Precise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced

Similar documents
Multipette / Repeater M4 allows for fast, precise and sterile liquid transfer in cell culture

Technical Report Exact dispensing of whole blood using the Eppendorf positive displacement system Multipette Xstream /Repeater Xstream

N) justment. Register your instrument! Eppendorf Research plus. Factory Adjustment

December 2010 No 036. The electronic Eppendorf Xplorer pipette Versatile adjustment

Spa & Wellness for Pipettes

Superior well-to-well Consistency with Eppendorf Cell Culture Plates

Eppendorf ept.i.p.s. LoRetention Comparison of low retention pipette tips via simple and fast absorbance measurements in the BioPhotometer plus

96 Well? Done! Eppendorf epmotion 96 semi-automated electronic pipette for fast and precise 96 channel microplate processing

ept.i.p.s. pipette tips are available with various degrees of purity, with and without filters, and as special tips.

The New Legend. Eppendorf Reference 2 Pipette

Well within Reach. The Eppendorf Pipette Holder System

Simply Better Pipetting

Spa & Wellness for Pipettes

The New Legend. Eppendorf Reference 2 Pipette

The Pure Truth. Eppendorf Forensic DNA Grade according to ISO 18385

N) manual. Biomaster Operating manual

1 Product description. Material. 3 Warnings for intended use. Instructions for use ept.i.p.s. Version B. English (EN)

Get the Eppendorf Advantage!

Applications Note 228 September 2010

Larger Display Easier to Use. Generation E. Repeater E3 and E3x: Dispense without compromising accuracy or precision

ept.i.p.s. Instructions for use Version B

Eppendorf Xplorer Eppendorf Xplorer plus. Ordering information. Register your instrument!

NEW! ep Dualfilter T.I.P.S. Peak of perfection

The New Eppendorf X-Drive - Maximum Performance, Flexibility and Longevity for Demanding Shaker Tasks

Simply better pipetting!

Cleaning and Decontamination of the ept.i.p.s. Box

PhysioCare Concept improved. The feeling of a new philosophy

PhysioCare Concept. Make your Lab a better place

Automated purification of high quality genomic DNA

Raise the limits. Eppendorf Research plus

PhysioCare Concept improved. The feeling of a new philosophy

No 022 December Application-Specific Requirements for Consumables in the Laboratory

Automated Illumina TruSeq DNA PCR-Free library construction with the epmotion 5075t/TMX

Eppendorf PCR Consumables Compatibility Guide for PCR and qpcr Cyclers

Eppendorf Reference : The pipette.

ept.i.p.s. Version A

Now also in Eppendorf LoBind. Conical Competence. Eppendorf Conical Tubes 15 ml and 50 ml

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

Automation of Routine Microplate Pipetting and Dispensing

NGS Made Easy. Optimize your NGS library preparation with the epmotion Automated liquid handling system

Promotion period: September 15 - December 31, Super Sale. Quality at unbeatable prices

Easy Illumina Nextera DNA FLEX Library Preparation using the epmotion 5075t automated liquid handler

Comparative Analysis of Protein Recovery Rates in Eppendorf LoBind and Other Low Binding Tubes

Electrical safety testing Checklist

Applications Note 164 October 2007

stance et Helper Eppendorf Pipet Helper Chemical Resistance

Technology that counts

Eppendorf UVette. Instructions for use

The Plus for your lab!

Comparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011

Automated pipetting made easy new ep Motion 5073 family members for perfect PCR set-up and nucleic acids purifi cation or general liquid handling

CHEM 334 Quantitative Analysis Laboratory

Technology that counts

Eppendorf BioSpectrometer 6135/6136/6137

Automated Illumina TruSeq Stranded Total RNA library construction with the epmotion 5075t/TMX

New Promotions July 1 December 31, Take Good Care of Your Cells. More great offers for your cell biology processes

Technology that counts

Faster Isolation of PBMC Using Ficoll-Paque Plus in the Eppendorf Multipurpose Benchtop Centrifuges 5920 R and 5910 R

Eppendorf SmartBlock. Instructions for use

NEW: 5.0 ml screw cap tube. The Missing Link. Eppendorf Tubes 5.0 ml Discover a new sample handling system

Automated Illumina TruSight HLA v2 Sequencing Panel Library Preparation with the epmotion 5075t

