Dynamics of laser induced thermoelastic expansion of native and coagulated ex-vivo soft tissue samples and their optical and thermomechanical

Similar documents
A FEM STUDY ON THE INFLUENCE OF THE GEOMETRIC CHARACTERISTICS OF METALLIC FILMS IRRADIATED BY NANOSECOND LASER PULSES

Bayesian approach to image reconstruction in photoacoustic tomography

Photoacoustic radio-frequency spectroscopy (PA-RFS): A technique for monitoring absorber size and concentration

For more information, please contact

Nanoacoustics II Lecture #2 More on generation and pick-up of phonons

Quantitative in vivo measurements of blood oxygen saturation using multiwavelength photoacoustic imaging

Pulsed photoacoustic Doppler flowmetry using a cross-correlation method

Time Scale of the Quaternary Structural Change in Hemoglobin Revealed by Transient Grating Technique

BETTER DESIGN AND NEW TECHNOLOGIES IMPROVE LASER POWER MEASUREMENT INSTRUMENTATION

Excitation-Wavelength Dependent and Time-Resolved Photoluminescence Studies of Europium Doped GaN Grown by Interrupted Growth Epitaxy (IGE)

Numerical Modeling of Thermal Effects during Irreversible Electroporation Treatments

New Concept of DPSSL

ULTRAFAST LASER PULSE TRAIN RADIATION TRANSFER IN A SCATTERING-ABSORBING 3D MEDIUM WITH AN INHOMOGENEITY

Noncontact measurement technique for wide range of viscosity of µl-order liquid sample

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Laser-Induced Incandescence of Rough Carbon Surfaces

UNIVERSITY OF CINCINNATI

Real-time optoacoustic monitoring of temperature in tissues

Finite Element Simulation of the Liquid Silicon Oriented Crystallization in a Graphite Mold

Advanced holographic nondestructive testing system for residual stress analysis

Digital Holographic Measurement of Nanometric Optical Excitation on Soft Matter by Optical Pressure and Photothermal Interactions

1 Introduction. IWBBIO Proceedings Granada, March,

A faster, more accurate way of characterizing cube beamsplitters using the Agilent Cary 7000 Universal Measurement Spectrophotometer (UMS)

Microwave-induced thermoacoustic tomography using multi-sector scanning

30 MM CAGE CUBE-MOUNTED CIRCULAR POLARIZERS

EXPERIMENTAL STUDY OF THE OUT-OF-PLANE DISPLACEMENT FIELDS FOR DIFFERENT CRACK PROPAGATION VELOVITIES

Numerical Simulation of the Bubble Cloud Dynamics in an Ultrasound Field

Measurement of local elastic modulus with CLUE

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a).

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

On the beam deflection method applied to ultrasound absorption measurements

Behavior and Energy States of Photogenerated Charge Carriers

Opto-Mechanical I/F for ANSYS

A numerical model for heat and pressure propagation for temperature controlled retinal photocoagulation

Towards quantitative tissue absorption imaging by combining photoacoustics and acousto optics

Two-Dimensional simulation of thermal blooming effects in ring pattern laser beam propagating into absorbing CO2 gas

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006

Quality Factor Thickness (nm) Quality Factor Thickness (nm) Quality Factor 10

Observing the Doppler Absorption of Rubidium Using a Tunable Laser Diode System

Integrating MD Nastran with Optical Performance Analysis

Ultrafast X-Ray-Matter Interaction and Damage of Inorganic Solids October 10, 2008

Computer Modelling and Numerical Simulation of the Solid State Diode Pumped Nd 3+ :YAG Laser with Intracavity Saturable Absorber

Ultrasonic Measurement of Minute Displacement of Object Cyclically Actuated by Acoustic Radiation Force

Determining thermal noise limiting properties of thin films

Investigation of the Energy Shielding of Kidney Stones by Cavitation Bubble Clouds during Burst Wave Lithotripsy

Multiple-source optical diffusion approximation for a multilayer scattering medium

Density Field Measurement by Digital Laser Speckle Photography

Analysis, simulation, and experimental studies of YAG and CO 2 laserproduced plasma for EUV lithography sources

Transient Newton rings in dielectrics upon fs laser ablation

Measurement of the optical absorption coefficient of a liquid by use of a time-resolved photoacoustic technique

Absorption spectra variations of EBT radiochromic film from radiation exposure

Chapter 1 INTRODUCTION AND BASIC CONCEPTS

How to measure packaging-induced strain in high-brightness diode lasers?

