Inform is a series of white papers designed to provide advice on material characterization issues. Mie theory The rst 100 years

Similar documents
10 REASONS THAT THE ZETASIZER NANO IS CHOSEN FOR SCIENTIFIC SUCCESS

Morphologically directed chemical identification Coupling particle characterization by morphological imaging with Raman spectroscopy

next generation Welcome to the Dispersion options Smarter particle sizing Your sample, expertly processed and prepared Particle size

ZS90. The ultimate in desktop particle characterization. Particle size. Zeta potential. Molecular weight

Component specific particle characterisation of the active components in a pharmaceutical topical formulations

GPC/SEC Analysis of Polymer Degradation

Determining cure profile and post-cure shrinkage of photopolymers using UV accessory on a rotational rheometer

Discriminating between polymorphs of acetaminophen using Morphologically-Directed Raman Spectroscopy (MDRS)

The ultimate in desktop particle characterization

PARTICLE SIZE CHARACTERISATION OF AGRICULTURAL SPRAYS USING LASER DIFFRACTION

The power of multi-detector SEC in the analysis of antibodies

Method validation for laser diffraction measurements

The Diffusion Barrier Technique, Practical Aspects and Data interpretation

Introduction to the calculators in the Zetasizer software

Overcoming and quantifying Wall Slip in measurements made on a rotational rheometer

The ultimate in desktop particle characterization

Nanobubble Applications and Characterization by Nanoparticle Tracking Analysis

How to use GPC/SEC for compositional analysis

Measurements of protein electrophoretic mobility using the Zetasizer Nano ZSP

GPC/SEC Analysis of Polymer Solutions Used in Inkjet Printing

Optimization of PLA/PLGA molecular weight and structure to suit final application requirements

Studying the Effect of Steric Layer Thickness on Emulsion Stability

Measuring the flow properties of powders. FT4 Powder Rheometer. freemantechnology

PARTICLE SIZE ANALYTICAL RANGES AND APPLICABILITY. m mm (10-6 m) nm (10-9 m)

DEVELOPING A BIOFORMULATION STABILITY PROFILE

Using the Zetasizer Nano to optimize formulation stability

Differentiation of polymer branching and composition using the Mark Houwink plot

How switching mobile phases can improve your OMNISEC results

Evaluating the rheological properties of hyaluronic acid hydrogels for dermal filler applications

Understanding the colloidal stability of protein therapeutics using dynamic light scattering

Using rheology to study the hardness and spreadability of butter

Laser diffraction and automated imaging: complementary techniques for nasal spray development

Pigment Particle Size Using Microtrac Laser Technology

Sample measurements to demonstrate Zetasizer specifications (Zetasizer Nano, Zetasizer APS, Zetasizer μv)

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20

REVERSE ENGINEERING PHARMACEUTICAL FORMULATION USING THE MORPHOLOGI G3-ID.

Particle Characterization of Pharmaceutical Products by Dynamic Image Analysis

Use of DLS/Raman to study the thermal unfolding process of lysozyme

MASTERSIZER 3000 SMARTER PARTICLE SIZING

STABILITY OF PIGMENT INKJET INKS

MicroBrook Eyetech. Particle Size, Shape, and Concentration Analyzer. Particle Sizing. Particle Concentration. Particle Shape

Characterization of protein aggregates in suspension and on a filter membrane by Morphologically-Directed Raman Spectroscopy

Particle Size Determination with Dynamic Light Scattering DLS

MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

CH5716 Processing of Materials

Electrophoretic Light Scattering Overview

A Basic Guide to INFORM WHITE PAPER. Abstract:

Executive Summary. Introduction WHITEPAPER

Characterization of Single Wall Carbon Nanotubes (SWNTs) by combining Dynamic Light Scattering and Raman Spectroscopy

Measurement of Particle Size Distributions

Measuring the rheology of thermoplastic polymer melts

A basic guide to particle characterization

The Elzone II Particle Count and Size Analyzer

Method Development. Creating the Perfect Standard Operating Procedure (SOP) 2007 HORIBA, Ltd. All rights reserved.

