PFIS Camera and Collimator CDR Conceptual Design Report. March 5, 2003 J. Alan Schier

Similar documents
VACUUM SUPPORT FOR A LARGE INTERFEROMETRIC REFERENCE SURFACE

Tutorials. 1. Autocollimator. Angle Dekkor. General

Opto-Mechanical Design of Altair, the Gemini Adaptive Optics System

Athermal design of nearly incompressible bonds

Automatic Level Maintenance Manual SAL-XX W/ AIR DAMPENED COMPENSATOR

HERSCHEL/UVCI ALIGNMENT PLAN

New concept of a 3D-probing system for micro-components

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Instrument Acceptance Testing Plan

Use of computer generated holograms for alignment of complex null correctors

GP-B Telescope Measurement of Telescope Transmission P0450 Rev -

Physics 476LW Advanced Physics Laboratory Michelson Interferometer

Magnet Buildability Review report : C-Beta Magnets

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements

Picometre metrology. The Gaia mission will create an ultra-precise three-dimensional map of about one billion stars

Metrology Prof. Dr Kanakuppi Sadashivappa Bapuji Institute of Engineering and Technology Davangere

arxiv:astro-ph/ v1 28 Oct 2002

ilearnalignment Alignment Training Subjects

A Compliant Optical Mount Design

Table of Contents. iii

2.4m Space Telescopes

Mounts for large lens in cooled environment

Compact, lightweight, and athermalized mirror and mount for use on JWST's NIRCam instrument

Development of surface metrology for the Giant Magellan Telescope primary mirror

ABB temperature measurement Radiation thermometry. Measurement made easy. Process temperature measurement practice--non-contacting

Coating Requirements for Pipelines Installed by Horizontal Directional Drilling John D. Hair, P.E.*

Galileo Telescope Solar Viewer Joseph Hora, Elizabeth Hora 2017/09/18

AME Model. (1) Rated flow (l/min (ANR)) Port size (Nominal size B) Weight (kg)

Fall 2014 Mott Target Replacement and Ladder Position Re- Calibration JLAB- TN Joe Grames. September 17, 2014

TE 75R RESEARCH RUBBER FRICTION TEST MACHINE

Hanle Echelle Spectrograph (HESP)

Operating Instructions Spectro-Goniometer Student. 1 Functional Elements. 2 Safety Precautions. Figure 1: Spectro-Goniometer Student

OWL: Further steps in designing the telescope mechanical structure and in assessing its performance

Positive Signal The most technologically advanced. double coated dichroic filters are used, which transmit

SIR C.R.REDDY COLLEGE OF ENGINEERING ELURU

Universal type Port size AS13 1F-M3. M3 x 0.5 M5 x 0.8 AS23 1F-01 AS23 1F-02 AS33 1F-02 AS33 1F-03 AS43 1F-04 R 1/8 R 1/4 R 1/4 R 3/8 R 1/2

ANAMORPHOTIC TELESCOPE FOR EARTH OBSERVATION IN THE MID- INFRARED RANGE

Atmospheric dispersion correction for the Subaru AO system

ITTC Recommended Procedures and Guidelines

Auto collimator. Introduction. Objectives: Apparatus: Theory:

Newtonian 17.5-inch Optical Tube Assembly

JDV CONTROL VALVES CO., LTD. JCT/JCTM Series 3/4 Way Trunnion Type Valve Manual

The Distributed Defining System for the Primary Mirrors

DEVELOPMENT OF MEASURING SYSTEM FOR STRESS BY MEANS OF IMAGE PLATE FOR LABORATORY X-RAY EXPERIMENT

The IPIE Adaptive Optical System Application For LEO Observations

saxon Instruction Manual saxon Grandeur Brass Telescope High quality optics

A deeper investigation into the thermal drift of a linear axis

Important Instructions

Collimation Instructions Learn how to accurately collimate the optics in your telescope.

