BINOCULAR STEREOSCOPIC MICROSCOPE MBS-10. User Manual

Similar documents
Richter Optica. Instructions for Model: S850 Stereo Zoom Microscope

Electron Microscopy Sciences INSTRUCTION MANUAL. Stereo Microscopes Non-illuminated Stands

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

Kalinka Optics Warehouse User Manual Kobra AK Side Mount Red Dot Sight Manual

Z650HR SERIES MANUAL ZOOM STEREO MICROSCOPE. 73 Mall Drive, Commack, NY (P) (F)

Flexbar 6.3:1 MANUAL ZOOM STEREO MICROSCOPE. 250 Gibbs Road, Islandia, NY (P) (F)

EE 1/3.5х14 U EE 1/3.5х14 U

Owner's Manual NSL100B BUILDERS LEVEL NSL500B TRANSIT LEVEL

Federal State Unitary Enterprise КО-60. Registration certificate

State Unitary Enterprise Production Amalgamation Novosibirsk Instrument-Making Plant PN-6K. Servise Manual

Automatic Level Maintenance Manual SAL-XX W/ AIR DAMPENED COMPENSATOR

Directions for use

Altazimuth Mount. Altazimuth Mount

ZENITHSTAR 61 APO. Owner s Manual ver

Telescope Sight 4 X 24 for G3 Rifle

FSUE PA Novosibirsk. PO3-9x24, PO3-9x24-1. Instruction manual

SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL

COMMODORE BRASS TELESCOPE ZHUMELL COMMODORE BRASS TELESCOPE

product manual H-3220A Benkelman Beam

6XMag-1 Magnifier User manual

ZFK 4 X 25. Description and Utilization Instructions

22X Builder s Transit Level Model No Instruction Manual

5-25x56 PMII PSR With Double Turn

ZOOM STEREO MICROSCOPE

RP-200 TELESCOPE Instruction Manual

Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2

22X Builder s Level Model No Instruction Manual

Operating Instructions Spectro-Goniometer Student. 1 Functional Elements. 2 Safety Precautions. Figure 1: Spectro-Goniometer Student

saxon Instruction Manual saxon Grandeur Brass Telescope High quality optics

2-7x32. Instruction manual

RT-6 Riflescopes. User Guide

LEAPERS, INC. RED/GREEN DOTS

Spotting Scope Instruction Manual

2. OPERATIONAL CONDITIONS

ER2 Short-head Electric Chain Hoist

Operating Manual 1.1-4x24 Zenith x24 Zenith

RECON. How to Setup the Telescope to Observe. This guide will show you how to setup your telescope for observing. Written By: Brittany McCrigler

EPOCSELFIR LACITCATiX5T TECHNICAL MANUAL

FIELD GUIDE FOR HORUS SCOPES

CAUTION! NEVER ATTEMPT TO LOOK AT THE SUN WITH UNLESS YOU HAVE A FULL-APERTURE SOLAR FILTER (NOT INCLUDED WITH

5-25x56 PMII PSR Illuminated with Double Turn

RP-100 TELESCOPE Instruction Manual

Telescope. 1. Read these instructions carefully and familiarise yourself with the procedure before assembling the unit.

3B SCIENTIFIC PHYSICS

Instruction Manual. Model #: (Reflector) Lit #: / 06-07

AGES CF600PM. 50 mm Telescope w/ AZ Pan Tilt Mount Instruction Manual. WARNING! SUN HAZARD: Never look directly at the sun with this device.

