Electrolytic Tiltmeter Temperature Coefficients: Source, Definition and Use to Improve Accuracy

Similar documents
Dam Monitoring With Tiltmeters

STRAIN GAUGES YEDITEPE UNIVERSITY DEPARTMENT OF MECHANICAL ENGINEERING

Model 791 Calibration Plate for Tiltmeters and Inclinometers

Module 2 Mechanics of Machining. Version 2 ME IIT, Kharagpur

Mechanical Motion Measurement System Design, Initial Results and Experiments with Orbit Feedback

Glossary Innovative Measurement Solutions

Budge Road Landslide Jackson Extensometer installation report April 2014 with data to 20 May

Instruction Manual. HI Refractometer for Sucrose in Wine and Grape Products Measurements

Freescale Semiconductor

Load Washers. Force. for Forces of 7, kn. F z. Type 9001A A 9081B, 9091B

DESIGN AND CONSTRUCTION OF A WATER TUNNEL. Stephen C. Ko

Temperature Measurement

Outline. 4 Mechanical Sensors Introduction General Mechanical properties Piezoresistivity Piezoresistive Sensors Capacitive sensors Applications

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

Introduction to Strain and Borehole Strainmeter Data

Smartec Pressure Sensor (bridge output)

Source Wave Design for Downhole Seismic Testing

Zeroth Law of Thermodynamics

Thermal Equilibrium. Zeroth Law of Thermodynamics 2/4/2019. Temperature

1. Mark the correct statement(s)

Physics 3 Summer 1989 Lab 7 - Elasticity

THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY

ENG1001 Engineering Design 1

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

CE 320 Structures Laboratory 1 Flexure Fall 2006

CE415L Applied Fluid Mechanics Laboratory

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity

ENSC387: Introduction to Electromechanical Sensors and Actuators LAB 3: USING STRAIN GAUGES TO FIND POISSON S RATIO AND YOUNG S MODULUS

Microdeformation: combining direct fracture height measurement with microseismic response Natalia Verkhovtseva, Greg Stanley

Model 3024 Albedometer. User s Manual 1165 NATIONAL DRIVE SACRAMENTO, CALIFORNIA WWW. ALLWEATHERINC. COM

CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT. Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements

SITRANS F flowmeters. SITRANS F US System information SITRANS F US Ultrasonic flowmeters 4/181

Because the third wire carries practically no current (due to the voltmeter's extremely high internal resistance), its resistance will not drop any

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

Piezo Theory: Chapter 1 - Physics & Design

User s Manual: MODEL 904-T CLINOMETER PAK

QAM-I-111 Operation and Calibration of the Conductivity Meter

Catalytic bead sensors are used primarily to detect

Copyright 2017 by Select Calibration Inc. All rights reserved. Temperature Compensation

Custody Transfer Measurement and Calibration Round Robin Testing for Natural Gas with Coriolis Meters

The Strain Gauge. James K Beard, Ph.D. Rowan Hall Auditorium November 2, 2006

Development of a Precision Tilt Sensor for a Cryogenic Torsion Balance Experiment

Human Arm. 1 Purpose. 2 Theory. 2.1 Equation of Motion for a Rotating Rigid Body

Structures - Experiment 3B Sophomore Design - Fall 2006

Author(s) Sawamura, Yasuo; Kishida, Kiyoshi;

*These are the vinyl for MAX Bepop Cutting/printing machines. We have 100mm (4in) and 200mm(8in) width models available in North America.

Strain Gages. Approximate Elastic Constants (from University Physics, Sears Zemansky, and Young, Reading, MA, Shear Modulus, (S) N/m 2

1. Introduction. 1.1 Overview of Study:

Strain Measurements. Isaac Choutapalli

The Development of a High-Temperature PRT Calibration Process Based On Dry- Block Calibrators

APPLICATIONS OF VIBRATION TRANSDUCERS

Microscopy Cryostat System

Earlier Lecture. In the earlier lecture, we have seen non metallic sensors like Silicon diode, Cernox and Ruthenium Oxide.

1.060 Engineering Mechanics II Spring Problem Set 8

Paroscientific, Inc. Digiquartz Technology Overview. Paroscientific, Inc.

