MODELING, DESIGN AND EXPERIMENTAL CARACHTERIZATION OF MICRO-ELECTRO ELECTRO-MECHANICAL- SYSTEMS FOR GAS- CHROMATOGRAPHIC APPLICATIONS

Similar documents
Développement de micro-préconcentrateurs pour l'analyse de traces de gaz et explosifs.

2D Simulations and Electro-Thermal Analysis of Micro-Heater Designs Using COMSOL TM for Gas Sensor Applications

Time-of-Flight Flow Microsensor using Free-Standing Microfilaments

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC


Sensors and Actuators B: Chemical

Microfabricated Passive Vapor Preconcentrator/Injector Designed for Microscale Gas Chromatography

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

A Micro-GC Based Chemical Analysis System

Resistance Thermometry based Picowatt-Resolution Heat-Flow Calorimeter

Odor Detection and Analysis using GC/SAW znose

Outline. Chemical Microsystems Applications. Microfluidic Component Examples Chemical Microsystems for Analysis Chemical Microsystems for Synthesis

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

Chapter 31 Gas Chromatography. Carrier Gas System

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

CHEM 429 / 529 Chemical Separation Techniques

Chromatography. Gas Chromatography

Gas Chromatography (GC)

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Chapter 3 Engineering Science for Microsystems Design and Fabrication

GC Instruments. GC Instruments - Sample Introduction

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Micro-sensors based on thermal transduction for steady and unsteady flow measurements

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

An Investigation on NEG Thick Film for Vacuum packaging of MEMS

Rapid Catalyst Screening Reactors

Proposal of A New Structure Thermal Vacuum Sensor with Diode-Thermistors Combined with a Micro-Air-Bridge Heater

Principles of Gas- Chromatography (GC)

Introduction to Gas Chromatography

Chip Integrated Micro Cooling System For High Heat Flux Electronic Cooling Applications

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Real Time On-Site Odor and VOC Emission Measurements Using a znose

Understanding Gas Chromatography

Chemistry Gas Chromatography: Separation of Volatile Organics

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

Introduction to Capillary GC

Analytical Chemistry

EE C245 ME C218 Introduction to MEMS Design Fall 2007

2401 Gas (liquid) Chromatography

GAS CHROMATOGRAPHY (GC)

Chip-Scale Mass Spectrometers for Portable Gas Analyzers Luis Fernando Velásquez-García. A. I. Akinwande, K. Cheung, and L.-Y Chen.

EE 5344 Introduction to MEMS CHAPTER 7 Biochemical Sensors. Biochemical Microsensors

Micro Cooling of SQUID Sensor

systems Dibyadeep Paul

Introduction to Chromatographic Separations

Technical know-how in thermal analysis measurement

Theory and Instrumentation of GC. Chromatographic Parameters

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

High Pressure/Performance Liquid Chromatography (HPLC)

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Lecture 18: Microfluidic MEMS, Applications

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Prototype of the gas chromatograph mass spectrometer to investigate volatile species in the lunar soil for the Luna-Resurs mission.

CHAPTER 6 GAS CHROMATOGRAPHY

Open Column Chromatography, GC, TLC, and HPLC

Liquid Chromatography

10/27/10. Chapter 27. Injector typically 50 C hotter than oven

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

Gas Chromatography. Introduction

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

G a s C h r o m a t o g r a p h y Determination of Ethanol in Wine by Head-Space Gas Chromatography

Foundations of MEMS. Chang Liu. McCormick School of Engineering and Applied Science Northwestern University. International Edition Contributions by

Zebra GC: A Fully Integrated Micro Gas Chromatography System

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

RESEARCH ON BENZENE VAPOR DETECTION USING POROUS SILICON

MPC-D403 MPC-D404. Ultra-small Peltier Coolers. High impedance Low control current High power efficiency

Luminescence transitions. Fluorescence spectroscopy

Chemistry Instrumental Analysis Lecture 31. Chem 4631

MM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE

Too many answers to list.