Foundations of Ultraprecision Mechanism Design

Where s the Blur? By Eric Chesak

Telescope Mechanical Design

APPENDIX G I-BEAM SUMMARIES 0.6-IN. STRAND G-1

Assembly Manual for the Brevard Astronomical Society 16 inch F4.5 Dobsonian Telescope Brevard Astronomical Society P.O. Box 1084 Cocoa, FL 32922

FOUR-POINT CONTACT SLEWING RINGS - without gear [ O ]

BINOCULAR STEREOSCOPIC MICROSCOPE MBS-10. User Manual

Development of a probe traversing system for an open test section wind tunnel

THE LOSMANDY G-11 MOUNT

Controlling Thermal Expansion

Athermal bond thickness for axisymmetric optical elements

GP-B Telescope Determine Metering Tube Length P0364 Rev -

Calibrating the Thermal Camera

Optics. Measuring the line spectra of inert gases and metal vapors using a prism spectrometer. LD Physics Leaflets P

MELOS 530 MEASURING EQUIPMENT FOR LENSES AND OPTICAL SYSTEMS. Combinations and Accessories. Indispensable for Quality Control of Optics

Lab Exercise #5: Tension and Bending with Strain Gages

GA A22722 CENTRAL THOMSON SCATTERING UPGRADE ON DIII D

Papers for the NAOMI Opto-Mechanical Chassis Critical Design Review

ME411 Engineering Measurement & Instrumentation. Winter 2017 Lecture 9

ALIGNMENT OF THE MSGC BARREL SUPPORT STRUCTURE

Technical description of photoelectron spectrometer Escalab 250Xi

RoHS. Specification CUD8AF1C. 서식 Rev: 00

Precision Ball Screw/Spline

MEMS Metrology. Prof. Tianhong Cui ME 8254

Strain Measurement. Prof. Yu Qiao. Department of Structural Engineering, UCSD. Strain Measurement

PRIMARY ARMS 3X COMPACT SCOPE WITH PATENTED.223/5.56, 5.45X39,.308 ACSS RETICLE

Strain Measurement Techniques for Composite Coupon Testing

Implementation of velocity measurement as intensity verification in production

TABLE OF CONTENT Scope 2 Alignment Systems and Procedures 2 Balancing 8 Lubrication System 10 Preservation of Spare Parts 12

H A R D R O C K T B M D E S I G N F E AT U R E S F O R D I F F I C U LT G R O U N D C O N D I T I O N S. Tyler Sandell The Robbins Company

LAB 10: OPTICAL MATERIALS AND DISPERSION I

PICTURE sounding rocket telescope

Optimization and Tolerance Scheme for a Mirror Mount Design Based on Optomechanical Performance

CRYOGENIC CONDUCTION COOLING TEST OF REMOVABLE PANEL MOCK-UP FOR ITER CRYOSTAT THERMAL SHIELD

The LAT Electronics consists of five distinct box assemblies as follows:

MODEL NAME : LLDMWW0-15K*0*A

The Alignment Concept For The Fermilab Recycler Ring Using Permanent Magnets

Project PAJ2 Dynamic Performance of Adhesively Bonded Joints. Report No. 3 August Proposed Draft for the Revision of ISO

TIR100-2 Measurement of Thermal Emissivity

Heating Beam Pattern Optical Design CO2 Laser Thermal Compensation Bench

5. STRESS CONCENTRATIONS. and strains in shafts apply only to solid and hollow circular shafts while they are in the

A 954 C HD. Technical Description Hydraulic Excavator. Machine for Industrial Applications

Determination of Cauchy s Contants

Purpose of this Guide: To thoroughly prepare students for the exact types of problems that will be on Exam 3.

INDEX OF ILLUSTRATIONS

Primary Mirror Cell Deformation and Its Effect on Mirror Figure Assuming a Six-zone Axial Defining System

APPENDIX 4.4.A STRAWMAN STRUCTURAL DESIGN OF A 30-M GSMT

Coma Correction. Diagram. Introduction

Developing a Compact Doppler/Magnetograph for Solar Diagnostics

Thermal Coatings for In-vacuum Radiation Cooling LIGO-T C R. Abbott, S. Waldman, Caltech 12 March, 2007

Seminar BELA STAR SIMULATOR

GEOMETRICAL OPTICS Practical 1. Part II. OPTICAL SYSTEMS

Transcription:

PFIS Camera and Collimator CDR Conceptual Design Report March 5, 2003 J. Alan Schier Contents I. Summary of Design Features II. Summary of Quantitative Results Mounting Tolerances Element Stresses Optical Element Gravity and Hydrostatic Deflections Thermal Behavior Weights Fluid Volumes Structural Deflection III. Alignment and Positioning IV. Shop Tests V. Technical Risks VI. Environmental Risk Mitigation VII. Open Issues I. Summary of Design Features Figures 1a and 1b show the concept layout of the PFIS camera and collimator. The layouts are shown without the eventual turning flat. Referring to figure 1a, the main elements are: --The right-most barrel containing the camera elements. --The left-most subassembly containing the first group of the collimator. --The second-to left subassembly comprising the main collimator barrel. --The remaining subassembly consisting of the last collimator group on its own stand. (N.B. Figures 1 through 4 show various 3D and 2D views of the Collimator and Camera conceptual designs. In those views, the 3D models are more representative of the final configuration with regard to overall sizes and envelopes. The 2D views are more accurate with regard to the particular details of the assemblies and subassemblies. Since the designs are largely axisymmetric, often only one half of the components is shown.) Page 1 of 25

The main features of the design are as follows. Elastomeric Bonding: All of the optical elements are bonded into their cells with and elastomer. For this design, the elastomer is Dow-Corning Sylgard 184. For the close coupled groups, a bezel is bonded to subsequent elements, and the bezel is then bolted to the cell. This is shown in Figure 3 where the orange ring is a bezel around element 8. This ring is bolted to the turquoise cell in which element 9 is bonded. Fluid-Coupled Multiplets: For this design, the multiplets are optically coupled with Cargille LL5610. Each fluid-containing cell is sealed with single Viton o- ring in face-flange configuration. The change of fluid volume with temperature is accommodated by a bladder external to the cell. An example bladder is shown in Figure 5. Most Element Spacing Set By Shims and Spacers: In all of the multiplets, the spacing for the elements is set when the elements are bonded in place. This is done by placing appropriate polyester shims between the elements to maintain the proper separation until the elastomeric bond is in place. The cell is then disassembled and the shims are removed before the cell's final assembly. For the groups assembled in the camera and collimator barrels, the separations are set by means of spacer rings that are cut to length during assembly. This permits compensation of the manufacturing tolerances in the cells by measuring the asassembled position of the optics within their cells. The spacer is then cut to account for the measured positions. The separation between the first collimator group and the main barrel is set by means of a gage bar that is cut to length at assembly. Due to the presence of the waveplates, a permanent spacer cannot be used. Rather, by inserting a gage bar between datum surfaces on the first and second cells, the position of the first cell can be properly adjusted. Deflections of Elastomeric Bonds Managed by Means of Engineered Bond Sizes: It is possible to limit the deflection of the elastomeric bonds that retain the optics to acceptably small values. These deflections are due to the weight of the elements and in the case of the multiplets, the hydrostatic pressure of the coupling fluid. In the radial direction, the deflections can be limited to 1.7 microns for any orientation. In the axial direction, the deflections are 14 ±2.5 microns, with the exception of elements 1 and 10. For these elements, weight-induced axial motions are 6.2 and 6.5 microns respectively. The tilt of an element caused by the gradient in the hydrostatic pressure is no more than 0.01 degrees (0.6 arcminutes). Page 2 of 25

