Discharge coefficient behaviour in presence of four perforated plates flow conditioners - Numerical investigation Part 1-

Similar documents
Discharge Coefficient Behaviour in Presence of Four Perforated Plates Flow Conditioners

QUALIFICATION OF FLOW CONDTIONING DEVICES ACCORDING TO THE NEW API 14.3 PART 2 PROCEDURE. Klaus J. Zanker and Dale Goodson. Daniel Industries, Inc.

The Expansibility Factor Equations in ISO and ISO : Do They Deliver What They Promise?

EXPERIMENTAL INVESTIGATION OF THE INSTALLATION EFFECTS ON THE VENTURI FLOWMETER PERFORMANCE

NORTH SEA FLOW MEASUREMENT WORKSHOP 2004 In. St Andrews, Scotland

Lecture23. Flowmeter Design.

Discharge Coefficient Prediction for Multi hole Orifice Plate in a Turbulent Flow through Pipe: Experimental and Numerical Investigation

ABSTRACT I. INTRODUCTION

Alden Test Results On AMITY Insert Venturi

Prediction of Performance Characteristics of Orifice Plate Assembly for Non-Standard Conditions Using CFD

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

AN 8-PATH ULTRASONIC MASTER METER FOR OIL CUSTODY TRANSFERS

FLOW DISTRIBUTION ANALYSIS IN A HEAT EXCHANGER WITH DIFFERENT HEADER CONFIGURATIONS

Conditioning of swirling and stratified pipe-flow: Analysis with laser-doppler velocimetry

Head loss coefficient through sharp-edged orifices

NUMERICAL EVALUATION OF DISTORTED VELOCITY PROFILE EFFECTS ON INSERTION FLOW METERS

[Prasanna m a*et al., 5(6): July, 2016] ISSN: IC Value: 3.00 Impact Factor: 4.116

Comparison between Numerical and Experimental for UVP Measurement in Double Bent Pipe with Out-of-Plane Angle

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH

Mechanical Measurements and Metrology Prof. S. P. Venkateshan Department of Mechanical Engineering Indian Institute of Technology, Madras

Predictionof discharge coefficient of Venturimeter at low Reynolds numbers by analytical and CFD Method

ME332 FLUID MECHANICS LABORATORY (PART I)

Experiment No.4: Flow through Venturi meter. Background and Theory

Smart DP Flow Metering A summary of Established Technology and Latest Innovation. Gary Fish

NUMERICAL ANALISYS OF FLOW CONDITIONER EFFICIENCY

Lecture 13 Flow Measurement in Pipes. I. Introduction

FINAL DRAFT: October 25, 2010 API TECHNICAL REPORT FUEL GAS MEASUREMENT. FINAL DRAFT: October 25, 2010 Page 0

PRESSURE DROP OF FLOW THROUGH PERFORATED PLATES

Experimental and CFD analysis of flow through venturimeter to determine the coefficient of discharge

Flow Measurement in Pipes and Ducts COURSE CONTENT

Throttling Characteristics of Multi-Hole Orifice in Multi Stage Assembly

Analysis of the possibility to optimize the passive pressure pulsations dampers using CFD

04/01/1998 Developments in DP Flowmeters By Jesse Yoder

ERRATA CRANE FLOW OF FLUIDS THROUGH VALVES, FITTINGS AND PIPE TECHNICAL PAPER NO. 410 METRIC VERSION

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Installation effects on an ultrasonic flow meter with implications for self diagnostics

Measurement of fluid flow in closed conduits

Wall Pressure Due to Turbulent Flow Through Orifice Plate

Experiment (4): Flow measurement

ABSTRACT I. INTRODUCTION

INTEGRATION OF CURRENT METER MEASUREMENTS IN GENERAL FLOW CROSS SECTIONS

Applied Fluid Mechanics

Applied Fluid Mechanics

Process Gas Flow Measurement and Performance Evaluation API TECHNICAL REPORT 2571 SECOND EDITION DRAFT, FEBRUARY 2018

