GENERAL INCUBATOR Instructin Manual Mdel : LGI-050T Please read this manual carefully befre using the instrument Labnics Equipment
Table f Cntent CHAPTER CONTENT PAGE NO. 1. Befre Use 1 2. Safety Precautin 1 2.1 Pwer Cnnectin 1 2.2 Installatin 2 2.3 Operatin 2 2.4 Maintenance 2 3 Feature and Advantage 3 4. Specificatins 3 5. Parts and Descriptin 4 6. Main Cntrller 4 7. Operatin 6 8. Frequently Asked Questins 8 9. Truble Shting 8 10. Setting Parameters 8 11. Inspectin Lg 11 12. Service Reprt 12
CHAPTER1. BEFORE USE: Thank yu fr chsing LABNICS Labratry Prducts Please read this peratin manual carefully befre use fr yur safety and ptimum perating perfrmance. If yu have any questin, please cntact sales representative r service department. CHAPTER 2. SAFETY PRECAUTION: This manual cntains imprtant perating and safety infrmatin. Yu must carefully read and understand the cntents f this manual prir t the use f this equipment. Warning:- Warning alert yu t a pssibility f persnal injury Cautin:- Cautin alerts yu t a pssibility f damage t the equipment. 2.1 Pwer Cnnectin: Cautin 1. Yur Incubatr is designed fr 110VAC 60Hz 1P r 220VAC 50 / 60Hz 1P. 2. Check electrical requirement n the name plate befre use. Name plate is lcated in electric crd cnnectin. 3. Cnnect t receptacle with grund cnnectin. 4. Make sure t supply enugh electrical current. - 1 -
2.2 Installatin: Cautin 1. D nt use in high humid envirnment. May cause Electrical leakage Water drplet may ccur n the dr Crrsin may ccur 2. D nt use in high temperature envirnment r s nt use beside instrument generate heat. 3. Install incubatr at least 50mm apart frm side wall. 4. Place flat, rigid and leveled surface. 5. When mving ven d nt up-side-dwn. 6. D nt place any bject n the tp f incubatr. 2.3 Operatin : Warning 1. D nt put vlatile, flammable and explsive material inside f incubatr. 2. D nt put vlatile, flammable and explsive material nearby incubatr. Cautin : 1. D nt spill liquid in the chamber. 2. Be careful nt t spill liquid n the cntrl panel. 2.4. Maintenance: Cautin 1. D nt pur water r any liquid when yu clean incubatr. 2. D nt use highly rganic slvent fr cleaning surface f incubatr. - 2 -
CHAPTER 3. FEATURE: Labnics LGI-Series Micrbilgy Incubatr is... ideal fr general labratry incubatin requires temperature range frm ambient +5 C t 70 C. Cntrller - Digital micrprcessr PID Cntrller Perfrmance - Temperature Range : ambient +5 C ~ 70 C Reslutin : ± 0.1 C Accuracy : ± 0.2 C Unifrmity : ± 0.5 C Material - Inner Chamber : Stainless Steel 304 (ANSI 304) Outer Cabinet : Steel with epxy pwder cating Glass Dr : Tempered Safety Glass 5mm thick Cnvenience - Rund crnered edge fr easy cleaning and decntaminatin Transparent glass inner dr Height adjustable shelve in 25mm increment Safety - Over current cut-ff Over temperature cut-ff CHAPTER 4.SPECIFICATIONS:- Mdel LGI-050T Chamber Vlume 50L N. f Rm 1 Temperature Range Ambient + 5 C ~ 70 C Accuracy ± 0.1 C at 37 C Unifrmity ± 0.5 C at 37 C Heater 100Watt Cntrller Digital PID Cntrller with Timer and Aut-Tuning Wait ff Timer mm:ss / hh:mm / Cntinuus Selectable Sensr Pt 100 Ω Safety Device Temperature Hydraulic ver Temperature prtectin safety Device Electrical Electrical Leakage Breaker Dimensins Inner 400 x 300 x 400 W x D x H mm Outer 580 x 530 x 600 Shelves PVC Cating wire shelf Electrical Requirement 120VAC 60Hz/220 VAC 50/60 Hz - 3 -
