code: a) cylindrical 7.62cm x 7.62cm model [9] and b) spherical 7.62cm x 7.62cm model [10]. The inner structure of the studied NaI detectors is shown in Figure 1. Figure 1: Cross section of the MCNP detector models. The spherical NaI detector (right) and the cylindrical one (left) are represented. Distances are in mm. The NaI crystal density was 3.667 g.cm -3, the MgO reflector density was 2 g.cm -3 [11] and the aluminium density was 2.7 g.cm -3. The estimated weight of the crystal volume in the spherical model ( 850 g) was in agreement with the value of 847 g from [10]. This confirms the exactness of the actual model used for the spherical detector. To account for backscattering, the window of the PM tube was modelled as being composed from glass (SiO 2 layer), having an effective density of 0.94 g.cm -3. Nevertheless in the present study this layer did not play a Mouhti et al., JMES, 2017, 8 (12), pp. 4560-4564 4561

significant role since it was located in the backside of the NaI crystal. All gamma ray emitting sources were modelled as point like isotropic sources. They were placed 10 cm from the detector surface on the centre axis. The full energy peak in the simulated spectra was treated as a Gaussian peak whose resolution ( E) i.e. Full Width at Half Medium (FWHM) was derived from the measured photopeak [12]. This broadening is required to take into account the statistical fluctuations in each step during pulse creation process. Among these are the following: efficiency of light collection and transfer of photons to photocathode (NaI sensitive volume), charge multiplication in the dynode structure (PM) and pulse amplification (electronics). It should be noted that the MC code does not account for all these effects that would contribute to the observed width. The Gaussian energy broadening (GEB) option was used to simulate the statistical variance in the photopeak centred at energy E (MeV). In MCNP4C code, this broadening is defined by the (FWHM) or E: FWHM = a + b E + ce 2 Where constants a, b and c are equal to -0.0137, 0.0752, and -0.1210 respectively. Figure 2: Pulse height distribution (PHD) of 137 Cs simulated by MCNP for a cylindrical NaI detector. The graph illustrates the effect of Gaussian energy broadening (GEB) option on the full energy peak. The spectral line at 662 kev represents the full energy peak while the peaks centred at 634 kev are attributed to Iodine x-rays escape (see page 331 of [1]). The effect of broadening the photopeak is better illustrated in Figure 2 where the Pulse Height Distribution (PHD) of 137 Cs point source is simulated for a cylindrical 7.62cm x 7.62cm NaI detector with and without GEB option (i.e. GEB option turned on/off). Typically, for a 7.62cm x 7.62cm NaI cylindrical detector the energy resolution (FWHM/E) at 662 kev energy is about 6-7% (see page 347 of [1]). 3. Results and Discussion All MC generated spectra were normalized to the highest measured value in the mean channel of the full energy peak. Results from MCNP4C simulations are shown in Figure 3 which represents the PHD of a cylindrical 7.62cm x 7.62cm NaI detector for 200 kev photons. The obtained spectrum does not consider the effects of GEB (which is the default option in the MCNP code). The spectral line (full energy peak) at 200 kev was reduced by a factor of 100 for better viewing. The peak centred at ~172 kev is attributed to Iodine K x-rays escape from the surface of the NaI crystal [13]. When this occurs, the energy of the x-ray (28 kev) is lost. The Compton continuum can be clearly identified below 70 kev. A comparison between empirical and MCNP4C simulated spectra of the cylindrical NaI(Tl) detector in the case of (a) 137 Cs and (b) 60 Co sources is shown in Figure 4. The plotted measurement data points were reproduced from [9], whereas MCNP4C simulation indicates present work. A good agreement between simulated and measured energy spectra for both point sources can be seen. While for the 137 Cs source the measured and MC calculated spectra match well, for the 60 Co source however we observe a systematic difference in the Compton continuum below 600 kev. This is most likely due to nonproportional scintillation efficiency of the NaI crystals at low energy region [14]. The latter which is defined as the light output per unit deposited energy, is not a constant value and is not considered in the MCNP4C code. Mouhti et al., JMES, 2017, 8 (12), pp. 4560-4564 4562

Figure 3: Pulse height distribution generated by MCNP code for the spherical NaI detector at 200 kev photons. For better viewing, the full energy peak was reduced by a factor of 100. The Iodine K x-rays escape peak (~172 kev [13]) and the Compton edge below (~70 kev) can be clearly identified. Figure 4: Comparison between empirical and simulated spectra of the cylindrical NaI (Tl) detector in the case of (a) 137 Cs and (b) 60 Co sources. The two point isotropic sources were placed on the center axis of the detector at 10 cm from the detector face. Data were from ref. [9], while MCNP calculations correspond to present work. It should be noted that the backscatter peak of Figure 4 in the vicinity of 0.2-0.25 MeV energy region was not reproduced by MC simulations. This peak arises from Compton scattering of gamma rays in the walls of the shield surrounding the detector [15]. The reason is that the actual MC model simulates only the detector-source system rather than the whole experimental setup including the shield. In table 1, MC simulated efficiencies are compared to measured values for both NaI detector models (cylindrical and spherical) in the photon energy range 60 kev-2750 kev. The relative error of the measured efficiencies was reported to be ( 2.5%) in the spherical NaI detector [10]. The uncertainty of the MC results was about 5%. Recall that for the cylindrical NaI detector, uncertainties on efficiency data were not provided in [9]. The results presented in table 1 indicate that the Monte Carlo efficiencies for the studied NaI detectors agreed well with the measured values. Nevertheless, for the cylindrical NaI detector, we observed a significant difference between MC values and efficiency data in particularly for the 60 Co gamma energies (1173 and 1332 kev) and similarly Mouhti et al., JMES, 2017, 8 (12), pp. 4560-4564 4563