VENTURIMETER EXPERIMENT

TransCanada / Daniel Using Advances Diagnostics to Identify Measurement Uncertainties

Flow rate and mass flow rate

ISO 9906 INTERNATIONAL STANDARD. Rotodynamic pumps Hydraulic performance acceptance tests Grades 1 and 2

DISCHARGE COEFFICIENT OF SMALL SONIC NOZZLES

Numerical and Experimental Study of Effects of Upstream Disturbance on Accuracy of Vortex-Shedding Flow Meter

Investigation of Flow Profile in Open Channels using CFD

EXPERIMENT No.1 FLOW MEASUREMENT BY ORIFICEMETER

Numerical Methods Applied to Fluid Flow Metering. Ernst von Lavante University of Duisburg-Essen

VELOCITY PROFILE EFFECTS IN CORIOLIS MASS FLOWMETERS: RECENT FINDINGS AND OPEN QUESTIONS

CFD ANALYSIS OF CD NOZZLE AND EFFECT OF NOZZLE PRESSURE RATIO ON PRESSURE AND VELOCITY FOR SUDDENLY EXPANDED FLOWS. Kuala Lumpur, Malaysia

INDUSTRIAL APPLICATION EXPERIENCES OF NEW TYPE FLOW- METERING SYSTEM BASED ON ULTRASONIC-DOPPLER FLOW VELOCITY-PROFILE MEASUREMENT

Hydraulics and hydrology

Experimental and Numerical Investigation of Two- Phase Flow through Enlarging Singularity

Flowmeter Discharge Coefficient Estimation

MEDIDAS EN FLUIDOS. INEL 5205 Instrumentación. Monday, November 19, 12

Piping Systems and Flow Analysis (Chapter 3)

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560

Lecture 24. Design of flow meters

DESIGN AND PERFORMANCE OF THE CONVERGING-DIVERGING VORTEX FLOWMETER

CFD Analysis of Forced Convection Flow and Heat Transfer in Semi-Circular Cross-Sectioned Micro-Channel

Computational Fluid Dynamics Based Analysis of Angled Rib Roughened Solar Air Heater Duct

Measurements using Bernoulli s equation

THE STRUCTURE OF FLOW FIELD IN SMALL SIZE VORTEX FLOW METERS

Bounding the Uncertainties of the Meter Factors for LEFM CheckPlus Flow Elements in Feedwater Flow Measurements

Canada. CFD-Experiments Integration in the Evaluation of Six Turbulence Models for Supersonic Ejectors Modeling

Chemical Engineering 3P04 Process Control Tutorial # 1 Learning goals

Why the Fluid Friction Factor should be Abandoned, and the Moody Chart Transformed

[Ramya, 4 (12): December, 2015] ISSN: (I2OR), Publication Impact Factor: 3.785

Effect of radius ratio on pressure drop across a 90 bend for high concentration coal ash slurries

Determination of the Blockage Effect on a Thermal Anemometer using a Small Open Jet Wind Tunnel

Orifice and Venturi Pipe Flow Meters

Chapter 6. Losses due to Fluid Friction

1-Reynold s Experiment

Investigation of Jet Impingement on Flat Plate Using Triangular and Trapezoid Vortex Generators

Experimental and numerical analysis of the flow inside a configuration including an axial pump and a tubular exchanger

ME411 Engineering Measurement & Instrumentation. Winter 2017 Lecture 11

Numerical Investigation of Multijet Air Impingement on Pin Fin Heat Sink with Effusion Slots

White Paper FINAL REPORT AN EVALUATION OF THE HYDRODYNAMICS MECHANISMS WHICH DRIVE THE PERFORMANCE OF THE WESTFALL STATIC MIXER.

Water Circuit Lab. The pressure drop along a straight pipe segment can be calculated using the following set of equations:

CHAPTER THREE FLUID MECHANICS

Calibration of Orifice Flow Meter and Venturi Flow Meter

Laboratory work No 2: Calibration of Orifice Flow Meter

Fluid flow simulation in a double L-bend pipe with small nozzle outlets

FLOW MEASUREMENT IN PIPES EXPERIMENT

PERFORMANCE ANALYSIS OF PARABOLIC TROUGH COLLECTOR TUBE WITH INTERNAL INTERMITTENT FINS

An Essential Requirement in CV Based Industrial Appliances.

