SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

Similar documents
APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

Joint Committee for Traceability in Laboratory Medicine Terminology. R. Wielgosz and S. Maniguet

Document No: TR 12 Issue No: 1

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

R SANAS Page 1 of 7

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

Precision estimated by series of analysis ISO and Approach Duplicate Approach

Uncertainty of Measurement A Concept

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Measurement Uncertainty, March 2009, F. Cordeiro 1

Part 5: Total stations

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement. Second Edition. Draft: June 1999

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

Practical Interpretation of Metrological Traceability on Gauge Block as to VIM 3

Comparability and Traceability Point of view from a metrological institute

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY

Uncertainty of Measurement

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis*

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

Introduction to the evaluation of uncertainty

A guide to expression of uncertainty of measurements QA4EO-QAEO-GEN-DQK-006

Measurement Uncertainty Principles and Implementation in QC

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Calibration and verification: Two procedures having comparable objectives and results

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s):

Guide to the Expression of Uncertainty in Measurement (GUM) and its supplemental guides

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

A Unified Approach to Uncertainty for Quality Improvement

A2LA. G118: Guidance for Defining the Scope of Accreditation for Calibration Laboratories

Vocabulary of Metrology

Essentials of expressing measurement uncertainty

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Validation and Standardization of (Bio)Analytical Methods

Uncertainty Propagation for Force Calibration Systems

Certificate of Accreditation to ISO/IEC 17025:2005

Part IVB Quality Assurance/Validation of Analytical Methods

A basic introduction to reference materials. POPs Strategy

Calibration Traceability Guidelines

Quantifying Uncertainty in Analytical Measurement

Unit 4. Statistics, Detection Limits and Uncertainty. Experts Teaching from Practical Experience

Quantities and Measurement Units

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

TITLE OF THE CD (English): New OIML Document

Measurement uncertainty and legal limits in analytical measurements

Decision rules applied to conformity assessment

Analysis of interlaboratory comparison when the measurements are not normally distributed

GUIDE Uncertainty of measurement Part 3: Guide to the expression of uncertainty in measurement (GUM:1995)

CALIBRATION CERTIFICATE # 503

Measurement & Uncertainty - Basic concept and its application

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

This document is a preview generated by EVS

Process Accuracy Ratio vs Process Uncertainty Ratio (Risk Mitigation: Calibration and the Customer s Process)

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

Technical Guide 4 A Guide on Measurement Uncertainty in Medical Testing

M3003 EDITION 2 JANUARY 2007

Certificate of Accreditation

Reference Materials and Proficiency Testing. CropLife International Meeting October 2015 Gina M. Clapper

Internal CQ. Performance criteria for internal QC how to define a significant deviation from the expected results

ISO Water quality Measurement of polonium 210 activity concentration in water by alpha spectrometry

Specific Accreditation Guidance

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis

CALIBRATION REPORT FOR THERMOHYDROMETER

HOW TO EXPRESS THE RELIABILITY OF MEASUREMENT RESULTS. The role of metrological traceability and measurement uncertainty-

Revision of the Guide to the expression of uncertainty in measurement impact on national metrology institutes and industry

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

Certificate of Accreditation

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

This document is a preview generated by EVS

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

Determination of particle size distribution Single particle light interaction methods. Part 4:

EA-10/14. EA Guidelines on the Calibration of Static Torque Measuring Devices. Publication Reference PURPOSE

Method of Estimating Uncertainty of Measurement Values in Case of Measurement Methods Involving Non-Linear Calibration

New Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018

Certificate of Accreditation

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

Measurement & Uncertainty - Concept and its application

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Conformity assessment of a high-performance liquid chromatography calibration.