Adjustable Groups: In the collimator, the first and last groups (groups 1 and 5) each have adjustments in 5 degrees of freedom. Figures 2b and 3b schematically indicate these adjustments with adjusting screws. The tip, tilt, and piston adjustments are accomplished with three screws that bear kinematically in the axial direction. The centering adjustments are accomplished with two screws that bear in the radial direction and cells that are flexured for motion in that direction. The layout of such a cell is shown in Figure 7. Flexured Focus Suspension: The system is focused by moving elements 14 through 17 in the camera as a group. To achieve this, these elements are supported by a diaphragm flexure at each end of the subassembly. This is best seen in Figures 4a and 4b. Low Element Stresses: The maximum stress in an optical element over the operating range of 5 to 20 C is under 50 psi. Over the survival range of 20 to 40 C, the maximum stress is under 100 psi. To achieve these levels, the elements must be assembled at 10 C (50 F). Passive Thermal Compensation: The design includes thermal compensation in the axial direction for elements in the collimator barrel, and for focus in the camera. For the collimator, this can be seen in Figure 2a as the two black rings that form parts of the barrel structure. These rings are made of Delrin and are the length necessary to compensate for other changes with temperature within the collimator. Focus Mechanism: Figure 9 shows the focus mechanism. It consists of a passive Delrin thermal compensator, a Physik Instrumente M230 stepper micrometer, a lever reducing mechanism, and a Schaevitz LVDT model 100DCSE. The arrangement is suitable for positioning and measurement with a resolution of less than 1 micron. Active Thermal Management: To work properly, the Delrin thermal compensators in the collimator barrel must be within about 2 C of each other. As an inexpensive precaution to insure that this is the case, the design provides thermocouple sensors and low-wattage foil heaters at each of the collimator's Delrin compensators. Should one compensator be significantly cooler than the other, the cooler one could be warmed. Primary Materials: Black anodized aluminum structure. Delrin thermal compensators. Delrin optics seats. Stainless steel focus flexure. ¼-80 adjustment screws. Cargille LL5610 optical coupling fluid. Dow-Corning Sylgard 184 bonding elastomer. Page 3 of 25

Ether-based polyurethane bladders. Stainless hardware. II. Summary of Quantitative Results Mounting Tolerances: In general, the allowable tolerance for the tilt of an element is ±0.032 (2 arcminutes), and for the positioning, ±0.075 mm (0.003") relative to the preceding element. In two exceptional cases, the tolerances are tighter. In the first case, lens 3 must be constrained to a tilt and decenter of less than ±0.016 and ±0.050 mm. In the second case, group 4 (lenses 5 through 7) must be centered to within ±0.050 mm. Element Stresses: The thermally induced stresses are tabulated below. Element No. Stress For 15 C Excursion (psi) (Operating Range) Stress For 30 C Excursion (psi) (Survival Range) 1 6.1 12.6 2 44 88 3 4.2 8.4 4 3.4 6.8 5 45 90 6 43 86 7 34 68 8 0.1 0.2 9 44 88 10-20 -40 11 44 88 12-20 -40 13 49 98 14 47 94 15-41 -82 16 46 92 17-11 -22 Note: Negative values indicate a tensile stress with rising temperature. Optical Element Gravity and Hydrostatic Deflections The optical elements will move in their elastomeric mounting depending on the forces exerted by gravity and the hydrostatic pressure of the coupling fluid within the cells. The motions due to gravity are an axial piston or a radial decenter, depending on the orientation of the element. The motions due to hydrostatic pressure are an Page 4 of 25

axial piston and a tilt induced by the gradient in the pressure. Note that the hydrostatic deflections occur only when the element is vertical. The table below lists the various deflections. For the elements up to and through the corrector barrel (elements 1 through 7), axial sag due to gravity is the dominant deflection during operation since the gravity orientation is confined to a small range. For the remaining elements, gravity may range over practically any direction during operation. For testing, all of the elements will be oriented vertically so that the hydrostatic motions and the radial decenter will be the relevant deflections. Element No. Axial Sag, Gravity (µm) Radial Sag, Gravity (µm) Axial Motion, Hydrostatic (µm) Element Tilt, Hydrostatic (degrees) 1 6.2 0.3 9.7 0.009 2 16.1 0.9-8.8-0.007 3 13.0 1.0 4 11.6 1.1 5 14.2 0.3 11.8 0.007 6 11.7 0.6 7 12.7 0.4-16.1-0.009 8 10.7 0.3 16.7 0.010 9 13.6 0.3-12.7-0.007 10 6.5 0.1 15.4 0.007 11 14.7 0.2 12 12.2 0.2 13 15.3 0.3-17.3-0.008 14 16.1 0.3 15 15.5 0.2 12.4 0.007 16 15.9 0.8 17 12.6 0.1-6.3-0.004 Thermal Behavior Estimates of various thermal time constants follow. The time constants assume an ambient temperature of 295 K, and a component temperature of 305 K. Collimator Barrel Aluminum Portion Convective Time Constant Horizontal...1 h Vertical...0.9 h Radiative Time Constant...0.4 h Internal Conduction Page 5 of 25