Lab 3 Guide: Gas Chromatography (GC) (Sept 8-14)

LMM-H04 Mass Air Flow Sensor

Introduction to Capillary GC

Integrated measuring system for MEMS

Slide 1. Temperatures Light (Optoelectronics) Magnetic Fields Strain Pressure Displacement and Rotation Acceleration Electronic Sensors

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Design And Analysis of Microcantilevers With Various Shapes Using COMSOL Multiphysics Software

Self-study problems and questions Processing and Device Technology, FFF110/FYSD13

TEPZZ A_T EP A1 (19) (11) EP A1 (12) EUROPEAN PATENT APPLICATION. (51) Int Cl.:

Headspace Technology for GC and GC/MS: Features, benefits & applications

The Effects of Carrier Gas Viscosity and Diffusion on Column Efficiency in Capillary Gas Chromatography

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Gas Chromatography. A schematic diagram of a gas chromatograph

Fabrication Technology, Part I

Design and Fabrication of a Micro-size Thermionic Ionization/Flame Ionization Detector for Gas Phase Chemical Analytes

Instrumental Chemical Analysis

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

Control of the fabrication process for the sensors of the CMS Silicon Strip Tracker. Anna Macchiolo. CMS Collaboration

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Making Improvements to Today s Natural Gas Analysis Systems

Transcription:

MODELING, DESIGN AND EXPERIMENTAL CARACHTERIZATION OF MICRO-ELECTRO ELECTRO-MECHANICAL- SYSTEMS FOR GAS- CHROMATOGRAPHIC APPLICATIONS ENRICO COZZANI DEIS DOCTORATE CYCLE XXIII 18/01/2011 Enrico Cozzani 1

SCENARIO GAS-CHROMATOGRAPHY CHROMATOGRAPHY: analytical analytical technique based on separative principle, applied to mixtures of pollutants at unknown concentrations. IN PARTICULAR: The principle of chromatographic separation is based on the specific affinity between the sample mixture and a stationary phase housed in a separation column. Carrier gas (mobile phase) Sample mixture A + B AB Stationary phase Separation column Signal 18/01/2011 Enrico Cozzani 2 t A Time t B t A, t B = retention time

BLOCKS OF A GAS- CHROMATOGRAPHIC MODULE INJECTION SYSTEM SEPARATION COLUMN DETECTOR CHROMATOGRAM sample inlet loop injection system sampling pump detector carrier gas GC separation column Signal Time 18/01/2011 Enrico Cozzani 3

RESEARCH @ IMM-CNR BOLOGNA Since several years, the Institute for Microelectronics and Microsystems (IMM) - Section of Bologna is involved in an intense research activity focused on air quality monitoring. In particular, a great concern is referred to the analysis of volatile organic compounds (eg. BENZENE) because of their strong toxicity even at extremely low concentrations (as a matter of fact, risk threshold is in the order of parts per billion (ppb)). In this scenario, a thorough knowledge of MEMS (Micro-Electro-Mechanical- Systems) fabrication techniques provides a great contribution for the design of innovative microsystems employed in a gas-chromatographic module, charachterized by reduced dimensions (and costs) and improved performances with respect to the actually disposable products TARGET: fabrication of the three blocks (injection system, separation column, detector) by exploiting silicon micro-machining techniques. 18/01/2011 Enrico Cozzani 4

ELECTRO-THERMAL THERMAL SIMULATIONS The starting point of my doctorate activity was the development and the validation of a Finite Element Model (FEM) aimed at simulating the coupled electro-thermal behavior of the microdevices, in order to support and optimize their design. The validation of the model was carried out by comparing the simulation results with the experimental characterization of a particular detector developed since 2005 at IMM Section of Bologna, i.e. a Metal OXide (MOX) sensor, whose detection principle is based on the fact that when a thin film of metal oxide (specifically SnO 2 ), wired between two electrodes, interacts with a pollutant at unknown concentration, in proximity of 400 C mechanisms of oxidation-reduction take place and cause changes in the chemical and physical properties of the film (specifically, ELECTRICAL RESISTIVITY). The analysis of the pollutant can be easily performed by monitoring the variations of the electrical current flowing through the electrodes. 18/01/2011 Enrico Cozzani 5