Philip E. Plantz, PhD Application Note. SL-AN-36 Revision A. Provided By: Microtrac, Inc. Particle Size Measuring Instrumentation

KINEXUS SERIES REDEFINING RHEOMETER CAPABILITIES

KINEXUS DSR SERIES REDEFINING RHEOMETER CAPABILITIES FOR ASPHALT TESTING

FRITSCH NANO PARTICLE SIZER IDEAL FOR PIGMENTS CHEMICALS PETROCHEMISTRY FOODSTUFFS PHARMACEUTICALS ENVIRONMENT COSMETICS DYNAMIC LIGHT SCATTERING DLS

Zeta potential - An introduction in 30 minutes

Machine Learning Using Machine Learning to Improve Demand Response Forecasts

Setting Attainable and Practical Particle Size Specifications

KINEXUS SERIES REDEFINING RHEOMETER CAPABILITIES

HIGH PERFORMANCE PARTICLE SIZER THE FIRST SYSTEM TO COMBINE HIGH SENSITIVITY WITH A HIGH CONCENTRATION CAPABILITY

429 LIGHT DIFFRACTION MEASUREMENT OF PARTICLE SIZE

Your partner in Science!

A comparative study of particle size analysis in fine powder: The effect of a polycomponent particulate system

The solution for all of your

MASTERSIZER 3000 SMARTER PARTICLE SIZING

Bohlin. Rheological instruments backed with rheological experience. Rheological properties

MASTERSIZER 3000 SMARTER PARTICLE SIZING

Particle size analysis -Chapter 3

Thermal Imaging Best Practices: The Measuring Object. Page 2

NanoParticle Tracking Analysis The NANOSIGHT system.

Scattering by Groups of Particles

Zeta Potential Analysis using Z-NTA

FRITSCH "analysette 22" NanoTec The Result of 30 Years of Experience in Particle Size Measurement. 1 The Inventor of "Reverse Fourier Optics"

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers

Introducing the Morphologi G3 ID The future of particle characterization

IN QUALITATIVE ANALYSIS,

How to measure the zeta potential of individual Nanoparticles (NPs)? To anticipate the answer: With the

CPS Instruments Europe P.O. Box 180, NL-4900 AD Oosterhout, The Netherlands T: +31 (0) F: +31 (0) E:

BGD 611 Falling Rod Viscometer

Introduction to Fourier Transform Infrared Spectroscopy

OPTIMAL WAVELENGTH SELECTION ALGORITHM OF NON-SPHERICAL PARTICLE SIZE DISTRIBUTION BASED ON THE LIGHT EXTINCTION DATA

White Paper Optimization of Dry Powder Particle Size Analysis

Introduction to Fourier Transform Infrared Spectroscopy

Protocol Particle size distribution by centrifugal sedimentation (CPS)

Fundamentals of Particle Counting

F r a U n h o F E r I n s T I T U T E F o r C h E M I C a l T E C h n o l o g y I C T EnErgEtic MatErials ParticlE technology

PhysioCare Concept improved. The feeling of a new philosophy

ISO INTERNATIONAL STANDARD. Particle size analysis Laser diffraction methods. Analyse granulométrique Méthodes par diffraction laser

ENVR 116 Aerosol Technology Laboratory Session Fall 2005

ASSESSMENT. Industry Solutions Harness the Power of GIS for Property Assessment

Perseverance. Experimentation. Knowledge.

Humidity Calibration Solutions

PhysioCare Concept improved. The feeling of a new philosophy

ZETASIZER NANO Series PERFORMANCE, SIMPLICITY, VERSATILITY

BIO & PHARMA ANALYTICAL TECHNIQUES. Chapter 5 Particle Size Analysis

Application Note # AN112 Failure analysis of packaging materials

PTG-NIR Powder Characterisation System (not yet released for selling to end users)

Transcription:

Inform is a series of white papers designed to provide advice on material characterization issues Mie theory The rst 100 years

Mie theory the rst 100 years One hundred years ago Gustav Mie developed an analytical solution for the equations derived by Maxwell to describe the scattering of light by spherical particles. The resulting Mie theory has various uses, one of the most important industrial applications being laser diffraction particle size analysis. Laser diffraction is a well-established particle sizing technique, widely used across a range of manufacturing sectors for the measurement of many different materials. With this method, an optical model unlocks the diffraction pattern produced as light is transmitted through a sample, generating particle size information. Measurement accuracy and range is directly dependent on the model selected. As Mie theory celebrates its centenary it is timely to refl ect on its role in the application of laser diffraction particle size analysis and the benefi ts it brings. Choosing an optical model for laser diffraction RI Contrast Incident light Absorption RI Contrast Figure 1 : Image showing light scattering by a particle Light transmitted through a wet or dry suspension is scattered by the particles in it. Generally speaking, smaller particles scatter light at relatively low intensities to wide angles while larger particles scatter light more strongly at narrow angles. Laser diffraction analyzers capture the diffraction pattern produced by this scattering and then use an optical model to derive the size distribution of the particles that produced it.