Circumferential...260 s Longitudinal...14 s Delrin Portion Convective Time constant Horizontal...2.3 h Vertical...1.9 h Radiative Time Constant...0.32 h Internal Conduction Circumferential...31 h Longitudinal...1.4 h Through thickness...113 s Camera Barrel (Aluminum) Convective Time Constant Horizontal...2.6 h Vertical...2.1 h Radiative Time Constant...4.4 h Internal Conduction Circumferential...342 s Longitudinal...433 s Camera Focus Compensator (Delrin) Convective Time Constant Horizontal...1.3 h Vertical...1.3 h Radiative Time Constant...0.1 h Internal Conduction Circumferential...2.7 h Longitudinal...1.4 h Through thickness...113 s Calcium Fluoride Slab (example optical element, 20 mm thickness, 300 mm diameter) Internal conduction, through thickness...11 s Convection, vertical...5.0 h With the exception of the internal conduction values, all of these time constants are rough approximations. The radiative time constants will be considerably longer since as the temperatures equilibrate, the heat transfer decreases. The convective time constants will effectively be shorter since the values above are for truly free convection in otherwise still air. The slightest breeze can easily reduce a time constant by a factor of 2. The important features are: Page 6 of 25

--Every component has an internal conduction time constant that is short compared to any convective or radiative time constant. This means the component will have only small internal gradients as temperatures equilibrate. --The convective and radiative time constants of the structural components are of similar scale. This means the temperatures of the components will equilibrate at similar rates. --The time constant for an optical element can be considerably longer than that for the structural components. This means that the temperature of the optics can be considerably different from that of the structure if exposed to a sudden temperature change. Fortunately, even if we assume a rather extreme case in which the time constant for the structure is 1/2 hour and that for the optic is 5 hours, the effect is benign. For a 20 C mismatch between an unfavorable element (element 13) and the structure, the induced stresses are still below 300 psi. This is due the relatively soft bonding material that largely isolates the element from its cell. The temperature behavior resulting from a 1 degree step change in ambient is shown in Figure 8. Masses All masses in kilograms. Collimator: Front Cell...3.8 Lens 1...0.55 Lens 2...1.38 Subtotal...5.73 kg Main Barrel...10.5 Lens 3...0.99 Lens 4...0.86 Lens 5...3.0 Lens 6...1.55 Lens 7...2.9 Subtotal...19.8 kg Last Cell, incl. support...8.1 Lens 8...2.1 Lens 9...3.5 Subtotal...5.5 kg Camera Structure...19.2 Lens 10...2.3 Page 7 of 25

Lens 11...6.1 Lens 12...4.2 Lens 13...5.0 Lens 14...4.6 Lens 15...2.6 Lens 16...2.3 Lens 17...3.8 Subtotal...50.1 kg Total...81.1 kg Fluid Volumes (Fluid volumes will be calculated when cell details are finalized.) Structural Deflections The largest bending and shear deflections occur in the outer Delrin compensator in the collimator. For this element the total deflection is less than 5 microns, and the resulting angle is less than 5 arcseconds. III. Alignment and Positioning The general approach to alignment of the optics is to first accurately mount the elements in their cells, and then taking into account the measured cell dimensions, align the cells in their respective assemblies. The details are as follows: 1. Alignment of Elements in Cells: a). Fabricate accurate cells. Accurate means that there is a Delrin seat in the cell that is concentric to a datum diameter on the outer surface of the cell to within +/-10 µm, and parallel to datum face to +/-10 µm. b). Fixture the cell on a rotary table so that the Delrin seat is concentric to the axis of the table. c). Place the optical element on seat. Since the seat is centered, the optical surface resting on the seat is also centered. d). Tilt the element to align the upper vertex with the axis of the rotary table. This is done by observing a reflection from the upper surface with a long-working distance microscope. (Sensitive to about 0.2 arcseconds). e). Bond the element in place. f). For multiplets, repeat the procedure by setting next element in place on spacing shims, aligning as above, and then bonding the element into its bezel. Page 8 of 25