Lasers & Holography. Ulrich Heintz Brown University. 4/5/2016 Ulrich Heintz - PHYS 1560 Lecture 10 1

D. BARD DIVISION OF ENGINEERING ACOUSTICS, LUND UNIVERSITY

Optical Coherence Tomography detection of shear wave propagation. in MCF7 cell modules

Supporting Information for:

Thermal Imaging at Multiple Time Instants for Study of Self-Heating and ESD Phenomena

gives rise to multitude of four-wave-mixing phenomena which are of great

Measurements of the optical mirror coating properties.

phase retardance THz intensity ratio THz filling factor in air : 0.2 filling factor in si : 0.8 length of air : 4um length of si : 16um depth : 27.

arxiv: v1 [physics.med-ph] 21 Jan 2018

Degree of polarization in laser speckles from turbid media: Implications in tissue optics

Laser heating of noble gas droplet sprays: EUV source efficiency considerations

PHYS 160 Astronomy Test #2 Fall 2017 Version A

Publication I Institute of Physics Publishing (IOPP) Reprinted by permission of Institute of Physics Publishing.

ABLATION ON POLYMETHYLMETHACRYLATE BY ARGON FLOURIDE EXCIMER LASER ABSTRACT

Transducer design simulation using finite element method

Growth and collapse of laser-induced bubbles in glycerol water mixtures

Simulation of Laser-Material Interactions for Dynamic Transmission Electron Microscopy Experiments

DEVELOPMENT OF MEASURING SYSTEM FOR STRESS BY MEANS OF IMAGE PLATE FOR LABORATORY X-RAY EXPERIMENT

Multimodal multiplex Raman spectroscopy optimized for in vivo chemometrics

Output intensity measurement on a diagnostic ultrasound machine using a calibrated thermoacoustic sensor

Optical Interface for MSC.Nastran

Gain dependence of measured spectra in coherent Brillouin optical time-domain analysis sensors

NUMERICAL ANALYSIS OF OPTIMAL FOURTH-HARMONIC- GENERATION CRYSTAL LENGTH FOR WAVELENGTH CONVERSION INTO ULTRAVIOLET LIGHT

Lecture #2: Split Hopkinson Bar Systems

In Situ Imaging of Cold Atomic Gases

Elements of Quantum Optics

Diffuse Optical Spectroscopy: Analysis of BiophotonicsSignals Short Course in Computational Biophotonics Albert Cerussi Day 4/Lecture #2

OPTICAL DIAGNOSTICS TO STUDY SUPERCRITICAL CO 2 PROCESSES. A. Braeuer

Detection of laser generated ultrasonic waves in ablation regime by plane and confocal Fabry- Perot interferometers: Simulation

Chem Homework Set Answers

Technical University of Denmark

5.5. Calibration and Test Procedures

High-Power SrBr 2 Vapour Laser

10. OPTICAL COHERENCE TOMOGRAPHY

B 2 P 2, which implies that g B should be

Fatigue Testing of Materials by UV Pulsed Laser Irradiation

Optical Properties of Tissues after Laser Treatments in the Wavelength Range of nm

Supplemental material for Bound electron nonlinearity beyond the ionization threshold

An autonomous underwater telescope for measuring the scattering of light in the deep sea

VACUUM SUPPORT FOR A LARGE INTERFEROMETRIC REFERENCE SURFACE

CHIPP Plenary Meeting University of Geneva, June 12, 2008 W. Lustermann on behalf of the AX PET Collaboration

Check the LCLS Project website to verify 2 of 6 that this is the correct version prior to use.

SPARCLAB. Source For Plasma Accelerators and Radiation Compton with Laser And Beam

On the possibility to create a prototype of laser system for space debris movement control on the basis of the 3-meter telescope.

