Quantifying and Verifying trace stability of vector network analyzers at frequencies below 1GHz

Similar documents
Lecture 13. Vector Network Analyzers and Signal Flow Graphs

Advancements in mm-wave On-Wafer Measurements: A Commercial Multi-Line TRL Calibration Author: Leonard Hayden Presenter: Gavin Fisher

VNA Tools II S-parameter uncertainty calculation

Accuracy Estimations of S-Parameter Measurements at the Wafer-Level

Estimation of the Cosmic Microwave Background Radiation

Keysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement. Application Note

Ripple Method for Evaluating Residual Errors

Performance of Antenna Measurement Setup

Solutions to Problems in Chapter 5

SHENZHEN MORLAB COMMUNICATIONS TECHNOLOGY CO., LTD

Experiment 06 - Extraction of Transmission Line Parameters

5730A High Performance Multifunction Calibrator. Extended specifications

SHENZHEN SEM.TEST TECHNOLOGY CO., LTD. 1/F, BUILDING A, HONGWEI INDUSTRIAL PARK, LIUXIAN 2ND ROAD BAO'AN DISTRICT SHENZHEN (518101), CHINA

Application Note. Electrical. Optical

THE search for higher accuracy in measurements of S-

Calibration Uncertainty Estimation. at the Wafer Level

Front End Electronics Internal Noise Source Characterization

Practical Considerations and Solutions for Temperature-Dependent S-Parameter Measurement for Accurate Parameter Extraction of

SPARQ S-Parameter Measurements with Impulse Response Time Limiting

This section reviews the basic theory of accuracy enhancement for one-port networks.

Solder Joint Reliability Report

Nonlinear Vector Network Analyzer Applications

<Insert Picture Here> Inter-Dependence of Dielectric and Conductive Losses in Interconnects

Qualification of tabulated scattering parameters

Broad-band space conservative on wafer network analyzer calibrations with more complex SOLT definitions

Light Work Memo 4 rev 3: System Temperature Measurements July 7 1

Temperature Coefficients for DC Resistance of Match and Reference Attenuators

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard

Microwave Impedance Measurement for Nanoelectronics

Characterising Properties and Loss in High Powered Metamaterials

Omar M. Ramahi University of Waterloo Waterloo, Ontario, Canada

Physical Noise Sources

Electricity and Light Pre Lab Questions

Advances in Radio Science

RAPID growth in satellite-communications and mobile-communications

Solutions to Problems in Chapter 6

Observation of neutral hydrogen using FFT spectrometer Argos on a 5m telescope

Winds on Titan: First results from the Huygens Doppler Wind Experiment

Lecture 36 Date:

Solder Joint Reliability Lucent X Summary Report

Determination of sound absorption coefficient in impedance tubes according to ISO (2 appendices)

Surface Mount Chip Capacitors

Measurement and Calibration of a High-Sensitivity Microwave Power Sensor with an Attenuator

Measurement of Galactic Rotation Curve

Design Reference Mission for SKA1 P. Dewdney System Delta CoDR

The LM741C Integrated Circuit As A Differential Amplifier Building The Electronic Thermometer

IPC-TM-650 TEST METHODS MANUAL

Experiment 9 Equivalent Circuits

Contents. Transmission Lines The Smith Chart Vector Network Analyser (VNA) ü structure ü calibration ü operation. Measurements

The Polarimeter for Observing Inflationary Cosmology at the Reionization Epoch (POINCARE)

Radio Observation of Milky Way at MHz. Amateur Radio Astronomy Observation of the Milky Way at MHz from the Northern Hemisphere

Appendix D Probe Calibration Data Sheets

SRV02-Series Rotary Experiment # 1. Position Control. Student Handout

FREQUENTLY ASKED QUESTIONS RF & MICROWAVE PRODUCTS

DesignCon Evaluation of Relative Humidity and Temperature Effects on Scattering Parameters in Transmission Systems 1

Temperature measurement

III. Spherical Waves and Radiation

Importance and estimation of mismatch uncertainty for RF parameters in calibration laboratories

Applications. Broadband Gain Block Mobile Infrastructure Cellular, GSM PCS, WCDMA, WiBro, WiMax W-LAN / ISM RFID / Fixed Wireless

