Earlier Lecture. In the earlier lecture, we have seen non metallic sensors like Silicon diode, Cernox and Ruthenium Oxide.

Similar documents
Cryogenic Engineering Prof. M. D. Atrey Department of Mechanical Engineering Indian Institute of Technology, Bombay

Industrial Instrumentation Prof. A. Barua Department of Electrical Engineering Indian Institute of Technology - Kharagpur

Control Engineering BDA30703

Vacuum. Kai Schwarzwälder, Institut für Physik Universität Basel October 6 th 2006

Angular Correlation Experiments

Vacuum Pumps. Two general classes exist: Gas transfer physical removal of matter. Mechanical, diffusion, turbomolecular

DEPOSITION OF THIN TiO 2 FILMS BY DC MAGNETRON SPUTTERING METHOD

SENSORS AND TRANSDUCERS

Generation of vacuum (pumps) and measurements

PhysicsAndMathsTutor.com 1

Sputter Ion Pump (Ion Pump) By Biswajit

14CH406. Chemical Engineering Scheme of Valuation Process Instrumentation Maximum : 60 Marks (1X12 = 12 Marks) Answer ONE question from each unit.

QPR No. 68 RADIO PHYSICS I. PHYSICAL ELECTRONICS. Prof. W. B. Nottingham J. L. Coggins B. L. Blackford L. E. Sprague RESEARCH OBJECTIVES

( KS A ) (1) , vapour, vapor (USA) , saturation vapour pressure. , standard reference conditions for gases. , degree of saturation

INSTRUMENTATION ECE Fourth Semester. Presented By:- Sumit Grover Lect., Deptt. of ECE

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements

I. PHYSICAL ELECTRONICS

Rough (low) vacuum : Medium vacuum : High vacuum (HV) : Ultrahigh vacuum (UHV) :

Previous Lecture. Electron beam lithoghraphy e - Electrons are generated in vacuum. Electron beams propagate in vacuum

AE 3051, Lab #16. Investigation of the Ideal Gas State Equation. By: George P. Burdell. Group E3

CHAPTER 4 THERMAL CONDUCTIVITY AND VISCOSITY MEASUREMENTS

Vacuum. Residual pressure can thwart the best cryogenic design. Each gas molecule collision carries ~kt from the hot exterior to the cold interior.

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Generation of vacuum (pumps): Vacuum (pressure) measurements:

Lecture 2: Zero law of thermodynamics

CHARGED PARTICLES IN FIELDS

Homework 2: Forces on Charged Particles

Lecture 3 Vacuum Science and Technology

ION Pumps for UHV Systems, Synchrotrons & Particle Accelerators. Mauro Audi, Academic, Government & Research Marketing Manager

1 Written and composed by: Prof. Muhammad Ali Malik (M. Phil. Physics), Govt. Degree College, Naushera

Metal Deposition. Filament Evaporation E-beam Evaporation Sputter Deposition

A 200 kev Electrostatic Accelerator

VALLIAMMAI ENGINEERING COLLEGE

EXPERIMENT 5. The Franck-Hertz Experiment (Critical Potentials) Introduction

Discovered by German scientist Johann Hittorf in 1869 and in 1876 named by Eugen Goldstein.

Siddharth Institute of Engineering & Technology

Vacuum II. G. Franchetti CAS - Bilbao. 30/5/2011 G. Franchetti 1

Electric Deflection of Electrons

III B.Tech. II Semester Regular Examinations, April/May INSTRUMENTATION & CONTROL SYSTEMS (Mechanical Engineering) Time: 3 Hours Max Marks: 75

Technology Brief 9: Capacitive Sensors

Strain, Force, and Pressure

D- Charge Exchange Ionizer for the JINR Polarized Ion Source POLARIS

Sometimes cars that would normally travel at 65 km hour 1 may be limited to about 20 km hour 1 by a cyclist.

TECHNISCHE UNIVERSITÄT MÜNCHEN. Vacuum (VAK) Group B412. Patrick Christ and Daniel Biedermann

Transducer. A device to which change or converts physical quantity in a more easily measurable quantity. Transducer. (Input) Sensor.

KCSE PHYSICS PAPER SECTION A (25 marks) Answer all the questions in this section in the spaces provided.

Most matter is electrically neutral; its atoms and molecules have the same number of electrons as protons.