International Journal of Scientific & Engineering Research, Volume 8, Issue 2, February-2017 ISSN

A tunable corner-pumped Nd:YAG/YAG composite slab CW laser

31545 Medical Imaging systems

Transcription:

Dynamics of laser induced thermoelastic expansion of native and coagulated ex-vivo soft tissue samples and their optical and thermomechanical properties Behrouz Soroushian* a, William M. Whelan b, Michael C. Kolios a a Department of Physics, Ryerson University, Toronto, Canada; b Department of Physics, University of PEI, Charlottetown, Canada ABSTRACT The interferometric measurement of laser induced thermoelastic expansion of tissue samples can be used to estimate their optical, thermal and mechanical properties. This method was used to assess the Gruneisen coefficient and optical attenuation depth for native and coagulated ex-vivo bovine liver and porcine kidney samples. The results demonstrate decreases of 54% and 60% in the optical attenuation depth in bovine liver and porcine kidney after coagulation, respectively. The Gruneisen coefficient of native porcine kidney was determined to be 58% smaller (p < 0.05) that native bovine liver. The measured Gruneisen coefficients for native and coagulated ex-vivo porcine kidney were 0.07 ± 0.03 and 0.105 ± 0.0, respectively, whereas the Gruneisen coefficients for native and coagulated liver were 0.16 ± 0.036 and 0.17 ± 0.04, respectively. Our measurements indicate significant inter sample variability due likely to inherent variations in tissue optical absorption and surface preparation. Keywords: Thermoelastic expansion of tissue sample, surface displacement, optical attenuation depth, interferometry measurement. 1. INTRODUCTION Optoacoustic imaging (OA) relies on the detection of transient pressure waves induced in a target after its irradiation by short laser pulses. The characteristics of the laser induced optoacoustic signal depend upon the interplay of the optical, thermal and mechanical properties of the target. Optoacoustic techniques can exploit changes in these properties when a tissue undergoes thermal treatment and therefore can be used as an effective method for assessment and help minimize damage to normal tissue [1,]. Several previous studies have shown that changes in the optical properties of thermally treated tissues (for instance the increase in the reduced scattering coefficient) play an important role in contrast between laser induced optoacoustic signals from treated and untreated tissues [1, 3]. However the thermal treatment of tissues can also lead to changes in their thermo-mechanical properties [4]. The role of these property changes on optoacoustic contrast is less known. We recently employed an inteferometry method for studying laser induced thermoelastic deformation of tissue and tissue mimicking phantom samples. The mathematical equation governing laser induced thermoelastic expansion of a target was numerically solved. By using this technique we were able to estimate optical and thermo-mechanical properties of samples simultaneously at the same point of the sample which is very difficult to achieve with other techniques.. THEORY When a short laser pulse reaches the surface of a turbid target like tissue or a tissue mimicking phantom, it enters into the medium and will be distributed inside the target because of scattering. The final distribution of light inside the target can be approximated by an axial symmetric function with an exponentially decaying axial profile and a Gaussian radial profile: r 0 z w D I( r, I e e (1) Photons Plus Ultrasound: Imaging and Sensing 011, edited by Alexander A. Oraevsky, Lihong V. Wang, Proc. of SPIE Vol. 7899, 78990Z 011 SPIE CCC code: 1605-74/11/$18 doi: 10.1117/1.876011 Proc. of SPIE Vol. 7899 78990Z-1 Downloaded from SPIE Digital Library on 04 Mar 011 to 14.150.190.39. Terms of Use: http://spiedl.org/terms