Requirements of a reference measurement procedure and how they relate to a certified reference material for ctni that is fit for purpose

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Maha Ireland Ltd. 629 Jordanstown Avenue, Greenogue Business Park, Rathcoole, Co. Dublin. Calibration Laboratory Registration number: 287C

STATISTICAL PROCESS CONTROL - THE IMPORTANCE OF USING CALIBRATED MEASUREMENT EQUIPMENT. CASE STUDY

Measurement Uncertainty - How to Calculate It In The Medical Laboratory

EA 4/02. Expression of the Uncertainty of Measurement in Calibration. Publication Reference

Measurement Uncertainty Knowing the Unknown

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

Certified Reference Material - Certificate of Analysis

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ALLOMETRICS, INC Bayport Blvd. Seabrook, TX Terry Baldwin Phone: CALIBRATION

TNI V1M Standard Update Guidance on Detection and Quantitation

Transcription:

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM) LABORATORY ACCREDITATION SCHEME OF MALAYSIA SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014) JABATAN STANDARD MALAYSIA Department of Standards Malaysia

TABLE OF CONTENTS Page 1 Introduction 1 2 Scope 1 3 Definitions 1 4 Requirements of MS ISO/IEC 17025 & MS ISO 15189 2 5 Policy 4 6 Tests for which uncertainty does not apply 5 7 Tests for which uncertainties applies 5 8 Evidence of compliance 5 9 Guidance on steps to be taken by laboratory 6 Bibliography 7

1 Introduction 1.1 This document specifies the policy of the Department of Standards Malaysia (Standards Malaysia) with regard to measurement uncertainty requirements imposed on applicant and accredited testing laboratories under the Skim Akreditasi Makmal Malaysia (SAMM). 1.2 This policy document should be read in conjunction with other SAMM requirements. 2 Scope This policy provides guidelines on the estimation and expansion of measurement uncertainty and applies to all applicants and accredited testing laboratories. 3 Definitions For the purpose of the document, the following definitions shall apply: 3.1 Calibration (VIM 3 clause 2.39) Operation that, under specified conditions, in a first step, establishes a relation between the quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication NOTE 1 A calibration may be expressed by a statement, calibration function, calibration diagram, calibration curve, or calibration table. In some cases, it may consist of an additive or multiplicative correction of the indication with associated measurement uncertainty. NOTE 2 Calibration should not be confused with adjustment of a measuring system, often mistakenly called self-calibration, nor with verification of calibration. NOTE 3 Often, the first step alone in the above definition is perceived as being calibration. 3.2 Metrological traceability (VIM 3 clause 2.41) Property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty. NOTE: For this definition, a reference can be a definition of a measurement unit through its practical realization, or a measurement procedure including the measurement unit for a non-ordinal quantity, or a measurement standard. The ILAC considers the elements for confirming metrological traceability to be an unbroken metrological traceability chain to an international measurement SAMM Policy 5, Issue 2, 28 February 2007 (Amd.1, 11 August 2014) Page 1 of 7

standard or a national measurement standard, a documented measurement uncertainty, a documented measurement procedure, accredited technical competence, metrological traceability to the SI, and calibration intervals. 3.3 Measurement uncertainty (VIM 3 clause 2.26) Non-negative parameter characterizing the dispersion of the quantity values being attributed to a measurand, based on the information used NOTE: The parameter may be, for example, a standard deviation called standard measurement uncertainty (or a specified multiple of it), or the half-width of an interval, having a stated coverage probability. Measurement uncertainty comprises, in general, many components. Some of these may be evaluated by Type A evaluation of measurement uncertainty from the statistical distribution of the quantity values from series of measurements and can be characterized by standard deviations. The other components, which may be evaluated by Type B evaluation of measurement uncertainty, can also be characterized by standard deviations, evaluated from probability density functions based on experience or other information. In general, for a given set of information, it is understood that the measurement uncertainty is associated with a stated quantity value attributed to the measurand. A modification of this value results in a modification of the associated uncertainty. 3.4 Measurand (VIM 3 clause 2.3) Quantity intended to be measured. 3.5 Measurement (VIM 3 clause 2.1) Process of experimentally obtaining one or more quantity values that can reasonably be attributed to a quantity. 4 Requirements of MS ISO/IEC 17025 and MS ISO 15189 4.1 MS ISO/IEC 17025, clause 5.4.6 states the following: 5.4.6.1 A calibration laboratory, or a testing laboratory performing its own calibrations, shall have and shall apply a procedure to estimate the uncertainty of measurement for all calibrations and types of calibrations. 5.4.6.2 Testing laboratories shall have and shall apply procedures for estimating uncertainty of measurement. In certain cases the nature of the test method may preclude rigorous, metrologically and statistically valid, calculation of uncertainty of measurement. In these cases the laboratory shall at least attempt to identify all the components of uncertainty and make a reasonable estimation, and shall ensure that the form of reporting of the result does not give a wrong impression of the uncertainty. Reasonable estimation shall be based on knowledge of the performance of the method and on the measurement scope and shall make use of, for example, previous experience and validation data. SAMM Policy 5, Issue 2, 28 February 2007 (Amd.1, 11 August 2014) Page 2 of 7