CHAPTER 5. PARTS AND DESCRIPTIONS : Bath Cver Dr Latch Outer Dr AC Inlet fr pwer Crd CHAPTER 6. MAIN CONTROLLER Pilt Lamp LED Display Over Temp. Cut-Off Pwer Switch 3 7. 0 MODE Timer Heat Set Parameter Shift / AT MAX 120 0 40 MODE Increment 80 SAFETY POWER MODE Buttn INC (DSP) Buttn SHIFT (AT) Buttn - 4 -
[DISPLAYS] 20.0 Digital LED Display Display current temperature f the incubatr when turned n. Press INC buttn t shw temperature and timer alternatively. MODE < Display Mde Lamp > TIMER HEATER OUTPUT Display mde Lamp TIMER : If user set wait-ff timer, lamp turned n. Lamp blinks until temperature reach t set temperature. Lamp stp blinking and start cunt dwn when temperature reach t set temperature. HEATER OUTPUT: Red lamp n and ff when cntrller heat up incubatr. [ BUTTONS ] MODE MODE SHIFT (AT) INC (DSP) Mde Buttn Push t Change temperature and time. Push t change parameters. Push and hld fr 20 secnds t enter parameter mde. (Remarks) Operating parameters are set befre shipment. D nt change withut understanding f each functin f parameters. SHIFT (AT) Buttn Shift t left digit AUTO-TUNING Functin Press and hld fr 30 secnds t start aut-tuning. Aut-tuning was finished befre shipment. INC (DSP) Buttn Increase value when setting temperature and timer. Press t display temperature and timer alternatively in nrmal display mde. - 5 -
0 MAX 320 200 SAFETY 100 SAFETY Safety dials fr ver temperature cut-ff. Set dial 10 ~ 20% higher than perating temperature. If temperatures rise ver perating temperature, safety autmatically cut-ff heater utput t prevent ver heating. POWER SWITCH Main pwer switch. POWER CHAPTER 7. OPERATION:- Befre peratin 1) Check electrical requirement n the name plate befre cnnect t cnsent. 2) Remve packing material in the chamber and install shelves. 3) Cnnect pwer plug between AC inlet cnsent n the left side panel and wall munt receptacle. Hw t Set Temperature 1) In nrmal display mde. 2) Press MODE buttn. 3) LED displays "TEMP "symbl and prmpt user input. 4) Press MODE buttn again. 5) LED displays current perating temperature. 6) Input new perating temperature by using SHIFT and INC buttn. 7) Yu can input perating temperature range frm ambient +5 C t 70 C. 8) Press MODE buttn t set timer. 9) Press mdel buttn three times t skip timer setting. - 6 -
Hw t Set Timer 1) Press MODE buttn twice t set timer. 2) LED displays "TIME "symbl and prmpt user input. 3) Press MODE buttn again. 4) LED displays current timer setting (ex. 00:00 ). 5) Input new wait-ff timer value by using SHIFT and INC buttn 6) Default time scale is HH:MM. 7) Fr cntinuus peratin set timer at 00.00. 8) Yu can input timer range frm 1min t 99hr 59min. 9) Press MODE buttn back t nrmal display mde. (Remarks) Timer can be changed in different scale by changing Parameter 1) mm : ss 1 sec ~ 99 min 59 sec 2) hh : mm 1 min ~ 99 hr 59 min 3) dd : hh 1 hr ~ 99 day 23 hr Refer parameter setting t change time scale MODE 0, N1: time scale selectin parameter 0 : mm:ss 1 : hh:mm 2 : dd: hh 0 Over Temperature Prtectin MAX 320 200 SAFETY 100 1) User must set the ver temperature prtectin befre peratin. 