Adaptation of PTB s analytical modelling for TDCR-Cherenkov activity measurements at LNE-LNHB. LSC 2017, Copenhagen, 1-5 May 2017

Similar documents
RADIOPHARMACEUTICAL 11 C ACTIVITY MEASUREMENTS BY MEANS OF THE TDCR-CERENKOV METHOD BASED ON A GEANT4 STOCHASTIC MODELING

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

Čerenkov counting and liquid scintillation counting of 36 Cl

TDCR in a nutshell. P. Cassette, Laboratoire National Henri Becquerel, France

Activity determination of 88 Y by means of 4πβ(LS)-γ coincidence counting

Quantum Mechanics. An essential theory to understand properties of matter and light. Chemical Electronic Magnetic Thermal Optical Etc.

APPLICATION OF A FREE PARAMETER MODEL TO PLASTIC SCINTILLATION SAMPLES

BUREAU INTERNATIONAL DES POIDS ET MESURES

Fundamentals of Radionuclide Metrology

Activity measurement of 55 Fe within the scope of the BIPM comparison 2006

C2-05: Creation of national standards for some emerging pharmaceutical radionuclides to ensure the radioprotection of patients and medical staffs

Scintillators General Characteristics

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

Metrology for Molecular Radiotherapy MetroMRT V. Smyth (coordinator) Lena Johansson, Andrew Fenwick, Kelley Ferreira et al.

Summary of the ICRM Life Sciences Working Group Meeting

CHAPTER 5 Wave Properties of Matter and Quantum Mechanics I

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2

Radionuclide metrology using LSC. - Current status and limitations -

Standardization of 18 F by Digital β(ls)-γ Coincidence Counting

Precise Measurement of the Absolute Yield of Fluorescence Photons in Atmospheric Gases

The metrology of radioactivity. C. Michotte, BIPM

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

Determination of the activity of radionuclides

Guidance for best practice from MetroMRT. Lena Johansson, NPL

Chemical Engineering 412

Lecture 16 Light transmission and optical detectors

Wave Motion. Chapter 14 of Essential University Physics, Richard Wolfson, 3 rd Edition

TDCR and CIEMAT/NIST Liquid Scintillation Methods applied to the Radionuclide Metrology

Interaction with matter

IMPROVED TECHNIQUES FOR THE ACTIVITY STANDARDIZATION OF 109 Cd BY MEANS OF LIQUID SCINTILLATION SPECTROMETRY

Photons in the universe. Indian Institute of Technology Ropar

Recent Primary Standardisations and Traceability in Nuclear Medicine. Lena Johansson, NPL Radioactivity Group

NEW DEVELOPMENTS OF AUTORADIOGRAPHY TECHNIQUE TO IMPROVE ALPHA AND BETA MEASUREMENTS FOR DECOMMISSIONING FACILITIES

CHAPTER 4 Structure of the Atom

Disproof of solar influence on the decay rates of 90 Sr/ 90 Y

Scintillation Detectors

Acronyms, Abbreviations, and Symbols Foreword to the First Edition Foreword to the Second Edition Preface to the First Edition Preface to the Second

arxiv:physics/ v1 3 Aug 2006

MEDICINSK STRÅLNINGSFYSIK

WbLS measurements at BNL

Chapter 4 Scintillation Detectors

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

Standardization of Tritium by CIEMAT/NIST Method and TDCR Method

Radioactivity standardization in South Africa

SOME ASPECTS OF MONTE CARLO SIMULATION FOR EFFICIENCY CALIBRATION OF GERMANIUM DETECTORS

Development of a High Precision Axial 3-D PET for Brain Imaging

Determination of the boron content in polyethylene samples using the reactor Orphée

TIME-RESOLVED LIQUID SCINTILLATION COUNTING

Time-Resolved Liquid Scintillation Counting

Aerogel counter with a Fresnel lens. Guy Paic Instituto de Ciencias Nucleares UNAM Mexico

INTRODUCTION TO QUANTUM MECHANICS

Interaction of particles with matter - 2. Silvia Masciocchi, GSI and University of Heidelberg SS2017, Heidelberg May 3, 2017

Validation of detection efficiency of alpha particles in commercial liquid scintillation counters

Overview Physical processes PMT and electronics response Some results Plans

Elastic light scattering

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data

Radionuclide Imaging MII Detection of Nuclear Emission

PHYS 3650L - Modern Physics Laboratory

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Monte Carlo Simulations for Future Geoneutrino Detectors