D 1 [ 3 ( μ + μ )] μ () a a s in which μ a and μ' s are the optical absorption and reduced scattering coefficients of the material, respectively. The absorption of light with such a distribution profile results in generation of heat inside the target which leads to a temperature rise. One can assume that the initial temperature distribution and distribution of the absorbed light in the target are the same: r z Φμ w D a (3) T( r, T0 e e and T0 ρc where C V is the heat capacity in constant volume, Φ is the laser fluence, ρ is mass density. A nonuniform temperature distribution leads to the creation of mechanical stresses inside the target which in turn causes thermoelastic deformation of the target. The resulting thermoelastic deformation of the sample can be described by a wave equation in the form of: u E E Eβ ρ u ( u) T ( r, t (1 + σ ) (1 + σ )(1 σ ) 3(1 σ ) where u is the displacement vector, E is Young s modulus, σ is Poisson s ratio and β is the thermal expansion coefficient of the material. Numerical solutions to equation (4) are explored using the finite-difference time domain (FDTD) method using an appropriate set of boundary conditions. The details of the method are previously described elsewhere [8]. Briefly a staggered mesh was used to discretize the space in cylindrical coordinates. The temperature distribution inside the target was assumed to be axial symmetric to simplify the calculations. An absorbing boundary layer at the distal edge of the solution geometry with rigid constraints (displacement zero at this boundary) was used to reduce artificial wave reflections from the boundary. Since the numerical solutions of the wave equation require excessive computation time and resources, they are not used to describe all of dynamics for the thermoelastic expansion. However a solution to the thermoelastic wave equation when a range of properties for biological tissues are applied leads to three distinguishable time regimes [5]; the first regime is shortly after the pulse when the internal stresses resulting from the nonuniform temperature distribution cause thermoelastic deformation of the absorbing area. The time scale of this regime is determined by the optical attenuation depth and speed of sound in the sample (for soft tissues like liver around few hundred nanoseconds). The target surface displacement during this period can be approximated by [6]: u( t) u0 1 e c L t D where u 0 which is the maximum displacement of the target surface which is determined by thermo-mechanical properties of the sample and c L is longitudinal speed of sound in the target. After reaching its maximum displacement, the target surface sustains a quasi-steady state (the second regime). During this period, the system reaches equilibrium and although the individual stress components inside the target are not zero and even can be appreciable, the net force in any direction will be zero. For an initial temperature distribution in the form of (3) an analytical expression for the equilibrium surface displacement on the axis z during the quasi-steady-state time regime is given by [7]: u z (1 + σ ) Γ g0 ( R) Φ 3 ρc L where g 0 (R) is a function dependent upon the aspect ratio R, which is given by the ratio of the laser beam radius, w, to the optical attenuation depth, D (R w/d). The parameter Γ c Lβ C is the Grüneisen coefficient, an important thermomechanical property of the sample, which in addition to its dependence on the thermal expansion coefficient and the V specific heat capacity is also dependent on the longitudinal speed of sound in the target, c L. The Grüneisen coefficient can be determined using equation (6) by performing measurements for the equilibrium displacement of the sample surface after its exposure to laser pulses with different energies. V (4) (5) (6) Proc. of SPIE Vol. 7899 78990Z- Downloaded from SPIE Digital Library on 04 Mar 011 to 14.150.190.39. Terms of Use: http://spiedl.org/terms

