Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Similar documents
Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Trace Analyses in Metal Matrices Using the ELAN DRC II

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

Analysis Repeatability of Trace and Major Elements in a Water Sample

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

OES - Optical Emission Spectrometer 2000

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Samples compliant with the WATER

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Fergus Keenan Shona McSheehy Julian Wills

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Agilent Technologies at TIAFT 2013

ICP-OES Application Note Number 35

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry: acquiring basic skills and solid sample meausurements

Analysis of noble metals at low levels in geological reference materials and ores

Agilent ICP-MS Journal September 2009 Issue 39

IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Agilent 8900 Triple Quadrupole ICP-MS

The New Agilent 7700 Series ICP-MS

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical

ICP-Mass Spectrometer

Introduction to Laser Ablation ICP-MS for the Analysis of Forensic Samples Application Note

Autenticity control, provenance testing and fraud detection using mass spectrometry

Florida State University Libraries

Analysis of domestic sludge using the Agilent 4200 MP-AES

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Sample Analysis Design PART II

Elemental Scientific. Spike Recovery. Elemental Scientific

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Accurate Determination of TiO 2 Nanoparticles in Complex Matrices using the Agilent 8900 ICP-QQQ

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Advanced ICP-MS techniques for overcoming interferences in LA-ICP-MS bioimaging

Reporting Category 1: Matter and Energy

Sample Analysis Design PART III

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Outsmart Interferences with Triple Quadrupole ICP-MS. The world leader in serving science

Transcription:

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode Application Note Pharmaceutical Authors Rebeca Santamaria-Fernandez, SheilaMerson, and Ruth Hearn LGC, Queens Road Teddington, Middlesex, TW11 0LY UK Abstract Rapid semiquantitative elemental screening of pharmaceutical samples is described using the Agilent 7500ce ICP-MS with helium as collision gas. Polyatomic interferences are eliminated in the He pressurized cell, enabling rapid elemental profiling of pharmaceutical tablets for screening purposes. Elements which normally suffer from polyatomic interferences have been included in the semiquantitative analysis and trace metal profiles obtained for the pharmaceutical tablets from three different batches agreed with those obtained by fully quantitative analysis. The method has potential therefore as a fast analytical tool in the pharmaceutical manufacture chain.

Introduction The control of pharmaceutical impurities is currently a critical issue to the pharmaceutical industry [1]. Most pharmaceutical products are specified to be essentially free of inorganic impurities. The U.S. Food and Drug Administration (FDA) and the British Pharmacopeia (BP) strongly advise that contamination problems be fully investigated in a timely fashion; therefore, fast screening methods for the detection of inorganic contaminant impurities is required. The presence of contaminants in pharmaceutical products may be related to the active ingredient, excipient materials, or colorant. Contamination can also result from the manufacturing process; examples include detergents and lubricant oils and catalyst residues [1]. The production chain involves, in many cases, the use of mixing/reaction tanks, water, filters, or other equipment. Failure of any of those to be sufficiently clean might lead to inorganic contamination of the end product. Thus, trace elemental composition of a substance is commonly used to fingerprint unknown samples and can also give information regarding the manufacturing process. It has also been shown that in some pharmaceutical products, traces of inorganic impurities can clearly reduce drug stability and shelf life [1]. Methods for the rapid assessment of inorganic impurities are therefore of high interest to the pharmaceutical industry for rapid quality control and trace elemental profiling of a large number of samples of unknown composition. Thus the presence of inorganic contaminants can be quickly detected and further action can be taken. In this work, the potential of semiquantitative elemental screening by inductively coupled plasma mass spectrometry (ICP-MS) with collision/reaction cell technology (CRC) using helium gas is demonstrated. Semiquantitative analysis has proved useful in situations where either calibration standards are not available or a rapid screening of a large number of unknown samples is required [2 4]; however, little has been published about its applicability to real-world samples. Recent advances in reaction cell technology have shown the ability of the Octopole Reaction System (ORS) to eliminate polyatomic interferences using controlled kinetic energy discrimination (KED) in helium collision mode, enabling relatively accurate semiquantification in unknown matrices [5]. Semiquantitative elemental screening has been performed for a suite of 29 elements, and quantitative measurements have been performed for comparison. Experimental Sample Preparation Six pharmaceutical tablets from different production batches of an antiviral drug were used in the study. The tablets were crushed with a pestle and mortar. Approximately 0.06 g of sample was accurately weighed and microwave digested with 4 ml of high-purity nitric acid (Romil, Cambridge, UK) and 2 ml of high-purity hydrogen peroxide (Romil, Cambridge, UK). The microwave program consisted of heating the samples to 180 C over 10 minutes and holding for a further 20 minutes. Once cool, digests were made up to 30 g using deionized water, and the solutions were subjected to element determination by ICP-MS. Blanks and quality control materials were included in each digestion run. Aquacheck solutions used in the Aquacheck proficiency testing scheme, with known concentrations of the analytes (ranging from 2 ng/g [ppb] of Cd to 11 µg/g [ppm] of Ca), were analyzed as independent quality checks (QC) on the accuracy and precision of each ICP-MS run. Instrumentation Sample digestion was performed in a Multiwave 3000 microwave (Perkin Elmer, Beaconsfield, UK). Measurements were performed using an Agilent 7500ce collision/reaction cell inductively coupled plasma mass spectrometer (CRC-ICP- MS) operating in helium semiquantitative mode for the elemental screening. Tuning conditions are given in Table 1. The Integrated Sample Introduction System (ISIS) was used with the nebulizer pump speed set at 0.1 rps during the analysis and washout in order to minimize overloading of the sample introduction system and the plasma with matrix components. The torch was equipped with a 2.5 mm diameter injector. A 30 second stabilization time was used for each cell gas mode when performing quantitative measurements in different gas modes. Measurements were performed in triplicate and a QC sample was included in each run. 2