Eppendorf Vis Cuvettes. Instructions for use

ISO INTERNATIONAL STANDARD. Piston-operated volumetric apparatus Part 7: Non-gravimetric methods for the assessment of equipment performance

Metrological Methods for Low Volume Liquid Handling in Drug Delivery and In-Vitro Diagnostics

Automated Illumina TruSeq Stranded mrna library construction with the epmotion 5075t/TMX

Natural Winners. Eppendorf Liquid Handling Instruments: Unrivaled in technology, efficiency and handling for over 50 years

High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC

NEW! Your bacteria s favorite! Eppendorf Eporator

Peak of Perfection. Made for a faster and easier daily routine: Eppendorf Liquid Handling Consumables

Natural Winners. Eppendorf Liquid Handling Instruments: Unrivaled in technology, efficiency and handling for over 50 years

The Next Stage. The new Mastercycler X50

Micropipetting Basics

HPLC Sample Preparation

Applications Note 215 September 2009

The first address for Titration. TITRONIC 300 and TitroLine 5000

Application Note. Automated pipetting and analysis of polymer suspensions for process control in chemical production

LAB 15. Traceability: Volumetric Apparatus. Edition 3 February 2019 Draft for Consultation. LAB 15 / Issue: 3 Page 1 of 12

Food & Beverage by Eppendorf

cobas c 111 analyzer Big analyzer performance in a small package

Technology that counts

Applications Note 202 October 2008

Still Making History. Over 50 years of pipetting

Evaluating the functionality of Eppendorf BioPhotometer and Eppendorf BioSpectrometer using a Secondary UV-VIS Filter Set

Practice Lab. Balances and calibration of volumetric tools

CHEMISTRY WORKSTATION. The power of proven quality GIVING YOU PEACE OF MIND

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 1

Eppendorf Plate Deepwell 96 and 384: RecoverMax

Quad-Z 215 with Disposable Tips: Implementing the Use of ZipTips for the Purification of Biological Samples

Growth in Motion. The next evolution in biological shakers the Innova S44i

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Catalog Liquid Handling, Sample Handling, Cell Handling

Certificate of Accreditation

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

DELFIA Thyroxine (T 4 )

CLINICAL CHEMISTRY ANALYSER. The economical solution for all your needs GIVING YOU PEACE OF MIND

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Modified Adams Assay for Phenolics in Wine

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities

Transcription:

APPLICATION NOTE No. 386 Precise and Accurate whole Blood Dispensing with Multipette E3x and Combitips advanced Vincent Dufey¹, Muriel Art¹, Hanaë Henke² ¹Eppendorf Application Technologies S.A., Namur, Belgium ²Eppendorf AG, Hamburg, Germany 1 Eppendorf AG, Hamburg, Germany; 2 Eppendorf Application Technologies SA, Namur, Belgium Abstract Dispensing of whole blood samples is challenging and needs careful consideration of the appropriate pipetting tool. We showed that using a positive displacement system, such as the Eppendorf, is suitable for dispensing whole blood samples. Compared to an electronic air cushion pipette the results obtained with the were accurate and precise. Furthermore, the and Combitips advanced system was compared to two competitor instruments. This evaluation led to the conclusion, that one competitor instrument already failed to meet the supplier s specifications when dispensing water. Therefore, only the second competitor instrument was tested with whole blood. This instrument showed deviation from the manufacturers precision limits. The only instrument within supplier s specifications when dispensing whole blood was the. Introduction During the past 40 years, liquid delivery by pipetting became a common laboratory process. To fulfill this routine task, the researcher can select the right device among instruments operating according to two different dispensing principles: via air cushion or positive displacement [1]. In both cases liquid transfer is realized by a piston that moves within a cylinder. In an air cushion pipette, air is present between the aspirated sample and the piston inside the pipette. In a positive displacement instrument the piston is directly integrated in the consumable. Consequently, no compressible air cushion is present between the piston and the sample liquid in those instruments. In this case, the sample is in direct contact with the piston like in a syringe. The presence of this air cushion clearly differentiates both kind of instruments. Being elastic, the air cushion can be significantly influenced by environmental parameters (relative humidity, room temperature, air pressure) as well as by physical properties of the pipetted liquid (density, viscosity, liquid temperature, vapor pressure) [2]. For this reason, and especially when accurate pipetting is crucial for the application, air cushion instruments should only be used for handling aqueous solutions. Having no elastic air cushion, a positive displacement pipette is not affected by the characteristics of the sample or laboratory conditions. Therefore, those instruments are ideal for dispensing all complex liquids having physical properties different than water.