Coherence and width of spectral lines with Michelson interferometer

USER MANUAL REFLEX SIGHT. Ultra Shot Plus

ULTRA SHOT PLUS - DARK EARTH REFLEX SIGHT SM26008DE

Owner's Manual. for AUTOMATIC LEVEL. NBL Series. FOR CUSTOMER SERVICE, PARTS & REPAIR, CALL Toll Free:

Self-Testing Prevectron 2 Millenium Lightning Conductor Set-up and Installation Manual

100 mm NEWTONIAN TELESCOPE (The Galileoscope) INSTRUCTION MANUAL IIA BANGALORE

Physics 476LW Advanced Physics Laboratory Michelson Interferometer

instruction Manual #9974 Customer Support (800) IN 293 Rev. C 02/09

BAADER MAXBRIGHT BINOCULAR VIEWER INSTRUCTION MANUAL

ML9000 SERIES POLARIZING MICROSCOPES FOR TRANSMITTED LIGHT MICROSCOPY

Instruction Manual. Made in Japan

HS515G-M. PARALOW Circle Dot Sight. User's Manual. Holosun Technologies Inc. Phone: (909) Fax: (909)

Table of Contents. Specifications 1 Adjustments 2 Using the Reticle 8 Maintenance 16 Troubleshooting 18 SS Warranty 19

Lamp and Control Panel (Lamp Panel)

NG114mm 114mm Newtonian Telescope w/ Equatorial Mount Instruction Manual

Melton Series. Owners Manual. Lab Centrifuge. Model # s A C, A C, B C, B C, C C, C C, D C, D C

Eclipse Refractometer General Models

FLT5A FLT5A. abc. abc. Thermometer User Guide. Thermometer User Guide. Internet:

ASTRO-PHYSICS, INC. POLAR ALIGNMENT TELESCOPE

Premier Heritage 3-15X50 User Manual

Optics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P

DIGITAL ABBE REFRACTOMETER INE-WYA-2S OPERATING INSTRUCTION PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

GEOMETRICAL OPTICS Practical 1. Part II. OPTICAL SYSTEMS

Platinum Series. 6-30x56 Front Focal Plane Scope. with Patent Pending DEKA MIL Reticle. DEKA MIL reticle for unsurpassed accuracy at long range

USER MANUAL. Wolverine FSR and CSR Series SM26020 SM26020DE SM26020-LQD SM26021 SM26021DE SM26021-LQD

P4Xi TACTICAL Riflescope TECHNICAL MANUAL

USER MANUAL. Wolverine FSR and CSR Series SM26020 SM26021 SM26020-LQD SM26020-LQD

C4 PLUS RIFLESCOPES User Guide

Instructions Manual Omegon Dobson Advanced 203

LABORATORY WRITE-UP MICHELSON INTERFEROMETER LAB AUTHOR S NAME GOES HERE STUDENT NUMBER:

SCHIEBER TELESCOPES. Unique, High-Quality Telescopes

'-. MICROMEr II. The MICROMET8 II is shipped assembled except for the Filar Eyepiece, Micrometer Stage, and Standard Vise.

Barograph 1.080x

Instructions Manual Zoomion Apollo 80

product manual H-4204 Proving Ring Penetrometer

BINOCULAR. 50 mm Tactical Binocular Manual HD R/T

Commercial Generator Parts Catalog. PROTEC Series

Sitting height... 6 Back tilt and back height... 6 Armrest... 7 Brakes... 7 Charging VELA Tango 300E... 7

INSTALLATION INSTRUCTIONS FRONT PLOW MOUNT KIT Part Number: Application: Polaris Sportsman 400, 500, 800

Basic Operation Adjust the eyecup

TELESCOPE TAL-150K TELESCOPE TAL-200K

HT600 Dat e Code: 2 - HT MW Purchase Date: PLEASE RETAIN THIS INSTRUCTION MANUAL FOR FUTURE REFERENCE. All Rights Reserved

Coma Correction. Diagram. Introduction

u s e r m a n u a l w w w. v e l a. E U

INSTRUCTION MANUAL FOR. TMM.41.X1/X2 Industrial Measuring Microscope. User s manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Instruction Manual. AC 70/700 AZ-2 English version Rev A

ASTRO-PHYSICS, INC. POLAR ALIGNMENT TELESCOPE (PASILL2)