MOX SENSOR DESIGN By exploiting silicon micro-machining and photolitography definition techniques, chips containing arrays of four sensors are fabricated. Each sensor is formed by: A suspended thermallyinsulated dielectric membrane (SiO 2 Si 3 N 4 SiO 2 th = 1µm) A horseshoe-shaped platinum heater (th = 200nm) Two concentric platinum electrodes contacting the sensitive film (th = 200nm) 400 µm 80 µm 18/01/2011 Enrico Cozzani 6

FINITE ELEMENT MODEL The use of FEMLAB software enabled the coupling of the electrical behavior described by Poisson s equation (σ V) = 0 (1) with the thermal behavior described by Fourier s equation Q = - (k ht ct T)+ (T - Ta) + (T da da 4 Ta 4 ) (2) by means of Joule s equation upon power dissipation in the form of heat Q = ρ(t) ІJІ 2 = 1 σ(t) (σ(t) ІEІ)2 = σ(t) ІEІ 2 = σ(t) І VІ 2 (3) Putting eq. (3) in eq. (2), the relationship representing the fully coupled electro-thermal model can be easily disclosed: 2 ht T - ct (T - Ta) - (T k da 4 - Ta 4 ) + k da σ k І VІ 2 = 0 18/01/2011 Enrico Cozzani 7

Electrical, thermal and optical properties of all the thin film materials were experimentally measured by using appropriate test structures and the obtained values were employed in the model ( no fitting parameters). Simulations were carried out in two subsequent steps: Initially, sensor behavior in vacuum was considered (i.e. neglecting the contribution of air as heat conducting material) 2D simulation Err < 1% 18/01/2011 Enrico Cozzani 8

Following, the procedure was repeated considering the influence of air upon heat conduction 3D simulation 450 400 Err < 5% The increase in relative error can be explained with the uncertainty in the evaluation of thermal conductivity of air (the formula of k(t) used in the model was taken from literature) and with the fact that its dependence from relative humidity was neglected. 18/01/2011 Enrico Cozzani 9

PRE-CONCENTRATION UNIT A limitation of valve-based devices is the small quantity of injected sample: preconcentration devices have been designed to collect volatile compounds from large-volume vapor sample and inject them into the gas chromatographic column as narrow vapor plugs. Our unit is a micro-fabricated cartridge formed by channels filled with cavitands, able to trap the volatile compounds, and to subsequently purge them by a rapid thermal desorption controlled through an integrated platinum resistor. FABRICATION PROCESS Silicon Wafer DRIE etch Oxide deposition Pt deposition Bonding pyrex PYREX (300µm) OXIDE (100nm) SILICON (500µm, 300µm etched) PLATINUM (240nm) 18/01/2011 Enrico Cozzani 10

CHIP WIDTH = 1.4cm CHIP LENGTH = 2.5cm CHANNEL WIDTH = 1mm CHANNEL SEP = 0.2mm Rheater @ Tamb = 12Ω RTsensor @ Tamb = 100Ω TEMPERATURE ( C) 204.3 204 ELECTRO-THERMAL 3D SIMULATION 600 mw @ 200 C 203.5 203 202.5 202 201.5 18/01/2011 Enrico Cozzani 11 201

SEPARATION COLUMN The doctorate activity proceeded focusing the attention on the design and fabrication of micro-machined silicon packed columns that specifically execute chromatographic separation 1st PROTOTYPE: 50 cm long double-spiral shaped column (1 inlet & 1 outlet), rectangular section, fabricated on a 2.5 x 2.5 cm2 chip: Sep CROSS-SECTION W Step = W + Sep W = channel width = 800 µm Sep = 100 µm Step 18/01/2011 Enrico Cozzani 12

ELECTRO-THERMAL THERMAL 3D SIMULATION TEMPERATURE ( C) 203.3 12Ω @RT 850 mw @ 200 C 203 202.5 202 201.5 201 200.5 200 18/01/2011 Enrico Cozzani 13

2 nd PROTOTYPE: 1m long double-spiral shaped column (1 inlet & 1 outlet), rectangular section, fabricated on a 2.5 x 4.5 cm 2 chip: 207.5 207 206.5 12Ω @ RT 1.27 W @ 200 C 206 205.5 205 204.5 18/01/2011 Enrico Cozzani 14 204 203.5 203