Radioactivity. Lecture 6 Detectors and Instrumentation

Cherenkov Detector. Cosmic Rays Cherenkov Detector. Lodovico Lappetito. CherenkovDetector_ENG - 28/04/2016 Pag. 1

Basics of a/b Discrimination for Liquid Scintillation Counting

MONTE CARLO POWER ITERATION: ENTROPY AND SPATIAL CORRELATIONS

Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado

Proton decay and neutrino astrophysics with the future LENA detector

MAGNETIC ANISOTROPY IN TIGHT-BINDING

A Liquid Argon Scintillation Veto for the GERDA Experiment

Characterization of high temperature solar thermal selective absorber coatings at operation temperature

EXAMEN GÉNÉRAL DE SYNTHÈSE ÉPREUVE ÉCRITE Programme de doctorat en génie physique. Jeudi 18 juin Salle B-508.

Gamma ray coincidence and angular correlation

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

Radiation Detection and Measurement

Status report on parity violation in the (1232) resonance

Photoneutron reactions studies at ELI-NP using a direct neutron multiplicity sorting method Dan Filipescu

Radioactivity and Ionizing Radiation

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Liquid Scintillation Counting Performance Using Glass Vials in the Wallac 1220 QuantulusTM

Development of Radioactivity Standards for Quantitative Positron Emission Tomography

Primary Standardisation of I-125

MEASUREMENT AND DETECTION OF RADIATION

The Bose Einstein quantum statistics

ETNA (Efficiency Transfer for Nuclide Activity measurement)

So, you want to build a neutrino detector?

Development of Continuous Inflow Tritium in Water Measurement Technology by using Electrolysis and Plastic Scintillator

INORGANIC crystal scintillators are widely used in high

New Results from the DREAM project

19:00 21:30 Registration and reception at Hotel ETAP Altinel. Welcome

Scintillation detectors

ARTICLE IN PRESS. Applied Radiation and Isotopes

Nuclear Physics and Astrophysics

J. Keightley 1 P. Cassette 2, L. Johansson 1

Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber

EDS User School. Principles of Electron Beam Microanalysis

Platinum resistance. also wirewound versions. eg

SCINTILLATION DETECTORS AND PM TUBES

Neutrino Physics. Neutron Detector in the Aberdeen Tunnel Underground Laboratory. The Daya Bay Experiment. Significance of θ 13

Radioactivity. The Nobel Prize in Physics 1903 for their work on radioactivity. Henri Becquerel Pierre Curie Marie Curie

Transcription:

Adaptation of PTB s analytical modelling for TDCR-Cherenkov activity measurements at LNE-LNHB Cheick THIAM, Christophe BOBIN and Jacques BOUCHARD LSC 2017, Copenhagen, 1-5 May 2017

General introduction TDCR method (Triple to Double Coincidence Ratio) based on Cherenkov counting Cherenkov radiation: discovery by Cerenkov (1937). Phys. Rev. 52, 378. Theoretical interpretation by Frank and Tamm (1937) J. Phys. 1, 439 454. Electromagnetic shockwave resulting from a charged particle moving in a material faster than the velocity of light in that medium In radionuclide metrology only electron or positron can produce Cherenkov light Photons emitted as a cone with spanning angle θθ: cccccc θθ = 1 Condition of Cherenkov emission: nnnn 1 wwwwwwwww ββ = vv cc Main physical proprieties: nnnn Threshold effect (for electron in water: n ~ 1.33; E th ~ 260 kev) Directional character (not isotropic emission) Large spectral bandwidth (comprised between UV and visible wavelengths) 2

General introduction TDCR method (Triple to Double Coincidence Ratio) based on Cherenkov counting Applied for β-emitting radionuclides, use existing LS counters: Direct measurements with aqueous solutions (easy source preparation) Drawback: counting efficiencies lower than LS counting (due to the threshold effect) Natural discrimination for low-energy β particles and for alpha particles Vial A B C Coincidences AB, BC, AC D, T RCTD=T/D Free parameter model of light emission + Radionuclide data PMT Preamplifiers TDCR Calculation algorithm ε = f(rctd) Activity As for LS, the free parameter is adjusted for the calculation of the detection of double coincidences using experimental TDCR values The model has to take into account the physical properties of Cherenkov emission 3

General introduction Towards model of Cherenkov counting for radionuclides standardization: A free parameter model of light emission (to compute Cherenkov counting efficiencies) first proposed by Grau Carles and Grau Malonda (2006) Developed for Cherenkov counting standardization with detection systems using two PMTs Kossert (2010) and Kossert et al. (2014): extends this model for standardizations with TDCR detection systems Based on adaptation the statistical model usually carries our for LS counting New empirical formula to account for anisotropy of Cherenkov emission Bobin al.(2010) and Thiam et al.(2011) describe a stochastic approach based on Monte Carlo simulation using Geant4 code 4