Vacuum techniques (down to 1 K)

STRONG DOUBLE LAYER STRUCTURE IN THERMIONIC VACUUM ARC PLASMA *

Chapter 3: Electric Current And Direct-Current Circuits

Mass of the Electron

PHYSICS HIGHER LEVEL

Chapter 1 The discovery of the electron 1.1 Thermionic emission of electrons

US06CPHY06 Instrumentation and Sensors UNIT 2 Part 2 Pressure Measurements

Warning!! Chapter 5 Gases. Chapter Objectives. Chapter Objectives. Chapter Objectives. Air Pollution

Lab 1: Determination of e/m for the electron

ELECTROMAGNETIC WAVES

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

Temperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale

Stepwise Solution Important Instructions to examiners:

Western Electric A V a c u u m T u b e

Roll No. :... Invigilator s Signature :.. CS/B.Tech (EIE)/SEM-5/EI-501/ INDUSTRIAL INSTRUMENTATION II

e453.eps 1 Change (or the absolute value) in the measured physical variable 2 Change in the sensor property is translated into low-power-level

The Q Machine. 60 cm 198 cm Oven. Plasma. 6 cm 30 cm. 50 cm. Axial. Probe. PUMP End Plate Magnet Coil. Filament Cathode. Radial. Hot Plate.

Superfluid helium and cryogenic noble gases as stopping media for ion catchers Purushothaman, Sivaji

Discharge Cell Design. Some initial explorations, considerations and open questions

Answer all the questions in this section in the spaces provided.

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium?

POLARIZED DEUTERONS AT THE NUCLOTRON 1

Temperature Measurement

DUBLIN INSTITUTE OF TECHNOLOGY Kevin Street, Dublin 8.

UNIT 2 STRAIN MEASURMENTS, FORCE MEASUREMENTS. Therefore, resistance strain gauges are also known as piezo-resistive gauges.

A 3 Vapor pressure of volatile liquids

Biosensors and Instrumentation: Tutorial 2

Lecture 4. Ultrahigh Vacuum Science and Technology

A DETAILED EXAMINATION OF THE PRINCIPLES OF ION GAUGE CALIBRATION

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS International General Certificate of Secondary Education

Huashun Zhang. Ion Sources. With 187 Figures and 26 Tables Э SCIENCE PRESS. Springer

The secondary winding have equal no. of turns. The secondary windings are placed identically on either side of the primary winding.

VACUUM PUMPING METHODS

Vacuum gas dynamics investigation and experimental results on the TRASCO-ADS Windowless Interface

Repetition: Practical Aspects

TMT4320 Nanomaterials November 10 th, Thin films by physical/chemical methods (From chapter 24 and 25)

PHYSICS ORDINARY LEVEL

LECTURE 5 SUMMARY OF KEY IDEAS

Ultra-High Vacuum Technology. Sputter Ion Pumps l/s

Western Electr/c A V a c u u m T u b e. The 301A vacuum tube is designed to supply direct current from an alternating-current supply.

Characterization of the operation of RITs with iodine

The Franck-Hertz Experiment

Study of DC Cylindrical Magnetron by Langmuir Probe

Experiment 6: Franck Hertz Experiment v1.3

BALKAN PHYSICS LETTERS Bogazici University Press 15 November 2016 BPL, 24, , pp , (2016)

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

LYOPHILIZATION. First of Four Lectures. Introduction to Freeze-Drying. Introduction to Freeze-Drying. J. Jeff Schwegman, Ph.D. AB BioTechnologies


PhysicsAndMathsTutor.com

1. Distinguish the important characteristics of instrument that are totally electrical and totally electronic in nature. [16]

Sensors and Actuators Sensors Physics

Outline. 4 Mechanical Sensors Introduction General Mechanical properties Piezoresistivity Piezoresistive Sensors Capacitive sensors Applications

Transducers. ME 3251 Thermal Fluid Systems

Transcription:

41 1

Earlier Lecture In the earlier lecture, we have seen non metallic sensors like Silicon diode, Cernox and Ruthenium Oxide. Silicon diodes have negligible i 2 R losses. Cernox RTDs offer high response time and have low magnetic field induced errors. Some of the sensors used for liquid level monitoring are Dipstick, Hydrostatic gauge, Electric Resistance/Capacitance gauge, Thermodynamic gauge and Superconducting LHe gauge. 2

Outline of the Lecture Topic : Instrumentation in Cryogenics Pressure Measurement Conclusion 3

Thermo physical Properties There are various thermo physical properties that are measured or monitored in Cryogenics. They are Temperature Liquid Level Pressure Mass Flow Rate Viscosity and Density Electrical and Thermal Conductivity In this topic, only the first three properties are covered, which are very important. 4