取扱説明書 /INSTRUCTION MANUAL 自動レベル /AUTOMATIC LEVEL AT-B2/B3/B4 FC10386-A012-02

Model 791 Calibration Plate for Tiltmeters and Inclinometers

Ballistic pendulum Operating Instructions Fig. 1: Ballistic pendulum SAFETY PRECAUTIONS

RANGE-FINDERS FOR USE OF INFANTRY AND CAVALRY HANDBOOK OF WITH DESCRIPTION AND INSTRUCTIONS FOR THEIR CARE AND USE (TWELVE PLATES) DECEMBER 9, 1915

INTRODUCTION TO ENGINEERING SURVEYING (CE 1305)

Assembly Instructions for the 1-Wire Weather Station V2.0/V3.0

???? The Vortex Crossfire II Riflescopes

Transcription:

BINOCULAR STEREOSCOPIC MICROSCOPE MBS-10 User Manual

ATTENTION! Мicroscope design may differ somewhat from description in this manual. 2

1 - PURPOSE The MBS-10 microscope is intended for observing both three- dimensional and thin-film and transparent objects as well as for performing preparatory operations. The observation can be performed both in artificial and natural illumination in reflected and transmitted light. Field of application: botany, biology, medicine, mineralogy, criminalistics, archeology, mechanical engineering, instrument making industry and other fields of science and technology. 2 - SPECIFICATIONS 2.1 Magnification, times, in the range 2.2 Linear field of vision, mm, in the range 2.3 Operating distance, mm, no less 2.4 Illumination source 2.5 Overall dimensions in operating state (disregarding elbow-rest, eyepiece, illuminator), 4-100 39-2.4 95 halogen lamp 12V/20W mm, no more: length 26 width 5 height 16 2.6 Mass of device, kg, at most 0 2.7 Mass of device in packing, kg, at most 47 Information of precious and non-ferrous metals content Aluminum - 4.092 kg Copper - 0.34 kg Zinc - 0.053 kg 3 - COMPONENTS 3.1 Body with drum 3.2 Binocular attachment 3.3 Table for operating in reflected light 3.4 Table for operating in transmitted light 3.5 Power supply unit 3.6 Objective f'=90mm 3.7 Illuminator 3.8 Elbow-rest 2 pcs 3

3.9 Lamp-holder (to fix the lamp when operated in reflected light) 3.10 Eyepiece 8 X 2 pc 3.11 Eyepiece 14 x 2 pc 3.12 Eyepiece 8 X with scale 3.13 Eye shield 2 pc 3.14 Halogen lamp 12V/20W 3.15 Light filter 3.16 Clump 2 pc 3.17 Preparation glass 3.18 Plate 3.19 Rubber strap 3.20 Flannel cloth 3.21 Maintenance manual 1 copy 3.22 Case 3.23 Case for spare parts 3.24 Packing box 3.25 Fused plug VP-IV 1.0 A/250 V 2 pc 3.26 Fused plug VP-IV 3.15 A/250 V 2 pc 4 - DESIGN AND OPERATING PRINCIPLE 4.1 Microscope Optical Train An object image is formed due to the successive rays transmission through the f : =90mm objective 16; pair Galilean systems, set in the drum placed in the body placed in the body 1; tubules lenses and Schmidt prisms, been in the binocular attachment 5 and changeable eyepieces, which installed into the eyepieces tubes 11 (Figure 1). Galilean systems are alternatively switched on into the light beam path and allow realizing four versions of the microscope object part magnifications. The fifth version of magnification comes out if the Galilean systems are switched off from the light beam path. Magnifications values of the objective part of the microscope are given in Table 1. Table 1 - Magnification of objective part Linear magnification of Galilean system, mult. General linear magnification of objective part 1/3.5 1/2 1 2 3.5 0.6 1 2 4 7 The microscope is delivered with two pairs of changeable eyepieces (characteristics are given in Table 2) and eyepiece 8 x with changeable scale and 4