2. Alignment of Cells in Assemblies: For the camera barrel: Spacings are set with spacers that are cut to length at assembly. The asbuilt dimensions of the cell are measured on a milling machine with suitable metrology. Contact measurements can be made on hard-coated elements. Measurements with a video camera and microscope objective are made on surfaces with sol-gel coatings. (Accurate to about ±4 microns over 10", repeatable to ±2 microns.) Axial alignment of the cells in the assembly is again done on a rotary table with long working distance microscope, similar to the alignment of the optics in their cells. For the collimator: Alignment of the collimator is complicated by the fact that it mounts in three sections on the optical bench: field lens, main barrel, and last group. The following is one possible alignment procedure: a). The main barrel is first assembled in a manner similar to camera barrel, as outlined above. b). Mount the main barrel in a fixed (nominally final) position on the optical bench. c). Mount the alignment telescope (K&E 2202) ahead of field lens and cell. d). Align the telescope with the axis formed by front and rear vertices of main barrel. e). Install the last group and align the front and rear vertices with the telescope axis (tilt and decenter). f). Install field lens and align front and rear vertices with telescope g). Set field lens spacing with gage bar. h). Set last group spacing in autocollimation: Place a mirrored flat behind last group and align with the telescope. Place pinhole and camera (or camera with microscope objective) at the focal surface. Move the last group axially until return spot is in sharp focus at the focal plane. IV. Shop Tests We propose that acceptance testing consist of verification of the correct back focal distance and spot size using two colors for the camera and collimator. Testing will be accomplished by introducing a collimated beam and measuring the spot size and location with a measuring microscope. The camera will be tested in the forward direction; the collimator will be tested in reverse. V. Technical Risks Page 9 of 25

Compared to other systems we have produced, the primary technical risk in the PFIS system is damage to the sol-gel coatings and damage to the NaCl optical elements. These risks are addressed in section VI below. VI. Environmental Risk Mitigation The PFIS system will be assembled in an environmentally controlled pseudoclean room. This in not intended to be a true clean room by industrial standards. Rather, it is a room where dust is kept to a practical minimum through use of smooth surfaces, air filters, and good personal practice. The room will also be dehumidified and temperature controlled. Additionally, the system design includes features to minimize the risk of damage during handling and operation. This includes guards on the cells either integral or removable that prevent the optical element from making contact when the cell is set on a flat surface. Covers will also be provided for all exposed surfaces when not in use. The camera and collimator barrels will include ports for the introduction of a filtered dry-air purge. VII. Open Issues The selection of the coupling fluid, bonding compound, and bladder material has not been finalized. The materials assumed for the conceptual design (Cargille LL5610, Sylgard 184, and ether-based polyurethane) are know to be compatible. We do not know if LL5610 is acceptable optically. Should another coupling fluid be selected, it may be necessary to change other materials. Page 10 of 25

Figure 1a 3D PFIS Camera and Collimator General Layout From left to right, the subassemblies are: Collimator Group 1, Collimator Main Barrel, Collimator Group 5, Camera Barrel Page 11 of 25

Figure 1b 2D PFIS Camera and Collimator General Layout Page 12 of 25

Figure 2a 3D Sectioned Layout of Collimator Group 1 and Main Barrel Page 13 of 25

Figure 2b 2D Sectioned Layout of Collimator Group 1 and Main Barrel Page 15 of 25

Figure 3a 3D Sectioned Layout of Collimator Group 5 Page 16 of 25

Figure 3b 2D Sectioned Layout of Collimator Group 5 Page 18 of 25

Figure 4a 3D Sectioned View of Camera Layout Page 19 of 25

Figure 4b 2D Sectioned View of Camera Layout Page 20 of 25

Figure 5 Photo of Example Bladder Page 21 of 25

Figure 6 Example Flexure Model and Photograph Page 22 of 25

Figure 7 Page 23 of 25

Figure 8 Thermal Behavior for A Step Change in Ambient Temperature Page 24 of 25

Figure 9 Focus Mechanism Page 25 of 25