Pressure Measurement We know that the cryogenic vessels are insulated, closed containers. Besides, temperature and liquid level, pressure is also a vital aspect in Cryogenic Engineering. Pressure measurement is needed To check, whether the level of vacuum is maintained. To monitor the pressure rise inside a container, as there is a continuous heat in leak. 5

Pressure Measurement Pressure measurement in Cryogenics deals with both pressures above atmosphere and pressures below atmosphere (vacuum). In Cryogenic Insulations topic, we have seen that the cryogenics and vacuum go hand in hand. Every cryogenic equipment needs a vacuum gauge for pressure measurement. Various vacuum gauges and their working principles are discussed in this topic. 6

Pressure Measurement As seen in the earlier lecture, the levels of vacuum range from atmosphere to 10 12 mbar or less. For different levels of vacuum, we have different gauges, working on different principles. For example, up to a particular level of vacuum, thermal conductivity gauges are used. Therefore, the choice of a gauge for a particular application or for a particular vacuum level is an important aspect. 7

Pressure Measurement Different pressure/vacuum gauges which could be used are Hydrostatic Gauge McLeod Gauge Diaphragm Gauge Mechanical/Electrical Thermal Conductivity Gauge Thermocouple Gauge Ionization Gauge Thermionic Ionization Gauge Cold Cathode Gauge 8

Hydrostatic Gauge McLeod Hydrostatic Gauge is one of the oldest type of vacuum gauges. It is also called as McLeod Gauge. The schematic of this gauge is as shown in the figure. It works on the principle of Boyle s Law. The gauge consists of a glass U tube, whose left arm has a spherical bulb of a known volume. 9

Hydrostatic Gauge McLeod The right arm is branched into a capillary tube, to monitor the Zero ref. point minute changes in pressure. The capillary is marked with a zero reference point. It culminates back into the right arm as shown in the figure. The lower end of the U tube is connected to a mercury reservoir equipped with a piston. 10

Hydrostatic Gauge McLeod p 1 Initially, the apparatus is filled with mercury up to the indicated level. Let the vacuum pressure to be measured be p 1. It is applied on the right arm as shown in the figure. In this situation, the pressure at any point in the system is p 1. 11

Hydrostatic Gauge McLeod p 1 With the application of piston load, the mercury level in the apparatus rises. When the mercury crosses the junction, a known volume of gas is trapped inside bulb and tube. Let this volume of the gas be V 1 as shown in the figure. Initial Cond. Pressure p Therefore, 1 Volume V 1 12

Hydrostatic Gauge McLeod p 1 With the further application of piston load, the mercury rises to fill up both the arms. The load is applied until the mercury level in the capillary tube reaches the zero reference point. The mercury levels in the arms are adjusted to suit to Applied vacuum in right arm Compressed gas in left arm 13

Hydrostatic Gauge McLeod p 1 In this condition, the volume of the gas in left arm is read directly from the available scale. h That is, the difference in the mercury levels in capillary and left arm represents volume and pressure of gas in left arm. Let a be the cross sectional area of the tube, we have Final Cond. Pressure p 1 +h Volume ah 14

h Hydrostatic Gauge McLeod p 1 Hence, applying the Boyle s Law to the left arm, we have Initial Cond. Pressure p 1 Volume V 1 Final Cond. Pressure p 1 +h Volume ah pv = p V i i f f 1 1 1 ( ) p V = p + h ah p V = p ah + ah 1 1 1 2 15

Hydrostatic Gauge McLeod p 1 Rearranging the above equation, we have h p 1 = 2 ah V ah 1 The term, ah, being very small as compared to V 1, is neglected. We have 2 ah p1 = V Therefore, the pressure p 1 is directly proportional to h 2. 1 16

Hydrostatic Gauge McLeod The advantages are The gauge reading is independent of gas. It serves as a reference standard to calibrate other low pressure gauges. There is no need of any zero error corrections. The disadvantages are The gas should obey the Boyle s law. It does not give a continuous output. 17

Diaphragm p ref Signal Test Pressure Diaphragm Gauge The schematic of a diaphragm gauge is as shown in the figure. It consists of a low stiffness corrugated Teflon diaphragm. On the left side of the diaphragm, a reference pressure (p ref ) is maintained. On the right side, the diaphragm is exposed to the test pressures. 18

Diaphragm p ref Signal Test Pressure Diaphragm Gauge In this device, a deflection is caused by a pressure difference across the corrugated diaphragm. This pressure signal or the deflection is amplified either by mechanical or electrical arrangements, to read the pressure directly. The amount of diaphragm s deflection decides the accuracy and sensitivity of the gauge. 19