MECHANICS OF STRUCTURES SCI 1105 COURSE MATERIAL UNIT - I

[5] Stress and Strain

Electricity. Measuring the force on current-carrying conductors in a homogeneous magnetic field. LEYBOLD Physics Leaflets P

DEPARTMENT OF MECHANICAL ENIGINEERING, UNIVERSITY OF ENGINEERING & TECHNOLOGY LAHORE (KSK CAMPUS).

Experiment Two (2) Torsional testing of Circular Shafts

Mechanics of Solids. Mechanics Of Solids. Suraj kr. Ray Department of Civil Engineering

Mechanical Engineering Ph.D. Preliminary Qualifying Examination Solid Mechanics February 25, 2002

Displacement and Settlement Monitoring in Large Geotechnical Structures with a Novel Approach to Distributed Brillouin Sensing

Temperature control for Varian Cary line of UV/Vis Spectrophotometers and the Eclipse Fluorometer

Elastic Properties of Solids (One or two weights)

Thermal movements of the Existing Gateway Bridge

Development of Laser Thickness Gauge in Steel Plate Shearing Line

Influence of residual stresses in the structural behavior of. tubular columns and arches. Nuno Rocha Cima Gomes

1 Written and composed by: Prof. Muhammad Ali Malik (M. Phil. Physics), Govt. Degree College, Naushera

Target Insertion Flowmeter

APPENDIX 1 DESCRIPTION OF HOT WIRE ANEMOMETER

Variable Capacitance Accelerometers: Design and Applications

Strength of Material. Shear Strain. Dr. Attaullah Shah

ph/o.r.p. probe Type 8619 multicell transmitter/controller General data Measuring range Bürkert ph probe Bürkert O.R.P. probe Medium temperature

473 Dew Point Hygrometer

CEE575 - Homework 1. Resistive Sensing: Due Monday, January 29

MET 301 EXPERIMENT # 2 APPLICATION OF BONDED STRAIN GAGES

D : SOLID MECHANICS. Q. 1 Q. 9 carry one mark each. Q.1 Find the force (in kn) in the member BH of the truss shown.

KINETIC EEFCT ON FLEXIBLE BODIES BEHAVIOR

Wheatstone Bridge Nonlinearity

INSTRUCTION MANUAL Milwaukee Refractometer MA886 Refractometer for Sodium Chloride Measurements

International Journal of Scientific & Engineering Research, Volume 5, Issue 1, January ISSN

An Essential Requirement in CV Based Industrial Appliances.

Model tests and FE-modelling of dynamic soil-structure interaction

EXPERIMENT 11 LINEAR EXPANSION OF A SOLID MATERIAL

EE C247B / ME C218 INTRODUCTION TO MEMS DESIGN SPRING 2014 C. Nguyen PROBLEM SET #4

Forces on piles preventing debris slope slips

INSTRUCTION MANUAL Milwaukee Refractometer MA888 Refractometer for Ethylene Glycol Measurements

473 Dew Point Hygrometer

SENSOR DEVICES MECHANICAL SENSORS

Strain, Force, and Pressure

Design and Development of Impact Load Sensor for Dynamic Testing Purposes

Stage 4 G-03: CONDUCTIVITY OF SOLUTIONS CP: USP. May 2015 INTRODUCTION

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano

Mechanics. Determining the gravitational constant with the gravitation torsion balance after Cavendish. LEYBOLD Physics Leaflets P

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

Instruction Manual HI Refractometer for Sodium Chloride Measurements

PIZZI FRANCO via Ripoli, 207/E Firenze - Italy. Instruments and Systems for Dam Monitoring and Structural Controls

Transcription:

Electrolytic Tiltmeter Temperature Coefficients: Source, Definition and Use to Improve Accuracy Introduction Environmental temperature changes alter the mechanical and electrical characteristics of all instrumentation. Metals expand and contract, and electrical properties such as resistance and capacitance rise and fall. These effects change instrument output and lessen the accuracy of the measured variable (pressure, flow, tilt, strain, etc.). This technical note describes the source of temperature dependence in one type of instrumentation, Jewell Instruments electrolytic tiltmeters, and explains how to remove this effect to maximize accuracy. The principles presented here also apply to many other instrument types. Just as instrumentation exhibits temperature-dependent behavior, so too do natural and engineered structures including slopes, embankments, and concrete and steel construction. Thermal expansion and contraction in response to daily and seasonal temperature fluctuations generate real movements that are detected by tiltmeters and other sensors. The magnitude of this effect, and ways of differentiating it from purely instrumental behavior, are discussed in this article. Sources of Temperature Coefficients The sensors in Jewell Instruments Geotech/Tech tiltmeters are known as electrolytic tilt sensors, a type of electronic spirit level comprised of a glass case and containing a conductive liquid (electrolyte), an air bubble and platinum electrodes. As the sensor tilts, the wetted area of each excitation electrode (Figure 1) increases or decreases, depending on the tilt direction. This change causes the electrical resistance between the central pickup electrode and each excitation electrode to rise or fall. It is these resistance changes that are sensed by the Figure 1: Electrolytic Tilt Sensor Movement of the bubble changes the output at the pick-up electrode when an AC voltage is applied across the excitation electrodes. tiltmeter electronics, which convert them to precise measurements of the magnitude and direction of tilt. Temperature fluctuations cause thermal expansion and contraction of the sensor liquid, shrinking or swelling the air bubble and changing the amount of liquid in contact with each excitation electrode. This process alters the scale factor (gain) of the sensor and can shift its zero point. Small changes in sensor output in the absence of any real tilt movement are the result. Experiments have shown that volumetric expansion and contraction of the liquid is the single biggest source of temperature coefficients in Jewell Instruments electrolytic tiltmeters. This effect is much greater than dimensional changes of the sensor s glass case, which has a thermal expansion coefficient 100 times smaller than that of the liquid.

Thermoelasticity of the tiltmeter housing, and of the mechanical connections between housing and sensor, is another source of tiltmeter zero shift. To minimize this effect, rigid housings are used and connections between the sensor and enclosure are made as few as possible. In many designs we pot the tilt sensor directly into the housing base, eliminating mechanical connections entirely and turning the sensor and base into one unified element. The temperature effects described above are partially removed (compensated) by the tiltmeter s electronic circuitry. The apparent tilt (residual error) remaining after such compensation is highly repeatable and is described by two linear temperature coefficients, the temperature coefficient of scale factor, K S, and the temperature coefficient of zero shift, K Z. These coefficients include contributions from all sources, including the tiltmeter electronics. There is one additional effect of temperature on electrolytic tilt sensors. The conductivity of the electrolyte changes more than five-fold over the typical operating range of a tiltmeter (typically -40 C to +70 C). By measuring sensor output ratiometrically (taking output as a percentage of input), Jewell Instruments electrolytic tiltmeters remove this effect entirely. However, in designs that incorporate the sensor as part of a Wheatstone bridge, electrolyte conductivity change can be a major source of measurement error. Temperature Coefficients Defined Scale factor is the proportionality constant between tilt angle and tiltmeter output. It is determined in the factory by calibrating the tiltmeter - rotating it through a range of known angles and recording the output voltage at each angle. The slope of the best-fit straight line through the calibration data is the scale factor S cal that is reported on the tiltmeter calibration certificate. In reality, the slope is slightly different at each temperature (Figure 2). The change of slope per unit temperature change is the temperature coefficient of scale factor: Equation 1: / where S cal is the scale factor at the calibration temperature T cal, and S is the scale factor at a different temperature T. Figure 2: Calibration lines at two different temperatures, T cal and T. Temperature change can also shift the zero crossing of the calibration line in the absence of any real tilt of the structure to which the tiltmeter is attached. In Figure 2 the zero offset voltage is V T which leads to an apparent tilt angle of θ at temperature T. The zero shift is therefore or. The zero shift per unit temperature change is defined as the temperature coefficient of zero shift, K Z : Equation 2: The coefficients K S and K Z are determined in the laboratory by performing calibrations at two or more temperatures and include contributions from all sources. Their values are specific to each of the several classes of tiltmeters made by Jewell Instruments and are available on request. For tiltmeters with the designation high gain, the ones most typically