NOTE 1: The degree of rigor needed in an estimation of uncertainty of measurement depends on factors such as: the requirements of the test method; the requirements by the customer; the existence of narrow limits on which decisions on conformity to a specification are based. NOTE 2: In those cases where a well-recognized test method specifies limits to the values of the major sources of uncertainty of measurement and specifies the form of presentation of calculated results, the laboratory is considered to have satisfied this clause by following the test method and reporting instructions (see 5.10). 5.4.6.3 When estimating the uncertainty of measurement, all uncertainty components which are of importance in the given situation shall be taken into account using appropriate methods of analysis. NOTE 1: Sources of uncertainty include, but are not necessarily limited to, the reference standards and reference materials used, methods and equipment used, environmental conditions, properties and condition of the item being tested or calibrated, and the operator. NOTE 2: The predicted long-term behavior of the tested and/or calibrated item is not normally taken into account when estimating measurement uncertainty. NOTE 3: For further information, see ISO 5725 and the Guide to the Expression of Uncertainty in Measurement (GUM). 4.2 MS ISO/IEC 17025 clause 5.10.3.1 c states the following: [Test reports shall, where necessary for the interpretation of the test results, include the following: ] c) where applicable, a statement on the estimated uncertainty of measurement; information on uncertainty is needed in test reports when it is relevant to the validity or application of the test results, when a customer's instructions so requires, or when the uncertainty affects compliance to a specification limit. 4.3 MS ISO 15189, clause 5.5.1.4 Measurement uncertainty of measured quantity values states the following: The laboratory shall determine measurement uncertainty for each measurement procedure in the examination phase used to report measured quantity values on patients samples. The laboratory shall define the performance requirements for the measurement uncertainty of each measurement procedure and regularly review estimates of measurement uncertainty. NOTE 1: The relevant uncertainty components are those associated with the actual measurement process, commencing with the presentation of the sample to the measurement procedure and ending with the output of the measured value. SAMM Policy 5, Issue 2, 28 February 2007 (Amd.1, 11 August 2014) Page 3 of 7

NOTE 2: Measurement uncertainties may be calculated using quantity values obtained by the measurement of quality control materials under intermediate precision conditions that include as many routine changes as reasonably possible in the standard operation of a measurement procedure, e.g. changes of reagent and calibrator batches, different operators, scheduled instrument maintenance. NOTE 3: Examples of the practical utility of measurement uncertainty estimates might include confirmation that patients values meet quality goals set by the laboratory and meaningful comparison of a patient value with a previous value of the same type or with a clinical decision value. The laboratory shall consider measurement uncertainty when interpreting measured quantity values. Upon request, the laboratory shall make its estimates of measurement uncertainty available to laboratory users. Where examinations include a measurement step but do not report a measured quantity value, the laboratory should calculate the uncertainty of the measurement step where it has utility in assessing the reliability of the examination procedure or has influence on the reported result. 5 Policy 5.1 General It is the policy of SAMM that testing laboratories accredited by Standards Malaysia shall comply with the requirements of MS ISO/IEC 17025 or MS ISO 15189 in relation to the estimation and reporting of uncertainty of measurement in testing. The complexity involved in estimation of uncertainty of measurement in the case of testing varies considerably from one test field to another and also within one field itself. Clause 5.4.6.2 of MS ISO/IEC 17025 allows a less rigorous process than that which can be followed for calibration. Standards Malaysia requires the testing laboratory to make a reasonable estimate of the uncertainty. NOTE: Uncertainty components / budgets are a combination of many factors that may include, but are not limited to: Reference standards; Reference materials; Test methods used; Equipment used; Environmental conditions; Properties and condition of item being tested; Calibration; Operator; and Known physical characteristics of components. SAMM Policy 5, Issue 2, 28 February 2007 (Amd.1, 11 August 2014) Page 4 of 7