Diaphragm p ref Diaphragm Gauge Mech. In a Mechanical Diaphragm Gauge, the diaphragm s deflection is magnified to a mechanical pointer and scale assembly. Mechanical Test Pressure The scale is directly calibrated in terms of pressure for direct reading. The operating range of this gauge is from 1000 to 1 mbar with a good accuracy. 20

Diaphragm Diaphragm Gauge Elec. The schematic of a Capacitance Diaphragm Gauge is as shown. p ref S Signal A D Test Pressure It consists of two capacitance electrodes in the form of Concentric circular disc D Circular annulus A. G These two electrodes are deposited on a ceramic substrate S as shown in the figure. 21

Diaphragm Diaphragm Gauge Elec. These two electrodes are placed in the closed vicinity of an inconel diaphragm. p ref S Signal G A D Test Pressure The circular annulus capacitor is grounded at G. The whole assembly is connected to an AC electrical bridge, in which a change in capacitance is calibrated directly in terms of pressure. 22

Diaphragm Diaphragm Gauge Elec. Therefore, in an Electrical Diaphragm Gauge, the deflection is fed to movable capacitance assembly. p ref S Signal G A D Test Pressure These gauges are more reliable and accurate as compared to the earlier designs. It is important to note that the accuracy and sensitivity of the gauge is independent of the composition of the gas. 23

100 Thermal Conductivity Gauge Residual Gas : N 2 The figure shows the variation of thermal conductivity with residual gas pressure of N 2. 10 The x axis denotes pressure in Torr and y axis denotes logarithm of thermal conductivity. 1 10 5 10 3 1 Torr 10 10 3 10 From the figure, it is clear that for the pressures between 10 to 10-2 Torr, the thermal conductivity decreases. 24

100 Thermal Conductivity Gauge Residual Gas : N 2 This decrease is approximately linear for this pressure range. 10 That is, k gas is directly proportional to the pressure in this range. Hence, Q cond through the gas is directly proportional to the gas pressure in this range. 1 10 5 10 3 1 Torr 10 10 3 10 Pirani Gauge works on the above principle. 25

Thermal Conductivity Gauge In this gauge, a tungsten filament is placed inside the residual gas of Filament which, the vacuum level is to be measured. It is heated to a high temperature by passing an electric current. The temperature of filament, thereby, its resistance, changes with Q cond. The Q cond is a function of k gas, which directly represents the pressure. 26

Thermal Conductivity Gauge This filament is connected to one of the arms of the Wheatstone bridge as Filament shown in the figure. V R f R c D R 1 R 2 With the change in resistance, the equilibrium of the bridge is disturbed at D, which is directly calibrated in terms of pressure. The bridge can either be a constant voltage type or a constant current type. 27

Thermocouple Gauge The Thermocouple Gauge functions on the same principle as that of Pirani Gauge. That is, the effect of residual gas in cooling a heated filament. The change in the temperature of the filament, due to the change in the surrounding gas pressure is measured directly by a very fine thermocouple. The thermocouple is attached at the center of the filament, which represents an average value. 28

Thermocouple Gauge This thermocouple voltage is magnified and it is calibrated to the denote the pressure reading. The operating range of this gauge is between 5 to 10-3 mbar. The applications of thermal conductivity gauges are widely found in Rotary and Sorption pumps. In theses pumps, these gauges are used for a continuous monitoring of backing line pressure. 29

Advantages Some of the advantages of these gauges are A fast response time offered by these gauges. They offer an appropriate solution in case of control applications. These gauges are often preferred due to their robustness. 30

Thermionic Ionization Gauge In an Thermionic Ionization Gauge, the residual gas molecules are ionized using an electron beam. A+ e A + 2e + In the above reaction, we have the following. A is the gas molecule from the residual gas. e - is the ionization electron beam. A + is the ionized gas molecule. 2e - are the electrons in the electric current. This reaction produces, two different types of current. They are I + and I -. 31

Ionization Gauges For the measurement of vacuum pressures, there are two types of Ionization Gauges. They are Thermionic Ionization Gauge Cold Cathode Gauge These gauges operate accurately up to very low pressures, typically, in the order of 10-3 to 10-10 mbar. 32

Thermionic Ionization Gauge The schematic of a Thermionic Ionization Gauge is as shown in the figure. It consists of Filament F Thermionic Filament F Cylindrical open mesh grid G Ion collector C Grid G The Thermionic Filament F emits the electrons to ionize the residual gas. Collector C The mesh grid G traps the electrons to measure electronic current. 33