2) Turn ver temperature prtectin dial with screw driver. 3) Set temperature abut 10~20% higher than the perating (user set) temperature. 4) If ver-temp. Prtectin value is under perating temperature, temperature cannt be reach up t perating temperature. Start Temperature Cntrl 1) After setting temperature and timer, incubatr start increase temperature t perating temperature. 2) Put samples t be incubated n the shelves. 3) D nt put samples n the bttm f the chamber. 4) Clse Glass dr carefully and firmly. - 7 -
CHAPTER 8. FREQUENTLY ASKED QUESTION 8.1 Temperature keep increasing and decreasing under perating temperature Cause: SAFETY setting is lwer than perating temperature Slutin: Turn SAFETY setting clckwise higher than perating temperature 8.2 Temperature Oversht Cause: High rm temperature. Slutin: Yur incubatr has n cling system. If rm temperature is higher than 30 C, yur incubatr cannt maintain 37.0 C. Use incubatr in ptimum perating cnditin. 8.3 LED displays uuuu and beep Cause: Over heat higher than 70 C. Slutin: Yur incubatr cannt be used ver 70 C If temperature increase ver 70 C, cntrller warning high temperature and cut-ff heater. CHAPTER 9. TROUBLE SHOOTING Errr Symbl Cause Slutin uuuu Over heat higher than 70 C Call fr Service nnnn Sensr discnnectin Call fr Service CHAPTER 10. SETTING PARAMETERS:- 10.1 Parameter T set parameters, press and hld MODE Buttn fr 5 secnds. Press SHIFT and INC Buttn t change values. Press MODE Buttn t g next parameter. T escape frm Parameter mde t nrmal display mde, press and hld MODE Buttn fr 5 secnds. Factry Parameter A Parameter Name f Parameter Setting Range Factry Default User Value Symbl and Descriptins ALH Alarm Limit High 00.0 ~ 99.9 C 2.0 ALL Alarm Limit Lw 00.0 ~ 99.9 C 21.0 HYS Hysteresis 00.0 ~ 99.9 C 0.2 BEEP Beep On Time 0 ~ 9999 Sec. 30 ADJ Temperature Calibratin -99.9 ~ 299.9 C LOC Key, Data, Parameter Lck 0000 ~ 1111 0000-8 -
10.2 Parameter 2 T set parameters, press and hld MODE Buttn fr 30 secnds. LED Display turns t ALH after 5 secnds. Keep press the buttn 25 secnds. Press SHIFT and INC Buttn t change values. Press MODE Buttn t g next parameter. T escape frm Parameter mde t nrmal display mde, press and hld MODE Buttn fr 5 secnds. Factry Parameter B Parameter Name f Parameter Setting Range Factry Default User Value Symbl and Descriptins rnt Maximum temperature limit t set -99.9 ~ 299.9 C 71 D Nt Change ACTP Temp. where timer activate -00.0 ~ 99.9 C 0 (Parameter can be changed nly when the N2 value f Mde0 is 1) Timer starts cunt dwn when, (current temp. - set temp.) > ACTP Prd Perid (Output Interval) 1 ~ 99 sec. 2 D Nt Change P Prprtin 0 ~ 6999 Aut-tuned value D Nt Change I Integral 0 ~ 6999 Aut-tuned value D Nt Change A Anti-Intergran 0 ~ 6999 Aut-tuned value D Nt Change D Differential 0 ~ 6999 Aut-tuned value D Nt Change Mde0 Operating Mde Cntrl 0 N3 N2 N1 N0 0001 D Nt Change N3 0 = KS, JIS Pt 100 0 0 0 0 1 = DIN Pt 100 1 1 1 1 N2 0 = ALH (Relative) 1 = AALH (Abslute) N1 0 = ALL (Relative) 1 = AALL (Abslute) N0 0 = 000 C 1 = Decimal display (000.0 C) Mde1 Operating Mde Cntrl 1 N3 N2 N1 N0 0111 N3 0 = PID Cntrl 0 0 0 0 1 = ON/OFF Cntrl 1 1 1 1 N2 0 = Timer Disable 1 = Timer N1 0 = Time Scale ( mm:ss) 1 = Time Scale ( hh:mm) 2 = Time Scale ( dd:hh) N0 0 = Pwer n restre disable 1 = Pwer n restre - 9 -