6 Tests for which uncertainty does not apply The following types of test methods do not require estimate of the measurement uncertainty: 6.1 Qualitative tests. (Test results are not numerical such as pass/fail; positive/negative or other qualitative expressions). 6.2 In those cases where a well-recognised test method specifies limits to the values of the major sources of uncertainty of measurement and specifies the form of presentation of calculated results, the laboratory is considered to have satisfied the clause by following the test methods and reporting instructions. 7 Tests for which uncertainty applies Where a test produces numerical results (or the reported result is based on numerical result), those uncertainty of those numerical results shall be estimated. In cases where the nature of the test method precludes rigorous, metrologically and statistically valid estimation, of the measurement uncertainty, a laboratory shall make a reasonable attempt to estimate the uncertainties of those results. This applies whether the test methods are rational or empirical. The following types of test methods need a reasonable estimate of the measurement uncertainty. The tests under this however, require varying degrees of rigor when estimating the measurement uncertainty. 7.1 Consensus tests methods or methods or published by regulatory bodies (examples: chemical, environmental, or microbiological test methods published by AOAC, ASTM, DOE) for which the major source of uncertainty of measurement is not stated in the method. 7.2 Test methods that need identification of the major components of uncertainty and a reasonable estimate of measurement uncertainty. (Such as laboratorydeveloped methods) [Laboratory-developed methods also require method validation] 7.3 Published methods (such as found in text books, journal and bulletins). 8 Evidence of compliance 8.1 Objective evidence of compliance, including the procedure for estimating measurement uncertainty, and the resulting documented uncertainty budgets and uncertainty values, where relevant, will be reviewed in accordance with the following schedule: a) For applicant laboratories, compliance will be verified by the assessor during the laboratory's on-site assessment. SAMM Policy 5, Issue 2, 28 February 2007 (Amd.1, 11 August 2014) Page 5 of 7

b) For accredited laboratories due for renewal, compliance will be verified by the assessor during laboratory s on-site assessment. All necessary measurement uncertainties, including the uncertainty budgets shall be documented and made available. 9 Guidance on steps to be taken by laboratory 9.1 The laboratory shall have a procedure(s) for calculating their measurement uncertainty. 9.2 Estimating measurement uncertainty The laboratory may adopt the following approach: a) Specifying the measurand i.e. relationship between the measurand and the variables of the analytical procedure. b) Describe the measurement/testing procedure. Flowchart with the description of all steps of analytical procedures. c) Identify all uncertainty sources. The uncertainty of each variable in the uncertainty budget is identified by mean of a cause and effect diagram. Identify major uncertainty components (Laboratories may adopt methods provided in the EURACHEM / CITAC / CSLI / NPAAC Guide. Other methods will also be considered, if they present a valid approach to the estimation of measurement uncertainty). d) Quantification and combination of uncertainties. The above identified uncertainties are quantified and combined to give the total uncertainty on the measurand. 9.3 Reporting expanded uncertainty The result x should be stated together with the expanded uncertainty U calculated using coverage factor k=2. The following form is recommended: (Result): (x U) (units) where the reported uncertainty is (an expanded uncertainty as defined in the International vocabulary of metrology - Basic and general concepts and associated terms (VIM), 3 rd ed.) calculated using a coverage factor of 2, (which gives a level of confidence of approximately 95%). The coverage factor should be adjusted to show the value actually used. SAMM Policy 5, Issue 2, 28 February 2007 (Amd.1, 11 August 2014) Page 6 of 7

Bibliography 1. MS ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories 2. MS ISO 15189, Medical laboratories - Requirements for quality and competence 3. EURACHEM/CITAC Guide CG 4:2012, Quantifying uncertainty in analytical measurement 4. JCGM 100:2008 Evaluation of measurement data - Guide to the expression of uncertainty in measurement (GUM) 5. JCGM 200:2012, International vocabulary of metrology - Basic and general concepts and associated terms (VIM), 3rd edition 6. ISO 5725 Accuracy of measurement methods and results package 7. Clinical and Laboratory Standard Institute (CLIS): Guideline describing the principles of estimating measurement uncertainty (C51-A) 8. The National Pathology Accreditation Advisory Council (NPAAC): Requirements for the estimation of measurement uncertainty (2007) SAMM Policy 5, Issue 2, 28 February 2007 (Amd.1, 11 August 2014) Page 7 of 7