Table 1. Tune Conditions Used for SemiQuant Analysis in Helium Collision Mode. Parameter RF power Sample depth Carrier gas Make-up gas Sample flow rate Nebulizer Value 1520 W 8.0 mm 0.88 L/min 0.27 L/min 0.5 ml/min Glass concentric, MicroMist Spray chamber Quartz cooled to 2 C Interface cones Cell gas Cell gas flow rate KED voltage Ni He 4.6 ml/min +2 V Sensitivity in helium semiquant mode is essentially the same as in full quant mode when similar integration times are used. While method detection limits were not specifically calculated in semiquant mode in this work, they can be estimated from the measured semiquant response factors and the response in low-level samples. For example, Figure 1 shows the response for Cd isotopes in one of the samples. The calibrated semiquant response factor for 111 Cd was 9,960 cps/ppb and the response for 111 Cd in this sample was 41 cps (with excellent isotopic fit), resulting in an estimated concentration of 4.1 ppt (pg/g) in solution. Based on this, a very conservative 100 ppt (50 ppb in original sample) was used as the limit of detection for all elements measured in semiquant mode. Semiquantitative acquisition parameters are listed in Table 2. Semiquantitative Measurements in He Mode For the most accurate semiquantitative determination, it is essential that all elements are acquired under the same conditions, so the consistent relationship between the responses of adjoining elements is maintained. The unique ability of the Agilent 7500 ORS to eliminate polyatomic interferences using carefully controlled kinetic energy discrimination (KED) in helium collision mode allows accurate semiquantification of a range of elements in the unknown samples. Helium is a non-reactive gas and thus no new interferences are formed in the cell, there are no analyte losses by reaction, and the interferences formed in the plasma are eliminated through a physical molecular filtering process [2]. Polyatomic interferences are eliminated in helium collision mode simply due to their size [6]. Regardless of the sample matrix, only a single set of ORS parameters is used. This approach is simple, fast, accurate, and interference free for all analytes (note that interference-free isotopes of elements suffering from isobaric interferences must be used). The instrument was therefore tuned for typical robust plasma conditions as indicated in Table 1. A single calibration standard containing 100 ppb of Al, Ce, Cd, Co, Li, Tl, U, and Y made up in 10 percent HNO 3 (w/w) was used to update the semiquantitative response factors for a set of elements across the mass range. This means that noncalibrated elements are updated by interpolating between calibrated isotopes. The ICP-MS ChemStation software does this automatically. Though any number of elements can be used for the semiquantitative calibration, increasing the number of elements reduces the mass range over which the factors are interpolated between measured reference elements and thus improves semiquantitative accuracy. Internal standardization was applied using a typical suite of internal standard elements across the mass range. 3 Figure 1. Cd isotopes in an antiviral drug measured in helium semiquant mode at approximately 5 ppt. Overlay shows expected isotope ratios normalized to m/z 111. Table 2. Semiquantitative ICP-MS Acquisition Parameters. Parameter Total acquisition time Acquisition mode Value 150 seconds Spectrum- peak hopping Number of masses m/z 250 Integration time 0.6 seconds Number of points per mass 6 Repetition 1 Uptake time Stabilization time Post-acquisition rinse 50 seconds 20 seconds 30 seconds