.1 Measuring setup The interferometric setup used for our experiments has been described previously [8]. Laser pulses at a 750 nm wavelength and 6.5 ns pulse duration are emitted from an optical parametric oscillator (OPO Vibrant B-53, Opotek Inc., Carlsbad, CA, USA). These pulses with a repetition rate of 10 Hz are directed onto the target surface. By measuring the beam profile on a photographic page, the final beam size on the target location has been determined to be an elliptical cross section with major and minor axes of 5 and 3 mm respectively. A part of the pump beam was directed onto a fast silicon photodiode (DET10A, 00-1100 nm, Thorlabs, Newton, USA) to measure the pulse energy. The thermoelastic movements of the samples surface were measured using a modified Michelson interferometer. It is built around a cw solid-state laser (GCL-05-S, CrystaLaser, Reno, USA) operating at the wavelength of 53 nm. The beam from the laser was split in two parts with one directed towards a corner cube mounted on a rotating table permitting a constant modulation on the interference pattern when the target surface is not moving. The interference signals were detected by a photomultiplier (Hamamatsu H773P-11, Hamamatsu City, Japan) and after amplification (Miteq AU- 191, Miteq, Hauppauge, USA) were digitized on an oscilloscope (Agilent DSO6054A, Agilent, Santa Clara, USA) and recorded to be analyzed later. The interference pattern was recorded during a period of 10 μs from μs prior to the laser pulse incident onto the target. The resulting data were exported to Matlab for analysis. Displacement of the targets was calculated using Hilbert transform on the interferogram.. Sample preparation Seven bovine liver samples and six porcine kidney samples (each on a different day) were purchased from a local market. The samples remained sealed in their packaging and kept at 4 C to avoid gassing and to slow tissue degradation. From each piece of liver, two samples of about 5 x 5 x 1 cm were prepared. One of the samples was placed in a 70 C water bath for 30 minutes. The heated samples were then removed and cut in smaller rectangular pieces with dimensions of approximately x 1 x cm. The untreated liver sample was also cut in smaller pieces with similar dimensions as of the heated one. The accuracy of our interferometry technique is improved for tissue samples with a smooth, flat surface. As these two surface qualities can be affected by the heating process, smaller pieces were cut from the heated samples, yielding better surface uniformity. One piece from each sample was placed in the center of two identical aluminum cylindrical containers. Larger pieces of tissue were thermally treated since the process can cause the loss of flatness and smoothness of the surfaces which are used for interferometry measurements. Since these two qualities are important for the outcome of the experiments, the quality of sample surfaces was ensured by cutting smaller pieces out of the treated larger pieces of the tissue. Gelatin (Type A, SIGMA Life Science) was dissolved in hot water and after cooling to a lower temperature around 40 C was poured around the samples in the containers. The containers then were placed in a refrigerator at temperature of 6 C for letting the gelatin to harden and fix the samples in place. After solidification of the gelatin, the front face of the tissue sample, which was exposed to air, was used for the interferometric experimental measurements. The same procedure was used to prepare the kidney samples. The only difference is that care was taken to ensure that pieces of kidney are cut from the cortex to minimize local inhomogeneities. In addition, the kidney samples were thermally treated for a longer time, 3 hours. 3. RESULTS AND DISSCUSSIONS Coagulated ex-vivo bovine liver has an average smaller optical attenuation depth (880 ± 17 µm) compared to the native liver (474 ± 110 µm). From the interferometric measurements we found that coagulated ex-vivo porcine kidney samples similarly have an average smaller optical attenuation depth (66 ± 101 μm) compared to native kidney samples (1095 ± 01 μm). The difference between the two groups is significant (t 3.55, p < 0.01). The difference between the optical attenuation depth for native ex-vivo bovine liver and porcine kidney is not significant (t -1.85, p 0.09). Proc. of SPIE Vol. 7899 78990Z-3 Downloaded from SPIE Digital Library on 04 Mar 011 to 14.150.190.39. Terms of Use: http://spiedl.org/terms

Fig 1. Average optical attenuation depth for native and coagulated ex-vivo bovine liver (N7), and native and coagulated ex-vivo porcine kidney (N6). Error bars are ±1 standard deviation. We earlier reported that the interferometric measurements were unable to indicate a significant difference between values of Grüneisen coefficient for the native and coagulated liver tissues [8]. The calculated Gruneisen coefficients for native and coagulated kidney and liver are shown as a box-and-whisker diagram in the Fig.. The average value of the Gruneisen coefficient for both groups of the native and coagulated liver tissue samples was estimated 0.1. The difference between Gruneisen coefficient for native and coagulated porcine kidney samples is more pronounced. The measurements yield an average value of 0.07 ± 0.03 for the Gruneisen coefficient in the case of native ex-vivo porcine kidney while this value for the coagulated samples is 0.105 ± 0.0. The difference between these two groups is significant only to the level of p 0.091 (t -1.87). However when comparing the native kidney samples with native liver tissues, one finds that the difference between the Gruneisen coefficients for this two groups is significant (t -.61, p < 0.05). The Gruneisen coefficient of native kidney tissue samples is significantly smaller that native liver samples. Fig. Calculated Gruneisen coefficients for native and coagulated ex-vivo bovine liver (N7), and native and coagulated ex-vivo porcine kidney (N6). The stresses induced in the target by the light absorption and temperature rise cause the tissue surface to expand and reach a maximum displacement. However after reaching this maximum displacement the target surface may contract to a new equilibrium position. In our experiments with tissue mimicking phantoms, ex-vivo bovine liver and porcine kidney tissues, we observed that the coagulated samples generally have shown more prominent contractions compared to untreated samples. We earlier related this retraction behavior to differences of the mechanical properties between native Proc. of SPIE Vol. 7899 78990Z-4 Downloaded from SPIE Digital Library on 04 Mar 011 to 14.150.190.39. Terms of Use: http://spiedl.org/terms

