Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers

Similar documents
Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers

Using Calibrated Specular Reflectance Standards for Absolute and Relative Reflectance Measurements

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Whole Tablet Measurements Using the Spectrum One NTS Tablet Autosampler System

RamanStation 400: a Versatile Platform for SERS Analysis

Instrumental Limitations for High Absorbance Measurements in Noise

Accurate Measurement of Transmittance and Reflectance for Engineering Applications

Traceable UV-Vis Reference Materials

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer

HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY

Model 556X User s Manual. Optical Isolator

A faster, more accurate way of characterizing cube beamsplitters using the Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS)

Characterization of the VIIRS Blackbody Emittance

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters

Experiment 3 1. The Michelson Interferometer and the He- Ne Laser Physics 2150 Experiment No. 3 University of Colorado

integrating-sphere-construction-kits

Integrating Spheres in Molecular Spectrophotometry

C101-E107B UV Shimadzu UV-VIS-NIR Spectrophotometer

Introduction to Fourier Transform Infrared Spectroscopy

Agilent Cary 630 FTIR sample interfaces combining high performance and versatility

The Plus for your lab!

Introduction to Fourier Transform Infrared Spectroscopy

Quantum Northwest, Inc. Product overview

Trace Analyses in Metal Matrices Using the ELAN DRC II

FOCUS 30/FOCUS 35 Field Calibration with Survey Pro Field Software

LED Measurement Instrumentation

Defining quality standards for the analysis of solid samples

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology

Interferometer for Squareness measurement

Regular Reflectance and Transmittance Measurements of Transmissive Materials Using a STAR GEM Optical Accessory

Whole Tablet Measurements Using the Frontier Tablet Autosampler System

AN AUTOMATED GONIOPHOTOMETER FOR LUMINAIRE CHARACTERIZATION

APPLICATION INFORMATION

Physics 476LW Advanced Physics Laboratory Michelson Interferometer

Your partner in Science!

LIBSlab ANALYZERS ANALYZERS

InfraRed Thermometer with Laser Pointer

AN INSTRUMENT FOR THE MEASUREMENT OF ROAD SURFACE REFLECTION PROPERTIES

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES)

novaa 800 D Atomic Absorption Spectrometer

THE MEASUREMENT OF SOLAR ULTRAVIOLET SPECTRAL IRRADIANCE PROBLEMS & SOLUTIONS

Physisorption of Antibodies using BioReady Bare Nanoparticles

High-throughput Quantification of DNA for NGS Library Prep with the Zephyr G3 Workstation and the VICTOR Nivo Plate Reader

Which Optistat is right for you?

custom reticle solutions

Eppendorf Vis Cuvettes. Instructions for use

ISO INTERNATIONAL STANDARD

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

Measuring Laser Diode Optical Power with an Integrating Sphere

Experiment 2: The Speed of Light

Nucleic Acid Quantitation in Microplates

Evaluating Labsphere s new UV-2000

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

SPECTROPHOTOMETERS. Visible and UV-Visible

RS-C Flexible Reticle Stocker

FLS980 Series Reference Guide

INSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7

Characterisation & Use of Array Spectrometers

HYDROBETA: A NEW INSTRUMENT FOR MEASURING IN-SITU PROFILES OF THE VOLUME SCATTERING FUNCTION FROM 10 TO 170 DEGREES

OPTICAL FIBRES IN ASTRONOMY (OP-006) Course Overview

ASMT-UYBG / ASMT-UYBH 0.25W Warm White Power PLCC4 Surface Mount LED. Datasheet

APPLICATION NOTE. Supercontinuum Generation in SCG-800 Photonic Crystal Fiber. Technology and Applications Center Newport Corporation

ICP-MS. High Resolution ICP-MS.

Inside Wall Temperature Measurements of DSTs Using an Infrared Temperature Sensor

Optics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P

application note LED measurement radiant power measurement

LAB DEMONSTRATION OF INTERFEROMETRIC

INTRODUCTION TO ENGINEERING SURVEYING (CE 1305)

HL-800K Infrared Laser Thermometer. -50 C to +800 C (13:1 ratio) User Manual

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Stray Light Rejection in Array Spectrometers

BIOLIGHT STUDIO IN ROUTINE UV/VIS SPECTROSCOPY

MELOS 530 MEASURING EQUIPMENT FOR LENSES AND OPTICAL SYSTEMS. Combinations and Accessories. Indispensable for Quality Control of Optics

Eppendorf UVette. Instructions for use

Polarized Light. Nikki Truss. Abstract:

CUSTOM RETICLE SOLUTIONS

Swamp Optics Tutorial. Pulse Compression

LASER TRAPPING MICRO-PROBE FOR NANO-CMM

Calibrating the Thermal Camera

Auto collimator. Introduction. Objectives: Apparatus: Theory:

EXPERIMENTAL DETERMINATION OF SPECTRAL AND ANGULAR DEPENDENT OPTICAL PROPERTIES OF INSULATING GLASSES

Water in Food 3 rd International Workshop. Water Analysis of Food Products with at-line and on-line FT-NIR Spectroscopy. Dr.