TDCR-Cherenkov model Monte Carlo Model using Geant4 code: Comprehensive modelling of detection setup Simulates interactions of ionizing radiation and propagation of Cherenkov photons from their creation to the production of photoelectrons in PMTs Geometry modelling, EM physics Optical processes: refraction, reflection, transmission and absorption Model works for measurements with glass vials Geometry model with Geant4 Detector parameter Material Optical parameter PMT-window (Ø = 52 mm) Fused silica Dispersive refractive index (DRI) ~ 1.47 at 400 nm and 1.64 at 160 nm Surface type (ST): dielectric-dielectric. Skin: polished PMT-photocathode (Ø = 46 mm) Bi-alkali (K 2 CsSb) DRI ~ 2.5 at 430 nm, Harmer et al. (2006). ST: dielectric-dielectric Optical chamber Teflon ST: dielectric-metal, Lambertien-type reflectivity 95% Vial (1 mm layer) Borosilicate DRI ~ 1.52 at 430 nm. ST: dielectric-dielectric, Skin: polished Aqueous solution (15 ml) Water DRI ~ 1.33 at 2600 nm 5

TDCR-Cherenkov model Monte Carlo Model using Geant4 code Detection efficiency variation implemented using the PMT-defocusing technique Activity calculated with TDCR-Geant4 model validated for an aqueous solution of 90 Y 0.70 ε D 0.69 kcounts.s -1.g -1 s -1.g -1 814 812 810 808 806 804 802 800 798 Detection efficiency 0.68 0.67 0.66 0.65 0.66 0.67 0.68 0.69 0.70 0.71 0.72 0.73 0.74 TDCR ε D =f(tdcr) given by Geant4 Activity concentration / kbq.g -1 kbq.g -1 1205 1204 1203 1202 796 0.675 0.680 0.685 0.690 0.695 0.700 0.705 TDCR Double coincidence rate according to TDCR More details on: Appl. Radiat. Isot. 68 (2010) 2366 2371. 1201 0.675 0.680 0.685 0.690 0.695 0.700 0.705 TDCR TDCR Activity concentration according to TDCR No significant trend 6

TDCR-Cherenkov model Monte Carlo Model using Geant4 code Well validated for several radionuclides 90 Y, 32 P, 11 C, 68 Ga/ 68 Ge. Appl. Radiat. Isot. 68 (2010) 2366 2371 TDCR Counter Applicable also for liquid scintillation counting 3 H, 63 Ni Appl. Radiat. Isot. 70 (2012) 2195 2199 4πβ(LS)γ coincidences method γ-detector 56 Fe, 54 Mn, 14 C. Appl. Radiat. Isot. 109 (2016) 319 324 Useful to study the contribution of several parameters and physical effects in counting Volume effect of the source and vial Contribution of bremsstrahlung, 511 γ-annihilation Variation of optical parameters C++ programing, long computing time 7

TDCR-Cherenkov model Adaptation of PTB s analytical model at LNHB Objective: test an extension of PTB model for TDCR-Cherenkov measurements by considering physical properties of the detection setup in use at LNHB Calculation implemented using MATLAB computing environment The three-pmts counter used for measurements: equipped with X2020Q PMTs (Photonis) Fused silica window, large spectral sensitivity (160 to 600 nm) Quantum efficiency ~ 24% (300 to 400 nm) 3-PMT TDCR at LNHB Spherical optical chamber made of Teflon (~ 95 % Lambertian reflexion) 8

TDCR-Cherenkov model Adaptation of PTB s analytical model at LNHB Measurements carried out with plastic vials to attenuate the geometry dependence of coincidence counting 88 s -1.g -1 86 Polyethylene vial Glass vial Comparison between glass-vials and plastic-vials for measurements of 90 Y Significant shift observed kcounts.s -1.g -1 84 82 80 78 76 0.66 0.68 0.70 0.72 0.74 0.76 0.78 0.80 0.82 0.84 TDCR TDCR Because of geometrical effect due to their diffusive wall, reflection and refraction process taking place at the wall/air boundary are modified Better detection efficiencies are obtained with usual LS plastics vials 9