used in geotechnical engineering, K S +0.0004/ C and K Z 1.5µradian/ C (= 0.2 arcsecond/ C). Temperature coefficient values should decline in the future as sensor and electronic designs advance. Equations for Temperature Compensation For uncompensated tiltmeter measurements, the tilt angle θ is simply: Equation 3: θ where V is the measured voltage and S cal is the scale factor measured at the calibration temperature T cal. For measurements at all temperatures, S cal, the scale factor is first adjusted using the temperature coefficient K S before computing θ. Equation 4:!! " The zero offset is then removed using the temperature coefficient K Z and the true compensated tilt angle, θ comp, is computed as follows: Equation 5: #$% where θ comp T T cal S S cal V = true angular position (tilt) = the temperature at which your measurement was made = the calibration temperature reported on the tiltmeter calibration certificate = scale factor at temperature T = the scale factor reported on the tiltmeter calibration certificate = the measured output voltage at temperature T Inserting Equation 4 into Equation 5, and expanding results in the following expression: Equation 6: #$% " Recognizing that S cal V is the uncompensated angle θ, we can rewrite Equation 6 as: Equation 7: #$% Equation 7 intuitively shows the compensation of the angular measurement consists of the uncompensated angular measurement, a slope compensation term that scales the uncompensated angular measurement and a zero offset term that is dependent on temperature only. Finally, this temperature compensation procedure is an option for all Jewell Instruments electrolytic digital tiltmeters. It may also be incorporated into spreadsheets and other user-written programs.

Example 1 Figures 3 and 4 show 10 days of data for two high-gain tiltmeters with a resolution of 1 microradian (Tiltmeter 1 is blue and Tiltmeter 2 is yellow). The upper graphs in each figure plots daily temperature oscillations at the ground surface in degrees Celsius measured by temperature sensors inside each tiltmeter. The lower graph shows tilt of the ground surface in microradians during the same period. The large change in tilt on the fourth day resulted from ground subsidence caused by a nearby pump test. The graph of tilt in Figure 3 contains daily oscillations that directly correlate with temperature, while Figure 4 shows temperature compensated results. Figure 4: Ten days of ground tilts during a pump test, without temperature compensation Figure 3: Same data as in Figure 3, with temperature compensation.

Thermoelasticity in Geotechnical Engineering Thermoelasticity is the elastic expansion and contraction of materials in response to changing temperature. Soil, steel and concrete structures each have their own temperature coefficient, the coefficient of thermal expansion α, which is expressed in units of strain (microinches per inch or microns per meter) per unit change in temperature. Thermoelasticity is a major source of structural movement, and precision tiltmeters easily measure this behavior. Thermoelastic deformation typically produces tilts that exceed the temperature-induced output changes of properly designed tiltmeters. The following example illustrates the effect of large thermoelastic movements. Example 2 Tiltmeters are commonly installed on bridge piers and columns to detect early signs of settlement and riverbed scour. Figure 5a shows a bridge with one span. Let us assume that the span is fixed at one end but can expand laterally at the other. Now if the slip bearings are seized at the movable end, thermal expansion of the span by an amount ΔL will result in a tilt of the right pier (Figure 5b) of '() *,/-. If the temperature change is 10 C, α is 10-5 / C and span length L is 30 meters, then: ΔL = (10 C)( 10-5 / C)(30,000 mm) = 3 mm. For a pier that is 3 meters high, the tilt will be: θ = 1000 microradians = 206 arc seconds. Now compare this 1000 microradian movement with the uncorrected temperature-induced error of a Jewell Instruments electrolytic tiltmeter. Our high-gain tiltmeters, typically used in geotechnical and structural monitoring, have temperature coefficients of K S 0.0004/ C and K Z 1.5 microradians/ C. A 10 C temperature change therefore produces a zero shift of 15 microradians, 1.5% of the actual pier movement. The error induced by the coefficient K S is proportional to the rotation angle of the tiltmeter and the temperature change, and is even smaller. If the tiltmeter was leveled (nulled) during installation, its angle after column rotation would be 1000 microradians and the K S error would be (0.0004/ C)(10 C)(1000 microradians) = 4 microradians. In this example the tiltmeter measures thermoelastic tilt of the pier to better than 2% accuracy with no temperature compensation. Compensating the readings for temperature change yields improved results. Although this is a hypothetical example, it is typical of real field projects involving Jewell Instruments electrolytic tiltmeters. Most of the correlation of tilt with temperature results from thermoelastic deformation. If your data still correlate with temperature change after compensating for temperature, you are observing real structural or ground movement. Figure 5a: Initial bridge geometry. Figure 5b: Bridge geometry after temperature increase.