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U

Measuring the Universal Gravitational Constant, G

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Michelson Interferometer

L-malic acid measurement in wines using the SpectraMax Plus 384 Microplate Reader

symphony Reference Electrodes

THERMOFOCUS, a technological innovation

Coherence and width of spectral lines with Michelson interferometer

Technology that counts

Ultra-Small Footprint N-Channel FemtoFET MOSFET Test EVM

473 Dew Point Hygrometer

Name Final Exam May 1, 2017

Physical models for color prediction

Transcription:

Design Considerations for a Variable Angle Absolute Reflectance Accessory For the LAMBDA 950/850/650 UV/Vis/NIR and UV/Vis Spectrophotometers UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Introduction The accurate measurement of specular reflectance over a substantial range of angles and a wide range of wavelengths from the UV through NIR is a prerequisite to the design and manufacture of a wide variety of modern optical components. In designing a system for making these measurements, PerkinElmer set out to address the various limitations that effect currently available commercial spectrophotometer accessories. This application note discusses these limitations and describes how we have attempted to reduce them with the design of a variable angle absolute reflectance accessory. This device provides absolute reflectance over a wide range of angles of incidence for samples sizes from 5 mm to greater than 150 mm.

Requirements One of the primary design objectives of the variable angle absolute reflectance accessory was to address market requirements for the best possible accuracy, versatility, and convenient operation at an acceptable cost, without having to make any compromises. For routine measurements, an accuracy of 1% with repeatability of 0.2% was identified as an appropriate target specification. The desired measurement range was 250 to 3200 nm with the angle of incidence variable between 8 and 68 degrees. Due to the increased need to measure small samples, the minimum size was specified to be 5 mm, but at the same time, the need to handle large samples continues. It is difficult to specify convenience, but the need is for anyone used to handling optical components to be able to achieve the specified performance with reasonable care. VARIABLE ANGLE RELATIVE 45 Mirror 90 -D Prism Sample Figure 1. A variable angle relative reflectance sampling arrangement. b sample and reference are measured under identical conditions and the ratio of these measurements, after dark signal correction, gives the required value. A typical arrangement for variable angle reflectance V 45 V is shown in Figure 1. The major problem with this approach is the need for the reference mirror, with the possibilities of ageing and contamination creating uncertainty. At the same time, imperfections in the sample and variations in the sample positioning introduce errors associated with beam and detector homogeneity. Absolute measurements avoid the need for a reference mirror at the expense of introducing more complex optical arrangements. The principle is to make two measurements that differ only in that one includes reflection at the sample. Provided that propagation losses are identical, the ratio of the two values leads to the required reflectance. Current commercial spectrophotometer accessories are based primarily on either the V-N or V-W configura- N Figure 2. The V-N absolute reflectance geometry (simplified). N Sample Current methods There are two classes of measurement, relative and absolute. Relative measurements require a reference mirror of known reflectance. The Figure 3. The V-N absolute reflectance geometry. 2

V W been to use an integrating or collecting sphere to scramble the light input to the detector. For the present development an alternative method for light scrambling was investigated. This is to use a cylindrical light pipe with multiple internal reflections as the light is propagated. Figure 4. The V-W absolute reflectance sampling geometry. tions, named for their distinctive optical paths. In the V-N arrangement, rotation and translation of the output mirror allows a single reflection at the sample to be introduced without altering the direction of the output beam (Figure 2). In practice, more elaborate schemes are used to maintain the length of the optical path through the accessory (Figure 3). However, the output beam is inverted between the two configurations. This makes the accuracy critically dependent upon alignment and the beam and detector homogeneity. To overcome these problems an integrating sphere detector is often needed. The Strong V-W arrangement introduces two reflections at the sample (Figure 4). This has been seen as advantageous for samples with high reflectivity since the signal change to be measured is twice as large. However, the need for two reflections makes this unsuitable for small samples, especially at higher angles of incidence. It is also unsuitable for samples of low reflectivity. It is sometimes suggested that these problems can be addressed by using a mirror of known reflectivity alongside the sample to provide one of the reflections, but this is very inconvenient and removes the absolute nature of the measurement. Although there is no beam inversion the alignment is again critical. Although these absolute accessories are simple in principle, in practice they do not readily achieve the highest accuracy. Manufacturing tolerances limit the matching of the two configurations so that realignment between the configurations may be needed. This is typically done by incorporating a visible laser and adjusting the moveable mirror to maintain the spot from this at a constant position on a screen. The configuration of a commercial variable angle V-W accessory using this approach is shown in Figure 5. Although this method is successful, some users regard the inconvenience of the absolute technique as impractical for a day-to-day calibration service. Constraints The principal constraint for the absolute reflectance accessories is to match the output beam between two configurations. This matching is required because the beam intensity is not uniform and the PbS detector responsitivity varies across its area. Experience with current systems suggests that normal mechanical tolerances result in arrangements that require frequent readjustment. The usual way to address this has As well as the performance in light scrambling, the practicability of implementing this alternative has to be considered. The issues here are convenience and cost. For both simplicity and reproducibility, it is highly desirable that samples should be measured in a horizontal position. While a light pipe could represent a lower cost it might also result in an unacceptably large optical arrangement. LM1 Polarizer W k1 S4 V-W Battery unit M1 Laser unit S1 k3 M2 k2 M3 S3 LM2 Screen Reference beam Sample beam Laser alignment beam Marker 1 Marker 2 Figure 5. A commercial variable angle V-W accessory. S2 k4 3