APPLICATION NOTE I No. 386 I Page 2 Whole blood is particularly complex. This biological fluid is composed of plasma and cells (red blood cells, white blood cells and platelets). Whole blood also contains a significant amount of proteins like albumin and globulins. All those components make blood denser and more viscous than water. As several applications (forensic analysis, blood cell culture, RNA/DNA extraction, etc.) require exact and reproducible blood pipetting, the appropriate dispensing system has to be carefully selected. In this study, we compared an electronic positive displacement instrument, the, to the Eppendorf Xplorer plus, a more commonly used electronic air cushion pipette. We demonstrated that a positive displacement system is the most appropriate solution to handle whole blood accurately and precisely. The Eppendorf dispenser was also compared to two major competitors pointing out the impact of high quality instruments and excellent instrumentconsumable association on the application. Materials and Methods Materials Instruments > Eppendorf, electronic pipette, 50 1000 µl (order no. 4861 000.732) > Eppendorf, electronic pipette, 0.5 10 ml (order no. 4862 000.759) > Eppendorf, (order no. 4987 000.029) > Competitor B, electronic dispenser > Competitor R, electronic dispenser Consumables > Eppendorf, ept.i.p.s. standard 50 1000 µl (order no. 0030 000.927) > Eppendorf, ep T.I.P.S. standard 0.5 10 ml (order no. 0030 000.765) > Eppendorf, Combitips advanced 1 ml standard (order no. 0030 089.430) > Eppendorf, Combitips advanced 10 ml standard (order no. 0030 089.464) > Competitor B, dispenser tips, 1 ml and 10 ml > Competitor R, dispenser tips, 1 ml and 10 ml Instrument Calibration > Mettler-Toledo, micro balance Excellence plus XP26PC (Mettler-Toledo, Ohio, USA; order no. 11106021) > Sartorius, Analytical balance Cubis MSE224S-100-DA (Sartorius, Göttingen, Germany; order no. MSE-224S-100-DA) > VWR, Water Molecular Biology Grade (VWR International, Pennsylvania, USA; order no. 733-1090) > Human whole blood (pool of 4 blood samples) Methods Blood density determination The density of human blood used in this study was determined by using a pycnometer of 10 cm 3 (BRAND BLAUBRAND ; BRAND, GmbH + Co KG, Wertheim, Germany; order no 43308). Also called density bottle, this calibrated device is provided with a certificate notifying its exact capacity when filled with water at a specific temperature (20 C). Here, the exact volume was 9.8422 cm 3. To determine the blood density, the pycnometer was filled with the liquid of interest and weighed on an analytical balance. From the volume (specified on the certificate) and the measured mass, the density was determined. The obtained (1.058 kg/liter) was used to determine the Ζ factor. Imprecision and inaccuracy determination by gravimetric analysis Systematic and random errors were determined using gravimetric methods in accordance with the EN ISO 8655 standard [3]. As requested by the norm, tests were carried out in a draught-free room. During testing, relative humidity was above 50 % and temperature was constant. Instruments, consumables and sample liquids were equilibrated to the test room for at least 2 hours before starting the test. For determining errors, the sample liquid was dispensed ten times into a vessel and weighed. For each condition, three series of ten pipetting/dispensing steps were performed. A new consumable was used for each series. The systematic error (inaccuracy) and the random error (imprecision) were determined for each series of 10 measurements. Three s were obtained per condition, from which the average and standard deviation were calculated.

