Tips & Tricks GPC/SEC: Inter-Detector Delay

Similar documents
Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Tips & Tricks GPC/SEC: From a Chromatogram to the Molar Mass Distribution

Tips & Tricks GPC/SEC: Protein Analysis with Size-Exclusion Chromatography

Tips & Tricks GPC/SEC: Branching Analysis

Differentiation of polymer branching and composition using the Mark Houwink plot

An Introductions to Advanced GPC Solutions

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

Characterization of polyphenylene sulphide using the Agilent PL-GPC 220 High Temperature GPC System with triple detection

Lecture 4 : Gel Permeation or Size Exclusion Chromatography

GPC/SEC An essential tool for polymer analysis

GPC / SEC Theory and Understanding

How to use GPC/SEC for compositional analysis

Polymer analysis by GPC-SEC. Technical Note. Introduction

GPC/SEC Troubleshooting Volume 1

Optimizing GPC Separations

Comprehensive Polymer Analysis Strategies

EcoSEC - semi-micro GPC/SEC system for high throughput polymer analysis

Analysis of Star Polymers Using the Agilent 1260 Infinity Multi-Detector GPC/SEC System

Gel Permeation Chromatography Basics and Beyond eseminar March 13, Jean Lane Technical and Applications Support LSCA, Columns and Supplies

GPC - Gel Permeation Chromatography. aka Size Exclusion Chromatography- SEC

Macromolecular Chemistry

Practical Steps in GPC Method Development

The power of multi-detector SEC in the analysis of antibodies

Molecular Weights of Copolymers Obtained by Gel Permeation Chromatography Light Scattering

Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection

Sem /2007. Fisika Polimer Ariadne L. Juwono

GPC/SEC standards. Product guide

GPC/SEC standards. Product guide

Latest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments

Comparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation

GPC/SEC Column Selection & Method Development

c) fitting of the NMR intensity in dependence of the recycle delays 4

Benefit of light scattering technologies (RALS/LALS/MALS) and multidetection characterization in life science research?

Size-exclusion Chromatography as a Useful Tool for the Assessment of Polymer Quality and Determination of Macromolecular Properties

Gel Permeation Chromatography

Progress toward reliable NC molecular mass distribution by GPC

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20

Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i

PAPER No. 6: PHYSICAL CHEMISTRY-II (Statistical

Gel Permeation Chromatography

MALVERN ADVANCED GPC/SEC DETECTORS SETTING THE STANDARD MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

Setting the Standard for Advanced GPC/SEC Systems

COURSE MATERIAL: Unit 3 (Part 1) Polymer Science LT8501 (Click the link Detail to download)

GPC/SEC Analysis of Polymer Solutions Used in Inkjet Printing

How switching mobile phases can improve your OMNISEC results

Waters GPC User Guide and Tutorial for Using the GPC in the Reynolds Research Group 2 nd Edition: April 2012

Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques

Polymer Analysis by Gel Permeation Chromatography

GPC/SEC Analysis of Polymer Degradation

Polymer Separations by Chemical Composition Using SEC-Gradients

Speeding up 2D chromatography (HPLCxSEC) at it s limits. ..fast, faster ups broken...

Investigating the Relationship Between the Rheological Properties of Hyaluronic Acid and its Molecular Weight and Structure using Multidetector

Maximizing Performance Through GPC Column Selection

Application compendium. Authors. Greg Saunders, Ben MacCreath Agilent Technologies, Inc. A guide to multi-detector gel permeation chromatography

GPC Streamliner. » Continues on Page 2. Focus: accurate GPC/SEC. Sieve Curves. In This Issue

How Molecular Weight and Branching of Polymers Influences Laser Sintering Techniques

Optimization of PLA/PLGA molecular weight and structure to suit final application requirements

Gel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC)

Contents. List of Contributors. Structure of the Book. Part 1 Evaluation and Estimation of Chromatographie Data 1

Physical Polymer Science Lecture notes by Prof. E. M. Woo adapting from: Textbook of Physical Polymer Science (Ed: L. H.