Investigating light scrambling performance Two types of light pipe were considered: a hollow gold-plated tube where light is guided by reflection at the gold walls; and a glass rod which guides light by total internal reflection. As well as modeling the uniformity of output for an inhomogeneous beam, the sensitivity to lateral displacement and tilt of the beam were investigated. The results showed that a gold-plated light pipe would be too sensitive to displacement of the beam, but that a solid glass rod, 50mm long, has suitable characteristics. While an integrating sphere can achieve very good beam scrambling, the light pipe approach offers much higher optical throughput. Following this investigation, the glass rod light scrambler was adopted. To minimise the overall size of the optics this is incorporated in a Z-fold configuration with simple spherical mirrors to focus the input beam on to one end of the light pipe and then focus the output on to the detector. Figure 6. The URA sample measurement configuration at 68 0. The Universal Reflectance Accessory (URA): an innovative system design The basic operation of the system consists of first making a baseline measurement and then a sample measurement that involves one additional reflection at the sample. Switching between baseline and sample configurations involves rotating one mirror and translating the detector assembly automatically. This is illustrated below for measurement at 68 degree incidence (Figures 6 and 7). The input mirror rotates to direct the beam onto the measurement sample. At the same time the detector assembly is translated to maintain the same Figure 7. The URA baseline measurement configuration at 68 0. relationship with the input beam. The total optical path remains the same so that the ratio of the two measurements gives the reflectance of the sample. For measurements at different angles of incidence the input mirror is translated and rotated with the detector arc assembly being rotated, again under software control. The baseline and sample configurations for 8 degree incidence are illustrated below (Figures 8 and 9). There are additional optics to maintain the pathlength as the input mirror is moved for different angles of incidence. In addition there is a silicon detector to cover the UV-visible range. The beam scrambling light pipe is not 4

any measurement angle can be selected from within the system software. This allows reproducible measurement at multiple angles without manual adjustment to the optics or the sample. Only one accessory is required to cover the full range of angles between 8 and 68 degrees. Figure 8. The URA baseline measurement configuration at 8 0. The accessory also switches automatically between sample and baseline configurations without manual intervention. Additionally, the sample is simply placed on a horizontal sampling plate rather than being clamped vertically. This increases reproducibility and prevents delicate samples being scratched by the clamping mechanism. The whole URA optical unit is housed within a compact sampling module which protects the optics from dirt and dust and ensures that optical alignment is maintained. This unit can easily be interchanged with other sampling modules such as integrating spheres and transmission optics when other types of samples need to be measured. (Figure 10.) Figure 9. The URA sample measurement configuration at 8 0. required with the silicon detector which is extremely uniform. Switching between the detectors is automatic. The complete optical scheme with the various motions that are involved is shown in the previous figures. Despite the apparent complexity all the user sees is the aperture over which the sample is placed. No adjustments are required in order to make measurements over the full range of angles of incidence from 8 to 68 degrees. Conclusion There are several significant advantages to the URA design, compared with traditional approaches. Firstly, Figure 10. The URA sampling module can be simply interchanged depending on sampling need. 5

PerkinElmer Life and Analytical Sciences 710 Bridgeport Avenue Shelton, CT 06484-4794 USA Phone: (800) 762-4000 or (+1) 203-925-4602 For a complete listing of our global offices, visit /lasoffices 2004 PerkinElmer, Inc. All rights reserved. The PerkinElmer logo and design are registered trademarks and LAMBDA is a trademark of PerkinElmer, Inc. All other trademarks not owned by PerkinElmer, Inc. or its subsidiaries that are depicted herein are the property of their respective owners. PerkinElmer reserves the right to change this document at any time and disclaims liability for editorial, pictorial or typographical errors. 006991A_06 Printed in USA