dioptric setting. The eyepieces are used for observing the image reproduced by the objective part of the microscope. On the eyepieces bodies are indicated the magnifications in round numbers. Table 2 - Characteristics of changeable eyepieces Magnification of Diam. of linear field of Range of output pupil, eyepiece, mult. vision, mm mm 8 23.0 18.0 14 16.0 16.0 Optical characteristics of the microscope with the use of each pair of the changeable eyepieces and of all the magnifications of the objective part are given in Table 3. Table 3 - Optical characteristics of the microscope Characteristics of the Magnification of Field of vision in the microscope eyepiece, mult. plane of object, mm Magnification of eyepiece Magnification objective part, 8 14 8 14 0.6 4.8 8.4 39.6 29.7 1 8.0 14.0 23.0 16.8 2 16.0 28.0 11.2 8.4 4 32.0 56.0 5.6 4.2 7 56.0 100.0 3.2 2.4 Schmidt prisms present a direct image of the object and allow to change the instrument interpupilary distance from 56 to 72 mm depending on the observer's eye. 4.2 Construction The general view of the microscope is given in Figure 1. The microscope consists of the following basic units: - optical head; - microscope table; - power supply unit. 5

Figure 1 - General view of the microscope I - body with drum; 2 - table for operating in reflected light; 3 - table for operating in transmitted light; 4 - dioptric setting ring; 5 - binocular attachment; 6 - handle of interpupilary distance change mechanism; 7 - table lock; 8,17 - screws, fixing the binocular attachment and objective f=90mm; 9 - illuminator bush; 10 - illuminator nut; II - eyepiece tube; 12 - mirror and clouded glass in mount; 13 - handle of magnification switching; 14 - light filter; 15 - stand; 16 - objective f=90mm; 18 - preparation glass; 19 - arm; 20 - elbow-rest; 21 - clamp; 22 - focusing handle; 23 - throw control; 24 - power supply unit; 25 ring The optical head consists of the body with the drum, the objective f'=90mm, the binocular attachment, the illuminator. The microscope table consists of the table for operating in reflected light and the table for operating in transmitted light. The necessary magnification is setting by rotation the handle 13 (Figure 1) to match the digit on the handle with the index point on the ring. The focusing of the microscope on the object is carried out by the optical head traveling relative to the microscope table along the guide line of "dovetail" type by rotation the handles 22 (Figure 1). ATTENTION! Rotation of focusing handles and magnification change handles in the opposite directions is prohibited. Throw control of focusing handles from slight to stiff is fulfilled by revolving the handle 23 (Figure 1). 6

4.2.1 Body with drum In the body of the optical head incorporates the drum with the Galilean systems. Rotating the handles 13 (Figure 1) which are fixed on the drum shaft changes over the value of magnification. On the handles 13 are indicated the magnifications in round numbers (7, 4, 2, 1 and 0.6 mult.). 4.2.2 Objective f'=90mm The objective is fixed to the body of the drum by the bayonet joint. The fixing of the objective is fulfilled by the screw 17. ATTENTION! To prevent dropping of the objective the screw 17 should always be screwed against the stop. 4.2.3 Binocular attachment The objectives and Schmidt prisms are installed in the binocular attachment 5 (Figure 1). The changing of interpupilary distance is fulfilled by rotation the Schmidt prisms in the opposite direction by the screw mechanism that is put in motion by handle 6 (Figure 1). ATTENTION! It is absolutely prohibited to change the interpupilary distance by moving the eyepiece tubes toward and apart by hand. The changeable eyepieces are installed in the eyepiece tubes. On the left eyepiece tube is a mechanism of dioptric focusing which is fulfilled in the limits of ±5 diopters by rotation the ring 4. The diopters nought is installed at the combination of index on the dioptric ring 4 with a mark on eyepiece tube. The binocular attachment is fixed to the body by the bayonet joint. Its fixing in the seat is fulfilled by the screw 8. 4.2.4 The microscope table On the stand 15 of the microscope table is fixed the optical head by the clamping screw which should always be tightened. To prevent an accident dropping of the optical head and for improvement of illumination in the right and left branches of the microscope on the stand is a clip 28 (Figure 2), fixed in the required position by screw 26 (Figure 2). The table has a round window (see Figure 1), in which the preparation glass or the plate 18, two holes for the clamp 21 and three holes for the substage ST-12 setting are installed, the latter is 7