APPLICATION NOTE I No. 386 I Page 3 Results and Discussion Selection of the right dispensing system Accuracy and precision of a liquid handling instrument are defined with water and can only be guaranteed with this liquid. When the physical properties of the used liquid differ from water, this can highly impact pipetting performances and have to be taken into account when selecting a dispensing system. Human blood density depends on the proportion of its components, in particular the number of red blood cells and the amount of protein. For this reason, the blood density is higher than water density (1.000 kg/liter). The density of blood used in this study was experimentally determined to be 1.058 kg/liter. Blood is also more viscous than water. Value of blood viscosity is 4.5 Pa/s on average. Whereas water has a viscosity of 1 Pa/s. In order to determine the most appropriate dispensing system to deliver an accurate and precise volume of human whole blood, air cushion pipettes were compared to a positive displacement instrument. To minimize operatorsʹ influence, electronic instruments were preferred. Two air cushion pipettes with a maximum volume of 1 ml and 10 ml were used in combination with the appropriate tips (1 ml and 10 ml). As positive displacement instrument, the, was used in combination with Combitips advanced 1 ml and 10 ml. Pipetting and dispensing were evaluated for both systems. Pipetting mode As described by Ewald in the Eppendorf Userguide No. 21 [2], when used in pipetting mode, an air displacement pipette should deliver smaller volumes of high density liquids as compared to aqueous solutions. This theory was confirmed with whole blood. As shown in figures 1 A and 1 B, independent of the volume tested, the volume delivered by the pipette is significantly less than expected inducing the instrument accuracy not being in compliance with the technical limit. On the opposite, by using a positive displacement system, an accurate pipetting of blood is feasible within the systematic error limits (0.52 % and 0.26 % respectively). 1 A 1,0 Percentage 0,0-1,0-2,0-3,0 Systematic error Random error Systematic error Random error -4,0-5,0 1 B 2,0 1,0 0,0 Percentage -1,0-2,0-3,0 Systematic error Random error Systematic error Random error -4,0-5,0 Figure 1: Results of blood pipetting using and an air cushion pipette. Systematic error (blue) and random error (rose) for 10 x 1 ml (A) and 10 x 10 ml (B) are compared to technical specifications (n=3).

APPLICATION NOTE I No. 386 I Page 4 The instrument precision is not affected by the liquid density. This was demonstrated by 10 reproducible indvidual s as displayed in figure 2 A and B. With random error s below 0.2 %, both instruments generated a random error lower than the acceptable error limit. 2 A Xplorer Volume dispensed (µl) 1030 1020 1010 1000 990 980 970 960 950 0 1 2 3 4 5 6 7 8 9 10 11 10,3 2 B Volume dispensed (ml) 10,2 10,1 10,0 9,9 9,8 9,7 9,6 9,5 0 1 2 3 4 5 6 7 8 9 10 11 Figure 2: Results of blood pipetting using and an air cushion pipette. Individual s for 10 x 1 ml (A) and 10 x 10 ml (B) (n=3).

APPLICATION NOTE I No. 386 I Page 5 Dispensing mode Data obtained for dispensing mode is shown in Figure 3. Again, systematic errors obtained with the air cushion pipette (1.52 % and 1.62 % respectively) are significantly higher than the errors obtained with the dispenser (0.45 % and 0.39 % respectively). In this case, the volume delivered by the air cushion pipette is higher than expected opposing data obtained from pipetting mode. This could be attributed to the reduced dead volume. The tip is filled completely to dispense the volume (100 µl or 1 ml) in multiple steps leading to a small air cushion. While in pipette mode 100 µl or 1 ml were aspirated into the tip leading to a bigger air dead volume above the liquid. Percentage (%) 2,5 2,0 1,5 1,0 0,5 0,0 pipette pipette Systematic error Random error 10 x 100 µl 10 x 1 ml Figure 3: Results of blood dispensing using and an air cushion pipette (n=3). Besides an accurate blood dispensing, the positive displacement instrument also permits a reproducible liquid delivery (random errors of 0.26 % and 0.14 % respectively), whereas the air cushion system is more variable (random errors of 0.56 % and 1.16 % respectively). As shown by the individual s in Figure 4, the dispensing variability is particularly prominent when a larger volume is dispensed. In this case, the instrument precision is even above the acceptable error limit. 1,10 Volume dispensed (ml) 1,05 1,00 0,95 0,90 0 1 2 3 4 5 6 7 8 9 10 11 Figure 4: Individual s of blood dispensing (10 x 1mL) using and an air cushion pipette (n=3).