Mie theory is based on Maxwell s electromagnetic fi eld equations and predicts the scattering intensity induced by all particles within the measurement range, whether they are transparent or opaque. It is based on the following assumptions: The particles being measured are spherical. The suspension is dilute, so that light is scattered by one particle and detected before it interacts with other particles. The optical properties of the particles, and the medium surrounding them, are known. The particles are homogeneous. Mie theory uses the refractive index difference between the particle and the dispersing medium to predict the intensity of the scattered light. It also describes how the absorption characteristics of the particle affect the amount of light which is transmitted through the particle and either absorbed or refracted. This capability to account for the impact of light refraction within the particle is especially important for particles of less than 50μm in diameter and/or those that are transparent. In the late 1970 s, when laser diffraction systems were fi rst introduced, limited computing power made it diffi cult, and impractical, to rigorously apply Mie theory. The Fraunhofer approximation of the Mie theory was a much easier model to use and was therefore widely adopted at this stage. It provides a simpler approach by additionally assuming that: The particles being measured are opaque discs. Light is scattered only at narrow angles. Particles of all sizes scatter light with the same effi ciency. The refractive index difference between the particle and surrounding medium is infi nite. These approximations hold for large particles (>50μm) but increasingly they break down as particle size decreases. Fraunhofer is particularly inaccurate below 2μm and also fails to properly characterize systems containing transparent particles. When the refractive index difference between the particles and the surrounding medium is low, inaccuracies also tend to increase. An important point to recognise is that Fraunhofer may inaccurately predict the size of particles or the amount of material within a given size range, or both. The nature and extent of any inaccuracies is not predictable. Accurate characterization of the overall particle size distribution demands precise determination of both the size and quantity of material present. The later development of more powerful computers eliminated any

need for compromise, enabling application of the more rigorous and comprehensive Mie theory in its entirety. In very many cases this resulted in much improved measurement accuracy. However, the simplicity of Fraunhofer remains a major attraction, particularly the fact that refractive index data are not required. Today, both models are in use but the ISO standard for laser diffraction particle size analysis, ISO13320, recognises the fundamental superiority of Mie. Mie theory is recommended for analysis across the complete measurement range (0.1 to 3000 μm) while the use of Fraunhofer is limited to systems containing particles that are 50μm or more in size. Instruments that routinely measure particles smaller than this should therefore now support Mie. Modern analyzers, such as the Mastersizer 2000 from Malvern Instruments, make it easier to access the accuracy of Mie through the inclusion of an extensive database of refractive indices within the software. Long term improvements to the user interface have also enhanced the transparency of the calculation process, demystifying application of the theory. The following industrial examples highlight the performance offered by these systems. Measurement accuracy: sizing calcium carbonate The particle size distribution of calcium carbonate, used as a fi ller in paper production, is important because within a narrow particle size band there is an increase in optical scattering effi ciency that enhances whiteness. Measurements of a calcium carbonate sample made using Mie and Fraunhofer are shown in Figure 2. Fraunhofer measures the particle size distribution as larger than it is. This error arises from the assumption that the scattering effi ciency of particles is the same for all particle sizes. In reality, scattering power decreases rapidly below 0.5μm, so the amount of light scattered from fractions in this size range is very small. When using the Fraunhofer approximation, the scattering observed for these fi ne particles is assumed to be insignifi cant, hence the reported volume distribution ends at 0.3μm. Mie Theory, in contrast, predicts the changes in scattering effi ciency which occur at small particle sizes, and therefore correctly interprets the low intensity scattering

observed from particles in this range as being from a signifi cant volume of particles. This enables the true extent of the distribution to be shown, allowing accurate production control during calcium carbonate milling. Figure 2 : Measurement of calcium carbonate using Mie and Fraunhofer Broad dynamic range: tracking ink milling The very broad dynamic range accessible through the use of Mie extends the applicability of laser diffraction and is particularly useful when tracking dramatic changes in particle size, such those which occur during milling. Pigment-based inks for ink jet printers are increasingly common, especially for applications where there is exposure to moisture or a demand for high durability. For these suspensions, wet milling to a particle size of 500 nm or less prevents problems such as blockage of the printer nozzle, defect formation within the printed fi lm and inconsistent colour density. Dispersing the pigment in a suitable mobile phase breaks up lumps and produces a pre-mix which is then milled to a suitable endpoint (see fi gure 3).