Probe Tip Characterization Using Self-Calibration. Lab. RF Modeling

EE Power Gain and Amplifier Design 10/31/2017

Series CCR-39S Multi-Throw DC-12 GHz, SP9T & SP10T Latching Coaxial Switch

Microwave Loss Reduction in Cryogenically Cooled Conductors R. Finger and A. R. Kerr

Estimation of Complex and Linear Uncertainities in S-Parameter Measurements for Metrology Applications

An Introduction to Radio Astronomy

Biol/Chem 4900/4912. Forensic Internship Lecture 5

S-Parameter Calibration of Two-Port Setup: Gavin Fisher Cascade Microtech

Error Correction in Vector Network Analyzers

ALMA Memo 373 Relative Pointing Sensitivity at 30 and 90 GHz for the ALMA Test Interferometer M.A. Holdaway and Jeff Mangum National Radio Astronomy O

Search for 21cm Reionization

db: Units & Calculations

Thermal Conductivity and Electrical Loss of Thin Wall Millimeter Wave Stainless Steel Waveguides

LightWork Memo 18: Galactic Spectra Data Overview 1

An Introduction to ASKAP Bringing Radio Interferometers Into the Multi-pixel Era

Pulses in transmission lines

Traceable UV-Vis Reference Materials

Series CCR-39S Multi-Throw DC-12 GHz, SP9T & SP10T Latching Coaxial Switch

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

Orbit and Transmit Characteristics of the CloudSat Cloud Profiling Radar (CPR) JPL Document No. D-29695

Technique for the electric and magnetic parameter measurement of powdered materials

Unknown Thru Calibration Algorithm

A UNEQUAL COUPLED-LINE WILKINSON POWER DI- VIDER FOR ARBITRARY TERMINATED IMPEDANCES

Ballistic transport at GHz frequencies in ungated HEMT structures

SHOCK AND VIBRATION RESPONSE SPECTRA COURSE Unit 6B. Notes on the Fourier Transform Magnitude

Steady State Frequency Response Using Bode Plots

Microstrip Antennas. Prof. Girish Kumar Electrical Engineering Department, IIT Bombay. (022)

Radiation noise source modeling and near-field coupling estimation in RF interference

Lecture 32: Astronomy 101

CE Test Report. : abn (1x1) + Bluetooth (2.1) module

Contents. Transmission Lines The Smith Chart Vector Network Analyser (VNA) ü structure ü calibration ü operation. Measurements

Electromagnetic Wave Absorption Technology for Stub Effects Mitigation

MFJ 259/259B Operation & Simplified Calibration

Impact of adapters on LISN's input impedance calibration: How to improve accuracy and reliability of the measurements? Part 2

TECHNO INDIA BATANAGAR

Theory and Operation of Cold Field-effect Transistor (FET) External Parasitic Parameter Extraction

Mimir NIR Spectroscopy Data Processing Cookbook V2.0 DPC

Nonlinear kinetic inductance in TiN/NbTiN microresonators and Transmission lines. Peter Day Byeong-Ho Eom Rick Leduc Jonas Zmuidzinas

NSG 4070C APPLICATION FOR MIL-STD 461G CS 114

Transcription:

1 Quantifying and Verifying trace stability of vector network analyzers at frequencies below 1GHz Hamdi Mani, Chris Groppi, Judd Bowman Abstract we describe measurements done to check the low frequency stability of network analyzers from different vendors. The data on this report shows that the Agilent PNA N5230A and the PNA-X N5242A magnitude and phase measurements are more stable at frequencies below 100MHz than the R&S ZVA24. The stability of the PNA VNA is needed for low frequency measurements including- but not limited to- video amplifiers, low loss transmission lines, low frequency dipole antennas, measurement of VHF components Index Terms Vector Network Analyzers, test equipment precision, noise in instruments, VHF band. I. INTRODUCTION here is a need for characterizing radio telescope Tcomponents below 1GHz in experimental cosmology such as the experiments to detect the epoch of re-ionization. These experiments are trying to measure the faint power spectrum of the highly red shifted 21cm hydrogen line. The spectral line frequency is red shifted (z>6) from the microwave L Band region of around 1420 MHz to the 75-200MHz frequency range [1]. Measuring components in this frequency range requires precise and accurate test equipment.the requirement for precision is not hard to satisfy since there are well established calibration techniques and NIST traceable calibration standards. The accuracy requirement deals with repeatability of the measurement over some period of time or what is referred to as trace noise. It defines the maximum resolution of the measurement. The latter requirement is investigated in this report. The trace noise of both magnitude and phase of the reflection coefficient are studied since the impedance is a complex quantity and the uncertainties of the magnitude and phase contribute to the uncertainty of the impedance. II. MEASUREMENT PROCEDURE To quantify the stability of the trace of a vector network analyzer at low frequencies, we need to measure the statistical dispersion over time of few sample readings. Few quantities can describe this dispersion: Standard deviation Interquartile range Range Mean difference Median absolute deviation Average absolute deviation Distance standard deviation Coefficient of variation Quartile coefficient of dispersion The author choose standard deviation as the statistical parameter. But other parameters might be used. Physical scientists and especially electrical engineers use the term "Root Mean Square" (RMS) as a synonym for standard deviation. The 2 quantities are actually different. The standard deviation is the root mean square of the deviation of the individual samples from the mean.the RMS of samples is equal to the standard deviation only if the mean of the samples (i.e. the expected value) is equal to 0. The definition of the standard deviation for N samples: whose arithmetic mean is defined by : Hamdi Mani is a physics student at Arizona state university and works at the Terahertz instrumentation (http://thz.asu.edu/cgroppi/ )and low frequency Cosmology Labs. He can be reached at: hamdi.mani@asu.edu Chris Groppi is an assistant professor at ASU and is the Terahertz astronomy group leader.(cgroppi@asu.edu) Judd Bowman is an assistant professor at ASU and is the group leader for the Low Freq radio astronomy and Cosmology group (Judd.Bowman@asu.edu) A large standard deviation from the mean is an indication of higher trace noise; a lower standard deviation indicates more stable trace. The network analyzers settings are: Freq: 10MHz-1GHz

2 Number of Points: 991(1MHz Step) Test port power: 0dBm IF Bandwidth: 500Hz (reduce IF Bandwidth at low frequency option was enabled) A 1-port SOLT Calibration was performed on all analyzers, the cal kit used is the HP 85052B 3.5mm manual kit (broadband load used). The PNA N5230A [2] was calibrated at the end of test cables after a long warm up time (more than 12 hours).the PNA-XN5242A was calibrated at the test port after 10 minutes warm up time. The ZVA24 [3] was calibrated at the test port with no cables after 10 minutes warm up time. The 2 network analyzers were terminated by good 50 ohm load, the averaging was turned off and 10 traces were saved within 5 minutes. The data was saved as a touchstone s1p file and analyzed and plotted in Microsoft Excel and in AWR Microwave Office. Data was also collected with the averaging set to 10 and then to 100. It is important to mention that measuring 50ohm load (instead of a short, open or any other termination value) is most demanding since the reflected power is very small and thus is a tough test of the dynamic range and stability of the instruments. III. TRACE NOISE ON THE MAGNITUDE DATA The 10 s11[db] measurements taken with the Agilent PNA-L N5230A are shown on Fig1a.The 10 measurements taken with the R&S ZVA24 are plotted with the same scale and shown on Fig1b.The measurements by the PNA-X N5242A are shown on Fig1.c Fig1b. 10 traces measured by the R&S ZVA24. The 3 plots 1.a, 1.b and 1.c show differences both in absolute measurement and on trace noise especially below 200MHz. The standard deviation is computed at all frequency points and plotted on Fig2a and Fig2.b for the PNA and ZVA VNAs respectively. Fig3a and Fig3b are the same plots than 2a and 2b just zoomed on the frequency range of interest. The standard deviation at 10MHz is very high for both instruments; this is due to coupler roll off at 10MHz. At 100MHz, the 10 measurements taken with the PNA N5230A are about 0.1dB from the mean while the 10 measurements taken with the ZVA24 are within 0.9dB. Fig1a. 10 traces measured by the Agilent PNA