GPC/MS. Analysis of PET

Analysis of polyolefins by GPC/SEC

Advanced GPC. GPC On Tour, Barcelona, 28 th February The use of Advanced Detectors in GPC

Polymers Reactions and Polymers Production (3 rd cycle)

Use of SEC-MALS. (Size Exclusion Chromatography - Multi Angle. Light Scattering) for protein quality and characterization

Setting the Standard for GPC. Complete Guide for GPC / SEC / GFC Instrumentation and Detection Technologies. The Right Instrument for Your Application

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Dr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1

Clearing the Confusion: GPC, SEC, GFC What, When, Why, and How?

Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Offline Infrared Detection

Aqueous and polar GPC/SEC columns. Product guide

CALIBRATING GPC COLUMNS

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Macromolecular Chemistry

SELF-CONDENSING VINYL POLYMERIZATION: THEORETICAL ASPECTS AND APPLICATION TO GROUP TRANSFER POLYMERIZATION OF METHACRYLATES.

GPC/SEC Start-up Kit

Chapter 14. Molar Mass Distribution.

GPC-Analytics of (Bio)Polymers with Light scattering detection

MODERN HPLC FOR PRACTICING SCIENTISTS

Calibration and Normalization of MALS Detectors

OVERVIEW INTRODUCTION

The ultimate in desktop particle characterization

The ph-responsive behaviour of aqueous solutions of poly(acrylic acid) is dependent on molar mass

Size exclusion chromatography of branched polymers: Star and comb polymers

Branching Revealed: Characterizing Molecular Structure in Synthetic and Natural Polymers by Multi-Angle Light Scattering

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Gel Permeation Chromatography - GPC

Marine bio-inspired underwater contact adhesion

Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis

Supporting Information. for. Hexene. Clark R. Landis*,

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

food testing & agriculture

Puzzled? Look it up in our GPC Glossary

Evaluation of round robin test results for a size exclusion chromatography method for the analysis of nitrocellulose

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

Amended Specifications as per the Pre Bid meeting held on in Chapter 4

MALVERN OMNISEC RESOLVE, REVEAL, REALIZE MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

MALVERN OMNISEC RESOLVE, REVEAL, REALIZE MOLECULAR WEIGHT MOLECULAR SIZE MOLECULAR STRUCTURE

Transcription:

Tips & Tricks GPC/SEC: Inter-Detector Delay Daniela Held and Wolfgang Radke, PSS Polymer Standards Service GmbH, Mainz, Germany Combinations of detectors are often used in gel permeation chromatography/size-exclusion chromatography (GPC/SEC) to measure absolute molar masses or chemical compositions as a function of elution volume. Such multidetector setups require the correction of the delay volume between the detectors for correct data processing. This instalment of Tips & Tricks explains more. Photo Credit: saicle/shutterstock.com Different detectors can be applied in gel permeation chromatography/size-exclusion chromatography (GPC/SEC) depending on the analytical task to be solved. Refractive index detectors (RID) or (less often) evaporative light scattering detectors (ELSDs) are used to measure the concentration of samples lacking chromophores. Ultraviolet (UV) detectors or diode array detectors (DADs) can be applied alone to determine the concentrations of UV-active samples or in combination with RID to derive the chemical composition distribution of a sample. On-line viscometers or light scattering detectors are applied in conjunction with an RID or UV detector to calculate true molar masses or to derive structural information on the sample. The order in which the different detectors should be installed has already been covered in a previous instalment of GPC/SEC Tips & Tricks (1). However, if several detectors are used either daisy-chained or in parallel, the sample fraction leaving the column will reach the detectors at different times. The difference in time (or volume) between the detectors is commonly referred to as inter-detector delay. Figure 1 shows an example of a setup with a UV detector (first detector) and an RI detector installed in series. The tubing between the two detectors means that the sample reaches the RI detector cell (red trace) 0.15 ml (equal to 0.15 min or 7.5 s at 1 ml/min flow-rate) after it has passed the UV cell (blue trace). GPC/SEC software packages are often able to correct for this time and volume difference. This is of advantage for multidetector setups because it reduces the calibration efforts. For example, if the inter-detector delay is properly corrected for, all detector signals can be 9