and coagulated liver tissue samples. Further work is undertaken to quantify and analyze the similar behavior in the case of native and coagulated ex-vivo porcine kidney tissue samples. 4. CONCLUSION Interferometric measurements of the laser induced thermoelastic expansion of soft tissues can provide valuable information about their optical, thermal and mechanical properties. With this technique, the determination of the optical attenuation depth and Grüneisen coefficient of the tissue samples using the same tissue displacement dataset, ensures that the optical and thermomechanical properties are measured at the same tissue location. We found that the optical attenuation depth decreases significantly for coagulated porcine kidney and bovine liver compared to native tissues. In addition, the Gruneisen coefficient for coagulated porcine kidney was 33% higher than native kidney. However, no significant difference in the Gruneisen coefficient was observed for native and coagulated bovine liver. The considerable variability observed in our experiments are likely due to intrinsic differences between properties of tissues from different animals, intra sample heterogeneities and qualities of prepared sample surfaces. Thanks are due to Arthur Worthington for his assistance in the experiments. Financial support was provided by the Atlantic Innovation Fund, the Natural Sciences and Engineering Research Council of Canada, the Canadian Institutes of Health Research (grant CHRPJ33745-60) and the Canada Foundation for Innovation (CFI). This research was undertaken, in part, thanks to funding from the Canada Research Chairs Program awarded to Michael Kolios and Whilliam Whelan. REFERENCES [1] Larin, K.V., Larina, I.V., Esenaliev, R.O. "Monitoring of tissue coagulation during thermotherapy using optoacoustic technique", Journal of Physics D: Applied Physics, 38 (15), 645-653, (005). [] Arsenault M. G., Kolios M. C. and Whelan W. M., Optoacoustic detection of thermal lesions, Proc. of SPIE Vol. 7177, 71771V, (009). [3] Iizuka, M.N., Sherar, M.D. and Vitkin, I.A. "Optical phantom materials for near infrared laser photocoagulation studies", Lasers in Surgery and Medicine 5, 159-169 (1999). [4] Kiss, M.Z., Varghese, T., Hall, T.J. "Viscoelastic characterization of in vitro canine tissue", Physics in Medicine and Biology, 49 (18), 407-418, (004). [5] Albagli, D., Dark, M., Von Rosenberg, C., Perelman, L., Itzkan, I. and Feld, M. S., Laser-induced thermoelastic deformation: A three-dimensional solution and its application to the ablation of biological tissue. Med. Phys., 1(8), 133-1331 (1994). [6] Payne B. P., Venugopalan V., Mikic B. B. and Nishioka N. S., "Optoacoustic determination of optical attenuation depth using interferometric detection", J. Biomed. Optics 8(), 64 7 (003). [7] Dark M. L., Perelman L. T., Itzkan I., Schaffer J. L. and Feld M. S., "Physical properties of hydrated tissue determined by surface interferometry of laser-induced thermoelastic deformation", Phys. Med. Biol., 45, 59-539 (000). [8] Soroushian B., Whelan, W. M. and Kolios M. C., Study of laser-induced thermoelastic deformation of native and coagulated ex-vivo bovine liver tissues for estimating their optical and thermomechanical properties, J. Biomed. Opt. 15, 06500 (010). Proc. of SPIE Vol. 7899 78990Z-5 Downloaded from SPIE Digital Library on 04 Mar 011 to 14.150.190.39. Terms of Use: http://spiedl.org/terms