PICTURES METAL (chip front) HOLES (chip rear) STATIONARY PHASE: Carbograph 2 + 0.2% Carbowax 18/01/2011 Enrico Cozzani 15

EXPERIMENTAL RESULTS Chromatogram of a VOC s mixture (BENZENE, TOLUENE, ETHYL- BENZENE, M-XYLENE @ equal concentrations). Time of analysis: 10 minutes 50 cm column 18/01/2011 Enrico Cozzani 16

FAST GC: CAPILLARY COLUMNS Miniaturized, portable fast-gc systems based on very low inner diameter ( 100µm) capillary columns are emerging, allowing for vastly improved separations with higher resolution. Other merits include complete chromatogram acquisitions in few minutes, smaller sample size requirements and often higher sensitivities Chip = 2 x 2 cm 2 Column length = 2m W Sep Step = W + Sep (W = 20 µm, Sep = 120 µm) Step Circular section Diameter = 100µm ISOTROPIC ETCH 18/01/2011 Enrico Cozzani 17

EXPERIMENTAL RESULTS VOC s MIXTURE INJECTED IN A 2m-LONG CAPILLARY COLUMN: 1 Benzene; 2 Toluene; 3 ethyl-benzene; 4 & 5 Xilene-isomers 1 2 4 Time of analysis: 2 minutes 3 5 18/01/2011 Enrico Cozzani 18

FAST GC: DETECTOR Analysis cycle should be around [1 5] minutes and typical column flow rates for fast-gc stand inside the [0.1 0.5] sccm range. For these reasons, fast-gc requires detectors with extremely low volumes (absence or reduction of dead volumes). The adopted solution consisted in the fabrication of a integrated pyrex microchamber in order to encapsulate the detector in a ultra-low volume ( 0.1µL). Preliminary simulations were performed on MOX sensors, implementing the electro-thermal model with a microfluidic part governed by the Navier-Stokes equations: u x i i ρ u = 0 j ui xj + p xi = µ 2 ui xi x j 18/01/2011 Enrico Cozzani 19

S I M U L A T E D E X P E R I M E N T 18/01/2011 Enrico Cozzani 20

THERMAL CONDUCTIVITY DETECTOR (TCD) Although chemical MOX sensors showed an excellent sensitivity towards VOC s, they can not be considered as general purpose detectors. TCD, whose operation principle is purely physical, is still today the most used detector in portable GC systems because of its universality. If R1 R4 = R2 R3 V = 0 (NO pollutant) R1 R2 If R1 R4 R2 R3 V 0 (YES pollutant) R3 R4 Reference gas (carrier) flow comes from an empty capillary column, while analytical (carrier + pollutant) flow comes from a capillary column filled with stationary phase. 2 columns needed 18/01/2011 Enrico Cozzani 21

FABRICATION PROCESS Si Wafer (500µm) membrane deposition RIE etch Pt deposition TMAH etch pyrex bonding Reference channel 650 µm 170µm Analytical channel 18/01/2011 Enrico Cozzani 22

Chip: 5 x 5 mm 2 405 TEMPERATURE ( C) Diss Pow = 40mW for AIR carrier Diss Pow = 200mW for HELIUM carrier 400 350 300 250 200 150 100 50 25 E X P E R I M E N T 18/01/2011 Enrico Cozzani 23

PRELIMENARY RESULT ON TCD injection R1 R2 R3 R4 2% toluene in He TCD connected to a commercial GC-system: not its own control electronics for biasing and reading + extra dead volumes introduced 43mV 18/01/2011 Enrico Cozzani 24

FUTURE ACTIVITIES Improve performances of the capillary columns (increase length or decrease inner diameter). Characterize TCDs with their own electronics. Exploiting the same fabrication process, design a mixed array of sensors, to combine the high sensitivity of MOX and the universality of TCD. Design and fabricate the first prototype of a silicon micromachined injector (ongoing..). Assembly the whole in order to fabricate a portable micro-gc with high performances and reduced dimensions (and costs). THANKS FOR YOUR ATTENTION 18/01/2011 Enrico Cozzani 25