3 Fig2a. Standard deviation in [db] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A VNA and plotted on the 10MHz-1GHz Fig3a. Standard deviation in [db] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A VNA and plotted on the 50MHz-300MHz Fig2b. Standard deviation in [db] of the 10 samples reflection coefficient (S11) measured by the ZVA24 VNA and plotted on the 10MHz-1GHz Fig3b. Standard deviation in [db] of the 10 samples reflection coefficient (S11) measured by the ZVA24 VNA and plotted on the 50MHz-300MHz Fig2c. Standard deviation in [db] of the 10 samples reflection coefficient (S11) measured by the PNAX N5242A VNA and plotted on the 10MHz-1GHz range Fig3c. Standard deviation in [db] of the 10 samples reflection coefficient (S11) measured by the PNAX N5242A VNA and plotted on the 50MHz-300MHz The reflection coefficient value is converted from db to linear and the standard deviation is plotted in Fig4a and Fig4b.

4 Fig4a. Standard deviation in [linear] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A VNA and plotted on the 10MHz-1GHz Fig5a. Standard deviation in [linear] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A VNA and plotted on the 50MHz- 300MHz Fig4b. Standard deviation in [linear] of the 10 samples reflection coefficient (S11) measured by the ZVA24 VNA and plotted on the 10MHz-1GHz Fig5b. Standard deviation in [linear] of the 10 samples reflection coefficient (S11) measured by the ZVA24 VNA and plotted on the 50MHz-300MHz Fig4b. Standard deviation in [linear] of the 10 samples reflection coefficient (S11) measured by the PNA-X N5242A VNA and plotted on the 10MHz- 1GHz The standard deviation in linear s11 is plotted on the 50MHz-300MHz (Fig5a and Fig5b) Fig5c. Standard deviation in [linear] of the 10 samples reflection coefficient (S11) measured by the PNA-X N5242A VNA and plotted on the 50MHz- 300MHz The standard deviation is then plotted in % Reflection. The reflection coefficient S11 varies from 0% ( total absorption for an ideal matched load) and 100% for a total reflection (perfect open or short).fig6a and Fig6b are plots in the 10MHz-1GHz range and Fig7a and Fig7B are zoomed plots in the 50MHz- 300MHz band.

5 Fig5a. Standard deviation in [% reflection] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A VNA and plotted on the 10MHz-1GHz Fig6a. Standard deviation in [% reflection] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A VNA and plotted on the 50MHz-300MHz Fig5b. Standard deviation in [% reflection] of the 10 samples reflection coefficient (S11) measured by the ZVA24 VNA and plotted on the 10MHz- 1GHz Fig6b. Standard deviation in [% reflection] of the 10 samples reflection coefficient (S11) measured by the ZVA24 VNA and plotted on the 50MHz- 300MHz Fig5c. Standard deviation in [% reflection] of the 10 samples reflection coefficient (S11) measured by the PNAx N5242A VNA and plotted on the 10MHz-1GHz Fig6c. Standard deviation in [% reflection] of the 10 samples reflection coefficient (S11) measured by the PNAX N5242A VNA and plotted on the 50MHz-300MHz The standard deviation of S11 in [db], [Linear] and [% Reflection] are plotted on the same graphs for both instruments for comparison: Fig7a, Fig7b and Fig7c.

6 Fig7d. This plot shows how the standard deviation varies in a 10MHz bandwidth starting at 100MHz for both instruments. Fig7a. Standard deviation in [db] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A,PNAX N5242A and ZVA VNAs and the plotted on the 50MHz-300MHz Data was also taken with the averaging was turned on. Data was taken after averaging 10 measurements for the first measurement (Fig8a and Fig8b) and 100measurements (Fig9a and Fig9c). Fig7b. Standard deviation in [Linear] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A,the PNAX N5242A and ZVA VNAs and the plotted on the 50MHz-300MHz Fig8a. 10 averaged S11 [db] measurements recorded on the ZVA24 and PNA N5230A and N5242A from 10MHz to 1GHz. Fig7c. Standard deviation in [% Reflection] of the 10 samples reflection coefficient (S11) measured by the PNA N5230A and ZVA VNAs and the plotted on the 50MHz-300MHz Fig8b. 10 averaged S11 [db] measurements recorded on the ZVA24 and PNA L and X from 50MHz to 300MHz. It is interesting to note that since the trace noise is a random phenomenon, the standard deviation of the ZVA is higher and varies more in frequency as is shown on Fig7d.