Figure 6 presents a real-life example of thermoelastic deformation of a thin-arch concrete dam. The high-gain Jewell Instruments electrolytic tiltmeter is installed in a gallery inside the dam, where temperatures do not cycle on a daily basis because of the insulating effect of the thick concrete. The plot shows gallery temperatures and raw (uncompensated) upstream-downstream tilting of the dam during a two-week period in early October 1993. Although temperatures do not vary, the real tilt angle fluctuates by 15 microradians daily as the result of daily heating and cooling of the downstream face of the dam a few meters away. Figure 6: Thermoelastic dam tilt caused by heating and cooling of the downstream face.

How to Minimize Temperature-Induced Measurement Errors without Temperature Compensation There are several ways to minimize temperature-induced measurement errors that do not involve any data processing at all. In many cases these methods eliminate the need for the temperature compensation procedures outlined above. 1. Reduce Temperature Extremes. When possible, instruments should be installed underground or in shaded locations where temperature extremes are minimized. If temperatures do not vary, they can have no effect on your measurements. If your instruments must be installed in locations exposed to direct sunlight, set up a hood that keeps them shaded while maintaining good ventilation. 2. Choose Light Colors. When other specifications are equal, light-colored instruments stay cooler and are preferable to dark-colored ones. 3. Establish Your Accuracy Requirements. Before selecting the tiltmeters for your project, decide on the accuracy that is required and estimate the temperature range that the instruments will experience. Then get temperature coefficients for the tiltmeters under consideration from their manufacturers. Use the temperature range and coefficients to compute potential errors, following the procedure in the previous section. If these errors are smaller than your accuracy requirements, no temperature compensation is necessary. 4. Use a Mechanically Stable Tiltmeter Design. Choose a tiltmeter design that minimizes thermoelastic deformation of the instrument itself. Compact, stiff housings are more stable and less likely to bend or vibrate than elongated beam designs with fixed ends. Also, the fewer the mechanical linkages between internal sensor and outer enclosure, the better. 5. Use a Mechanically Stable Mounting Method. Use a mounting method that maximizes thermoelastic stability. Three-point mounting is best because it is the most rigid and prevents bending and torsion that can occur with 2-point mountings. Mounting studs (typically threaded rods) that attach the tiltmeter to the structure should be as short as possible, of the same length and of the same material. In special cases thermally stable, but more expensive, invar studs can be used.

Conclusions 1. All instruments exhibit some degree of temperature-dependent behavior. Thermal expansion and contraction of the sensor liquid is the largest source of temperature dependency in Jewell Instruments electrolytic tiltmeters. 2. The effect of temperature change on tiltmeter output is predictable and repeatable. It is quantified by two constants, the temperature coefficient of scale factor, K S, and the temperature coefficient of zero shift, K Z. These constants enable the user to predict the magnitude of potential temperature-induced errors and to correct (compensate) for such errors during data analysis. 3. The large thermoelastic movements of civil engineering structures are easily detected by tiltmeters and are sometimes mistaken for measurement errors. 4. Before beginning an instrumentation project, the user should first establish the required measurement accuracy then estimate the measurement error over the expected temperature range using the instrument s temperature coefficients. If the error is smaller than the accuracy requirements, then no temperature compensation is necessary. 5. Simple precautions such as installing tiltmeters in the shade or underground can reduce or eliminate temperature effects. 6. Temperature compensation of Jewell Instruments electrolytic tiltmeter readings is performed using Equation 6 or Equation 5 if the uncompensated angle has already been calculated using Equation 3. Compensation is available as an option on all Jewell Instruments electrolytic digital tiltmeters and may also be built into spreadsheets and user-defined programs. Angle Conversion Factors 1 degree = 60 arc minutes = 3600 arc seconds = 17453 microradians = 0.01745 radians 1 arc second = 4.85 microradians 1 microradian = 1 microinch per inch = 1 micron per meter = 1 mm per km If Canada had a hinge at Winnipeg and a man standing in Vancouver lifted the west coast to chest height, he would tilt the western half of the country by 1 microradian.