for 22 Na photon energy at 1272 kev: The measured efficiencies are about 47% lower than the MCNP values in the 60 Co case and about 10% lower in the 22 Na. The origin of these discrepancies would be attributed to the larger uncertainties in the experimental efficiencies. Indeed, when re-observing the experimental 60 Co spectrum of figure 4 b), the two peaks are poorly separated (at such energy, NaI energy resolution is about 12%) and consequently a bad subtracting of the Compton background from the two peaks results in both a wrong net count as well as an inaccurate efficiency determination. Table 1: NaI(Tl) detector s efficiency values obtained by MC simulation and measurements. Sources were located at 10 cm from the face of the detector. Efficiency data were from [10] for the spherical NaI detector and from [9] for the cylindrical NaI detector. For the spherical NaI detector, efficiency data corresponding to 22 Na, 54 Mn and 241 Am sources were not given (n.g.) Cylindrical NaI detector Spherical NaI detector Source Photon energy (KeV) Measured MCNP Measured MCNP 241 Am 60 n.g. 2.83 n.g. 1.271±0.001 22 Na 510.6 1.431 1.435 1272.7 0.590 0.649 24 Na 2754 n.g. 0.329 n.g. 0.191±0.006 54 Mn 834.7 0.812 0.861 137 Cs 661.7 1.07 1.04 0.684±0.01 0.690±0.02 60 Co 1173.2 0.470 0.698 0.416±0.01 0.415±0.01 1332.5 0.455 0.625 0.369±0.008 0.370±0.02 Conclusion In the present work, Monte Carlo simulations of a two NaI(Tl) detectors using MCNP4C code were performed. Two mathematical models were constructed for the NaI crystal: 7.62cm x 7.62cm cylindrical NaI and 7.62cm diameter spherical NaI. Detector efficiency was calculated for both NaI detectors in the photon energy range from 60 kev to 2750 kev. The resulting overall agreement between MC efficiency values and empirical efficiency values shows that the MCNP4C code can be successfully used for modelling response functions of these two NaI(Tl) scintillators. References 1. Knoll G.F. Radiation detection and measurement, John Wiley & Sons, New York (2000). 2. Johansson S.A.E., Campbell J.L. PIXE A Novel Technique For Elemental Analysis, John Wiley & Sons, New York (1988). 3. Mesradi M., Elanique A., Nourreddine A., Pape A., Raiser D. Appl. Radiat. Isot. 66 (2008) 780. 4. Elanique A., Marzocchi O., Leone D., Hegenbart L., Breustedt B., Oufni L. Appl. Radiat. Isot. 70 (2012) 538. 5. Briesmeister J.F. MCNP: A General Monte Carlo N-Particle Transport Code, Version 4C, LA-13709-M, Los Alamos National Laboratory (2000). 6. Hubbell J.H. Phys. Med. Biol. 44 (1999) R1. 7. Boson J., Agren G., Johansson L. Nucl. Instr. Meth., A 587 (2008) 304. 8. Shultis J. K., Faw R. E. An Mcnp Primer, Kansas state university, USA (2004). 9. Akkurt I., Tekin H.O., Mesbahi A. Acta Physica Polonica A, Vol. 128 (2015) B332. 10. Chul-Young Yi, Suck-Ho Hah Appl. Radiat. Isot. 70 (2012) 2133. 11. Saito K., Moriuchi S. Nucl. Instrum. Methods 185 (1981) 299. 12. Shi H.X., Chen B.X., Li T.Z., Yun D. Appl. Radiat. Isot. 57 (2002) 517. 13. Heath, R.L. Scintillation Spectrometry Gamma-Ray Spectrum Catalogue, IDO-16880, vols 1 and 2 AEC Research and Development Report, Physics (updated Feb. 1997) 11. 14. Zerby C.D., Meyer A., Murray R.B. Nucl. Instrum. Methods 12 (1961) 115. 15. Berger M. J.; Seltzer S. M. Nucl. Instr. Meth.104 (1972) 317. (2017) ; http://www.jmaterenvironsci.com / Mouhti et al., JMES, 2017, 8 (12), pp. 4560-4564 4564