TDCR-Cherenkov model Description of the analytical modelling Using Frank and Tamm (1937) theory; the number of Cherenkov photons dddd emitted by an e - or e + along a path dddd for an energy EE and λλ is given by: αα FFFF : fine structure constant nn(λλ): refractive-index of transparent medium dddd dddddddd = 2ππαα 1 FFFF λλ 2 1 1 ββ 2 nn 2 (λλ) In term of energy variation ( dddd dddd from ESTAR database) the equation became: dddd(ee, λλ) dddddddd = 2ππαα FFFF 1 λλ 2 1 1 ββ 2 nn 2 (λλ) 1 ρρ dddd dddd The mean number of photoelectrons distributed in PMTs is obtained by the integration of the equation between EE cc and the energy released in the aqueous solution. Considering PMT-spectral response Taking into account the variation of the Cherenkov threshold E c with decreasing photon wavelengths NOTE: Optical properties of vials are not taken into account 10

TDCR-Cherenkov model Description of the analytical modelling The empirical probabilities RR 1, RR 2 and RR 3 to have at least one count obtained as for LS counting i.e. we can apply an adjusting free parameter (q) (based on a Poisson distribution assumption) RR 1 EE el = 1 ee qqαα 1kk EE el αα 1 = αα EE el RR 2 EE el = 1 ee qqαα 2kk EE el αα 2 = 3 2 1 αα EE el αα EE el cos θθ = 1 nnnn RR 3 EE el = 1 ee qqαα 3kk EE el αα 3 = 1 αα 1 αα 2 Ω Anisotropy described with αα EE el, expressed as a function of Ω PTB LNHB Ω = 2ππ (1+ ssssss θθ ) αα EE el = xx 2 3 1 Ω 3 4ππ = xx 1 2 + 1 cccccc θθ 6 αα EE el = xx 2 3 1 Ω 3 4ππ = xx 1 2 1 ssssss θθ 6 if θ 0, Ω = 0 if θ 0, Ω = 2ππ Formula slightly different so as to get an equi-probable distribution of photons between PMTs by setting the free parameter xx = 1 11

TDCR-Cherenkov model Description of the analytical modelling The calculation of detection efficiencies of triple and double coincidences εε TT and εε DD Implemented as for LS measurements using the probabilities RR 1, RR 2 and RR 3 and the probability density function SS EE el related to the energy distribution of particle εε TT = EEccEE max SS EEel RR 1 RR 2 RR 3 ddee el εε DD = EEccEE max SS EEel RR 1 RR 2 + RR 2 RR 3 + RR 1 RR 3 2. RR 1 RR 2 RR 3 The detection efficiency determined by adjusting the free parameter to match the measured TDCR values Related publications: Kossert, Appl. Radiat. Isot. 68 (2010) 1116 1120. Kossert, Grau Carles, Nähle, Appl. Radiat. Isot. 86 (2014) 7 12. 12

Application of TDCR-Cherenkov modelling Activity measurements of solution of dissolved 90 Y-labelled microspheres Standardization of 90 Y-labeled microspheres (SIR-Spheres, Sirtex) Medical device used in Selective Internal Radiation Therapy (SIRT) Biocompatible microspheres (20-60 µm) containing 90 Y high-energy β - ; main branch (99.98%); Emax ~ 2.28 MeV, Emean ~ 927 kev Measurements carried out after dissolution of microspheres Avoid problems of non-homogeneity Dissolved solution subsequently diluted to reduce colour quenching NOTE: More details in Lourenço et al. (2015) Appl. Radiat. Isot. 97, 170-176. Sources preparation with plastic vials 15 ml of carrier solution (25 mg/g of Y in 0.04 M HCl) 9 mg of dissolved and diluted solution of 90 Y-microspheres Results compared to classical TDCR LS counting with glass vials 10 ml of Hionic-Fluor +9 mg of dissolved solution of 90 Y-microspheres After dissolution (no residues) 13

Application of TDCR-Cherenkov modelling Activity measurements of solution of dissolved 90 Y-labelled resign microspheres Three sources measured in standard plastic vials (with out coated Teflon in wall) The TDCR values: comprised between 0.8-0.815, corresponding detection efficiencies : ~ 72% Two sources measured using Teflon coated plastic vials The TDCR values: ~ 0.8, corresponding detection efficiencies : ~ 71.5 The activity concentration with TDCR-Cherenkov consistent with classical TDCR LS Relative difference: 0.2% with standard plastic vials; 0.26% with Teflon coated plastic vials Uncertainty budget of 90 Y measurements using the analytical modelling Uncertainty component Comments u (%) Measurement variability Standard deviation of the measured sources 0.1 Weighing Gravimetric measurements (pycnometer method) 0.1 Live time 1-MHz frequency clock used for the live-time 0.01 Decay correction Half-life of 90 Y: 2.6684 (13) d 0.05 Analytical modelling Conservative estimation 0.5 Background 0.05 Relative combined standard uncertainty 0.52 14