Figure 2 - The microscope table not included in the microscope standard equipment and is purchased separately. The one side of the plate 18 is white and is intended for observing dark objects, the other side is black and is intended for observing light objects. The table 2 for operating in reflected light is mounted on the table 3 for operating in transmitted light and is fixed by the lock 7, faced to the front open side of the table body. There are mirror and clouded plate in mount 12 in the table 3. Their rotation is fulfilled by knob 27 (Figure 2). The table rear wall has a seat for fixing the illuminator in operating with reflected light. The side walls of the table have seats for mounting the elbow-rests. 4.2.5 Illuminator If operating at reflected light the illuminator is installed in the arm seat 19 (Figure 1) which makes possible to illuminate the object under different angles and from different sides. The arm 19 can be fixed in the required position by rotation the ring 25. The disassembled illuminator is represented in Figure 3 and consists of the lamp 1, the body 2, the seat flange 3, the chuck 4, the spring 5, the flex 6, and the bush 7. The microscope equipment has a light filter that is united through the thread to the seat flange 3 of the illuminator. The illuminator power can be controlled changing lamp power voltage. 8

Figure 3 - Illuminator When replacing the lamp it is necessary, keeping safety precautions listed in the paragraph 5.3, to weaken the flex 6 fixing (see Figure 3) by the bush 7 (counterclockwise rotation), screw the seat flange 3 out, remove the flex with the lamp 1 and with the chuck 4 out of the body 2, replace the lamp in the chuck. The assembling should by made vise versa. 4.2.6 Power supply unit The power of the illuminator lamp is provided through the power supply unit 24 (Figure 1) from alternating current mains Of 220 V using the flex with the plug of the side ground contact. The cover of the power supply unit has a connecter of 12V for the illuminator switching on (by means of two prongs). The side walls of the power supply unit have toggle switch, lamp voltage regulator, holders of safety devices (of fused plugs) of 1.0A and 3.15A. 5 SAFETY PRECAUTIONS 5.1 The instrument is intended for indoor operation under conditions of absence of increased electrical danger. The conditions increasing the mentioned danger are: a) increased humidity and dustiness of air; b) current-conducting floor: metal, brick, reinforced-concrete; c) temperatures above 40 0 C. 5.2 Regularly, before connecting the instrument to the power mains, the safety of the cord insulation, power supply unit and illuminator should be checked. 5.3 The lamp and fused plugs (safety devices) removing is required to make under de-energized illuminator and power supply unit. 9

6 PREPARATIONS FOR USE If the instrument is delivered indoors from cold, unpack it six hours after bringing it inside. After unpacking the instrument is needed to be prepared for the operating for which is purpose: a) install the body with the drum on the stand 15 and tighten the screw; b) install the binocular attachment in the bayonet seat of the body with drum and tighten screw 8. In operating with artificial lighting the binocular attachment 5 should be installed as it is shown in Figure 1. At the natural lighting it is necessary to rotate the binocular attachment at 180 0 C; c) install the objective f'=90mm and tighten screw 17; d) install the lamp in the lamp holder of illuminator and adjust illumination as instructed in paragraph 4.2.5; e) choose the pair of eyepieces of required magnification (see Table 2); f) install the eye shield that is included in the microscope standard equipment; g) check up the correspondence of magnification to the reading on the handle 13. Proceed as follows: 1 Take the binocular attachment 5 off. 2 Install the drum so that through the seat under the binocular attachment should be seen the apertures in the drum without optics and at both sides of these apertures should be seen the mounts with the small optics of the drum. 3 The index on the ring must coincide with figure 2 at the handle of the drum that is between 7and 4. If index don't coincide with the indicated figure, return the ring with the index around its axis to combine with the indicated figure 2. 7.1 General Indication 7 OPERATION PROCEDURE Focus the microscope on the object by rotation the handle 22 (Figure1). Install the interpupilary distance according to the base of observer's eye, by turning the handle 6. When operating at the large magnifications use the handle of throw control 23 at the focusing on the object. Choose the position of illuminator by turning it with the arm about the objective and choose the handy angle of slope. The dioptric focusing must be used after the microscope has been focus on the object on the right branch that doesn't have the dioptric focusing. 7.2 Operation with Eyepiece 8 x with the Scale 10