Fully Quantitative Measurements for Comparison Purposes Fully quantitative analyses in standard (no gas), helium, and hydrogen modes were performed for comparison purposes. Multielemental standards were prepared and a five-point calibration curve was obtained. Calibration standards containing the following elements: Al, As, Au, Ba, Be, Bi, Ca, Cd, Cr, Co, Cu, Fe, Ge, Li, Mg, Mo, Nd, Ni, Pb, Pd, Pt, Rh, Ru, Se, Si, Sb, Sn, Sr, Te, Ti, Tl, V, Zn, and Zr from 1 ppb to 2,000 ppb were prepared in HNO 3 1 percent (w/w). Results and discussion Six tablets of the antiviral drug from three different production batches were analyzed. Results are shown for each batch: B1, B2, and B3. Figure 2 shows the trace metal profile obtained from the semiquantitative analysis in He mode and that obtained from the full quantitative analysis results. In each case, the two profiles show very good agreement. Figure 1 shows the calibration curves for each analyte in the clean seawater matrix. Corresponding internal DLs (3σ of the calibration blank) and external MDLs (from the 40 blank seawater measurements throughout the run) are shown in Table 2. It is important to highlight that the suite of elements analyzed in the drugs includes Co, Cr, Cu, Fe, Ni, V, and Zn, all elements with isotopes that can suffer from polyatomic overlaps in complex matrices. Figure 3 is the full-scan mass spectrum obtained from a single semiquant run, including an expanded inset showing the region containing Ni, Cu, and Zn. The excellent fit with expected isotope ratios for all three elements confirms that no polyatomic interferences are present. Those elements that were found to be in the pharmaceutical tablets below the limit of detection are also included in the plots. Since this is important information for the screening of impurities of pharmaceutical products, it is also important to know which elements are not present above LOD levels. Results for the internal QC certified for the elements being analyzed in the tablets were within ± 30% of the certified concentration, from as low as 2 ppb for Cd to over 11 ppm for Ca in the aqueous solutions Figure 2. Trace metal profiling results for the pharmaceutical tablets from batches 1 to 3. Semiquantitative results are plotted with fully quantitative results for comparison. Fe concentrations have been divided by 1,000 and results are expressed in ng/g. [1] Spectrum No.1 [ 146.739 sec]:sq.d# / Tune #1 [Count] [Linear] Figure 3. Full scan semiquant mass spectrum of antiviral drug representing entire elemental composition of sample. Inset shows expanded region between m/z 60 to 70, including all isotopes for Ni, Cu, and Zn with an overlay of theoretical isotope ratios. 4

Conclusions The control of inorganic impurities in pharmaceutical tablets is of high importance in the pharmaceutical manufactory chain and therefore rapid, elemental screening tools are needed. In this work the potential of semiquantitative elemental screening by inductively coupled plasma mass spectrometry (ICP-MS) with collision/reaction cell technology (CRC) using ------- He gas has been shown. Elements which normally suffer from polyatomic interferences have been included in the semiquantitative analysis and trace metal profiles obtained for the pharmaceutical tablets from three different batches agreed with those obtained by fully quantitative analysis. Semiquantitative screening in He mode has proved to effectively remove polyatomic interferences, providing rapid screening of real pharmaceutical samples. References 1. J. Roy, Pharmaceutical Impurities A Mini Review, AAPS PharmSciTech, 3(2) article 6 (http://www.aapspharmscitech. org) 2. G. Woods, E. McCurdy, and S. Wilbur, Interference-Free Semiquantitative Analysis Using the Agilent 7500ce ICP-MS, Agilent Technologies publication 5989-0741EN 3. M. Oishi and K. Fukuda, Analysis of Electro-ceramics by Laser Ablation ICP-MS, Agilent Technologies publication 5989-0321EN 4. S. Wilbur, Faster, Simpler, More Accurate Semiquantitative Analysis Using the Agilent 7500cx ICP-MS, Agilent Technologies publication 5989-6662EN 5. E. McCurdy and G. Woods, J. Anal. At. Spectrom., 2004, 19, 607 615. 6. Noriyuki Yamada, Junichi Takahashi and Ken ichi Sakata, J. Anal. At. Spectrom., 2002, 17, 1213 1222 www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc. 2017 Published March 12, 2017 Publication number: 5989-9443EN