ABSTRACT I. INTRODUCTION

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Practical Analysis Of Turbulent Flow In A Pipe Using Computational Fluid Dynamics

Experimental Investigations on the Local Distribution of wall static pressure coefficient Due To an Impinging Slot Air Jet on a Confined Rough Surface

ALASTAIR MCLACHLAN Applications Cameron

A Computational Fluid Dynamics Investigation of Solar Air Heater Duct Provided with Inclined Circular Ribs as Artificial Roughness

N. Zhang, W.-J. Yang and Y. Xu Department of Mechanical Engineering and Applied Mechanics, The University of Michigan, Ann Arbor, MI 48109, USA

LECTURE 6- ENERGY LOSSES IN HYDRAULIC SYSTEMS SELF EVALUATION QUESTIONS AND ANSWERS

Transcription:

17 International Congress of Metrology, 5 (215) DOI: 1.151/ metrolo gy/ 2155 C Owned by the authors, published by EDP Sciences, 215 Discharge coefficient behaviour in presence of four perforated plates flow conditioners - Numerical investigation Part 1- Boualem Laribi 1,a and Abdellah Abdellah Hadj 2 1 FIMA Laboratory, University of Djillali Bounaama, Khemis Miliana, Algeria 2 LMPEM Laboratory, University of Feres Yahia, Medea, Algeria Abstract. This paper present a part 1 of numerical experimentation of the behaviour of the discharge coefficient and the effect of four perforated plates flow conditioners on the discharge coefficient for flow measurement accuracy. Three of the plates are described by the Standard ISO517 and the fourth one is proposed for study. In this part 1, the disturber used is double bend in perpendicular planes while in the part 2 the flow is subject to disturbers namely % and 5% closed valves. The turbulent flow is examined in conduit with an inner diameter of D=1mm. The diameter of orifice meters are respectively d=5,, 7 and 75mm which done for =d/d respectively the values of.5,.,.7 and.75. The orifice meters are located in conduit at different stations downstream the disturber. The flow is examined with air at Reynolds number Re=2.5x1 5. The software used for this simulation is ANSYS package with k- like turbulence model. A numerical investigation was done before (CIM 21) with the same conditions but without perforated plates and the results showed that when the diameter of the orifice meter increases the shifts deviation of the discharge coefficient increases this causes a great error in flow measurement. Contrary, when the diameter of the orifice meter decreases the shifts deviation in the discharge coefficient decreases and the error in flow measurement is reduced. These results are the same with the two disturbers double bend and 5% closed valve used separately in conduit. In this paper the results showed that the perforated plates have significantly reduced the error on the discharge coefficient. Indeed, the errors recorded downstream disturbers are superior to 2%. Downstream the perforated plates used separately the errors on the discharge coefficient are reduced to a value inferior to 1% for the four plates. It is noted that the standards ISO517 and AGA stipulate that the error on the discharge coefficient Cd must be less than for better flow measurement accuracy. By comparing our results with this condition we found that the results obtained with the four plates are substantially reduced especially downstream station z=25d (z=2.5d downstream flow conditioners). However the fourth proposed plate with its high porosity produces less losses pressures than those of the other three plates. This is good condition of exploitation for some installation where high losses pressures are not tolerated. 1 Introduction Flow meters flowmeters are calibrated and characterized under fully developed pipe flow conditions. Disturbances such as swirl, cross-flow, and asymmetry can produce relevant metering errors [1]. In practical applications, the fully developed conditions can hardly be obtained. In fact, flow disturbances are caused by the elements of the piping itself, such as bends, Tee and valves. Theoretically, it could be possible to reduce the influence of these disturbances using an adequate straight of pipe between the disturbers and the flowmeter. In practice, due to the reduced dimensions of the piping, it is useful to use flow conditioners [1,2]. The flow conditioners used in the oil and gas industry such as the Etoile, the tube bundle, the Sprinkle, the Zanker and the perforated plate, have been described in the international standards [1]. The Etoile flow straightener has been the subject of several studies from Wauters and al [], Laws and al [4] to Laribi and al [5,]. Standards ISO 517[1] and AGA-[2] define a satisfactory condition on the discharge coefficient errors while high accuracy about in mass flow measurement is required, disturbers in the flow introduced by contractions, bends, valves and other components introduce errors more % []. Flow conditioners have as a main role the acceleration of the formation of the established flow, i.e., obtaining a fully developed velocity and turbulent intensity profiles [7] over a length as reduced as possible in order to a boualemlaribi@yahoo.com This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4., which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Article available at http://cfmetrologie.edpsciences.org or http://dx.doi.org/1.151/metrology/2155