APPLICATION NOTE I No. 386 I Page 6 Comparing both liquid handling systems, the results indicate that a positive displacement instrument is the right choice when accurate and precise dispensing of whole blood is required. Indeed, this dispensing system is not impacted by the blood's physical properties independent of the selected operating mode. It delivers exact and reproducible volumes even in the high volume range. Selection of the most reliable dispenser Knowing that a positive displacement system is the best option to guarantee a precise and accurate dispensing of complex solutions such as blood, customers have to select their instrument carefully. As dispenser tips are specifically designed to be used with its dedicated instrument, dispenser and consumable have to be considered as an inseparable combination. Indeed, with a positive displacement system the accuracy depends on the disposable tip quality, even more than with an air cushion pipette. In order to determine the most reliable equipment to dispense whole blood, the Eppendorf was compared to two major competitors: dispenser B and dispenser R. Each instrument was used with their own consumables and was tested in both operating modes (pipetting and dispensing). As shown on tables 1 and 2, accuracy and precision claimed by the supplier are not always reached, even when the dispensed liquid is water. In pipetting mode as well as in dispensing mode, the dispenser R combined with its dedicated 10 ml dispenser tips generates a systematic error above the upper limit defined by the supplier. In contrast the Eppendorf and the dispenser B deliver accuracy within declared specifications. Table 1 and 2: Accuracies and precisions obtained with three dispensers used in dispensing mode to deliver water. Values are compared to supplier specifications. Green highlighting indicates that the is within the limits defined by the supplier. Red highlighting indicates that the is outside the limits defined by the supplier (n=3). Table 1 Pipetting Mode Multipette E3x Dispenser B Dispenser R Volume pipetted Accuracy 10 x 1 ml ± 0,6% 0,24% ± 0,6% -0,21% ± 0,8% 0,41% 10 x 10 ml ± 0,4% 0,13% ± 0,4% -0,07% ± 0,3% 0,59% Precision Volume pipetted 10 x 1 ml 0,3% 0,04% 0,3% 0,05% 0,2% 0,02% 10 x 10 ml 0,15% 0,02% 0,15% 0,01% 0,2% 0,12% Table 2 Dispensing Mode Multipette E3x Dispenser B Dispenser R Accuracy Precision Volume dispensed 10 x 100 µl ± 0,9% 0,25% ± 1,0% -0,11% ± 0,8% 0,20% 10 x 1 ml ± 0,5% 0,16% ± 0,7% -0,09% ± 0,25% 0,64% Volume dispensed 10 x 100 µl 0,55% 0,21% 0,8% 0,34% 0,55% 0,26% 10 x 1 ml 0,25% 0,10% 0,8% 0,17% 0,5% 0,26%

APPLICATION NOTE I No. 386 I Page 7 1,04 Dispenser B Volume dispensed (ml) 1,02 1,00 0,98 0,96 0,94 0,92 0,90 0 1 2 3 4 5 6 7 8 9 10 11 Figure 5: Individual s of blood dispensing (10 x 1mL) using and dispenser B (n=3). Tested with water, the Eppendorf and the dispenser B were the most reliable systems. Consequently, only those instruments were afterwards evaluated for whole blood handling. As mentioned previously, besides the instrument quality, the consumable itself has to be taken into account. During dispensing, viscous liquids can leave a film on the walls of the tip. This may induce a large volume remaining inside the tip. Air bubbles can also be created, reducing the liquid volume. When dispenser B combined with its 10 ml dispenser tips is used to dispense whole blood, this phenomenon is observed. This results in an aspirated blood volume insufficient to guarantee 10 successive reproducible dispensing steps. As a consequence the last volume dispensed is significantly lower than expected, as demonstrated in Figure 5. This last dispensing step highly impacts the instruments precision. As shown in Table 3, the random error of dispenser B is, in this case, significantly above the acceptable error limit defined by the supplier. This phenomenon is also observed with the dispenser B 1 ml tips. However using those tips this effect is less promiment, allowing the instrument to stay just below the defined precision limit. For an exact and reproducible blood dispensing, the Eppendorf qualifies as the only option. Only this instrument, combined with Combitips advanced, ensures repeated reproducible blood dispensing in every tested case. Table 3: Accuracies and precisions obtained with two dispensers used in dispensing mode to deliver whole blood. Values are compared to supplier specifications. Green highlighting indicates that the is within the limits defined by the supplier. Red highlighting indicates that the is outside the limits defined by the supplier (n=3). Pipetting Mode Multipette E3x Dispenser B Accuracy Precision Volume pipetted 10 x 100 µl ± 0,9% 0,45% ± 1,0% -0,16% 10 x 1 ml ± 0,5% 0,39% ± 0,7% -0,53% Volume pipetted 10 x 100 µl 0,55% 0,26% 0,8% 0,77% 10 x 1 ml 0,25% 0,14% 0,8% 2,02%