Application of TDCR-Cherenkov modelling Activity measurements of 89 Sr Suited radionuclide for Cherenkov measurements High-energy β - ; main branch (~ 99.99%) with E max ~ 1495 kev, E mean ~ 585 kev Sources preparation Four sources measured in standard plastic vials 15 ml of carrier solution (10 mg/g of Sr in HCl 0.1 M) + 30 to 100 mg of 89 Sr solution Results: Maximum TDCR values: ~ 0.665 Corresponding detection efficiencies computed with the analytical modelling: ~ 53% Activity concentration given by TDCR-Cherenkov: 0.26% lower than the result with TDCR LS Main uncertainty component due to the analytical modelling: 0.5% 15

Application of TDCR-Cherenkov modelling Activity measurements of 68 Ga in a solution of 68 Ge/ 68 Ga in equilibrium The activity measurements carried out in the framework of a BIPM international comparison using a radioactive solution prepared at NIST 68 Ge/ 68 Ga: favorable to standardization by Cherenkov measurements via 68 Ga Mainly decaying by β + emissions ~ 88.9%; E max ~ 1899 kev; T 1/2 ~ 67.83 min Take advantage of the Cherenkov threshold to avoid the contribution of EC emissions of 68 Ge Suitable for PET imaging as a surrogate of 18 F Simplified decay scheme of 68 Ge - 68 Ga Sources preparation Six sources in Teflon coated plastic vials 15 ml of carrier solution (65 mg/g of Ge 4+ and Ga 3+ in 0.5 M HCl) + 10 mg of 68 Ge/ 68 Ga aliquot 16

Application of TDCR-Cherenkov modelling Activity measurements of 68 Ga in a solution of 68 Ge/ 68 Ga in equilibrium Results: Efficiency calculation with analytical model take into account the two β + spectra and related branching ratios (ββ + 0,0, EE mmmmmm ~ 1899 kev; ~ 87.7 %; ββ + 0,1, EE mmmmmm ~ 821.7 kev, 1.2 %) The contribution of 511 kev annihilation photons also considered using the resulting energy spectrum in the aqueous solution (obtained by Monte Carlo simulation) Maximum TDCR value: 0.746 Corresponding detection efficiency computed with the analytical modelling: 73.4% Contribution of 511 kev γ-photon interactions: ~ 0.2% TDCR-Cherenkov compared with 4π(LS)β γ anticoincidence measurements Both results are in good agreement with 0.16% relative difference 17

Discussion and conclusion First developed at PTB for TDCR-Cherenkov measurements, the analytical model has been well adapted to operate with a TDCR counter in use at LNHB Implementation inspired from the classical statistical model used for LS counting The modelling is based on several approximations Optical properties of the vials are not taken into account (optical transmittance, refractive indices) The anisotropy of emission is considered by means of empirical expressions and by using an additional free parameter Depending on E(e - /e + ), xx and q, the mean number of photoelectrons (therefore probability to obtain a count in PMTs) is based on Poisson-distribution The contribution of different physical effects may compensated by additional diffusing-effect when using plastics vials 18

Discussion and conclusion Despite approximations, Cherenkov measurements using the analytical model give consistent results with a conservative model-uncertainty of 0.5% The model successfully tested at LNHB for three standardizations of 1.008 1.006 TDCR-Cherenkov vs expected activity value three radiopharmaceuticals: 90 Y, 89 Sr and 68 Ge Reliable activity measurements Activity ratios 1.004 1.002 1.000 0.998 0.996 0.994 90 Y (1) 90 Y (1,*) 90 Y (2) 89 Sr (*) 68 Ga already reported by Kossert et al. 0.992 (2014): 0.990 32 P, 89 Sr, 90 Y, 204 Tl, 106 Rh (1) dissolved and diluted solution of 90 Y microspheres (2) aqueous solution of 90 Y (*) plastic-vial without Teflon coating in inner wall 19

Thank you. Commissariat à l énergie atomique et aux énergies alternatives Institut List CEA SACLAY NANO-INNOV BAT. 861 PC142 91191 Gif-sur-Yvette Cedex - FRANCE www-list.cea.fr Établissement public à caractère industriel et commercial RCS Paris B 775 685 019