Web of Conferences attenuate the disturbance of the flow other wise to reduce the error on flow measurement [8,]. Our paper is a numerical analysis, with a commercial CFD code [1], which examines the effect of disturber the discharge coefficient errors. Four perforated plates are examined to show the capability of plate to reduce this error. Three of the perforated plates are described in the standard ISO 517 namely the NOVA, the NEL and the ZANKER. The fourth one is a new design and proposed plate. The study is focused on conduit with diameter D of 1mm at a Reynolds number of 2.5x15. The measuring stations of the discharge coefficient are respectively 1.5D, 7D, D, 17.5D, 25D, 5D downstream the disturber. - A ring of 1 holes of diameter.d on a pitch circle diameter of.8d The plate thickness is.d with a pressure loss coefficient approximately equal.2. 2 Experimental simulation 2.1 Conduit The flow in the conduit, Fig. 1, is produced by air with a Reynolds number Re=2.5x15. The air leads from the entrance on the length 1D for good flow conditions upstream disturber. The disturber is installed at 1D from the entrance. The four plates flow conditioners are used separately at station =14.5 downstream of the entrance of conduit (4.5D downstream disturber). The measuring stations of the discharge coefficient are respectively 1.5D downstream disturber and 2.5D, 7.5D, 1D, 2.5D,.5D and 1.5D downstream flow conditioners. NOVA F.C. (P1) ZANKER F.C. (P) NEL F.C. (P2) NEW PLATE F.C. (P4) Figure 2. The perforated plates used Figure 1. Facility for numerical simulation 2.2 Perforated plates used The flow in The four perforated plates, P1, P2, P and P4 used in this simulation are described as bellow, Fig.2, P1 : NOVA [1] flow conditioner is provided in our case - A central hole of diameter.22d - A ring of 8 holes of diameter.1d on a pitch circle diameter of.5d - A ring of 1 holes of diameter.4d on a pitch circle diameter of.85d The plate thickness is.5d with a pressure loss coefficient approximately equal 2.. P2 : NEL PLATE [1] flow conditioner is provided - A ring of 4 holes of diameter.1d on a pitch circle diameter of.18d - A ring of 8 holes of diameter.1d on a pitch circle diameter of.48d P : ZANKER PLATE [1] flow conditioner is provided - A ring of 4 holes of diameter.141d on a pitch circle diameter of.25d - A ring of 8 holes of diameter.1d on a pitch circle diameter of.5d - A ring of 4 holes of diameter.1d on a pitch circle diameter of.75d - A ring of 8 holes of diameter.11d on a pitch circle diameter of.85d - A ring of 8 holes of diameter.77d on a pitch circle diameter of.d The plate thickness is.5d with a pressure loss coefficient approximately equal.. P4 : NEW PLATE flow conditioner is provided - A central hole of diameter.24d - A ring of holes of diameter.2d on a pitch circle diameter of.5d - A ring of holes of diameter.d on a pitch circle diameter of.d - A ring of holes of diameter.17d on a pitch circle diameter of.8d 5-p.2