7 Fig9a. 100 averaged S11 [db] measurements recorded on the ZVA24 and PNA L and X from 10MHz to 1GHz. Fig10a. The 10 measurements of the phase of S11 taken with the PNA N5230A with no averaging Fig9b. 100 averaged S11 [db] measurements recorded on the ZVA24 and PNA L and X from 50MHz to 300Hz. IV. TRACE NOISE ON THE PHASE DATA The 10 measurements of the Phase of the reflection coefficient of the 50 ohm load with No averaging is plotted in Fig10a and Fig10b for the PNA N5230A and ZVA VNAs respectively. Fig10b. The 10 measurements of the phase of S11 taken with the ZVA24 with no averaging The standard deviation of both measurements is shown on Fig11a and Fig11b. The standard deviation of the phase for both instruments is shown on Fig12. We can see that the noise of the phase measurements is higher for the ZVA24 VNA and also varies a quicker in frequency. This latter is shown on Fig12 by the width of the trace of the standard deviation curve. Fig10c. The 10 measurements of the phase of S11 taken with the PNAX N5242A with no averaging The averaging was turned on and phase data was taken with 10 and 100 averaging factors. The data for the 10 and 100 averaging is shown on Fig13a and Fig13b.

8 Fig11a. Standard deviation from the mean for the 10 measurements of the phase of S11 taken with the PNA N5230A with no averaging Fig12a. Standard deviation from the mean for the 10 measurements of the phase of S11 taken with all instruments and plotted on 50MHz-300MHz band with no averaging. On Fig12b.the standard deviation is plotted on a 10MHz band starting from 100MHz.As for the magnitude data, it shows variation of the standard deviation on a narrow bandwidth Fig11b. Standard deviation from the mean for the 10 measurements of the phase of S11 taken with the ZVA24 with no averaging Fig12b. This plot shows how the standard deviation varies in a 10MHz bandwidth starting at 100MHz for both instruments. Fig11c. Standard deviation from the mean for the 10 measurements of the phase of S11 taken with the PNAX N5242A with no averaging Fig13a. 10 averaged S11 PHASE [deg] measurements recorded with the ZVA24 and PNA N5230A from 50MHz to 300MHz.

9 VII. CONCLUSION It seems that, although the PNA VNA showed better trace stability, both instruments are stable enough for the needed measurements. APPENDIX ACKNOWLEDGMENT Fig13b. 100 averaged S11 PHASE [deg] measurements recorded with the ZVA24 and PNA N5230A from 50MHz to 300MHz. V. SUMMARY OF MAIN RESULT VNA Agilent PNA-L N5230A S11[dB] STD DEV @100MHz S11 Phase[deg] STD DEV @100MHz 0.05 2.1 We like to thank the R&S application engineers and the agilent application engineers as well for giving us the opportunity to try their instruments. REFERENCES [1] http://www.astro.caltech.edu/~jdbowman/ [2] http://www.home.agilent.com/agilent/techsupport.jspx?pid=343726&pag emode=mn&cc=us&lc=eng [3] http://www2.rohdeschwarz.com/en/products/test_and_measurement/network_analysis/zva 24_%284-Port%29.html [4] R&S ZVA24 Agilent PNA-X N5242A 0.9 6.1 0.05 0.3 VI. DISCUSSION As mentioned earlier, using a matched load to check the stability of the vector network analyzer is an extreme case. In practical measurements, it is rare to encounter components or systems that have better than 25dB return loss or a reflection coefficient of 0.06 (6% reflection). The trace noise is due to random errors and can e reduce by: Increasing the averaging Reducing the IF bandwidth: this reduces the noise floor of the VNA's receivers and improves the signal to noise ratio. Increasing the test port power: this improves the signal to noise ratio at the input of the VNA receivers.