APPLICATION NOTE I No. 386 I Page 8 Conclusion Liquids having physical properties different from water are difficult to dispense and clearly affect pipette accuracy as well as precision. Having a higher density and viscosity, whole blood is a liquid very different from water. Positive displacement instruments, based on a direct contact between liquid and piston, should not be affected by liquid properties and should be the consequent choice to handle complex solutions such as blood. Therefore, the Eppendorf, a positive displacement dispenser, has been compared to a classic air displacement pipette ( electronic pipette) to handle whole blood. With an air cushion pipette the systematic error increases significantly, using both pipetting and dispensing mode. Depending to the operating mode selected, the volume delivered can be either smaller or higher than expected. At the opposite, the performance stays stable in all circumstances ensuring an accurate and precise blood multi-dispensing. By comparing both liquid handling systems, this study confirms that a positive displacement system is the right instrument to be employed when exact and reproducible dispensing of blood is critical. In this case, customers should select their instrument carefully among all options available on the market. To determine the most reliable dispenser, the Eppendorf was compared to two major competitors: dispenser B and dispenser R. We firstly demonstrated that even when the liquid dispensed is water, accuracy and precision claimed by the supplier are not always reached. In pipetting mode as well as in dispensing mode, dispenser R combined with its dedicated 10 ml dispenser tip reproducibly generates a systematic error above the limit defined by the supplier. For dispensing blood, an instrument of good quality alone is not sufficient. Indeed, the consumable itself has to be considered as well. With dispenser B associated with its own consumable, the aspirated blood volume is insufficient to guarantee 10 consecutive reproducible dispensing steps. When the last volume dispensed is significantly lower than expected, the instrument precision can be outside acceptable limits defined by the supplier. For an exact and reproducible blood dispensing, the Eppendorf is the right choice. As a matter of fact, this instrument comparison demonstrated, that among all dispensers tested only the Eppendorf, combined with Combitips advanced, reaches accuracy and precision specifications in every case. Literature [1] Ewald K. Fundamentals of dispensing. Eppendorf Userguide N 19, 2015. www.eppendorf.com [2] Ewald K. Influence of physical parameters on the dispensed volume of air-cushion pipette. Eppendorf Userguide N 21, 2015. www.eppendorf.com [3] EN ISO 8655, Parts 1 6: Piston-operated volumetric apparatus. ISO 2002.

APPLICATION NOTE I No. 386 I Page 9 Ordering Information Description Order no. international Order no. North America Multipette E3/ E3x, with charging cable 4987 000.029 4987000029 and 2 Combitips advanced assortment packs, 1 µl - 50 ml Eppendorf Xplorer plus, single channel, 50-1,000 µl 4861 000.732 4861000732 Eppendorf Xplorer plus, single channel, 0.5-10 ml 4862 000.759 4862000759 ept.i.p.s., 50-1,000 µl, Eppendorf Quality 0030 000.927 0030000927 ept.i.p.s., 0.5-10 ml, Eppendorf Quality 0030 000.765 0030000765 Combitips advanced, 1.0 ml, Eppendorf Quality 0030 089.430 0030089430 Combitips advanced, 10 ml, Eppendorf Quality 0030 089.464 0030089464 Your local distributor: www.eppendorf.com/contact Eppendorf AG 22331 Hamburg Germany eppendorf@eppendorf.com www.eppendorf.com www.eppendorf.com Mettler-Toledo is a registered trademark of Mettler-Toledo Inc., USA. Sartorius and Cubis are registered trademarks of Sartorius AG, Germany. BRAND and BLAUBRAND are registered trademarks of BRAND GmbH + CO KG, Germany Eppendorf, the Eppendorf Brand Design, Multipette, Combitips advanced, Eppendorf Xplorer and ept.i.p.s. are registered trademarks of Eppendorf AG, Germany. Eppendorf Quality is a trademark of Eppendorf AG, Germany All rights reserved, including graphics and images. Copyright 2017 by Eppendorf AG, Germany.