17 International Congress of Metrology The plate thickness is.5d. 2. Basic equations The general equation used in CFD and by Fluent code is given by: Where is the general dependent variable which can be the mean velocity, the turbulent kinetic energy k or the rate of dissipation of the turbulent kinetic energy. S is the term source of the variable is the coefficient of diffusion of The k-ε model is used like turbulent model. It is based on the Boussinesq hypothesis. It is a semi-empirical model. Two transport equations are used, one for the turbulence kinetic energy k and the other for its dissipation rate ε. The reader can found more information in []. 2.4 Variation of discharge coefficient For testing the effect of disturbers on the discharge coefficients of the orifice meters, the shift deviation for the discharge coefficient Cd(%) is calculated by using the difference pressure P obtained by the simulation at different locations of the orifice meter in the pipe and Po at the same time at station =7 were the flow is supposed fully developed. Eq. 2 shows the calculus formula Po (%) 1 P Cd (2) The difference pressure is calculated according to the standard ISO 517 at D upstream and D/2 downstream the orifice meter. This formula was applied for all orifice plates with the disturber. Results and discussion The examination of the Fig. to Fig. shows the discharge coefficient error in the case of a double bend in perpendicular planes used with the four flow conditioners. The flow meter is located at different stations downstream the flow conditioners. The ISO 517 limits of the discharge coefficient is included for comparison. Fig. shows the discharge coefficient errors at several stations downstream double bend in perpendicular planes with the NOVA F.C on line. At =1.5 we register an error values around -2% for the orifice meters with =.7 and.75 and a minimum values around 1% for =.5 and.. These results are in good agreement with results obtained in before research works and exposed in the International Congress of Metrology 21 [8]. We can see the effectiveness of the NOVA flow conditioner at station =7 (2.5D downstream the NOVA F.C.) to reduce these errors around ±1% for all flow meters except for =.75 where the error closed 5% at station =7. The best results where the discharge coefficient errors are closed the ISO limits are obtained for =.7 in all stations downstream the NOVA F.C. Fig. 4 shows the discharge coefficient errors at several stations downstream double bend in perpendicular planes with the NEL F.C on line. At =1.5 we register an error values around -2% for the orifice meters with =. and a minimum values around -.5% for =.7 and =.75. We can see the effectiveness of the NEL flow conditioner at station =17 (.5D downstream the NEL F.C) to reduce these errors around ±1% for all flow meters. The best results where the discharge coefficient errors are closed the ISO limits are obtained for all the flow meters except for =. where the error is around 1% at station =25 downstream. For the NEL C.F it seems all deviations on the discharge coefficient are greater than.5% as stipulate by the standards. Fig. 5 shows the discharge coefficient errors at several stations downstream double bend in perpendicular planes and the ZANKER F.C on line. At =1.5 we register an error values around -5% for the orifice meters with =. and a minimum values around -.5% for =.75. We can see the effectiveness of the ZANKER flow conditioner at station =17 (.5D downstream the ZANKER F.C) to reduce these errors around ±1% for all flow meters. The best results where the discharge coefficient errors are closed the ISO limits are obtained for all the flow meters where the error is around 1%. For the ZANKER C.F it seems that all deviations on the discharge coefficient are greater than as stipulate by the standards. For station =7 downstream the disturber we register an error around % for =.7 and around -8% for =.75. For this plate all the values of the discharge coefficient are out of ISO limits. Fig. shows the discharge coefficient errors at several stations downstream double bend in perpendicular planes with the new plate F.C on line. At =1.5 we register an error values around -2% for all orifice meters except with =.75 where the error is closed the ISO limits. We can see the effectiveness of the new plate F.C at station =1 and 25 (8.5D and 21.5D downstream F.C) to reduce these errors around for flow meter with =.7. The best results where the discharge coefficient errors are closed the ISO limits are obtained for all the flow meters except for =.75 where the error is around 1.5% at station z=25d. 5-p.