Figure 3: Data collected during milling using the Mastersizer 2000 laser diffraction system. Dv10 (particle size below which 10% of the volume particles exists), Dv50 and Dv90 are shown These results demonstrate the suitability of Mie-based laser diffraction measurements for monitoring the complete milling process, and identifying the endpoint. After 180 minutes the suspended pigment has a narrow particle size distribution, with a mode of 137nm and a volume median diameter of around 250nm, and should therefore give problem-free operation. Fraunhofer is not suitable for sizing the fi nal product, yielding a median particle size of ca. 800nm instead of 250nm, thereby failing to detect the endpoint. Accurate Volume Response: measuring blended materials In the manufacture of refractory products a suspension is either pumped into a mould or coated onto a form with the required shape. Suspensions with high solids loadings are benefi cial in terms of fi nished product strength (post sintering), but may have high viscosity so making handling more problematic. Controlling particle size distribution is one way of optimising

suspension properties since increasing polydispersivity allows solid loadings to be raised without increasing viscosity. Figure 4: Size distributions of mixed refractory samples Often manufacturers mix materials with different size distributions in order to control the particle packing density and obtain the desired suspension properties. Figure 4 shows data for samples of refractory materials made of up coarse and fi ne material fractions. In this study the aim was to determine the amount of coarse fraction in an unknown sample, in order to verify the suitability of laser diffraction for detecting changes in the mixed sample. Using Mie theory, the laser diffraction technique was shown to correctly respond to changes in the mass of coarse particles within the product. The trends observed enable the amount of coarse material within the unknown sample to be identifi ed as 21.6%; this compares with the real fi gure of 22%. Such accuracy was not possible using the Fraunhofer approximation, because of its inability to account for refraction of light within the particle phase (the particles were transparent in this case). Fraunhofer overestimated the fi ne fraction (fi gure 5), leading to an incorrect assessment of the sample mix.

Figure 5: Mie and Fraunhofer results obtained for unknown sample mix shown in fi gure 4 In conclusion Advances in computing power allow modern laser diffraction-based particle analyzers to fully exploit the description of light scattering developed by Mie 100 years ago. The examples included here demonstrate how the ability of Mie theory to correctly predict the effect of particle transparency and changes in scattering effi ciency make it superior to the Fraunhofer approximation, particularly for particles less than 50μm in diameter. ISO13320 recognises these benefi ts, concluding that the Mie theory provides an appropriate optical model across the full laser diffraction measurement range. Modern measurement systems enable easier access to the powerful capabilities of the Mie theory through the inclusion of, for example, a database of refractive indices. These systems provide the greatest accuracy for the widest possible range of materials.

Mastersizer 2000 The Mastersizer 2000 is a laser diffraction-based system from Malvern Instruments that uses Mie theory to provide particle size measurement for many different materials, both wet and dry. It offers a broad measurement range, 0.02μm to 2000μm, and exemplary accuracy of +/- 1%. Full automation and software-driven Standard Operating Procedures eliminate user variability and streamline analysis, making measurement a simple and routine task. A wide range of plug and play dispersion cassettes ease the switchover from one sample type to another and optimise fl exibility.

We hope you find Mie Theory - The first 100 years useful. This is one in a series of publications designed to help decision makers in your industry make more informed choices. Malvern Instruments Limited Grovewood Road Malvern Worcestershire UK WR14 1XZ Telephone: +44 (0)1684 892456 Facsimile: +44 (0)1684 892789 Malvern Instruments Worldwide Sales and service centres in over 50 countries. For details visit www.malvern.com/contact Advanced technology made simple distributor details Malvern Instruments is part of Spectris plc, the Precision Instrumentation and Controls Company. All information supplied within is correct at time of publication, Malvern and the green hills logo are international Trade Marks owned by Malvern Instruments Ltd. 2009 MRK1304-01