Figure 1: UV (blue) and RI (red) trace. The additional tubing between UV and RI means that the sample reaches the RI detector cell 0.15 ml (7.5 s) after passing the UV cell. An inter-detector delay correction can solve this issue. evaluated using the same calibration curve, while if the inter-detector delay correction is not supported by the software, separate calibration curves need to be established for each detector. However, using separate calibration curves is not applicable if the desired results require correlating signal intensities of different detectors corresponding to the same elution volume. Typical examples are the determination of: absolute molar mass using light scattering or triple detection; structure parameters using viscometry, triple detection, or light scattering; chemical composition using dual detection; low-molecular-weight heparin according to the European Pharmacopoeia (Ph. Eur.). In cases like these, the signal intensities of different detectors corresponding to the 10

Figure : Traces of daisy-chained UV (blue) and RI (red) detector after correcting for inter-detector delay. The RI trace is slightly broader than the UV trace from band broadening because of the additional tubing and the size of the (analytical) RI cell. same GPC/SEC-slice are related to each other. A correctly determined inter-detector delay is therefore a prerequisite for such applications. Figure shows how the data from Figure 1 look after properly accounting for the inter-detector delay. The delay for the RI has been determined so that both detector traces overlay and the maxima of both signals match each other. Figure also shows that the RI trace is slightly broader than the UV trace because of band broadening resulting from the additional tubing between the detectors and from the size of the RI cell. How Can the Inter-Detector Delay Be Determined? The most accurate and precise approach to determine the inter-detector delay is using a truly monodisperse substance, which yields sufficient signal intensities in all detectors used. This is easy for the vast 11

Figure 3: Example of a correctly fine-tuned inter-detector delay in a viscometer setup. The Mark Houwink exponent is a very good parameter to verify the inter-detector delay and should be obtained correctly when measuring a standard with a broad molar mass distribution. majority of setups where only concentration detectors, such as RI and UV, are applied. In organic solvents, substances like butylated hydroxytoluene (BHT) or toluene can be used. In aqueous solvents, a protein (for example, BSA) or an amino acid might be suited to determine the inter-detector delay. If a truly monodisperse material is not available, reference materials with a narrow molar mass distribution can be used (). To determine the inter-detector delay the substance is run on the system without correction. The result should look similar to Figure 1. The volume difference between the peak maxima of detector trace 1 and trace is the inter-detector delay. For detector setups with more than two detectors, an inter-detector delay of 0 ml is usually assigned to the first detector, and positive inter-detector delays, with respect to the first detector, are given to all following detectors. Simply using the elution volume difference of the peak maxima works nicely if truly monodisperse materials are available because the detector traces are directly related to the concentration profile of the material. However, the situation is more difficult when molar mass sensitive detectors (light scattering or viscometer) are used. There are two aspects to be considered. First, monodisperse low molar mass substances, such as BHT, toluene, or an amino acid, will usually not provide sufficient signal intensities for molar mass sensitive detectors because of their low molar mass. Since the signal intensity of molar mass sensitive detectors increases with molar mass, narrowly distributed reference materials with molar masses between 50 100 KDa should be applied to measure the volume between the peak maxima. Second, certain results obtained using molar mass sensitive detectors are very sensitive to the inter-detector delay. In GPC/SEC-viscometry the determined Mark Houwink coefficient, α, is strongly influenced by an incorrect inter-detector delay (3). The sensitivity of the Mark Houwink exponent on the selected inter-detector delay can be used to fine-tune the inter-detector delay, once a first estimate is obtained by overlaying the peak maxima. To fine-tune the inter-detector delay, a broadly distributed linear standard with a known Mark Houwink exponent is run. The Mark Houwink exponent is determined from SEC-viscometry using the previously estimated inter-detector delay. If the exponent is not yet correctly determined, the inter-detector delay of the viscometer can be slightly adjusted until the correct Mark Houwink exponent α is obtained. Figure 3 shows an example of this. As described above, the inter-detector delays were first estimated using a standard with a narrow molar mass distribution. A broadly distributed polystyrene standard was then analyzed. The inter-detector delay for the viscometer was slightly varied until the resulting Mark Houwink exponent α shows the expected value of 0.714 (THF, 30 C). A similar approach is applicable in GPC/ SEC-light scattering. It is already possible to verify or correct the determined inter-detector delay by using a narrowly distributed molar mass reference standard. Physically meaningful inter-detector volumes result in a slightly decreasing molar mass with increasing elution volume. A positive example is shown in Figure 4, where the 1