Web of Conferences - 5-5 - - - - Figure : Discharge coefficient errors with the NOVA F.C and double bend on line Figure 5 : Discharge coefficient errors with the ZANKER F.C and double bend on line - 5-5 - - - - Figure 4 : Discharge coefficient errors with the 4. Conclusion NEL F.C and double bend on line The examination of the results shows that the double bend produces an error, generally in the case of no flow conditioners, less than %. Downstream the perforated plates used separately the errors on the discharge coefficient are reduced to a value inferior to 1% for the four plates. It is noted that the standards stipulate that the error on the discharge coefficient Cd must be less than.5% for better flow measurement accuracy. We show that the perforated plates have significantly reduced the error on the discharge coefficient. Also these results show that the orifice meters are very sensitive to the type of flow conditioners used. Indeed, there are some restrictions to use a particular flow conditioner with a particular orifice meter. Based on the results we see that the NOVA FC is better suited with the Figure : Discharge coefficient errors with the New plate F.C and double bend on line orifice meter with =.7 where the error on the discharge coefficient is within the limits set by the standards. The NEL FC is better adapted with the orifice meter which =.75. The ZANKER C.F is poorly adapted than these other flow meters but the error on the discharge coefficient Cd is around of ±1%. Looking at these results obtained by the proposed new plate we see that it fits better with the orifice meter which =.7 where the error on the discharge coefficient is around the values set by the standards. By comparing our results with this condition we found that the results obtained with the four plates are substantially reduced especially downstream station z=25d (z=2.5d downstream flow conditioners). However the fourth proposed plate with its height 5-p.4

17 International Congress of Metrology porosity produce less losses pressures than those of the other three plates. This is good condition of exploitation for some installation where height losses pressures are not tolerated. Views these differences of these results obtained we thought to undertake another series of numerical works with other disturbers to clarify this divergence. This will be the subject of Part 2. References 1. ISO 517, Measurement of fluid flow by means of orifice plates nozzles and ventury tubes inserted in circular cross section conduits running fuel, (2) 2. AGA-, Orifice Metering of Natural Gas and the Related Hydrocarbone Fuels, (18). Yeh, T.T and Mattingly, G. E, Flow meter installation effects due to a generic header. NIST Technical note 141, (1) 4. Gallagher J. Flow conditioning, Why, What, When and where, Siminarion Avanzado de Medition de Flujo de fluidos, Aguascalientes, Ags, Mexico, September 2 5. Rans, R. & al, Flow Conditioning and Effects on Accuracy for Fluid measurement, 7th East Asia Hydrocarbon Flow Measurement Workshop 5th 7th, Malisia, March (28). Laribi B. and al, Numerical simulation of flow development downstream four perforated plates for flow measurement accuracy, ASME Fluids Engineering Division Summer Meeting, FEDSM'14, Chicago, USA, July 214. 7. Bukhari Manshoor and Amir Khalid, Numerical Investigation of the Circle Grids Fractal Flow Conditioner for Orifice Plate Flowmeters, Applied Mechanics and Materials, Vol.22-21, pp. 7-74, Nov 2. 8. Laribi B. and Ait-Amrane A, Numerical analysis of the discharge coefficient with disturbers for flowmetring accuracy on flow measurement, 1th International Metrology Congress Paris, October 21.. Darin L. & Bowles E. B., Conditioning On Gas Measurement, Pipeline & Gas Journal, PP 52 (28) 1. Fluent v., Fluent incorporated, Journal Centerra source park, USA, 2. 11. B. Laribi and al, Numerical investigation of contribution of three flow conditioners in the development and establishment of turbulent flows, Proceedings of ASME 21 rd joint US-European Fluids Engineering Summer Meeting, and 8th International Conference on Nanochannels, Microchannels and Minichannels, FEDSM21- ICNMM21, Montreal, Canada, 21.. B. Laribi and al, Numerical investigation of turbulent models in the development and establishment of turbulent flows with flow conditioners, Proceedings of ASME 21 rd joint US-European Fluids Engineering Summer Meeting, and 8th International Conference on Nanochannels, Microchannels and Minichannels, FEDSM21-ICNMM21, Montreal, Canada, 21.. 5-p.5