Mde2 Operating Mde Cntrl 2 N3 N2 N1 N0 0000 D Nt Change N Functin 0 0 0 0 1 1 1 1 DrAn Fix drift f temperature display within the set value 0.5 Temperature drifts during peratin wing t several reasns. T eliminate temperature drift, set DrAn value t fix temperature within the value. LbAT N Functin 0000 1) RNT : Maximum temperature limit t set. User cannt set temperature higher than this value. D nt alter the value. Factry default is 251. 2) ADJ : Temperature adjustment. Smetimes the actual temperature f the water is slightly the different frm displayed temperature. User can adjust the displayed temperature by cmpensate the difference by Adj value. Example : 1) Measure temperature f Incubatr with calibrated thermmeter. 2) Read LED display. 3) Change Adj value. Thermmeter Cntrller Adj Value Temperature Reading Temperature Reading 38.0 C 37.0 C 1 36.0 C 37.0 C - 1-10 -
Chapter 11: Inspectin Lg Fr Digital Water Bath Mdel # : LGI-050T Serial # : Date & Time: Amb. Temp. : Client : Electricity : VAC Hz Labnics Pretest check list (Initial after each Checkpint) If Nn-Applicable enter N/A Terminatin f each electrical cnnectins Terminatin f remte alarm N/A Battery terminatin light N/A Surge prtectin terminatin N metal cntact with refrigeratin tubes N/A N gaps in insulatin Cntrl panel key switch tight Equipped with prper accessries Esthetics k Labnics TEST AREA, TEST LOG (NOTE) If units fails any prtin f the test enter "NG" in the blank space adjacent t that check pint Technician Vlts at plug / terminal strip / Hz Fans-Crrect blade, n cntact Unusual vibratin during run Unusual nise during run Unusual fan nise r vibratin Battery Check 12 vlts r better N/A Display indicatin and warning functins Battery % reading frm display N/A Leakage r cndensatin n the tubes N/A Cut in vltage Back Up system N/A Display vltage N cndensatin n cabinet at BO N/A Surge Check frst n dr and interir N/A Check Fuse LED Lid/dr seal at pull dwn RS232 N/A Dr clse and latches Alarm/remte peratin N/A Set pint security verified Unusual nise at startup N excessive dummy panel vibratin Labeling n cmpnents crrect Mlded plug nt verheated Recrder check and calibrated N/A N frst n refrigeratin cmpnents N/A Leakage n the water pipe line N/A Breaker fit at lw temperature N/A Verify slenid valve N/A Leak check while running Verify ver temp. prtectin f chamber Verify mtr vltage (verify clr cde) Verify ver temp. prtectin f humidity heater N/A Check flurescent lamps and lights N/A - 11 -
SERVICE REPORT Custmer s Address : Cntact Persn / Designatin : Tel.N.: Fax N.: Weekly Off.: Dept.: Date Frm Time T System Cnfiguratin Mdel Serial N. Date : Status : OK SR. N. Nt OK Installatin Warranty Demnstratin Maintenance Repairs Cntract Applicatin Calibratin Billable Validatin Curtesy Nature f Prblem : Observatin & Actin Taken : Custmer s Remarks : Parts Replaced : Parts Recmmended / Actin Required : Yes Service Engineer s Name & Signature N Requisitin Number : Custmer s Name, Signature, Date & Stamp Page Of - 12 -