Figure 4: Example of a correctly fine-tuned inter-detector delay in a light scattering detector setup. The molar mass (green) decreases slightly with the elution volume. There are no kinks or upward curvatures at the extrema. measured molar mass (green trace) shows the expected behaviour. There are no kinks and no unexpected upward curvatures at low concentrations (red trace). Samples with Broad Molar Mass Distributions in Multidetection GPC/SEC It is important to understand that the inter-detector delay should be determined with a monodisperse standard or a standard with a narrow molar mass distribution in multidetection setups. The different molar mass responses of molar mass sensitive detectors and concentration detectors result for broadly distributed samples in different detector traces. The light scattering trace and the viscosity trace usually appear to be shifted to lower elution volumes, even when correcting properly for inter-detector delay. This results from the fact that at identical concentrations, that is, identical signal 13

Figure 5: Detector signals for a sample with a broad molar mass distribution. Although the inter-detector delays are corrected, the light scattering signal (green, 90 only), viscometer delta pressure signal (black), and the RI (red) appear to be shifted. However, this is because the detectors respond differently to molar mass. The effect is fully expected for samples having a broad molar mass distribution (4). heights in the RID, the molar mass on the left-hand side of the RID peak is higher than on the right-hand side. Consequently, the light scattering signal at the left-hand side is of a higher intensity than on the right-hand side of the peak. The higher the polydispersity, the higher the effect. Figure 5 shows light scattering (green), viscometry (black), and refractive index (red) traces for a broadly distributed polystyrene sample (PDI approximately.3). The peak maximum for the light scattering trace is observed at lower elution volumes (higher molar masses) compared to the RI trace. The dependence of the viscosity signal on molar mass is not as strong as that of the light scattering signal. That is why the peak maximum of the viscosity signal is found at higher elution volumes (lower molar masses) when compared to the light scattering trace. However, both peak maxima are shifted to lower elution volumes compared to the RI-trace. The order of the detector signals agrees fully with the expectation and the effect clarifies why narrow standards should be applied when determining the inter-detector delay. Summary GPC/SEC is a powerful multidetector method. However, the time and volume delay between the detectors must be properly accounted for. Inter-detector delays can be measured using low molar mass, monodisperse substances. When molar mass sensitive detectors are used, the inter-detector delay is determined using a standard with a narrow molar mass distribution and a molar mass between 50 100 kda. When molar mass sensitive detectors are used, fine-tuning of the inter-detector delay is recommended. The narrow standard is sufficient for light scattering detectors, while viscometers often require the additional running of a standard with a broad molar mass distribution. References 1. P. Kilz and D. Held, The Column 8(18), 9 1 (01).. D. Held, J. Preis, and F. Gores, The Column 11(1), 0 3 (015). 3. P. Kilz and D. Held, in Quantification in LC and GC, S. Kromidas and J. Kuss, Eds. (Wiley, 009). 4. D. Held and P. Kilz, The Column 4(10), 17 0 (008). Daniela Held studied polymer chemistry in Mainz, Germany, and works in the PSS software and instrument department. She is also responsible for education and customer training. Wolfgang Radke studied polymer chemistry in Mainz, Germany, and Amherst, Massachusetts, USA, and is head of the PSS application development department. He is also responsible for instrument evaluation and for customized trainings. E-mail: Website: DHeld@pss-polymer.com www.pss-polymer.com 14