Thermionic Ionization Gauge The filament is charged with a positive potential of +30 V. The grid is maintained at +180 V. Filament F Grid G Collector C This large positive potential difference is required to accelerate the electrons in least possible time. The ion collector C is earthed, in order to maintain a zero potential. The electrons are emitted from F and are accelerated towards the grid. 34

Thermionic Ionization Gauge A majority of electrons strike the grid. Filament F Grid G Collector C However, a few of the electrons move beyond the grid, due to porosity of the grid and high velocity of electrons. These electrons enter a region of decelerating field in between mesh grid G and collector C. 35

Thermionic Ionization Gauge They oscillate back and forth and are eventually collected on the grid. Filament F Grid G Collector C During this phase, the electrons have a maximum probability to hit the residual gas molecules, which produce ionic current. This ionization current represents the ions in residual gas. This is directly calibrated to read the gas pressure. 36

Thermionic Ionization Gauge These gauges are used from 10-3 to 10-7 mbar. Filament F Grid G Collector C The advantages of these gauges are It offers a high reliability and ease of operation. It can be easily degassed by electron bombardment (Grid power = 35 W). These gauges offer a linear calibration current and pressure. 37

Thermionic Ionization Gauge The disadvantages are The use of hot filament increases the risk of burring out, when exposed to atmospheric air. Filament F Grid G Collector C As a result, an extra filament is provided as a standby. The chemical reaction within the residual gas at high temperatures, produces undesirable gases. 38

Cold Cathode Ionization Gauge As mentioned in the earlier slide, the Thermionic Gauges exhibit a risk of burning out of hot filament. B G Anode Cathode This led to the development of Cold Cathode Ionization Gauges. These are also called as Penning Gauges, which are widely used at very low pressures. The schematic of Penning Gauge is as shown in the figure. 39

Cold Cathode Ionization Gauge It consists of an anode ring as shown in the figure. G Anode Cathode It is placed between the two symmetrical cathode plates. The cathode plates are grounded at G. The anode is charged with a potential difference of 2 kv. 40

Cold Cathode Ionization Gauge An axial magnetic field of about 0.05 T is maintained across the entire setup. B G Anode Very often, permanent magnets are used to provide this field. Cathode 41

Cold Cathode Ionization Gauge The combined crossed electric and magnetic fields increase the length of travel of electrons. B G Anode This increases the probability of ionization. The electrons are finally collected at anode. Cathode This ionization current represents the ions in residual gas. This is directly calibrated to read the gas pressure. 42

Cold Cathode Ionization Gauge This gauge is widely used for many scientific applications in the range of 10-3 to 10-7 mbar. B G Anode Cathode The advantages of this gauge are It is very robust No thermionic filament No thermal radiation. The disadvantage of this gauge is It is normally less accurate than the thermionic gauge. 43

Conclusion In Cryogenics, there is a need to monitor various properties like pressure, temperature, liquid level, etc. for safe operation. Thermocouple works on Seebeck effect. PT 100, PT 1000 are some of the commonly used RTDs in Cryogenics. Some of the commonly used non metallic sensors are Silicon diode, Cernox and Ruthenium Oxide. 44

Conclusion Sensors used to monitor liquid level are Dipstick, Hydrostatic gauge, Electric Resistance/Capacitance level gauge, Thermodynamic level gauge and Superconducting LHe level gauge. Different pressure/vacuum gauges used are McLeod Gauge, Diaphragm Gauge, Thermal Conductivity Gauge, Thermocouple Gauge, Thermionic Ionization Gauge and Cold Cathode Gauge. Pirani and Penning gauges are used for high vacuum levels. For less vacuum levels, other gauges are used. 45

A self assessment exercise is given after this slide. Kindly asses yourself for this lecture. 46

Self Assessment 1. The hydrostatic gauge works on. 2. In McLeod Gauge, piston load is applied until the mercury level reaches. 3. In a diaphragm gauge, diaphragm is made of. 4. In a Capacitance diaphragm gauge, deflections alters. 5. In a Pirani Gauge, k gas is proportional to. 6. In a Thermocouple Gauge, temperature of filament is an indication of. 7. In an Ionization Gauge, the filament the electrons. 8. In a cold cathode gauge, the is calibrated to read gas pressure. 47

1. Boyle s Law 2. Zero reference point 3. Teflon 4. Capacitance 5. Gas pressure 6. Gas pressure 7. Emits 8. Ionization current Answers 48

Thank You! 49