The eyepiece has a mechanism of dioptric setting. The scale is set in the focal plane; grid, which is delivered under the special order, can be set instead of the scale. The name of grid for special order "Grid in box. IAZU. 305648.010" The scale and the grid are actually plane-parallel glass round plates. The scale division value is 0.1mm; the grid square side value is 1.0mm To perform linear measurements or measurements of various object areas, insert the eyepiece 8 x with the scale into the one of the eyepiece tubes of the microscope. By the mechanism of dioptric setting of the eyepiece obtain a sharp image of the scale or the grid (depending on what is installed). Then, rotating the knob of the focusing mechanism obtain the sharp image of the object. The table 4 shows to what actual value of the object corresponds one division of the scale or the grid at all values of magnification of the microscope. Table 4 - Conversion table of the microscope magnifications Magnifications in round numbers on the handles of the drum, mult. Scale one division, Square side, 1mm 0.1mm Corresponds to actual value of the object 0.6 0.17 1.7 1 0.1 1.0 2 0.05 0.5 4 0.025 0.25 7 0.014 0.14 In order to determine the approximate dimensions of the object (its linear dimensions or area) it will suffice to count the number of the division of the scale which are laid in the measured section of the object and to multiply the obtained result by the number indicated in the conversion. Table corresponding to the magnification of the microscope at which the measurement is being performed. To replace the scale (or the grid), proceed as follows: 1 turn the scale rim out of the eyepiece body. The rim is found in the bottom part of the eyepiece; 2 turn out the nut fastening the scale (or the grid) and remove carefully the scale from the rim; 3 install the grid into the rim and tighten up the nut; 4 turn the rim into the body of eyepiece. 8. MAINTENANCE When receiving the microscope pay attention if the Manufacturer's seal is intact. The microscope leaves the Manufacturer duly tested and can operate free of trouble for a long time, provided it is kept clean, properly cared of and 11

prevented from any kind of mechanical damage. The Manufacturer's package ensures safety of the microscope during its transportation. When not using, the microscope should be covered by the case. To keep the microscope in good appearance, it should be wiped periodically, excepting optical surface, by a soft cloth impregnated by acid-free vaseline after which the instrument should be dried by a soft clean cloth. If after a certain time the grease in the focusing mechanism guides becomes dirty and thick, remove it by using xylene or benzene, wipe the friction surfaces with a clean soft cloth and greasy slightly the guides with the acid-free vaseline or special grease. Pay special attention to the cleanness of the microscope optical elements. During the operation the optical surface has to be prevented from moisture. To prevent the prisms surfaces from dust, the eyepieces should be always left in the eyepieces tubes of the microscope. The eyepieces have to be also prevented from dust. The lenses surfaces should not be touched by fingers and hard objects in order to avoid a spoilage of coated surfaces. When cleaning the lenses surfaces, dust and other soiling should be removed by using a soft clean cloth. If after that the optical surfaces are insufficiently clean, use a soft batiste cloth slightly moistened by spirit. 9 ACCEPTANCE CERTIFICATE The MBS-10 microscope, serial No.. meets the technical specifications and has been found fit for service. QCD Representative.. «..» 20 10 PRESERVATION AND PACKING REPORT The MBS-10 stereoscopic microscope, serial No has been subjected to preservation treatment and duly packed at the Manufacturing factory in compliance with the requirements prescribed by the technical documentation. Preserved and packed on Preservation and packing were made by 295-03 12